K Number
K250553
Manufacturer
Date Cleared
2025-07-18

(143 days)

Product Code
Regulation Number
886.1570
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CASIA2 is a non-contact, high-resolution tomographic and biomicroscopic device intended for the in vivo imaging and measurement of ocular structures in the anterior segment. CASIA2 measures corneal thickness, anterior chamber depth and lens thickness.

Device Description

The Tomey Cornea/Anterior Segment OCT CASIA2 (CASIA2) is a non-contact, high resolution tomographic and biomicroscopic device indicated for in vivo imaging of ocular structures in the anterior segment. The Tomey Cornea/Anterior Segment OCT CASIA2 is indicated as an aid in the visualization and measurement of anterior segment findings. CASIA2 measures corneal thickness, anterior chamber depth and lens thickness.

This medical device product has functions subject to FDA premarket review (corneal thickness, curvature, anterior chamber depth and lens thickness) as well as functions that are not subject to FDA premarket review. For this application, for the (510(k) exempt functions that are not subject to FDA premarket review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the overall device.

CASIA2 consists of several components: the main unit, AC input power source, a touch panel LCD monitor, an external hard drive (HDD), a mouse and a keyboard.

AI/ML Overview

Here’s a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) clearance letter for the Tomey CASIA2.

Device: Tomey Cornea/Anterior Segment OCT (CASIA2)
Measurements evaluated: Central Corneal Thickness (CCT), Anterior Chamber Depth (ACD), and Lens Thickness (LT).

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined quantitative acceptance criteria (e.g., "CCT agreement must be within X µm"). Instead, it focuses on demonstrating agreement and precision compared to a legally marketed reference device (LENSTAR LS900). The "acceptance criteria" can be inferred from the study's objective to show substantial equivalence through these performance metrics. The reported device performance is presented as the actual agreement (mean difference and 95% Limits of Agreement - LOA) and precision (Repeatability and Reproducibility %CV).

Therefore, the table below reflects the demonstrated performance and implicitly what was considered acceptable for substantial equivalence.

MeasurementAcceptance Criteria (Implicit)Reported Device Performance (CASIA2 vs. LS900) - All Subjects Pooled
Agreement - Central Corneal Thickness (CCT)Agreement with reference device (LS900) demonstrated by Bland-Altman analysis with narrow 95% LOA.Bland-Altman plot shows data points clustered around zero difference, indicating good agreement. (Specific numerical LOA for CCT not provided in Table 21, but visually presented in Figure 14.4.1.12).
Agreement - Anterior Chamber Depth (ACD)Agreement with reference device (LS900) demonstrated by Bland-Altman analysis with narrow 95% LOA.Bland-Altman plot shows data points clustered around zero difference, indicating good agreement. (Specific numerical LOA for ACD not provided in Table 21, but visually presented in Figure 14.4.1.13).
Agreement - Lens Thickness (LT)Agreement with reference device (LS900) demonstrated by Bland-Altman analysis with narrow 95% LOA.Mean Difference (CASIA2 - LS900): 0.16 mm (SD 0.719 mm) 95% LOA: (-1.26 mm, 1.59 mm) (Visually represented in Figure 14.4.1.14 shows data points clustered, supporting agreement)
Precision (Repeatability) - CCTHigh repeatability (low %CV) of CASIA2 measurements.CASIA2: 0.18% CV LS900 (for comparison): 0.36% CV
Precision (Repeatability) - ACDHigh repeatability (low %CV) of CASIA2 measurements.CASIA2: 1.01% CV LS900 (for comparison): 3.09% CV
Precision (Repeatability) - LTHigh repeatability (low %CV) of CASIA2 measurements.CASIA2: 1.05% CV LS900 (for comparison): 1.01% CV
Precision (Reproducibility) - CCTHigh reproducibility (low %CV) of CASIA2 measurements.CASIA2: 0.32% CV LS900 (for comparison): 0.53% CV
Precision (Reproducibility) - ACDHigh reproducibility (low %CV) of CASIA2 measurements.CASIA2: 1.13% CV LS900 (for comparison): 4.35% CV
Precision (Reproducibility) - LTHigh reproducibility (low %CV) of CASIA2 measurements.CASIA2: 1.39% CV LS900 (for comparison): 2.35% CV

