(29 days)
The Deepblue Multi-Drug Urine Test Cup is a rapid lateral flow immunoassay for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, fentanyl, Methadone, d-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off(ng/mL) |
|---|---|---|
| 6-MAM | 6-Monoacetylmorphine | 10 |
| AMP | d-Amphetamine | 500 or 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 150 or 300 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrolidine | 300 |
| FYL | Fentanyl | 1 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | d-Methamphetamine | 500 or 1000 |
| MTD | d/l-Methadone | 300 |
| MOP/OPI | Morphine | 300 or 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | d-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-Δ9-THC-COOH | 50 |
The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Deepblue Home Multi-Drug Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of one or more of the following drugs.
| Code | Substance | Cut-off(ng/mL) |
|---|---|---|
| AMP | Amphetamine | 1000 or 500 |
| BUP | Buprenorphine | 10 |
| BAR | Secobarbital | 300 |
| BZO | Oxazepam | 300 |
| COC | Cocaine | 300 or 150 |
| EDDP | EDDP | 300 |
| FYL | Fentanyl | 1 |
| MET | Methamphetamine | 1000 or 500 |
| MDMA | Ecstasy | 500 |
| OPI | Morphine | 2000 or 300 |
| MTD | Methadone | 300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | Marijuana | 50 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
This drug test cup may contain any combination of the drug tests listed in the table above.
This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
Deepblue Home Muti-Drug Urine Test Cup and Deepblue Muti-Drug Urine Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
The provided FDA 510(k) clearance letter details the performance characteristics of the Deepblue Multi-Drug Urine Test Cup and Deepblue Home Multi-Drug Urine Test Cup. This information allows us to describe the acceptance criteria and the study that proves the device meets these criteria.
It's important to note that this device is a qualitative lateral flow immunoassay for initial drug screening, not a diagnostic imaging AI, so the criteria and study methodology differ significantly from those for an AI-powered diagnostic tool. Specifically, there are no references to AI assistance, human readers, or image adjudication, as these are not relevant to this type of device.
Acceptance Criteria and Device Performance for Deepblue Multi-Drug Urine Test Cups
The acceptance criteria for this type of qualitative immunoassay are primarily based on its analytical performance, specifically its ability to correctly identify the presence or absence of target drugs at specified cutoff concentrations. This is demonstrated through precision/reproducibility studies, analytical specificity (cross-reactivity and interference), and method comparison studies against a gold standard (LC-MS/MS).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly derived from the successful demonstration of performance in the analytical and method comparison studies. The goal is to achieve high agreement rates with the confirmatory method (LC-MS/MS) and consistent results at and around the cutoff concentrations.
| Performance Metric Category | Specific Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Precision/Reproducibility | Consistent results across multiple lots and runs, especially at and near cutoff concentrations. High percentage of correct results for spiked samples at various concentrations. | Across three lots and 25 days of testing (50 runs per concentration), the device showed excellent performance. For concentrations at +50%, +75%, and +100% of cutoff, all results were positive (0-/50+). For concentrations at -50%, -75%, and -100% of cutoff, all results were negative (50-/0+). At the cutoff concentration, there was a mix of positive and negative results, indicating the assay's sensitivity at the threshold (e.g., AMP 1000: Lot 1 had 8-/42+, meaning 8 negatives and 42 positives out of 50 runs). Near the cutoff (e.g., -25% and +25%), expected mixed results were observed, demonstrating the assay's ability to differentiate. |
| Analytical Specificity (Cross-Reactivity) | Minimal cross-reactivity with non-target compounds and sufficient cross-reactivity with known metabolites/analogs to ensure broad detection. | Detailed tables provided showing specific cross-reactivity percentages for various compounds. For example, for AMP 1000, D,L-Amphetamine and D-Amphetamine showed 100% cross-reactivity as expected. For BAR 300, Alphenal showed 200% cross-reactivity, and Phenobarbital showed 150%. For FYL 1, Acetyl fentanyl showed 100%, and Butyryl Fentanyl showed 50%. Norfentanyl showed <0.001%. Many structurally unrelated compounds showed <1% cross-reactivity, demonstrating high specificity. |
| Analytical Specificity (Interference) | No significant interference from common exogenous or endogenous substances (e.g., pH, specific gravity, other medications) that would alter results. | pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 did not affect assay results. Over 100 non-structurally related compounds (e.g., Acetaminophen, Ibuprofen, Caffeine, etc.) showed no interference at 100 µg/mL. |
| Method Comparison (Concordance with LC-MS/MS) | High agreement (concordance) rates with a validated confirmatory method (LC-MS/MS) for both negative and positive samples, especially around the cutoff. | High agreement was observed. For each drug, 80 clinical samples (40 negative, 40 positive based on LC-MS/MS) were tested by three operators. The detailed tables show the number of results identified as positive (+) or negative (-) by each operator compared to the LC-MS/MS classification (Drug-Free, Low Negative, Near Cutoff Negative, Near Cutoff Positive, High Positive). For example, for AMP (1000), Operator A correctly identified 10 drug-free samples as negative, 14 low negative as negative, and 22 high positive as positive. There were discordant results for samples near the cutoff, which is expected for qualitative assays; these are detailed in the "Discordant Results" table. |
| Lay Person Study Acceptance | High agreement rates with expected results when tested by lay users, and clear, easy-to-understand instructions. | High agreement rates were demonstrated across all drug types and concentrations in the lay person study (e.g., AMP 500 showed 90-100% agreement depending on concentration closest to cutoff). All 280 participants found the instructions easy to understand and follow. Flesch-Kincaid read-ability score of Grade 7. |
2. Sample Size Used for the Test Set and Data Provenance
- Analytical Performance (Precision): For each drug, 50 runs were performed per concentration (9 concentrations tested: +100%, +75%, +50%, +25%, cutoff, -25%, -50%, -75%, -100%). This involved using 3 lots of test cups, with 2 runs per day for 25 days. The sample provenance is not explicitly stated as patient data, but rather drug-free urine spiked with controlled concentrations of target drugs. This is typical for analytical validation.
- Analytical Specificity: Urine samples were spiked with various drug metabolites, analogs, and interfering substances. The number of samples per compound tested is not explicitly stated but implied to be sufficient for demonstrating cross-reactivity thresholds.
- Method Comparison Study: For each drug, 80 unaltered urine clinical samples were used (40 negative, 40 positive). The data provenance is "in-house," implying these were from a controlled setting, likely in China where the submitter is located. It is labeled as "clinical samples," suggesting they were derived from human urine. The study was retrospective in nature, as pre-collected samples were tested.
- Lay Person Study: 280 lay persons participated. The samples were prepared by spiking drug-free pooled urine specimens with known drug concentrations and blind-labeled.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- For Analytical Performance (Precision & Specificity): The ground truth was established by precise spiking of drugs into drug-free urine samples. The concentrations were confirmed by LC-MS/MS, an advanced analytical chemistry technique. LC-MS/MS is considered a gold standard for quantifying drug concentrations and establishing ground truth in these types of studies, requiring skilled analytical chemists or lab technicians to perform. The number of experts or their specific qualifications (e.g., number of years of experience) are not provided in the document, but the use of LC-MS/MS inherently implies expert execution and interpretation.
- For Method Comparison Study: The ground truth for the 80 clinical urine samples per drug was established by LC-MS/MS results. These results served as the reference against which the device performance was compared. As above, this relies on the expertise of those performing and interpreting LC-MS/MS.
- For Lay Person Study: The ground truth was based on precisely prepared spiked samples with confirmed concentrations by LC-MS/MS.
4. Adjudication Method for the Test Set
- For Analytical Performance (Precision & Specificity): No multi-reader adjudication method (e.g., 2+1, 3+1) is described as relevant. The results are quantitative (spiked concentrations) confirmed by LC-MS/MS, and the device provides a qualitative output (positive/negative) that is directly compared to the expected qualitative outcome from the established concentration and cutoff.
- For Method Comparison Study: Three operators performed the tests. The results from each operator are reported individually ("Operator A," "Operator B," "Operator C"). Discordant results (where the device outcome did not match the LC-MS/MS ground truth, or where operators' results differed) are explicitly listed. There is no explicit mention of an adjudication process among the operators; rather, the data from each operator is presented against the LC-MS/MS reference. This indicates that their individual interpretations are the "reads," and the LC-MS/MS is the adjudication method.
