K Number
K251283
Device Name
Once-Fil Flow
Manufacturer
Date Cleared
2025-06-06

(42 days)

Product Code
Regulation Number
872.3690
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
- Temporary inlay and onlay treatments of the cavity. - Sealing of openings for implant screws. - Relining material for temporary crowns and bridges. - Covering of the gingival margin - Fixing of resin matrix during filling placement. - Block-out material for retentive ares in the dental arch, e.g. before taking impressions.
Device Description
Once-Fil Flow is a light-curing filling material designed for temporary restorations. It serves as an interim solution before the placement of a final prosthesis, providing functions such as pulp protection and maintaining tooth position. The material is suitable for temporary fillings in prepared cavities and inlay/onlay cases. To accommodate different clinical needs, Once-Fil Flow is available in two shades. The Yellow shade is ideal for anterior teeth, where esthetics are a priority, while the Blue shade provides clear visual contrast, making it easier to identify and remove before permanent restoration. The application process involves preparing the tooth surface, applying an appropriate amount of material, and light-curing it. Once cured, the material can be finished and polished using conventional instruments.
More Information

No
The device description and performance tests indicate it is a light-curing dental filling material and does not mention any AI, DNN, or ML components.

Yes
The device is a temporary dental filling material designed for therapeutic purposes such as pulp protection and maintaining tooth position, which are direct treatments for dental conditions.

No

The device description indicates it is a light-curing filling material for temporary restorations, and its indicated uses are primarily for filling, sealing, relining, and covering, none of which involve diagnosing a medical condition.

No

The device is a light-curing filling material, which is a physical substance used for dental restorations. The description mentions physical properties, shades, application processes involving material, and light-curing, along with performance tests related to material characteristics (e.g., Shore D Hardness, Water Sorption/Solubility). These are all indicators of a hardware/material-based medical device, not a software product.

No.
The device is a restorative dental material for temporary use, applied directly to the patient's teeth. It is not intended for the examination of specimens derived from the human body for the purpose of providing information for diagnostic, monitoring, or compatibility purposes.

N/A

Intended Use / Indications for Use

  • Temporary inlay and onlay treatments of the cavity.
  • Sealing of openings for implant screws.
  • Relining material for temporary crowns and bridges.
  • Covering of the gingival margin
  • Fixing of resin matrix during filling placement.
  • Block-out material for retentive ares in the dental arch, e.g. before taking impressions.

Product codes

EBF

Device Description

Once-Fil Flow is a light-curing filling material designed for temporary restorations. It serves as an interim solution before the placement of a final prosthesis, providing functions such as pulp protection and maintaining tooth position. The material is suitable for temporary fillings in prepared cavities and inlay/onlay cases.
To accommodate different clinical needs, Once-Fil Flow is available in two shades. The Yellow shade is ideal for anterior teeth, where esthetics are a priority, while the Blue shade provides clear visual contrast, making it easier to identify and remove before permanent restoration.
The application process involves preparing the tooth surface, applying an appropriate amount of material, and light-curing it. Once cured, the material can be finished and polished using conventional instruments.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Licensed Dentist or Dental Professional

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance Tests including Appearance, Weight, Packaging, Sensitivity to Ambient Light, Curing Depth, Water Sorption/Solubility, Color and Color stability, Shore D Hardness in accordance with ISO 4049, ISO 868.
Biocompatibility Tests:
ISO 10993-1 Biological evaluation of medical devise – Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009, test for in vitro cytotoxicity
ISO 10993-21:2021, Tests for irritation
ISO 10993-11:2017, Tests for systemic toxicity
ISO 10993-10:2021, Tests for skin sensitization

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Curing Depth: 4.3 mm
Water Absorption: 32 μg/mm³
Solubility: 5 μg/mm³
Color and color stability: All samples have a matching shade guide
Sensitivity to ambient light: All samples are physically homogeneous
Shore D Hardness: Standard: ISO 868 Result: 53

Predicate Device(s)

K153493

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.

FDA 510(k) Clearance Letter - Once-Fil Flow

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 6, 2025

Mediclus Co., Ltd.
Ku Da Hyeon
RA Associate
No. 1210, 134, Gongdan-ro, Heungdeok-gu
Cheongju-si, 28576
Korea, South

Re: K251283
Trade/Device Name: Once-Fil Flow
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth Shade Resin Material
Regulatory Class: Class II
Product Code: EBF
Dated: April 25, 2025
Received: April 25, 2025

Dear Ku Da Hyeon:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K251283 - Ku Da Hyeon Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K251283 - Ku Da Hyeon Page 3

Sincerely,

Bobak Shirmohammadi - S

For Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K251283

Device Name: Once-Fil Flow

Indications for Use (Describe)

  • Temporary inlay and onlay treatments of the cavity.
  • Sealing of openings for implant screws.
  • Relining material for temporary crowns and bridges.
  • Covering of the gingival margin
  • Fixing of resin matrix during filling placement.
  • Block-out material for retentive ares in the dental arch, e.g. before taking impressions.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

No. 1210, 134, Gongdan-ro, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, Korea
TEL +82-43-211-2877 FAX +82-43-211-2866 e-mail : sales@mdclus.com www.mediclus.co.kr

510(k) summary 1 / 4 page

510(k) Summary

(K251283)

This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: Apr 25,2025

1. Submitter/Contact Person

Da-Hyeon, Ku
MEDICLUS Co., Ltd.
No. 1210, 134, Gongdan-ro, Heungdeok-gu,
Cheongju-si, Chungcheongbuk-do, Republic of Korea
TEL : +82(43)211-2877 FAX : +82(43)211-2866
Email: ra@mdclus.com

