K Number
K250150
Device Name
VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood cardioplegia set); VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood cardioplegia set with shunt); VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood Cardioplegia Set Mini)
Date Cleared
2025-05-30

(129 days)

Product Code
Regulation Number
870.4240
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Vanguard 4:1 Blood cardioplegia sets are cardiopulmonary bypass devices allowing cardioplegia fluid delivery. They also regulate cardioplegia fluid temperature and trap air emboli. The devices are intended to be used for 6 hours or less.
Device Description
The Vanguard Blood Cardioplegia Systems are single-use, non-toxic, non-pyrogenic and supplied sterile in individual packs. They are cardiopulmonary bypass devices allowing cardioplegia fluid delivery; they also regulate cardioplegia fluid temperature and trap air emboli. Vanguard devices are indicated for use for patients undergoing surgical procedures requiring cardiopulmonary bypass procedures. These Vanguard Blood Cardioplegia Systems are circuits designed to administer cardioplegic solution and they can be used up to 6 hours. The Vanguard Blood Cardioplegia Systems are the modified version of the disposables currently marketed in the SORIN BLOOD CARDIOPLEGIA SYSTEM (K934763) and the NGBCD & NGBCDP CARDIOPLEGIA DELIVERY SYSTEMS (K934847).
More Information

No
The provided text describes a medical device for delivering cardioplegia fluid during cardiopulmonary bypass procedures. It focuses on the device's function, components, and technical specifications, with no mention of AI models or related terms.

No.
A therapeutic device is one that treats a disease or condition. This device is used during a surgical procedure to deliver cardioplegia fluid and regulate its temperature, but it does not directly treat a disease or condition itself.

No

Explanation: The device is described as a cardiopulmonary bypass device used for delivering cardioplegia fluid, regulating its temperature, and trapping air emboli during surgical procedures. It is a therapy delivery device, not one that gathers or analyzes information about a patient's health condition for diagnosis.

No

The device is described as single-use, non-toxic, non-pyrogenic, sterile, and consisting of circuits designed to administer cardioplegic solution, regulate temperature, and trap air, indicating it is a hardware-based medical device (disposable sets).Yes

The device is a software-only medical device because the 510(k) summary makes no mention of any physical hardware components, indicating the device's functionality is derived purely from software.

No
The device is a cardiopulmonary bypass device for delivering cardioplegia fluid during surgery, not a device used to examine specimens taken from the human body for diagnostic purposes.

N/A

Intended Use / Indications for Use

Vanguard 4:1 Blood cardioplegia sets are cardiopulmonary bypass devices allowing cardioplegia fluid delivery. They also regulate cardioplegia fluid temperature and trap air emboli. The devices are intended to be used for 6 hours or less.

Product codes

DTR

Device Description

The Vanguard Blood Cardioplegia Systems are single-use, non-toxic, non-pyrogenic and supplied sterile in individual packs.

They are cardiopulmonary bypass devices allowing cardioplegia fluid delivery; they also regulate cardioplegia fluid temperature and trap air emboli.

Vanguard devices are indicated for use for patients undergoing surgical procedures requiring cardiopulmonary bypass procedures.

These Vanguard Blood Cardioplegia Systems are circuits designed to administer cardioplegic solution and they can be used up to 6 hours.

The Vanguard Blood Cardioplegia Systems are the modified version of the disposables currently marketed in the SORIN BLOOD CARDIOPLEGIA SYSTEM (K934763) and the NGBCD & NGBCDP CARDIOPLEGIA DELIVERY SYSTEMS (K934847).

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

No clinical testing was conducted in support of the Vanguard Blood Cardioplegia Systems, as the indications for use and technical characteristics are equivalent to those of the predicate devices, which have been on the market for several years with proven safety and efficacy of use. The non-clinical testing summarized in this submission supports the substantial equivalence of the subject devices with the predicate devices when used according to their intended use.

Key Metrics

Not Found

Predicate Device(s)

K934763, K934847

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.4240 Cardiopulmonary bypass heat exchanger.

