K Number
K243801
Device Name
FRONT-LINE Field Sterilizer (FL135)
Date Cleared
2025-03-07

(86 days)

Product Code
Regulation Number
880.6880
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Fort Defiance Industries FRONT-LINE Field Sterilizer (FL135) is an autoclave designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.
Device Description
The Fort Defiance Industries FRONT-LINE Field Sterilizer, model FL135, is an autoclave designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities. The device uses electrical resistive heaters to produce steam as the sterilizing agent. The device adheres to sterilization processes and requirements that are universally accepted sterilization standards for Tabletop Steam Sterilizers. A touchscreen controller is used for monitoring and control of the device.
More Information

Not Found

No
The summary describes a standard autoclave with a touchscreen controller and mentions adherence to universally accepted sterilization standards. There is no mention of AI or ML in the device description, intended use, or performance studies.

No
The device is an autoclave used for sterilizing medical instruments, not for directly treating a condition or disease in a patient.

No

The device is an autoclave designed for sterilizing medical materials, not for diagnosing diseases or conditions.

No

The device description clearly states it is an autoclave that uses electrical resistive heaters to produce steam, indicating it is a hardware device with a physical sterilization mechanism. While it has a touchscreen controller (software component), the core function and description are of a physical sterilizer.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to sterilize medical, dental, and surgical materials. This is a process performed on instruments and equipment, not on biological samples taken from the human body.
  • Device Description: The device is an autoclave that uses steam for sterilization. This aligns with the intended use of sterilizing instruments.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body to provide information for diagnosis, monitoring, or treatment.

IVD devices are specifically designed to perform tests on biological samples (like blood, urine, tissue) to provide diagnostic information. This device's function is entirely focused on sterilizing instruments for reuse.

N/A

Intended Use / Indications for Use

The Fort Defiance Industries FRONT-LINE Field Sterilizer (FL135) is an autoclave designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

Indications for Use (Describe)

The Fort Defiance Industries FRONT-LINE Field Sterilizer (FL135) is an autoclave designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

CYCLEMAXIMUM LOADEXPOSURE TEMP* (°F/°C)EXPOSURE TIME (MINUTES)DRY TIME (MINUTES)
IMMEDIATE USE (IUSS)25 LBS.270 / 13240
275 / 13530
TEXTILES3 TEXTILE PACKS**250 / 1213015
270 / 132415
275 / 135315
WRAPPED INSTR./POUCHES25 LBS.250 / 1213035
270 / 132430
275 / 135330
HANDPIECES12 HANDPIECES270 / 132430
275 / 135330

NOTES:

  • THE ANSI/AAMI ST55:2016 MINIMUM TEMPERATURE IS LISTED FOR ALL CYCLES.

** AAMI STANDARD TEXTILE PACK (9 TOWELS - ST 55 SECTION 5.7.1.1).

Product codes (comma separated list FDA assigned to the subject device)

FLE

Device Description

This change to the FRONT-LINE Field Sterilizer, model FL135, reduces the dry times of the wrapped instrument, textiles, and handpieces cycles from the original dry times to the following new dry times.

ModelCycleExposure Temperature (°F)Original Dry Time (mins)New Dry Time (mins)
FL135Immediate Use Steam Sterilization (IUSS)270 & 27500
Textiles250, 270, & 2756015
Wrapped Instruments/Pouches2506035
270 & 2756030
Handpieces270 & 2756030

To allow for the reduction in dry time, the dry rate of the sterilizer was increased by the following changes: (1) added a surface heater on the outside of the chamber to increase heat delivered to the chamber and load, (2) replaced the existing chamber load rack with an improved rack that allows for better heat and mass transfer, (3) implemented a new valve control scheme to improve condensate drainage and air flow during drying, and (4) increased the speed of the air pump to increase airflow through the chamber during drying. FDA consensus standard, AAMI ST55:2016, was referenced to validate dry performance of the predicate and new device. Acceptance criteria were also derived from the above standard.