Note on "Acceptance Criteria": The document implies that meeting or exceeding the performance of the LS900 in terms of precision, and demonstrating good agreement via Bland-Altman analysis, constituted the acceptance for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: A total of 224 subjects were enrolled and completed the study for precision and agreement testing.
    • 55 subjects in the normal group
    • 60 subjects in the cataract group
    • 109 subjects in the special eyes group (eyes without a natural lens or eyes containing artificial materials)
      N for specific analyses (e.g., Agreement Analysis for LT) varied based on acceptable scans (e.g., N=122 for LT agreement, N=138 for CCT/ACD precision, N=76 for LT precision).
  • Data Provenance: The document does not explicitly state the country of origin. It indicates "The subjects of this study had no notable or unexpected/untoward assessments..." which suggests a single clinical site. However, no specific country is mentioned.
  • Retrospective or Prospective: The study was a prospective clinical study, as subjects were "enrolled," "randomized," and "assigned" to configurations and sequences, and data was collected during the study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • The document implies that the LENSTAR LS900 device itself served as the reference standard for establishing "ground truth" (or more accurately, the comparator for agreement) for the measured parameters.
  • It states that "The clinical site had 3 device operators trained on the devices used in the study."
  • Qualifications of Experts: The specific qualifications (e.g., radiologist, ophthalmologist, optometrist expertise, years of experience) of these 3 device operators are not explicitly stated in the provided text.

4. Adjudication Method for the Test Set

  • The document states, "Additional scans were taken at the operator's discretion if image quality was unacceptable based on the device DFU and the Tomey CASIA2 Reference Guide and included, missing scans, truncated scans, image defocus, floaters, presence of eye blinks, eye motion, etc. Each device operator had up to 3 attempts to obtain an acceptable scan for each of the required scans."
  • This suggests an operational approach to ensure data quality rather than a formal, independent adjudication process (e.g., 2+1/3+1 consensus by experts) for the measurements themselves. The "ground truth" was derived from the in-device measurements of the LS900, not a separate expert review. Therefore, there was no expert consensus-based adjudication method for the measurements.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a traditional MRMC comparative effectiveness study was not done.
  • The study design was focused on device-to-device agreement and precision (CASIA2 vs. LENSTAR LS900) rather than evaluating how human readers' performance (e.g., diagnostic accuracy) improved with or without AI assistance.
  • The CASIA2 is described as a "tomographic and biomicroscopic device intended for the in vivo imaging and measurement of ocular structures," with software providing "quantitative outputs." It does not appear to be an AI-assisted diagnostic aid for image interpretation that would typically require an MRMC study to show human reader benefit.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

  • The study primarily assessed the measurement performance of the CASIA2 device (algorithm/system) in generating quantitative outputs (CCT, ACD, LT) and compared these directly with a reference device. It's implied that these measurements are generated automatically by the device's software.
  • The role of the human operators was to acquire an "acceptable scan" based on predefined image quality criteria, not to interpret the images or provide a human "answer" for comparison with an AI-generated reading.
  • Therefore, the precision and agreement studies essentially represent the standalone performance of the CASIA2's measurement capabilities compared to the LS900.

7. Type of Ground Truth Used

  • The "ground truth" (or clinical reference standard) for comparison was the measurements obtained from the legally marketed predicate/reference device, LENSTAR LS900.
  • This is a device-based comparative ground truth, not expert consensus, pathology, or outcomes data. The study aimed to show that the CASIA2's measurements were interchangeable or highly agreeable with those from an established, cleared device.

8. Sample Size for the Training Set

  • The document describes a clinical study for validation/testing of the updated software. It does not provide any information about a separate training set size for the development of the algorithms generating these quantitative measurements.
  • It only mentions: "The device is a software upgraded version of the predicate K213265 that provides quantitative measurements. All quantitative measurements are derived from OCT images acquired with optical coherence tomography." This implies the software update incorporated algorithms to derive these measurements, but details on their development (including training data) are not provided in this 510(k) summary.

9. How the Ground Truth for the Training Set Was Established

  • As the document does not describe the training set or its development, there is no information provided on how the ground truth for any training set was established.