- For Lay Person Study: No explicit adjudication method is described. The results from the lay persons were compared against the known concentrations of the spiked samples.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This type of study (MRMC) is typically performed for diagnostic imaging devices where human readers interpret images with and without AI assistance to assess the AI's impact on human performance in a clinical setting. The Deepblue device is a rapid lateral flow immunoassay for drug screening, not an imaging device or an AI-powered system designed to assist human interpretation. Therefore, this section is not applicable.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, in essence, a "standalone" or "device-only" performance study was done. The "Analytical Performance" section (Precision/Reproducibility and Analytical Specificity/Interference) and portions of the "Method Comparison Study" demonstrate the device's performance based on its inherent physical and chemical properties and its interaction with the sample, irrespective of human interpretation variability (though human operators handle the device and read the result). The "Method Comparison" explicitly compares the device's output to the gold standard (LC-MS/MS) directly. The "Lay Person Study" assesses the ease of use and reads by untrained users, ensuring the device works as intended without specialized human expertise. Since it's a simple qualitative result (presence of a line), "human-in-the-loop" "performance" is primarily about correct reading of a physical characteristic.
7. The Type of Ground Truth Used
The primary type of ground truth used for both the analytical performance and method comparison studies was confirmatory analytical chemistry data, specifically Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS/MS). These methods are considered the gold standard for accurate and quantitative identification of drugs and their metabolites in biological samples.
8. The Sample Size for the Training Set
No specific "training set" or "training" is described, as this device does not use machine learning or AI that requires a distinct data training phase. The development of the immunoassay itself relies on chemical and biological design principles, not data-driven model training.
9. How the Ground Truth for the Training Set Was Established
As there is no "training set" in the context of an AI/ML model for this device, a corresponding ground truth establishment process is not applicable. The immunoassay itself is developed and validated through iterative biochemical and engineering processes to ensure sensitivity and specificity to the target analytes.
FDA 510(k) Clearance Letter - K250803
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
April 15, 2025
ANHUI Deepblue Medical Technological Co., Ltd.
℅ Jenny Xia
Director
LSI International, Inc.
504 East Diamond Ave, Suite H
Gaithersburg, Maryland 20877
Re: K250803
Trade/Device Name: Deepblue Multi-Drug Urine Test Cup; Deepblue Home Muti-Drug Urine Test Cup
Regulation Number: 21 CFR 862.3650
Regulation Name: Opiate test system
Regulatory Class: Class II
Product Code: NGL, PTH, NFT, NFV, NFY, PTG, NGG, NGM, QBF, QAW, NFW
Dated: March 16, 2025
Received: March 17, 2025
Dear Jenny Xia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
Page 2
K250803 - Jenny Xia Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K250803 - Jenny Xia Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOSEPH A. KOTAREK -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2025.04.15 15:16:37 -04'00'
Joseph Kotarek
Branch Chief for Toxicology
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 2
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K250803
Device Name: Deepblue Multi-Drug Urine Test Cup; Deepblue Home Multi-Drug Urine Test Cup
Indications for Use (Describe)
The Deepblue Multi-Drug Urine Test Cup is a rapid lateral flow immunoassay for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, fentanyl, Methadone, d-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off(ng/mL) |
|---|---|---|
| 6-MAM | 6-Monoacetylmorphine | 10 |
| AMP | d-Amphetamine | 500 or 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 150 or 300 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrolidine | 300 |
| FYL | Fentanyl | 1 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | d-Methamphetamine | 500 or 1000 |
| MTD | d/l-Methadone | 300 |
| MOP/OPI | Morphine | 300 or 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | d-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-Δ9-THC-COOH | 50 |
The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Deepblue Home Multi-Drug Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of one or more of the following drugs.
| Code | Substance | Cut-off(ng/mL) |
|---|---|---|
| AMP | Amphetamine | 1000 or 500 |
| BUP | Buprenorphine | 10 |
| BAR | Secobarbital | 300 |
| BZO | Oxazepam | 300 |
| COC | Cocaine | 300 or 150 |
| EDDP | EDDP | 300 |
Page 5
FORM FDA 3881 (8/23) Page 2 of 2
| FYL | Fentanyl | 1 |
|---|---|---|
| MET | Methamphetamine | 1000 or 500 |
| MDMA | Ecstasy | 500 |
| OPI | Morphine | 2000 or 300 |
| MTD | Methadone | 300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | Marijuana | 50 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
This drug test cup may contain any combination of the drug tests listed in the table above.
This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 6
510(k) SUMMARY
- Date: March 9, 2025
- Submitter: ANHUI Deepblue Medical Technological Co., Ltd.
Pearl Industrial Park, 106 Innovation Ave.
Hefei, Anhui 230088
China - Contact person: Jenny Xia
LSI International Inc.
504 East Diamond Ave., Suite H
Gaithersburg, MD 20877
Telephone: 301-525-6856
Email: jxia@lsi-consulting.org - Device Name: Deepblue Multi-Drug Urine Test Cup
Deepblue Home Muti-Drug Urine Test Cup - Classification: Class II
| Product Code | Target Drug | Regulation Section | Panel |
|---|---|---|---|
| NFT | Amphetamine (AMP) | 862.3100, Amphetamine Test System | Toxicology |
| PTH | Secobarbital (BAR) | 862.3150, Barbiturate Test System | Toxicology |
| NGL | Buprenorphine (BUP)Morphine (OPI)Oxycodone (OXY)6-Monoacetylmorphine(6-MAM)Fentanyl (FEN) | 862.3650, Opiate Test System | Toxicology |
| NFV | Oxazepam (BZO) | 862.3170, Benzodiazepine Test System | Toxicology |
| NFY | Cocaine (COC) | 862.3250, Cocaine and cocaine metabolite test system | Toxicology |
| PTG | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)Methadone (MTD) | 862.3620, Methadone Test System | Toxicology |
| NGG | Methylenedioxymethamphetamine (MDMA)Methamphetamine (MET) | 862.3610, Methamphetamine Test System | Toxicology |
| NGM | Unclassified | Toxicology |
Page 7
| Phencyclidine (PCP) | ||
|---|---|---|
| QBF | Propoxyphene(PPX) | 862.3700 Propoxyphene test system. |
| QAW | Nortriptyline (TCA) | 862.3910 Tricyclic antidepressant drugs test system |
| NFW | Cannabinoids (THC 50) | 862.3870, Cannabinoids Test System |
-
Predicate Devices:
Healgen Accurate Urine Drug Screen Dip Card (K240686) -
Intended Use
The Deepblue Multi-Drug Urine Test Cup is a rapid lateral flow immunoassay for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, fentanyl, Methadone, d-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| 6-MAM | 6-Monoacetylmorphine | 10 |
| AMP | d-Amphetamine | 500 / 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 150 / 300 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrolidine | 300 |
| FYL | Fentanyl | 1 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | d-Methamphetamine | 500 / 1000 |
| MTD | d/l-Methadone | 300 |
| OPI | Morphine | 300 / 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | d-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-∆9-THC-COOH | 50 |
The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-
Page 8
spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Deepblue Home Multi-Drug Urine Test Cup is a rapid qualitative immunoassay.
The device provides preliminary results for the detection of one or more of the following drugs.
| CODE | SUBSTANCE | CUT-OFF (ng/mL) |
|---|---|---|
| AMP | Amphetamine | 1000 or 500 |
| BUP | Buprenorphine | 10 |
| BAR | Secobarbital | 300 |
| BZO | Oxazepam | 300 |
| COC | Cocaine | 300 or 150 |
| EDDP | EDDP | 300 |
| FYL | Fentanyl | 1 |
| MET | Methamphetamine | 1000 or 500 |
| MDMA | Ecstasy | 500 |
| OPI | Morphine | 2000 or 300 |
| MTD | Methadone | 300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | Marijuana | 50 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
This drug test cup may contain any combination of the drug tests listed in the table above.