The Fort Defiance Industries FRONT-LINE Field Sterilizer, model FL135, is an autoclave designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities. The device uses electrical resistive heaters to produce steam as the sterilizing agent. The device adheres to sterilization processes and requirements that are universally accepted sterilization standards for Tabletop Steam Sterilizers. A touchscreen controller is used for monitoring and control of the device.

The device is equipped with multiple safety features. The door includes a safety locking mechanism and door switch. The door switch does not permit cycle operation unless the door is sealed and does not permit the door to be opened until the chamber gauge pressure is near zero. Additional safety features include electronic overtemperature and overpressure protection, independent thermal cutoffs, and a pressure safety relief valve for the chamber.

Only United States Food and Drug Administration cleared sterilization accessories should be used with this autoclave.

The FRONT-LINE Field Sterilizer, model FL135, is a steam sterilizer composed of a stainless-steel chamber, an aluminum chamber door, electric resistive heaters, a plumbing system with a clean water tank, control system, and an aluminum frame providing a housing for all components. The device is electronically controlled and includes 10 preprogrammed sterilization cycles. All materials in the sterilization system and plumbing system are deemed acceptable for steam sterilizers.

The FRONT-LINE FL135 autoclave is a steam sterilizer composed of a pressure vessel that utilizes electric resistive heaters to produce the sterilizing agent, steam, from a chamber boiler. The autoclave contains an electronic control system that automatically initiates programmed Steam Flush Pressure Pulse (SFPP) phases to purge atmospheric air from the chamber followed by a temperature-controlled, timed steam exposure. A final drying phase, assisted by chamber internal and external heaters, completes the sterilization cycle. Exhaust air purged during the dry phase passes through a heat exchanger to condense water vapor and reclaim as much condensate as possible for reuse.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

healthcare facilities.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Fort Defiance Industries conducted validation studies in accordance with AAMI/ANSI ST55:2016 (FDA Recognition Number 14-518). Design outputs and testing demonstrate that the FRONT-LINE Model FL135 sterilizer meets all requirements of the standard.

Performance effectiveness of the sterilizer cycles and exposure time recommendations were demonstrated by the overkill method by passing 3 successive cycles at half the programmed exposure time to guarantee a sterility assurance level (SAL) of at least 10t6 probability of survival. Fort Defiance Industries validates its sterilization cycles using recommended practices, standards and guidelines developed by independent organizations such as the Association for the Advancement of Medical Instrumentation (AAMI). The Fort Defiance Industries FL135 Sterilizer has been validated to meet the requirements of ANSI/AAMI ST55:2016, Tabletop Steam Sterilizers (FDA Recognition Number 14-518). Table 2 summarizes the FL135 effectiveness testing and results performed by Fort Defiance Industries.

Table 2: Performance Testing Summary

Test ProcedureAAMI ST55:2016 RequirementsAAMI ST55:2016 TestsMeasured Performance
Sterilizer Temperature Control & Pressure Measurement TP-50704.4.3 -0/+6°F temp control5.4.3 - Temperature instrumented chamber tests to confirm steady state temperature control range and documentation of cold pointPASS
Software Verification and Validation4.4.7 Sterilizer fault conditions5.4.7 - Verified by inspection
TP-5110: Sterilizer Fault Conditions
TP-5120: General FunctionsPASS
TP-5130: Full Cycle Verification
Biological Performance - Wrapped Instrument TP-51604.5 Biological performance of sterilizers5.5.4 Biological performance with wrapped instrumentsPASS
Moisture Retention - Textiles TP-51904.7 Moisture Retention5.7.1 Textile test packsPASS
Moisture Retention - Wrapped Instrument TP-52004.7 Moisture Retention5.7.2 Wrapped Instr. Test traysPASS
Moisture Retention - Pouches TP-52104.7 Moisture Retention5.7.3 Paper-plastic peel pouchesPASS

