K Number
K250856
Manufacturer
Date Cleared
2025-06-12

(83 days)

Product Code
Regulation Number
866.1640
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ETEST® is a manual, quantitative technique for the determination of antimicrobial susceptibility of non‑fastidious Gram‑negative and Gram‑positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in μg/mL) of different antimicrobial agents against microorganisms tested on agar media after overnight incubation.

Testing with ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) is indicated for Enterobacterales, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).

The ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) demonstrated acceptable performance with the following microorganisms:

Enterobacterales:

  • Escherichia coli
  • Klebsiella pneumoniae
  • Klebsiella aerogenes
  • Citrobacter freundii complex
  • Citrobacter koseri
  • Enterobacter cloacae complex
  • Proteus mirabilis
  • Proteus vulgaris
  • Morganella morganii
  • Providencia stuartii
  • Serratia marcescens
Device Description

ETEST® is a thin, inert and non-porous plastic strip carrying the MIC reading scale in μg/mL on one side and a predefined antibiotic gradient on the other side.

When the strip is applied to an inoculated agar surface, the preformed antibiotic gradient immediately transfers into the agar matrix, then forming a stable, continuous and exponential gradient of antibiotic concentrations directly underneath the strip. Bacterial growth becomes visible during incubation, and a symmetrical inhibition ellipse centered along the strip appears. The MIC value is read from the scale in terms of μg/mL at complete inhibition of bacterial growth, where the pointed end of the ellipse intersects the strip.

ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL) contains a range of Aztreonam from (0.016-256 µg/mL), overlaid with a fixed concentration of 4 µg/mL of Avibactam.

AI/ML Overview

The provided FDA 510(k) clearance letter describes the acceptance criteria and the study proving the ETEST Aztreonam/Avibactam (AZA) device meets these criteria for determining antimicrobial susceptibility.

Here's a breakdown of the requested information:


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for antimicrobial susceptibility test (AST) systems, as outlined by the FDA and CLSI (Clinical and Laboratory Standards Institute), are primarily based on Essential Agreement (EA) and Category Agreement (CA) with a reference method.

The acceptance criteria are implicitly met if the reported performance metrics achieve or exceed the established thresholds (though the exact numerical acceptance thresholds for EA and CA are not explicitly stated as "acceptance criteria" values in this document, standard FDA guidance for AST devices typically requires EA ≥ 90% and CA ≥ 90% for overall performance).

Performance MetricAcceptance Criteria (Implicit, based on typical FDA AST guidance)Reported Device Performance (ETEST Aztreonam/Avibactam)
Essential Agreement (EA) (overall species)≥ 90% (typical guidance)95.7% (within ± 1 dilution of the reference method)
Category Agreement (CA) (overall species)≥ 90% (typical guidance)98.0%
ReproducibilityAcceptable (e.g., ≥ 95% within defined range)Best case = 100%, Worst case = 98.5%
Quality Control> 95% within expected QC results range> 95% within the expected QC results range

Note: The document also states that the device "demonstrated substantially equivalent performance when compared with the CLSI M07-11th Ed (January 2018) broth microdilution reference method, following rules as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, issued on August 28, 2009 and following specifications as defined in CLSI M100 35th Ed. (January 2025)." This indicates adherence to established regulatory and standard-setting body guidelines for performance.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: A total of 602 strains of Enterobacterales were used for the performance evaluation.
    • This included:
      • Escherichia coli (150 strains)
      • Klebsiella pneumoniae (118 strains)
      • Enterobacter cloacae complex (67 strains, including E. cloacae, E. hormaechei, E. asburiae, E. roggenkampii, E. ludwigii)
      • Citrobacter freundii complex (50 strains, including C. freundii, C. braakii, C. murliniae, C. portucalensis)
      • Citrobacter koseri (30 strains)
      • Klebsiella aerogenes (32 strains)
      • Morganella morganii (30 strains)
      • Proteus mirabilis (32 strains)
      • Providencia stuartii (30 strains)
      • Proteus vulgaris (30 strains)
      • Serratia marcescens (33 strains)
    • Additionally, 12 Enterobacterales resistant isolates with high MIC values (≥ 16 µg/mL) were specifically included: Klebsiella pneumoniae (2), Escherichia coli (8), Serratia marcescens (1) and Enterobacter cloacae complex (1).
  • Data Provenance: The document states "External evaluations were conducted with fresh and stock clinical isolates, as well as a set of challenge strains." This implies a mix of retrospective (stock clinical isolates, challenge strains) and prospective (fresh clinical isolates) data. The country of origin of the data is not specified in this document.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This document describes the validation of an in vitro diagnostic device for antimicrobial susceptibility testing, not an imaging AI device requiring expert human reads. Therefore, the concept of "experts" to establish ground truth in the context of human readers is not directly applicable.

