(58 days)
Koala Intrauterine Pressure Catheter is indicated for use on patients requiring intrapartum intrauterine pressure monitoring.
The Koala Intrauterine Pressure Catheter is a disposable, sterile, single patient use intrauterine pressure monitoring catheter, designed for measurement of intrauterine contraction pressures. The catheter is designed with a pressure-sensing membrane cavity at the tip (mounted externally), a port for amnioinfusion and amniotic fluid sampling, and an introducer which is removed after placement. The Koala Intrauterine Pressure Catheters utilizes a reusable cable with a reusable pressure transducer which is connected at the other end of the device. The sensor, located on the tip of the catheter is a 360° sensor that registers pressure in all directions. The tip is made of soft, malleable urethane. There are no electronics in the catheter. Clinicians can disconnect the catheter while it is in the uterus and zero the monitor while the transducer is exposed to atmospheric pressure, providing a true zero reading. The Koala Intrauterine Pressure Catheter has two different, clear lumens, one for the amnioinfusion, to confirm proper placement via amniotic fluid flashback, and another channel for sensor-charging. The amnioinfusion port has a hydrophobic and a tethered replacement cap.
The FDA Clearance Letter for the Koala Intrauterine Pressure Catheter (IPC-5000E) indicates that the device has met the requirements for substantial equivalence to a predicate device. However, the provided document does not contain information about clinical studies with specific acceptance criteria, sample sizes for test/training sets, or expert involvement for ground truth adjudication, particularly in the context of AI/ML performance. This is because the device is a physical medical device (catheter) for measuring intrauterine pressure, not an AI/ML-driven software device.
Therefore, many of the requested points related to AI/ML performance evaluation (e.g., sample sizes for test/training sets, expert ground truth establishment, MRMC studies, standalone performance) are not applicable to this specific device as described in the provided clearance letter.
The document focuses on demonstrating substantial equivalence through:
- Comparison of intended use and technological characteristics with a predicate device.
- Non-clinical performance tests, including biocompatibility, sterility, shelf-life, and functional testing.
Below is a table summarizing the functional acceptance criteria that were met by the device during non-clinical testing, as described in the document.
1. Table of Acceptance Criteria and Reported Device Performance
While the document doesn't explicitly list a "table of acceptance criteria" with numerical values alongside "reported device performance" in the typical AI/ML sense (e.g., sensitivity/specificity targets), it states that "All device samples met their acceptance criteria" for various functional tests. These functional tests are the "performance" demonstrated.
| Acceptance Criteria Category | Specific Test | Reported Device Performance (as stated in document) |
|---|---|---|
| Biocompatibility | Cytotoxicity (ISO 10993-5) | Non-cytotoxic |
| Sensitization (ISO 10993-10) | Non-sensitizing | |
| Irritation/Intracutaneous Reactivity (ISO 10993-23) | Non-irritating | |
| Acute Systemic Toxicity (ISO 10993-11) | Not systemically toxic | |
| Material Mediated Pyrogenicity (ISO 10993-11, USP<151>) | Non-pyrogenic | |
| Sterility & Shelf-life | Gamma Radiation Sterilization & Validation (ISO 11737-1, ISO 11737-2, ANSI AAMI ST67, FDA Guidance) | Achieved sterility. All samples met acceptance criteria. |
| Simulated Transportation & Package Integrity (ASTM D4169-22 DC-13) | Maintained package integrity. All samples met acceptance criteria. | |
| Accelerated Aging (ASTM F1980:21) including: - Package Integrity (ASTM F1929-15, ASTM F88/F88M-21, ASTM F1886/F1886M-16) | Maintained package integrity and functional performance for proposed 6-month shelf-life. All samples met acceptance criteria. | |
| Functional Testing | Simulated Use (Balloon Pressure, Tip Pull-off force, Monofilament removal) | Functions as intended. All samples met acceptance criteria. |
| Pressure Accuracy | Functions as intended. All samples met acceptance criteria. | |
| Leak Rate/Membrane Integrity | Functions as intended. All samples met acceptance criteria. | |
| T-connector Bond Strength | Functions as intended. All samples met acceptance criteria. | |
| Tip Bond Strength | Functions as intended. All samples met acceptance criteria. | |
| Catheter Flexibility | Functions as intended. All samples met acceptance criteria. | |
| Introducer Peel-force | Functions as intended. All samples met acceptance criteria. | |
| Performance Specifications | Operation Pressure Range | -50 to +150mmHg (Same as predicate) |
| (with reusable cable) | Over-pressure Protection | -400 to +1200 mmHg (Same as predicate) |
| Operating Temperature | 15º C to 40º C (Same as predicate) | |
| Amnio Lumen Flow | Min. 20 cc/min at a pressure of 65 mmHg (Same as predicate) |
Note: The phrase "All device samples met their acceptance criteria" is used consistently, indicating successful performance against predetermined thresholds, even if numerical specifics for each functional test aren't detailed in this summary.
