(61 days)
The Serim™ DISINTEK™ PA Test Strip is a chemical indicator for use in determining whether the concentration of peracetic acid, the active ingredient in the high-level disinfectant RAPICIDE™ PA HLD solution, is above the 850 ppm PAA minimum recommended concentration (MRC) established for RAPICIDE™ PA HLD at both 30°C and 20°C.
The device is a qualitative, single use, reagent chemical indicator made up of a 0.2-inch x 0.2-inch indicator pad that has been chemically treated to detect peracetic acid above or below the Minimum Recommended Concentration established for RAPICIDE™ PA High-Level Disinfectant. The indicator pad is affixed to one end of a 3.25-inch by 0.2-inch white plastic handle.
Here's an analysis of the acceptance criteria and study information based on the provided FDA 510(k) Clearance Letter for DISINTEK™ PA Test Strips:
Acceptance Criteria and Device Performance for DISINTEK™ PA Test Strips
1. Table of Acceptance Criteria and Reported Device Performance
| Study Parameter | Acceptance Criteria (Implicit from "Met Acceptance Criteria") | Reported Device Performance |
|---|---|---|
| Dynamic Range | Negative response at concentrations at or below MRC (850 ppm PAA), positive responses at higher concentrations. | 100% FAIL results at and below the 850 ppm PAA (MRC) and PASS results above the MRC at both 30°C and 20°C. |
| Instructions for Use Validation | Negative response at concentrations at or below MRC (850 ppm PAA), positive responses at higher concentrations, when used according to specified dip time (1-2 seconds) and sideblot time (25-30 seconds). | 100% FAIL results at and below the MRC and PASS results above the MRC with a 1-2 second dip time and a sideblot time at 25 seconds + 5 seconds for 30°C solutions and 30 seconds + 5 seconds for 20°C solutions. |
| Closed Bottle Shelf-Life Stability | Acceptable performance (100% FAIL at/below MRC, 100% PASS above MRC) for minimum claimed shelf-life (18 months) at specified storage conditions (≥ 10°C and up to 32°C). | Acceptable performance out to a minimum of 18 months shelf-life from ≥ 10°C storage up to 32°C, with 100% FAIL results at and below the MRC and PASS results above the MRC. (Data collection ongoing for extended shelf-life). |
| Open Bottle Use-Life Stability | Acceptable performance (100% FAIL at/below MRC, 100% PASS above MRC) for minimum claimed open bottle use-life (18 months) at specified storage conditions (constant high humidity, repeated openings). | Acceptable performance out to a minimum of 18 months open bottle use-life with 100% FAIL results at and below the MRC and PASS results above the MRC. (Data collection ongoing for extended open bottle use-life). |
| Comparative Sensitivity and Specificity | 100% comparative sensitivity and 100% comparative specificity against the predicate device. | Comparative sensitivity of 100% and comparative specificity of 100% was found for the DISINTEK PA test strips at both 30°C and 20°C. |
| Analytic Specificity – Bioburden | Acceptable performance (100% FAIL at/below MRC, 100% PASS above MRC) in the presence of bioburden (fetal bovine serum and hard water). | 100% FAIL results at and below the MRC and 100% PASS results above the MRC in the presence of fetal bovine serum and hard water. |
| Analytic Specificity – Other Disinfectants | Test strip yields PASS results when exposed to other peracetic acid disinfectants at their use concentrations. | Test strip yielded PASS results using these disinfectants. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific numerical sample sizes (e.g., number of test strips, unique disinfectant solutions tested, or replicates) used for each test set within the "Dynamic Range," "Instructions for Use Validation," "Closed Bottle Shelf-Life Stability," "Open Bottle Use-Life Stability," "Analytic Specificity – Bioburden," or "Analytic Specificity – Other Disinfectants" studies.
For Comparative Sensitivity and Specificity, the document also does not specify the numerical sample size.
