K Number
K244010

Validate with FDA (Live)

Device Name
ExamVue Apex
Date Cleared
2025-02-24

(60 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ExamVue Apex flat panel x-ray detector system is indicated for use in general radiology including podiatry, orthopedic, and other specialties, and in mobile x-ray systems. The Exam Vue Apex flat panel x-ray detector system is not indicated for use in mammography.

Device Description

The ExamVue Apex flat panel x-ray detector system consists of a line of 3 different models of solid state x-ray detectors, of differing size and characteristics, combined with a single controlling software designed for use by radiologists and radiology technicians for the acquisition of digital x-ray images. The ExamVue Apex flat panel x-ray detector system captures digital images of anatomy through the conversion of x-rays to electronic signals, eliminating the need for film or chemical processing to create a hard copy image. ExamVue Apex flat panel x-ray detector system incorporates the ExamVue Duo software, which performs the processing, presentation and storage of the image in DICOM format. All models of the ExamVue Apex flat panel x-ray detector system use a Si TFTD for the collection of light generated by a CsI scintillator, for the purpose of creating a digital x-ray image.

The three available models are:

EVA 14W, with a 14x17in (35x43cm) wireless cassette sized panel
EVA 17W, with a 17x17in (43x43cm) wireless cassette sized panel
EVA 10W, with a 10x12 (24x30cm) wireless cassette sized panel

AI/ML Overview

The provided text describes the regulatory clearance of a digital X-ray detector system, the "ExamVue Apex," and its substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria or an analytical study proving the device meets those criteria, as one would typically find for an AI/ML medical device submission with defined performance metrics (e.g., sensitivity, specificity, AUC).

Instead, the submission focuses on demonstrating substantial equivalence through:

  • Bench Testing: Comparing engineering specifications like resolution, sensitivity, and dynamic range to the predicate device.
  • Software Validation: Ensuring the software adheres to relevant standards (IEC 62304) and performs expected functions.
  • Clinical Testing: An ABR-certified radiologist visually evaluating image quality as equivalent or better than the predicate device.

Therefore, many of the requested fields regarding a detailed statistical study (sample size, ground truth, expert adjudication, MRMC study, standalone performance) cannot be filled from the provided text because such a study, with quantitative acceptance criteria, does not appear to have been performed or reported in this 510(k) summary. The submission relies more on demonstrating equivalence through technical specifications and expert opinion on image quality rather than rigorous statistical performance criteria for an AI algorithm.

Here's a breakdown of what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

The provided text does not define explicit quantitative acceptance criteria for device performance in the typical AI/ML sense (e.g., a target sensitivity of X% or specificity of Y%). Instead, the "acceptance criteria" are implied by demonstrating "similar or greater imaging characteristics" compared to the predicate device and that the software "performs the same required basic functions."

Acceptance Criteria (Implied)Reported Device Performance
Bench Testing (Comparison to Predicate):
a. Resolution equivalent or greater"product performs with similar or greater imaging characteristics" (general statement). Specific comparison metrics for ExamVue Apex vs. Predicate: Pixel Pitch (99um vs 143/140/143um), DQE @ 0 lp/mm (73% @ 6 μGy / 70% @ 2 μGy vs 57%/60%/58%), MTF @ 1 lp/mm (68% vs 63%/68%/65%) - all indicate equal or improved performance for Apex.
b. Sensitivity equivalent or greater"product performs with similar or greater imaging characteristics" (general statement).
c. Dynamic range in image acquisition equivalent or greater"product performs with similar or greater imaging characteristics" (general statement).
d. Software performs the same basic functions"software performs the same required basic functions as the predicate device."
Software Validation:
a. Designed and developed according to IEC 62304"The software was designed and developed according to a software development process in compliance with IEC 62304."
b. Performs all functions of the predicate's software"tested to show that it performs all the functions of the software in the predicate device." The software "performs the same functions as the software for the predicate device with some additional features."
Clinical Testing:
a. Image quality equivalent or better than predicate device."The images were evaluated by an ABR certified radiologist who evaluated the image quality to be of equivalent or better to the predicate device."

