K Number
K250273
Date Cleared
2025-06-13

(134 days)

Product Code
Regulation Number
866.3982
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BinaxNOW COVID-19 Ag Card is a lateral flow immunochromatographic assay for the rapid, qualitative detection of the SARS-CoV-2 nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory tract infection (i.e., symptomatic). The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the BinaxNOW COVID-19 Ag Card and followed up with a molecular test.

A negative test is presumptive and does not preclude SARS-CoV-2 infection; it is recommended these results be confirmed by a molecular SARS-CoV-2 assay.

Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Device Description

The BinaxNOW COVID-19 Ag Card is an immunochromatographic membrane assay that uses antibodies to detect SARS-CoV-2 nucleocapsid protein from anterior nasal swab specimens. SARS-CoV-2 specific antibodies and a control antibody are immobilized onto a membrane support as two distinct lines and combined with other reagents/pads to construct a test strip. This test strip and a well to hold the swab specimen are mounted on opposite sides of a cardboard, book-shaped hinged test card.

To perform the test, an anterior nasal swab specimen is collected from the patient, 6 drops of extraction reagent from a dropper bottle are added to the top hole of the swab well. The patient sample is inserted into the test card through the bottom hole of the swab well, and firmly pushed upwards until the swab tip is visible through the top hole. The swab is rotated 3 times clockwise and the card is closed, bringing the extracted sample into contact with the test strip. Test results are interpreted visually at 15 minutes based on the presence or absence of visually detectable pink/purple colored lines. Results should not be read after 30 minutes.

AI/ML Overview

The provided document is a 510(k) summary for the BinaxNOW COVID-19 Ag Card. It does not describe a study proving a device meets acceptance criteria in the manner typically associated with AI/ML-driven medical devices, which would involve measures like sensitivity, specificity, or AUC against a ground truth, often with human readers involved (MRMC studies).

Instead, this document describes the validation of an immunochromatographic assay (a rapid antigen test) for COVID-19. The "acceptance criteria" here are typically performance targets for analytical and clinical characteristics (e.g., Limit of Detection, cross-reactivity, Positive Percent Agreement, Negative Percent Agreement). The "study" refers to the analytical and clinical studies conducted to demonstrate these performance characteristics.

Therefore, the following response will interpret "acceptance criteria" as the performance benchmarks for a diagnostic assay and describe the validation studies for the BinaxNOW COVID-19 Ag Card based on the provided text.

Here's a breakdown of the information requested, interpreted in the context of a rapid antigen test (not an AI/ML device):