Data Provenance: The studies were conducted by Serim Research Corporation themselves, implying the data is generated internally from their own testing facilities. The data appears to be prospective, as it describes the testing performed on the new DISINTEK™ PA Test Strips to demonstrate their performance against defined criteria. No information regarding country of origin of the data beyond "Serim Research Corporation" is provided, but as an FDA submission, the testing would likely adhere to US regulatory standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not provided in the document. For chemical indicators like this, "ground truth" typically refers to the analytically determined concentration of PAA in the test solutions. This would be established by validated chemical testing methods (e.g., titration or spectroscopy) rather than expert consensus on visual interpretation. The document implies that the "ground truth" for FAIL/PASS results is based on a known PAA concentration (below/at 850 ppm MRC as FAIL, above as PASS), which would be determined analytically.
4. Adjudication Method for the Test Set
The document describes chemical indicator functionality (color change) and the interpretation of those changes (PASS/FAIL based on visual comparison to color blocks or a "solid black color" for the predicate). There is no indication of an adjudication method involving multiple human readers for interpreting the results of the DISINTEK™ PA Test Strips. The expectation is that the visual interpretation by a single user, following the provided instructions and color chart, is sufficient.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is usually conducted for imaging diagnostics or devices where human interpretation directly impacts effectiveness and can be improved by AI assistance. This device is a chemical indicator with a direct visual interpretation, not a diagnostic imaging AI system.
6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)
Yes, the studies described are essentially standalone in terms of the device's chemical performance. The studies evaluate the chemical indicator's ability to accurately reflect the PAA concentration, independent of human variability in reading.
However, the "Instructions for Use Validation" section does implicitly involve human interaction, as it validates the device when used as directed by a human with a specific dip and read time. The "performance" being measured (100% FAIL below MRC, 100% PASS above MRC) is the device's inherent chemical reaction, and the human's role is to correctly perform the test and interpret the color chart. So, while it's not an algorithm only in the AI sense, the core data is about the chemical functionality.
7. Type of Ground Truth Used
The ground truth used is based on analytically determined peracetic acid (PAA) concentrations in the RAPICIDE™ PA HLD solutions. Solutions were prepared with known concentrations below, at, and above the 850 ppm PAA Minimum Recommended Concentration (MRC). The "FAIL" result would be correlated with PAA concentrations ≤ 850 ppm, and the "PASS" result with PAA concentrations > 850 ppm.
8. Sample Size for the Training Set
This information is not applicable/not provided. The DISINTEK™ PA Test Strips are a chemical indicator, not an AI/machine learning device. Therefore, there is no "training set" in the context of an algorithm that learns from data.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable/not provided for the reasons stated in point 8.
FDA 510(k) Clearance Letter - DISINTEK™ PA Test Strips
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
June 3, 2025
Serim Research Corporation
Jackie Nelson
Director of Quality Operations
3506 Reedy Drive
Elkhart, Indiana 46514
Re: K251035
Trade/Device Name: DISINTEK™ PA Test Strips
Regulation Number: 21 CFR 880.2800
Regulation Name: Sterilization Process Indicator
Regulatory Class: Class II
Product Code: JOJ
Dated: April 3, 2025
Received: April 3, 2025
Dear Jackie Nelson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K251035 - Jackie Nelson Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K251035 - Jackie Nelson Page 3
Sincerely,
Stephen A. Anisko -S
Digitally signed by Stephen A. Anisko -S
Date: 2025.06.03 16:06:06 -04'00'
for: Christopher K. Dugard, M.S.
Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
Indications for Use
See PRA Statement below.
