K Number
K250138
Date Cleared
2025-04-02

(75 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Small Volume 0.2mL Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

Device Description

The Small Volume 0.2mL Syringe is intended for manual, single use. It is a small volume 0.2 mL syringe. It is comprised of a PP barrel, ABS plunger, TPE plunger stopper, PP plunger shield and PP Luer protective cap. The device has a Luer lock connection type that enables connection to a hypodermic needle for liquid injection or withdrawal. The device has a plunger stopper to prevent any liquid residual from remaining in the syringe barrel, maximizing the efficiency of liquid usage. The feature is called Low Dead Space/Volume. The Luer protective cap serves to safeguard the small Luer connector from potential damage during transportation. Similarly, the plunger shield is utilized to ensure the small plunger remains undamaged during transportation. The Luer protective cap and the plunger shield are optional to be used. The device has been manufactured without any silicone oil. The device has six models, each equipped with a 0.2 mL Luer lock syringe. Among these models, three include an additional feature for enhanced dosing technology (EDT), which comprises a colored reference line on the plunger surface to improve visibility. These include identifying markings to distinguish the EDT models from the regular 0.2 mL syringe.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the "Small Volume 0.2mL Syringe" do not contain any information related to AI/ML or software-driven medical devices. The submission is for a physical medical device (syringe) and, as such, the acceptance criteria and study descriptions are entirely focused on the physical and material properties, performance, biocompatibility, and sterility of the syringe itself.

Therefore, I cannot provide details on:

  • A table of acceptance criteria for AI/ML performance
  • Sample sizes for a test set or data provenance for AI/ML
  • Number of experts and their qualifications for establishing AI/ML ground truth
  • Adjudication methods for AI/ML test sets
  • MRMC comparative effectiveness study for AI with human readers
  • Standalone AI algorithm performance
  • Type of ground truth for AI/ML
  • Training set sample size for AI/ML
  • How ground truth for AI/ML training was established

The document explicitly states "9. Clinical testing: Not applicable for this submission." This further confirms that no studies involving human interaction or clinical performance for a software/AI component were conducted or required for this purely mechanical device.

If the request was intended to be about the acceptance criteria and studies for the physical device as described in the 510(k) summary, here is the relevant information extracted:


Acceptance Criteria and Study for the "Small Volume 0.2mL Syringe" (Physical Device)

The device, a "Small Volume 0.2mL Syringe," is a physical medical device. The acceptance criteria and the studies proving it meets these criteria are related to its mechanical performance, material safety, and sterility, in accordance with recognized medical device standards. There is no AI/ML component described in this 510(k) submission.

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary references compliance with specific standards, rather than providing explicit numerical acceptance criteria directly in a table. The device performance is generally stated as "met the standards' requirements."

CategoryAcceptance Criteria (Standard Reference)Reported Device Performance
Material SafetyComplies with ISO 10993-1: - Cytotoxicity - Irritation - Sensitization - Acute systemic toxicity - Hemocompatibility - Pyrogenicity - Particulate Matter per USP <788> - Bacterial Endotoxins per USP <85>All tested endpoints met the acceptance criteria (Implied: complied with ISO 10993-1, USP <788>, and USP <85>).
Syringe PerformanceComplies with ISO 7886-1:2017 (e.g., accuracy, force, freedom from leakage, residual volume)Met all necessary requirements of the standard.
Luer Connector PerformanceComplies with ISO 80369-7:2016 and ISO 80369-20:2015 (e.g., connection integrity, leakage)Met the standards' requirements.
Dead Space Specification≤ 0.0058 mL (95% confidence/ 95% reliability)Meets requirements (stated as "a lower dead space volume is achieved" compared to predicate).
SterilizationComplies with ISO 11135 (EO sterilization process validation)Validation results confirm compliance.
Ethylene Oxide (EO) & Ethylene Chlorohydrin (ECH) ResidualsDo not exceed limits according to ISO 10993-7Residues do not exceed the limits.
Shelf-LifeDevice functionality maintained through 3 years (accelerated aging per ASTM F1980-16)Demonstrated device functionality through the claimed shelf-life of 3 years. (Real-time testing is ongoing to confirm).
Packaging IntegrityComplies with ISO 11607-1, ISO 11607-2, ASTM D4169-22, ASTM F88/F88-15, ASTM F1929-15All packaging was deemed acceptable for protection of product and sterility maintenance.
Sterile Barrier TestingComplies with USP <71>Confirmed compliance.

2. Sample Size Used for the Test Set and Data Provenance

The exact sample sizes used for each specific test (e.g., biocompatibility endpoints, performance testing per ISO 7886-1) are not detailed in the 510(k) summary. The summary generally states that testing was conducted on "samples" or "unconditioned samples as well as samples aged... and samples transit conditioned."

