K Number
K250724
Manufacturer
Date Cleared
2025-07-30

(142 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intraosseous Infusion Needles are sterile, disposable devices used primarily during pediatric emergencies as an alternative to unsuccessful intravenous access to allow for effective infusion of resuscitative drugs or fluids.

Device Description

The Intraosseous Infusion Needles are intended for use as an alternative to intravenous access during pediatric emergencies, permitting infusion of drugs and fluids. Intraosseous Infusion Needles consist of two basic components - the needle (composed of a knob, a stylet luer-lock, and a beveled stylet) and the cannula (comprised of a hub, a base plate, and a cannula shaft). The cannula attaches to the needle by the stylet luer-lock and is supplied attached and ready to use. The Intraosseous Infusion Needles are available in the following configurations:

Intraosseous Infusion Needle (with Dieckmann Modification – Standard Hub Design)

Cannula Gauge: 14, 16, 18
Cannula/Stylet Length (cm): 14 Gauge: 3.0; 16 Gauge: 2.5, 3.0, 4.0; 18 Gauge: 2.5, 3.0
Stylet Length (cm): 14 Gauge: 4.6; 16 Gauge: 4.2 (for 2.5cm cannula), 4.65 (for 3.0cm cannula), 5.65 (for 4.0cm cannula); 18 Gauge: 4.2 (for 2.5cm cannula), 4.65 (for 3.0cm cannula)
Stylet Bevel Style: Trocar
Hub Material: Clear Polycarbonate (for 14 and 16 Gauge), Pink Polycarbonate (for 18 Gauge)

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the Intraosseous Infusion Needles does not contain the detailed information necessary to answer most of the questions you've posed regarding acceptance criteria and study particulars.

This document focuses on justifying "substantial equivalence" to a predicate device, primarily by demonstrating that changes to the sterilization process do not negatively impact safety or effectiveness. It explicitly states that "No additional performance testing was conducted because the design of the subject device is identical to the predicate, K160887."

Therefore, I can only provide limited information based on the text.

Here's what can be inferred and what cannot:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria or performance metrics for the device's primary function (intraosseous infusion). The "performance testing" described is related only to the sterilization process.

Acceptance Criteria CategorySpecific Criteria (from document)Reported Device Performance (from document)
Sterilization EfficacySterility Assurance Level (SAL) of $10^{-6}$Not explicitly stated as "met" an acceptance criterion, but the Ethylene Oxide Sterilization Performance Qualification was conducted to validate changes to the sterilization cycle parameters. This implies the SAL of $10^{-6}$ was achieved with the new process.
ResidualsCompliance with ISO 10993-7 for Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) residualsEvaluation of EO and ECH residuals was conducted, implying compliance.
BiocompatibilityIdentical to predicate in formulation, processing, sterilization method, geometry, and no other chemicals added.Device is identical to the predicate in these aspects, addressing biocompatibility by proxy with the predicate.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: Not specified for any performance testing related to the device's functional use or even for the sterilization validation beyond general statements of "performance testing was conducted."
  • Data Provenance: Not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The document does not describe studies requiring expert ground truth for device performance related to patient outcomes or imaging. The testing focuses on sterilization process validation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. The document does not describe studies requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical medical needle, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical medical needle, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the sterilization validation, the "ground truth" would be objective laboratory measurements and adherence to international standards (e.g., ANSI/AAMI/ISO 11135 and ISO 10993-7) for sterility and residuals. It's not related to clinical "ground truth" like pathology or outcomes.

8. The sample size for the training set

Not applicable. This is a physical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established

Not applicable.

FDA 510(k) Clearance Letter - Intraosseous Infusion Needles

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

July 30, 2025

Spectrum Vascular
Sharon Klugewicz
Chief Operating Officer / SVP Regulatory Affairs
50 Main Street
Suite 1000
White Plains, New York 10606

Re: K250724
Trade/Device Name: Intraosseous Infusion Needles
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: FMI
Dated: June 30, 2025
Received: June 30, 2025

Dear Sharon Klugewicz:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K250724 - Sharon Klugewicz
Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250724 - Sharon Klugewicz
Page 3

Sincerely,

Shruti N. Mistry -S

Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250724

Device Name
Intraosseous Infusion Needles

Indications for Use (Describe)
Intraosseous Infusion Needles are sterile, disposable devices used primarily during pediatric emergencies as an alternative to unsuccessful intravenous access to allow for effective infusion of resuscitative drugs or fluids.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K250724 - 510(k) Summary

Spectrum Vascular
Page 1 of 4

1. SUBMITTER INFORMATION

Applicant: Spectrum Vascular
Contact: Sharon Klugewicz
Phone: 516-425-4446
Email: sklugewicz@spectrumvascular.com
Address: 50 Main Street, Suite 1000
White Plains, NY 10606

2. CORRESPONDENT INFORMATION

Contact: Sharon Klugewicz
Title: Chief Operating Officer/Sr. VP Regulatory Affairs
Firm: Spectrum Vascular

3. DATE PREPARED: JULY 29, 2025

4. DEVICE INFORMATION

Device Name: Intraosseous Infusion Needles
Common Name: Hypodermic single lumen needle
Regulation Number: 880.5570
Regulation Name: Hypodermic single lumen needle
Product Code: FMI
Regulatory Class: Class II

5. PREDICATE DEVICE INFORMATION

Device Name: Cook® Intraosseous Infusion Needle
Common Name: Hypodermic single lumen needle
510(k) Number: K160887
Manufacturer: Cook Medical

6. DEVICE DESCRIPTION

The Intraosseous Infusion Needles are intended for use as an alternative to intravenous access during pediatric emergencies, permitting infusion of drugs and fluids.