2. Sample size used for the test set and the data provenance:
- The document mentions "All device samples" for functional and shelf-life testing without specifying exact numerical sample sizes per test. These were non-clinical bench and material tests, not human patient data or a clinical test set in the AI/ML context.
- Data Provenance: Not applicable in the context of clinical data for AI/ML. The tests conducted are laboratory-based, demonstrating physical and material properties of the device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. The ground truth for this medical device (a catheter) is established through engineering specifications, material science, and established international standards (e.g., ISO, ASTM). It does not involve expert readers adjudicating clinical "ground truth" to evaluate an algorithm's performance on patient data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This applies to expert review of clinical cases, which was not part of this device's non-clinical testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a physical medical instrument, not an AI-assisted diagnostic tool. Therefore, MRMC studies involving human readers and AI assistance are irrelevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is not an algorithm; it is a physical catheter designed for direct measurement.
7. The type of ground truth used:
- The "ground truth" for this device's performance is based on engineering specifications, material properties, and adherence to established international and national standards (e.g., ISO 10993 for biocompatibility, ISO 11737 for sterility, ASTM standards for package integrity and material testing). It's a "known" physical and chemical standard, not derived from expert consensus or patient outcomes data in a diagnostic sense.
8. The sample size for the training set:
- Not applicable. This is a physical device, not an AI/ML model that requires a training set. The device's design is based on engineering principles and previous iterations (as indicated by the predicate device K974389).
9. How the ground truth for the training set was established:
- Not applicable. See point 8.
FDA Clearance Letter for Koala Intrauterine Pressure Catheter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
May 22, 2025
Clinical Innovations LLC
Alex Garrett
Regulatory Affairs Specialist
747 W 4170 S
Murray, Utah 84047
Re: K250897
Trade/Device Name: Koala Intrauterine Pressure Catheter (IPC-5000E)
Regulation Number: 21 CFR 884.2700
Regulation Name: Intrauterine Pressure Monitor and accessories
Regulatory Class: III
Product Code: HFN
Dated: March 24, 2025
Received: March 25, 2025
Dear Alex Garrett:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Page 2
K250897 - Alex Garrett Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K250897 - Alex Garrett Page 3
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K250897
Device Name
Koala Intrauterine Pressure Catheter (IPC-5000E)
Indications for Use (Describe)
Koala Intrauterine Pressure Catheter is indicated for use on patients requiring intrapartum intrauterine pressure monitoring.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20) Page 1 of 1
Page 5
510(k) Summary – K250897
Koala Intrauterine Pressure Catheter
Applicant: Clinical Innovations, LLC
747 West 4170 South,
Murray, UT 84123, USA
Contact Person: Alex Garrett
Regulatory Affairs Specialist II
Phone: 1(800) 522-6743
Email: agarrett@laborie.com
Date Prepared: May 19, 2025
Device Information
Device Trade Name: Koala Intrauterine Pressure Catheter (IPC-5000E)
Common Name: Intrauterine Pressure Catheter
Regulation Name: Intrauterine pressure monitor and accessories
Regulation Number: 21 CFR 884.2700
Product Code: HFN (Transducer, Pressure, Intrauterine)
Regulatory Class: Class II
Predicate Device Information
Device Name: Koala Intrauterine Pressure Catheter
510(k) Number: K974389
Manufacturer: Clinical Innovations LLC
The predicate device has not been subject to a design-related recall.