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified. The clinical testing merely states "Clinical data was provided with the submission to demonstrate equivalence with the predicate device. This data includes images of all the relevant ROI." It doesn't quantify the number of images or patients.
  • Data Provenance: Not specified (country, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Number of Experts: "an ABR certified radiologist" (singular, implied to be one).
  • Qualifications: "ABR certified radiologist." No mention of years of experience.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Adjudication Method: Not specified. Since only one radiologist is mentioned, it's likely "none" in the sense of a consensus or adjudication process among multiple readers. The single ABR-certified radiologist provided the evaluation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study was not explicitly stated or described. This submission is for a general X-ray detector system, not specifically an AI-powered diagnostic algorithm designed to assist human readers. The "AI" mentioned is the software components related to image acquisition, processing, and management, not a diagnostic AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable in the context of an AI diagnostic algorithm. This device is a digital X-ray detector system. Its "performance" refers to image quality and functionality, not a diagnostic output from an algorithm that would then require standalone performance metrics (e.g., sensitivity/specificity for disease detection). The software handles image processing, presentation, and storage, not automated diagnosis.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth: The "ground truth" for the clinical testing was the visual evaluation of image quality by an ABR-certified radiologist. It's not a diagnostic ground truth (like pathology for cancer detection) but rather an assessment of whether the images produced by the new device are diagnostically acceptable and equivalent/superior to those from the predicate device.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable/not specified. This device is an X-ray detector and associated software for image acquisition and processing, not a deep learning AI model that requires a "training set" to learn features for interpretation. The software's development (as per IEC 62304) involves validation and verification, but not "training" in the machine learning sense.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable. (See #8).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

February 24, 2025

JPI Healthcare Co, Ltd % William Little Senior Product Manager JPI Healthcare Solutions, Inc 3555 Veterans Memorial Highway Unit D RONKONKOMA NY 11779

Re: K244010

Trade/Device Name: ExamVue Apex Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: October 2, 2024 Received: December 26, 2024

Dear William Little:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}------------------------------------------------

See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K244010

Device Name ExamVue Apex

Indications for Use (Describe)

The ExamVue Apex flat panel x-ray detector system is indicated for use in general radiology including podiatry, orthopedic, and other specialties, and in mobile x-ray systems. The Exam Vue Apex flat panel x-ray detector system is not indicated for use in mammography.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

ExamVue Apex Digital X-ray Detector

K244010

    1. Company and Correspondant Making the Submission:
      Name: JPI Healthcare Co., LTD Address: 608-ho, 28, Digital-ro 33-gil, Guro-gu, 08377 Seoul, Korea Telephone: +82-2-2108 - 2580 Fax: +82-2-2108-1180 Contact: Lee Jang Won Website: http://www.jpi.co.kr/
    1. Identification of Device
      Classification Name: Stationary x-ray system Common Name: Digital X-ray Acquisition System Trade/Proprietary Name: ExamVue Apex
    1. Predicate Device
      Manufacturer: JPI Healthcare, Inc Device: ClearVision ExamVue Flat Panel Detector 510(k) Number: K160143 Classification Name: Stationary x-ray system Common Name: Digital X-ray System Regulatory Number:21 CFR 892.1680 Regulatory Class: II Product Code: 90 MQB
    1. Product Classification Names and Citations
      Regulatory Number:21 CFR 892.1680 Regulatory Class: II 90 MQB Product Code:

{5}------------------------------------------------

Description: 5.

The ExamVue Apex flat panel x-ray detector system consists of a line of 3 different models of solid state x-ray detectors, of differing size and characteristics, combined with a single controlling software designed for use by radiologists and radiology technicians for the acquisition of digital x-ray images. The ExamVue Apex flat panel x-ray detector system captures digital images of anatomy through the conversion of x-rays to electronic signals, eliminating the need for film or chemical processing to create a hard copy image. ExamVue Apex flat panel x-ray detector system incorporates the ExamVue Duo software, which performs the processing, presentation and storage of the image in DICOM format. All models of the ExamVue Apex flat panel x-ray detector system use a Si TFTD for the collection of light generated by a CsI scintillator, for the purpose of creating a digital x-ray image.

