K Number
K250432

Validate with FDA (Live)

Date Cleared
2025-05-15

(90 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The COLONOVIDEOSCOPE OLYMPUS CF-EZ1500DL & CF-EZ1500DI are intended to be used with a video system center, endoscope position detecting unit, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The COLONOVIDEOSCOPE OLYMPUS CF-EZ1500DL/I are indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-EZ1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-EZ1500 is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

Device Description

The endoscope consists of three parts: the control section, the insertion section, and the connector section.

  1. Control section: The UP/DOWN angulation control knob and the RIGHT/LEFT angulation control knob on the control section are connected to the tip of the bending section by a series of wires. By operating the angulation control knobs, the bending section at the distal end bends vertically or parallel to guide the distal end for insertion and observation. The observation mode can be selected by focus switching function, "near focus mode" featuring innovative resolving power for close observation or "normal focus mode" for normal observation. To realize the dual focus mechanism, Voice Coil Motor (VCM) is incorporated as an actuator. The endoscope contains a cylinder to attach a suction valve for suction and air/water valve. Depressing the suction valve will allow the physician to use the endoscope to suction any fluids which are obscuring a good view of the tissue. Therapeutic instruments can be passed through the instrument channel for performing endoscopic biopsy and other therapies. Depressing the air/water valve will allow the doctor to feed water through the endoscope for lens washing. It also can be operated to feed air for removing any fluids or debris adhering to the objective lens.

  2. Insertion section: The insertion section has main parts including the image guide, light guides that bring light from the video system center through the endoscope, and instrument channel where therapeutic tools can be pushed in and out (also the suction channel).

  3. Connector section: The connector section connects the endoscope with the video system center (CV-1500) through the universal cord.

AI/ML Overview

It appears crucial information regarding the AI or software components' acceptance criteria and the specific study proving the device meets these criteria is missing from the provided FDA 510(k) clearance letter for the Olympus Endoscopes (K250432).

The document details the substantial equivalence of the new endoscopes (CF-EZ1500DL/DI and GIF-EZ1500) to their predicates based on physical characteristics, intended use, and general performance. It mentions extensive non-clinical testing, including:

  • Reprocessing validation
  • Biocompatibility testing
  • Software verification and validation (mentioning FDA guidance for software in medical devices, but not detailing specific performance criteria or tests related to AI/image analysis capabilities)
  • Electrical safety and EMC
  • Bench testing (performance characteristics like DOF/FOV, Resolution, Noise, etc.)
  • Animal study to confirm WLI, NBI, RDI, and TXI performance.
  • Risk management

However, the prompt asks for specifics related to an AI/Software component's acceptance criteria and its performance study. The provided text focuses on the endoscope hardware itself and its basic image capture and display functionalities. There is no mention of any AI-driven image analysis, diagnostic augmentation, or automated detection capabilities that would require the kind of clinical performance study details (e.g., sample size, ground truth, expert adjudication, MRMC studies) outlined in the prompt.

The "Indications for Use" for these endoscopes clearly state their purpose is for "endoscopy and endoscopic surgery" for visualization. They are tools for observation and facilitating therapeutic accessories. They are not described as having an AI component for automated analysis or diagnosis.

Therefore, since the provided document does not indicate the presence of an AI or advanced software feature requiring a separate performance study with associated acceptance criteria, I cannot fulfill most of the requested points.

Here's what can be inferred or stated based on the provided document and the information that is explicitly NOT present:


Based on the provided document, there is no evidence of an AI-driven or advanced image analysis software component in the Olympus Endoscopes (CF-EZ1500DL/DI and GIF-EZ1500) that would necessitate the type of acceptance criteria and performance study described in the prompt.

The 510(k) clearance focuses on demonstrating substantial equivalence to predicate endoscopes by comparing physical and operational characteristics, and through general performance testing of the device hardware and its fundamental image capture capabilities.

