(212 days)
The eCO2 3D Laser System with fractional handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; rhytides, furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia.
The eCO2 3D Laser System using the non-fractional handpieces (F100, F50, Zoom, and 500 micron tip) is also indicated for use in skin resurfacing and surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, and podiatry.
Dermatology & Plastic Surgery:
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- Laser skin resurfacing
- Treatment of wrinkles, rhytids and furrows
- Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the reduction, removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi, xanthelasma palpebrarum, syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
Dermatology, Plastic Surgery & General Surgery: Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for hair transplantation, treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
Podiatry:
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares.
The eCO2 3D Laser System consists of a self-contained console, with a 10.6μm wavelength delivered through an articulated arm. The emitted laser beam is then irradiated to the treatment area through a handpiece and a tip connected to the handpiece.
The eCO2 3D is used by a trained physician for procedures requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue.
The eCO2 3D Laser System utilizes an articulated arm to generate a laser beam with a wavelength of 10.6μm by applying patented chaos scanning technology. It radiates a micro-laser beam with an aiming beam of 520nm to the surgical area. Micro laser beams are radiated onto the treatment area, generating MAC (Micro Ablative Column), which improves the skin condition as it recovers. By changing the handpiece according to various application areas, the user can change the beam spot size to optimize the treatment effect. The physician can optimize the effect for different applications by controlling the power of the laser pulse and using a different handpiece tip.
The system console is the heart of the eCO2 3D Laser System and contains the Touch LCD, handpieces, system control module, and power supply module. The main console also includes a key switch used to turn the power on and off, an emergency stop push button that quickly de-energizes the system in emergency situations, and the Touch LCD. There are 4 casters in the console base that can be used when moving the system.
This FDA 510(k) clearance letter pertains to the eCO2 3D Laser System, an ablative laser, not an AI/Software as a Medical Device (SaMD). Therefore, the typical acceptance criteria and study designs associated with AI/SaMDs (such as sensitivity, specificity, MRMC studies, ground truth establishment, etc.) are not applicable to this document.
The document describes the device's technical specifications and compares them to predicate devices to demonstrate substantial equivalence, as is standard for a 510(k) submission for a physical medical device. It explicitly states: "No clinical tests were performed for this submission." This means there is no data provided within this document regarding device performance against typical AI/SaMD acceptance criteria.
However, I can extract the information that is relevant to the device's clearance and structure it to address your request as best as possible within the limitations of the provided text.
Analysis of the Provided Document for Acceptance Criteria and Study Proof:
The eCO2 3D Laser System is a physical medical device (laser system) for dermatological and surgical procedures. The 510(k) clearance process for such a device primarily relies on demonstrating substantial equivalence to predicate devices through technical characteristics, safety standards compliance, and intended use alignment, rather than performance metrics on a 'test set' as would be seen for an AI diagnostic algorithm.
Therefore, many of the requested categories (related to AI performance metrics, test sets, ground truth, and human reader studies) are explicitly not present in this document because they are not relevant to the 510(k) clearance of this type of device.
Here's a breakdown based on the provided text, indicating where information is present and where it's not applicable:
1. Table of Acceptance Criteria and Reported Device Performance
For this device, "acceptance criteria" are primarily related to safety, electrical performance, laser specifications, and functional equivalence to predicate devices. There are no performance metrics like sensitivity/specificity for a diagnostic AI.
