K Number
K250294
Date Cleared
2025-08-12

(193 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Straumann® RidgeFit Implants ∅ 2.4 mm are indicated for oral endosteal implantation in the fully edentulous mandible and/or maxilla for the stabilization of removable dentures. The implants can be placed with immediate function when primary stability is achieved for all implants or with conventional loading if primary stability is not achieved on all implants.

  • Mandibular restorations require at least 4 Straumann® RidgeFit Implants ∅ 2.4 mm.
  • Maxillary restorations require at least 6 Straumann® RidgeFit Implants ∅ 2.4 mm.
Device Description

The Straumann® RidgeFit Implants are tapered implants manufactured from Roxolid® (Titanium-Zirconium alloy, TiZ) with a finished SLA surface. The implant neck is machined with an Optiloc® attachment portion, which is coated in Titanium Nitride (TiN) coating, resulting in a one-piece implant system acting as a retention feature for dentures. The Straumann® RidgeFit Implants have an external diameter of 2.4 mm and are available in implant lengths 10 mm, 12 mm and 14 mm with a gingival height of 2.8 mm and implant lengths 10 mm and 12 mm for gingival heights 3.8 mm and 4.8 mm. The Straumann® RidgeFit Implants ∅ 2.4 mm are suitable for oral endosteal implantation in the upper and lower jaw of fully edentulous patients. Conventional loading is recommended if primary stability cannot be achieved immediately on all implants.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the Straumann® RidgeFit Implants contains information regarding the device's technical specifications and non-clinical testing. However, it does not include any information about acceptance criteria or a study that uses a test set with ground truth established by experts to prove the device meets these criteria.

The document focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and non-clinical engineering tests (fatigue, insertion torque, biocompatibility, sterilization, packaging stability, MRI compatibility). These are performance tests on the device itself, not evaluations based on human-in-the-loop or standalone AI performance using a study with a test set and expert-established ground truth.

Therefore, I cannot provide the requested information for the following points as they are not present in the given text:

  • A table of acceptance criteria and the reported device performance: The document implicitly states that passing the non-clinical tests demonstrates "substantial equivalence," but does not define explicit acceptance criteria or performance metrics directly from a comparative study with a test set and ground truth.
  • Sample sized used for the test set and the data provenance: No such test set or study is described within the provided text.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no such test set or expert ground truth establishment is described.
  • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  • If a multi reader multi case (MRMC) comparative effectiveness study was done: No MRMC study is mentioned. The clearance is for an endosseous dental implant, which is a physical device, not an AI software intended for interpretation or diagnostic assistance that would typically undergo such studies.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
  • The sample size for the training set: Not applicable.
  • How the ground truth for the training set was established: Not applicable.

What the document does provide regarding testing:

The device's performance is proven through a summary of non-clinical testing, which includes:

  1. Dynamic fatigue testing: Conducted according to FDA guidance (Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments) and ISO 14801 ("Dentistry — Implants — Dynamic loading test for endosseous dental implants"). This testing demonstrated the subject devices are equivalent to primary predicate and reference devices.
  2. Insertion torque and Torque-to-failure test: Conducted and demonstrated substantial equivalence to the primary predicate device.
  3. Biocompatibility evaluation: Referenced K211052 and K191895 for evaluation in accordance with ISO-10993-1:2018 and FDA guidance.
  4. Sterilization validation: Referenced K211052 and K191895 for validation in accordance with ISO 11137-1:2006 (VDmax25 method).
  5. Packaging stability and shelf-life study: Referenced K211052 and K191895 for studies in accordance with ISO 11607-1:2019, ASTM F1886, ASTM F1929, and ASTM F88.
  6. MRI simulations: Conducted to demonstrate that the devices are MR Conditional.

These tests are primarily physical and chemical performance assessments of the implant materials and design, aimed at establishing safety and effectiveness, and equivalence to existing cleared devices, not at evaluating an AI's diagnostic performance against human experts.

