K Number
K250116

Validate with FDA (Live)

Date Cleared
2025-04-24

(98 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.

Device Description

Arm blood pressure monitor, models(models ARM-30H, ARM-30J, ARM-30K and AOJ-90B), are designed as a battery driven automatic no-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa. All the models included in this submission follow the similar software, same measurement principle and same specifications. All the models can be used with one cuff size 2242 cm (8.616.5 inches). AOJ90B has extra 4G network except for other models.

AI/ML Overview

Here's an analysis of the provided FDA 510(k) clearance letter to extract information on acceptance criteria and the study proving device performance:

Summary of Device and Context:

The submission K250116 concerns several models of Arm Blood Pressure Monitors (ARM-30H, ARM-30J, ARM-30K, ARM-90B) manufactured by Shenzhen AOJ Medical Technology Co., Ltd. These devices are intended for non-invasive oscillometric measurement of systolic and diastolic blood pressure and pulse rate in adults, for both medical facilities and home use. The predicate device is the Arm Blood Pressure Monitor, model AOJ-30B (K222125).


1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Predicate)Reported Device Performance (Proposed Device)Study Type Supporting Performance
Blood Pressure Accuracy± 3 mmHg± 3 mmHgClinical
Heart Rate Accuracy± 5%± 5%Clinical
BiocompatibilityCytotoxicity, Sensitization, Irritation (ISO 10993-1, 10993-5, 10993-10)Cytotoxicity, Sensitization, Irritation (ISO 10993-1, 10993-5, 10993-10, 10993-23)Non-clinical (Bench)
Electrical SafetyIEC 60601-1, IEC 60601-1-11, ISO 80601-2-30IEC 60601-1, IEC 60601-1-11, ISO 80601-2-30Non-clinical (Bench)
EMC ComplianceIEC 60601-1-2IEC 60601-1-2Non-clinical (Bench)
Wireless CoexistenceNot explicitly stated for predicate in table, but 4G for ARM-90B.ANSI C63.27:2017, AAMI TIR69:2017 risk management, FDA Guidance for RF Wireless Technology in Medical Devices.Non-clinical (Bench)
Clinical PerformanceMean error and standard deviation of differences for systolic/diastolic pressure not over limits of ISO 81060-2.Mean error and standard deviation of differences for systolic/diastolic pressure not over limits of ISO 81060-2.Clinical

2. Sample Size and Data Provenance for Test Set (Clinical Studies)

  • Clinical Study for ARM-90B:

    • Sample Size: 100 adult subjects (61 females, 39 males)
    • Age Range: 19 to 77 years
    • Data Provenance: Not explicitly stated (e.g., country of origin). Since Shenzhen AOJ Medical Technology Co., Ltd. is based in China, it's highly probable the study was conducted there, but this is not confirmed in the provided text. The study method was prospective, as subjects were enrolled for the purpose of the study.
  • Clinical Study for ARM-30H:

    • Sample Size: 100 adult subjects (54 females, 46 males)
    • Age Range: 18 to 80 years
    • Data Provenance: Not explicitly stated. Likely prospective, similar to ARM-90B.

3. Number of Experts and Qualifications for Ground Truth

The document does not provide information on the number of experts used to establish ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience"). For blood pressure monitor clinical studies, ground truth is typically established by trained technicians or clinicians using validated reference devices, but the specific details are not present.


4. Adjudication Method for Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1). For blood pressure studies following ISO 81060-2, the ground truth measurements are usually taken by trained observers using mercury sphygmomanometers or other validated reference methods. The standard itself outlines the procedures for taking these measurements, which inherently involves multiple measurements and comparisons, but not necessarily a formal adjudication process in the sense of resolving conflicting interpretations of images or complex diagnostic results. The "Same Arm Sequential Method" was chosen for both studies, indicating a direct comparison of the device's readings against reference readings on the same arm of the subject.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done.
This type of study is typically relevant for interpretative devices (e.g., AI in radiology) where human readers are making diagnoses. For a blood pressure monitor, the device provides a direct measurement, not an interpretation that a human would then refine. The clinical studies directly evaluate the device's accuracy against a recognized reference standard.


