K Number
K250940

Validate with FDA (Live)

Date Cleared
2025-06-26

(90 days)

Product Code
Regulation Number
876.1725
Age Range (0 - 120)
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atmo® Gas Capsule System measures whole gut and regional gut (stomach, small bowel, and colon) transit times. Measurements of gastrointestinal transit times are used for evaluating motility disorders.

Gastric Emptying Time (GET) is indicated for the evaluation of patients with suspected gastroparesis. Delayed gastric emptying is implicated in such disorders as idiopathic and diabetic gastroparesis and functional non-ulcer dyspepsia.

Colonic Transit Time (CTT) is indicated for the evaluation of colonic transit in patients with chronic constipation, to aid in differentiating slow and normal transit constipation. In the case where ileocecal junction transit cannot be determined, the system will report combined Small and Large Bowel Transit Time (SLBTT).

Transit times are derived from measures of temperature, hydrogen concentration, and carbon dioxide concentration, and indicators of oxygen level, Capsule tumble, and antenna reflectance.

Not for use in pediatric patients.

Device Description

The Atmo Gas Capsule System is a prescription only ingestible telemetric medical device system used to measure transit times of the gastrointestinal (GI) tract. The system consists of a single-use Capsule, reusable Receiver, Mobile Device with Clinician App, and an online Clinician Portal.

Sensors onboard the Capsule provide measurements for temperature, hydrogen concentration, and carbon dioxide concentration, along with indicators of oxygen level, capsule tumble, and antenna reflectance that are used to identify regional gastrointestinal anatomical landmarks. The Capsule data is transmitted from within the GI tract via radiofrequency communication to a Receiver worn by the patient. The Mobile Device runs the Atmo Clinic App, a software application which guides the clinician through the pairing process, Capsule administration, and subsequent transfer of stored data from the Receiver to a cloud server. The Atmo Clinician Portal, accessible through a web browser, displays the resulting data and provides guidance for the physician to review, confirm and create a downloadable Motility Study Report containing whole and regional gut transit times.

AI/ML Overview

The provided FDA 510(k) clearance letter and its associated 510(k) summary pertain to the Atmo Gas Capsule System. However, this document does not contain specific acceptance criteria or the detailed results of a study designed to prove the device met those criteria in the format requested.

The information provided largely describes the device, its intended use, comparison to a predicate device, and general details about the clinical study (e.g., multi-site, prospective, sample size, Bland-Altman analysis for agreement) and non-clinical testing. It states that "all pre-determined end points successfully met" for agreement with the predicate device, but it does not specify what those pre-determined endpoints/acceptance criteria were (e.g., a specific agreement percentage, limits of agreement, maximum allowable difference, or statistical p-value for non-inferiority).

Therefore, based on the provided text, I cannot complete most of the requested table and details about the study that proves the device meets the acceptance criteria. I can only extract what is present.

Here's what can be inferred and what information is missing:


Acceptance Criteria and Device Performance

Since the document does not explicitly state the acceptance criteria for GET and CTT agreement, I cannot fill this table as requested. It only says "all pre-determined end points successfully met."

Table of Acceptance Criteria and Reported Device Performance (Information Not Provided in Source):

MetricAcceptance Criteria (e.g., % agreement, LOA, etc.)Reported Device Performance (Atmo Gas Capsule System)Result (Met/Not Met)
Gastric Emptying Time (GET) AgreementNot explicitly stated in the document."Bland-Altman analysis showed agreement between Atmo Gas Capsule System and the Predicate device for GET.""all pre-determined end points successfully met."
Colonic Transit Time (CTT) AgreementNot explicitly stated in the document."Bland-Altman analysis showed agreement between Atmo Gas Capsule System and the Predicate device for CTT.""all pre-determined end points successfully met."
SafetyMinimal Adverse Events"No serious adverse events (SAE) were associated with the Atmo Gas Capsule system. No device deficiencies led to SAEs or serious adverse device effects. Minimal adverse events (AE) were observed with all reported AEs being expected and common to the study population."Met