However, if we were to hypothetically assume there was an AI/Software component (e.g., for polyp detection), and based on common FDA expectations for such devices, here's how the answer would look, highlighting what's missing:


Acceptance Criteria and Study Proving Device Performance (Hypothetical AI Component)

Note: The provided FDA 510(k) clearance letter for K250432 does not detail any specific AI or advanced image analysis software functionality, nor does it contain the performance study metrics for such a component. The following table and points are constructed under the assumption of such a component existing, to illustrate what an ideal response would contain if the information were available in the document.

1. Table of Acceptance Criteria and Reported Device Performance (Hypothetical AI Component)

Since no AI/software component is described in the provided document, this table cannot be filled with actual data. Below is a hypothetical example of what such a table would contain if an AI component for, e.g., polyp detection, were present.

Acceptance Criterion (Hypothetical)Threshold (Hypothetical)Reported Device Performance (Hypothetical)Pass/Fail (Hypothetical)
Sensitivity for Polyp Detection≥ 90%92.5%Pass
Specificity for Polyp Detection≥ 80%85.1%Pass
False Positive Rate per Case≤ 2.0 per case1.8 false positives per casePass
Latency for AI Annotation≤ 100 ms50 msPass

2. Sample Size and Data Provenance for Test Set

  • Sample Size for Test Set: Not specified (as no AI performance study is detailed). Typical AI/CADe clearances involve hundreds to thousands of cases.
  • Data Provenance: Not specified. For AI devices, this would typically include details like:
    • Country of Origin (e.g., multi-center, US, Europe, Asia)
    • Retrospective or Prospective collection
    • Diversity of patient demographics, disease prevalence, equipment used, etc.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not specified. For AI/CADe devices, often 3 or more independent experts are used.
  • Qualifications of Experts: Not specified. Typically, this would involve board-certified gastroenterologists or colorectal surgeons with significant experience (e.g., 5-10+ years) in endoscopic procedures, potentially sub-specialized in screening/surveillance or interventional endoscopy.

4. Adjudication Method for Test Set

  • Adjudication Method: Not specified. Common methods for AI medical device clearances include:
    • Majority Rule: E.g., 2 out of 3, or 3 out of 5 experts agree.
    • Consensus Reading: Experts review and discuss cases to reach a unanimous decision.
    • 2+1 or 3+1: Initial reads by 2 or 3 experts, with a senior adjudicator resolving discrepancies.
    • Pathology Correlation: Where possible, biopsy/histology results serve as the definitive ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study Conducted?: No mention of an MRMC study. The document states "No clinical study was performed to demonstrate substantial equivalence." This implies that any human-in-the-loop performance evaluation of an AI component was not required or not conducted for this 510(k).
  • Effect Size of Human Improvement (if applicable): Not applicable as no MRMC study is mentioned. If performed, this would report metrics like the change in reader sensitivity, specificity, or reading time with AI assistance compared to without.

6. Standalone (Algorithm Only) Performance Study

  • Standalone Performance Conducted?: No explicit mention of a standalone algorithm performance study with specific metrics (e.g., AUC, sensitivity, specificity on a defined dataset). The "Software verification and validation" generally refers to software engineering and quality assurance, not necessarily clinical performance of an AI algorithm itself.

7. Type of Ground Truth Used

  • Type of Ground Truth: Not specified. If an AI component were present for lesion detection, ground truth would typically be established by:
    • Expert Consensus: As described in point 3.
    • Pathology/Histology: Biopsy results confirming the presence or absence of lesions. This is often considered the gold standard for many gastrointestinal applications.
    • Clinical Outcomes Data: Long-term follow-up to confirm diagnoses, though less common for initial device clearance.

8. Sample Size for Training Set

  • Sample Size for Training Set: Not specified. This information is typically not included in FDA clearance letters unless directly relevant to a novel AI claim. Training set sizes for medical AI can range from thousands to hundreds of thousands of images/cases.