This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
- Device Description
Deepblue Home Muti-Drug Urine Test Cup and Deepblue Muti-Drug Urine Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
- Substantial Equivalence Information
Page 9
Similarities
| Item | Device | Predicate (K240686) |
|---|---|---|
| Intended use | Qualitative detection of drugs of abuse in urine. For prescription use or over-the-counter use | Same. |
| Methodology | Competitive binding, lateral flow immunochromatographic assay based on antigen-antibody reaction | Same |
| Type of Test | Qualitative | Same |
| Specimen Type | Human urine | Same |
| Target Drug and Cut Off Values | Target Drugs - Cutoff (ng/mL) Amphetamine(AMP) - 1000 or 500Secobarbital (BAR) - 300Buprenorphine (BUP) - 10Oxazepam (BZO) - 300Cocaine (COC) - 300 or 1502-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) - 300Fentanyl - 1Methylenedioxymethamphetamine (MDMA) - 500Methamphetamine (MET) - 1000 or 500Morphine (MOP300/OPI2000) - 2000 or 300Methadone (MTD) - 300Oxycodone (OXY) - 100Phencyclidine (PCP) - 25Propoxyphene(PPX) - 300Nortriptyline (TCA) - 1000Cannabinoids (THC) - 506-Monoacetylmorphine(6-MAM) - 10 | Same except no Fentanyl |
Differences
| Configurations | Test cup | Test Dip Card |
|---|
Page 10
-
Standard/Guidance Document Reference (if applicable)
None referenced. -
Test Principle
Deepblue Home Muti-Drug Urine Test Cup or Deepblue Muti-Drug Urine Test Cup is a competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody conjugate binding sites.
When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug within the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.
When sample drug levels are at or above the detection level of the test, the free drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.
To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.
- Performance Characteristics
A. Analytical performance
a. Precision/Reproducibility:
Precision studies were carried out for samples with concentrations of +100% cutoff, +75% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Other samples were prepared by spiked target drug in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test cups. The results obtained are summarized in the following tables:
| Drug | Lot Number | +100% cutoff | +75% cutoff | +50% cutoff | +25% cutoff | Cutoff | -25% cutoff | -50% cutoff | -75% cutoff | -100% cut-off |
|---|---|---|---|---|---|---|---|---|---|---|
| AMP 1000 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 8-/42+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 4-/46+ | 11-/39+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 7-/43+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| BAR 300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 4--/46+ | 17-/33+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 14-/36+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 15-/35+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
Page 11
| BUP 10 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 19-/31+ | 46-/4+ | 50-/0+ | 50-/0+ | 50-/0+ |
|---|---|---|---|---|---|---|---|---|---|---|
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 18-/32+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 18-/32+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| BZO 300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 15-/35+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 16-/34+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 14-/36+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| COC 300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 9-/41+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 8-/42+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 12-/38+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| EDDP 300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 10-/40+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 10-/40+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 10-/40+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| FYL 1 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 18-/32+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 4-/46+ | 20-/30+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 18-/32+ | 46-/4+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MDMA 500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 12-/38+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 14-/36+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 15-/35+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MET 1000 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 10-/40+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 8-/42+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 7-/43+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MOP 300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 4-/46+ | 10-/40+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 12-/38+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 13-/37+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MTD 300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 17-/33+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 19-/31+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 20-/30+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| OXY 100 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 10-/40+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 9-/41+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 10-/40+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| PCP 25 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 17-/33+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 20-/30+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 18-/32+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| PPX 300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 22-/28+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 24-/26+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 23-/27+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| TCA 1000 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 20-/30+ | 46-/4+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 18-/32+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 21-/29+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| THC 50 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 24-/26+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 4-/46+ | 26-/24+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
Page 12
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 4-/46+ | 22-/28+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
|---|---|---|---|---|---|---|---|---|---|---|
| 6-MAM 10 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 11-/39+ | 46-/4+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 14-/36+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 9-/41+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| AMP 500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 11-/39+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3--/47+ | 12-/38+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 11-/39+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| COC 150 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 12-/38+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 14-/36+ | 46-/4+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 10-/40+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MET 500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 9-/41+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 4-/46+ | 10-/40+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 8-/42+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| OPI 2000 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 12-/38+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 3-/47+ | 14-/36+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 4-/46+ | 15-/35+ | 47-/3+ | 50-/0+ | 50-/0+ | 50-/0+ |