Acceptance Criteria and Device Performance for BinaxNOW COVID-19 Ag Card

The BinaxNOW COVID-19 Ag Card is a lateral flow immunochromatographic assay, not an AI/ML diagnostic device. Therefore, the "acceptance criteria" are based on the analytical and clinical performance characteristics typical for such an in-vitro diagnostic (IVD) device, rather than metrics like AUC, sensitivity/specificity of an AI algorithm, or human reader improvement with AI assistance.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Metric/StudyPerformance Target (Implicit/Typical for IVDs)Reported Device Performance
Analytical PerformanceLimit of Detection (LOD)Lowest virus concentration detected ≥ 95% of the time (e.g., 19/20 replicates positive)USA-WA1/2020: 3.5 x 10³ TCID50/mL (70 TCID/swab) B.1.1.529 (Omicron): 1.6 x 10³ TCID50/mL (32.06 TCID/swab) WHO International Standard (NIBSC 21/368): 375 IU/mL (7.5 IU/swab), with 100% detection (20/20) at this concentration.
Analytical Reactivity (Inclusivity)Detection of various SARS-CoV-2 strains at specified concentrations (all 5 replicates positive for a given concentration)Detected 19 different SARS-CoV-2 variants (Alpha, Beta, Delta, Gamma, Iota, Italy-INMI1, Kappa, Zeta, Omicron variants including BA.2.3, BA.2.12.1, BA.2.75.5, BA.4.6, BA.5, BA.5.5, BF.5, BF.7, BQ.1, BQ.1.1, XBB, JN.1) at concentrations ranging from 8.75 x 10² TCID50/mL to 5.60 x 10⁴ TCID50/mL (or IU/mL for JN.1).
Analytical Specificity (Cross Reactivity) & Microbial InterferenceNo cross-reactivity or interference with common respiratory pathogens/commensals.No cross-reactivity or interference observed with 28 tested microorganisms (9 bacteria, 17 viruses, 1 yeast, pooled human nasal wash, and 4 Coronavirus HKU1 clinical specimens). In silico analysis for P. jirovecii showed very low potential for cross-reactivity. Possible susceptibility to SARS-CoV (due to homology) noted, but deemed low clinical likelihood.
High Dose Hook EffectNo hook effect at high viral concentrations.No high dose hook effect observed up to 1.4 x 10⁶ TCID50/mL.
Interfering SubstancesNo interference from specified endogenous or exogenous substances (e.g., common nasal medications, blood, mucin).No effect on test performance found at specified concentrations for 25 substances (e.g., throat lozenges, various nasal sprays, hand sanitizer, blood, mucin).
Reproducibility/Near the Cut OffHigh agreement across sites for negative, low, moderate positive, and high negative samples.Moderate Positive: 100% (135/135) overall agreement (95% CI: 97.2%-100.0%). Low Positive: 94.1% (127/135) overall agreement (95% CI: 88.7%–97.0%). High Negative: 99.2% (132/133) overall agreement (95% CI: 95.9%-99.9%). True Negative: 99.3% (134/135) overall agreement (95% CI: 95.9%-99.9%).
Clinical PerformancePositive Percent Agreement (PPA)High PPA against a molecular comparator (RT-PCR) in symptomatic individuals.Overall (Combined Studies): 86.9% (186/214) with 95% CI: 81.7%, 90.8% (within 5 days symptom onset). Original Study: 81.6% (71/87) with 95% CI: 72.2%, 88.4%. Omicron Study: 90.6% (115/127) with 95% CI: 84.2%, 94.5%.
Negative Percent Agreement (NPA)High NPA against a molecular comparator (RT-PCR) in symptomatic individuals.Overall (Combined Studies): 98.5% (384/390) with 95% CI: 96.7%, 99.3% (within 5 days symptom onset). Original Study: 98.6% (205/208) with 95% CI: 95.8%, 99.5%. Omicron Study: 98.4% (179/182) with 95% CI: 95.3%, 99.4%.
Performance by Days Post Symptom Onset (DPSO)Performance maintained within the specified window.PPA ranged across DPSO: - Day 0: 69.23% (Omicron Study) - Day 1: 94.12% (Original), 88.24% (Omicron) - Day 2: 73.33% (Original), 97.22% (Omicron) - Day 3: 76.00% (Original), 100.00% (Omicron) - Day 4: 88.89% (Original), 66.67% (Omicron) - Day 5: 100.00% (Original), 100.00% (Omicron)
Invalid RateLow invalid rate.0.68% overall (5/730).
User/Environmental FactorsFlex Studies (Robustness)Device performs accurately under various usage and environmental conditions.Demonstrated robustness to usage variation and environmental factors. Identified that direct exposure of test strip to wet cleaning solutions or excessive glove powder may cause erroneous results, leading to specific instructions for use.

2. Sample Sizes and Data Provenance (Clinical Studies)

  • Clinical Test Set Sample Size:
    • Study 1 (Original): 295 evaluable subjects.
    • Study 2 (Omicron): 309 evaluable subjects.
    • Combined Clinical Data: 604 evaluable nasal swabs from symptomatic patients (within 5 days of symptom onset).
  • Data Provenance: Clinical studies were conducted within the United States.
    • Study 1: November 2020 through March 2021 (when Delta and Omicron were dominant).
    • Study 2: February 2022 to July 2022 (when Omicron and its variants were prevalent).
  • Retrospective/Prospective: Both clinical studies were prospective.

3. Number of Experts and Qualifications for Ground Truth for Test Set

This type of diagnostic device (lateral flow immunoassay) does not typically utilize human experts in the same way an AI/ML device would for image interpretation or clinical diagnosis. For the BinaxNOW COVID-19 Ag Card, the ground truth for the clinical studies was established by a comparator molecular test (RT-PCR). The experts involved would be the laboratory personnel performing and interpreting the RT-PCR assays. Their specific qualifications are not detailed in this summary but are implicitly assumed to be standard for clinical laboratory professionals performing EUA-authorized RT-PCR tests.