The Fort Defiance FRONT-LINE Field Sterilizer FL135 has been designed, constructed, and tested to meet the safety and performance requirements of various national safety codes and standards. The device complies with the following standards:

  • ANSI/AAMI ST55:2016, Tabletop Steam Sterilizers (FDA Recognition Number 14-518).
  • IEC 61010-1 Edition 3.1 2017-01, "Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use -- Part 1 General Requirements". (FDA Recognition Number 19-34)
  • IEC 61326-1:2020, "Electrical Equipment for Measurement, Control, and Laboratory Use – EMC Requirements – Part 1: General Requirements" (General Use)
  • IEC 61010-2-040:2020, "Safety Requirements for Electrical Equipment for . Measurement, Control and Laboratory Use – Part 2-040: Particular Requirements for Sterilizers and Washer-disinfectors used to Treat Medical Materials" (General Use)
  • ASME Boiler and Pressure Vessel Code, Section VIII, Division I, 2023 Edition, . Rules for Construction of Pressure Vessels (General Use)

Table 3 summarizes FRONT-LINE Field Sterilizer FL135 safety testing results performed by Fort Defiance Industries or a Third Party where indicated.

FDI Test Procedure or 3rd PartyStandardTestsMeasured Performance
3rd Party Testing Electrical SafetyIEC 61010-1 Edition 3.1 2017-013rd Party test protocolPASS
3rd Party Testing EMC/EMIIEC 61326-1:20203rd Party test protocolPASS

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K213457

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.

March 7, 2025

Fort Defiance Industries LLC Chris Coleman Quality Management Representative 2411 Maremont Parkway Loudon, Tennessee 37774

Re: K243801

Trade/Device Name: FRONT-LINE Field Sterilizer (FL135) Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: February 12, 2025 Received: February 12, 2025

Dear Chris Coleman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

1

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

2

Sincerely,

Stephen A. Anisko -S

Digitally signed by Stephen A. Anisko -S Date: 2025.03.07 16:32:38 -05'00'

for: Christopher K. Dugard Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

510(k) Number (if known) K243801

Device Name

FRONT-LINE Field Sterilizer (FL135)

Indications for Use (Describe)

The Fort Defiance Industries FRONT-LINE Field Sterilizer (FL135) is an autoclave designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

| CYCLE | MAXIMUM
LOAD | EXPOSURE
TEMP* (°F/°C) | EXPOSURE
TIME
(MINUTES) | DRY TIME
(MINUTES) |
|-------------------------------|----------------------|---------------------------|-------------------------------|-----------------------|
| IMMEDIATE
USE (IUSS) | 25 LBS. | 270 / 132 | 4 | 0 |
| | | 275 / 135 | 3 | 0 |
| TEXTILES | 3 TEXTILE
PACKS** | 250 / 121 | 30 | 15 |
| | | 270 / 132 | 4 | 15 |
| | | 275 / 135 | 3 | 15 |
| WRAPPED
INSTR./
POUCHES | 25 LBS. | 250 / 121 | 30 | 35 |
| | | 270 / 132 | 4 | 30 |
| | | 275 / 135 | 3 | 30 |
| HANDPIECES | 12 HANDPIECES | 270 / 132 | 4 | 30 |
| | | 275 / 135 | 3 | 30 |

NOTES:

  • THE ANSI/AAMI ST55:2016 MINIMUM TEMPERATURE IS LISTED FOR ALL CYCLES.

** AAMI STANDARD TEXTILE PACK (9 TOWELS - ST 55 SECTION 5.7.1.1).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

4

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

5

Image /page/5/Picture/0 description: The image shows the logo for FDI. The logo is orange and consists of the letters "FDI" in a bold, sans-serif font. A white line cuts through the top of the letters, giving the logo a modern and dynamic look. The logo is simple and memorable, and the orange color gives it a sense of energy and enthusiasm.