The three available models are:

EVA 14W, with a 14x17in (35x43cm) wireless cassette sized panel EVA 17W, with a 17x17in (43x43cm) wireless cassette sized panel EVA 10W, with a 10x12 (24x30cm) wireless cassette sized panel

6. Indication for use

The ExamVue Apex flat panel x-ray detector system is indicated for use in general radiology, specialist radiology including podiatry, orthopedic, and other specialties, and in mobile x-ray systems. The ExamVue Apex flat panel x-ray detector system is not indicated for use in mammography.

7. Comparison with Predicate Device:

JPI Healthcare Co., Ltd, believes that the ExamVue Apex is substantially equivalent to the ClearVision ExamVue Flat Panel Detector registered by JPI Healthcare Co. Ltd.

The ExamVue Apex and the predicate device both

-Are a combination of a digital x-ray detector with an image acquisition software -Include digital x-ray detectors that are built upon the same fundamental technology (Amorphous Silicon collectors and Cesium lodide scintillator) to turn incoming x-rays into a digital image.

-Provide a user interface for the registration, acquisition, and evaluation of x-ray studies.

-Perform the functions of image transfer, image acquisition, image processing, and maintaining a patient database.

-Use the DICOM 3.0 standard for medical imaging

-Are intended for installation on Windows operating systems for use in a medical environment.

-Interface with and process images from multiple models of hardware.

{6}------------------------------------------------

ExamVue Apex and the predicate device share the same essential functions of image acquisition, transfer, and processing. The detector hardware differs in resolution and the substrate material for the electronics. The software on the ExamVue Apex is ExamVue Duo, previously registered with 510(k) number K213057 for use with flat panel detectors. The software represents a moderate level of concern, which is addressed in the software validation and verification and bench testing provided with the submission. The software performs the same functions as the software for the predicate device with some additional features, including expanded tool sets to simplify measurement and image viewing and management tools, that do not impact the core functions of the device. We believe these differences do not represent a substantial difference between the two devices, as the indications for use, core technologies, and user interface presents the same essential data and supports similar workflow as the predicate device.