b. Linearity/assay reportable range:
Not applicable. This device is intended for qualitative use only.
c. Stability:
The device is stable at 2-30℃ for 24 months based on real time stability study.
d. Analytical specificity/Interference:
To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine. These urine samples were tested using three lots of the device.
Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100. If no cross-reactivity was observed, the highest concentration tested is shown.
| Drug/Cutoff | Compound | Minimum concentration required to obtain a positive result (ng/mL) | % Cross-Reactivity |
|---|---|---|---|
| AMP 1000 | Hydroxyamphetamine | 8000 | 12.5% |
| (+/-)-Methylenedioxyamphetamine (MDA) | 10000 | 10% | |
| D,L-Amphetamine | 1,000 | 100% | |
| D-Amphetamine | 1,000 | 100% | |
| Diethylstilbestrol | >100000 | <1% |
Page 13
| L-Amphetamine | 50000 | 2% |
|---|---|---|
| Phentermine | 10000 | 10% |
| β-Phenylethylamine | 100000 | 1% |
| Tyramine | 100000 | 1% |
| p-Hydroxynorephedrine | 100000 | 1% |
| D,L-Norephedrine | 100000 | 1% |
| p-Hydroxyamphetamine | 100000 | 1% |
| D-Methamphetamine | >100000 | <1% |
| L-Methamphetamine | >100000 | <1% |
| Ephedrine hydrochloride | >100000 | <1% |
| (+/-)3,4-Methylenedioxymethamphetamine(MDMA) | >100000 | <1% |
| Phenylpropanolamine | >100000 | <1% |
| Benzphetamine | >100000 | <1% |
| L-Ephedrine | >100000 | <1% |
| L-Epinephrine | >100000 | <1% |
| D,L-Epinephrine | >100000 | <1% |
| (+/-)3,4-Methylenedioxyethylamphetamine (MDEA) | >100000 | <1% |
| BAR 300 | Alphenal | 150 | 200% |
|---|---|---|---|
| Amobarbital | 300 | 100% | |
| Aprobarbital | 300 | 100% | |
| Butabarbital | 100 | 300% | |
| Butethal | 100 | 300% | |
| Cyclopentobarbital | 500 | 60% | |
| Pentobarbital | 200 | 150% | |
| Phenobarbital | 200 | 150% | |
| Secobarbital | 300 | 100% | |
| Butalbital | 2500 | 12% |
| BUP 10 | Buprenorphine | 10 | 100% |
|---|---|---|---|
| Buprenorphine -3-D-Glucuronide | 10 | 100% | |
| Norbuprenorphine | 20 | 50% | |
| Norbuprenorphine-3-D-Glucuronide | 200 | 5% | |
| Morphine | >100000 | <0.01% |
Page 14
| Oxymorphone | >100000 | <0.01% |
|---|---|---|
| Hydromorphone | >100000 | <0.01% |
| BZO 300 | a-Hydroxyalprazolam | 1000 | 30% |
|---|---|---|---|
| Alprazolam | 200 | 150% | |
| Bromazepam | 1000 | 30% | |
| Chlordiazepoxide | 1500 | 20% | |
| Clobazam | 100 | 300% | |
| Clonazepam | 2500 | 12% | |
| Clorazepate Dipotassium | 200 | 150% | |
| Desalkylflurazepam | 500 | 60% | |
| Diazepam | 200 | 150% | |
| Estazolam | 500 | 60% | |
| Flunitrazepam | 400 | 75% | |
| D,L-Lorazepam | 1000 | 30% | |
| Midazolam | 10000 | 3% | |
| Nitrazepam | 100 | 300% | |
| Norchlordiazepoxide | 200 | 150% | |
| Nordiazepam | 200 | 150% | |
| Oxazepam | 300 | 100% | |
| R,S-Lorazepam glucuronide | 200 | 150% | |
| Temazepam | 100 | 300% | |
| Triazolam | 2500 | 12% | |
| Demoxepam | 2000 | 15% | |
| Flurazepam | 500 | 60% | |
| Delorazepam | 10000 | 3% |
| COC 300 | Benzoylecogonine | 300 | 100% |
|---|---|---|---|
| Cocaethylene | 12500 | 2.4% | |
| Cocaine hydrochloride | 200 | 150% | |
| Ecgonine | >100000 | <0.3% | |
| Norcocaine | 100000 | 0.3% | |
| Ecgonine methyl ester | >100000 | <0.3% |
| EDDP 300 | EDDP | 300 | 100% |
|---|---|---|---|
| Methadone | >100000 | <0.3% | |
| EMDP | >100000 | <0.3% |
Page 15
| Doxylamine | >100000 | <0.3% |
|---|---|---|
| Disopyramide | 100000 | 0.3% |
| LAAM (Levo-alpha-acetylmethadol) hydrochloride | >100000 | <0.3% |
| Alpha Methadol | >100000 | <0.3% |
| FYL 1 | Norfentanyl | >100000 | <0.001% |
|---|---|---|---|
| Fentanyl | 1 | 100% | |
| Acetyl fentanyl | 1 | 100% | |
| Acetyl Norfentanyl | >100000 | <0.001% | |
| (±)-β-Hydroxythiofentanyl HCl | 1 | 100% | |
| Acryl Fentanyl | 1 | 100% | |
| Butyryl Fentanyl | 2 | 50% | |
| Cis-d, I 3-Methylfentanyl | 50 | 2% | |
| Furanyl Fentanyl | 5 | 20% | |
| Para-fluoro butyrl Fentanyl (P-FBF) | 3 | 33.3% | |
| Para-fluoro Fentanyl | 3 | 33.3% | |
| 9-Hydroxy Risperidone | 10000 | 0.01% | |
| Alfentanil | 5000 | 0.02% | |
| Isobutyryl Fentanyl | 1 | 100% | |
| 4-Fluoro-isobutyryl Fentanyl | 50 | 2% | |
| Norcarfentail Oxalate | 10000 | 0.01% | |
| Remifentanil | >100000 | <0.001% | |
| Valeryl Fentanyl | 10 | 10% | |
| Thienyl Fentanyl | 1000 | 0.1% | |
| Trans-d, I 3-Methylfentanyl | 1000 | 0.1% | |
| Despropionyl fentanyl (4-ANPP) | 2500 | 0.04% | |
| MT-45 | 10000 | 0.01% | |
| Ocfentanil | 2 | 50% | |
| Risperidone | >100000 | <0.001% | |
| Sufentanil | 10 | 10% | |
| Carfentanil Oxalate | 500 | 0.2% | |
| Labetalol Hydrochloride | >100000 | <0.001% | |
| Trazodone | >100000 | <0.001% | |
| U-47700 | >100000 | <0.001% |
Page 16
| ω-1-Hydroxyfentanyl | 20000 | 0.005 |
|---|---|---|
| 6-Acetyl morphine | >100000 | <0.001% |
| Amphetamine | >100000 | <0.001% |
| Buprenorphine | >100000 | <0.001% |
| Buprenorphineglucuronide | >100000 | <0.001% |
| Codeine | >100000 | <0.001% |
| Dextromethorphan | >100000 | <0.001% |
| Dihydrocodeine | >100000 | <0.001% |
| EDDP | >100000 | <0.001% |
| EMDP | >100000 | <0.001% |
| Fluoxetine | >100000 | <0.001% |
| Heroin | >100000 | <0.001% |
| Hydrocodone | >100000 | <0.001% |
| Hydromorphone | >100000 | <0.001% |
| Ketamine | >100000 | <0.001% |
| Levorphanol | >100000 | <0.001% |
| Meperidine | >100000 | <0.001% |
| Methadone | >100000 | <0.001% |
| Morphine | >100000 | <0.001% |
| Morphine-3-glucuronide | >100000 | <0.001% |
| Naloxone | >100000 | <0.001% |
| Naltrexone | >100000 | <0.001% |
| Norbuprenorphine | >100000 | <0.001% |
| Norcodeine | >100000 | <0.001% |
| Norketamine | >100000 | <0.001% |
| Normeperidine | >100000 | <0.001% |
| Normorphine | >100000 | <0.001% |
| Noroxycodone | >100000 | <0.001% |
| Oxycodone | >100000 | <0.001% |
| Oxymorphone | >100000 | <0.001% |
| Pentazocine (Talwin) | >100000 | <0.001% |
| Pipamperone | >100000 | <0.001% |
| Tapentadol | >100000 | <0.001% |
| Thioridazine | >100000 | <0.001% |
Page 17
| Tilidine | >100000 | <0.001% |
|---|---|---|
| Tramadol | >100000 | <0.001% |
| Tramadol-O-Desmethyl | >100000 | <0.001% |
| Tramadol-N-Desmethyl | >100000 | <0.001% |
| MDMA 500 | (+/-)3,4-Methylenedioxy-nEthylamphetamine (MDEA) | >100000 | <0.5% |
|---|---|---|---|
| (+/-)-Methylenedioxyamphetamine(MDA) | 5000 | 10% | |
| (±)-MDMA | 500 | 100% | |
| L-Methamphetamine | 100000 | 0.5% | |
| d-methamphetamine | 100000 | 0.5% | |
| d-amphetamine | >100000 | <0.5% | |
| l-amphetamine | >100000 | <0.5% |
| MET 1000 | (+/-)3,4-Methylenedioxy-nEthylamphetamine (MDEA) | >100000 | <1% |
|---|---|---|---|
| (±)-MDMA | 2000 | 50% | |
| D-Methamphetamine | 1000 | 100% | |
| L-Methamphetamine | 25000 | 4% | |
| Fenfluramine | 50000 | 2% | |
| p-Hydroxymethamphetamine | 10000 | 10% | |
| D,L-Methamphetamine | 1000 | 100% | |
| β-Phenylethylamine | 50000 | 2% | |
| Mephetermine | 50000 | 2% | |
| Methoxyphenamine hydrochloride | 10000 | 1% | |
| L-Amphetamine | 75000 | 1.3% | |
| D-Amphetamine | >100000 | <1% | |
| D,L-Amphetamine | >100000 | <1% | |
| Chloroquine | 100000 | 1% | |
| Ephedrine hydrochloride | 50000 | 2% | |
| (+/-)3,4-Methylenedioxyamphetamine (MDA) | 100000 | 1% | |
| Trimethobenzamide | 100000 | 1% | |
| l-phenylephrine | 50000 | 2% | |
| (1R,2S)-(-)-Ephedrine | 100000 | 1% | |
| Procaine hydrochloride | >100000 | <1% |
Page 18
| MOP 300 | 6-acetylmorphine | 100 | 300% |