| Acceptance Criteria (Implied from technical comparison and standards) | Reported Device Performance (from the table in 807.92(a)(6)) |
|---|---|
| Indications for Use Alignment | "See Indications for Use Above" (matches predicate and reference devices, indicating equivalent scope of use) |
| Laser Wavelength (10.6µm) | 10.6µm (=10,600 nm) (Matches predicate and reference devices) |
| Laser Medium (CO2) | CO2 (Matches predicate and reference devices) |
| Laser Transfer Method | Articulated arm with Handpiece (Matches predicate and reference devices) |
| Aiming Beam Wavelength (within acceptable range) | 520nm (Green) +/- 5nm (predicate uses 637nm, reference uses 630-650nm. This is a technical difference but deemed acceptable for substantial equivalence). |
| CO2 RF Module Maximum Power (within range of predicates) | 40Watt at Continuous Wave (Max. 30W with surgical handpiece, Max. 40W with Fractional handpiece and Bella V handpiece) (Predicate eCO2 Plus is 30W, Reference Finexel is 30/40W. The 40W capability is a new maximum for the eCO2 3D fractional handpiece, but aligns with the range of the reference device). |
| Protection against shock (CLASS I/ B-type) | CLASS I/ B-type (Matches predicate and reference devices) |
| Laser Class (CLASS IV) | CLASS IV (Matches predicate and reference devices) |
| User Interface (Touch LCD Display) | 10.1inch Touch LCD Display (Matches predicate and reference devices – "Touch LCD Display") |
| Dimensions/Weight (within acceptable manufacturing variation) | Dimensions: 391mm(W) x 581mm(L) x1931mm(H) / Weight: 45kg (These vary slightly from predicates but represent the device's physical specifications, not a performance metric). |
| Device Cooling Method (Air cooling) | Air cooling (Matches predicate and reference devices) |
| Electrical Rating (compliant) | Single Phase AC100~230V, 50Hz/60Hz (FUSE 250V/6.3A or 125V/6.3A), Power consumption: 550VA (Comparable to predicate and reference electrical specifications). |
| Compliance with recognized safety standards | Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1, EN ISO 14971, ISO/TR 24971, IEC 62366-1, IEC 62304, ISO 10993 series (biocompatibility), ISO 15223-1, ISO 20471, ISO 15883 series (washer disinfectors), ISO 17664-1, ISO 11607 series (packaging). (These are "acceptance criteria" and the submission implicitly confirms compliance). |
Since this is not an AI/SaMD, the following sections are Not Applicable (N/A) based on the provided document:
- Sample size used for the test set and the data provenance: N/A (No clinical test set described for performance evaluation; clearance based on substantial equivalence to predicate device specifications and non-clinical testing for safety/electrical performance).
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (No external 'ground truth' establishment for diagnostic performance).
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: N/A (No adjudication required for this type of device clearance).
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A (Not an AI-assisted device).
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A (Not an AI algorithm).
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): N/A (No diagnostic ground truth needed).
- The sample size for the training set: N/A (No AI training set).
- How the ground truth for the training set was established: N/A (No AI training set).
Conclusion from the Document:
The acceptance for the eCO2 3D Laser System is based on demonstrating substantial equivalence to its predicate device (Lutronic eCO2 Plus, K100610) and a reference device (SNJ Finexel, K213557). This demonstration relies on comparing technical specifications and compliance with relevant national and international safety and performance standards (as listed in 807.92(b)(1)). The document explicitly states: "No clinical tests were performed for this submission." Therefore, there is no performance study data in the context of diagnostic accuracy, human reader improvement, or algorithm-specific metrics available in this 510(k) clearance letter.
FDA 510(k) Clearance Letter - eCO2 3D Laser System
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.08.00
Silver Spring, MD 20993
www.fda.gov
July 31, 2025
Lutronic Corporation
℅ Sean Reynolds
Senior Manager, Global Regulatory Affairs
Cynosure Inc
3/5 Carlisle Rd
Westford, Massachusetts 01886
Re: K244060
Trade/Device Name: eCO2 3D
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: ONG, GEX
Dated: July 2, 2025
Received: July 2, 2025
Dear Sean Reynolds:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K244060 - Sean Reynolds Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K244060 - Sean Reynolds Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by TANISHA L. HITHE -S
Date: 2025.07.31 23:56:18 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (6/20) Page 1 of 2
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known): K244060
Device Name: eCO2 3D
Indications for Use (Describe)
The eCO2 3D Laser System with fractional handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; rhytides, furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia.
The eCO2 3D Laser System using the non-fractional handpieces (F100, F50, Zoom, and 500 micron tip) is also indicated for use in skin resurfacing and surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, and podiatry.