8. Non-Clinical Test Summary

  • a) Electrical Safety, Electromagnetic Compatability and Performance The ExamVue Apex Digital X-ray Detectors comply with voluntary standards for electrical safety, electromagnetic compatibility and product design. Safety and quality testing and documentation have been performed and provided following IEC standards.
Standards NoStandards TitleVersionPublicationYear
ISO 10993Biological Evaluation ofMedical Devices20182018
IEC 60529Degrees of protectionprovided by enclosures (IPCode)IEC60529:1989+AMD1:1999+AMD2:2013CSV2019
IEC 62304Medical device software -Software life cycleprocessesIEC 62304Ed. 1.12015
IEC 62366Medical devices - Part 1:Application of usabilityengineering to medicaldevicesIEC 62366-1Ed. 1.1b:20202020
IEC 60068-2-27Environmental testing - Part2-27: Tests - Test Ea andguidance: ShockIEC60068-2-27Ed. 4.0b:2008▷2008
IEC 60601-1-2Medical electricalequipment - Part 1-2:General requirements forbasic safety and essentialperformance - CollateralStandard: Electromagneticdisturbances -Requirements and testsIEC60601-1-2 Ed.4.1 en:20202020
IEC 60601-1-3Medical electricalequipment - Part 1-3:General requirements forbasic safety and essentialperformance - CollateralStandard: Radiationprotection in diagnosticIEC60601-1-3:2008+AMD1:2013+AMD2:2021 CSV2021
IEC 60601-1-6Medical electricalequipment - Part 1-6:General requirements forbasic safety and essentialperformance - Collateralstandard: UsabilityIEC60601-1-6:2010+AMD1:2013+AMD2:2020 CSV2020
IEC 60601-1-8mendment 2 - Medicalelectrical equipment - Part1-8: General requirementsfor basic safety andessential performance -Collateral Standard:General requirements,tests and guidance foralarm systems in medicalelectrical equipment andmedical electrical systemsIEC60601-1-8:2006/AMD2:20202020
IEC 60601-2-54Medical electricalequipment - Part 2-54:Particular requirements forIEC60601-2-54:20222022
the basic safety andessential performance ofX-ray equipment forradiography andradioscopy
IEC 61000-3-2Electromagneticcompatibility (EMC) - Part3-2: Limits - Limits forharmonic current emissions(equipment input current≤16 A per phase)IEC61000-3-2:20182018
IEC 61000-3-3Electromagneticcompatibility (EMC) - Part3-3: Limits - Limitation ofvoltage changes, voltagefluctuations and flicker inpublic low-voltage supplysystems, for equipmentwith rated current ≤16 Aper phase and not subjectto conditional connectionIEC61000-3-3:2013+AMD1:2017+AMD2:2021 CSV2017
61000-4-2Electromagneticcompatibility (EMC) - Part4-2: Testing andmeasurement techniques -Electrostatic dischargeimmunity testIEC61000-4-2:20082008
IEC 61000-4-3Electromagneticcompatibility (EMC) - Part4-3 : Testing andmeasurement techniques -Radiated, radio-frequency,electromagnetic fieldimmunity testIEC61000-4-3:20202020
IEC 61000-4-4Electromagneticcompatibility (EMC) - Part4-4: Testing andmeasurement techniques -Electrical fasttransient/burst immunityIEC61000-4-4:20122012
test
IEC 61000-4-5Electromagneticcompatibility (EMC) - Part4-5: Testing andmeasurement techniques -Surge immunity testIEC61000-4-5:20142014
IEC 61000-4-6Electromagneticcompatibility (EMC) - Part4-6: Testing andmeasurement techniques -Immunity to conducteddisturbances, induced byradio-frequency fieldsIEC61000-4-6:20232023
IEC 61000-4-8Electromagneticcompatibility (EMC) - Part4-8: Testing andmeasurement techniques -Power frequency magneticfield immunity testIEC61000-4-8:20092009
IEC 61000-4-11Electromagneticcompatibility (EMC) - Part4-11: Testing andmeasurement techniques -Voltage dips, shortinterruptions and voltagevariations immunity testsfor equipment with inputcurrent up to 16 A perphaseIEC61000-4-11:20202020
IEC 61000-4-39Electromagneticcompatibility (EMC) - Part4-39: Testing andmeasurement techniques -Radiated fields in closeproximity - Immunity testIEC61000-4-39:20172017
IEC 60950-1Information technologyequipment - Safety - Part 1:General requirementsIEC 60950-1Ed. 2.2b:20132013
CONSOLIDATED EDITION
IEC 62133-2Secondary cells andbatteries containingalkaline or other non-acidelectrolytes - Safetyrequirements for portablesealed secondary cells,and for batteries made fromthem, for use in portableapplications - Part 2:Lithium systemsIEC62133-2:20172017
IEC 62304Medical device software -Software life cycleprocessesIEC 62304Ed. 1.1b:20152015
IEC 62366-1Medical devices - Part 1:Application of usabilityengineering to medicaldevicesIEC 62366-1Ed. 1.1b:20202020

{7}------------------------------------------------

2/24/2025 510(k) Summary K244010

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

  • b) Additional bench testing was performed detailing the direct comparison of functions between the ExamVue Apex and predicate device. These testing demonstrated that the product performs with similar or greater imaging characteristics
    • i) Resolution
    • ii) Sensitivity
    • iii) Dynamic range in image acquisition
    • iv) That the software performs the same required basic functions as the predicate device.
  • c) Software Validation

The ExamVue Apex Digital X-ray Detectors contain software of a MODERATE level of concern. The software was designed and developed according to a software development process in compliance with IEC 62304 and tested to show that it performs all the functions of the software in the predicate device. Software development and testing documentation has been provided in accordance with the FDA guidance "The Content of Premarket Submissions for Software Contained in Medical Devices."