4. Adjudication Method for the Test Set

Not applicable in the typical sense for an AI/ML study involving human interpretation. The comparator method (RT-PCR) serves as the reference standard. The document mentions for the serial testing study's composite comparator method that in cases of discordant RT-PCR results, a third RT-PCR test was performed, and the final result based on majority rule.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This is a rapid antigen test, not an AI/ML system where human readers would interpret results "with vs. without AI assistance." The test is visually read by the user, and its performance is assessed against a molecular gold standard.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

This question is not applicable in the context of this device. The BinaxNOW COVID-19 Ag Card is a manually read, qualitative visual assay. There is no AI algorithm to evaluate for standalone performance. The "performance" tables provided in the document (PPA and NPA) essentially represent the "standalone" performance of the rapid antigen test itself when interpreted visually.

7. Type of Ground Truth Used

  • For Clinical Studies: The primary ground truth for clinical performance (PPA, NPA) was an FDA Emergency Use Authorized real-time Polymerase Chain Reaction (RT-PCR) assay for the detection of SARS-CoV-2.
  • For Serial Testing Study: A composite comparator method was used, involving at least two highly sensitive EUA RT-PCRs. If discordant, a third RT-PCR was performed, and the final result was based on majority rule.
  • For Analytical Studies: Ground truth was established by known concentrations of heat-inactivated SARS-CoV-2 virus or WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) for LoD and inclusivity studies, and known presence/absence of specific microorganisms for cross-reactivity.

8. Sample Size for the Training Set

This information is not applicable for this type of IVD device. The BinaxNOW COVID-19 Ag Card is a laboratory-developed lateral flow assay, not an AI/ML model that is 'trained' on data. Its 'training' is the fundamental assay development and optimization process, not a computational training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable for this device type. The manufacturing process and quality control of the reagents and test strip govern its 'performance' characteristics, which are then analytically and clinically validated.

U.S. Food & Drug Administration

10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 13, 2025

Abbott Diagnostics Scarborough, Inc.
Kristen Cyr
Regulatory Affairs Specialist
10 Southgate Road
Scarborough, Maine 04074

Re: K250273
Trade/Device Name: BinaxNOW COVID-19 Ag Card
Regulation Number: 21 CFR 866.3982
Regulation Name: Simple Point-Of-Care Device To Directly Detect SARS-Cov-2 Viral Targets From Clinical Specimens In Near-Patient Settings
Regulatory Class: Class II
Product Code: QVF
Dated: January 30, 2025
Received: January 30, 2025

Dear Kristen Cyr:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250273 - Kristen Cyr Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250273 - Kristen Cyr Page 3

Sincerely,

JOSEPH BRIGGS -S

Joseph Briggs, Ph.D.
Deputy Division Director
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250273

Device Name: BinaxNOW COVID-19 Ag Card

Indications for Use (Describe)

The BinaxNOW COVID-19 Ag Card is a lateral flow immunochromatographic assay for the rapid, qualitative detection of the SARS-CoV-2 nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory tract infection (i.e., symptomatic). The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the BinaxNOW COVID-19 Ag Card and followed up with a molecular test.

A negative test is presumptive and does not preclude SARS-CoV-2 infection; it is recommended these results be confirmed by a molecular SARS-CoV-2 assay.

Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Performance characteristics for SARS-CoV-2 were established from November 2020 to July 2022, when SARS-COV-2 Delta and Omicron were dominant. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(K) SUMMARY

Preparation Date: June 11, 2025

CONTACT DETAILS

Applicant Name: Abbott Diagnostics Scarborough, Inc.
Applicant Address: 10 Southgate Road, Scarborough, Maine 04074 USA
Applicant Contact: Ms. Kristen Cyr
Applicant Contact Email: kristen.cyr@abbott.com
Applicant Contact Phone: (207) 210-4311

DEVICE NAME

Device Trade Name: BinaxNOW COVID-19 Ag Card
Common Name: BinaxNOW COVID-19
Classification Name: Simple Point-Of-Care Device To Directly Detect Sars-Cov-2 Viral Targets From Clinical Specimens In Near-Patient Settings
Regulation Number: 866.3982
Product Code: QVF

PREDICATE DEVICE

K231187, Nano-Check COVID-19 Antigen Test

DEVICE DESCRIPTION SUMMARY

The BinaxNOW COVID-19 Ag Card is an immunochromatographic membrane assay that uses antibodies to detect SARS-CoV-2 nucleocapsid protein from anterior nasal swab specimens. SARS-CoV-2 specific antibodies and a control antibody are immobilized onto a membrane support as two distinct lines and combined with other reagents/pads to construct a test strip. This test strip and a well to hold the swab specimen are mounted on opposite sides of a cardboard, book-shaped hinged test card.