FDA 510(k) Clearance Letter - Tomey CASIA2

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

July 18, 2025

TOMEY Corporation
℅ Roger Albright
Associate Director, Medical Devices
Ora Inc.
138 Haverhill Street, Suite 102
Andover, Massachusetts 01810

Re: K250553
Trade/Device Name: Tomey Cornea/Anterior Segment OCT (CASIA2)
Regulation Number: 21 CFR 886.1570
Regulation Name: Ophthalmoscope
Regulatory Class: Class II
Product Code: OBO
Dated: February 25, 2025
Received: February 25, 2025

Dear Roger Albright:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250553 - Roger Albright
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250553 - Roger Albright
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Elvin Y. Ng -S

Elvin Ng
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250553

Device Name: Tomey Cornea/Anterior Segment OCT (CASIA2)

Indications for Use (Describe):

CASIA2 is a non-contact, high-resolution tomographic and biomicroscopic device intended for the in vivo imaging and measurement of ocular structures in the anterior segment. CASIA2 measures corneal thickness, anterior chamber depth and lens thickness.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

K250553
Page 1 of 12

This summary of the 510(k) premarket notification for the Tomey Cornea/Anterior Segment OCT CASIA2 is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR§807.92.

Submitter Information: (21.CFR 807.92(a)(1))

Owner/Company name, address
Tomey Corporation
2-11-33 Noritakeshinmachi
Nishi-ku, Nagoya, Aichi Nagoya 451-0051
Japan

Contact person
Yuko Matsushita
Telephone: [+81] 52-581-5327
E-mail: ymatsushita@tomey.co.jp

Contact/Application Correspondent
Roger Albright
Ora, LLC
138 Haverhill Street, Suite 102
Andover, MA 01810, USA
Telephone: (603) 518-0422
E-mail: ralbright@oraclinical.com

Date Prepared
July 8, 2025

1. Device Name (21.CFR 807.92(a)(2))

Trade Name: Tomey Cornea/Anterior Segment OCT (CASIA2)
Common Name: Ophthalmoscope.
Classification Name: Tomography, optical coherence
Product Code: OBO
Classification Regulation: 21 CFR 886.1570

2. Legally Marketed Predicate Devices (21.CFR 807.92(a)(3))

The CASIA2 is substantially equivalent to the following legally marketed device:

510(k) NumberTrade nameProduct code
K213265Cornea/Anterior Segment OCT CASIA2OBO

3. Device Description Summary (21.CFR 807.92(a)(4))

The Tomey Cornea/Anterior Segment OCT CASIA2 (CASIA2) is a non-contact, high resolution tomographic and biomicroscopic device indicated for in vivo imaging of ocular structures in the anterior

Page 6

K250553
510(k) Summary
Page 2 of 12

segment. The Tomey Cornea/Anterior Segment OCT CASIA2 is indicated as an aid in the visualization and measurement of anterior segment findings. CASIA2 measures corneal thickness, anterior chamber depth and lens thickness.

This medical device product has functions subject to FDA premarket review (corneal thickness, curvature, anterior chamber depth and lens thickness) as well as functions that are not subject to FDA premarket review. For this application, for the (510(k) exempt functions that are not subject to FDA premarket review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the overall device.

CASIA2 consists of several components: the main unit, AC input power source, a touch panel LCD monitor, an external hard drive (HDD), a mouse and a keyboard.

4. Intended Use/Indications for Use (21.CFR 807.92(a)(5))

CASIA2 is a non-contact, high-resolution tomographic and biomicroscopic device intended for the in vivo imaging and measurement of ocular structures in the anterior segment. CASIA2 measures corneal thickness, anterior chamber depth and lens thickness.

5. Indications for Use Comparison (21.CFR 807.92(a)(5))

The CASIA2 maintains the same intended use as the predicate CASIA2 (K213265) in vivo imaging and measurement of anterior segment ocular structures using identical swept-source OCT technology (1310 nm, 50,000 A-scans/s). It adds quantitative outputs (e.g., corneal thickness, curvature, anterior chamber depth and lens thickness) that align with K213265's capabilities, with no new intended use or safety/effectiveness risks. While some outputs overlap with LENSTAR LS900 (K082891) as a reference device, CASIA2 omits axial length and uses OCT, not optical low coherence reflectometry (OLCR), with no new concerns raised.

6. Technological Comparison (21.CFR 807.92(a)(6))

The device is a software upgraded version of the predicate K213265 that provides quantitative measurements. All quantitative measurements are derived from OCT images acquired with optical coherence tomography. This is the same technology used in the reference device K082891 utilizes to acquire measurements of various ocular structures of the eye.