FDA 510(k) Clearance Letter - Straumann® RidgeFit Implants

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

Institut Straumann AG
℅ Jennifer Jackson
Senior Director of Regulatory Affairs and Quality
Straumann USA, LLC
60 Minuteman Road
Andover, Massachusetts 01810

Re: K250294
Trade/Device Name: Straumann® RidgeFit Implants
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE
Dated: January 31, 2025
Received: July 17, 2025

Dear Jennifer Jackson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

August 12, 2025

Page 2

K250294 - Jennifer Jackson Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K250294 - Jennifer Jackson Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

510(k) Number (if known): K250294
Device Name: Straumann® RidgeFit Implants

Indications for Use (Describe)

Straumann® RidgeFit Implants ∅2.4mm are indicated for oral endosteal implantation in the fully edentulous mandible and/or maxilla for the stabilization of removable dentures. The implants can be placed with immediate function when primary stability is achieved for all implants or with conventional loading if primary stability is not achieved on all implants.

  • Mandibular restorations require at least 4 Straumann® RidgeFit Implants ∅ 2.4 mm.
  • Maxillary restorations require at least 6 Straumann® RidgeFit Implants ∅ 2.4 mm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

Traditional 510(k) Submission

Straumann® RidgeFit Implants

K250294 510(k) Summary

Institut Straumann AG 1

510(k) Summary

Submitter's Contact Information

Submitter: Straumann USA, LLC
60 Minuteman Road
Andover, MA 01810
Registration No.: 1222315 Owner/Operator No.: 9005052

On the behalf of:
Institut Straumann AG
Peter Merian weg 12
4052 Basel, Switzerland

Contact Person: Jennifer M. Jackson, MS
Sr.Director of Regulatory Affairs & Quality
Phone Number: +1-978-747-2509
Fax Number: +1-978-747-0023

Prepared By: Ninda Syam, PhD
RA & Compliance Manager
Institut Straumann AG
Phone number: +41619651411

Date Prepared: August 12, 2025

Name of the Device

Trade Names: Straumann® RidgeFit Implants
Common Name: Endosseous dental implant
Classification Name: Endosseous dental implant
Regulation Number: 21 CFR 872.3640
Device Classification: II
Product Code(s): Primary product code - DZE

Page 6

Traditional 510(k) Submission

Straumann® RidgeFit Implants

K250294 510(k) Summary

Institut Straumann AG 2

Predicate Device(s)

Primary Predicates:

  • K211052- Straumann® RidgeFit Implants

Reference Devices:

  • K191895 - Straumann® Mini Implants
  • K031106 - IMTEC Sendax MDI and MDI Plus Endosseous Implant

Device Description

The Straumann® RidgeFit Implants are tapered implants manufactured from Roxolid® (Titanium-Zirconium alloy, TiZ) with a finished SLA surface. The implant neck is machined with an Optiloc® attachment portion, which is coated in Titanium Nitride (TiN) coating, resulting in a one-piece implant system acting as a retention feature for dentures. The Straumann® RidgeFit Implants have an external diameter of 2.4 mm and are available in implant lengths 10 mm, 12 mm and 14 mm with a gingival height of 2.8 mm and implant lengths 10 mm and 12 mm for gingival heights 3.8 mm and 4.8 mm. The Straumann® RidgeFit Implants ∅ 2.4 mm are suitable for oral endosteal implantation in the upper and lower jaw of fully edentulous patients. Conventional loading is recommended if primary stability cannot be achieved immediately on all implants.

Indications for Use

Straumann® RidgeFit Implants ∅ 2.4 mm are indicated for oral endosteal implantation in the fully edentulous mandible and/or maxilla for the stabilization of removable dentures. The implants can be placed with immediate function when primary stability is achieved for all implants or with conventional loading if primary stability is not achieved on all implants.

  • Mandibular restorations require at least 4 Straumann® RidgeFit Implants ∅ 2.4 mm.
  • Maxillary restorations require at least 6 Straumann® RidgeFit Implants ∅ 2.4 mm.

Technological Characteristics

The indications for use and technological characteristics of the subject devices are consistent with those of the primary predicate and reference devices as detailed on the following table.

Page 7

Traditional 510(k) Submission

Straumann® RidgeFit Implants

K250294 510(k) Summary

Institut Straumann AG 3

Table 1. Comparative Summary of the Indications for Use and Technological Characteristics.