To perform the test, an anterior nasal swab specimen is collected from the patient, 6 drops of extraction reagent from a dropper bottle are added to the top hole of the swab well. The patient sample is inserted into the test card through the bottom hole of the swab well, and firmly pushed upwards until the swab tip is visible through the top hole. The swab is rotated 3 times clockwise and the card is closed, bringing the extracted sample into contact with the test strip. Test results are interpreted visually at 15 minutes based on the presence or absence of visually detectable pink/purple colored lines. Results should not be read after 30 minutes.

INTENDED USE/INDICATIONS FOR USE

The BinaxNOW™ COVID-19 Ag Card is a lateral flow immunochromatographic assay for the rapid, qualitative detection of the SARS-CoV-2 nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory tract infection (i.e., symptomatic). The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the BinaxNOW COVID-19 Ag Card and followed up with a molecular test.

A negative test is presumptive and does not preclude SARS-CoV-2 infection; it is recommended these results be confirmed by a molecular SARS-CoV-2 assay.

Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Page 6

510(K) SUMMARY
Page 2 of 10

Performance characteristics for SARS-CoV-2 were established from November 2020 to July 2022, when SARS-COV-2 Delta and Omicron were dominant. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

INDICATIONS FOR USE COMPARISON

The BinaxNOW COVID-19 Ag Card and the predicate device, Nano-Check COVID-19 Antigen Test, have the same intended use. The test systems are intended for the rapid, qualitative detection of SARS-CoV-2 directly in anterior nasal swab specimens as an aid in the diagnosis of SARS-CoV-2 infection. The devices are intended for use in near-patient settings.

TECHNOLOGICAL COMPARISON

The BinaxNOW COVID-19 Ag Card and the predicate device, Nano-Check COVID-19 Antigen Test, have the same technological characteristics. Both test systems are visually read lateral flow immunoassays to detect nucleocapsid protein antigen from SARS-CoV-2 in anterior nasal swab samples.

NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY AND CONCLUSIONS

ANALYTICAL STUDIES

Analytical Sensitivity (Limit of Detection)

The BinaxNOW COVID-19 Ag Card limit of detection (LOD) was determined by evaluating different concentrations of two heat inactivated strains of SARS-CoV-2 virus. Presumed negative natural nasal swab specimens were eluted in PBS. Swab eluates were combined and mixed thoroughly to create a clinical matrix pool to be used as the diluent. Inactivated SARS-CoV-2 virus was diluted in this natural nasal swab matrix pool to generate virus dilutions for testing. Contrived nasal swab samples were prepared by absorbing 20 microliters of each virus dilution onto the swab. The contrived swab samples were tested according to the test procedure.

For each strain, the LOD was determined as the lowest virus concentration that was detected ≥ 95% of the time (i.e., concentration at which at least 19 out of 20 replicates tested positive).

The BinaxNOW COVID-19 Ag Card LOD in natural nasal swab matrix was confirmed 3.5 x 10³ TCID50/mL for USA-WA1/2020 and 1.6 x 10³ TCID50/mL for B.1.1.529 (Omicron).

Limit of Detection Study Results

StrainConcentration TCID50/mLConcentration TCID/swab
USA-WA1/20203.5 x 10³70
B.1.1.529 (Omicron)1.6 x 10³32.06

WHO Standard Testing

A study was performed to also determine the LoD for the BinaxNOW COVID-19 Ag Card Test in nasal samples using the WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) as a standardized material. As per the WHO instructions, the international standard material was reconstituted in 0.25 mL of ultra-pure water. Following reconstitution, the ampule was left at ambient temperature for 20 minutes and then mixed thoroughly, avoiding generation of excess foam. The reconstitution of the material yielded a final stock concentration equal to 2.0 x 10⁴ IU/mL.

Page 7

510(K) SUMMARY

For each replicate, 20 μL of virus dilution was applied to a swab and the swab was tested according to the IFU. A preliminary LoD concentration was determined by testing a series of 2-fold dilutions of the antigen spiked into negative nasal matrix (NNM) in replicates of three (3). The lowest concentration with 3 out of 3 positive replicates was considered to be the preliminary LoD. The results of the preliminary LoD study are shown in the Table below:

WHO SARS-CoV-2 Standard LoD Range Finding Results

Concentration (IU/ml)BinaxNOW COVID-19 Ag Card Results
# Detected% Detected
20003/3100
10003/3100
5003/3100
2500/30
1250/30

The preliminary LoD was confirmed by testing an additional seventeen (17) replicates per dilution for a total of (20) until less than 100% positive results were obtained. The results of this testing, as shown in the table below confirmed the LoD for the WHO International Standard Antigen to be 375 IU/mL (7.5 IU/swab).