9. How Ground Truth for Training Set Was Established

  • Ground Truth for Training Set: Not specified. It would generally be established through similar methods as the test set (e.g., expert annotations, pathological confirmation), often with additional data augmentation or curation processes.

FDA 510(k) Clearance Letter - Olympus Endoscopes

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 15, 2025

Olympus Medical Systems Corporation
℅ Teffany Hutto
Regulatory Affairs Program Manager
Olympus Corporation of the Americas
800 West Park Drive
Westborough, Massachusetts 01581

Re: K250432
Trade/Device Name: Colonovideoscope (CF-EZ1500DL); Colonovideoscope (CF-EZ1500DI); Gastrointestinal Videoscope (GIF-EZ1500)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FDF, FDS, NWB
Dated: February 14, 2025
Received: February 14, 2025

Dear Teffany Hutto:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250432 - Teffany Hutto Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250432 - Teffany Hutto Page 3

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K250432

Device Name
Colonovideoscope (CF-EZ1500DL); Colonovideoscope (CF-EZ1500DI); Gastrointestinal Videoscope (GIF-EZ1500)

Indications for Use (Describe)

The COLONOVIDEOSCOPE OLYMPUS CF-EZ1500DL & CF-EZ1500DI are intended to be used with a video system center, endoscope position detecting unit, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The COLONOVIDEOSCOPE OLYMPUS CF-EZ1500DL/I are indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-EZ1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-EZ1500 is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 1 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Date Prepared: 04/11/2025

510(k) Summary

CONTACT DETAILS

■ 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP.
2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan

■ Contact Person: Teffany Hutto
Olympus Corporation of the Americas
3500 Corporate Parkway PO Box 610
Center Valley, PA 18034-0610, USA
Phone: 512-508-6550
Fax: 484-896-7128
Email: teffany.hutto@olympus.com

• Manufacturing site: Aizu Olympus Co., Ltd.
3-1-1 Niiderakita, Aizuwakamatsu-shi, Fukushima 965-8520, Japan

DEVICE NAME

Device Trade NameOLYMPUS CF-EZ1500DL
Common NameColonovideoscope
Regulation Number876.1500
Classification NameEndoscope and accessories
Regulatory ClassII
Product CodeFDF, NWB
Device Trade NameOLYMPUS CF-EZ1500DI
Common NameColonovideoscope
Regulation Number876.1500
Classification NameEndoscope and accessories
Regulatory ClassII
Product CodeFDF, NWB

Page 6

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 2 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Device Trade NameOLYMPUS GIF-EZ1500
Common NameGastrointestinal Videoscope
Regulation Number876.1500
Classification NameEndoscope and accessories
Regulatory ClassII
Product CodeFDS, NWB

LEGALLY MARKETED PREDICATE DEVICE

Device Name510(k) Submitter510(k) No.Product Code
Colonovideoscope Olympus CF-HQ1100L/IOlympus MedicalK222584FDF
Gastrointestinal Videoscope OLYMPUS GIF-1100Olympus MedicalK222584FDS

DEVICE DESCRIPTION SUMMARY

• General Description of the subject device

The COLONOVIDEOSCOPE OLYMPUS CF-EZ1500DL & CF-EZ1500DI is intended to be used with a video system center, Endoscope position detecting unit, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The Olympus COLONOVIDEOSCOPE CF-EZ1500DL/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-EZ1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Note: The only differences between CF-EZ1500DL and CF-EZ1500DI are the insertion section working length of each scope and the total length of each scope.

The Olympus GASTROINTESTINAL VIDEOSCOPE GIF-EZ1500 is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

Page 7

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 3 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

• Principle of Operation

The endoscope consists of three parts: the control section, the insertion section, and the connector section. The basic principle including user interface and operation for the procedure of the endoscope is identical to that of the predicate device.

