K Number
K250259
Date Cleared
2025-06-04

(127 days)

Product Code
Regulation Number
870.2910
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TeleRehab® Aermos Cardiopulmonary Rehabilitation System is intended to acquire and condition the ECG signal from a patient so that it can be transmitted wirelessly from a radiofrequency transmitter to a workstation in a hospital or a clinical setting where the data is displayed and analyzed. This device also measures heart rate and provides visual and audible alarms if the patient's heart rate goes out of a prescribed range. This device is for use with ambulatory adult patients who need monitoring while undergoing cardiovascular and/or pulmonary rehabilitation. The physiological data from monitoring and other patient information (such as patient demographics, exercise protocol and medical information) is stored in a database for tracking and reporting of the patients' progress through rehabilitation.

Device Description

The TeleRehab® Aermos Cardiopulmonary Rehabilitation System ("Aermos") provides the ECG monitoring functionality required for performing rehabilitation of cardiovascular and/or pulmonary patients. Patients' ECG may be monitored using the Aermos system during exercise under clinical supervision. During monitoring, Aermos provides both visual and audible alarms if the patient's heart rate goes out of a prescribed range. The heart rate alarm indication is one of multiple inputs a clinician may use to modify and adjust rehabilitation activities such as decreasing the patient's level of physical exertion or halting the exercise entirely.

Aermos also provides the ability to plan a patient's rehabilitation program and document the patient's progress through the creation of various types of reports. The report types supported in Aermos include individual treatment plan reports, daily exercise session reports and various patient information reports. Additionally, the Aermos system provides the ability to transfer various report types to the hospital Electronic Medical Records system.

The main components of Aermos are Argus ECG transmitters, the Aermos Workstation and associated networking equipment.

AI/ML Overview

This FDA 510(k) clearance letter pertains to the TeleRehab Aermos Cardiopulmonary Rehabilitation System, which is a device for monitoring ECG signals and heart rate during patient rehabilitation. The provided documentation (the 510(k) Summary) details non-clinical bench testing for performance and safety but explicitly states that clinical testing was not applicable.

Therefore, based on the provided document, the following information regarding acceptance criteria and a study that proves the device meets those criteria, specifically concerning an AI/algorithm-driven component with clinical performance metrics, cannot be fully extracted or is explicitly stated as not performed:

Here's an analysis of the provided information relative to your request:

Acceptance Criteria and Device Performance (Based on Non-Clinical Bench Testing)

Since no clinical study was performed, the "reported device performance" would pertain to the results of non-clinical bench testing against established performance standards. The document does not provide specific quantitative acceptance criteria or reported numerical performance results for the device. Instead, it states that the device's specifications were "verified through internal verification testing" and its usability "evaluated through internal validation testing," and that it complies with various international standards.

Acceptance Criteria CategoryAcceptance Criteria (General, Inferred from Standards Compliance)Reported Device Performance
ECG Signal AcquisitionCompliance with IEC 60601-2-27 (electrocardiographic monitoring equipment) for frequency response and dynamic range.Verified through compliance with IEC 60601-2-27. Specific values (e.g., 0.05 - 100 Hz, ±5.0 mV) are stated as specifications but detailed test results against specific acceptance criteria for these are not provided in this summary.
Heart Rate MeasurementAccurate heart rate calculation.Part of ECG signal processing; compliance with IEC 60601-2-27 implies performance. Exact accuracy metrics not reported.
Alarm FunctionalityVisual and audible alarms for out-of-range heart rate; compliance with IEC 60601-1-8 (alarm systems).Compliance with IEC 60601-1-8 for alarm systems.
Wireless TransmissionReliable and safe wireless data transmission (WiFi); compliance with ANSI C63.27 and IEC 60601-1-2.Verified through compliance with ANSI C63.27 and IEC 60601-1-2, and applicable FDA guidance/consensus documents for RF wireless and cybersecurity.
Software FunctionalitySoftware verification, validation, and adherence to FDA guidance for device software functions (Enhanced Documentation level).Software V&V conducted at unit, integration, system levels, documentation as per FDA guidance (June 2023).
CybersecurityCompliance with FDA guidance on cybersecurity in medical devices.Complete risk-based cybersecurity assessment and testing performed per FDA guidance (Sept. 2023).
Cleaning & DisinfectionVerification and validation of cleaning and disinfection processes.Internal and external testing performed as per FDA guidance (March 2015).
General Safety & PerformanceCompliance with IEC 60601-1 (general safety), IEC 60601-1-6 (usability), ISO 14971 (risk management), etc.Compliance with a comprehensive list of IEC, ANSI/AAMI, and ISO standards is reported.