FEATURESUBJECT DEVICEPRIMARY PREDICATE DEVICEREFERENCE DEVICE 1REFERENCE DEVICE 2EQUIVALENCE DISCUSSION
K NumberK250294K211052K191895K031106*
Intended UseStraumann® RidgeFit Implants are intended for oral implantation to provide a support structure for connected prosthetic devices.The Straumann® RidgeFit Implants are intended for the stabilization of removable dentures. The removable dentures are connected to the RidgeFit Implants through the incorporated Optiloc® attachment element.The Straumann® Mini Implants are intended for the stabilization of removable dentures.Information not availableSimilar. The subject device has essentially the same intended use and Optiloc® connection for the removable dentures as the predicate device. The last sentence of Intended Use in K211052 was removed in the subject device for wording adaptation reasons.
Indications for UseStraumann® RidgeFit Implants ∅ 2.4 mm are indicated for oral endosteal implantation in the fully edentulous mandible and/or maxilla for the stabilization of removable dentures. The implants can be placed with immediate function when primary stability is achieved for all implants or with conventional loading if primary stability is not achieved on all implants. • Mandibular restorations require at least 4 Straumann® RidgeFit Implants ∅ 2.4 mm. • Maxillary restorations require at least 6 Straumann® RidgeFit Implants ∅ 2.4 mm.Straumann® RidgeFit Implants ∅2.4 mm are for oral endosteal implantation in the upper and/or lower jaw of fully edentulous patients. The implants can be placed with immediate function when primary stability is achieved for all implants or with conventional loading if primary stability is not achieved for all implants. Straumann® RidgeFit Implants are intended for the stabilization of removable dentures. • For mandibular restorations, at least 4 Straumann® RidgeFit Implants ∅2.4 mm should be placed. • For maxillary restorations, at least 6 Straumann® RidgeFit Implants ∅2.4 mm should be placed.Straumann® Mini Implants ∅2.4 mm are suitable for oral endosteal implantation in the upper and lower jaw of fully or partially edentulous patients. The implants can be placed with immediate function when good primary stability is achieved. Straumann® Mini Implants are intended for the stabilization of removable dentures.The MDI and MDI PLUS are self-tapping titanium threaded screws indicated for long-term intra-bony applications. Additionally, the MDI may also be used for interradicular transitional applications. These devices will permit immediate splinting stability and long-term fixation of new or existing crown and bridge installations, for full partial edentulism, and employing minimally invasive surgical intervention.Similar. The subject device has a more limited indication, which is only for fully and partially edentulous patients compared to the reference device (K191895), which is for fully and partially edentulous patients. Other than this, the indication of the subject device is similar to the predicate device (K211052) and reference device 1 (K191895). Additionally, new bench tests and comparisons of engineering drawings provide evidence that the subject device is substantially equivalent to the primary predicate device (K211052) and the reference device 1 (K191895).

Page 8

Traditional 510(k) Submission

Straumann® RidgeFit Implants

K250294 510(k) Summary

Institut Straumann AG 4

FEATURESUBJECT DEVICEPRIMARY PREDICATE DEVICEREFERENCE DEVICE 1REFERENCE DEVICE 2EQUIVALENCE DISCUSSION
K NumberK250294K211052K191895K031106*
Implant Diameter2.4 mm2.4 mm2.4 mm2.4 mmIdentical to K031106, K191895 and K211052.
Implant Length10, 12, 14 mm10, 12, 14 mm10, 12, 14 mm10, 13, 15, 18 mmIdentical to K191895 and K211052. Similar to K031106.
Number of cutting flutes2330Substantially equivalent to K191895 and K211052. Even though there is one flute fewer in the subject device compared to the predicate device and reference device 1, their global geometry and dimensions stay the same. Data from performance testing (see discussion further below) show that the subject device has substantially equivalent performance to the predicate device and reference device 1.
Depth of cutting flutesAverage: 0.43mmAverage: 0.52mmAverage: 0.52mmInformation not availableSubstantially equivalent to K191895 and K211052. Even though the depth of the cutting flutes of the subject device is reduced by 18% compared to the predicate device and reference device 1 in general, their global geometry and dimensions stay the same. Data from performance testing (see discussion further below) show that the subject device has substantially equivalent performance to the predicate device and reference device 1.
Depth of thread at the apical implant tipConstant/uniform Average: 0.33mmVarying/non-uniform Average: 0.37mmVarying/non-uniform Average: 0.37mmInformation not availableSubstantially equivalent to K191895 and K211052. Even though the depth of the thread at the tip of the subject device is reduced by 11% compared to the predicate device and reference device 1 in general, their global geometry and dimensions stay the same. Data from performance testing (see discussion further below) show that the subject device has substantially equivalent performance to the predicate device and reference device 1.