(a)
Identification. A cardiopulmonary bypass heat exchanger is a device, consisting of a heat exchange system used in extracorporeal circulation to warm or cool the blood or perfusion fluid flowing through the device.(b)
Classification. Class II (performance standards).

FDA Approval Letter - VANGUARD Blood Cardioplegia Systems

Page 1

U.S. Food & Drug Administration

10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

May 30, 2025

Sorin Group Italia S.r.l.
Martina Carlini
Regulatory Affairs Specialist
Via Statale 12 Nord, 86
Mirandola (Modena), IT 41037
Italy

Re: K250150
Trade/Device Name: VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood cardioplegia set); VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood cardioplegia set with shunt); VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood Cardioplegia Set Mini)
Regulation Number: 21 CFR 870.4240
Regulation Name: Cardiopulmonary bypass heat exchanger
Regulatory Class: Class II
Product Code: DTR
Dated: January 18, 2025
Received: January 21, 2025

Dear Martina Carlini:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250150 - Martina Carlini Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250150 - Martina Carlini Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support,
Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known)
K250150

Device Name
VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood cardioplegia set);
VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood cardioplegia set with shunt);
VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood Cardioplegia Set Mini)

Indications for Use (Describe)
Vanguard 4:1 Blood cardioplegia sets are cardiopulmonary bypass devices allowing cardioplegia fluid delivery. They also regulate cardioplegia fluid temperature and trap air emboli. The devices are intended to be used for 6 hours or less.

Type of Use (Select one or both, as applicable)

  • Prescription Use (Part 21 CFR 801 Subpart D)
  • Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

(in accordance with 21 CFR 807.92)

510(k) Number: K250150

I. Applicant Information

Applicant:
SORIN GROUP ITALIA S.R.L.
Via Statale 12 Nord, 86
Mirandola, MO 41037 Italy

Contact Person: Martina Carlini
Regulatory Affairs Specialist
Tel: +39 0535 29811
e-mail: martina.carlini@livanova.com

Application Correspondent:
SORIN GROUP ITALIA S.R.L.
Via Statale 12 Nord, 86
Mirandola, MO 41047 Italy

Contact Person: Martina Carlini
Regulatory Affairs Specialist
Tel: +39 0535 29811
e-mail: martina.carlini@livanova.com

Date Prepared: May 30th 2025

II. Subject Device Identification

Device Trade Name:

  • VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood cardioplegia set);
  • VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood cardioplegia set with shunt);
  • VANGUARD Blood cardioplegia Systems (Vanguard 4:1 Blood Cardioplegia Set Mini)

Classification Name: Heat-Exchanger, Cardiopulmonary Bypass
Regulation Number: 21 CFR 870.4240
Product Code: DTR
Classification: Class II
Classification Panel: Cardiovascular

Page 6

III. Predicate Device

The Vanguard Blood Cardioplegia Systems are substantially equivalent to the following cleared predicate devices. Both modified and unmodified models have the same fundamental scientific technology and intended use:

Vanguard 4:1 Blood cardioplegia set (with and without shunt)
510(k) Number: K934763
Device Trade Name: SORIN BLOOD CARDIOPLEGIA SYSTEM
Classification Name: Reservoir, Blood, Cardiopulmonary Bypass
Regulation Number: 21 CFR 870.4400
Product Code: DTN
Classification: Class II
Classification Panel: Cardiovascular

Vanguard 4:1 Blood Cardioplegia Set Mini
510(k) Number: K934847
Device Trade Name: NGBCD & NGBCDP CARDIOPLEGIA DELIVERY SYSTEMS
Classification Name: Reservoir, Blood, Cardiopulmonary Bypass
Regulation Number: 21 CFR 870.4400
Product Code: DTN
Classification: Class II
Classification Panel: Cardiovascular

Note: These 2 predicates were classified as Reservoir, Blood, Cardioplumonary Bypass linked to the regulation number 21 CFR 870.4400 and to the product code DTN, but since the subject products are cardioplegia systems containing heath-exchangers, we believe that the correct classification should be Heat-Exchanger, Cardiopulmonary Bypass linked to the regulation number 21 CFR 870.4240 and to the product code DTR.

