K Number
K244054

Validate with FDA (Live)

Date Cleared
2025-02-28

(59 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 16CH Flex SPEEDER Medium is intended for use with Canon 1.5T MR systems to produce diagnostic images of general human anatomy that can be interpreted by a trained physician.

The 16ch Flex SPEEDER Large coil is intended for use with Canon 1.5T MR systems to produce diagnostic images of general human anatomy that can be interpreted by a trained physician.

Device Description

The 16ch Flex Speeder Medium and 16ch Flex Speeder Large coils are a receive-only, 16-channel phased array coil designed for magnetic resonance imaging (MRI) using the Canon 1.5T MR systems. The 16ch Flex Speeder Medium and 16ch Flex Speeder Large coils are intended to be used for general anatomies.

AI/ML Overview

This document describes the acceptance criteria and the study conducted to prove that the 16ch Flex SPEEDER Medium 1.5T and 16ch Flex SPEEDER Large 1.5T coils meet these criteria.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
Signal-to-Noise Ratio (SNR)Analyzed per NEMA MS-9 (using alternate method 2.5 from MS-6) and met pre-determined acceptance criteria.
Image UniformityAnalyzed per NEMA MS-9 (primary method from MS-6) and met pre-determined acceptance criteria.
Diagnostic Quality ImagesClinical images obtained from volunteer scanning of general human anatomy (upper and lower extremities, chest, abdomen, pelvis, head, neck, and spine) demonstrated diagnostic quality.
Electrical SafetyTested in accordance with IEC standards, data supports safety.
Electromagnetic Compatibility (EMC)Tested in accordance with IEC standards, data supports safety.
Surface HeatingMeasured temperature of the coil surface never exceeded the maximum limit of 41°C, tested in accordance with AAM/ANSI EN60601-1.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Clinical images were obtained from "volunteer scanning of general human anatomy." The exact number of volunteers is not specified.
  • Data Provenance: The document does not explicitly state the country of origin. The study appears to be prospective, as clinical images were "obtained for the 16CH Flex SPEDER Medium and Large 1.5T from upper and lower extremities, chest, abdomen, pelvis, head, neck and spine."

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • The document states that the diagnostic images are "interpreted by a trained physician." It does not specify the number of physicians, their specific qualifications (e.g., years of experience), or how ground truth was established by these physicians for the test set.

4. Adjudication Method for the Test Set

  • The document does not describe any specific adjudication method (e.g., 2+1, 3+1). It only mentions that images are "interpreted by a trained physician," implying a single interpretation per image for diagnostic quality assessment.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done. The study compares the new device to pre-determined acceptance criteria and implicitly to the predicate device in terms of performance (e.g., "performs as well as or better than the predicate device"), but not in terms of human reader improvement with AI assistance. This is an MRI coil, not an AI-driven device.

6. Standalone Performance Study

  • A standalone performance study was done for the device itself. The document describes non-clinical bench testing (SNR, uniformity, electrical safety, EMC, surface heating) and clinical imaging with volunteers to assess diagnostic image quality. This refers to the performance of the coil as a medical device without explicit human-in-the-loop performance measurement.

7. Type of Ground Truth Used

  • For technical performance (SNR, uniformity, electrical safety, EMC, surface heating): Ground truth was established against engineering specifications and industry standards (NEMA MS-9, IEC, AAM/ANSI EN60601-1).
  • For diagnostic image quality: The ground truth was established through visual assessment by a "trained physician" of "diagnostic quality images." This is essentially expert consensus/interpretation of the acquired images.

8. Sample Size for the Training Set

  • This information is not applicable as this is a physical medical device (MR coil) and not an AI/machine learning algorithm requiring a separate training set.

9. How the Ground Truth for the Training Set Was Established

  • This information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 28, 2025

Quality Electrodynamics Jovanna Boudreaux Senior Quality/ Regulatory Engineer 6655 Beta Drive Ste 100 Mayfield Village, Ohio 44122

Re: K244054

Trade/Device Name: 16ch Flex SPEEDER Medium 1.5T (Q7000246); 16ch Flex SPEEDER Large 1.5T (07000247) Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: December 30, 2024 Received: December 31, 2024

Dear Jovanna Boudreaux:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Signature

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K244054

Device Name

16ch Flex SPEEDER Medium 1.5T (Q7000246); 16ch Flex SPEEDER Large 1.5T (Q7000247)

Indications for Use (Describe)

The 16CH Flex SPEEDER Medium is intended for use with Canon 1.5T MR systems to produce diagnostic images of general human anatomy that can be interpreted by a trained physician.