1) Control section

The UP/DOWN angulation control knob and the RIGHT/LEFT angulation control knob on the control section is connected to the tip of the bending section by a series of wires. By operating the UP/DOWN angulation control knob and the RIGHT/LEFT angulation control knob, the bending section at the distal end bends vertically or parallel to guide the distal end for insertion and observation.

The observation mode can be selected by focus switching function, "near focus mode" featuring innovative resolving power for close observation or "normal focus mode" for normal observation. To realize the dual focus mechanism, Voice Coil Motor (VCM) is incorporated as an actuator.

The endoscope contains a cylinder to attach a suction valve for suction and air/water valve. Depressing the suction valve will allow the physician to use the endoscope to suction any fluids which are obscuring a good view of the tissue. Therapeutic instruments can be passed through the instrument channel for performing endoscopic biopsy and other therapies. Depressing the air/water valve will allow the doctor to feed water through the endoscope for lens washing. It also can be operated to feed air for removing any fluids or debris adhering to the objective lens.

2) Insertion section

The insertion section has main parts including the image guide, light guides that bring light from the video system center through the endoscope, and instrument channel where therapeutic tools can be pushed in and out (also the suction channel).

3) Connector section

The connector section connects the endoscope with the video system center (CV-1500) through the universal cord.

Page 8

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 4 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

INTENDED USE/INDICATIONS FOR USE

The COLONOVIDEOSCOPE OLYMPUS CF-EZ1500DL & CF-EZ1500DI are intended to be used with a video system center, endoscope position detecting unit, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The COLONOVIDEOSCOPE OLYMPUS CF-EZ1500DL/I are indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-EZ1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-EZ1500 is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

INDICATIONS FOR USE COMPARISON

CF-EZ1500DL/I have the same intended use and indications for use as compared to the predicate device. The subject device is also equivalent to the predicate device for direction of view, field of view, geometric distortion, diameters of insertion section, diameters of instrument channel, same control section, configuration of control section and connector section, bending section angulation range, total length, functions (excluding electronic shutter function), Scope ID, and RFID. The subject device also utilizes the same manufacturing materials, and the same methods of reprocessing.

GIF-EZ1500 is has the same intended use and indications for use as the predicate devices. The subject device also is equivalent to the predicate devices for direction of view, field of view, geometric distortion, distal end feature, insertion section working length, instrument channel diameters, configuration of control section and connector section, bending section of angulation range, total length, functions (excluding electronic shutter function and focus switching function), Scope ID, and RFID. The subject device also utilizes the same manufacturing materials, and the same methods of reprocessing.

Page 9

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 5 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

TECHNOLOGICAL COMPARISON

CF-EZ1500DL/I differs from the predicate device for image sensor type and number of pixels, depth of field, and airflow rate. CF-EZ1500DL/I also does not include an electronic shutter function where the predicate does include it.

GIF-EZ1500 differs from the predicate devices for image sensor type, number of pixels, and pixel size, depth of field, outer diameters of distal end, insertion section, insertion tube, and airflow rate. The subject device does not include electronic shutter function where the predicate device does include it. The subject device has a focus switching function, whereas the predicate device does not.

Subject Device CF-EZ1500DL/IPredicate Device CF-HQ1100L/I (K222584)Reference Device CF-HQ190L (K131780)
510(k) numberThis submissionK222584K131780
Regulation number876.1500876.1500876.1500
Regulatory classClass IIClass IIClass II
Product codeFDF (colonoscope and accessories, flexible/rigid) NWB (endoscope, accessories, narrow band spectrum)FDF (colonoscope and accessories, flexible/rigid) NWB (endoscope, accessories, narrow band spectrum)FDF (colonoscope and accessories, flexible/rigid) NWB (endoscope, accessories, narrow band spectrum)
Classification panelGastroenterology and urologyGastroenterology and urologyGastroenterology and urology
Common nameCOLONOVIDEOSCOPECOLONOVIDEOSCOPECOLONOVIDEOSCOPE
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.OLYMPUS MEDICAL SYSTEMS CORP.OLYMPUS MEDICAL SYSTEMS CORP.
Indications for useThe COLONOVIDEOSCOPE OLYMPUS CF-EZ1500DL & CF-EZ1500DI is intended to be used with a video system center, Endoscope position detecting unit, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.This instrument is intended to be used with an Olympus video system center, Endoscope position detecting unit, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The OlympusThis instrument is intended to be used with an Olympus video system center, endoscope position detecting unit, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The EVIS EXERA III

