(58 days)
Disposable Blood Lancet is used for capillary blood sampling.
Disposable Blood Lancet is a single use, sterile medical device designed to be used for capillary blood sampling. The device comprises a stainless needle encapsulated with a plastic needle body and protective cap, the protective cap is twisted off to expose the needle for use. The device was sterilized by Radiation. The needle body and protective cap form a sterile barrier to maintain the needle sterile.
The provided document is a 510(k) summary for the SteriLance Medical (Suzhou) Inc. Disposable Blood Lancet. It focuses on demonstrating substantial equivalence to a predicate device and does not contain information about acceptance criteria and a study proving a device meets them in the context of an AI/ML medical device.
The document discusses:
- Regulatory classification: Class II, product code QRK/QRL (Blood Lancets)
- Device description: Single-use, sterile device for capillary blood sampling.
- Comparison to predicate device: Highlights similarities in intended use, sterilization, gauges, components, and materials, with a noted difference in exposed needle length for one model.
- Performance data: Mentions biocompatibility testing (cytotoxicity, skin sensitization, intracutaneous reactivity, acute systemic toxicity) and non-clinical bench testing (appearance, dimensions, sharpness, binding strength, double needles, empty needle, reverse needle, and compatibility with lancing device).
Therefore, I cannot provide the requested information for acceptance criteria and a study proving a device meets them for an AI/ML perspective as this document describes a physical medical device (blood lancet) and not an AI/ML-driven device.
If you have a document pertaining to an AI/ML medical device, please provide it, and I will be happy to extract the relevant information.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 26, 2025
SteriLance Medical (Suzhou) Inc. Susan Sun Quality Manager No.168 PuTuoShan Road, New District Suzhou, Jiangsu 215253 China
Re: K244031
Trade/Device Name: Disposable Blood Lancet (Soft Pro); Disposable Blood Lancet (Softsure); Disposable Blood Lancet (Softsure Pro); Disposable Blood Lancet (Softsense) Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: QRK Dated: December 23, 2024 Received: December 30, 2024
Dear Susan Sun:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Figure/3 description: The image shows the name James H. Jang -S, along with the words "Digitally signed by". The date 2025.02.26 is also present, along with the time 21:52:11 -05'00'. The name James H. Jang -S appears twice in the image.
For
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
Device Name
Disposable Blood Lancet (Soft Pro);
Disposable Blood Lancet (Softsure);
Disposable Blood Lancet (Softsure Pro);
Disposable Blood Lancet (Softsense)
Indications for Use (Describe)
Disposable Blood Lancet is used for capillary blood sampling.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(K) Summary - K244031
Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: 2024/12/11
1. Submission sponsor
Name: SteriLance Medical (Suzhou) Inc. Address: No.168 PuTuoShan Road, New District, 215153 Suzhou, Jiangsu, P. R. China Contact person: Susan Sun Title: Quality Manager E-mail: registration@sterilance.com Tel: 86-0512-65799308 Ext 8301
| Trade/Device Name | Disposable Blood Lancet |
|---|---|
| Model | Softsure, Softsure Pro, Softsense, Soft Pro |
| Common Name | Blood Lancet |
| Regulatory Class | Class II |
| Classification | 21CFR 878.4850QRK (Single use only blood lancet without an integral sharps injury prevention feature)QRL (Multiple use blood lancet for single patient use only) |
| Submission type | Traditional 510(K) |
2. Subject Device Information
3. Predicate Device
SteriLance Medical (Suzhou) Inc., Disposable Blood Lancet, under K221507.
4. Device Description
Disposable Blood Lancet is a single use, sterile medical device designed to be used for capillary blood sampling. The device comprises a stainless needle encapsulated with a plastic needle body and protective cap, the protective cap is twisted off to expose the needle for use.
The device was sterilized by Radiation. The needle body and protective cap form a sterile barrier to maintain the needle sterile.
Intended use & Indication for use 5.
Disposable Blood Lancet is used for capillary blood sampling.
Comparison to the Predicate Device 6.
{5}------------------------------------------------
| Features | Subject Device: Disposable BloodLancet | Predicate Device: Disposable BloodLancet (K221507) | Comparison |
|---|---|---|---|
| ProductCode | QRK; QRL | QRK; QRL | Same |
| RegulationNumber | 21 CFR § 878.4850 | 21 CFR § 878.4850 | Same |
| Classification | Class II | Class II | Same |
| Type of use | OTC | OTC | Same |
| Indications forUse | Disposable Blood Lancet is used forcapillary blood sampling. | Disposable Blood Lancet is used forcapillary blood collection | Similar |
| Reusedurability | Single use | Single use | Same |
| Sterilizationmethod andSAL | Sterilized by Radiation SAL=10-6 | Sterilized by Radiation SAL=10-6 | Same |
| Gauge | 21G, 23G, 26G, 28G, 30G, 32G, 33G | 21G, 23G, 26G, 28G, 30G, 32G, 33G | Same |
| Component | NeedleNeedle bodyProtective cap | NeedleNeedle bodyProtective cap | Same |
| ExposedNeedle Length | Soft Pro; Softsure; Softsure Pro3.20±0.60mmSoftsense: 2.50±0.60mm | 3.20±0.60mm | Different |
| PenetrationDepth | Depend on the Puncture depth ofCompatible lancing device. | Depend on the Puncture depth ofCompatible lancing device. | Same |
| Materials ofparts incontact withhuman body | Needle: stainless steelNeedle body and cap: Polyethylene | Needle: stainless steelNeedle body and cap: Polyethylene | Same |
Similar: The intended use is same. The subject device is used for draw capillary blood and regulated under 21 CFR 878.4850 . The difference does not raise any safety and effectiveness questions.
Different: The physical dimension difference does not raise any safety and effectiveness questions. The Exposed Needle Length were verified in the performance test report.
7. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility testing
The biocompatibility evaluation for the proposed device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1:
{6}------------------------------------------------
K244031
Evaluation and testing within a risk management process". The battery of testing included the following tests:
- Cytotoxicity
- Skin Sensitization
- Intracutaneous reactivity
- Acute systemic toxicity
Non-clinical data
The bench testing performed verifies that the performance of the subject devices are substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:
- Appearance and type, Cleanliness
- Basic Dimensions, Exposed length of product
- Cap site, Needle tip sharpness and Binding Strength
- Double needles, Empty needle and Reverse needle
- Compatibility test between Disposable blood lancet and Lancing Device
8. Conclusion
Performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.
§ 878.4850 Blood lancets.
(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.