Study Details (Based on the provided 510(k) Summary)

  1. A table of acceptance criteria and the reported device performance:

    • See the table above. Specific quantitative acceptance criteria beyond "compliance with standard" are not provided in this regulatory summary.
  2. Sample size used for the test set and the data provenance:

    • The document explicitly states "Clinical Testing: Not applicable."
    • For the non-clinical bench testing, specific sample sizes (e.g., number of devices tested, number of test cases) are not detailed in this 510(k) summary.
    • Data provenance for non-clinical testing would typically be internal laboratory data generated during device development and verification. There is no mention of geographical origin or retrospective/prospective nature as this was not clinical data.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable, as no clinical study with human interpretation/ground truth was performed. The "ground truth" for bench testing would be defined by validated test equipment and reference standards.
  4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable, as there was no study involving human readers or interpretation requiring adjudication.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done, as clinical testing was "Not applicable." The device is a physiological signal monitor, not an AI-assisted diagnostic tool that interprets images or signals requiring human reader comparison.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The core functionality of the device (ECG acquisition, heart rate measurement, alarms) is algorithmic. The performance of these algorithms would have been assessed during the non-clinical bench testing, which is essentially "standalone algorithm" testing against known inputs and expected outputs. Specific quantitative results (e.g., algorithm accuracy for heart rate) are not provided in this summary beyond "compliance with IEC 60601-2-27" and "ANSI/AAMI EC57: 2012, Testing and Reporting Performance Results of Cardiac Rhythm and ST-Segment Measure Algorithms."
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For non-clinical bench testing, the "ground truth" is typically established by:
      • Reference standards and calibrated test equipment: For electrical performance, signal acquisition accuracy, frequency response, etc.
      • Simulated physiological signals: For testing heart rate calculation and alarm thresholds.
      • Design specifications and established engineering principles: For software functionality and cybersecurity.
  8. The sample size for the training set:

    • Not applicable. The device is a monitoring system and not primarily driven by a deep learning or machine learning algorithm that requires a "training set" in the sense of a large dataset for model development. The algorithms for heart rate calculation, etc., are likely traditional signal processing algorithms.
  9. How the ground truth for the training set was established:

    • Not applicable, as there was no training set for a machine learning model.

Summary of Device Nature and Regulatory Pathway:

The TeleRehab Aermos Cardiopulmonary Rehabilitation System is a Class II device (Product Codes DRG, DRT) which functions as a physiological signal transmitter and receiver. It monitors ECG and heart rate and provides alarms. Its 510(k) clearance relied on demonstrating substantial equivalence to predicate devices primarily through non-clinical bench testing against recognized industry standards (e.g., IEC 60601 series, ANSI/AAMI, ISO standards) and adherence to FDA guidance documents (e.g., for software, cybersecurity, reprocessing). The explicit statement "Clinical Testing: Not applicable" indicates that the nature of the device and its intended use, combined with the comprehensive non-clinical data, satisfied the FDA's requirements for demonstrating safety and effectiveness without the need for a human-subject clinical study. This is common for devices that are evolutionary improvements on existing technologies with well-understood performance parameters.

FDA 510(k) Clearance Letter - TeleRehab Aermos Cardiopulmonary Rehabilitation

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 4, 2025

The ScottCare Corporation
Matthew Tarler
Director of Quality and Regulatory Affairs
4791 West 150th Street
Cleveland, Ohio 44135

Re: K250259
Trade/Device Name: TeleRehab Aermos Cardiopulmonary Rehabilitation
Regulation Number: 21 CFR 870.2910
Regulation Name: Radiofrequency Physiological Signal Transmitter And Receiver
Regulatory Class: Class II
Product Code: DRG, MWI, DRT
Dated: May 1, 2025
Received: May 2, 2025

Dear Matthew Tarler:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250259 - Matthew Tarler Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250259 - Matthew Tarler Page 3

Sincerely,

Kimberly N. Crowley -S Digitally signed by Kimberly N. Crowley -S

For: Jennifer Kozen
Assistant Director
DHT2A: Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known): K250259

Device Name: TeleRehab Aermos Cardiopulmonary Rehabilitation

Indications for Use (Describe)