Page 10

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 6 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device CF-EZ1500DL/IPredicate Device CF-HQ1100L/I (K222584)Reference Device CF-HQ190L (K131780)
The Olympus COLONOVIDEOSCOPE CF-EZ1500DL/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).COLONOVIDEOSCOPE CF-HQ1100DL/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).COLONOVIDEOSCOPE CF-HQ190L/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).
Type- CMOS - RGB color filter - Sequential read image signal- CCD - CYM color filter - Sequential read image signal- CCD - CYM color filter - Sequential read image signal
Direction of View
Field of ViewNormal focus mode: 170° Near focus mode: 160°Normal focus mode: 170° Near focus mode: 160°Normal focus mode: 170° Near focus mode: 160°
Depth of FieldNormal focus mode: WLI, NBI: 3 - 100 mm RDI: 7 - 100 mm Near focus mode: WLI, NBI: 1.5 - 5.5 mm RDI: 2.5 - 5.5 mmNormal focus mode: WLI, NBI, RDI: 5 - 100 mm Near focus mode: WLI, NBI, RDI: 2 - 6 mmNormal focus mode: WLI, NBI, RDI: 5 - 100 mm Near focus mode: WLI, NBI, RDI: 2 - 6 mm *RDI is available only connected to CV-1500
Magnification× 90 (OEV321UH) × 75 (OEV262H)× 75 (OEV321UH) × 60 (OEV262H)× 75 (OEV321UH) × 60 (OEV262H)
Distal end outer diameterø 13.2 mmø 13.2 mmø 13.2 mm
Maximum insertion section outer diameterø 14.9 mmø 14.9 mmø 14.9 mm

Page 11

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 7 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device CF-EZ1500DL/IPredicate Device CF-HQ1100L/I (K222584)Reference Device CF-HQ190L (K131780)
Distal end enlarged1 Air/water nozzle 2 Light guide lens 3 Objective lens 4 Instrument channel outlet 5 Auxiliary water channel1 Air/water nozzle 2 Light guide lens 3 Objective lens 4 Instrument channel outlet 5 Auxiliary water channel1 Air/water nozzle 2 Light guide lens 3 Objective lens 4 Instrument channel outlet 5 Auxiliary water channel
Insertion tube outer diameterø 12.8 mmø 12.8 mmø 12.8 mm
Insertion section working lengthL: 1680 mm I: 1330 mmL: 1680 mm I: 1330 mmL: 1680 mm I: 1330 mm
Range of the flexibility adjustmentThe rigidity in the most rigid condition is about twice that in the most flexible condition.The rigidity in the most rigid condition is about twice that in the most flexible condition.The rigidity in the most rigid condition is about twice that in the most flexible condition.
Channel inner diameterø 3.7 mmø 3.7 mmø 3.7 mm
Minimum channel inner diameterø 3.7 mmø 3.7 mmø 3.7 mm
Minimum visible distance4 mm (Normal focus mode)4 mm (Normal focus mode)4 mm (Normal focus mode)
Direction from which EndoTherapy accessories enter and exit the endoscopic image

Page 12

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 8 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device CF-EZ1500DL/IPredicate Device CF-HQ1100L/I (K222584)Reference Device CF-HQ190L (K131780)
Configuration
Airflow rate *At the setting of the air pressure as "high."30 cm³/s25 cm³/s25 cm³/s
Angulation rangeUP 180° DOWN 180° RIGHT 160° LEFT 160°UP 180° DOWN 180° RIGHT 160° LEFT 160°UP 180° DOWN 180° RIGHT 160° LEFT 160°
Total lengthL: 2005 mm I: 1655mmL: 2005 mm I: 1655mmL: 2005 mm I: 1655mm
Pre-freeze functionAvailableAvailableAvailable
Electronic zoom functionAvailableAvailableAvailable
Electronic shutter functionNot AvailableAvailableAvailable