510(k) Number (if known)
K251035
Device Name
DISINTEK™ PA Test Strips
Indications for Use (Describe)
The Serim™ DISINTEK™ PA Test Strip is a chemical indicator for use in determining whether the concentration of peracetic acid, the active ingredient in the high-level disinfectant RAPICIDE™ PA HLD solution, is above the 850 ppm PAA minimum recommended concentration (MRC) established for RAPICIDE™ PA HLD at both 30°C and 20°C.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1
Page 5
SERIM Research Corporation
510(k) Premarket Notification
Rapicide PA Test Strips
510(K) SUMMARY
510(k): K251035
Prepared: April 01, 2025
Revised: June 2, 2025
Submitter: Serim Research Corporation
Address: 3506 Reedy Drive
Elkhart IN 46514
Phone: 574-264-3440
Fax: 574-266-6222
Contact: Jackie Nelson
Director of Regulatory Operations
Device Trade Name: DISINTEK™ PA Test Strip
Common or Usual Name: Indicator for RAPICIDE™ PA HLD High-Level Disinfectant
Device Classification Name: Chemical Indicators for Liquid Chemical Germicide.
(b) Class II (Physical/Chemical Sterilization Process Indicator).
Product Code: JOJ
Class: II
Regulation Number: 21CFR 880.2800
Predicate Device: MEDIVATORS™ RAPICIDE™ PA Minimum Required Concentration (MRC) Test Strips, REF ML02-0118; K152394.
Device Description: The device is a qualitative, single use, reagent chemical indicator made up of a 0.2-inch x 0.2-inch indicator pad that has been chemically treated to detect peracetic acid above or below the Minimum Recommended Concentration established for RAPICIDE™ PA High-Level Disinfectant. The
K251035 1
Page 6
SERIM Research Corporation
510(k) Premarket Notification
Rapicide PA Test Strips
indicator pad is affixed to one end of a 3.25-inch by 0.2-inch white plastic handle.
Intended Use: The Serim™ DISINTEK™ PA Test Strip is a chemical indicator for use in determining whether the concentration of peracetic acid, the active ingredient in the high-level disinfectant RAPICIDE™ PA HLD solution, is above the 850 ppm PAA minimum recommended concentration (MRC) established for RAPICIDE™ PA HLD at both 30°C and 20°C.
Indications for Use: The Serim™ DISINTEK™ PA Test Strip is a chemical indicator for use in determining whether the concentration of peracetic acid, the active ingredient in the high-level disinfectant RAPICIDE™ PA HLD solution, is above the 850 ppm PAA minimum recommended concentration (MRC) established for RAPICIDE™ PA HLD at both 30°C and 20°C.
Technological Characteristics: The DISINTEK PA Test Strip contains colorimetric indicators, a catalyst, a reducing agent, and other non-reactive components. The reaction process involved with the test pad is based on a multi-step reaction. The first step involves a reaction in which the reducing agent reacts with PAA in the RAPICIDE PA HLD up to the Minimum Recommended Concentration (MRC) and neutralizes it. The next step is a reaction of the remaining peracetic acid with a catalyst, which then reacts with colorimetric indicators at peracetic acid levels above 850 ppm, resulting in a green pad coloration. The device will reliably indicate if the peracetic acid concentration is above or below the MRC level of 850 ppm peracetic acid at both 30°C and 20°C. Refer to Table 1 for more information.