6. Standalone (Algorithm Only) Performance Study

Yes, a standalone performance study was done for the device. The clinical studies performed according to ISO 81060-2 directly assess the performance of the device itself (the algorithm within the device) against a reference standard without human intervention in interpreting the device's output. The device is the algorithm in this context, automatically performing inflation, deflation, and measurement.


7. Type of Ground Truth Used

The ground truth used for the clinical studies was established by comparison with a reference method defined by ISO 81060-2. This standard outlines precise procedures for obtaining reference blood pressure measurements (typically using a mercury sphygmomanometer or other validated reference devices operated by trained observers) against which the automated device's readings are compared. The criteria state "mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2," which implies the ground truth is based on reference measurements adhering to this international standard.


8. Sample Size for the Training Set

The document does not provide information on the sample size for the training set.
510(k) submissions typically focus on the performance of the final device, not the details of its development or training data for its internal algorithms, unless the device explicitly uses machine learning algorithms that are updated or adapt. For a standard oscillometric blood pressure monitor, the algorithm is generally fixed and designed based on established physiological principles and extensive validation, rather than being "trained" on a dataset in the modern sense of machine learning.


9. How the Ground Truth for the Training Set Was Established

As with the training set sample size, the document does not provide information on how the ground truth for any potential training set was established. Given the nature of a blood pressure monitor, the "training" (if it occurs in a machine learning context) would likely involve extensive data collection with reference ground truth from validated sphygmomanometers, similar to the clinical validation studies, but specifically for internal algorithm development and refinement. However, this is speculative as the document does not detail the development process.

FDA 510(k) Clearance Letter - K250116

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

April 24, 2025

Shenzhen AOJ Medical Technology Co., Ltd
℅ Jie Yang
Consultant
Chonconn Consulting Co., Ltd.
Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District
Shenzhen, Guangdong 518067
China

Re: K250116
Trade/Device Name: Arm Blood Pressure Monitor (ARM-30H); Arm Blood Pressure Monitor (ARM30J); Arm Blood Pressure Monitor (ARM-30K); Arm Blood Pressure Monitor (ARM-90B)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: January 16, 2025
Received: March 21, 2025

Dear Jie Yang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K250116 - Jie Yang Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

K250116 - Jie Yang Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K250116

Device Name
Arm Blood Pressure Monitor (ARM-30H);Arm Blood Pressure Monitor (ARM-30J);
Arm Blood Pressure Monitor (ARM-30K);Arm Blood Pressure Monitor (ARM-90B)

Indications for Use (Describe)
The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: 2025/01/17

1. Submission sponsor

Name: Shenzhen AOJ Medical Technology Co., Ltd.
Address: Room 301&4F, Block A, Building A, Jingfa Intelligent Manufacturing Park, Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an District, 518126, Shenzhen, China
Contact person: Jack Wang
Title: Deputy Chief
TEL: 86 755-27786026

2. Submission correspondent

Name: Chonconn Consulting Co., Ltd.
Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China
Contact person: Yang Jie
E-mail: yangjie@chonconn.com
Tel: +86-755 33941160

3. Subject Device Information

Trade/Device NameArm Blood Pressure Monitor
Modelmodels ARM-30H, ARM-30J, ARM-30K and AOJ-90B
Common NameAutomatic Blood Pressure Monitor
Regulatory ClassClass II
Product CodeDXN
Submission typeTraditional 510(K)

4. Predicate Device

Manufacturer: Shenzhen AOJ Medical Technology Co., Ltd.
Device name: Arm Blood Pressure Monitor, model AOJ-30B
510(K) Number: K222125

5. Device Description

Arm blood pressure monitor, models(models ARM-30H, ARM-30J, ARM-30K and AOJ-90B), are designed as a battery driven automatic no-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa.

K250116

Page 6

All the models included in this submission follow the similar software, same measurement principle and same specifications. All the models can be used with one cuff size 2242 cm (8.616.5 inches). AOJ90B has extra 4G network except for other models.

6. Intended use & Indication for use

The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.