The TeleRehab Aermos Cardiopulmonary Rehabilitation System is intended to acquire and condition the ECG signal from a patient so that it can be transmitted wirelessly from a radiofrequency transmitter to a workstation in a hospital or a clinical setting where the data is displayed and analyzed. This device also measures heart rate and provides visual and audible alarms if the patient's heart rate goes out of a prescribed range. This device is for use with ambulatory adult patients who need monitoring while undergoing cardiovascular or pulmonary rehabilitation. The physiological data from monitoring and other patient information (such as patient demographics, exercise protocol and medical information) is stored in a database for tracking and reporting of the patients' progress through rehabilitation.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

510(k) Summary K250259

Page 1 of 6
1619692645.2

1. Sponsor/Applicant

The ScottCare Corporation
4791 West 150th Street
Cleveland, Ohio 44135

Dr. Matthew Tarler
Phone Number: (216) 362-0550 x113
Email: regulatory@scottcare.com
Fax: (216) 267-6129

Summary Preparation Date: May 01, 2025

2. Device

Trade NameTeleRehab® Aermos Cardiopulmonary Rehabilitation System
Model Number102206
ClassificationClass II
Product CodeDRG, DRT
Regulation Number21 CFR 870.2910, 21 CFR 870.2300
Review PanelCardiovascular

3. Predicate Device

ers2 - ergoline Rehabilitation System ("ers2") - K212883

Reference Device
iT20 Telemetry Transmitter - Edan Instruments, Inc. - K161056

Page 6

Page 2 of 6
1619692645.2

4. Device Description

The TeleRehab® Aermos Cardiopulmonary Rehabilitation System ("Aermos") provides the ECG monitoring functionality required for performing rehabilitation of cardiovascular and/or pulmonary patients. Patients' ECG may be monitored using the Aermos system during exercise under clinical supervision. During monitoring, Aermos provides both visual and audible alarms if the patient's heart rate goes out of a prescribed range. The heart rate alarm indication is one of multiple inputs a clinician may use to modify and adjust rehabilitation activities such as decreasing the patient's level of physical exertion or halting the exercise entirely.

Aermos also provides the ability to plan a patient's rehabilitation program and document the patient's progress through the creation of various types of reports. The report types supported in Aermos include individual treatment plan reports, daily exercise session reports and various patient information reports. Additionally, the Aermos system provides the ability to transfer various report types to the hospital Electronic Medical Records system.

The main components of Aermos are Argus ECG transmitters, the Aermos Workstation and associated networking equipment.

5. Indications for Use

The TeleRehab® Aermos Cardiopulmonary Rehabilitation System is intended to acquire and condition the ECG signal from a patient so that it can be transmitted wirelessly from a radiofrequency transmitter to a workstation in a hospital or a clinical setting where the data is displayed and analyzed. This device also measures heart rate and provides visual and audible alarms if the patient's heart rate goes out of a prescribed range. This device is for use with ambulatory adult patients who need monitoring while undergoing cardiovascular and/or pulmonary rehabilitation. The physiological data from monitoring and other patient information (such as patient demographics, exercise protocol and medical information) is stored in a database for tracking and reporting of the patients' progress through rehabilitation.