|---|---|---|---|
| Codeine | 300 | 100% | |
| Dihydrocodeine | 1000 | 30% | |
| EthylMorphine | 300 | 100% | |
| Heroin | 500 | 60% | |
| Hydrocodone | 1000 | 30% | |
| Hydromorphone | 10000 | 3% | |
| Levorphanol tartrate | 1500 | 20% | |
| Morphine | 300 | 100% | |
| Nalorphine hydrochloride | 300 | 100% | |
| Thebaine | 10000 | 3% | |
| s-Monoacetylmorphine | 300 | 100% | |
| Morphine-3-β-d-glucuronide | 300 | 100% | |
| 6-Monoacetylmorphine | 300 | 100% | |
| Codeine-6-β-D-glucuronide | 25000 | 1.2% | |
| Morphine-6-β-D-glucuronide | 1000 | 30% | |
| 6-Acetylcodeine | 25000 | 1.2% | |
| Normorphine | 50000 | 0.6% | |
| Oxycodone | >100000 | <0.3% | |
| Oxymorphone | >100000 | <0.3% | |
| Norcodeine | 25000 | 1.2% | |
| Procaine hydrochloride | >100000 | <0.3% | |
| Norpropoxyphene | >100000 | <0.3% |
| MTD 300 | (±)-Methadone | 300 | 100% |
|---|---|---|---|
| EDDP | >100000 | <0.3% | |
| EMDP | >100000 | <0.3% | |
| LAAM | >100000 | <0.3% | |
| Alpha Methadol | >100000 | <0.3% | |
| Doxylamine | >100000 | <0.3% |
| OXY 100 | Ethyl Oxycodone | 50000 | 0.2% |
|---|---|---|---|
| Hydrocodone | 5000 | 2% | |
| Hydromorphone | 100000 | 0.1% | |
| Levorphanol tartrate | 50000 | 0.2% | |
| Naloxone hydrochloride | 50000 | 0.2% |
Page 19
| Naltrexone hydrochloride | 50000 | 0.2% |
|---|---|---|
| Oxycodone | 100 | 100% |
| Oxymorphone | 200 | 50% |
| Oxymorphone-3β-D-glucuronide | 500 | 20% |
| Noroxycodone | 1000 | 10% |
| Noroxymorphone | 2500 | 4% |
| Dihydrocodeine | 20000 | 0.5% |
| Codeine | 100000 | 0.1% |
| Morphine | >100000 | <0.1% |
| Buprenorphine | >100000 | <0.1% |
| Ethylmorphine | >100000 | <0.1% |
| Thebaine | >100000 | <0.1% |
| 6-acetylmorphine | >100000 | <0.1% |
| PCP 25 | Phencyclidine | 25 | 100% |
|---|---|---|---|
| 4-Hydroxy Phencyclidine | 12,500 | 0.2% |
| PPX 300 | d-Propoxyphene | 300 | 100% |
|---|---|---|---|
| d-Norpropoxyphene | 300 | 100% |
| TCA 1000 | Amitriptyline | 1000 | 100% |
|---|---|---|---|
| Chlorpheniramine | 50000 | 2% | |
| Clomipramine | 1000 | 100% | |
| Cyclobenzaprine Hydrochloride | 1000 | 100% | |
| Desipramine | 300 | 333% | |
| Doxepine | 1000 | 100% | |
| Duloxetine | >100000 | <1% | |
| Imipramine | 500 | 200% | |
| Norclomipramine | 500 | 200% | |
| Nordoxepine | 10000 | 10% | |
| Nortriptyline | 1000 | 100% | |
| Promazine | 2000 | 50% | |
| Trimipramine | 3000 | 33.3% | |
| Maprotiline | >100000 | <1% | |
| Promethazine hydrochloride | 50000 | 2% |
| THC 50 | 11-nor-Δ8-THC -9-COOH | 30 | 166% |
|---|---|---|---|
| (-)-11-nor-9-carboxy-Δ9-THC | 50 | 100% |
Page 20
| (±)-11-nor-9-Carboxy-Δ9-THC | 50 | 100% |
|---|---|---|
| 11-nor-Δ9-THC -carboxy glucuronide | 50 | 100% |
| 11-hydroxy-Δ9-Tetrahydrocannabinol | 5000 | 1% |
| Δ8- Tetrahydrocannabinol | 5000 | 1% |
| Δ9- Tetrahydrocannabinol | 10000 | 0.5% |
| Cannabinol | 10000 | 0.5% |
| Cannabidiol | >100000 | <0.05% |
| (±)-11-Hydroxy-Δ9-THC | >100000 | <0.05% |
| 6-MAM 10 | 6-acetylmorphine | 10 | 100% |
|---|---|---|---|
| Heroin | 50 | 20% | |
| Morphine | 10000 | 0.1% | |
| Normorphine | 100000 | 0.01% | |
| Nalorphine HCl | 100000 | 0.01% | |
| Hydrocodone | 100000 | 0.01% | |
| Hydromorphone | 100000 | 0.01% | |
| Chlordiazepoxide | 100000 | 0.01% | |
| Clobazam | 100000 | 0.01% | |
| D-Amphetamine | >100000 | <0.01% | |
| (±)-Amphetamine | >100000 | <0.01% | |
| Levorphanol tartrate | 100000 | 0.01% | |
| Codeine | >100000 | <0.01% | |
| Ethylmorphine | >100000 | <0.01% | |
| Morphine3-β-D-glucuronide | 100000 | 0.01% | |
| Norcodeine | 100000 | 0.01% | |
| Oxycodone | >100000 | <0.01% | |
| Oxymorphone | >100000 | <0.01% | |
| Procaine hydrochloride | >100000 | <0.01% | |
| Thebaine | >100000 | <0.01% | |
| 6-Acetylcodeine | 100000 | 0.01% | |
| Buprenorphine | >100000 | <0.01% | |
| Dihydrocodeine | >100000 | <0.01% | |
| Dextromethorphan | >100000 | <0.01% | |
| Imipramine hydrochloride | >100000 | <0.01% |
Page 21
| Meperidine | >100000 | <0.01% |
|---|---|---|
| (±)-Methadone | >100000 | <0.01% |
| Mitragynine(kratom) | 100000 | 0.01% |
| Morphine-6-β-D-glucuronide | 100000 | 0.01% |
| Naloxone hydrochloride | >100000 | <0.01% |
| Naltrexone hydrochloride | >100000 | <0.01% |
| Naproxen | >100000 | <0.01% |
| Norbuprenorphine | >100000 | <0.01% |
| Norbuprenorphine-3-D-Glucuronide | 100000 | 0.01% |
| Noroxycodone HCL | >100000 | <0.01% |
| Noroxymorphone HCL | >100000 | <0.01% |
| Norpropoxyphene | >100000 | <0.01% |
| Oxymorphone-3β-D-glucuronide | >100000 | <0.01% |
| Tapentadol hydrochloride | >100000 | <0.01% |
| Tramadol hydrochloride | >100000 | <0.01% |
| AMP 500 | Hydroxyamphetamine | 4000 | 12.5% |
|---|---|---|---|
| (+/-)-Methylenedioxyamphetamine(MDA) | 200 | 250% | |
| D,L-Amphetamine | 500 | 100% | |
| D-Amphetamine | 500 | 100% | |
| Diethylstilbestrol | >100000 | <0.5% | |
| L-Amphetamine | 10000 | 5% | |
| Phentermine | 4000 | 12.5% | |
| β-Phenylethylamine | 50000 | 1% | |
| Tyramine | 10000 | 5% | |
| p-Hydroxynorephedrine | 50000 | 1% | |
| p-Hydroxyamphetamine | 50000 | 1% | |
| D,L-Norephedrine | 50000 | 1% | |
| (+/-)3,4-Methylenedioxy-nethylamphetamine(MDEA) | >100000 | <0.5% | |
| d-Methamphetamine | >100000 | <0.5% | |
| l-Methamphetamine | >100000 | <0.5% | |
| (+/-)3,4-Methylenedioxymethamphetamine (MDMA) | >10000 | <5% | |
| Ephedrine hydrochloride | >100000 | <0.5% |
Page 22
| Phenylpropanolamine | >100000 | <0.5% |
|---|---|---|
| Benzphetamine | >100000 | <0.5% |
| L-Ephedrine | >100000 | <0.5% |
| L-Epinephrine | >100000 | <0.5% |
| D,L-Epinephrine | >100000 | <0.5% |
| COC 150 | Benzoylecogonine | 150 | 100% |
|---|---|---|---|
| Cocaethylene | 150 | 100% | |
| Cocaine hydrochloride | 100 | 150% | |
| Ecgonine | >100000 | <0.15% | |
| Norcocaine | >100000 | <0.15% | |
| Ecgonine methyl Ester | >100000 | <0.15% |
| MET 500 | (+/-)3,4-Methylenedioxy-nethylamphetamine (MDEA) | >100000 | <0.5% |
|---|---|---|---|
| (±)-MDMA | 1000 | 50% | |
| D-Methamphetamine | 500 | 100% | |
| L-Methamphetamine | 10000 | 5% | |
| Fenfluramine | 25000 | 2% | |
| p-Hydroxymethamphetamine | 5000 | 10% | |
| D,L-Methamphetamine | 500 | 100% | |
| β-Phenylethylamine | 10000 | 5% | |
| Mephetermine | 25000 | 2% | |
| Methoxyphenamine hydrochloride | 50000 | 1% | |
| L-Amphetamine | 100000 | 0.5% | |
| Ephedrine hydrochloride | 100000 | 0.5% | |
| (1R,2S)-(-)-Ephedrine | >100000 | <0.5% | |
| D-Amphetamine | >100000 | <0.5% | |
| Chloroquine | 100000 | 0.5% | |
| (+/-)3,4-Methylenedioxyamphetamine(MDA) | >100000 | <0.5% | |
| L-Phenylephrine | 100000 | 0.5% | |
| Trimethobenzamide | 100000 | 0.1% | |
| Procaine hydrochloride | >100000 | <0.5% | |
| d/l-Amphetamine | 100000 | 0.5% |
Page 23
| OPI 2000 | 6-acetylmorphine | 10000 | 20% |
|---|---|---|---|
| Codeine | 1000 | 200% | |
| Dihydrocodeine | 1500 | 133.3% | |
| EthylMorphine | 2500 | 80% | |
| Codeine-6-β-D-glucuronide | 2500 | 80% | |
| Heroin | 5000 | 40% | |
| Hydrocodone | 5000 | 40% | |
| Hydromorphone | 2500 | 80% | |
| Levorphanol tartrate | 10000 | 20% | |
| Morphine | 2000 | 100% | |
| Nalorphine hydrochloride | 5000 | 40% | |
| Norcodeine | 4000 | 50% | |
| Normorphine | 50000 | 4% | |
| Oxymorphone | 25000 | 8% | |
| s-Monoacetylmorphine | 5000 | 40% | |
| Morphine-6-β-D-glucuronide | 2000 | 100% | |
| Thebaine | 10000 | 20% | |
| Morphine 3-β-D-glucuronide | 2000 | 100% | |
| 6-Monoacetylmorphine (6-MAM) | 1500 | 133.3% | |
| 6-Acetylcodeine | 10000 | 20% | |
| Oxycodone | >100000 | <2% | |
| Procaine | >100000 | <2% | |
| Norpropoxyphene | >100000 | <2% |
To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 50% below and 50% above each corresponding cutoff.