Page 7

K250553
510(k) Summary
Page 3 of 12

ClassificationNew Device: Tomey CASIA2 with Updated SoftwarePrimary Predicate: Tomey Cornea/Anterior Segment OCT CASIA2 (K213265)Reference Device: LENSTAR, MODEL LS900 (K082891)Discussion Analysis of Differences Affect Safety or Effectiveness
ManufacturerTomey CorporationTomey CorporationHaag StreitSame as predicate
ClassificationClass IIClass IIClass IISame as both predicate and reference
Product CodeOBOOBOHJOSame as predicate
Regulation Number21 CFR 886.157021 CFR 886.157021 CFR 886.1850Same as predicate
Regulation NameOphthalmoscopeOphthalmoscopeAC-powered Slitlamp BiomicroscopeSame as predicate
Indications for UseCASIA2 is a non-contact, high-resolution tomographic and biomicroscopic device intended for the in vivo imaging and measurement of ocular structures in the anterior segment. CASIA2 measures corneal thickness, anterior chamber depth and lens thickness.The Tomey Cornea/Anterior Segment OCT CASIA2 is a non-contact, high resolution tomographic and biomicroscopic device indicated for the in vivo imaging of ocular structures in the anterior segment. The Tomey Cornea/Anterior Segment OCT CASIA2 is indicated as an aid in the visualization of anterior segment findings.The LENSTAR LS 900 is a non-invasive, non-contact OLCR (Optical Low Coherence Reflectometry) Biometer used for obtaining ocular measurements and performing calculations to assist in the determination of the appropriate power and type of IOL (intraocular lens) for implantation after removal of the natural crystalline lens following cataract removal. The LENSTAR LS 900 measures: *Axial eye length Corneal thickness Anterior chamber depth *Aqueous depth *Lens thickness *Radii of curvature of flat and steep meridian *Axis of the flat meridian *White to white distance *Pupil diameterNo new intended use; measurements align with K213265's imaging, using identical swept-source OCT. Differences from LENSTAR (OCT vs. OLCR, no axial length) are immaterial, focusing on anterior segment, with no safety/effectiveness impact.
Technological Characteristics - OCT ImagingSwept Source OCTSwept Source OCTOptical Low Coherence Reflectometry (OLCR)Same as predicate

Page 8

K250553
510(k) Summary
Page 4 of 12

ClassificationNew Device: Tomey CASIA2 with Updated SoftwarePrimary Predicate: Tomey Cornea/Anterior Segment OCT CASIA2 (K213265)Reference Device: LENSTAR, MODEL LS900 (K082891)Discussion Analysis of Differences Affect Safety or Effectiveness
- Scan Rate50,000 A-Scan/s50,000 A-Scan/sUnknownSame as predicate
- Wavelength1310 nm1310 nm820 nm for eye length, 950 nm for keratometrySame as predicate
- Power Output<6 mW<6 mW<0.6 mW for eye length, not specified for keratometrySame as predicate
- Field of View/Measurement RangeDepth: 13mm Transverse: Radial scan: 16 mm, Raster scan: 12mm x 12mmDepth: 13mm Transverse: Radial scan: 16 mm, Raster scan: 12mm x 12mmN/ASame as predicate
- Focus Range-18 D to +15 D-18 D to +15 DNot specifiedSame as predicate
SoftwareUpdated for quantitative measurementsExisting softwareNot detailedSame as predicate; new outputs (e.g., CCT, ACD) derived from K213265's OCT images, no new risks.
Measures Axial LengthNoNoYesSame as predicate; omission vs. LENSTAR immaterial, focuses on anterior segment, no safety impact.
Obtains radiii of curvature of flat and steep meridians and axis of the flat meridian and steep meridianUses Swept Source OCT TechnologyN/A. Does not obtain measurementsUses keratometry dotsSimilar to reference device. Both devices obtain radii of curvature of flat and steep merdians and axis of the flat merdian and steep meridian
Image Display / Scan TypesVisualizes the anterior segment structures by OCT technology.Visualizes the anterior segment structures by OCT technology.Provides front image of the eyeSame as Predicate
Qualitative imagingVisualizes the anterior segment of the eye such as the cornea, anterior chamber, angle, iris, and lens by OCT technology. Also provides the front image of the eye.Visualizes the anterior segment of the eye such as the cornea, anterior chamber, angle, iris, and lens by OCT technology. Also provides the front image of the eye.Provides the front image of the eyeSame as Predicate