The 16ch Flex SPEEDER Large coil is intended for use with Canon 1.5T MR systems to produce diagnostic images of general human anatomy that can be interpreted by a trained physician.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) #:K244054
510(k) Summary
Prepared on: 2025-02-27
21 CFR 807.92(a)(1)
Contact Details
Applicant NameQuality Electrodynamics
Applicant Address6655 Beta Drive STE 100 Mayfield Village OH 44122 United States
Applicant Contact Telephone4404842340
Applicant ContactMrs. JoVanna Boudreaux
Applicant Contact Emailjovanna.boudreaux@qualedyn.com
Device Name
21 CFR 807.92(a)(2)
Device Trade Name16ch Flex SPEEDER Medium 1.5T (Q7000246); 16ch Flex SPEEDER Large 1.5T (Q7000247)
Common NameMagnetic resonance diagnostic device
Classification NameCoil, Magnetic Resonance, Specialty
Regulation Number892.1000
Product Code(s)MOS
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)
K1213621.5T 16ch Flex SPEEDER Coil Medium/Large
Product Code
MOS
Device Description Summary
21 CFR 807.92(a)(4)
The 16ch Flex Speeder Medium and 16ch Flex Speeder Large coils are a receive-only, 16-channel phased array coil designed for magnetic resonance imaging (MRI) using the Canon 1.5T MR systems. The 16ch Flex Speeder Medium and 16ch Flex Speeder Large coils are intended to be used for general anatomies.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The 16CH Flex SPEEDER Medium is intended for use with Canon 1.5T MR systems to produce diagnostic images of general human anatomy that can be interpreted by a trained physician.
The 16ch Flex SPEEDER Large coil is intended for use with Canon 1.5T MR systems to produce diagnostic images of general human anatomy that can be interpreted by a trained physician.
Indications for Use Comparison
21 CFR 807.92(a)(5)
The Indications for Use Statements for the 16ch Flex SPEEDER Medium and Large 1.5T are not identical to that of the predicate devices (1.5T 16ch Flex SPEEDER Medium and Large); however, the differences do not affect the safety or effectiveness of the device relative to the predicate device. Both Indications for Use statements for the proposed 16ch Flex SPEEDER Medium and Large 1.5Tand predicate 1.5T 16ch Flex SPEEDER Medium and Large indicate that the device is intended to be used in conjunction with a MR system to produce images of human anatomy and that the images can be interpreted by a trained physician. The indications for use statements for the proposed 16ch Flex SPEEDER Medium and Large 1.5T and predicate devices differ only in that the proposed devices are indicated for use

{5}------------------------------------------------

in general human anatomy whereas the predicate devices are indicated for use in various specific anatomies that comprise general human anatomy.

Technological Comparison

21 CFR 807.92(a)(6)

At a high level, the proposed and predicate device are based on the following same technological elements:

  • · Receive-only phased array RF coil
  • · Active PIN diode switching blocking circuitry. Passive blocking circuitry.
  • · Materials used for impact protection and biocompatibility: Polycarbonate with a polyurethane coated nylon fabric cover

No technological differences exist between the proposed and predicate device.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The signal-to-noise ratio (SNR) and image uniformity of the 16ch Flex Speeder Coils were measured on a 1.5T Canon MR System, manufactured by Canon Medical System. The SNR and uniformity of the 16ch Flex Speeder Coils were analyzed per NEMA MS-9 (using alternate method 2.5 from MS-6) and uniformity was analyzed using NEMA MS-9 (primary method from MS-6) using pre-determined acceptance criteria.

In accordance with the FDA Guidance for the Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, clinical images from volunteer scanning of general human anatomy were obtained from a Canon 1.5T MR system. Clinical images were obtained for the 16CH Flex SPEDER Medium and Large 1.5T from upper and lower extremities, chest, abdomen, pelvis, head, neck and spine. These images were used to demonstrate that the 16ch Flex Speeder Coils produce diagnostic quality images of the intended anatomy. No adverse events were reported or recorded.

The electrical safety and electromagnetic compatibility data support the safety of the 16ch Flex Speeder Coils and the bench testing per the IEC standards and diagnostic quality sample clinical images demonstrates the performance and effectiveness of the device under the specified use conditions. This testing demonstrates that the 16ch Flex Speeder Coils performs as well as or better than the predicate device.

Surface heating was tested in accordance with AAM/ANSI EN60601-1. The device was tested plugged. The measured temperature of the surface of the coil never exceeded the maximum limit of 41C.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.