Dermatology & Plastic Surgery:
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- Laser skin resurfacing
- Treatment of wrinkles, rhytids and furrows
- Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the reduction, removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi, xanthelasma palpebrarum, syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
Dermatology, Plastic Surgery & General Surgery: Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for hair transplantation, treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
Podiatry:
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
Page 5
FORM FDA 3881 (6/20) Page 2 of 2
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 6
510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
807.92(a)(1) - Submitter Information
| Field | Information |
|---|---|
| Name | Lutronic Corporation |
| Address | Lutronic Center219, Sowon-RoDeogyang-Gu, Goyang-Si, KR 410220 |
| Phone number | 888-588-7644 |
| Fax number | N/A |
| Establishment Registration Number | 3004483538 |
| Name of contact person | Sean ReynoldsSr. Manager, Regulatory Affairs3/5 Carlisle RoadWestford, MA, 01886 |
| Date prepared | 29 July 2025 |
807.92(a)(2) - Name of device
| Field | Information |
|---|---|
| Trade or proprietary name | eCO2 3D |
| Common or usual name | Surgical Laser |
| Classification name | Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology |
| Classification panel | General and Plastic Surgery |
| Regulation | 21 CFR 878.4810 |
| Product Code(s) | GEX, ONG |
807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed
| Manufacturer | Device | 510(k) Number | Date | Type |
|---|---|---|---|---|
| Lutronic Corporation | eCO2 Plus | K100610 | (01/20/2011) | Predicate |
| SNJ Co. Ltd | Finexel | K213557 | (04/12/2022) | Reference |
807.92(a)(4) - Device description
The eCO2 3D Laser System consists of a self-contained console, with a 10.6μm wavelength delivered through an articulated arm. The emitted laser beam is then irradiated to the treatment area through a handpiece and a tip connected to the handpiece.
The eCO2 3D is used by a trained physician for procedures requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue.
Page 7
The eCO2 3D Laser System utilizes an articulated arm to generate a laser beam with a wavelength of 10.6μm by applying patented chaos scanning technology. It radiates a micro-laser beam with an aiming beam of 520nm to the surgical area. Micro laser beams are radiated onto the treatment area, generating MAC (Micro Ablative Column), which improves the skin condition as it recovers. By changing the handpiece according to various application areas, the user can change the beam spot size to optimize the treatment effect. The physician can optimize the effect for different applications by controlling the power of the laser pulse and using a different handpiece tip.
The system console is the heart of the eCO2 3D Laser System and contains the Touch LCD, handpieces, system control module, and power supply module. The main console also includes a key switch used to turn the power on and off, an emergency stop push button that quickly de-energizes the system in emergency situations, and the Touch LCD. There are 4 casters in the console base that can be used when moving the system.
807.92(a)(5) Intended use of the device
Indications for use: The eCO2 3D Laser System with fractional handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; rhytides, furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia.
The eCO2 3D Laser System using the non-fractional handpieces (F100, F50, Zoom and 500 micron tip) is also indicated for use in skin resurfacing and surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, and podiatry.
Dermatology & Plastic Surgery:
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- Laser skin resurfacing
- Treatment of wrinkles, rhytids and furrows
- Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the reduction, removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi, xanthelasma palpebrarum, syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
Dermatology, Plastic Surgery & General Surgery: Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for hair
Page 8
transplantation, treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
Podiatry:
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares.
807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate
| Characteristic | eCO2 3D | eCO2 Plus K100610 | Finexel CO2 Laser System K213557 |
|---|---|---|---|
| Indications for Use | See Indications for Use Above | See Indications for Use in Comparison Table | See Indications for Use in Comparison Table |
| Laser wavelength | 10.6µm (=10,600 nm) | 10.6µm (=10,600 nm) | 10.6µm (=10,600 nm) |
| Medium of transmission | CO2 | CO2 | CO2 |
