K Number
K251798

Validate with FDA (Live)

Device Name
RCT700
Manufacturer
Date Cleared
2025-11-25

(166 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified heath care professionals

Device Description

RCT700 provides 3D computed tomography for scanning hard tissues such as bone and teeth. By rotating the C-arm, which houses a high-voltage generator, an X-ray tube and a detector on each end, CBCT images of dental maxillofacial structures are obtained by recombining data scanned from the same level at different angles. Functionalities include panoramic image scanning for obtaining images of whole teeth, and a cephalometric option for obtaining cephalometric images.
The software of RCT700 saves the patient and image data and offers an inquiry function, in addition, supports the image generate function intended to obtain images using the RCT700 equipment and various sensors for diagnosis.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - RCT700

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

November 25, 2025

RAY Co., Ltd.
℅ Suyeon Baek
RA Team
1F~3F, 4F(Part), 5F, 265, Daeji-Ro, Suji-gu
Yongin-si, Gyeonggi-do 16882
REPUBLIC OF KOREA

Re: K251798
Trade/Device Name: RCT700
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: OAS
Dated: July 1, 2025
Received: October 30, 2025

Dear Suyeon Baek:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251798 - Suyeon Baek Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251798 - Suyeon Baek Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K251798

Device Name: RCT700

Indications for Use (Describe):

RCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified heath care professionals

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Submission- RCT700

510(k) Summary

1. 510(k) Summary

The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

2. Date

July 4, 2025

3. Administrative Information

ApplicantRay Co., Ltd.
Address1F~3F, 4F(Part), 5F, 265, Daeji-ro, Suji-gu, Yongin-si, 16882, Korea
Manufacturer
NameRay Co., Ltd.
Address1F~3F, 4F(Part), 5F, 265, Daeji-ro, Suji-gu, Yongin-si, 16882, Korea
Tel+82-31-605-1000
Fax+82-2-6280-5534
Contact Person
NameSuyeon Baek
Emailsuyeon.baek@raymedical.com

4. Device Information

Trade/Proprietary NameRCT700
Common NameDental Panoramic/Tomography and Cephalometric X-ray System
Classification NameComputed tomography x-ray system
Regulation Number21 CFR 892.1750
Class2
Product CodeOAS
Review PanelRadiology

K251798

Page 6

510(k) Submission- RCT700

5. Predicate device

ParameterPredicate Device
Device NameRCT700
ManufacturerRay Co., Ltd
510(K) NumberK213226
Classification nameComputed tomography x-ray system
Regulation number892.1750
Primary product codeOAS

6. Device Description

RCT700 provides 3D computed tomography for scanning hard tissues such as bone and teeth. By rotating the C-arm, which houses a high-voltage generator, an X-ray tube and a detector on each end, CBCT images of dental maxillofacial structures are obtained by recombining data scanned from the same level at different angles. Functionalities include panoramic image scanning for obtaining images of whole teeth, and a cephalometric option for obtaining cephalometric images.

7. Indications for Use

RCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified heath care professionals

8. Patient population

The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure.

There is no restriction for ethnic group, Gender, age, weight, health, or condition.

We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.

Page 7

510(k) Submission- RCT700

9. Comparison with predicate device

The following table provides a summary of the technological characteristics of RCT700 compared to the predicate device.

ParameterSubject DevicePredicate Device
ManufacturerRAY Co., Ltd.RAY Co., Ltd.
Device nameRCT700RCT700
510(K) NumberTraditional 510KK213226 (Traditional 510K)
Common NameDental panoramic/tomography and cephalometric x-ray systemDental panoramic/tomography and cephalometric x-ray system
Indications for useSame as predicate deviceRCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals
Mode of OperationSame as predicate deviceContinuous operation with intermittent, stated permissible loading
3D technologySame as predicate deviceCBCT Cone beam Computed Tomography
Performance SpecificationSame as predicate device1) CBCT Computed tomography - Patient 2) Panoramic 3) Cephalometric(optional) - One shot type - Scan type
Functional OptionSame as predicate deviceBase CT+PANO Option(CEPH) CT + PANO + SCAN CEPH CT + PANO + One shot(One shot, Standard Type) CT + PANO + One shot(One shot, Large Type).
Detector Type CTSame as predicate deviceFXDD-0606CA
Same as predicate deviceJupi0606X1

