K Number
K251246

Validate with FDA (Live)

Device Name
Sparrow Ascent
Date Cleared
2025-08-25

(125 days)

Product Code
Regulation Number
882.5896
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.

Device Description

The Sparrow Ascent is a transcutaneous auricular neurostimulation (tAN) system intended to provide non-invasive, transcutaneous stimulation of the nerves around the auricle (ear). The device is indicated as an aid in the reduction of opioid withdrawal symptoms in adult patients. The Sparrow Ascent is a battery operated, prescription device that delivers mild electrical stimulation to the nerves around the auricle (ear), which carry information to the central nervous system. The Sparrow Ascent is to be used in clinical environments (e.g., doctor's office, clinics, rehab centers, and hospitals) and/or at home. Users of the subject device include adults experiencing opioid withdrawal symptoms. Stimulation parameters (i.e., the strength of stimulation) are set by the user's clinician, and users can only adjust stimulation intensity (amplitude) at home. The system consists of three main components 1) a disposable Earpiece, 2) a Cable, and 3) the External Pulse Generator (EPG).

AI/ML Overview

The provided FDA 510(k) clearance letter pertains to the Sparrow Ascent Transcutaneous Nerve Stimulator. The key aspect of this submission is a modification to the device's earpiece, specifically changing the stimulation site for the auricular branch of the vagus nerve (ABVN) from the cymba concha to the mastoid. The manufacturer aims to demonstrate that this change does not raise new questions of safety or effectiveness and that the modified device remains substantially equivalent to its predicate.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this 510(k) submission are primarily based on demonstrating functional equivalence of the new mastoid stimulation site to the previously cleared cymba concha site, specifically in terms of brain activation patterns, and overall safety and efficacy as compared to the predicate and reference devices.

Acceptance CriteriaReported Device Performance (from the document)
Brain Activation Equivalence: Spatial distribution of brain activity following mastoid stimulation is equivalent to cymba concha stimulation of the ABVN.fMRI study showed activation maps for mastoid and cymba concha stimulation were significantly correlated (r = 0.61, p < 0.001). This indicates consistent central responses between the two stimulation sites and reliable stimulation-induced brain activation at the mastoid.
Similar Current Density: The output related to nerve activation for the subject device is similar to the reference device."[T]he current density (i.e., the output related to nerve activation) is similar between the subject and the reference devices and helps demonstrate equivalent effectiveness." (No specific numerical range provided in the summary, but stated as similar.)
Similar Power and Charge Density: The output related to stimulation site impact for the subject device is similar to the predicate and reference devices."[P]ower and charge density (i.e., the output related to the impact on the stimulation site) is similar between the subject, the predicate, and the reference devices, and helps demonstrate equivalent safety." (No specific numerical range provided in the summary, but stated as similar.)
Non-clinical Performance: Device operates as expected, meets product requirements, delivers safe and effective stimulation."Non-clinical performance testing verified that, like the predicate, the subject device operates as expected when used as intended, and both meets and satisfies product requirements." "concluded that the subject device can safely and effectively deliver transcutaneous peripheral nerve stimulation when used as intended."
Biocompatibility: Materials used remain biocompatible."In addition to testing performed to support the clearance of K230796 (e.g., biocompatibility), which remains applicable to the subject device..." (Implies original biocompatibility data sufficient and still valid).
Service Life: Demonstrated 3-year service life."extended service life testing was completed for the subject device to verify a service life of 3 years."

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set (Clinical Study): 24 healthy adults.
  • Data Provenance: The document does not explicitly state the country of origin. It describes a prospective clinical study (fMRI study) conducted to compare brain activation patterns.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • The document does not mention the use of external experts to establish ground truth for the fMRI study. The "ground truth" in this context is the objective physiological response (brain activation patterns) measured by the fMRI, which is interpreted by the study's researchers based on established neuroimaging analysis techniques (e.g., GLM).
  • The qualifications of the researchers/analysts interpreting the fMRI data are not specified in the provided text.