Page 13

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 9 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device CF-EZ1500DL/IPredicate Device CF-HQ1100L/I (K222584)Reference Device CF-HQ190L (K131780)
Records of endoscope's informationAvailableAvailableAvailable
NBI observationAvailableAvailableAvailable
RDI observationAvailableAvailableNot Available *RDI is available only connected to CV-1500
High frequency cauterization treatmentAvailableAvailableAvailable
Endoscope position detecting functionAvailableAvailableAvailable
Passive bending functionAvailableAvailableAvailable
Flexibility adjustment functionAvailableAvailableAvailable
Zoom functionNot AvailableNot AvailableNot Available
Focus switching functionAvailableAvailableAvailable
Auxiliary water feeding functionAvailableAvailableAvailable
Transnasal insertionNot AvailableNot AvailableNot Available

Page 14

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 10 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device GIF-EZ1500Predicate Device GIF-1100 (K222584)Reference Device GIF-1TH190 (K232997)
510(k) numberThis submissionK222584K232997
Regulation number876.1500876.1500876.1500
Regulatory classClass IIClass IIClass II
Product codeFDS (gastroscope and accessories, flexible/rigid) NWB (endoscope, accessories, narrow band spectrum)FDS (gastroscope and accessories, flexible/rigid) NWB (endoscope, accessories, narrow band spectrum)FDS (gastroscope and accessories, flexible/rigid) NWB (endoscope, accessories, narrow band spectrum)
Classification panelGastroenterology and urologyGastroenterology and urologyGastroenterology and urology
Common nameGASTROINTESTINAL VIDEOSCOPEGASTROINTESTINAL VIDEOSCOPEGASTROINTESTINAL VIDEOSCOPE
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.OLYMPUS MEDICAL SYSTEMS CORP.OLYMPUS MEDICAL SYSTEMS CORP.
Indications for useThe GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-EZ1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The Olympus GASTROINTESTINAL VIDEOSCOPE GIF-EZ1500 is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).This instrument is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The Olympus GASTROINTESTINAL VIDEOSCOPE GIF-1100 is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE GIF-1TH190 is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

Page 15

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 11 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device GIF-EZ1500Predicate Device GIF-1100 (K222584)Reference Device GIF-1TH190 (K232997)
Type- CMOS - RGB color filter - Sequential read image signal- CCD - CYM color filter - Sequential read image signal- CCD - CYM color filter - Sequential read image signal
Direction of View
Field of ViewNormal focus mode: 140° Near focus mode: 140°140°140°
Depth of FieldNormal focus mode: WLI, NBI: 3 - 100 mm RDI: 7 - 100 mm Near focus mode: WLI, NBI: 1.5 - 5.5 mm RDI: 2.5 - 5.5 mmWLI, NBI, RDI: 2 - 100 mmWLI, NBI: 2 - 100 mm
Magnification× 100 (OEV321UH) × 85 (OEV262H)× 70(OEV321UH) × 55(OEV262H)× 65(OEV321UH) × 55(OEV262H)
Distal end enlarged1 Air/water nozzle 2 Light guide lens 3 Objective lens 4 Instrument channel outlet 5 Auxiliary water channel1 Air/water nozzle 2 Light guide lens 3 Objective lens 4 Instrument channel outlet 5 Auxiliary water channel1 Air/water nozzle 2 Light guide lens 3 Objective lens 4 Instrument channel outlet 5 Auxiliary water channel
Distal end outer diameterø 9.9 mmø 8.9 mmø 10.0mm
Maximum insertion section outer diameterø 12.5mmø 10.7mmø 13.0mm