Instead, the ground truth was established by a reference method: the CLSI (Clinical and Laboratory Standards Institute) M07-11th Ed (January 2018) broth microdilution reference method. This method is considered the gold standard for determining Minimum Inhibitory Concentrations (MICs) in microbiology. The qualifications of personnel performing this reference method would be standard microbiology laboratory technicians or scientists trained in CLSI methods, rather than clinical experts like radiologists. The number of individuals/laboratories involved in performing the reference method is not specified.


4. Adjudication Method for the Test Set

Not applicable in the context of in vitro diagnostic device validation using a reference laboratory method. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies where human readers provide subjective interpretations, and conflicts need resolution. Here, the ground truth is a quantitative, standardized microbiology assay.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI Assistance

Not applicable. This is an in vitro diagnostic device (a test strip) for determining antimicrobial susceptibility, not an AI-assisted diagnostic tool for human image interpretation. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted.


6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The device itself is a manual, quantitative technique. It is a physical strip that produces a visual inhibition zone from which a Bacterial MIC value is read. It does not involve a software algorithm in the way an AI diagnostic tool would. Therefore, the concept of "standalone performance" of an "algorithm" is not directly applicable in the same sense as for an AI product.

The performance evaluation (which can be considered analogous to "standalone performance" in that it assesses the device's accuracy without a human "interpretation" component beyond reading the scale) was done by comparing the ETEST Aztreonam/Avibactam (AZA) strip's results to the CLSI broth microdilution reference method. This demonstrates the device's inherent capability to accurately determine MIC values.


7. The Type of Ground Truth Used

The ground truth used was the CLSI M07-11th Ed (January 2018) broth microdilution reference method. This is a highly standardized and accepted laboratory reference method for determining the Minimum Inhibitory Concentration (MIC) of an antimicrobial agent against a microorganism. It serves as the definitive 'truth' against which the performance of the ETEST device is measured.


8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning model development. This is an in vitro diagnostic device, not a software algorithm that undergoes a distinct training phase. The development of such devices typically involves extensive R&D, formulation optimization, and internal testing before external clinical evaluations. The "training" in this context would refer to the development and optimization of the test strip's chemical gradient to accurately reflect MIC values, rather than training a data-driven model. The document focuses on the validation (test set) for regulatory clearance.


9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a "training set" and associated ground truth establishment methods for it, as typically applied to machine learning, is not directly relevant to this type of in vitro diagnostic device. The 'ground truth' for validating the performance of this device is consistently the CLSI broth microdilution reference method, which is the gold standard for antimicrobial susceptibility testing. Any internal development or optimization of the ETEST strip formulation would also rely on this or similar established reference methods to guide product development.

FDA 510(k) Clearance Letter - ETEST Aztreonam/Avibactam (AZA)

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.07.05
Silver Spring, MD 20993
www.fda.gov

June 12, 2025

bioMérieux
Stéphanie Millon-Serme
Regulatory Affairs Specialist
376 chemin de l'Orme
Marcy-l'Etoile, 69280
France

Re: K250856
Trade/Device Name: ETEST Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)
Regulation Number: 21 CFR 866.1640
Regulation Name: Antimicrobial Susceptibility Test Powder
Regulatory Class: Class II
Product Code: JWY
Dated: March 20, 2025
Received: March 21, 2025

Dear Stéphanie Millon-Serme:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an

Page 2

K250856 - Stéphanie Millon-Serme Page 2

established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Page 3

K250856 - Stéphanie Millon-Serme Page 3

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D. (ABMM)
Chief
General Bacteriology and Antimicrobial Susceptibility Branch
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K250856

Device Name
ETEST Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)

Indications for Use (Describe)

ETEST is a manual, quantitative technique for the determination of antimicrobial susceptibility of non-fastidious Gram-negative and Gram-positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in μg/mL) of different antimicrobial agents against microorganisms tested on agar media after overnight incubation.

Testing with ETEST Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) is indicated for Enterobacterales, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).

The ETEST Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) demonstrated acceptable performance with the following microorganisms:

Enterobacterales:

  • Escherichia coli
  • Klebsiella pneumoniae
  • Klebsiella aerogenes
  • Citrobacter freundii complex
  • Citrobacter koseri
  • Enterobacter cloacae complex
  • Proteus mirabilis
  • Proteus vulgaris
  • Morganella morganii
  • Providencia stuartii
  • Serratia marcescens

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)

A. 510(k) Submission Information:

Submitter's Name: bioMerieux SA

Address: 376 Chemin de l'Orme
69280 Marcy-l'Etoile, FRANCE

Contact Person: Stéphanie MILLON-SERME
Regulatory Affairs Specialist

Phone Number: (314) 731-8666

Date of Preparation: June, 11th, 2025

B. Device Name:

Formal/Trade Name: ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)

Classification Name: 21 CFR 866.1640
Manual Antimicrobial Susceptibility Test Systems
Product Code: JWY

Common Name(s): ETEST® Aztreonam/Avibactam (AZA)

C. Predicate Device:

ETEST® Ceftazidime/Avibactam (CZA) (0.016-256/4 µg/mL)

1/7

Page 6

D. Device Description:

ETEST® is a thin, inert and non-porous plastic strip carrying the MIC reading scale in μg/mL on one side and a predefined antibiotic gradient on the other side.