  • Data Provenance: The studies were conducted by the manufacturer, Prosum Medical Limited, and their testing partners, likely in the United Kingdom or through contract laboratories that adhere to international standards. The data is prospective as it involves new testing specifically for this device submission.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not Applicable. For a physical device like a syringe, "ground truth" as it relates to expert consensus for image interpretation or clinical outcomes is not relevant. The ground truth here is derived from objective measurements and compliance with established engineering and biological standards. Testing is performed by qualified laboratory personnel following validated protocols.

4. Adjudication Method for the Test Set

Not Applicable. Adjudication typically applies to subjective assessments (e.g., expert reads). For objective measurements on a physical device, statistical analysis of measured values against predefined acceptance limits determines compliance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

Not Applicable. MRMC studies are used for assessing the impact of AI/software on human performance in diagnostic tasks and do not apply to a physical syringe device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not Applicable. This refers to AI algorithm performance, which is not part of this submission.

7. The type of ground truth used

The "ground truth" for this device's performance and safety is based on objective, quantifiable measurements and compliance with internationally recognized consensus standards (e.g., ISO 7886-1 for syringe performance, ISO 10993-1 for biocompatibility). These standards define acceptable ranges for various parameters (e.g., dead space, force to push plunger, leakage, material toxicity).

8. The sample size for the training set

Not Applicable. There is no "training set" as this is not an AI/ML device.

9. How the ground truth for the training set was established

Not Applicable. There is no "training set" as this is not an AI/ML device.

FDA 510(k) Clearance Letter - Small Volume 0.2mL Syringe

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 2, 2025

Prosum Medical Limited
℅ Scott Zawko
Manager, Regulatory Affairs
Kymanox
430 Davis Dr
Ste #300
Morrisville, North Carolina 27560

Re: K250138
Trade/Device Name: Small Volume 0.2mL Syringe
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: QNQ
Dated: January 15, 2025
Received: January 17, 2025

Dear Scott Zawko:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250138 - Scott Zawko
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250138 - Scott Zawko
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Rong Guo -S

for
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250138

Device Name: Small Volume 0.2mL Syringe

Indications for Use (Describe):
The Small Volume 0.2mL Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Page 1 of 5

1. Date Prepared

March 31, 2025

2. Submitter

Manufacturer: Prosum Medical Limited
Address: Unit 14 Trade City Business Park, Cowley Mill Road, Uxbridge, UB8 2DB, United Kingdom
Contact Person: Yanshen Sun
Phone: +44 1895 913302
Email: info@prosummedical.co.uk

3. Device

Trade name: Small Volume 0.2mL Syringe
Common Name: Syringe
Classification name: Low Dead Space Piston Syringe
Regulation: 21 CFR 880.5860
Product code: QNQ
Classification: Class II

4. Predicate Device

Predicate device: K210443, PLPT LDV Sterile Syringe

5. Device Description

The Small Volume 0.2mL Syringe is intended for manual, single use. It is a small volume 0.2 mL syringe. It is comprised of a PP barrel, ABS plunger, TPE plunger stopper, PP plunger shield and PP Luer protective cap. The device has a Luer lock connection type that enables connection to a hypodermic needle for liquid injection or withdrawal. The device has a plunger stopper to prevent any liquid residual from remaining in the syringe barrel, maximizing the efficiency of liquid usage. The feature is called Low Dead Space/Volume. The Luer protective cap serves to safeguard the small Luer connector from potential damage during transportation. Similarly, the plunger shield is utilized to ensure the small plunger remains undamaged during transportation. The Luer protective cap and the plunger shield are optional to be used. The device has been manufactured without any silicone oil.

The device has six models, each equipped with a 0.2 mL Luer lock syringe. Among these models, three include an additional feature for enhanced dosing technology (EDT), which comprises a colored reference line on the plunger surface to improve visibility. These include identifying markings to distinguish the EDT models from the regular 0.2 mL syringe.

Page 6

Page 2 of 5

Table 1. Device Models

No.Syringe Volume (mL)Luer typeDescription
10.2Luer LockSmall Volume 0.2mL Syringe, Luer Lock, EDT
20.2Luer LockSmall Volume 0.2mL Syringe, Luer Lock, EDT, Luer protective cap
30.2Luer LockSmall Volume 0.2mL Syringe, Luer Lock, EDT, Luer protective cap, Plunger shield
40.2Luer LockSmall Volume 0.2mL Syringe, Luer Lock
50.2Luer LockSmall Volume 0.2mL Syringe, Luer Lock, Luer protective cap
60.2Luer LockSmall Volume 0.2mL Syringe, Luer Lock, Luer protective cap, Plunger shield

6. Indications for Use/Intended Use

The Small Volume 0.2mL Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

7. Technological Characteristics

The table below compares the technological characteristics of the subject device and those of the predicate device.

