K Number
K250410

Validate with FDA (Live)

Date Cleared
2025-06-02

(109 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GORE® Tri-Lobe Balloon Catheter is intended to assist in the dilatation of self-expanding endoprostheses in large diameter vessels.

Device Description

The GORE® Tri-Lobe Balloon Catheter is a compliant, tri-lobed polyurethane balloon catheter. The lobed design of the balloon catheter is designed for inflation without complete blockage of aortic blood flow. The three polyurethane balloons are mounted on the leading end of a multi-lumen catheter shaft. Radiopaque markers indicate the balloon edges. Each of the three inflation lumens is in communication with one of the balloons. The inflation port is in communication with all of the inflation lumens and is affixed with a luer lock. The guidewire lumen allows introduction of a 0.035" (0.89 mm) diameter guidewire for over-the-wire access. The trailing end of the guidewire lumen is affixed with a flushing/guidewire port with a luer lock, used for flushing the guidewire lumen. A Tuohy-Borst valve is integrated into the trailing end of the guidewire lumen.

The GORE® Tri-Lobe Balloon Catheter is available in two sizes. The smaller balloon can be inflated to diameters of 16 mm to 32 mm and the larger balloon can be inflated to diameters of 26 mm to 48 mm.

AI/ML Overview

The provided FDA 510(k) clearance letter and associated summary pertain to a medical device, the GORE® Tri-Lobe Balloon Catheter, not an AI/ML-driven software device. Therefore, the information typically requested for AI/ML device acceptance criteria and study design (such as confidence intervals for performance metrics, sample sizes for training/test sets, expert adjudication methods, MRMC studies, or ground truth establishment) is not applicable or present in this document.

The document describes the device's technical characteristics, indications for use, and summaries of performance testing and clinical data to demonstrate substantial equivalence to a predicate device.

Here's a breakdown of the relevant information provided, structured to highlight what is and isn't applicable to your request:

Acceptance Criteria and Device Performance (Not Applicable as per AI/ML context)

While the document details performance testing, it doesn't present a table of quantitative acceptance criteria and corresponding reported device performance values in the way one would for an AI/ML device (e.g., Sensitivity, Specificity thresholds). Instead, safety and effectiveness are demonstrated through various tests and clinical observations.

Table 1. Technological Characteristics (This is a comparison of intended balloon diameters and inflation volumes, not performance metrics like accuracy or sensitivity for an AI/ML device.)

Model NumbersPredicate Device (K081799) ConfigurationSubject Device Configuration
BCM1634/TBCM1634Intended Balloon Diameter: 16mm – 32mmRecommended Inflation Volume: 4mL-13mLSame
BCL2645/TBCL2645Intended Balloon Diameter: 26mm – 42mmRecommended Inflation Volume: 12mL-25mLIntended Balloon Diameter: 26mm – 48mmRecommended Inflation Volume: 12mL-35mL

Key observations for performance (qualitative as per the document):

  • Clinical Outcomes (Day 0 Adverse Events): Among the 44 subjects with known GORE® Tri-Lobe Balloon Catheter use in the Aortic Arch substudy:
    • Zero (0) Adverse Events of Aortic Rupture, Aortic Dissection, Balloon Related Mortality, and Type III Endoleak.
    • One (1) reported Type I Endoleak at Day 0.
  • Conclusion: The device is deemed "substantially equivalent" to its predicate based on bench, pre-clinical, and clinical testing. The changes did not raise new concerns related to safety or effectiveness.

Study Details (Interpreted for a medical device rather than AI/ML)

  1. Sample Size Used for the Test Set and Data Provenance:

    • Test Set (Clinical Data): 77 total subjects were treated for lesions of the aortic arch in the GORE® TAG® Thoracic Branch Endoprosthesis Clinical Trial (G130120). Of these, 44 subjects (57.1%) had known use of the GORE® Tri-Lobe Balloon Catheter during the index procedure.
    • Data Provenance: The study was a "prospective, non-randomized, multicenter study." The country of origin is not explicitly stated but implied to be related to FDA clearance (likely US-based).
    • Retrospective/Prospective: Prospective.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • This is not applicable as this is a medical device study, not an AI/ML performance evaluation requiring expert labeling for ground truth. Clinical outcomes and observations from medical professionals involved in the trial serve as the "ground truth" for device safety and performance here.
  3. Adjudication Method for the Test Set:

    • Not applicable in the context of AI/ML ground truth adjudication. Clinical adverse events and outcomes were reported and presumably reviewed as part of the clinical trial protocol.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC study is relevant for evaluating the impact of AI on human readers' diagnostic performance. This document describes the performance of a physical medical device.
  5. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done:

    • Not applicable as this is not an algorithm. The device's performance is inherently linked to its use by medical professionals in a clinical setting ("human-in-the-loop" in a very different sense).
  6. The Type of Ground Truth Used:

    • Clinical Outcomes/Events: The "ground truth" for the device's safety and effectiveness was established through direct observation of clinical adverse events (e.g., Aortic Rupture, Aortic Dissection, Type I Endoleak) and the device's ability to facilitate dilatation of endoprostheses during the clinical trial. Bench and In Vivo Animal Testing also contributed to this "ground truth" indirectly by demonstrating physical properties and performance.
  7. The Sample Size for the Training Set:

    • Not applicable. This is not an AI/ML device that requires a training set.
  8. How the Ground Truth for the Training Set was Established:

    • Not applicable.