Study Details Proving Device Meets Acceptance Criteria

  1. Sample size used for the test set and the data provenance:

    • Sample Size: 213 participants.
    • Data Provenance: Multi-site, prospective study across 12 US sites and 1 OUS (Outside US) site. Participants had "confirmed or suspected dysmotility issues of the gastric and/or colonic regions."
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not provided. The study assessed agreement between the Atmo Gas Capsule System and a predicate device (SmartPill GI Monitoring System, version 2.0). It doesn't explicitly mention external expert consensus for ground truth on transit times, but rather the comparative performance against an already cleared device. For a "gastrointestinal motility monitoring system," the ground truth for transit times is typically derived from the measurements themselves, not from expert interpretation of images or other subjective data.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not provided. This type of adjudication (e.g., for image interpretation or diagnosis) is generally not applicable to the direct measurement of physiological parameters like transit times by a capsule system. The study focused on agreement between two devices.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This device is a measurement system, not an AI-assisted diagnostic tool that humans interpret. The study compared the device's measurements to those of a predicate device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in the context of the device itself. The Atmo Gas Capsule System and the predicate device both automatically identify transit markers, with the user then manually confirming them. The study evaluated the agreement of the Atmo system's derived transit times against the predicate, implying the performance of the system's algorithm in calculating these times. The "human-in-the-loop" aspect (manual confirmation of markers) is part of the system's operation, but the core measurement and algorithm for transit time derivation is standalone.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the Atmo Gas Capsule System's performance was the measurements obtained by the legally marketed predicate device, the SmartPill GI Monitoring System, version 2.0. The study's primary objective was assessing agreement between the investigational device and the predicate device.
  7. The sample size for the training set:

    • Not provided. The document describes a clinical validation study (test set) but does not mention details about the training data used for the device's algorithms.
  8. How the ground truth for the training set was established:

    • Not provided, as training set details are absent.

FDA 510(k) Clearance Letter - Atmo Gas Capsule System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

June 26, 2025

Atmo Biosciences Ltd
Ian Macfarlane
Regulatory Manager
436 Elgar Rd
Box Hill, VIC 3128
Australia

Re: K250940
Trade/Device Name: Atmo Gas Capsule System
Regulation Number: 21 CFR 876.1725
Regulation Name: Gastrointestinal Motility Monitoring System
Regulatory Class: Class II
Product Code: NYV
Dated: March 28, 2025
Received: March 28, 2025

Dear Ian Macfarlane:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"


U.S. FOOD & DRUG ADMINISTRATION

June 26, 2025

Atmo Biosciences Ltd
Ian Macfarlane
Regulatory Manager
436 Elgar Rd
Box Hill, VIC 3128
Australia

Re: K250940
Trade/Device Name: Atmo Gas Capsule System
Regulation Number: 21 CFR 876.1725
Regulation Name: Gastrointestinal Motility Monitoring System
Regulatory Class: Class II
Product Code: NYV
Dated: March 28, 2025
Received: March 28, 2025

Dear Ian Macfarlane:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 2

K250940 - Ian Macfarlane Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250940 - Ian Macfarlane Page 3

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

510(k) Number (if known)
K250940

Device Name
Atmo Gas Capsule System

Indications for Use (Describe)

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

The Atmo® Gas Capsule System measures whole gut and regional gut (stomach, small bowel, and colon) transit times. Measurements of gastrointestinal transit times are used for evaluating motility disorders.

Gastric Emptying Time (GET) is indicated for the evaluation of patients with suspected gastroparesis. Delayed gastric emptying is implicated in such disorders as idiopathic and diabetic gastroparesis and functional non-ulcer dyspepsia.

Colonic Transit Time (CTT) is indicated for the evaluation of colonic transit in patients with chronic constipation, to aid in differentiating slow and normal transit constipation. In the case where ileocecal junction transit cannot be determined, the system will report combined Small and Large Bowel Transit Time (SLBTT).

Transit times are derived from measures of temperature, hydrogen concentration, and carbon dioxide concentration, and indicators of oxygen level, Capsule tumble, and antenna reflectance.