  • d) Cybersecurity

{11}------------------------------------------------

Cybersecurity documentation and procedures have been provided according to the FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions."

  • e) Clinical Testing
    Clinical data was provided with the submission to demonstrate equivalence with the predicate device. This data includes images of all the relevant ROI. The images were evaluated by an ABR certified radiologist who evaluated the image quality to be of equivalent or better to the predicate device.

In conclusion, the clinical testing indicate adequate imaging performance of the ExamVue Apex Digital X-ray Detectors.

    1. Conclusions:
      In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification JPI Healthcare Co., Ltd. concludes that the ExamVue Apex is safe and effective and substantially equivalent to predicate devices as described herein.

A detailed comparison supporting this conclusion can be found in Exhibit 1, Substantial Equivalence Chart.

EXHIBIT 1

SUBSTANTIAL EQUIVALENCE CHART

Device NameExamVue ApexClearVision ExamVue FPD
Hardware ManufacturerThalesToshiba
Software ManufacturerJPI Healthcare Co, LtdJPI Healthcare Co, Ltd
FDA 510(k)K244010K160143
ModelsEVA 10W/ EVA 14W/ EVA17W1417T / 1417W / 1717
Intended UseThe ExamVue Apex flatpanel x-ray detector systemis indicated for use ingeneral radiology, specialistradiology including podiatry,orthopedic, and otherspecialties, and in mobileThe ClearVision ExamVueFlat Panel detector isindicated for use in generalradiology, specialistradiology including podiatry,orthopedic, and otherspecialties. and in mobile
x-ray systems.The ExamVue Apex flatpanel x-ray detector systemis not indicated for use inmammography.x-ray systems. TheClearVision ExamVue FlatPanel detector is notindicated for use inmammography.
ConfigurationThis submission is for adigital panel and software. Itis intended for use withunconnected or softwareintegrated x-ray generatorsand associated equipment.This submission is for thedigital panel and softwareonly, no generator or standprovided.
Detector TypeAmorphous SiliconAmorphous Silicon
Pixel Pitch (um)99143 / 140 / 143
A/D Conversion (bits)1616 / 14 / 14
Active Area222x285mm345x425mm423x423mm14x17 inch14x17 inch16.9x17.3 inch
Dimensions (mm)267.5(W) x 327.5(D) x15(H)383.5(W) x 459.5(D) x15(H)459.5(W) x 459.5(D) x15(H)384(W) x 460(D) x 15(H)384(W) x 460(D) x 15(H)512(W) x 495(D) x 43(H)
Weight (kg)1.6 / 2.6 / 33 / 3 / 9
DQE @ 0 lp/mm73 % @ 6 μGy / 70 % @ 2μGy57% / 60% / 58%
MTF @ 1 lp/mm68%63% / 68% / 65%
DICOM FunctionalityYesYes
ScintillatorCsICsI
InterfaceWifi (802.11ac/ax) or cableGigabit Ethernet or Wireless
Automatic ExposureDetectionThales Full-Field AEDYes
Generator ExposureTrigger or SensorOptionalYes
Power SourceAC LineAC Line
Acquisition and ControlSoftwareExamVue Duo (K213057)ExamVueDR (K142930)

{12}------------------------------------------------

Summary

  • The software used in ExamVue Apex is the direct successor of the ExamVueDR software used in the predicate device, including similar image acquisition,

{13}------------------------------------------------

processing, and other features. Both softwares are independently registered with 510(k) submissions.

  • -The new features for general medicine, chiropractic and podiatry applications in ExamVue Apex are additional features not included in the predicate device. However, they are similar to the image management and measurement functions already incorporated in the software for the predicate device and are included in the previously registered 510(k) for the included software.
  • In Conclusion, the intended use and environment of the device is the same as the predicate devices, with only minor differences in features that are not integral to the function of the device. The differences do not introduce a fundamentally new scientific technology.

The intended use and environment of the device is the same as the predicate devices, with only minor differences in features that are not integral to the function of the device. For this reason, we believe it is substantially equivalent to the predicate devices.

N/A