WHO SARS-CoV-2 Standard Confirmatory LoD Results

Concentration (IU/mL)BinaxNOW COVID-19 Ag Card Results
# Detected% Detected
50020/20100
37520/20100
2504/2020

Analytical Reactivity (Inclusivity)

An Analytical Reactivity (Inclusivity) study was performed to determine whether the BinaxNOW COVID-19 Ag Card is able to detect a variety of SARS-CoV-2 strains.

Vendor provided stocks of SARS-CoV-2 strains were diluted in natural nasal swab matrix to generate virus dilutions for testing. Contrived swab samples were prepared by coating 20 microliters of virus dilution onto each swab.

Each dilution of virus strain was tested n = 5 replicates. A concentration level was considered "reactive/positive" in this study if all five replicates generated a positive result.

The BinaxNOW COVID-19 Ag Card detected all strains tested at the concentrations indicated in the table below:

Page 8

510(K) SUMMARY
Page 4 of 10

Analytical Reactivity Study Results

VariantConcentration (TCID50/ml in the test)
Alpha (B.1.1.7)3.50 x 10⁴
Beta (B.1.351)7.00 x 10³
Delta (B.1.617.2)1.75 x 10³
Gamma (P.1)1.75 x 10³
Iota (B.1.526)7.00 x 10³
Italy-INMI13.50 x 10⁴
Kappa (B.1.617.1)1.05 x 10⁴
Zeta (P.2)3.50 x 10³
Omicron (BA.2.3)2.63 x 10³
Omicron (BA.2.12.1)1.31 x 10³
Omicron (BA.2.75.5)8.75 x 10²
Omicron (BA.4.6)3.50 x 10³
Omicron (BA.5)5.60 x 10⁴
Omicron (BA.5.5)1.10 x 10²
Omicron (BF.5)3.50 x 10³
Omicron (BF.7)1.40 x 10⁴
Omicron (BQ.1)7.00 x 10³
Omicron (BQ.1.1)8.75 x 10²
Omicron (XBB)2.80 x 10⁴
Omicron (JN.1)3.63 x 10² IFU/mL

Analytical Specificity (Cross Reactivity) and Microbial Interference

Cross reactivity and potential interference of BinaxNOW COVID-19 Ag Card was evaluated by testing 28 commensal and pathogenic microorganisms (9 bacteria, 17 viruses, 1 yeast and pooled human nasal wash) that may be present in the nasal cavity. Each organism, virus, and yeast was tested n=5 in the absence or presence of heat inactivated SARS-CoV-2 virus at a concentration of 3X LOD (210 TCID50/swab). No cross-reactivity or interference was seen with the microorganisms in the table below when tested at 1 x 10⁶ CFU/mL for bacteria and yeast and at 1 x 10⁵ TCID50/mL for viruses.

TypePanel Member
VirusesHuman Adenovirus 1
Human Coronavirus 229E
Human Coronavirus NL63
Human Coronavirus OC43
Enterovirus 70
Human Metapneumovirus (hMPV)
Human Parainfluenza virus 1
Human Parainfluenza virus 2
Human Parainfluenza virus 3
Human Parainfluenza virus 4
RSV A
Rhinovirus 1A

Page 9

510(K) SUMMARY
Page 5 of 10

TypePanel Member
MERS-coronavirus
Human Influenza A/California/07/09
Human Influenza A/New Caledonia/20/99
Human Influenza A/Brisbane/02/18
Human Influenza B/Wisconsin/1/10
BacteriaBordetella pertussis
Chlamydia pneumoniae
Haemophilus influenzae
Legionella pneumophila
Mycoplasma pneumoniae
Staphylococcus aureus
Staphylococcus epidermidis
Streptococcus pneumoniae
Streptococcus pyogenes
YeastCandida albicans
OtherPooled Human Nasal Wash

Additionally, four (4) clinical specimens containing Coronavirus HKU1 were tested n=5 in the absence or presence of heat inactivated SARS-CoV-2 virus at a concentration of 3X LOD (210 TCID50/swab). No cross-reactivity or interference was seen with the Coronavirus HKU1 clinical specimens.

To estimate the likelihood of cross-reactivity with SARS-CoV-2 virus in the presence of organisms that were not available for wet testing, In silico analysis using the Basic Local Alignment Search Tool (BLAST) managed by the National Center for Biotechnology Information (NCBI) was used to assess the degree of protein sequence homology.