Table 2. Device Comparison

Subject Device (Small Volume 0.2mL Syringe)Predicate Device (PLPT LDV (Low Dead Volume) Sterile Syringe, K210443)Comments
Indications for Use / Intended useThe Small Volume 0.2mL Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.The PLPT LDV (Low Dead Volume) sterile syringe is intended to be used for medical purposes to inject fluid into or withdraw fluid from the human body.Same
Product CodeQNQQNQSame
Rx Only or OTCPrescription Use OnlyPrescription Use OnlySame

Page 7

Page 3 of 5

Subject Device (Small Volume 0.2mL Syringe)Predicate Device (PLPT LDV (Low Dead Volume) Sterile Syringe, K210443)Comments
Mechanism of ActionManual useManual useSame
Components and Materials- Barrel: PP- Plunger: ABS- Stopper: TPE- Luer protective Cap: PP- Plunger shield: PP- Barrel: PP- Plunger: PP- Stopper: RubberSee Comment #1 below
Syringe Volume0.2 mL1 mLSee Comment #2 below
Connection TypeLuer-lockLuer-lockSame
SterilizationEO gasSAL: 10⁻⁶EO gasSAL: 10⁻⁶Same
Shelf-life3 years3 yearsSame
BiocompatibilityComplies with ISO 10993-1- Cytotoxicity- Irritation- Sensitization- Acute systemic toxicity- Hemocompatibility- PyrogenicityComplies with ISO 10993-1- Cytotoxicity- Irritation- Sensitization- Acute systemic toxicity- Hemocompatibility- PyrogenicitySame
Syringe Performance TestingComplies with the following standards- ISO 7886-1- ISO 80369-7Complies with the following standards- ISO 7886-1- ISO 80369-7Same
Dead Space Specification≤ 0.0058 mL95% confidence/ 95% reliability meets requirements≤ 0.023 mL95% confidence/ 95% reliability meets requirementsSee Comment #2 below

Page 8

Page 4 of 5

Comment #1, Components and Materials: The proposed device features an additional Luer protective cap and a plunger shield that safeguard the luer connector and the plunger from potential damage during transportation. The Luer protective cap and plunger shield are discarded by the user prior to use and do not alter the mechanism of action or functionality of the proposed device in comparison to the predicate. Protection of the device through transportation has been demonstrated through packaging validation per ISO 11607. The proposed device utilizes different polymer materials for the components in comparison to the predicate device. The difference of materials has been addressed through demonstration of biocompatibility in accordance with the ISO 10993 standards.

Comment #2, Syringe Volume and Dead Space Specification: The proposed device features a lower syringe volume in comparison to the predicate device. In addition to the lower total syringe volume, a lower dead space volume is achieved. These differences have been assessed through syringe performance testing conducted in accordance with ISO 7886-1 that demonstrates that the proposed device meets all necessary requirements. The differences in syringe volume and dead volume specifications do not raise new questions of safety and effectiveness when compared to the predicate device.

8. Non-clinical testing

8.1 Performance testing

The proposed device was tested in compliance with ISO 7886-1:2017 for evaluating the overall non-clinical performance. In addition, Luer connector testing in compliance with ISO 80369-7:2016 and ISO 80369-20:2015 was conducted for evaluating the performance of connection to hypodermic needle. The performance and design testing results met the standards' requirements to demonstrate the device's substantial equivalence. Testing was conducted on unconditioned samples as well as samples aged according to ASTM F1980 and samples transit conditioned according to ASTM D4169-22.

8.2 Biocompatibility testing

The proposed device was tested in compliance with FDA Guidance Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process", as an Externally Communicating Device, Blood Path Indirect, Limited Contact (< 24 h).

The following endpoints were tested in accordance with ISO 10993-1: cytotoxicity, irritation, sensitization, acute systemic toxicity, hemocompatibility, and pyrogenicity. Particulate matter was tested in compliance with USP <788> and bacterial endotoxins were tested in compliance with USP <85> and met the acceptance criteria.

8.3 Sterility, packaging, and shelf-life

The sterilization process of the proposed device has been validated in compliance with ISO 11135. The EO and ECH residuals do not exceed the limits according to ISO 10993-7. The shelf-life verification was conducted under accelerated aging conditions in compliance with ASTM F1980-16 to demonstrate device functionality through the claimed shelf-life of 3 years.

Page 9

Page 5 of 5

Performance testing, package integrity testing, sterility testing, and particulate testing were conducted on samples accelerated aged to 3 years. Real time testing will be conducted to confirm device functionality through a shelf-life of 3 years.

Package integrity testing under simulated shipping conditions was conducted to satisfy the requirements in ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems. Simulated shipping conditions consisted of Distribution Cycle 13, Assurance Level 2. All packaging was deemed acceptable for protection of product and sterility maintenance.

Sterile barrier testing was conducted in compliance with the following FDA recognized consensus standards.

  • ASTM F88/F88-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
  • USP <71> Sterility Tests

A shelf life of 3 years is validated in accordance with ISO 11607-1 and ISO 11607-2.

9. Clinical testing

Not applicable for this submission.

10. Conclusion

The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use and technological characteristics.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).