Compounds that show no interference at a concentration of 100µg/mL are summarized in the following table.
| (-) Cotinine | Diflunisal | Norethindrone |
|---|---|---|
| 3-Hydroxytyramine(except AMP test) | Digoxin | Norpropoxyphene |
| 7-Aminoflunitrazepam(except MDMA test) | Dimethyl-aminoantipyrine | Norpseudoephedrine |
| 7-Aminonitrazepam | Diphenhydramine HCl | Nortriptyline(except TCA test) |
| Acetaminophen | Diphenylhydantoin | Noscapine |
| Acetone (1000 mg/dL) | Disopyramide | Octopamine |
Page 24
| Acetophenetidin | Dopamine HCl (except AMP TCA test) | O-Hydroxyhippuric acid |
|---|---|---|
| Acetylsalicylic acid | Doxepine(except TCA test) | Olanzapine |
| Acyclovir | Doxylamine | Omeprazole |
| Albumin(100mg/dL) | D-Pseudoephedrine | Oxalic acid |
| Albuterol | Duloxetine | Oxazepam (except BZO test) |
| Albuterol sulfate(Proair HFA) | Ecgonine methyl ester | Oxolinic acid |
| Alpha Methadol | EMDP | Oxymetazoline |
| Aminophylline | Ephedrine hydrochloride(except MET test) | Paliperidone |
| Aminopyrine | Erythromycin | Papaverine |
| Amitriptyline(except TCA test) | Esomeprazole Magnesium | Penicillin-G |
| Amlodipine besylate | Estradiol | PenicillinV Potassium |
| Amobarbital(except BAR test) | Estrone | Perphenazine(except TCA test) |
| Amoxicillin | Ethanol(1%) | Phenacetin |
| Ampicillin | Fenfluramine(except MET test) | Phencyclidine(except PCP test) |
| Apomorphine | Fenofibrate | Phenelzine |
| Aripiprazole | Fenoprofen | Phenethylamine(except AMP MDMA test) |
| Ascorbic acid | Fluoxetine Hydrochloride | Phenobarbital(except BAR test) |
| Aspartame | Fluphenazine | Phentermine(except AMP test) |
| Aspirin | Fotemustine | Phenylpropanolamine |
| Atomoxetine | Furosemide | Prednisone |
| Atorvastatin Calcium | Gabapentin | Pregablin |
| Atropine | Galactose | Procaine |
| Azithromycin | Gatifloxacin | Promazine(except TCA test) |
| Baclofen | Gemfibrozil | Promethazine |
| Benzilic acid | Gentisic acid | Propoxyphene(except PPX test) |
| Benzocaine | Glucose | Propranolol |
| Benzoic acid | Guaiacolglycerylether | Pseudoephedrine |
| Benzoylecogonine(except COC test) | Hemoglobin | Pyridoxine |
| Benzphetamine | Hexobarbital | Pyrilamine |
| Benzylpiperiazine | Hydralazine | Pyrogallol |
| Bilirubin | Hydrochlorothiazide | Quetiapine |
| Boric Acid (1%) | Hydrocortisone | Quinidine |
| Bromo-2,5,Dimethoxyphenethylamine | Hydroxybutyric Acid | Quinine |
| Bupropion | Ibuprofen | Quinolinic Acid |
| Caffeine | Imipramine(except TCA test) | Ranitidine |
| Cannabidiol | Isoproterenol | Rifampicin |
| Captopril | Isoxsuprine | Salicylic acid |
| Carbamazepine | Ketamine | Secobarbital(except BAR test) |
| Carfentanil | Ketoprofen | Serotonin(except AMP test) |
Page 25
| Carisoprodol | LAAM HCl | Serotonin (5-Hydroxytyramine)(except AMP test) |
|---|---|---|
| Cefradine | Labetalol | Sertraline |
| Cephalexin | Lamotrigine | Sildenafil Citrate |
| Cetirizine | L-Ephedrine | Simvastatin |
| Chloralhydrate | L-Epinephrine | Sodium Azide |
| Chloramphenicol | Levofloxacin Hydrochloride(except AMP test) | Sulfamethazine(except 6-MAM test) |
| Chlordiazepoxide(except BZO test) | Levonorgestrel | Sulindac(except BZO test) |
| Chloroquine | Levothyroxine Sodium | Telmisartan |
| Chlorothiazide | Lidocaine Hydrochloride | Tetracycline |
| Chlorpheniramine(except TCA test) | Lisinopril | Tetrahydrocortisone 3-(βDglucuronide) |
| Chlorpromazine(except TCA test) | Loperamide | Tetrahydrocortisone, 3-acetate |
| Cholesterol | Loratadine | Tetrahydrozoline |
| Ciprofloxacin Hydrochloride(except AMP test) | L-phenylephrine | THC (except THC test) |
| Citalopram(except TCA test) | LSD | Theophylline |
| Clarithromycin | L-thyroxine | Thiamine |
| Clofibrate | Magnesium | Thioridazine(except TCA test) |
| Clomipramine(except TCA test) | Maprotiline | Trazodone Hydrochloride |
| Clonidine | Meperidine | Triamteren |
| Clozapine | Meprobamate | Trifluoperazine |
| Conjugated Estrogens | Metformin | Trifluoromethylphenyl-piperazine |
| Cortisone | Methapyrilene | Trimethobenzamide(except AMP MET test) |
| Creatine Hydrate | Methaqualone | Trimethoprime |
| Creatinine | Methoxyphenamine (except MET test) | Tryptamine |
| Cyclobenzaprine(except TCA test) | Methylphenidate | Tyramine (except AMP test) |
| Cyproheptadine | Metoprolol Tartrate | Urea (2000 mg/dL) |
| D,L-Epinephrine | Metronidazole | Uric acid |
| D,L-Isoproterenol | Mifepristone | Valproic acid (250 μg/mL) |
| D,L-Lorazepam (except BZO test) | N-Acetylprocainamide | Venlafaxine HCl |
| D,L-Octopamine | Nacl (4000 mg/dL) | Verapamil |
| D,L-Propranolol | Nalidixic acid | Vitamin B2 |
| D,L-Tryptophan | Naloxone hydrochloride(except OXY test) | Vitamin C |
| D,L-Tyrosine | Naltrexone hydrochloride(except OXY test) | Zaleplon |
| Delorazepam | Naproxen | Zolpidem |
| Demoxepam(except BZO test) | N-desmethyl Tapentadol | Zomepirac |
Page 26
| Deoxycorticosterone | Niacinamide | β-Estradiol |
|---|---|---|
| Desloratadine(except TCA test) | Nicotine | γ-Cyclodextrin |
| Dextromethorphan | Nicotinic Acid | γ-Globulin (500mg/dL) |
| Diclofenac | Nifedipine | |
| Diclofenac sodium | Nitroglycerin |
Interference by pH and specific gravity were also evaluated using pooled urine specimens with concentrations of 0 (drug-free), at 50% below and 50% above each corresponding cutoff. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.
