K Number
K133442

Validate with FDA (Live)

Date Cleared
2014-03-11

(119 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The software's intended use is to assist diagnosis of chest pathology by minimizing anatomical distractions such as the ribs and clavicle in chest x-ray images.

Device Description

The Bone Suppression Software is a software component for use on diagnostic x-ray systems utilizing digital radiography (DR) or computed radiography (CR) technology. The software option suppresses bone anatomy in order to enhance visualization of chest pathology in a companion image that is delivered in addition to the original diagnostic image.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Bone Suppression Software, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly list quantitative acceptance criteria for the Bone Suppression Software's performance. It states that "Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as the predicate device." However, the specific metrics or thresholds for these criteria are not detailed.

Therefore, this section focuses on the qualitative claims made and the reported outcome.

Acceptance Criteria CategoryReported Device Performance (Qualitative)
Safety and Effectiveness"demonstrated that the device is as safe, as effective, and performs as well as the predicate device."
Image Acceptability"Clinical testing was conducted to evaluate the acceptability of the companion images for assisting diagnosis." (Implied: results were acceptable)
Design Output Compliance"Performance testing was conducted to verify the design output met the design input requirements." (Implied: met requirements)
User Needs/Intended Uses"to validate the device conformed to the defined user needs and intended uses." (Implied: conformed)
Substantial Equivalence"demonstrated that the device is as safe, as effective, and performs as well as the predicate device." (Supports a claim of substantial equivalence to predicates.)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified in the provided text. The document mentions "Clinical testing was conducted," but does not give a number of cases.
  • Data Provenance: Not specified. It's unclear what country the data came from or whether it was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified. The document only mentions "Clinical testing was conducted to evaluate the acceptability of the companion images for assisting diagnosis." It implies that qualified professionals reviewed the images, but their specific roles or experience levels are not detailed.

4. Adjudication Method for the Test Set

  • Not specified. The document does not describe any specific adjudication method (e.g., 2+1, 3+1 consensus) for establishing ground truth or evaluating the clinical images.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • MRMC Study: The document does not explicitly state that an MRMC comparative effectiveness study was done comparing human readers with and without AI assistance (bone suppression).
  • Effect Size: Not provided. Since an MRMC study is not confirmed, no effect size is discussed. The study's focus was on the "acceptability of the companion images for assisting diagnosis" and demonstrating equivalence to a predicate.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • The document describes the "Bone Suppression Software" as a "software component" that "suppresses bone anatomy in order to enhance visualization of chest pathology in a companion image that is delivered in addition to the original diagnostic image." This heavily implies that the software provides a processed image for human interpretation, rather than a standalone diagnostic output. While the algorithm performs its function independently, the "clinical testing" mentioned focuses on the acceptability of the companion images for assisting diagnosis, indicating an evaluation of the output intended for human review, not a standalone diagnostic performance evaluation against ground truth.

7. The Type of Ground Truth Used

  • The document implies that the ground truth for evaluation was based on the "acceptability of the companion images for assisting diagnosis" by clinical review. However, the specific type of ground truth (e.g., pathology, clinical follow-up, expert consensus on disease presence) against which the diagnostic enhancement was measured is not explicitly stated. It appears the ground truth was primarily related to the acceptability and utility of the suppressed images for assisting diagnosis rather than independently verifying the presence or absence of specific pathologies with a definitive gold standard.

8. The Sample Size for the Training Set

  • Not specified. The document does not provide any information about the training data or its sample size.

9. How the Ground Truth for the Training Set Was Established

  • Not specified. As no information on the training set or its ground truth is provided, the establishment method is unknown.