In summary, the provided document is a 510(k) clearance letter for a physical medical device. As such, the standard criteria and study designs relevant to AI/ML device performance evaluation (e.g., sensitivity, specificity, expert adjudication, MRMC studies, training/test sets) are not found within this content.

FDA 510(k) Clearance Letter - GORE® Tri-Lobe Balloon Catheter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 2, 2025

W. L. Gore & Associates, Inc.
Erika Grecco
Regulatory Affairs Associate
1505 N. Fourth Street
Flagstaff, Arizona 86004

Re: K250410
Trade/Device Name: GORE® Tri-Lobe Balloon Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: DQY
Dated: April 30, 2025
Received: May 1, 2025

Dear Erika Grecco:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250410 - Erika Grecco Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K250410 - Erika Grecco Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carmen G. Johnson -S

Carmen Gacchina Johnson, PhD
Assistant Director
DHT2B: Division of Circulatory Support, Structural and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K250410

Device Name
GORE® Tri-Lobe Balloon Catheter

Indications for Use (Describe)
The GORE® Tri-Lobe Balloon Catheter is intended to assist in the dilatation of self-expanding endoprostheses in large diameter vessels.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Traditional 510(k) GORE® Tri-Lobe Balloon Catheter
Page 1 of 5

510(K) SUMMARY

510(k) Submitter

W. L. Gore & Associates, Inc.
1505 N. Fourth Street
Flagstaff, AZ 86004

Regulatory Contact: Erika Grecco
Phone: (475) 202-2705
E-mail: egrecco@wlgore.com

Date Prepared

May 30, 2025

Device Names/Classification

Device Name: GORE® Tri-Lobe Balloon Catheter
Classification Name: Percutaneous Catheter
Regulation: 21 CFR 870.1250
Classification: Class II
Product Code: DQY

Predicate Devices

K081799, GORE® Tri-Lobe Balloon Catheter

Device Description

The GORE® Tri-Lobe Balloon Catheter is a compliant, tri-lobed polyurethane balloon catheter. The lobed design of the balloon catheter is designed for inflation without complete blockage of aortic blood flow. The three polyurethane balloons are mounted on the leading end of a multi-lumen catheter shaft. Radiopaque markers indicate the balloon edges. Each of the three inflation lumens is in communication with one of the balloons. The inflation port is in communication with all of the inflation lumens and is affixed with a luer lock. The guidewire lumen allows introduction of a 0.035" (0.89 mm) diameter guidewire for over-the-wire access. The trailing end of the guidewire lumen is affixed with a flushing/guidewire port with a luer lock, used for flushing the guidewire lumen. A Tuohy-Borst valve is integrated into the trailing end of the guidewire lumen.

The GORE® Tri-Lobe Balloon Catheter is available in two sizes. The smaller balloon can be inflated to diameters of 16 mm to 32 mm and the larger balloon can be inflated to diameters of 26 mm to 48 mm.

Page 6

Traditional 510(k) GORE® Tri-Lobe Balloon Catheter
Page 2 of 5

Indications for Use

The GORE® Tri-Lobe Balloon Catheter is intended to assist in the dilatation of self-expanding endoprostheses in large diameter vessels.

The Indication for Use is identical to that of the predicate.

Technological Characteristics

The technological characteristics (materials, design, etc.) compared to the predicate device remain largely unchanged. The only difference, from a technological perspective, between the subject devices and predicate is that the subject devices incorporate changes to the intended balloon diameters and associated recommended inflation volumes, as shown in Table 1 below.

Table 1. Technological Characteristics between predicate and subject devices

Model NumbersPredicate Device (K081799) ConfigurationSubject Device Configuration
BCM1634/TBCM1634Intended Balloon Diameter: 16mm – 32mmRecommended Inflation Volume: 4mL-13mLSame
BCL2645/TBCL2645Intended Balloon Diameter: 26mm – 42mmRecommended Inflation Volume: 12mL-25mLIntended Balloon Diameter: 26mm – 48mmRecommended Inflation Volume: 12mL-35mL

Summary of Performance Testing

The following testing was conducted to evaluate the changes that are the subject of this submission and demonstrate substantial equivalence to the predicate device:

  • Balloon Burst
  • Recommended Inflation Volume (RIV) Determination
  • Pushability, Trackability, and Kink Resistance
  • Torque
  • Balloon Inflate / Deflate Time
  • Balloon Fatigue
  • Simulated Use
  • Aortic Blood Flow
  • Use in the Ascending Aorta
  • In Vivo Animal Testing
  • Clinical use in the GORE® TAG® Thoracic Branch Endoprosthesis Clinical Trial

In Vivo Animal Testing was completed for the GORE® TAG® Thoracic Branch Endoprosthesis and included validation of the GORE® Tri-Lobe Balloon Catheter in the porcine aortic arch to support the use of the GORE® Tri-Lobe Balloon Catheter in the ascending aorta for the GORE® TAG® Thoracic Branch Endoprosthesis Clinical Trial. The clinical evidence from this study, in combination with the in vitro and in vivo testing, was provided to support the removal of an IFU warning precluding use in the ascending aorta. The data also supported the recommended increased inflation volume and balloon diameters.