4. Adjudication Method for the Test Set

  • No adjudication method is described for the fMRI study results. The analysis methods (e.g., General Linear Modeling, one-sample t-tests, correlation analysis) are standard scientific approaches for interpreting fMRI data, not an adjudication process by multiple human observers.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was mentioned or performed. This submission is not evaluating human reader performance with or without AI assistance. Instead, it focuses on the equivalence of a technological characteristic (stimulation site) for a device that delivers nerve stimulation for opioid withdrawal, supported by objective physiological data (fMRI).

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Performance

  • This is not an AI/Software as a Medical Device (SaMD) submission in the typical sense that would require standalone algorithm performance evaluation. The device is a hardware stimulator. The fMRI study evaluates the physiological effect of the stimulation, which is a standalone effect of the device itself, not an algorithm's diagnostic performance.

7. Type of Ground Truth Used

  • The primary "ground truth" for the clinical study was objective physiological data (brain activation patterns as measured by fMRI). This data was then analyzed statistically to demonstrate equivalence.

8. Sample Size for the Training Set

  • Not applicable / Not explicitly mentioned. This submission describes a modification to a hardware device and an fMRI study to demonstrate functional equivalence. There is no mention of an AI algorithm or a training set for such an algorithm. The 24 healthy adults were part of a test (or validation) set for the fMRI study, not a training set for an AI model.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As no AI algorithm training set is mentioned, the method for establishing its ground truth is not provided.

FDA 510(k) Clearance Letter - Sparrow Ascent

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 25, 2025

Spark Biomedical, Inc.
℅ Erin Gontang, PhD
Senior Consultant
Rqm+
2790 Mosside Blvd, Suite 800
Monroeville, Pennsylvania 15146

Re: K251246
Trade/Device Name: Sparrow Ascent
Regulation Number: 21 CFR 882.5896
Regulation Name: Percutaneous Nerve Stimulator For Substance Use Disorders
Regulatory Class: Class II
Product Code: PZR
Dated: April 22, 2025
Received: July 25, 2025

Dear Dr. Gontang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K251246 - Erin Gontang, PhD Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for

Page 3

K251246 - Erin Gontang, PhD Page 3

more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

ROBERT KANG -S

for Pamela Scott
Assistant Director
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K251246

Device Name
Sparrow Ascent

Indications for Use (Describe)
The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary

510(k) Summary
K251246

DATE PREPARED
August 25, 2025

MANUFACTURER AND 510(k) OWNER
Spark Biomedical, Inc.
18208 Preston Road, Suite D9 531, Dallas, TX 75252
Telephone: +1(844) 654-7775
Official Contact: Brent Croft, VP of Quality & Regulatory

REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., RAC
Erin A. Gontang, Ph.D.
RQM+
Telephone: +1 (412) 816-8253
Email: akomiyama@rqmplus.com, egontang@rqmplus.com

DEVICE INFORMATION
Proprietary Name/Trade Name: Sparrow Ascent
Common Name: Percutaneous Nerve Stimulator For Opioid Withdrawal
Classification Name: Percutaneous nerve stimulator for substance use disorders
Regulation Number: 21 CFR 882.5896
Class: Class II
Product Code: PZR
Premarket Review: Neurological and Physical Medicine Devices (OHT5)
Neuromodulation and Physical Medicine Devices (DHT5B)
Review Panel: Neurology

EXISTING DEVICE IDENTIFICATION
The Sparrow Ascent is substantially equivalent to the following predicate:

510(k) NumberPredicate Device Name / ManufacturerPredicate/ReferenceProduct Code /Reg Number
K230796Sparrow Ascent / Spark Biomedical, Inc.Primary PredicatePZR /21 CFR 882.5896
K233166NET Device / Net RecoveryReference DevicePZR /21 CFR 882.5896

Neither the predicate nor the reference device have been subject to a design related recall.