510(k) Summary K250259

Page 7

Page 3 of 6
1619692645.2

6. Technological Characteristics and Substantial Equivalence

Subject DevicePredicate DeviceReference DeviceComparison
DeviceAermosers2iT20 Telemetry Transmitter
ApplicantThe ScottCare Corporationergoline GmbHEdan Instruments, Inc.
510(k) #K250259K212883K161056
Device ClassClass IIClass IIClass II
Regulation21 CFR 870.291021 CFR 870.230021 CFR 870.291021 CFR 870.102521 CFR 870.230021 CFR 870.270021 CFR 882.132021 CFR 870.234021 CFR 870.2910
Classification CodeDRGDRTDRGMHX, DSI, MLD,DRT,DQA,GXY,DPS,DRGSimilar⁽¹⁾
Prescription/OTC UsePrescriptionPrescriptionPrescription
Indications for UseThe TeleRehab® Aermos Cardiopulmonary Rehabilitation System is intended to acquire and condition the ECG signal from a patient so that it can be transmitted wirelessly from a radiofrequency transmitter to a workstation in a hospital or a clinical setting where the data is displayed and analyzed. This device also measures heart rate and provides visual and audible alarms if the patient's heart rate goes out of a prescribed range. This device is for use with ambulatory adult patients who need monitoring while undergoing cardiovascular and/or pulmonary rehabilitation. The physiological data from monitoring and other patient information (such as patient demographics, exercise protocol and medical information) is stored in a database for tracking and reporting of the patients' progress through rehabilitation.The ers2 – ergoline Rehabilitation System is a device for recording of a single channel, bipolar surface ECG (frontal plane) acquired with two ECG electrodes. The ECG is transmitted to the ers2 software where it is processed for heartrate calculation, displayed for visual QRS complex assessment and to control the training load for a patient´s rehabilitation or preventive training activities. The ers2 system is not intended to be used to detect or diagnose cardiac conditions (e.g. arrhythmias, ST- elevation, etc. pp). The signal is acquired on the intact skin of adult patients. The medical device is intended for use in professional healthcare Institutions for inpatient and outpatient care.The iT20 telemetry transmitter is intended to monitor physiological parameters including: ECG, oxygen saturation of arterial blood (SpO2) and pulse rate (PR) for adults and pediatric patients. The iT20 requires the EDAN MFM-CMS (Central Monitoring Station) to provide full functionality of the device.The iT20 telemetry transmitter is intended to be used in clinical divisions of hospital environments, including CCU and general wards (as Cardiology Dept.).
Intended PopulationAdultAdultAdult / Pediatric
Intended EnvironmentHealthcare facility (e.g. hospital or clinic)Healthcare facility (e.g. hospital or clinic)Healthcare facility (e.g. hospital or clinic)

510(k) Summary K250259

Page 8

Page 4 of 6
1619692645.2

Number of ECG Channels11MultipleSame
ECG Transmitter Patient Cable3 lead wires2 lead wires3 or 5 lead wiresSimilar⁽³⁾
ECG Electrode TypeOff-the-shelf standard ECG electrodesOff-the-shelf standard ECG electrodesOff-the-shelf standard ECG electrodesSame
ECG Transmitter AcquisitionFrequency response: 0.05 - 100 HzDynamic range: ±5.0 mVFrequency response: 0.05 – 125 HzDynamic range: ±6.0 mVMonitor: 0.5Hz to 40HzRange: ±2.0 mVSimilar⁽⁴⁾
ECG Transmitter Data Transmission Method to WorkstationRadiofrequency - WiFi 802.11 b/g/n,2.4 GHz via off-the-shelf Wireless Access Point and Network RouterRadiofrequency Bluetooth 2402 – 2483.5 MHzRadiofrequency - WiFi 802.11 b/g/n, 2.4 GHzSimilar⁽⁵⁾
Pacemaker Pulse DetectionYesNoYesDifferent
Heart Rate DisplayYesYesYesSame
ECG DisplayYesYesYesSame
Heart Rate AlarmYesYesYesSame
Defibrillation ProtectionYesYesNot AvailableSame
WorkstationPC with Microsoft WindowsPC with Microsoft WindowsRequires EDAN MFM-CMA (Central Monitoring Station) - Windows based OSSame
Key Safety and Performance StandardsIEC 60601-1IEC 60601-1-2IEC 60601-2-27IEC 60601-1-8IEC 60601-1IEC 60601-1-2IEC 60601-2-27IEC 60601-1IEC 60601-1-2IEC 60601-2-27IEC 60601-1-8EC13ISO 80601-2-61IEC 60601-2-49ISO 10993-1,5,5-10Similar⁽⁶⁾

⁽¹⁾ The proposed Aermos device and predicate device both have the same indications for use, intended population, and intended environment, hence the DRT code. Based on the Aermos device having heart rate alarms the DRT code (regulation 870.2300) is also included in addition to the DRG code. The reference device includes the DRT code and serves as a predicate for presentation of heart rate alarms.

⁽²⁾ Like the predicate, the proposed Aermos device is intended for monitoring ambulatory adult patients who need ECG monitoring while undergoing cardiovascular and/or pulmonary rehabilitation in a hospital or clinic setting. Both systems capture ECG signal from the patient using an ECG transmitter, transmit it to a Windows-based workstation via RF transmission and display the ECG data on a computer monitor.