Page 7

Traditional 510(k) GORE® Tri-Lobe Balloon Catheter
Page 3 of 5

Summary of Clinical Data:

The GORE® Tri-Lobe Balloon Catheter was used as an accessory device in the GORE® TAG® Thoracic Branch Endoprosthesis Clinical Trial to assist in the dilatation of the self-expanding GORE® TAG® Thoracic Branch Endoprosthesis. The information that was collected on the use of the GORE® Tri-Lobe Balloon Catheter at the time of the index procedure (Day 0) is summarized below.

The GORE® TAG® Thoracic Branch Endoprosthesis Clinical Trial (G130120) was a prospective, non-randomized, multicenter study evaluating the GORE® TAG® Thoracic Branch Endoprosthesis in the treatment of lesions of the Aortic Arch and Descending Thoracic Aorta (DTA). The data provided specifically in support of this 510(k) focused on the Aortic Arch substudy, though the GORE® Tri-Lobe Balloon Catheter was used also for the DTA substudy.

In G130120, 77 subjects were treated for lesions of the aortic arch; please refer to Table 2 for Baseline Demographic Information. Forty four 44 (57.1%) total subjects had known use of the GORE® Tri-Lobe Balloon Catheter during the index procedure; the balloon was used in the ascending aorta and/or aortic arch in these cases. Those 44 subjects with known GORE® Tri-Lobe Balloon Catheter use experienced zero (0) Day 0 Adverse Events of Aortic Rupture, Aortic Dissection, Balloon Related Mortality, and Type III Endoleak. There was one (1) reported Type I Endoleak at Day 0. The clinical data supports the ability to use the GORE® Tri-Lobe Balloon Catheter in ascending aorta and aortic arch.

Table 2. Baseline Demographic Information

AneurysmDissectionOther Isolated LesionAll
Number of Enrolled Subjects5024377
Sex
Male32 (64.0%)17 (70.8%)2 (66.7%)51 (66.2%)
Female18 (36.0%)7 (29.2%)1 (33.3%)26 (33.8%)
Ethnicity
Not Hispanic or Latino47 (94.0%)23 (95.8%)3 (100.0%)73 (94.8%)
Hispanic or Latino3 (6.0%)1 (4.2%)0 (0%)4 (5.2%)
Unknown0 (0%)0 (0%)0 (0%)0 (0%)
Race¹
White27 (54.0%)14 (58.3%)3 (100.0%)44 (57.1%)
Black or African American6 (12.0%)6 (25.0%)0 (0%)12 (15.6%)
Asian16 (32.0%)2 (8.3%)0 (0%)18 (23.4%)
American Indian or Alaska Native0 (0%)0 (0%)0 (0%)0 (0%)
Hawaiian or Pacific Islander1 (2.0%)1 (4.2%)0 (0%)2 (2.6%)
Other0 (0%)2 (8.3%)0 (0%)2 (2.6%)
Age (yrs)
n5024377
Mean (Std Dev)74.3 (8.56)63.0 (11.31)73.7 (9.02)70.8 (10.75)

Page 8

Traditional 510(k) GORE® Tri-Lobe Balloon Catheter
Page 4 of 5

AneurysmDissectionOther Isolated LesionAll
Median75.063.073.072.0
Range(50, 89)(41, 86)(65, 83)(41, 89)
BMI²
n4924376
Mean (Std Dev)27.3 (5.94)29.9 (4.44)23.4 (2.01)27.9 (5.57)
Median25.629.024.426.9
Range(16.8, 44.4)(21.7, 40.2)(21.1, 24.8)(16.8, 44.4)

¹More than one option can be selected.
²Site was queried and the weight for one Subject is unavailable.

Additional information on the clinical study can be found in the Summary of Safety and Effectiveness for the GORE® TAG® Thoracic Branch Endoprosthesis (P210032/S015).

Conclusion

The GORE® Tri-Lobe Balloon Catheter is substantially equivalent to the predicate device, as supported by the bench, pre-clinical and clinical testing provided. The removal of the warning against use in the ascending aorta, additional larger balloon diameters/recommended inflation volumes, and other minor changes (e.g., addition of new model numbers, burst testing specification) since the last clearance did not raise new concerns related to safety or effectiveness.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).