K251035 2
Page 7
SERIM Research Corporation
510(k) Premarket Notification
Rapicide PA Test Strips
Table 1: Summary of Technological Characteristics Table
| Parameters | DISINTEK™ PA Test Strips | MEDIVATORS™ RAPICIDE™ PA Minimum Required Concentration (MRC) Test Strips, REF ML02-0118; K152394 | Comparison |
|---|---|---|---|
| Analyte | PAA in RAPICIDE™ PA HLD Solution. | PAA in RAPICIDE™ PA HLD Solution. | Same |
| Indicator strip | 0.2" x 0.2" test paper attached to a plastic handle | 0.2" x 0.2" test paper attached to a plastic handle | Same |
| Chemistry | A reducing agent reacts with PAA up to the Minimum Recommended Concentration (MRC) and neutralizes it. If the PAA concentration in the solution is > MRC, the remaining PAA reacts with a halide and redox indicators, creating a colored compound. The test pad changes from yellow to green. | The chemical principle of the test is based on the oxidation of iodide to iodine by peracetic acid. A dark complex is obtained in the presence of starch. | Different |
| Indicator Agents | Redox indicators | Iodide/Starch | Different |
| Chemicals Present in the Device | Polymers, halide, reducing agent, redox indicators, background dye. | Starch, iodide | Different |
| Minimum Time of Color Change | 25-30 seconds | 30 seconds | Same |
K251035 3
Page 8
SERIM Research Corporation
510(k) Premarket Notification
Rapicide PA Test Strips
| Parameters | DISINTEK™ PA Test Strips | MEDIVATORS™ RAPICIDE™ PA Minimum Required Concentration (MRC) Test Strips, REF ML02-0118; K152394 | Comparison |
|---|---|---|---|
| Details of Chemical Formulation | A reducing agent reacts with PAA up to the Minimum Recommended Concentration (MRC) and neutralizes it. If the PAA concentration in the solution is > MRC, the remaining PAA reacts with a halide and redox indicators, creating a colored compound. The test pad changes from yellow to green. | The chemical principle of the test is based on the oxidation of iodide to iodine by peracetic acid. A dark complex is obtained in the presence of starch. | Different |
| Results | A test pad which is similar in color to the FAIL color blocks indicates the solution is ≤ 850 ppm PAA (MRC). If the color of the test pad is similar to the PASS blocks, the solution is ≥ 1100 ppm PAA. | Any test strip color other than solid black is a FAIL result (i.e., ≤ 850 ppm PAA MRC). A solid black color indicates a PASS result (i.e., > 850 ppm PAA). | Same |
| "PASS" Indication | Green | Solid Black | Different |
Bold Italic text indicates differences between the proposed and predicate device.
The DISINTEK PA Test Strips have the same intended use as the predicate device. Both devices measure the potency of peracetic acid in the disinfectant solution to determine whether it is above or below the Minimum Recommended Concentration at both 30°C and 20°C. The differences between the proposed peracetic acid test strip and the predicate device are the colorimetric indicators. The differences do not impact the safety and effectiveness of the subject device. The risk to safety and effectiveness of the color differences between the two devices is mitigated by using validated directions for use and a validated color chart that accurately conveys how a user is to use the device and interpret the color at the appropriate read time.
K251035 4
Page 9
SERIM Research Corporation
510(k) Premarket Notification
Rapicide PA Test Strips
Summary of Non-Clinical Performance Data: The performances of the DISINTEK PA Test Strips were evaluated in non-clinical tests in a range of concentrations below, at, and above the MRC at both 30°C and 20°C. The results show that the proposed device is an effective monitor of the concentration of peracetic acid in the RAPICIDE PA HLD solution when used as directed by the manufacturer.
Table 2 summarizes the non-clinical testing performed by Serim Research to demonstrate safety and effectiveness of the DISINTEK PA Test Strips in monitoring the concentration of RAPICIDE PA HLD high-level disinfectant solutions. A further discussion of each individual study follows the table.
Table 2: DISINTEK™ PA Test Strips: Performance Testing Summary
| Study | Results |
|---|---|
| Dynamic Range | Met Acceptance Criteria: Negative response at concentrations at or below MRC, positive responses at higher concentrations. |
| Instructions for Use Validation | Met Acceptance Criteria: Negative response at concentrations at or below MRC, positive responses at higher concentrations. |
| Closed Bottle Shelf-Life Stability | Met Acceptance Criteria: Met specifications for 18 months (unopened). |
| Open Bottle Use-Life Stability | Met Acceptance Criteria: Met specification for 18 months (opened). |
| Comparative Sensitivity and Specificity | Met Acceptance Criteria: Comparative sensitivity and specificity of 1 |
Dynamic Range: The dynamic range of the DISINTEK PA Test Strips was evaluated below, at, and above the MRC using RAPICIDE PA HLD high-level disinfectant solutions. The test strip yielded acceptable performance with 100% FAIL results at and below the 850 ppm PAA (MRC) and PASS results above the MRC at both 30°C
K251035 5
Page 10
SERIM Research Corporation
510(k) Premarket Notification
Rapicide PA Test Strips
and 20°C.