Page 8

510(k) Submission- RCT700

ParameterSubject DevicePredicate Device
PANOSame as predicate deviceFXDD-0606CA
Same as predicate deviceJupi0606X1
Ceph (Scan)Same as predicate deviceXID-C24DC
Pluto0900XN/A
Ceph (One shot)Same as predicate deviceFXDD-1012CA
Same as predicate deviceFXRD-1717VA
Exposure switch TypeSame as predicate device"Deadman" Button type
Main ComponentsSame as predicate deviceCeph Apparatus
Same as predicate deviceVertical Carriage
Same as predicate deviceRotator
Same as predicate deviceX-RAY Generator
Same as predicate deviceX-ray tube
Same as predicate deviceHigh Frequency Generator
Same as predicate deviceColumn
Control panelTouch monitor (panel)

Detector

  • CT
    • FXDD-0606CA
    • Jupi0606X1
  • PANO
    • FXDD-0606CA
    • Jupi0606X1
  • Ceph
    • Pluto0900X(Scan)
    • XID-C24DC(Scan)
    • FXDD-1012CA(One shot, Standard Size)
    • FXRD-1717VA(One shot, Large Size)

Detector

  • CT
    • FXDD-0606CA
    • Jupi0606X1
  • PANO
    • FXDD-0606CA
    • Jupi0606X1
  • Ceph
    • XID-C24DC(Scan)
    • FXDD-1012CA(One shot, Standard Size)
    • FXRD-1717VA(One shot, Large Size)

| | Same as predicate device | Chinrest |
| | Same as predicate device | Head rest |
| | Same as predicate device | Automatic Collimator |
| | Same as predicate device | Exposure switch |
| | Same as predicate device | Emergency stop switch |
| | Same as predicate device | Console PC set |
| Automatic Collimator | Same as predicate device | CT exams Panoramic exams Cephalometric exams |

Page 9

510(k) Submission- RCT700

ParameterSubject DevicePredicate Device
ClassSame as predicate deviceClass I with type B applied parts according to IEC 60601-1
Focal sizeSame as predicate device0.5
Field of View(CT)Same as predicate deviceMax.160x100 mm
X-ray VoltageSame as predicate device60~100kVp
X-ray CurrentSame as predicate device1~17mA
Total FiltrationSame as predicate deviceMin. 2.8 ㎜ Al equivalent

Detector Pixel size

| CT | Same as predicate device | FXDD-0606CA: 119㎛ |
| | Same as predicate device | Jupi0606X1 : 100㎛ |
| PANO | Same as predicate device | FXDD-0606CA: 119㎛ |
| | Same as predicate device | Jupi0606X1 : 100㎛ |
| Ceph (Scan) | Same as predicate device | XID-C24DC: 100㎛ |
| | Pluto0900X: 100㎛ | N/A |
| Ceph(One shot) | Same as predicate device | FXDD-1012CA : 124㎛ |
| | Same as predicate device | FXRD-1717VA : 140㎛ |

| Format compatible | Same as predicate device | DICOM 3.0 Format compatible |
| Image Viewing Software | RayScan(version 2.5.0.0) | RayScan(version 2.4.0.0) |
| Image acquisition | Same as predicate device | Giga-Ethernet Network |
| Total Height | Same as predicate device | Max 2,296mm |
| Weight | Same as predicate device | 1) Computed Tomography(CT) + Panoramic(PANO)=185kg(407.9lb) ± 10% 2) Computed Tomography(CT) + Panoramic(PANO) + Ceph (Scan type)= 212.5kg (468.5lb) ± 10% 3) Computed Tomography(CT) + Panoramic(PANO) + Ceph (One shot type, installed in Standard size)= 211kg (465.2lb) ± 10% 4) Computed Tomography(CT) + Panoramic(PANO) + Ceph (One shot type, installed in Large size) 211kg (465.2lb) ± 10% |
| Type of installation | Same as predicate device | Wall or floor mount |
| Patient position | Same as predicate device | Standing / Wheelchair |
| Applicable Standards | Same as predicate device | IEC 60601-1 IEC 60601-1-3 IEC 60601-2-63 IEC 60601-1-2 |

Page 10

510(k) Submission- RCT700

The product is principally just the same as in the previous 510(k) #K213226.