⁽³⁾ The subject devices and the predicate device capture and display one channel of ECG using 3 lead wires and 2 lead wires respectively. The reference device also provides a 3 lead wire option. Safety and essential performance of the Aermos system was evaluated through compliance with the IEC 60601-2- 27 standard. Safety and performance of the Aermos patient cable was evaluated through compliance with the ANSI/AAMI EC53 standard. Therefore, this difference does not raise any new concerns for safety or effectiveness for the subject device's intended use.

510(k) Summary K250259

Page 9

Page 5 of 6
1619692645.2

⁽⁴⁾ The ECG acquisition performance of Aermos for its intended use has been verified through compliance with the IEC 60601-2-27 standard. Therefore, any differences in this regard do not raise new concerns of safety or effectiveness for the subject device's intended use.

⁽⁵⁾ The Aermos system and the reference device uses WiFi for transmission of ECG data from its ECG transmitter to its workstation, while the predicate uses Bluetooth 2402 – 2483.5 MHz radiofrequency band for data transmission. The safety and performance of the WiFi transmission of Aermos was evaluated through compliance with the ANSI C63.27 and IEC 60601-1-2 standards. The device was also evaluated per FDA guidance documents "Radio Frequency Wireless Technology in Medical Devices 2013" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices 2023", and consensus document "AAMI TIR69:2017 Risk Management of Radio-Frequency Wireless Coexistence for Medical Devices and Systems." Therefore, the use of Wi-Fi does not raise new concerns of safety or effectiveness for its intended use.

7. Non-clinical Bench (Performance) testing

Aermos' specifications were verified through internal verification testing and its customer requirements and usability were evaluated through internal validation testing.

Software verification and validation testing were conducted at the unit, integration, and system levels, and documentation is provided as recommended in FDA's "Guidance for the Content of Premarket Submissions for Device Software Functions: Guidance for Industry and FDA Staff" (June 2023). The software for this device requires an Enhanced Documentation level, since the device "could present a hazardous situation with a probable risk of death or serious injury, either to a patient, user of the device, or others in the environment of use" and the guidance recommends that "[t]hese risks should be assessed prior to implementation of risk control measures."

Cybersecurity concerns and documentation were addressed as recommended by FDA's "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions: Guidance for Industry and FDA Staff" (Sept. 2023). A complete risk-based cybersecurity assessment and testing in accordance with the FDA guidance document was performed.

Internal and external testing for cleaning and disinfection verification and validation was performed as per FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling: Guidance for Industry and Food and Drug Administration Staff" (Mar. 2015).

Safety and effectiveness of Aermos is demonstrated through testing conducted in compliance with the following standards and guidance documents:

• IEC 60601-1:2020, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance

510(k) Summary K250259

Page 10

Page 6 of 6
1619692645.2

• IEC 60601-1-2:2020, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic disturbances – Requirements and Tests

• IEC 60601-1-6:2020, Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability

• IEC 60601-1-8:2020, General requirements for basic safety and essential performance – Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems.

• IEC 60601-2-27:2011, Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment

• ANSI/AAMI EC57: 2012, Testing and Reporting Performance Results of Cardiac Rhythm and ST-Segment Measure Algorithms

• ANSI C63.27-2017, American National Standard for Evaluation of Wireless Coexistence

• Applicable requirements from the Recognized Consensus Document "AAMI TIR69:2017 Risk Management of Radio-Frequency Wireless Coexistence for Medical Devices and Systems"

• IEC 62304:2015, Medical device software - Software life-cycle processes

• Applicable requirements from FDA Guidance Document "Radio Frequency Wireless Technology in Medical Devices:Guidance for Industry and FDA Staff" (Aug. 2013)

• Applicable requirements from FDA Guidance Document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions: Guidance for Industry and Food and Drug Administration Staff" (Sept. 2023)

• Argus Battery complies with the standard IEC 62133:2012, Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications

• Argus Patient Cable complies with the standard ANSI/AAMI EC53:2013, ECG trunk cables and patient lead wires

• ISO 14971:2019, Medical Devices - Application Of Risk Management To Medical Devices

• IEC 60601-4-2 - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems

8. Clinical Testing

Not applicable.

9. Conclusion:

Based on the intended use, technological characteristics, and performance data provided in the submission, the Aermos device is substantially equivalent to the predicate device. The differences between the Aermos device and the predicate do not raise any questions regarding safety and effectiveness.

510(k) Summary K250259

§ 870.2910 Radiofrequency physiological signal transmitter and receiver.

(a)
Identification. A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.(b)
Classification. Class II (performance standards).