B. Method comparison study
The method comparison studies for the device were performed in-house with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the table below:
| Drug test | Test Result | Drug-Free | Low Negative by LC-MS/MS (less than -50%) | Near Cutoff Negative by LC-MS/MS (Between -50% and the Cutoff) | Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%) | High Positive by LC-MS/MS (greater than +50%) |
|---|---|---|---|---|---|---|
| AMP (1000) | Operator A | + | 0 | 0 | 0 | 15 |
| - | 10 | 14 | 16 | 3 | ||
| Operator B | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| Operator C | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| AMP (500) | Operator A | + | 0 | 0 | 0 | 16 |
| - | 10 | 14 | 16 | 2 | ||
| Operator B | + | 0 | 0 | 0 | 16 | |
| - | 10 | 14 | 16 | 2 | ||
| Operator C | + | 0 | 0 | 1 | 18 | |
| - | 10 | 14 | 15 | 0 | ||
| BAR | Operator A | + | 0 | 0 | 1 | 16 |
| - | 10 | 14 | 15 | 2 | ||
| Operator B | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| Operator C | + | 0 | 0 | 2 | 17 | |
| - | 10 | 14 | 14 | 1 | ||
| BUP | Operator A | + | 0 | 0 | 0 | 16 |
Page 27
| - | 10 | 14 | 16 | 2 | 0 | |
|---|---|---|---|---|---|---|
| Operator B | + | 0 | 0 | 1 | 16 | |
| - | 10 | 14 | 15 | 2 | ||
| Operator C | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| BZO | Operator A | + | 0 | 0 | 0 | 15 |
| - | 10 | 14 | 16 | 3 | ||
| Operator B | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| Operator C | + | 0 | 0 | 2 | 18 | |
| - | 10 | 14 | 14 | 0 | ||
| COC (300) | Operator A | + | 0 | 0 | 1 | 15 |
| - | 10 | 14 | 15 | 3 | ||
| Operator B | + | 0 | 0 | 1 | 15 | |
| - | 10 | 14 | 15 | 3 | ||
| Operator C | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| COC (150) | Operator A | + | 0 | 0 | 2 | 18 |
| - | 10 | 14 | 14 | 0 | ||
| Operator B | + | 0 | 0 | 2 | 18 | |
| - | 10 | 14 | 14 | 0 | ||
| Operator C | + | 0 | 0 | 3 | 18 | |
| - | 10 | 14 | 13 | 0 | ||
| EDDP | Operator A | + | 0 | 0 | 2 | 17 |
| - | 10 | 14 | 14 | 1 | ||
| Operator B | + | 0 | 0 | 2 | 18 | |
| - | 10 | 14 | 14 | 0 | ||
| Operator C | + | 0 | 0 | 3 | 18 | |
| - | 10 | 14 | 13 | 0 | ||
| FYL | Operator A | + | 0 | 0 | 2 | 16 |
| - | 10 | 14 | 14 | 2 | ||
| Operator B | + | 0 | 0 | 2 | 16 | |
| - | 10 | 14 | 14 | 2 | ||
| Operator C | + | 0 | 0 | 3 | 16 | |
| - | 10 | 14 | 13 | 2 | ||
| MDMA | Operator A | + | 0 | 0 | 2 | 17 |
| - | 10 | 14 | 14 | 1 | ||
| Operator B | + | 0 | 0 | 1 | 16 | |
| - | 10 | 14 | 15 | 2 | ||
| Operator C | + | 0 | 0 | 2 | 17 | |
| - | 10 | 14 | 14 | 1 | ||
| MET | Operator A | + | 0 | 0 | 1 | 17 |
Page 28
| (1000) | - | 10 | 14 | 15 | 1 | 0 |
|---|---|---|---|---|---|---|
| Operator B | + | 0 | 0 | 0 | 17 | |
| - | 10 | 14 | 16 | 1 | ||
| Operator C | + | 0 | 0 | 2 | 18 | |
| - | 10 | 14 | 14 | 0 | ||
| MET (500) | Operator A | + | 0 | 0 | 0 | 16 |
| - | 10 | 14 | 16 | 2 | ||
| Operator B | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| Operator C | + | 0 | 0 | 2 | 17 | |
| - | 10 | 14 | 14 | 1 | ||
| MOP (300) | Operator A | + | 0 | 0 | 0 | 16 |
| - | 10 | 14 | 16 | 2 | ||
| Operator B | + | 0 | 0 | 0 | 15 | |
| - | 10 | 14 | 16 | 3 | ||
| Operator C | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| OPI (2000) | Operator A | + | 0 | 0 | 1 | 17 |
| - | 10 | 14 | 15 | 1 | ||
| Operator B | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| Operator C | + | 0 | 0 | 2 | 17 | |
| - | 10 | 14 | 14 | 1 | ||
| MTD | Operator A | + | 0 | 0 | 0 | 15 |
| - | 10 | 14 | 16 | 3 | ||
| Operator B | + | 0 | 0 | 0 | 15 | |
| - | 10 | 14 | 16 | 3 | ||
| Operator C | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| OXY | Operator A | + | 0 | 0 | 0 | 16 |
| - | 10 | 14 | 16 | 2 | ||
| Operator B | + | 0 | 0 | 0 | 16 | |
| - | 10 | 14 | 16 | 2 | ||
| Operator C | + | 0 | 0 | 1 | 16 | |
| - | 10 | 14 | 15 | 2 | ||
| PCP | Operator A | + | 0 | 0 | 0 | 16 |
| - | 10 | 14 | 16 | 2 | ||
| Operator B | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| Operator C | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| PPX | Operator A | + | 0 | 0 | 0 | 16 |
Page 29
| - | 10 | 14 | 16 | 2 | 0 | |
|---|---|---|---|---|---|---|
| Operator B | + | 0 | 0 | 0 | 16 | |
| - | 10 | 14 | 16 | 2 | ||
| Operator C | + | 0 | 0 | 0 | 17 | |
| - | 10 | 14 | 16 | 1 | ||
| TCA | Operator A | + | 0 | 0 | 0 | 16 |
| - | 10 | 14 | 16 | 2 | ||
| Operator B | + | 0 | 0 | 0 | 16 | |
| - | 10 | 14 | 16 | 2 | ||
| Operator C | + | 0 | 0 | 1 | 17 | |
| - | 10 | 14 | 15 | 1 | ||
| THC | Operator A | + | 0 | 0 | 0 | 16 |
| - | 10 | 14 | 16 | 2 | ||
| Operator B | + | 0 | 0 | 0 | 16 | |
| - | 10 | 14 | 16 | 2 | ||
| Operator C | + | 0 | 0 | 1 | 18 | |
| - | 10 | 14 | 15 | 0 | ||
| 6-MAM | Operator A | + | 0 | 0 | 1 | 16 |
| - | 10 | 14 | 15 | 2 | ||
| Operator B | + | 0 | 0 | 1 | 16 | |
| - | 10 | 14 | 15 | 2 | ||
| Operator C | + | 0 | 0 | 2 | 18 | |
| - | 10 | 14 | 14 | 0 |
Discordant Results are summarized below.