Not for use in pediatric patients.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) SUMMARY - Atmo Gas Capsule System

1 Submitter and Device Details

Date Prepared:19-June-2025
Manufacturer:Atmo Biosciences Ltd436 Elgar RdBox Hill, VIC 3128Australia
Contact Information:Phone: +61 3 9945 7510Cell: +61 400 807 893Email: ian.macfarlane@atmobiosciences.com
Contact Person:Ian MacfarlaneRegulatory Manager
Device Trade Name:Atmo® Gas Capsule System
Regulatory Class:Class II
Regulation Number:21 CFR 876.1725 Gastrointestinal motility monitoring system
Product Code:NYV Capsule Gastrointestinal Motility System
Predicate Device(s)SmartPill GI Monitoring System, version 2.0
Predicate Device 510(k) numberK092342

2 Device Description

The Atmo Gas Capsule System is a prescription only ingestible telemetric medical device system used to measure transit times of the gastrointestinal (GI) tract. The system consists of a single-use Capsule, reusable Receiver, Mobile Device with Clinician App, and an online Clinician Portal.

Sensors onboard the Capsule provide measurements for temperature, hydrogen concentration, and carbon dioxide concentration, along with indicators of oxygen level, capsule tumble, and antenna reflectance that are used to identify regional gastrointestinal anatomical landmarks. The Capsule data is transmitted from within the GI tract via radiofrequency communication to a Receiver worn by the patient. The Mobile Device runs the Atmo Clinic App, a software application which guides the clinician through the pairing process, Capsule administration, and subsequent transfer of stored data from the Receiver to a cloud server. The Atmo Clinician Portal, accessible through a web browser, displays the resulting data and provides guidance for the physician to review, confirm and create a downloadable Motility Study Report containing whole and regional gut transit times.

K250940
Page 1 of 4

Page 6

3 Indications for Use

The Atmo Gas Capsule System measures whole gut and regional gut (stomach, small bowel, and colon) transit times. Measurements of gastrointestinal tract transit times are used for evaluating motility disorders.

Gastric Emptying Time (GET) is indicated for the evaluation of patients with suspected gastroparesis. Delayed gastric emptying is implicated in such disorders as idiopathic and diabetic gastroparesis and functional non-ulcer dyspepsia.

Colonic Transit Time (CTT) is indicated for the evaluation of colonic transit in patients with chronic constipation, to aid in differentiating slow and normal transit constipation. In the case where ileocecal junction transit cannot be determined, the system will report combined Small and Large Bowel Transit Time (SLBTT).

Transit times are derived from measures of temperature, hydrogen concentration, and carbon dioxide concentration, and indicators of oxygen level, Capsule tumble, and antenna reflectance.

Not for use in pediatric patients.

4 Summary of Technological Characteristics

The Atmo Gas Capsule System has similar key physical attributes to the Predicate. The Atmo Capsule and Receiver are slightly smaller and the overall usability reflects current human factors engineering best practice. The Atmo Gas Capsule System incorporates different sensors and employs a different algorithm to make its determination of gastrointestinal tract transit times.

Data flow for the Atmo system is from the Capsule to the Receiver, then via the cloud to an online portal for analysis, viewing and interpretation by the clinician. The Predicate transmits data from the capsule to a receiver, then to a local PC for analysis, viewing and interpretation by the clinician.

A comparison of the Atmo Gas Capsule and Predicate device attributes is shown below.

Device CharacteristicAtmo Gas Capsule SystemSmartPill GI Monitoring System, Version 2.0 (Predicate Device, K092342)
Capsule
Sensors• Temperature and Relative Humidity Sensor• Thermal Conductivity Detector (TCD)• Volatile Organic Compound (VOC) Sensor• Triaxial Accelerometer• Antenna Reflectometer• Temperature Sensor• pH Sensor• Pressure Sensor
Transmission Signal to ReceiverRF Signal centered frequency of 433.9 MHz (radiating between 433.5-434.5 MHz)RF Signal centered frequency of 434.2 MHz (radiating between 426-445 MHz)¹
Data StreamContinuousContinuous
Energy usedSilver oxide batteriesSilver oxide Batteries
Battery Life>240 hrs. (4 days gas sensing, then a further 6 days of temperature only sensing)>120 hrs
Size27.8 mm (length)11.1 mm (diameter)< 35 mm (length)< 15 mm (diameter)
Weight3.8 g4.5 g¹