Page 9

Traditional 510(k) Submission

Straumann® RidgeFit Implants

K250294 510(k) Summary

Institut Straumann AG 5

FEATURESUBJECT DEVICEPRIMARY PREDICATE DEVICEREFERENCE DEVICE 1REFERENCE DEVICE 2EQUIVALENCE DISCUSSION
K NumberK250294K211052K191895K031106*
MaterialRoxolid® (TiZr alloy)Roxolid® (TiZr alloy)Roxolid® (TiZr alloy)Ti-6Al-4VIdentical to K191895 and K211052.
Neck Height2.8, 3.8, 4.8 mm2.8, 3.8, 4.8 mm2.8 mm2.5 mmIdentical to K211052. The neck heights 3.8 and 4.8 mm were introduced via a Memo-to-File to K191895.
CoatingTiN coatedTiN coatedTiN coatedUncoatedIdentical to K191895 and K211052.
Surface TreatmentSLASLASLAInformation not availableIdentical to K191895 and K211052.
Abutment to Restoration ConnectionAnchor ballAnchor ballAnchor ballO-ballIdentical to K191895 and K211052.
Type of Recommended RestorationStabilization of removable dentures (Optiloc®)Stabilization of removable dentures (Optiloc®)Stabilization of removable dentures (Optiloc®)Stabilization of removable denturesIdentical to K191895 and K211052.
Sterilization MethodGamma irradiationGamma irradiationGamma irradiationInformation not availableIdentical to K191895 and K211052.

*K031106 IMTEC Sendax MDI AND MDI Plus Endosseous Implant was acquired by 3M™ ESPE™ after its 510(k) clearance and is known since then as 3M™ ESPE™ MDI Mini Dental Implants with several variants, e.g., Ø2.4 mm 13 mm.

Page 10

Traditional 510(k) Submission

Straumann® RidgeFit Implants

K250294 510(k) Summary

Institut Straumann AG 6

Materials

The Straumann® RidgeFit Implants are made from Roxolid® (Titanium and 13-17% Zirconium alloy), finished with SLA surface (sandblasted, large grit, acid-etched surface treatment), and integrated with Optiloc® attachment element (machined and TiN coated).

Summary of Nonclinical Testing

Nonclinical testing data submitted or referenced to demonstrate substantial equivalence in this 510(k) includes:

  • Dynamic fatigue testing was conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and ISO 14801 "Dentistry — Implants — Dynamic loading test for endosseous dental implants" and demonstrated the subject devices are equivalent to the primary predicate and reference devices.

  • Insertion torque and Torque-to-failure test were conducted and demonstrated the subject devices are substantially equivalent to the primary predicate device.

  • K211052 and K191895 (Institut Straumann AG) referenced for biocompatibility evaluation in accordance with ISO-10993-1:2018 Biological evaluation of medical devices. Part 1: Evaluation and testing within a risk management process and the biocompatibility evaluation flowchart according to the FDA guidance document Use of International Standard ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff, issue on September 8, 2023.

  • K211052 and K191895 (Institut Straumann AG) referenced for sterilization validation in accordance with ISO 11137-1:2006 Sterilization of health care products – Radiation. Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices," using the VDmax25 method.

  • K211052 and K191895 (Institut Straumann AG) referenced for the packaging stability study and shelf-life study in accordance with ISO 11607-1: 2019, ASTM F1886, ASTM F1929, and ASTM F88.

Page 11

Traditional 510(k) Submission

Straumann® RidgeFit Implants

K250294 510(k) Summary

Institut Straumann AG 7

  • Non-clinical testing and MRI simulations were conducted, and demonstrate that these devices are MR Conditional.

Conclusion

The data included in this submission demonstrate substantial equivalence to the predicate and reference devices listed above. Performance testing and comparison to previous clearances show that the subject devices are substantially equivalent.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.