Page 6

DEVICE DESCRIPTION
The Sparrow Ascent is a transcutaneous auricular neurostimulation (tAN) system intended to provide non-invasive, transcutaneous stimulation of the nerves around the auricle (ear). The device is indicated as an aid in the reduction of opioid withdrawal symptoms in adult patients. The Sparrow Ascent is a battery operated, prescription device that delivers mild electrical stimulation to the nerves around the auricle (ear), which carry information to the central nervous system. The Sparrow Ascent is to be used in clinical environments (e.g., doctor's office, clinics, rehab centers, and hospitals) and/or at home. Users of the subject device include adults experiencing opioid withdrawal symptoms. Stimulation parameters (i.e., the strength of stimulation) are set by the user's clinician, and users can only adjust stimulation intensity (amplitude) at home. The system consists of three main components 1) a disposable Earpiece, 2) a Cable, and 3) the External Pulse Generator (EPG).

INDICATIONS FOR USE
The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.

COMPARISON OF INDICATIONS FOR USE STATEMENT
The Indications for Use statements of the subject device and the predicate device are identical. The intended use of both devices, to reduce the signs and symptoms associated with opioid withdrawal, is the same.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Spark Biomedical, Inc. believes that the Sparrow Ascent is substantially equivalent to the predicate device based on the information summarized here:

The subject device is identical to that cleared in K230796, with the exception of a modification to the device Earpiece. While both the subject and predicate devices have an electrode that stimulates the auricular branch of the vagus nerve (ABVN), the predicate device does so via an electrode within the cymba concha, whereas the subject device does so via an electrode on the mastoid. The electrodes provide identical output, thereby providing equivalent stimuli to the ABVN. The change in the Sparrow Ascent Earpiece design does not change the device principle of operation, and there has been no change to the materials used to manufacture the earpiece. Furthermore, although the overall size of the earpiece has been reduced, an assessment of the overall ratios of the patient contacting materials in the revised Sparrow Ascent Earpiece relative to the predicate device earpiece were determined to be equivalent.

Spark Biomedical also evaluated functional magnetic resonance imaging (fMRI) data to show that the spatial distribution of brain activity following mastoid and cymba concha stimulation of

Page 7

the ABVN was the same. The sham-controlled study looked at 24 healthy adults (mean age: 34.3 ± 11.2 years), and the brain activation maps showing mastoid and cymba concha stimulation were significantly correlated (r = 0.61, p < 0.001). The results indicated consistent central responses between the two stimulation sites, and the mastoid site demonstrated reliable stimulation-induced brain activation across participants.

The subject device is similar to the reference device (NET Device; cleared in K233166), which uses a similar ABVN stimulation site and has similar outputs to support the intended use of reducing the signs and symptoms associated with opioid withdrawal. The current density (i.e., the output related to nerve activation) is similar between the subject and the reference devices and helps demonstrate equivalent effectiveness. Additionally, power and charge density (i.e., the output related to the impact on the stimulation site) is similar between the subject, the predicate, and the reference devices, and helps demonstrate equivalent safety. This comparison helps demonstrate that the subject device's equivalence to the predicate is supported by the reference.

Page 8

Spark Biomedical Sparrow Ascent K251246 Subject DeviceSpark Biomedical Sparrow Ascent K230796 Primary PredicateNet Recovery NET Device K233166 Reference DeviceComparison to Predicate
Indications for UseThe Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.The NET Device is a transcutaneous alternating current stimulator (tACS) that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.Identical.
Stimulation Sites (electrode placement)Trigeminal & MastoidTrigeminal & Cymba ConchaMastoidSimilar. The subject device now stimulates the ABVN via the mastoid as opposed to the cymba concha. Clinical data shows equivalent brain activation following stimulation at both sites. The reference device also