The table 1 provides the summary of the technological characteristics of RCT700 compared to the predicate device.

The complete of differences of the subject device to the predicate device is as follows:

  • Detector (Added a detector of scan ceph type)
  • Software version upgrade

The software of RCT700 saves the patient and image data and offers an inquiry function, in addition, supports the image generate function intended to obtain images using the RCT700 equipment and various sensors for diagnosis. And the structure and function of the software is the same as that of the predicate device(K213226).

The 510(k) for the existing detector used in our equipment is provided below.

ModalityManufacturerDetector ModelCleared510(k) No.
CTIrayJupi0606X1No PMAK213226
CTVieworksFXDD-0606CANo PMAK181452
PanoVieworksFXDD-0606CANo PMAK181452
PanoIrayJupi0606X1No PMAK213226
Scan Cephi3systemXID-C24DCNo PMAK181452
Scan CephIrayPluto0900XNo PMAK211159
One shot CephVieworksFXRD-1717VANo PMAK181003
One shot CephVieworksFXDD-1012CANo PMAK182805

Page 11

510(k) Submission- RCT700

10. Safety and Effectiveness Information

RCT700 system described in this 510(k) is similar to the predicate device in terms of indications for use, materials, safety characteristics, and X-ray source.

The following information further substantiates the substantial equivalence between the subject device and predicate device. The fundamental technological characteristics of the subject and predicate device are similar. The imaging modes are similar; PANO, CEPH (Optional), CBCT. The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the performance of the subject and the predicate device for comparison. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.

The complete of differences of the subject device to the predicate device is as follows:

  • Detector (Added a detector of scan ceph type)
  • Software version upgrade

Electrical, mechanical and environmental safety testing according to the standard of IEC 60601-1:2005/AMD1:2020 (3.2 Edition), IEC 60601-1-3:2008/AMD1:2013/AMD2:2021 (2.2 Edition), IEC 60601-1-6:2010/AMD1:2013/AMD2:2020 (3.2 Edition), and IEC 60601-2-63:2012/AMD1:2017/AMD2:2021 (1.2 Edition) were performed.

EMC testing was conducted in accordance with the standard IEC 60601-1-2:2014+AMD1:2020 (Edition 4.1).

The software of RCT700 saves patient and image data and offers an inquiry function. In addition, it supports the image generate function intended to obtain images using the RCT700 equipment and various sensors for diagnosis. That has been validated according to the FDA Guidance for the "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" to assure substantial equivalence. The software for this device was considered a "Basic Documentation Level" of concern since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.

As a result, we identified the level of concern associated with a new device and provided documentation consistent with that level. Based on our risk analysis of software, the difference does not affect its safety and effectiveness.

Bench testing was conducted according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices". Bench testing is used to assess whether the parameters required to describe functionalities related to imaging properties of the dental X-ray device and patient dosage satisfy the designated tolerance.

Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4 and IEC 61223-3-7. All test results were satisfactory.

Non-clinical considerations were conducted in accordance with FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices". Because the subject device used the same detector as the predicate device, there is no significant difference between the two devices as a result of non-clinical testing.

Clinical considerations were conducted according to the FDA Guidance "Electronic Submission Template for Medical Device 510(k) Submissions: Guidance for Industry and Food and Drug Administration Staff," Section "Performance Testing – Clinical" Clinical

Page 12

510(k) Submission- RCT700

images were provided, and they provide further evidence, in addition to the laboratory performance data, to show that the complete system works as intended.

The features of RCT700 were clinically tested and approved by one licensed practitioners/clinicians. Clinical imaging samples were collected from new detectors on the proposed device at the two offices where the predicate device was installed for the clinical test images. These images were gathered from all detectors installed with RCT700 using protocols with random patient age, gender, and size. A licensed practitioner reviewed the sample clinical images and deemed them to be of acceptable quality for the intended use.

11. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Ray Co., Ltd. concludes that the new RCT700 is safe, effective and substantially equivalent to the predicate device as described herein.

N/A