  • For P. jirovecii, the potential for cross-reactivity is very low. No significant sequence homology was found.
  • BinaxNOW COVID-19 Ag Card may be susceptible to cross reactivity with SARS-CoV. However, there have been no reported cases of SARS since 2004, reducing the likelihood of a false positive result due to cross reactivity.

High Dose Hook Effect

No high dose hook effect was observed when tested with up to a concentration of 1.4 x 10⁶ TCID50/mL of inactivated SARS-CoV-2 virus with the BinaxNOW COVID-19 Ag Card.

Interfering Substances

The following substances, naturally present in respiratory specimens or that may be artificially introduced into the upper respiratory tract, were evaluated with the BinaxNOW COVID-19 Ag Card at the concentrations listed below and were found not to affect test performance.

SubstanceActive IngredientConcentration
Throat LozengeMenthol, Benzocaine3 mg/mL
Sore Throat SprayPhenol5% w/v
OTC Nasal Spray 1Mometasone Furoate15% v/v

Page 10

510(K) SUMMARY
Page 6 of 10

SubstanceActive IngredientConcentration
OTC Nasal Spray 2Triamcinolone15% v/v
OTC Nasal Spray 3Budesonide15% v/v
OTC Nasal Spray 4Fluticasone15% v/v
OTC nasal gelSodium Chloride with Preservatives15% v/v
OTC Nasal Spray 5Phenylephrine15% v/v
OTC Nasal Spray 6Oxymetazoline15% v/v
OTC Nasal Spray 7Cromolyn15% v/v
OTC Homeopathic Nasal SprayZicam (Galphimia glauca, Histaminum hydrochloricum, Luffa operculate, sulfur)15% v/v
OTC Homeopathic Nasal WashAlkalol15% v/v
Hand SanitizerEthyl Alcohol 62%1% w/v
Hand Soap1% w/v
EndogenousWhole Blood2.5% v/v
EndogenousMucin2.5 mg/mL
EndogenousLeukocytes5 x 10⁶ cells/mL
Antibiotic, Nasal OintmentMupirocin10 mg/mL
Nasal Corticosteroid 1Beclomethasone15% v/v
Nasal Corticosteroid 2Dexamethasone15% v/v
Nasal Corticosteroid 3Flunisolide15% v/v
Anti-Viral Drug 1Tamiflu (Oseltamivir Phosphate)5 mg/mL
Anti-Viral Drug 2Remdesivir5 mg/mL
Anti-Viral Drug 3Molnupiravir5 mg/mL
AntibioticTobramycin1.44 mg/mL
Anti-Viral DrugZanamivir281.5 ng/mL

Reproducibility/Near the Cut Off

A reproducibility/near the cut off study of BinaxNOW COVID-19 Ag Card was conducted by operators from three sites using panels of blind coded specimens containing negative, low positive (near the limit of detection), moderate positive (above the limit of detection), and high negative SARS-CoV-2 samples.

Participants tested each sample multiple times on five different days. The percent agreement relative to the expected results for the moderate positive samples was 100% (135/135). The percent agreement relative to the expected results for the low positive samples was 94.1% (127/135). The percent agreement relative to the expected results for the high negative samples was 99.2% (132/133) and the true negative samples were 99.3% (134/135).

The Reproducibility Study site-to-site qualitative results (agreements relative to the expected results) are presented in the table below:

Page 11

510(K) SUMMARY
Page 7 of 10

SampleSite 1Site 2Site 3Overall Agreement and 95% CI
Moderate Positive100% (45/45)100% (45/45)100% (45/45)100% (135/135) 97.2% - 100.0%
Low Positive93.3% (42/45)97.8% (44/45)91.1% (41/45)94.1% (127/135) 88.7% – 97.0%
High Negative100% (45/45)100% (45/45)97.7% (42/43)99.2% (132/133) 95.9% - 99.9%
True Negative100% (45/45)100% (45/45)97.8% (44/45)99.3% (134/135) 95.9% - 99.9%

CLINICAL STUDIES

Clinical performance characteristics of the BinaxNOW COVID-19 Ag Card were evaluated in two prospective studies conducted in symptomatic individuals within the United States. In the first study, conducted across five (5) investigational sites from November, 2020 through March, 2021, each subject either self-collected one (1) nasal swab (from both nostrils) or had one sample collected from him/her by another individual and performed the BinaxNOW COVID-19 Ag Card. A total of 397 individuals were enrolled in the study of which 102 were excluded for not meeting inclusions criteria or were unevaluable resulting in a total of 295 evaluable subjects.