7. Comparison to the Predicate Device

ITEMProposed Device ARM-30H, ARM-30J, ARM30K and AOJ-90B/K250116Predicate Device AOJ-30B/K222125Comparison Result
ManufacturerShenzhen AOJ Medical Technology Co., Ltd.Shenzhen AOJ Medical Technology Co., Ltd.Same
Intended Use/Indications for UseThe Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.Same
ContraindicationsNot knownNot KnownSame
Application scenarioMedical Facilities and Home UseMedical Facilities and Home UseSame

Operational Specifications

ITEMProposed DevicePredicate DeviceComparison Result
PrincipleOscillometricOscillometricSame
Measurement ItemSYS, DYS, Pulse RateSYS, DYS, Pulse RateSame
Patient populationAdultAdultSame
Measurement siteUpper armUpper armSame
Blood pressure measurement range30-255 mmHg30 - 255 mmHgSame

K250116

Page 7

Accuracy± 3 mmHg± 3 mmHgSame
Heart rate measurement range40-199 bpm40-199 bpmSame
Accuracy± 5%± 5%Same
Cuff size22 - 42 cm22 - 42 cmSame
DisplayBlood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record NumberBlood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record NumberSame
Auto shutdownYESYESSame
Operating environmentTemperature: 5°C~40°C Humidity: 15%-90% RH, Atmospheric pressure: 70 kPa -106 kPaTemperature: 5°C~40°C Humidity: 15%-90% RH, Atmospheric pressure: 70 kPa -106 kPaSame
Storage environmentAmbient Temperature: -20°C to 55°C Relative Humidity: 10-93% RH, Atmospheric pressure: 70 kPa -106 kPaAmbient Temperature: -20°C to 55°C Relative Humidity: 10-93% RH, Atmospheric pressure: 70 kPa -106 kPaSame
Battery typeLithium-ion battery, d.c.3.7V6Vdc (4*1.5VAAA batteries)Similar
WeightARM-90B: About 263g ARM-30J: About 281g ARM-30K: About 280g ARM-30H: About 195g(262 ±5)g without batterySimilar
DimensionsARM90B:130mm×100mm×60mm ARM-30J: 130mm×100mm×60mm ARM-30K: 129mm×112mm×58.5mm ARM-30H: 12 mm×89.8mm×30.8mm127mm×93mm×73mm
Patient ContactingSurfacecontacting, Less than 24 hSurfacecontacting, Less than 24 hSame

K250116

Page 8

Biocompatibility evaluationCytotoxicity, skin sensitization and irritationCytotoxicity, skin sensitization and irritationSame
Electrical safetyIEC 60601-1 IEC 60601-1-11 ISO 80601-2-30IEC 60601-1 IEC 60601-1-11 ISO 80601-2-30Same
EMCIEC 60601-1-2IEC 60601-1-2Same
BiocompatibilityISO 10993-1 ISO 10993-5 ISO 10993-10 ISO 10993-23ISO 10993-1 ISO 10993-5 ISO 10993-10Equivalent
Data transmissionARM-90B:4G ARM-30H, ARM-30J, ARM30K:N/ABluetoothSimilar

8. Non-clinical Data

Biocompatibility testing

The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

  • Cytotoxicity
  • Sensitization
  • Irritation

Bench testing

The device has been tested according to the following standards:

  • IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance
  • IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 80601-2-30: Medical electrical equipment – Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers.
  • IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • FDA Guidance for Non-Automated Sphygmomanometer.

K250116

Page 9

Wireless testing:

  • ANSI C63.27: 2017: American National Stand for Evaluation of Wireless Coexistence.
  • AAMI TIR69: 2017 Risk management of radio-frequency wireless coexistence for medical devices and systems.
  • Radio Frequency Wireless Technology in Medical Devices: Guidance for Industry and Food and Drug Administration Staff (August 14, 2013)

9. Clinical data

Two studies were performed on ARM-30H and ARM-90B separately according to ISO 81060-2 Third edition 2018-11 [Including AMD1:2020] Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type [Including: Amendment 1 (2020)]. The clinical validation data on ARM-90B can cover ARM-30K and ARM-30J. The Same Arm Sequential Method was chosen for both studies.

ARM-90B study included 100 adult subjects (61 females, 39 males) with an age range of 19 to 77 years.

ARM-30H study included 100 adult subjects (54 females, 46 males) with an age range of 18 to 80 years

All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2 Third edition 2018-11 [Including AMD1:2020]. No adverse effect and/or complication is found in this study.

10. Conclusion

It is concluded form the non-clinical and clinical tests that demonstrate that the subject devices are as safe, as effective, and performs as well as the legally marketed predicate device identified above.

K250116

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).