Device Description Summary
The Koala Intrauterine Pressure Catheter is a disposable, sterile, single patient use intrauterine pressure monitoring catheter, designed for measurement of intrauterine contraction pressures. The catheter is designed with a pressure-sensing membrane cavity at the tip (mounted externally), a port for amnioinfusion and amniotic fluid sampling, and an introducer which is removed after placement. The Koala Intrauterine Pressure Catheters utilizes a reusable cable with a reusable pressure transducer which is connected at the other end of the device.
The sensor, located on the tip of the catheter is a 360° sensor that registers pressure in all directions. The tip is made of soft, malleable urethane. There are no electronics in the catheter. Clinicians can disconnect the catheter while it is in the uterus and zero the monitor while the transducer is exposed to atmospheric pressure, providing a true zero reading.
Page 6
The Koala Intrauterine Pressure Catheter has two different, clear lumens, one for the amnioinfusion, to confirm proper placement via amniotic fluid flashback, and another channel for sensor-charging. The amnioinfusion port has a hydrophobic and a tethered replacement cap.
Indications for Use
Koala Intrauterine Pressure Catheter is indicated for use on patients requiring intrapartum intrauterine pressure monitoring.
Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Device & Predicate Device(s): | Subject Device (K250897) Koala Intrauterine Pressure Catheter (IPC 5000E) | Predicate Device (K974389) Koala Intrauterine Pressure Catheter | Comparison |
|---|---|---|---|
| Intended patient population | Patients requiring intrapartum, intrauterine pressure monitoring | Patients requiring intrapartum, intrauterine pressure monitoring | Same |
| Product Code | HFN | HFN | Same |
| Indication for Use | Koala Intrauterine Pressure Catheter is indicated for use on patients requiring intrapartum intrauterine pressure monitoring. | This catheter is for use on patients requiring intrapartum intrauterine pressure monitoring. | Similar |
| Prescription or OTC | Prescription | Prescription | Same |
| Principle of operation | Employs air coupling technology from a distally mounted flexible balloon in the uterus to an external, reusable transducer in the monitor cable and connector. | Employs air coupling technology from a distally mounted flexible balloon in the uterus to an external, reusable transducer in the monitor cable and connector. | Same |
| Air coupling channel | Lumen with integrated element | Single lumen with microfilament inside attached at the ends of the tube. | Different |
| Catheter Dimensions | Catheter Length: 30.5 Inches Catheter Outer Diameter: 0.136 Inches Catheter Inner Diameter: 0.103 Inches | Catheter Length: 30.5 Inches Catheter Outer Diameter: 0.136 Inches Catheter Inner Diameter: 0.103 Inches | Same |
| Use Condition provided | Sterile | Sterile | Same |
| Sterilization method | Gamma irradiation | Gamma irradiation | Same |
| Shelf-life | 6 months | 2 Years | Different |
| Use environment | Hospital or medical center | Hospital or medical center | Same |
| Storage conditions | -25º C to +50º C | -25º C to +65º C | Different |
Page 7
Performance specifications (with reusable cable)
| Subject Device | Predicate Device | Comparison | |
|---|---|---|---|
| Transducer Type | Silicone Diaphragm | Silicone Diaphragm | Same |
| Operation Pressure Range | -50 to +150mmHg | -50 to +150mmHg | Same |
| Over-pressure Protection | -400 to +1200 mmHg | -400 to +1200 mmHg | Same |
| Operating Temperature | 15º C to 40º C | 15º C to 40º C | Same |
| Amnio Lumen flow | Min. 20 cc/min at a pressure of 65 mmHg | Min. 20 cc/min at a pressure of 65 mmHg | Same |
Material composition
| Subject Device | Predicate Device | Comparison | |
|---|---|---|---|
| Tubing (Bump tube) | Polyurethane, Isothane 5065D | Polyurethane | Different |
| Monofilament | Removed (Polyurethane) | Nylon | Different |
| Adhesive | Cyanoacrylate | Cyanoacrylate | Same |
| Adhesive (UV) | Polyurethane Oligomer | Polyurethane Oligomer | Same |
| Band (elastic) | C-flex | C-flex | Same |
| Membrane | Polyethylene/ethylene vinyl acetate (external) | Polyethylene/ethylene vinyl acetate (external) | Same |
| Catheter Body | Polyurethane/Pebax, Isothane 5055D (as an alternative to Pellethane) | Polyurethane/Pebax (Polyamide based thermoplastic) co-extrusion. Pellethane 2363-55D. | Different |
| Connector ('T' Barb) | Polycarbonate | Polycarbonate | Same |
| Ink, Black, Visprox | Ink (added PrintSafe ink as an alternate material). | Ink | Different |
| Introducer | Ferro RxLOY polyolefin alloy, HIVAL 521054 | Polyethylene | Different |
| Cap | Polyvinyl Chloride | Polyvinyl Chloride | Same |
| Luer Cap (non- vented) | Polyethylene | Polyethylene | Same |
| Luer Cap (vented) | ABS | ABS | Same |
| Adhesive patch | Polyester | Polyester | Same |
The indications for use of the subject and predicate devices are similar; therefore, they have the same intended use (i.e., for intrapartum measurement of intrauterine pressure).