Page 9

K250553
510(k) Summary
Page 5 of 12

ClassificationNew Device: Tomey CASIA2 with Updated SoftwarePrimary Predicate: Tomey Cornea/Anterior Segment OCT CASIA2 (K213265)Reference Device: LENSTAR, MODEL LS900 (K082891)Discussion Analysis of Differences Affect Safety or Effectiveness
Quantitative measurementsSimilarities with Secondary Predicate Device: The subject device provides the following measurements using OCT imaging, which is visualized images derived from signals using optical coherence technology as the secondary predicate: - Corneal thickness - Anterior chamber depth - Aqueous depth (provided as "ACD(Endo)") - Lens thickness The subject device provides the following measurements using the front image of the eye: - White to white distance - Pupil diameter Differences The subject device does not provide the axial length, unlike the secondary predicate. The subject device uses OCT technology to obtain the radii of curvature of flat and steep meridian and axis of the flat meridian and steep meridian, while the secondary predicate utilizes keratometry dots. Agreement between devices were confirmed in a clinical study.NoneProvides the following measurements using optical coherence technology: - Axial eye length - Corneal thickness - Anterior chamber depth - Aqueous depth - Lens thickness Provides the following measurements using keratometry dots: - Radii of curvature of flat and steep meridian - Axis of the flat meridian Provides the following measurements from the front image of the eye: - White to white distance - Pupil diameterNo new risks; outputs use K213265's OCT, matching LENSTAR where applicable, with no safety/effectiveness impact.

Page 10

K250553
510(k) Summary
Page 6 of 12

ClassificationNew Device: Tomey CASIA2 with Updated SoftwarePrimary Predicate: Tomey Cornea/Anterior Segment OCT CASIA2 (K213265)Reference Device: LENSTAR, MODEL LS900 (K082891)Discussion Analysis of Differences Affect Safety or Effectiveness
ResolutionAxial: 10μm or less (in tissue) Transverse: 30μm or less (in air)Axial: 10μm or less (in tissue) Transverse: 30μm or less (in air)N/ASame as Predicate
Voltage and Frequency100 to 240V AC 50 / 60Hz100 to 240V AC 50 / 60Hz100 to 240V AC 50 / 60HzSame
ConclusionThe subject device has the same hardware as its predicate, Cornea/Anterior Segment OCT CASIA2 (K213265). The subject device has updated software to provide quantitative measurements, including CCT, ACD, and LT evaluated against LENSTAR, MODEL LS900 (K082891) in a clinical study, with no new safety or effectiveness risks.

Page 11

K250553
510(k) Summary
Page 7 of 12

7. Non-Clinical and/or Clinical Tests Summary and Conclusions

Software documentation was provided as recommended by FDA's Guidance "Guidance for the Content of Premarket Submissions for Software Contained inMedical Devices."

Documentation regarding cybersecurity was submitted as recommended by FDA's Guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". Risk assessment regarding security and safety of the device was conducted in accordance with ISO 14971:2019.

Laser safety testing for the light sources used in CASIA2 were conducted to demonstrate compliance to ANSI Z80.36:2021.

Bench tests were conducted to verify, OCT specifications such as axial and transverse resolution, accuracy of transverse length and accuracy of angle. The device met all predetermined acceptance criteria.

Biocompatibility of the device was demonstrated by ISO 10993-1:2018.

Tests for electrical safety (IEC60601-1 Ed.3.2) and electromagnetic compatibility (IEC 60601-1-2 Ed.4.1) were performed. The CASIA2 device met the relevant requirements of the applied standards.

Clinical Testing:

Precision and agreement testing was performed. A total of 224 subjects were enrolled in the study, including 55 subjects in the normal group, 60 subjects in the cataract group, and 109 subjects in the special eyes (eyes without a natural lens or eyes containing artificial materials) group. All the subjects enrolled in the study completed the study. Therefore, the Safety Analysis Set and Full Analysis Set each comprised of 224 subjects. The clinical site had 3 device operators trained on the devices used in the study. The site was provided with 3 CASIA2 devices and 3 LenStar LS900 devices. Each CASIA2 device was paired with 1 of the LenStar LS900 devices and a device operator was assigned to each device pair to create 3 distinct operator/device configurations. Subjects were randomized to the device configuration sequences and to the starting device within each configuration. Each randomized subject was randomly assigned to one of the 3 configuration sequences:

  • 1, 2, 3
  • 2, 3, 1, or
  • 3, 1, 2

Additional scans were taken at the operator's discretion if image quality was unacceptable based on the device DFU and the Tomey CASIA2 Reference Guide and included, missing scans, truncated scans, image defocus, floaters, presence of eye blinks, eye motion, etc. Each device operator had up to 3 attempts to obtain an acceptable scan for each of the required scans.