2. U.S Agent

Priscilla Chung
LK Consulting Group USA, Inc.
18881 Von Karman Ave STE 160, Irvine CA 92612
Phone: 714.202.5789 Fax: 714-409-3357
Email: juhee.c@lkconsultinggroup.com

3. Device

  • Trade Name: Once-Fil Flow
  • Common Name: Tooth Shade Resin
  • Classification Name: Tooth Shade Resin Material
  • Product Code: EBF
  • Classification regulation: 21 CFR 872.3690

4. Predicate Device:

Clip Flow by VOCO GmbH (K153493)

5. Description:

Once-Fil Flow is a light-curing filling material designed for temporary restorations. It serves as an interim solution before the placement of a final prosthesis, providing

Page 6

No. 1210, 134, Gongdan-ro, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, Korea
TEL +82-43-211-2877 FAX +82-43-211-2866 e-mail : sales@mdclus.com www.mediclus.co.kr

510(k) summary 2 / 4 page

functions such as pulp protection and maintaining tooth position. The material is suitable for temporary fillings in prepared cavities and inlay/onlay cases.

To accommodate different clinical needs, Once-Fil Flow is available in two shades. The Yellow shade is ideal for anterior teeth, where esthetics are a priority, while the Blue shade provides clear visual contrast, making it easier to identify and remove before permanent restoration.

The application process involves preparing the tooth surface, applying an appropriate amount of material, and light-curing it. Once cured, the material can be finished and polished using conventional instruments.

6. Indication for use:

  • Temporary inlay and onlay treatments of the cavity
  • Sealing of openings for implant screws
  • Relining material for temporary crowns and bridges
  • Covering of the gingival margin
  • Fixing of resin matrix during filling placement
  • Block-out material for retentive areas in the dental arch, e.g. before taking impressions

7. Basis for Substantial Equivalence

7.1. Comparison Chart

Subject DevicePredicate DeviceEquivalence evaluation
ManufacturerMEDICLUS Co., Ltd.VOCO GmbH
Product NameOnce-Fil FlowClip Flow
510k#-K153493
Product CodeEBFEBF
MaterialBis-GMA, UDMA, HEMA, Silicone dioxide, Camphorquinone, 2,6-Di-tert-butyl-4-methylphenol (BHT), CoAl2O4 (Pigment), Etc..BHT, 2-hydroxyethyl methacrylate, Urethane Methacrylate, Pyrogenic silicic acids catalyst

Page 7

No. 1210, 134, Gongdan-ro, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, Korea
TEL +82-43-211-2877 FAX +82-43-211-2866 e-mail : sales@mdclus.com www.mediclus.co.kr

510(k) summary 3 / 4 page

Subject DevicePredicate DeviceEquivalence evaluation
Curing typeLight CuringLight CuringSame
Indications for Use Statement1) Temporary inlay and onlay treatments of the cavity.
  1. Sealing of openings for implant screws.
  2. Relining material for temporary crowns and bridges.
  3. Covering of the gingival margin.
  4. Fixing of resin matrix during filling placement.
  5. Block-out material for retentive areas in the dental arch, e.g. before taking impressions. | 1) Temporary inlay and onlay treatments of the cavity.
  6. Sealing of openings for implant screws.
  7. Relining material for temporary crowns and bridges.
  8. Covering of the gingival margin.
  9. Fixing of resin matrix during filling placement.
  10. Block-out material for retentive areas in the dental arch, e.g. before taking impressions. | Same |
    | Intended User | Licensed Dentist or Dental Professional | Licensed Dentist or Dental Professional | Same |
    | Technological Characteristics | | | |
    | Standard | ISO 4049 | ISO 4049 | Similar |
    | Curing Depth | 4.3 mm | 4 mm | |
    | Water Absorption | 32 μg/mm³ | | |
    | Solubility | 5 μg/mm³ | 18.4 μg/mm³ | |
    | Color and color stability | All samples have a matching shade guide | - | |
    | Sensitivity to ambient light | All samples are physically homogeneous | - | |
    | Shore D Hardness | Standard: ISO 868
    Result: 53 | - | |
    | Biocompatibility | Biocompatible | Biocompatible | Same |
    | Delivery method | • Delivery System: Syringe
    • Weight: 1.2ml
    • Disposable tip | • Delivery System: Syringe
    • Weight: 1.8g
    • Disposable tip | Similar |
    | Period of Use | Prolonged exposure(B)
    (exceed 24 hours but not 30 days) | Prolonged exposure(B)
    (exceed 24 hours but not 30 days) | Same |
    | Shelf-Life | 3 years | 3 years | Same |

7.2. Comparison Chart

The subject device has the same indications for use and the technological characteristics as the predicate device. The minor raw materials are different between the devices but the performance and the biocompatibility test results show that it does not raise a concern in safety and effectiveness.

8. Non-Clinical Testing

  • Performance Tests including

Page 8

No. 1210, 134, Gongdan-ro, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, Korea
TEL +82-43-211-2877 FAX +82-43-211-2866 e-mail : sales@mdclus.com www.mediclus.co.kr

510(k) summary 4 / 4 page

  • Appearance, Weight, Packaging, Sensitivity to Ambient Light, Curing Depth, Water Sorption/Solubility, Color and Color stability, Shore D Hardness in accordance with ISO 4049, ISO 868.

  • Biocompatibility Tests

    • ISO 10993-1 Biological evaluation of medical devise – Part 1: Evaluation and testing within a risk management process
    • ISO 10993-5:2009, test for in vitro cytotoxicity
    • ISO 10993-21:2021, Tests for irritation
    • ISO 10993-11:2017, Tests for systemic toxicity
    • ISO 10993-10:2021, Tests for skin sensitization

9. Conclusion

The subject device and the predicate device have the same intended use and have the same technological characteristics. Based on the similarities and the test results, we conclude that the subject device is substantially equivalent to the predicate device.