In the second study, an additional all comers, real world, prospective clinical study was performed at a high-volume COVID community testing site from February 2022 to July 2022 when Omicron and its variants were prevalent in the United States. This study was led by Johns Hopkins Medicine in collaboration with the University of Maryland Medical Center and Maryland Department of Health. In this study, participants independently performed the BinaxNOW COVID-19 Ag Card, from nasal swab collection to result interpretation. A total of 333 individuals were enrolled in the study of which 24 were excluded for not meeting inclusion criteria or were unevaluable resulting in a total of 309 evaluable subjects.

In each study, a matched anterior nasal swab sample was taken from each study subject by a healthcare professional for testing on an FDA Emergency Use Authorized real-time Polymerase Chain Reaction (RT-PCR) assay for the detection of SARS-CoV-2 as the comparator method in each study.

Across both studies, the performance of BinaxNOW COVID-19 Ag Card was established with a total of 604 nasal swabs from symptomatic patients (within 5 days of symptom onset) who were suspected of COVID-19. Performance is shown in the tables below combined for both studies and individual by study.

BinaxNOW™ COVID-19 Ag Card Performance within 5 days of symptom onset against the Comparator Method – Overall/Combined

BinaxNOW COVID-19 Ag CardComparator MethodTotal
PositiveNegative
Positive1866192
Negative28384412
Total214390604

Positive Agreement: 186/214 86.9% (95% CI: 81.7, 90.8)

Page 12

510(K) SUMMARY
Page 8 of 10

Negative Agreement: 384/390 98.5% (95% CI: 96.7, 99.3)

Note: Five (5) samples generated an invalid BinaxNOW COVID-19 Ag Card result and are not included in the analysis. The invalid rate is 5/730, or 0.68% (95% CI from 0.29% to 1.59%). The denominator for the invalid rate is based on total study enrollment.

BinaxNOW COVID-19 Ag Card Performance within 5 days of symptom onset against the Comparator Method – Original Study (November 2020 – March 2021)

BinaxNOW COVID-19 Ag CardComparator MethodTotal
PositiveNegative
Positive71374
Negative16205221
Total87208295

Positive Agreement: 71/87 81.6% (95% CI: 72.2, 88.4)
Negative Agreement: 205/208 98.6% (95% CI: 95.8, 99.5)

Note: Three (3) samples generated an invalid BinaxNOW COVID-19 Ag Card result and are not included in the analysis. The invalid rate is 3/397, or 0.76% (95% CI from 0.26% to 2.20%). The denominator for the invalid rate is based on total study enrollment.

BinaxNOW COVID-19 Ag Card Performance within 5 days of symptom onset against the Comparator Method – Omicron Study (February 2022 – July 2022)

BinaxNOW COVID-19 Ag CardComparator MethodTotal
PositiveNegative
Positive1153118
Negative12179191
Total127182309

Positive Agreement: 115/127 90.6% (95% CI: 84.2, 94.5)
Negative Agreement: 179/182 98.4% (95% CI: 95.3, 99.4)

Note: Two (2) samples generated an invalid BinaxNOW COVID-19 Ag Card result and are not included in the analysis. The invalid rate is 2/333, or 0.60% (95% CI from 0.16% to 2.10%). The denominator for the invalid rate is based on total study enrollment.

BinaxNOW COVID-19 Ag Card Positive Agreement (PPA) against the Comparator Method – Stratified by Days Post Symptom Onset (DPSO)

DPSOPPA - Original Study (November 2020-March 2021)PPA - Omicron Study (February 2022-July 2022)
Day 0N/A (0/0)69.23% (9/13)
Day 194.12% (16/17)88.24% (45/51)
Day 273.33% (22/30)97.22% (35/36)
Day 376.00% (19/25)100.00% (20/20)
Day 488.89% (8/9)66.67% (2/3)
Day 5100.00% (6/6)100.00% (4/4)

Serial Testing

A prospective clinical study was also conducted between January 2021 and May 2022 as a component of the Rapid Acceleration of Diagnostics (RADx) initiative from the National Institutes of Health (NIH)¹,². A total of 7,361 individuals were enrolled via a decentralized clinical study design, with a broad geographical representation of the United States. Per inclusion criteria, all individuals were asymptomatic upon enrollment

¹ https://www.medrxiv.org/content/10.1101/2022.08.04.22278274v1
² https://www.medrxiv.org/content/10.1101/2022.08.05.22278466v1

Page 13

510(K) SUMMARY
Page 9 of 10

in the study and at least 14 days prior to it and did not have a SARS-CoV-2 infection in the three months prior to enrollment. Participants were assigned to one of three EUA authorized SARS-CoV-2 OTC rapid antigen tests to conduct serial testing (every 48 hours) for 15 days. If an antigen test was positive, the serial-antigen testing result is considered positive.