The subject and predicate devices have similar technological features, including device design, operating principle, interface, device dimensions etc. However, as shown in the table above, the subject and predicate devices differ in terms of minor design changes, materials, storage temperature and shelf-life. These differences of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness.
Non-Clinical Performance Tests Summary and Conclusions
Biocompatibility testing
Biocompatibility testing of the final finished device was conducted in accordance with the 2024 FDA guidance "Use of International standard ISO 10993-1, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process" for a surface devices contacting breached or
Page 8
compromised surfaces for a limited time (<24 hours) with the following endpoints:
- Cytotoxicity per ISO 10993-5
- Sensitization per ISO 10993-10
- Irritation or intracutaneous reactivity per ISO 10993-23
- Acute systemic toxicity per ISO 10993-11
- Material mediated pyrogenicity per ISO 10993-11, USP<151>
The subject device was found to be non-cytotoxic, non-sensitizing, non-irritating, not systemically toxic, and non-pyrogenic (material mediated).
The following testing was conducted to support sterility and shelf-life of the subject device:
- Gamma radiation sterilization and validation testing per ISO 11737-1:2015, ISO 11737-2:2015, ANSI AAMI ST67:2019 and the 2024 FDA guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile Guidance for Industry and Food and Drug Administration Staff".
- Simulated transportation and associated package integrity testing in accordance with ASTM D4169-22 DC-13.
- Shelf-life testing (accelerated aging) per ASTM F1980:21, including package integrity testing (dye penetration test per ASTM F1929-15, seal strength testing per ASTM F88/F88M-21 and visual assessment per ASTM F1886/F1886M-16) and mechanical performance, as outlined in the functional testing below.
All device samples met their acceptance criteria, demonstrating the sterility and package integrity is maintained through the transportation and proposed shelf-life of the subject device.
Functional Testing
Following testing was conducted to assess the device performance after initial manufacturing (T=0) and following accelerated aging (T=six months):
- Simulated use (Ballon Pressure and Tip Pull-off force, with monofilament removal)
- Pressure accuracy
- Leak rate/ Membrane integrity
- T-connector bond strength
- Tip bond strength
- Catheter flexibility
- Introducer peel-force
All device samples met their acceptance criteria, demonstrating that when manufactured to specification, the device functions as intended throughout its specified shelf-life.
Conclusion
The performance testing described above demonstrate that the Koala Intrauterine Pressure Catheter (IPC 5000E) is as safe and effective as the predicate device and supports determination of substantial equivalence.
§ 884.2700 Intrauterine pressure monitor and accessories.
(a)
Identification. An intrauterine pressure monitor is a device designed to detect and measure intrauterine and amniotic fluid pressure with a catheter placed transcervically into the uterine cavity. The device is used to monitor intensity, duration, and frequency of uterine contractions during labor. This generic type of device may include the following accessories: signal analysis and display equipment, patient and equipment supports, and component parts.(b)
Classification. Class II (performance standards).