When the strip is applied to an inoculated agar surface, the preformed antibiotic gradient immediately transfers into the agar matrix, then forming a stable, continuous and exponential gradient of antibiotic concentrations directly underneath the strip. Bacterial growth becomes visible during incubation, and a symmetrical inhibition ellipse centered along the strip appears. The MIC value is read from the scale in terms of μg/mL at complete inhibition of bacterial growth, where the pointed end of the ellipse intersects the strip.

ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL) contains a range of Aztreonam from (0.016-256 µg/mL), overlaid with a fixed concentration of 4 µg/mL of Avibactam.

E. Intended Use:

ETEST® is a manual, quantitative technique for the determination of antimicrobial susceptibility of non‑fastidious Gram‑negative and Gram‑positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in μg/mL) of different antimicrobial agents against microorganisms tested on agar media after overnight incubation.

Testing with ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) is indicated for Enterobacterales, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).

The ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) demonstrated acceptable performance with the following microorganisms:

Enterobacterales:

  • Escherichia coli
  • Klebsiella pneumoniae
  • Klebsiella aerogenes
  • Citrobacter freundii complex
  • Citrobacter koseri
  • Enterobacter cloacae complex
  • Proteus mirabilis
  • Proteus vulgaris
  • Morganella morganii
  • Providencia stuartii
  • Serratia marcescens

2/7

Page 7

F. Summary of the technological characteristics of the new device in comparison to those of the predicate device.

A summary of the similarities and differences of important features of the device and predicate are provided in the table below:

Table 1 - Similarities and Differences of Important Features of the Device and Predicate

ItemDeviceETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)PredicateETEST® Ceftazidime/Avibactam (CZA) (0.016-256/4 µg/mL)
General Device CharacteristicsSimilarities
Intended UseETEST® is a manual, quantitative technique for the determination of antimicrobial susceptibility of non‑fastidious Gram‑negative and Gram‑positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in μg/mL) of different antimicrobial agents against microorganisms tested on agar media after overnight incubation.Testing with ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) is indicated for Enterobacterales, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC)The ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) demonstrated acceptable performance with the following microorganisms:• Enterobacterales:◦ Escherichia coli◦ Klebsiella pneumoniae◦ Klebsiella aerogenes◦ Citrobacter freundiiETEST® is a manual, quantitative technique for the determination of antimicrobial susceptibility of non‑fastidious Gram‑negative and Gram‑positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in μg/mL) of different antimicrobial agents against microorganisms tested on agar media after overnight incubation.Ceftazidime/Avibactam has been shown to be active against the Gram-negative aerobic microorganisms listed below according to the FDA label for this antimicrobial agent.ETEST® CZA can be used to determine the MIC of Ceftazidime/Avibactam against the following microorganisms: (as listed in the "Differences" section)

3/7

Page 8

ItemDeviceETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)PredicateETEST® Ceftazidime/Avibactam (CZA) (0.016-256/4 µg/mL)
complex◦ Citrobacter koseri◦ Enterobacter cloacae complex◦ Proteus mirabilis◦ Proteus vulgaris◦ Morganella morganii◦ Providencia stuartii◦ Serratia marcescens
Test MethodologyQuantitative antimicrobial susceptibility test to determine the in vitro susceptibility of microorganismsSame
InoculumSuspension of organismSame
Type of testQuantitativeSame
Clinical & Challenge Performance DataEssential Agreement & Category Agreement (overall species)EA = 95.7%CA = 98.0%Essential Agreement & Category Agreement (overall species)EA = 99.1%CA = 99.6%
ReproducibilityBest case = 100%Worst case = 98.5 %Best case = 100 %Worst case = 100 %
Quality Controlwithin the expected QC results range >95% of the timewithin the expected QC results range >95% of the time
Meets Guidance Document Performance RequirementsYesYes
Differences
Antimicrobial AgentAztreonam/AvibactamCeftazidime/Avibactam
ConcentrationsAztreonam: 0.016 to 256 μg/mLAvibactam: 4 μg/mLCeftazidime: 0.016 to 256 μg/mLAvibactam: 4 μg/mL
Indicated organismsTesting with ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) is indicated for Enterobacterales, as recognizedCeftazidime/Avibactam has been shown to be active against the Gram-negative aerobic microorganisms listed below according to the FDA label for this antimicrobial agent.