Instructions for Use Validation: The instructions for use of the DISINTEK PA Test Strips were evaluated below, at, and above the MRC using RAPICIDE PA HLD high-level disinfectant solutions at both 30°C and 20°C. The chemical indicators yielded acceptable performance with a 1-2 second dip time and a sideblot time at 25 seconds + 5 seconds for 30°C solutions and 30 seconds + 5 seconds for 20°C solutions. The test strips yielded acceptable performance with 100% FAIL results at and below the MRC and PASS results above the MRC.
Closed Bottle Shelf-Life Stability: The closed bottle shelf-life of the DISINTEK PA Test Strips was evaluated below, at, and above the MRC with RAPICIDE PA HLD high-level disinfectant solutions with test strip storage ≥ 10°C and up to 32°C. The test strip yielded acceptable performance out to a minimum of 18 months shelf-life from ≥ 10°C storage up to 32°C, with data collection ongoing. The test strip yielded acceptable performance with 100% FAIL results at and below the MRC and PASS results above the MRC. An 18-month shelf-life is claimed for the DISINTEK PA Test Strips.
Open Bottle Use-Life Stability: The open bottle use-life of the DISINTEK PA Test Strips was evaluated below, at, and above the MRC with RAPICIDE PA HLD high-level disinfectant solutions at both 30°C and 20°C at constant high humidity storage (~ 80% RH) and repeated openings of the test strip bottle. The test strips yielded acceptable performance out to a minimum of 18 months open bottle use-life, with data collection ongoing. The chemical indicator yielded acceptable performance with 100% FAIL results at and below the MRC and PASS results above the MRC. An 18- month open bottle use-life is claimed for the DISINTEK PA Test Strips.
Comparative Sensitivity and Specificity: The DISINTEK PA Test Strips were compared to the predicate device, MEDIVATORS™ RAPICIDE™ PA Minimum Required Concentration (MRC) Test Strips, REF ML02-0118 using comparison studies. A comparative sensitivity of 100% and comparative
K251035 6
Page 11
SERIM Research Corporation
510(k) Premarket Notification
Rapicide PA Test Strips
specificity of 100% was found for the DISINTEK PA test strips at both 30°C and 20°C.
Analytic Specificity – Bioburden: The sensitivity of the DISINTEK PA Test Strips to bioburden was evaluated below, at, and above the MRC using RAPICIDE PA HLD high-level disinfectant solutions at both 30°C and 20°C using fetal bovine serum and hard water. The chemical indicator yielded acceptable performance with 100% FAIL results at and below the MRC and 100% PASS results above the MRC.
Analytic Specificity – Other Disinfectants: The sensitivity of the DISINTEK PA Test Strips to other peracetic acid disinfectants was evaluated at use concentrations similar to the RAPICIDE PA HLD disinfectant at normal dilution concentrations (1:1:48). The test strip yielded PASS results using these disinfectants. These results are documented in the Product Insert for the test strip.
Conclusion: Based on the non-clinical tests performed, the subject device, DISINTEK PA Test Strip, is as safe, as effective, and performs as well as or better than the legally marketed predicate device, MEDIVATORS™ RAPICIDE™ PA Minimum Required Concentration (MRC) Test Strips, REF ML02-0118, K152394, Class II (21CFR 880.2800), Product code JOJ.
MEDIVATORS™ and RAPICIDE™ are trademarks of Medivators, Inc.
K251035 7
§ 880.2800 Sterilization process indicator.
(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).