Population characteristics:

The mean (SD) age in the Full Analysis Set was 34.7 (9.53) years for the normal subjects, 63.6 (9.73) years for the cataract subjects, and 69.2 (10.23) years for the special eye subjects. The overall mean age for all subjects was 59.2 (17.33) years, with the majority of subjects aged <65 years old (n = 113, 50.4%). All the subjects had age < 65 years (n = 55, 100%) in normal group, while majority of subjects aged >= 65 years in cataract group (n = 31, 61.7%) and special eyes group (n = 80, 73.4%) respectively. A total of 100 males, 123 females, and 1 unreported gender participated (44.6%, 54.9%, and 0.4% respectively) in this study, and the majority of subjects were white (n = 180, 80.4%) and not Hispanic or Latino (n = 181, 80.8%).

Page 12

K250553
510(k) Summary
Page 8 of 12

The study subject population included a full range of central corneal thickness (CCT) and anterior chamber depth (ACD). The normal range of CCT is generally depicted as 524 µm (0.524 mm) to 564 µm (0.564 mm) and the study population included a range of 410 to 655, with more than 25% of subjects below the normal range and more than 25% of subjects above the normal range based upon the LS900 measurement for the first configuration. For ACD, the vast majority of measurements in the special eyes group exceeded 4 mm, and ~10% subjects in the normal and cataract eye groups combined have ACD values less than 3 mm based upon the LS900 measurement for the first configuration.

Safety Evaluation:

There were no adverse events reported in this study. The subjects of this study had no notable or unexpected/untoward assessments for the safety measurements for visual acuity, undilated fundus, and intraocular pressure. Two subjects in the special eyes group experienced abnormal clinically significant (CS) findings in the slit lamp examination during the study. One subject had a CS finding of arcus senilis in the cornea in the right eye. A separate subject had a CS finding of ptosis in the eyelid in the left eye. Both the subjects completed the study.

Effectiveness:

This study evaluated the agreement and precision of the CASIA2 OCT device with the CASIA2 (K213265) as the predicate device and the LENSTAR, MODEL LS900 (K082891) as the reference device.

Agreement:

CASIA2 Corneal Map Scan – Two (2) endpoints (CCT and ACD) for agreement analysis between CASIA2 corneal map compared to LS900 in each of the cohort including pooled population (or all subjects), normal subjects, cataract subjects, and subjects with special eyes are summarized in Table 18.

Page 13

K250553
510(k) Summary
Page 9 of 12

Figure 14.4.1.12
Agreement Analysis - Bland Altman Plot of Observed Data (All Configurations) - Paired Difference Between Devices for CASIA2 Cornea Map (Test) and LS900 (Predicate) for Central Corneal Thickness (μm)
Agreement Analysis Set

Subject Population: All Subjects (N=206)

[THIS IS FIGURE: Bland-Altman plot showing paired differences vs paired mean for central corneal thickness measurements, with data points scattered around zero and confidence limits marked]

Figure 14.4.1.13
Agreement Analysis - Bland Altman Plot of Observed Data (All Configurations) - Paired Difference Between Devices for CASIA2 Cornea Map (Test) and LS900 (Predicate) for Anterior Chamber Depth (mm)
Agreement Analysis Set

Subject Population: All Subjects (N=201)

[THIS IS FIGURE: Bland-Altman plot showing paired differences vs paired mean for anterior chamber depth measurements, with data points scattered around zero and confidence limits marked]

CASIA2 Lens scan. The endpoint, lens thickness (LT) for agreement analysis between CASIA2 Lens scan and LS900 is summarized in Table 21.

Page 14

K250553
510(k) Summary
Page 10 of 12

Precision Study:

CASIA2 Corneal Map - Two (2) endpoints, CCT and ACD were evaluated for precision analysis between CASIA2 Corneal Map and LS900 in each cohort (pooled population of all subjects, normal subjects, cataract subjects, and subjects with special eyes) are summarized in Table 27.