| Laser transfer method | Articulated arm with Handpiece | Articulated arm and scanner handpiece | Articulated arm and scanner handpiece |
| Aiming Beam | 520nm (Green) +/- 5nm | 637nm | 630nm-650nm |
| CO2 RF Module Maximum Power | 40Watt at Continuous Wave (Max. 30W with surgical handpiece, Max. 40W with Fractional handpiece and Bella V handpiece) | 30Watt at Continuous Wave | 30/40W |
| Type and Degree of protection against shock | CLASS I/ B-type | CLASS I/ B-type | CLASS I/ B-type |
| Laser class | CLASS IV | CLASS IV | CLASS IV |
| User Interface | 10.1inch Touch LCD Display | Touch LCD Display | Touch LCD Display |
| Dimensions | 391mm(W) x 581mm(L) x1931mm(H) | 360mm(W) x 450mm(L) x1170mm(H) | 350mm(W) x 400mm(L) x 1000mm (H) |
| Weight (excluding the arm) | 45kg | 48kg | 50kg |
| Device Cooling method | Air cooling | Air Cooling | Air Cooling |
| Electrical rating | Single Phase AC100~230V, 50Hz/60Hz (FUSE 250V/6.3A or 125V/6.3A), Power consumption: 550VA | Single phase AC100~240VAC (FUSE 250VAC/6.3A or 125/12.6A), 50/60Hz, Power consumption:400VA | 100-120V~/220-230V~, 50/60Hz |
807.92(b)(1) NON CLINICAL TESTS SUBMITTED
| Standard | Description | Version |
|---|---|---|
| IEC60601-1 | Medical electrical equipment - Part 1: General requirements for safety | 2012+AMD2: 2020 |
Page 9
| Standard | Description | Version |
|---|---|---|
| IEC 60601-1-2 | Medical electrical equipment - Part1-2: General requirements for safety –Collateral standard: Electromagnetic compatibility - Requirements and Tests | 2014/AMD1:2020 |
| IEC 60601-1-6: | Medical electrical equipment - Part 1-6 General requirements for basic safety and essential performance – Collateral Standard: Usability | 2010+AMD1:2013+AMD2:2020 |
| IEC 60601-2-22 | Medical electrical equipment - Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment | 2019 |
| IEC 60825-1 | Safety of laser products – Part 1: Equipment classification and requirements | 2014 |
| EN ISO 14971 | Medical devices -Application of risk management to medical devices | 2019 |
| ISO/TR 24971 | Medical devices-Guidance on the application of ISO 14971 (ISO/TR 24971:2020) | 2020 |
| IEC 62366-1 | Medical devices -Application of usability engineering to medical devices | 2015+AMD1: 2020 |
| IEC 62304 | Medical device software - Software life cycle processes | 2006/ A1:2015 |
| ISO 10993-1 | Biological evaluation of medical devices Part 1: Evaluation and testing | 2018 |
| ISO 10993-5 | Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity | 2009 |
| ISO 10993-10 | Biological evaluation of medical devices -- Part 10: Tests for skin sensitization | 2021 |
| ISO 10993-17 | Biological evaluation of medical devices - Part 17: Toxicological risk assessment of medical device constituents | 2023 |
| ISO 10993-18 | Biological evaluation of medical devices 4 Part 18: Chemical characterization of medical device materials within a risk management process 4 Amendment 1: Determination of the uncertainty factor | 2020 |
| ISO 10993-23 | Biological evaluation of medical devices - Part 23: Tests for irritation | 2021 |
| ISO 15223-1 | Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements | 2021 |
| ISO 20471 | Information Supplied by the manufacturer with medical device | 2021 |
| ISO 15883-5 | Washer disinfectors Part 5: Performance requirements and test method criteria for demonstrating cleaning efficacy | 2021 |
| ISO 15883-1 | Washer-disinfectors –Part 1: General requirements terms and definitions and tests | 2006/Amd 1:2014 |
| ISO 15883-4 | Washer disinfectors Part 4: Requirements and tests for washer disinfectors employing chemical disinfection for thermolabile endoscopes | 2018 |
| ISO 17664 1 | Processing of health care products Information to be provided by the medical device manufacturer for the processing of medical devices Part 1: Critical and semi critical medical devices | 2021 |
Page 10
| Standard | Description | Version |
|---|---|---|
| ISO 11607-1 | Packaging for terminally sterilized medical devices – Part1: Requirements for materials, sterile barrier systems and packaging systems | 2019 |
| ISO 11607-2 | Packaging for terminally sterilized medical devices – Part2: Validation requirements for forming, sealing and assembly processes | 2019 |
807.92(b)(2) CLINICAL TESTS SUBMITTED
No clinical tests were performed for this submission.
807.92(b)(3) Conclusion
Based on the comparison of the technological characteristics, the specifications for the eCO2 3D are the same or a subset of the eCO2 Plus specifications. The additional tips allows the users more options in how the same energy is shaped during delivery and are similar to the reference device, Finexel. Also, the indication is the same as the predicate device, eCO2 Plus.
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.