Page 16

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 12 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device GIF-EZ1500Predicate Device GIF-1100 (K222584)Reference Device GIF-1TH190 (K232997)
Insertion tube outer diameterø 9.6 mmø 8.9 mmø 10.9mm
Insertion section working length1030 mm1030 mm1030 mm
Channel inner diameterø 2.8 mmø 2.8 mmø 3.7 mm
Minimum visible distance3 mm3 mm3 mm
Direction from which EndoTherapy accessories enter and exit the endoscopic image
Configuration

Page 17

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 13 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device GIF-EZ1500Predicate Device GIF-1100 (K222584)Reference Device GIF-1TH190 (K232997)
Configuration
Airflow rate *At the setting of the air pressure as "high."30 cm³/s25 cm³/s25 cm³/s
Angulation rangeUP 210° DOWN 90° RIGHT 100° LEFT 100°UP 210° DOWN 90° RIGHT 100° LEFT 100°UP: 210° DOWN: 90° RIGHT:100° LEFT:100°
Total length1350 mm1350 mm1350 mm
Pre-freeze functionAvailableAvailableAvailable
Electronic zoom functionAvailableAvailableAvailable
Electronic shutter functionNot AvailableAvailableAvailable
Records of endoscope's informationAvailableAvailableAvailable
NBI observationAvailableAvailableAvailable
RDI observationAvailableAvailableNot available
High frequency cauterization treatmentAvailableAvailableAvailable
Endoscope position detecting functionNot availableNot availableNot available

Page 18

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 14 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Subject Device GIF-EZ1500Predicate Device GIF-1100 (K222584)Reference Device GIF-1TH190 (K232997)
Passive bending functionNot availableNot availableNot available
Flexibility adjustment functionNot availableNot availableNot available
Zoom functionNot AvailableNot AvailableNot Available
Focus switching functionAvailableNot availableNot available
Auxiliary water feeding functionAvailableAvailableAvailable
Transnasal insertionNot availableNot availableNot available

NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS

Reprocessing validation testing

Reprocessing instruction and reprocessing method validation testing were conducted and documentation were provided as recommended by Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling".

Biocompatibility testing

Biocompatibility testing were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

  • Cytotoxicity Study Using the Colony Assay
  • Intracutaneous Study in Rabbits
  • Guinea Pig Maximization Sensitization Test

Page 19

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 15 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Software verification and validation

Software verification and validation testing were conducted, and documentations were provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing was conducted in accordance with the ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] and IEC 60601-2-18:2009 standards for safety and IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION for EMC.

Performance testing – Bench

Bench testing as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.

  • Durability
  • Thermal Safety
  • Photobiological Safety
  • Transportation
  • Color Performance
  • DOF/FOV
  • Distortion
  • Resolution
  • DOF
  • Noise and Dynamic Range
  • Image Intensity Uniformity
  • Video Latency
  • Rates of Air, Water, and Suction
  • Magnification
  • Automatic brightness adjustment
  • Pre-freeze
  • Functional Validation
  • Stability of EDOF Structure
  • Human Factors Evaluation

Page 20

Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500
CONFIDENTIAL

K250432
Page 16 of 16
Traditional 510(k) Notification
CF-EZ1500DL/I and GIF-EZ1500

Performance testing - Animal

Animal study was performed to confirm the White Light Imaging (WLI) and Narrow Band Imaging (NBI) performance, and the effectiveness of Red Dichromatic Imaging (RDI) and TeXture and color enhancement Imaging (TXI).

Performance testing - Clinical

No clinical study was performed to demonstrate substantial equivalence.

Risk management

Risk management was performed in accordance with ISO 14971:2019. The design verification tests and their acceptance criteria were identified and performed as a result of this risk management.

All testing demonstrated that the subject devices met design specifications, performed as intended, and was equivalent to the subject and reference devices.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.