Table 21 Summary of the Limits of Agreement - CASIA2 Lens Scan (Test) versus LS900 (Predicate) – Agreement Analysis Set

NCASIA2 Mean (SD)LS900 Mean (SD)Mean Difference (SD)95% LOA95% CI Lower LOA95% CI Upper LOA
Lens Thickness (mm)
Subject Population: All Subjects (N=224)
All Configurations1224.03 (0.705)3.86 (1.049)0.16 (0.719)(-1.26, 1.59)(-1.48, -1.04)(1.36, 1.81)
Subject Population: Normal (N=55)
All Configurations533.83 (0.304)3.82 (0.310)0.01 (0.112)(-0.21, 0.24)(-0.27, -0.16)(0.18, 0.29)
Subject Population: Cataract (N=60)
All Configurations584.43 (0.328)4.47 (0.338)-0.04 (0.126)(-0.29, 0.21)(-0.35, -0.23)(0.16, 0.27)
Subject Population: Special Eyes (N=109)
All Configurations112.82 (1.472)0.85 (0.174)1.97 (1.468)(-1.30, 5.24)(-3.01, 0.41)(3.53, 6.95)

Abbreviations: CI = Confidence Interval, NC = Not Calculable, SD = Standard Deviation.
Note: The Agreement Analysis Set used the first acceptable paired (from both devices) scans for analyses. N was the number of eyes with acceptable scans in the Agreement Analysis Set. Mean Difference was calculated as test minus predicate device (CASIA2 – LS900). The SD of the Mean Difference was based on paired t-test differences. Limits of Agreement (LOA) were calculated as Mean Difference +/- t x Mean Difference SD, where t was the critical value from t-distribution with n-1 degrees of freedom. The 95% CI of the upper and lower bounds of the LOA were calculated as LOA ± t x (SQRT(3/n) x s; t was the 97.5% critical value of the t distribution with n-1 degrees of freedom, s was the standard deviation, and n was the number of subjects.
Reference: Table 14.2.1.2.2, Listing 16.2.6.2, 16.2.6.4.

Figure 14.4.1.14
Agreement Analysis - Bland Altman Plot of Observed Data (All Configurations) - Paired Difference Between Devices for CASIA2 Lens Thickness (Test) and LS900 (Predicate) for Lens Thickness (mm)
Agreement Analysis Set

Subject Population: All Subjects (N=122)

[THIS IS FIGURE: Bland-Altman plot showing paired differences vs paired mean for lens thickness measurements, with data points and confidence limits marked]

Page 15

K250553
510(k) Summary
Page 11 of 12

Table 1 Summary of Repeatability and Reproducibility - CASIA2 Corneal Map (Test) versus LS900 (Predicate) - Precision Analysis Set

Device Parameter (unit)NOverall Mean%CVRepeatabilityReproducibility
%CV Lower 95% CI%CV Upper 95% CI%CV%CV Lower 95% CI%CV Upper 95% CI
Subject Population: All Subjects (N=224)
CASIA2
Central Corneal Thickness (μm)138538.8610.18%0.17%0.19%0.32%0.31%0.33%
Anterior Chamber Depth (mm)1353.9491.01%0.97%1.07%1.13%1.08%1.18%
LS900
Central Corneal Thickness (μm)138545.0130.36%0.34%0.38%0.53%0.51%0.56%
Anterior Chamber Depth (mm)1353.8613.09%2.94%3.24%4.35%4.17%4.54%
Subject Population: Normal (N=55)
CASIA2
Central Corneal Thickness (μm)45534.210.18%0.17%0.20%0.33%0.31%0.36%
Anterior Chamber Depth (mm)443.6810.45%0.42%0.49%0.82%0.76%0.89%
LS900
Central Corneal Thickness (μm)45540.020.39%0.36%0.43%0.66%0.62%0.71%
Anterior Chamber Depth (mm)443.5810.81%0.74%0.88%1.41%1.32%1.53%
Subject Population: Cataract (N=60)
CASIA2
Central Corneal Thickness (μm)54543.0530.17%0.16%0.18%0.31%0.29%0.33%
Anterior Chamber Depth (mm)533.4590.15%0.14%0.16%0.37%0.35%0.40%
LS900
Central Corneal Thickness (μm)54549.290.32%0.29%0.34%0.47%0.44%0.51%
Anterior Chamber Depth (mm)533.3680.86%0.80%0.94%1.48%1.38%1.58%
Subject Population: Special Eyes (N=109)
CASIA2
Central Corneal Thickness (μm)39538.4220.20%0.18%0.21%0.32%0.30%0.35%
Anterior Chamber Depth (mm)384.9411.48%1.35%1.63%1.54%1.42%1.67%
LS900
Central Corneal Thickness (μm)39544.8520.38%0.35%0.42%0.45%0.42%0.49%
Anterior Chamber Depth (mm)384.8734.51%4.13%4.97%6.29%5.82%6.83%