| Drug | Operator | Sample Number | LC/MS/MS Result (ng/mL) | Accurate Result |
|---|---|---|---|---|
| 6-MAM 10 | Operator A、C | DBMC1307 | 8.77 | Positive |
| Operator B、C | DBMC0177 | 9.16 | Positive | |
| Operator A、B | DBMC0604 | 10.82 | Negative | |
| Operator A、B | DBMC0685 | 11.01 | Negative | |
| AMP 1000 | Operator B、C | DBMC0242 | 998.95 | Positive |
| Operator A、B、C | DBMC1670 | 1017.86 | Negative | |
| Operator A | DBMC1538 | 1033.17 | Negative | |
| Operator A | DBMC0629 | 1039.9 | Negative | |
| BAR 300 | Operator A、B、C | DBMC0762 | 298.98 | Positive |
| Operator C | DBMC0612 | 288.79 | Positive | |
| Operator A、B、C | DBMC0103 | 314.09 | Negative | |
| Operator A | DBMC0278 | 319.32 | Negative | |
| BUP 10 | Operator B | DBMC0867 | 9.95 | Positive |
| Operator C | DBMC0498 | 9.97 | Positive | |
| Operator A、B、C | DBMC0182 | 10.44 | Negative |
Page 30
| Operator A、B | DBMC0496 | 10.69 | Negative |
|---|---|---|---|
| BZO 300 | Operator C | DBMC1305 | 278.05 |
| Operator B、C | DBMC0603 | 285.73 | |
| Operator A | DBMC1300 | 310.58 | |
| Operator A、B | DBMC1022 | 305.35 | |
| Operator A | DBMC0514 | 309.44 | |
| COC 300 | Operator A、B、C | DBMC0058 | 295.45 |
| Operator A、B | DBMC1107 | 305.07 | |
| Operator A、B、C | DBMC0277 | 301.06 | |
| Operator B | DBMC0815 | 311.63 | |
| Operator A | DBMC0550 | 312.18 | |
| EDDP 300 | Operator C | DBMC0495 | 268.37 |
| Operator A、B、C | DBMC0728 | 284.94 | |
| Operator A、B、C | DBMC1634 | 288.76 | |
| Operator A | DBMC0961 | 301.82 | |
| FYL | Operator C | DBMC1630 | 0.94 |
| Operator A、B、C | DBMC0260 | 0.97 | |
| Operator A、B、C | DBMC0313 | 0.99 | |
| Operator A、B、C | DBMC0622 | 1.03 | |
| Operator A | DBMC1198 | 1.06 | |
| Operator C | DBMC1514 | 1.1 | |
| Operator B | DBMC1336 | 1.06 | |
| MDMA 500 | Operator A、C | DBMC0425 | 472.58 |
| Operator A、B、C | DBMC0522 | 471.3 | |
| Operator A、B、C | DBMC0804 | 527.24 | |
| Operator B | DBMC0464 | 540.66 | |
| MET 1000 | Operator A、C | DBMC0780 | 981.48 |
| Operator C | DBMC1290 | 978.54 | |
| Operator A、B | DBMC0723 | 1019.75 | |
| MOP 300 | Operator C | DBMC0162 | 290.78 |
| Operator A、B、C | DBMC0424 | 301.43 | |
| Operator A、B | DBMC1121 | 303.18 | |
| Operator B | DBMC0343 | 304.93 | |
| MTD 300 | Operator C | DBMC0452 | 286.81 |
| Operator A、B、C | DBMC0668 | 310.22 | |
| Operator A、B | DBMC0750 | 316.65 | |
| Operator A、B | DBMC0553 | 327.95 | |
| OXY 100 | Operator C | DBMC0525 | 92.84 |
| Operator A、B、C | DBMC0838 | 102.37 | |
| Operator A、B、C | DBMC1006 | 101.84 | |
| PCP 25 | Operator B、C | DBMC1636 | 24.98 |
| Operator A、 B | DBMC1007 | 25.72 | |
| Operator A、C | DBMC0548 | 25.82 |
Page 31
| PPX 300 | Operator A、B | DBMC1533 | 313.94 | Negative |
|---|---|---|---|---|
| Operator A、B、C | DBMC1383 | 308.97 | Negative | |
| TCA 1000 | Operator C | DBMC0768 | 988.94 | Positive |
| Operator A、B、C | DBMC1567 | 1029.78 | Negative | |
| Operator A、B | DBMC1439 | 1046.84 | Negative | |
| THC 50 | Operator C | DBMC1146 | 45.12 | Positive |
| Operator A、B | DBMC0105 | 50.68 | Negative | |
| Operator A | DBMC1442 | 51.24 | Negative | |
| Operator B | DBMC0884 | 52.32 | Negative | |
| AMP 500 | Operator C | DBMC0523 | 466.77 | Positive |
| Operator A, B | DBMC1578 | 505.67 | Negative | |
| Operator B | DBMC1490 | 518.94 | Negative | |
| Operator A | DBMC0050 | 517.74 | Negative | |
| COC 150 | Operator A、C | DBMC1580 | 148.45 | Positive |
| Operator B、C | DBMC0561 | 148.38 | Positive | |
| Operator A、B、C | DBMC1233 | 149.02 | Positive | |
| MET 500 | Operator C | DBMC0398 | 473.25 | Positive |
| Operator B、C | DBMC1029 | 482.02 | Positive | |
| Operator A | DBMC0673 | 520.94 | Negative | |
| Operator A、B、C | DBMC1667 | 508.23 | Negative | |
| OPI 2000 | Operator A、C | DBMC1103 | 1878.95 | Positive |
| Operator B、C | DBMC0369 | 1875.61 | Positive | |
| Operator A、B、C | DBMC1167 | 2029.55 | Negative |
C. Lay person study
A lay user study was performed at three intended user sites with 280 lay persons. 176 male and 104 female participated the study. They had diverse educational and professional backgrounds and their age range from 21 to > 50. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.
Result of Deepblue Muti-Drug Urine Test Cup Configuration 1:
| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|-------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| AMP | 500 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| BAR | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
Page 32
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |
|---|---|---|---|---|---|---|---|---|
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% | |
| BUP | 10 | Negative | 20 | 20 | 20 | 18 | 1 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | ||
| BZO | 300 | Negative | 20 | 20 | 20 | 18 | 1 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | ||
| COC | 150 | Negative | 20 | 20 | 20 | 19 | 1 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | ||
| EDDP | 300 | Negative | 20 | 20 | 20 | 18 | 1 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | ||
| FTY | 1 | Negative | 20 | 20 | 20 | 19 | 2 | 0 |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | ||
| MDMA | 500 | Negative | 20 | 20 | 20 | 19 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 100% | 100% | ||
| MET | 500 | Negative | 20 | 20 | 20 | 18 | 1 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | ||
| MOP | 300 | Negative | 20 | 20 | 20 | 20 | 1 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | ||
| MTD | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | ||
| OXY | 100 | Negative | 20 | 20 | 20 | 19 | 1 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% |
Page 33
| PCP | 25 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
|---|---|---|---|---|---|---|---|---|---|
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% | ||
| PPX | 300 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% | ||
| TCA | 1000 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | ||
| THC | 50 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% | ||
| 6-MAM | 10 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% |
Result of Deepblue Muti-Drug Urine Test Cup Configuration 2:
| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|-------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| AMP | 1000 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| BAR | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| BUP | 10 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% |
| BZO | 300 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% |
| COC | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
Page 34
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |
|---|---|---|---|---|---|---|---|---|
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% | |
| EDDP | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | ||
| FTY | 1 | Negative | 20 | 20 | 20 | 19 | 1 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | ||
| MDMA | 500 | Negative | 20 | 20 | 20 | 18 | 1 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | ||
| MET | 1000 | Negative | 20 | 20 | 20 | 20 | 1 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | ||
| MOP | 2000 | Negative | 20 | 20 | 20 | 19 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 100% | 100% | ||
| MTD | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | ||
| OXY | 100 | Negative | 20 | 20 | 20 | 19 | 1 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | ||
| PCP | 25 | Negative | 20 | 20 | 20 | 19 | 2 | 0 |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | ||
| PPX | 300 | Negative | 20 | 20 | 20 | 18 | 1 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | ||
| TCA | 1000 | Negative | 20 | 20 | 20 | 19 | 2 | 0 |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% |
Page 35
| THC | 50 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
|---|---|---|---|---|---|---|---|---|---|
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | ||
| 6-MAM | 10 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% |
Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.
Clinical Studies:
Not applicable.
- Conclusion
Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that Deepblue Home Muti-Drug Urine Test Cup and Deepblue Muti-Drug Urine Test Cup are substantially equivalent to the predicate devices.
§ 862.3650 Opiate test system.
(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).