Page 9

Spark Biomedical Sparrow Ascent K251246 Subject DeviceSpark Biomedical Sparrow Ascent K230796 Primary PredicateNet Recovery NET Device K233166 Reference DeviceComparison to Predicate
stimulates the ABVN via the mastoid. The cleared reference device shares the same IFU and the same electrode position as the subject device.
Product Code / RegulationPZR / 21 CFR 882.5846PZR / 21 CFR 882.5846PZR / 21 CFR 882.5846Identical.
Patient PopulationAdultAdultAdultIdentical.
Maximum Voltage (V)2.5 @ 500 Ω10 @ 2K Ω50 @ 10K Ω2.5 @ 500 Ω10 @ 2K Ω50 @ 10K Ω6.8 @ 500 Ω25 @ 2K Ω77 @ 10K ΩIdentical.
Maximum Current (mA)5.0 @ 500 Ω5.0 @ 2K Ω5.0 @ 10K Ω5.0 @ 500 Ω5.0 @ 2K Ω5.0 @ 10K Ω14.1 @ 500 Ω12.4 @ 2K Ω7.7 @ 10K ΩIdentical.
Maximum Pulse Width (μs)750750750Identical.
Maximum Frequency (Hz)1501503,000Identical.
Area of Smallest Electrode (cm²)1.1810.8532.080Similar. The observed parameter differences do not affect the substantial equivalence of the subject device to the predicate.
Power Density (rms W/cm²)0.002 @ 500 Ω0.007 @ 2K Ω0.051 @ 10K Ω0.003 @ 500 Ω0.01 @ 2K Ω0.07 @ 10K ΩNot Reported as RMSSimilar. The observed parameter differences do not affect the substantial equivalence of the subject device to the predicate.
Max Average Power Density (W/cm² @ 4Hz and 750 µs)Reported as RMSReported as RMS0.57 @ 2K Ω0.68 @ 10K ΩIdentical.

Page 10

SUMMARY OF NON-CLINICAL TESTING
In addition to testing performed to support the clearance of K230796 (e.g., biocompatibility), which remains applicable to the subject device, verification testing of the revised Sparrow Ascent Earpiece was conducted. Additionally, extended service life testing was completed for the subject device to verify a service life of 3 years. Non-clinical performance testing verified that, like the predicate, the subject device operates as expected when used as intended, and both meets and satisfies product requirements. As such, it is concluded that the subject device can safely and effectively deliver transcutaneous peripheral nerve stimulation when used as intended.

SUMMARY OF CLINICAL TESTING
To demonstrate similar spatial distribution of brain activity between mastoid and cymba concha stimulation sites, a functional magnetic resonance imaging (fMRI) study was conducted to compare central nervous system activation patterns during stimulation at each site. The fMRI study looked at 24 healthy adults (mean age: 34.3 ± 11.2 years) and was a sham-controlled, crossover study with three stimulation targets: the cymba concha, mastoid, and earlobe. Each subject underwent concurrent transcutaneous auricular vagus nerve stimulation (taVNS) and fMRI during three 8-minute stimulation blocks per site, randomized and counterbalanced, using a 30-second ON/OFF stimulation paradigm. The subject's brain responses were evaluated using general linear modeling (GLM), and functional activation maps were generated and compared across conditions.

The fMRI study showed that both mastoid and cymba concha stimulation elicited overlapping spatial patterns of brain activity. One-sample t-test analyses were performed across the 24 subjects, and activation maps showing mastoid and cymba concha stimulation were significantly correlated (r = 0.61, p < 0.001), indicating consistent central responses between the two stimulation sites. The findings of the fMRI study support the anatomical and functional equivalence of the mastoid site to the cymba concha for auricular branch of the vagus nerve (ABVN) activation. As such, the modification to the subject device Earpiece constitute a minor modification that does not change the intended use of the device, nor does it introduce different questions of safety or effectiveness.

CONCLUSION
Based on non-clinical and clinical testing, it can be concluded that the subject device does not raise different questions of safety or effectiveness compared to the predicate device. The identical indications for use, similar technological characteristics, and equivalent performance characteristics for the proposed Sparrow Ascent Earpiece demonstrate the subject device is substantially equivalent to the predicate device.

§ 882.5896 Percutaneous nerve stimulator for substance use disorders.

(a)
Identification. A percutaneous nerve stimulator for substance use disorders is a device that stimulates nerves percutaneously to aid in the reduction of withdrawal symptoms associated with substance use disorders.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Sterility testing of the percutaneous components of the device must be performed.
(6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the specified shelf life.
(7) Labeling must include the following:
(i) A detailed summary of the device technical parameters;
(ii) A warning stating that the device is only for use on clean, intact skin;
(iii) Instructions for use, including placement of the device on the patient; and
(iv) A shelf life.