4/7

Page 9

ItemDeviceETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)PredicateETEST® Ceftazidime/Avibactam (CZA) (0.016-256/4 µg/mL)
by the FDA Susceptibility Test Interpretive Criteria (STIC)The ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) demonstrated acceptable performance with the following microorganisms:• Enterobacterales:◦ Escherichia coli◦ Klebsiella pneumoniae◦ Klebsiella aerogenes◦ Citrobacter freundii complex◦ Citrobacter koseri◦ Enterobacter cloacae complex◦ Proteus mirabilis◦ Proteus vulgaris◦ Morganella morganii◦ Providencia stuartii◦ Serratia marcescensETEST® CZA can be used to determine the MIC of Ceftazidime/Avibactam against the following microorganisms:Active both in vitro and in clinical infections:▪Gram-negative aerobes:▪Enterobacterales:▪Citrobacter freundii▪Enterobacter cloacae▪Escherichia coli▪Klebsiella oxytoca▪Klebsiella pneumoniae▪Proteus mirabilis▪Pseudomonas aeruginosaIn vitro data are available for the following microorganisms, but clinical significance is unknown:•Citrobacter koseri•Enterobacter aerogenes•Morganella morganii•Providencia rettgeri•Providencia stuartii•Serratia marcescens
BreakpointsEnterobacterales:S ≤ 4/4I = 8/4R ≥ 16/4Enterobacterales:S≤ 8/4I= N/AR≥ 16/4Pseudomonas aeruginosa:S≤ 8/4I= N/AR≥ 16/4

5/7

Page 10

G. Performance Overview

ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL) demonstrated substantially equivalent performance when compared with the CLSI M07-11th Ed (January 2018) broth microdilution reference method, following rules as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, issued on August 28, 2009 and following specifications as defined in CLSI M100 35th Ed. (January 2025).

This Premarket Notification (510[k]) presents data in support of ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL).

External evaluations were conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to establish the performance of ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL) by comparing with the CLSI broth microdilution reference method.

ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL) demonstrated acceptable performance as presented in the table below:

Table 2- Performance Characteristics for ETEST® Aztreonam/Avibactam (AZA)

OrganismsStrains (N)% Essential Agreement (EA) a)% Category Agreement (CA)
Enterobacterales b), c), d), e)60295.798.0

Notes:
a) EA = % of MIC values within ± 1 dilution of the reference method.
b) The performance data presented for Enterobacterales include the following organisms: Escherichia coli (150), Klebsiella pneumoniae (118), Enterobacter cloacae complex* (67), Citrobacter freundii complex** (50), Citrobacter koseri (30), Klebsiella aerogenes (32), Morganella morganii (30), Proteus mirabilis (32), Providencia stuartii (30), Proteus vulgaris (30) and Serratia marcescens (33).

    • Enterobacter cloacae complex includes Enterobacter cloacae, Enterobacter hormaechei, Enterobacter asburiae, Enterobacter roggenkampii and Enterobacter ludwigii.
  • **Citrobacter freundii complex includes Citrobacter freundii, Citrobacter braakii, Citrobacter murliniae and Citrobacter portucalensis.

6/7

Page 11

c) A total of 12 Enterobacterales resistant isolates with high MIC values ≥ 16 µg/mL were included in the ETEST® Aztreonam/ Avibactam performance evaluation studies: Klebsiella pneumoniae (2), Escherichia coli (8), Serratia marcescens (1) and Enterobacter cloacae complex (1).

d) In the ETEST® Aztreonam/Avibactam clinical studies, swabs were used for plate inoculation/streaking and forceps were used for ETEST® strip application. Testing with the optional Inoculator RETRO C80™, Vacuum Pen NEMA C88™, and Applicator SIMPLEX C76™ was not evaluated during the clinical studies.

e) ETEST® Aztreonam/Avibactam MIC values tended to be in exact agreement or at least one doubling dilution lower when testing Klebsiella pneumoniae, Morganella morganii, Proteus vulgaris and Providencia stuartii compared to the CLSI reference broth microdilution method.

Limitations

Due to unacceptable essential agreement, Klebsiella oxytoca should not be tested with the ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 μg/mL) and should be tested by an alternative method.

The ability of ETEST® Aztreonam/Avibactam to detect the following resistant isolates is unknown because an insufficient number of resistant isolates were available at the time of comparative testing: Citrobacter freundii complex, Citrobacter koseri, Enterobacter cloacae complex, Klebsiella aerogenes, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Providencia stuartii and Serratia marcescens.

Reproducibility and Quality Control demonstrated acceptable results.

Conclusion:

The performance data presented in this submission support a substantial equivalence decision. ETEST® Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL) is substantially equivalent to ETEST® Ceftazidime/Avibactam (CZA) (0.016-256/4 µg/mL) (K172150).

7/7

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).