{0}------------------------------------------------

MAR 1 1 2014

K133442
Page 1 of 3

Traditional 510(k) for Bone Suppression Softwar

Carestream

"510(k) Summary"

510(k) Owner Name:Carestream Health, Inc.
510(k) Owner Address:150 Verona Street
Rochester, New York 14608
510(k) Owner Phone:585-627-6528
510(k) Owner Fax:585-323-7643
Contact Person & Info:Carolyn Wagner
Regulatory Affairs Manager, X-Ray Systemscarolyn.wagner@carestreamhealth.com
585-627-6557
Date Summary Prepared:October 30. 2013
Device Trade Name:Bone Suppression Software
Device Common Name:DR digital imager
Classification Name:Solid state x-ray imager (flat panel/digital imager)
Device Class:Class II
Device Code:MQB
Regulation Number:21 CFR 892.1680
Predicate Devices:Kodak Eclipse Image Processing Software
Manufactured by Carestream Health, Inc.
510(k) No. - K060137 (May 16. 2006)
Softview. Model 2.01
Manufactured by Riverain Medical Group
510(k) No. - K092363 (March 18, 2010)

Device Description:

The Bone Suppression Software is a software component for use on diagnostic x-ray systems utilizing digital radiography (DR) or computed radiography (CR) technology. The software option suppresses bone anatomy in order to enhance visualization of chest pathology in a companion image that is delivered in addition to the original diagnostic image.

{1}------------------------------------------------

Indications for Use / Intended Use:

The Indications for Use for the device, as described in its labeling, are: ··The software's intended use is to assist diagnosis of chest pathology by minimizing anatomical distractions such as the ribs and clavicle in chest x-ray images."

The intended use for this device. as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above. The Bone Suppression software component provides a bone suppressed companion image with posterior ribs and clavicle suppressed to aid diagnosis in chest radiographs. We believe that the Bone Suppression Software and the predicate devices have the same intended use.

The Indications for Use for the "Bone Suppression Software" is different than the Kodak Eclipse Image Processing Software predicate device. but these differences do not alter the intended diagnostic use of the software component. The Kodak Eclipse Image Processing Software optimizes radiographic images for diagnosis. The Bone Suppression companion view is generated using an additional image processing step that provides a bone-suppressed image to aid diagnosis in chest radiographs.

The Indications for Use for the "Bone Suppression Software" is worded differently than the Riverain "SoftView, Model 2.1" predicate device, but these differences are appropriately described as descriptive and do not alter the intended diagnostic use of the software component. Both devices provide a bone-suppressed companion view to minimize distractions in chest images. Both devices provide the original "standard of care` image along with the companion image.

Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination.

Comparison of Technological Characteristics:

The Bone Suppression Software is an extension of the predicate device, the Kodak Eclipse Image Processing Software. The Bone Suppression Software is an optional software component that allows a companion image to be formed with bone anatomy suppressed in order to enhance visualization of chest pathology. It is available as an integrated optional feature within the predicate device Kodak Eclipse Image Processing Software (DirectView software). The new software device (feature) raises no new issues of safety or effectiveness.

The Bone Suppression Software has the same technological characteristics as the predicate device, the SoftView, Model 2.01. Both devices provide an additional companion image with bone anatomy suppressed in order to enhance visualization of chest pathology. Both devices provide the companion view along with the original

{2}------------------------------------------------

Fraditional 510(k) for Bone Suppression Software

K133442
Page 3 of 3
ftware

"standard of care" chest radiograph. The Carestream "Bone Suppression Software" device raises no new issues of safety or effectiveness.

Discussion of Testing

Performance testing was conducted to verify the design output met the design input requirements and to validate the device conformed to the defined user needs and intended uses. Nonclinical testing was conducted under simulated use conditions. Clinical testing was conducted to evaluate the acceptability of the companion images for assisting diagnosis. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as the predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: the department's name encircling the image and an abstract symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The symbol is a stylized representation of a human figure embracing or being supported, composed of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 11, 2014

Carestream Health, Inc. % Ms. Carolyn Wagner Regulatory Affairs Manager 150 Verona Street ROCHESTER NY 14608

Re: K133442

Trade/Device Name: Bone Suppression Software Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: February 28, 2014 Received: February 28, 2014

Dear Ms. Wagner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of . devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Carolyn Wagner

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D. O'Hara

for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K133442

Device Name Bone Suppression Software

Indications for Use (Describe)

The software's intended use is to assist diagnosis of chest pathology by minimizing anatomical distractions such as the ribs and clavicle in chest x-ray images.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpant C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

Michael D. O'Hara

This section applies only to requirements of the Paperwork Reduction Act of 1995.

"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.