At each rapid antigen testing time point, study subjects also collected a nasal swab for comparator testing using a home collection kit (using a 15-minute normalization window between swabs). SARS-CoV-2 infection status was determined by a composite comparator method on the day of the first antigen test, using at least two highly sensitive EUA RT-PCRs. If results of the first two molecular tests were discordant a third highly sensitive EUA RT-PCR test was performed, and the final test result was based upon the majority rule.

Study participants reported symptom status throughout the study using the MyDataHelps app. Two-day serial antigen testing is defined as performing two antigen tests 36 – 48 hours apart. Three-day serial antigen testing is defined as performing three antigen tests over five days with at least 48 hours between each test.

Out of the 7,361 participants enrolled in the study, 5,609 were eligible for analysis. Among eligible participants, 154 tested positive for SARS-CoV-2 infection based on RT-PCR, of which 97 (62%) were asymptomatic on the first day of their infection, whereas 57 (39%) reported symptoms on the first day of infection.

Performance of the antigen test with serial testing in individuals is described in the table below.

Data establishing PPA of COVID-19 antigen serial testing compared to the molecular comparator single day testing throughout the course of infection with serial testing. Data is from all antigen tests in study combined.

Days After First PCR Positive Test ResultSymptomatic On First Day Of TestingAg Positive / PCR Positive (Antigen Test Performance % PPA)
1 Test2 Tests3 Tests
034/5747/5144/47
(59.6%)(92.2%)(93.6%)
258/6259/6043/43
(93.5%)(98.3%)(100%)
455/5853/5439/40
(94.8%)(98.1%)(97.5%)
627/3426/3322/27
(79.4%)(78.8%)(81.5%)
812/1712/177/11
(70.6%)(70.6%)(63.6%)
104/93/7
(44.4%)(42.9%)

1 Test = one (1) test performed on the noted days after the first PCR positive test result. Day 0 is the first day of documented infection with SARS-CoV-2.

2 Tests = two (2) tests performed an average of 48 hours apart. The first test performed on the indicated day and the second test performed 48 hours later.

3 Tests = three (3) tests performed an average of 48 hours apart. The first test performed on the indicated day, the second test performed 48 hours later, and a final test performed 48 hours after the second test.

Page 14

510(K) SUMMARY
Page 10 of 10

Flex Studies

Using risk analysis as a guide, analytical flex studies were conducted on BinaxNOW COVID-19 Ag Card. The testing evaluated various sources of potential human errors and environmental factors that could affect the accuracy of results, including those related to sample handling, reagent handling, and extremes of operational conditions. The studies demonstrated that the test is robust to usage variation and environmental factors that may be encountered. It was observed that exposure of the device test strip directly to wet cleaning solutions such as bleach, ethanol or isopropyl alcohol or excessive glove powder may cause erroneous results. Therefore, the Instructions for Use include a statement cautioning users to ensure the test card is placed on a clean, dry surface to initiate the test.

Conclusion

The data presented in this 510(k) premarket notification demonstrate that the subject device, BinaxNOW COVID-19 Ag Card, is substantially equivalent to the predicate device (Nano-Check COVID-19 Antigen Test, K231187). The differences in the BinaxNOW COVID-19 Ag Card (proposed device) and the Nano-Chek COVID-19 Antigen Test (predicate device, K231187) are limited to the Intended Use population (individuals with symptoms within five (5) days of symptom onset vs. individuals with symptoms within four (4) days of symptom onset). This difference does not affect the overall substantial equivalence of the proposed device to the predicate device in terms of the technological similarity, intended use, safety, and effectiveness. Furthermore, the information contained within this notification demonstrates BinaxNOW COVID-19 Ag Card compliance with the special controls applicable to a simple point-of-care device to directly detect SARS-CoV-2 viral target from clinical specimens in near-patient settings.

There is no known potential adverse effect to the operator when using this in vitro device according to the BinaxNOW COVID-19 Ag Card package insert.

N/A