Page 6

K250724 - 510(k) Summary
Spectrum Vascular
Page 2 of 4

Intraosseous Infusion Needles consist of two basic components - the needle (composed of a knob, a stylet luer-lock, and a beveled stylet) and the cannula (comprised of a hub, a base plate, and a cannula shaft). The cannula attaches to the needle by the stylet luer-lock and is supplied attached and ready to use. The Intraosseous Infusion Needles are available in the following configurations:

Intraosseous Infusion Needle (with Dieckmann Modification – Standard Hub Design)

Cannula GaugeCannula/Stylet Length (cm)Stylet Length (cm)Stylet Bevel StyleHub Material
143.04.6TrocarClear Polycarbonate
162.54.2TrocarClear Polycarbonate
3.04.65
4.05.65
182.54.2TrocarPink Polycarbonate
3.04.65

7. INDICATIONS FOR USE

Intraosseous Infusion Needles are sterile, disposable devices used primarily during pediatric emergencies as an alternative to unsuccessful intravenous access to allow for effective infusion of resuscitative drugs or fluids.

8. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Intraosseous Infusion Needles are identical to the Cook Intraosseous Infusion Needles. No changes have been made to the design of the device. This 510(k) is submitted to address changes to the sterilization process parameters and location.

Table 1: Device Comparison Table

Proposed DevicePredicate Device K160887
Regulation880.5570880.5570
Product CodeFMIFMI
ClassificationIIII
Indications for UseIntraosseous Infusion Needles are sterile, disposable devices used primarily during pediatric emergencies as an alternative to unsuccessful intravenous access to allow for effective infusion of resuscitative drugs or fluids.Cook Intraosseous Infusion Needles are sterile, disposable devices used primarily during pediatric emergencies as an alternative to unsuccessful intravenous access to allow for effective infusion of resuscitative drugs or fluids.

Page 7

K250724 - 510(k) Summary
Spectrum Vascular
Page 3 of 4

Proposed DevicePredicate Device K160887
Cannula Materials
Hub materialPolycarbonateNickel plated brass or polycarbonate
Base plate materialsPolycarbonateNickel plated brass or polycarbonate
Cannula shaft materialStainless steelStainless steel
Cannula shaft diameter (gauge)14, 16, 1814, 15.5, 16, 18
Cannula shaft length (cm)2.5, 3.0, 4.02.5, 3.0, 4.0
Needle Material
Knob materialNylon 6 (grey)Nylon 6 (grey)
Stylet Luer-lock materialPolypropyleneNickel plated brass or polypropylene
Beveled stylet materialStainless steelStainless steel
Stylet length (cm)4.2, 4.6, 4.65, 5.654.2, 4.6, 4.65, 4.7, 5.51, 5.6, 5.65, 5.7, 6.8, 7.0
PackagingPETG tray/TyvekPETG tray/Tyvek
SterilizationEtOEtO
SAL10⁻⁶10⁻⁶

Discussion of Similarities and Differences

The change in the sterilization process (e.g., lower EO concentration, cycle parameters, and location), as well as the change in name from the Cook Intraosseous Infusion Needles to the Intraosseous Infusion Needles, and chosen configurations, do not raise new or different questions of safety and effectiveness when compared to the predicate device.

9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING

Biocompatibility Testing

The medical device in its final finished form is identical to the Cook® Intraosseous Infusion Needle in formulation, processing, sterilization method, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).

Electrical Safety

Not applicable. The device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.

Electromagnetic Compatibility (EMC)

Page 8

K250724 - 510(k) Summary
Spectrum Vascular
Page 4 of 4

Not applicable. The device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.

Software

Not applicable. The device contains no software.

Performance Testing

The following performance testing was conducted to validate changes to the sterilization cycle parameters:

  • Ethylene Oxide Sterilization Performance Qualification
  • Evaluation of ethylene oxide (EO) and ethylene chlorohydrin (ECH) residuals

No additional performance testing was conducted because the design of the subject device is identical to the predicate, K160887.

Reference Standards conformed to:

  • ANSI/AAMI/ISO 11135 :2014, BS EN ISO 11135: Sterilization of health care products – Ethylene oxide – Requirements for the development, validation, and routine control of a sterilization process for medical devices
  • ISO 10993-7:2008, Amd. 1:2019: Biological evaluation of medical devices— Part 7: Ethylene oxide sterilization residuals

10. CONCLUSION

The results of the performance testing described above demonstrate that the modified sterilization cycle for the subject device does not raise new or different questions of safety and effectiveness when compared to the predicate device, and the subject device is substantially equivalent to the predicate device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).