IV. Device Description

The Vanguard Blood Cardioplegia Systems are single-use, non-toxic, non-pyrogenic and supplied sterile in individual packs.

They are cardiopulmonary bypass devices allowing cardioplegia fluid delivery; they also regulate cardioplegia fluid temperature and trap air emboli.

Vanguard devices are indicated for use for patients undergoing surgical procedures requiring cardiopulmonary bypass procedures.

These Vanguard Blood Cardioplegia Systems are circuits designed to administer cardioplegic solution and they can be used up to 6 hours.

The Vanguard Blood Cardioplegia Systems are the modified version of the disposables currently marketed in the SORIN BLOOD CARDIOPLEGIA SYSTEM (K934763) and the NGBCD & NGBCDP CARDIOPLEGIA DELIVERY SYSTEMS (K934847).

V. Indications for Use

Vanguard 4:1 Blood cardioplegia sets are cardiopulmonary bypass devices allowing cardioplegia fluid delivery. They also regulate cardioplegia fluid temperature and trap air emboli. The devices are intended to be used for 6 hours or less.

Page 7

VI. Summary of Technical Characteristics

The Vanguard Blood Cardioplegia Systems have the same fundamental technological characteristics, principles of operation and control mechanisms as the unmodified devices.

The tubing and connectors of the Vanguard devices made of PVC materials were changed within the modified Vanguard Blood Cardioplegia Systems in order to remove the diethylhexyl phthalate (DEHP) used as plasticizer currently present in the device.

Moreover, the rigid connectors of the Vanguard devices made of PVC materials were changed within the modified Vanguard Blood Cardioplegia Systems in order to remove the DOTE used as thermal stabilizer.

Then, the PP Purell material of line clamps of the Vanguard devices was replaced within the modified Vanguard Blood Cardioplegia Systems with another PP material (Bormed); and the mesh Versapor material for the vented spike component of the Vanguard devices was replaced within the modified Vanguard Blood Cardioplegia Systems with another kind of Versapor material.

Additionally, some technical specifications for Vanguard Blood Cardioplegia Systems were updated and tested to better reflect the state of the art of the devices.

No other design changes have been made to the devices.

The devices are ethylene oxide sterilized and have a non-pyrogenic fluid path. They are for single use only.

VII. Non-Clinical Performance Data

Sorin Group Italia S.r.l. has conducted extensive verification and validation testing of the Vanguard Blood Cardioplegia Systems, as disposables used to administer cardioplegic solution during surgical procedures requiring cardiopulmonary bypass procedures.

The Vanguard Blood Cardioplegia Systems comply with all the applicable voluntary standards related to cardiovascular systems. The devices passed all the testing in accordance with national and international standards.

VIII. Clinical Performance Data

No clinical testing was conducted in support of the Vanguard Blood Cardioplegia Systems, as the indications for use and technical characteristics are equivalent to those of the predicate devices, which have been on the market for several years with proven safety and efficacy of use. The non-clinical testing summarized in this submission supports the substantial equivalence of the subject devices with the predicate devices when used according to their intended use.

IX. Statement of Substantial Equivalence

Based on equivalent intended use and technological characteristics, as well as on equivalent performance testing, the Vanguard Blood Cardioplegia Systems can be deemed to be substantially equivalent to their predicate devices:

  • the unmodified NGBCW4 Warm Blood Cardioplegia Delivery System 4:1 blood-to-crystalloid ratio and NGBCW4C Warm Blood Cardioplegia Delivery System 4:1 blood-to-crystalloid ratio with shunt for delivering 100% blood, cleared under K934763;
  • the unmodified NGBCDP Reduced Priming Volume Blood Cardioplegia Delivery System 4:1 blood-to-crystalloid ratio, cleared under K934847;

The Vanguard Blood Cardioplegia Systems, as designed and manufactured, do not raise new questions regarding safety and effectiveness as compared to their predicate devices and are determined to be substantially equivalent to their predicate devices listed above.