K250940
Page 2 of 4

Page 7

Device CharacteristicAtmo Gas Capsule SystemSmartPill GI Monitoring System, Version 2.0 (Predicate Device, K092342)
SterilityNon-sterileNon-Sterile
ReuseSingle UseSingle Use¹
User Performed CalibrationNo user calibration is required. Capsule calibration is performed as part of the manufacturing process.pH sensor calibration is performed by the user at time of administration.
Receiver
ReceiverRechargeable, belt worn ReceiverRechargeable, belt clip or lanyard worn receiver
Data Communication from ReceiverUSB cable transfers data from the Receiver to the Mobile Device which uploads the data to the cloud server for analysisDocking station and USB
Analysis Software
Identification of Transit MarkersSoftware automatically identifies capsule ingestion, gastric emptying through gastroduodenal junction (GDJ), ileocecal junction (ICJ), and body exit. The user must manually confirm each marker.Software automatically identifies capsule ingestion, gastric emptying (GDJ), ileocecal junction (ICJ), and body exit. The user must manually confirm each marker.
GI Regional Transit Measurements• GET – Gastric emptying time• SBTT – Small bowel transit time• OCTT – Orocecal transit time• SLBTT – Combined small and large bowel transit time• CTT – Colonic transit time• WGTT – Whole gut transit time• GET – Gastric emptying (transit) time• SBTT – Small bowel transit time• SLBTT – Combined small and large bowel transit time• CTT – Colonic transit time• WGTT – Whole gut transit time
Recording EventsBowel movements recorded via single button press on the receiver. Additional information such as meals and symptoms can be recorded in a provided patient diary.Any event's timing recorded via single button press on the receiver. Event details are recorded in the patient diary (e.g., meals, bowel movement, gas, drank liquid, and went to bed)
PlatformClinician portal hosted on a secure cloud server accessed using an internet connected computerSoftware on a computer

¹ SmartPill® GI Monitoring System, User Manual, MotiliGI® v3.0, March 2013

5 Performance Data

5.1 Non-clinical Testing

Non-clinical testing was performed to verify device specifications and confirm the safety and performance of the device. In all instances, the Atmo Gas Capsule System functioned as intended. Non-clinical testing included:

  • Biocompatibility
  • Human factors
  • Electrical safety
  • Electromagnetic compatibility
  • Software, including cybersecurity

K250940
Page 3 of 4

Page 8

  • Systems and Mechanical
  • Cleaning and disinfection of reusable components
  • Packaging
  • Shelf-life

5.2 Clinical Testing

A multi-site, prospective study was conducted across 12 US sites and 1 OUS site, where a total of 213 participants with confirmed or suspected dysmotility issues of the gastric and/or colonic regions were recruited. The primary objective of this study was the assessment of the agreement of the Atmo Gas Capsule System and the Predicate measurements with regard to both GET and CTT. All participants were asked to ingest both the Atmo Gas Capsule System and the Predicate device concurrently. Bland-Altman analysis showed agreement between Atmo Gas Capsule System and the Predicate device for GET and CTT, with all pre-determined end points successfully met.

No serious adverse events (SAE) were associated with the Atmo Gas Capsule system. No device deficiencies led to SAEs or serious adverse device effects. Minimal adverse events (AE) were observed with all reported AEs being expected and common to the study population.

Based on the clinical performance as documented in the pivotal clinical study, the Atmo Gas Capsule System has a safety and effectiveness profile that is similar to the Predicate Device.

6 Substantial Equivalence

The Atmo Gas Capsule System is substantially equivalent to the SmartPill GI Monitoring System, version 2.0. The Atmo Gas Capsule System has the same intended uses and substantially similar indications, technological characteristics, and principles of operation as the Predicate device. The minor differences in indications do not alter the intended diagnostic use of the device and do not affect its safety and effectiveness when used as labelled. In addition, the minor technological differences between the Atmo Gas Capsule System and the Predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Atmo Gas Capsule System is as safe and effective as the SmartPill GI Monitoring System, version 2.0. Thus, the Atmo Gas Capsule System is substantially equivalent.

K250940
Page 4 of 4

§ 876.1725 Gastrointestinal motility monitoring system.

(a)
Identification. A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.(b)
Classification. Class II (performance standards).