Abbreviations: CI = Confidence Interval, CV = Coefficient of Variation, mm = millimeters, μm = microns.
Note: The Precision Analysis Set used subjects with 9 acceptable scans per device. N was the number of eyes with acceptable scans in the Precision Analysis Set. The crossed analysis model was used to calculate reproducibility limits, and repeatability limits included configuration, subject, and configuration x subject interaction as variance components using restricted maximum likelihood method (REML) for estimation. The repeatability limit was 2.8 times the repeatability SD, which was the square root of the residual within subject variance component. The reproducibility limit was 2.8 times the reproducibility SD, which was the square root of the sum of the variance components of configuration, configuration x subject interaction and residual within subject. CV% was calculated as the Repeatability or Reproducibility SD/Overall Mean.
Reference: Table 14.2.2.2.2, Listing 16.2.6.2, 16.2.6.4.

CASIA2 Lens scan. The endpoint lens thickness (LT) for precision analysis in each cohort is summarized in Table 28.

Page 16

K250553
510(k) Summary
Page 12 of 12

Table 28 Summary of Repeatability and Reproducibility - CASIA2 Lens Scan (Test) versus LS900 (Predicate) - Precision Analysis Set

RepeatabilityReproducibility
DeviceNOverall Mean%CV%CV Lower 95% CI%CV Upper 95% CI%CV%CV Lower 95% CI%CV Upper 95% CI
Subject Population: All Subjects (N=224)
Lens Thickness (mm)
CASIA2764.1161.05%0.99%1.13%1.39%1.32%1.48%
LS900764.1161.01%0.95%1.08%2.35%2.23%2.49%
Subject Population: Normal (N=55)
Lens Thickness (mm)
CASIA2363.7980.85%0.78%0.94%1.54%1.42%1.68%
LS900363.7821.02%0.93%1.12%2.48%2.30%2.71%
Subject Population: Cataract (N=60)
Lens Thickness (mm)
CASIA2404.4031.16%1.07%1.28%1.28%1.19%1.39%
LS900404.4171.00%0.92%1.10%2.25%2.08%2.43%

Abbreviations: CI = Confidence Interval, CV = Coefficient of Variation, mm = millimeters.
Note: The Precision Analysis Set used subjects with 9 acceptable scans per device. N was the number of eyes with acceptable scans in the Precision Analysis Set. The crossed analysis model was used to calculate reproducibility limits, and repeatability limits included configuration, subject, and configuration x subject interaction as variance components using restricted maximum likelihood method (REML) for estimation. The repeatability limit was 2.8 times the repeatability SD, which was the square root of the residual within subject variance component. The reproducibility limit was 2.8 times the reproducibility SD, which was the square root of the sum of the variance components of configuration, configuration x subject interaction and residual within subject. CV% was calculated as the Repeatability or Reproducibility SD/Overall Mean.
Reference: Table 14.2.2.2.3, Listing 16.2.6.3, 16.2.6.4.

Repeatability %CV and reproducibility %CV for CASIA2 Lens Scans could not be assessed for the special eye subjects due to unacceptable scans.

8. Conclusion

Therefore, based on the same intended use and similar technological characteristics with substantial equivalence to the reference device confirmed with performance testing, the Tomey CASIA2 is technologically and functionally equivalent to the predicate device, Tomey CASIA2 (K213265) and the reference device LENSTAR, MODEL LS900 (K082891). The differences between the proposed device, CASIA2, and the predicate and reference device are not significant and do not raise new issues of safety or effectiveness of the device. The Tomey CASIA2 is as safe and effective as its predicate and reference devices, and thus may be considered substantially equivalent.

§ 886.1570 Ophthalmoscope.

(a)
Identification. An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.(b)
Classification. Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.