510(k) Summary for Fort Defiance Industries FRONT-LINE Field Sterilizer Model FL135 K243801

Submitter / 510(k) Owner: Fort Defiance Industries LLC 2411 Maremont Parkway Loudon, TN 37774 Phone: (865) 408-0100 Fax No: (866) 732-8086

Contact: Chris Coleman Quality Management Representative Fort Defiance Industries LLC Phone: (865) 657-6108 (866) 732-8086 Fax:

Revised: 12 FEB 2025

6

Image /page/6/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all connected and have a modern, geometric design. The logo is simple and eye-catching.

1. Device Name

Trade Name:FRONT-LINE Field Sterilizer (FL135)
Model Name(s):FL135
Common/usual Name:Electronic autoclave, steam sterilizer
Classification Name:Steam Sterilizer
Class II Device - 21 CFR 880.6880
Product Code FLE

2. Predicate Device

K213457, FRONT-LINE Field Sterilizer, Model FL135, product code FLE, cleared March 2, 2022

3. Descrintion of Device

This change to the FRONT-LINE Field Sterilizer, model FL135, reduces the dry times of the wrapped instrument, textiles, and handpieces cycles from the original dry times to the following new dry times.

| Model | Cycle | Exposure
Temperature (°F) | Original Dry
Time (mins) | New Dry
Time (mins) |
|-------|------------------------------------------------|------------------------------|-----------------------------|------------------------|
| FL135 | Immediate Use Steam
Sterilization
(IUSS) | 270 & 275 | 0 | 0 |
| | Textiles | 250, 270, & 275 | 60 | 15 |
| | Wrapped Instruments/
Pouches | 250 | 60 | 35 |
| | | 270 & 275 | 60 | 30 |
| | Handpieces | 270 & 275 | 60 | 30 |

To allow for the reduction in dry time, the dry rate of the sterilizer was increased by the following changes: (1) added a surface heater on the outside of the chamber to increase heat delivered to the chamber and load, (2) replaced the existing chamber load rack with an improved rack that allows for better heat and mass transfer, (3) implemented a new valve control scheme to improve condensate drainage and air flow during drying, and (4) increased the speed of the air pump to increase airflow through the chamber during drying. FDA consensus standard, AAMI ST55:2016, was referenced to validate dry performance of the predicate and new device. Acceptance criteria were also derived from the above standard.

7

Image /page/7/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all connected and have a modern, geometric design. The logo is simple and eye-catching.

The Fort Defiance Industries FRONT-LINE Field Sterilizer, model FL135, is an autoclave designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities. The device uses electrical resistive heaters to produce steam as the sterilizing agent. The device adheres to sterilization processes and requirements that are universally accepted sterilization standards for Tabletop Steam Sterilizers. A touchscreen controller is used for monitoring and control of the device.

The device is equipped with multiple safety features. The door includes a safety locking mechanism and door switch. The door switch does not permit cycle operation unless the door is sealed and does not permit the door to be opened until the chamber gauge pressure is near zero. Additional safety features include electronic overtemperature and overpressure protection, independent thermal cutoffs, and a pressure safety relief valve for the chamber.

Only United States Food and Drug Administration cleared sterilization accessories should be used with this autoclave.

8

Image /page/8/Picture/0 description: The image shows the logo for FDI in orange. The logo is a stylized version of the letters FDI. The letters are bold and sans-serif. The logo is simple and modern.

4. Design and Materials

The FRONT-LINE Field Sterilizer, model FL135, is a steam sterilizer composed of a stainless-steel chamber, an aluminum chamber door, electric resistive heaters, a plumbing system with a clean water tank, control system, and an aluminum frame providing a housing for all components. The device is electronically controlled and includes 10 preprogrammed sterilization cycles. All materials in the sterilization system and plumbing system are deemed acceptable for steam sterilizers.

న. Technology

The FRONT-LINE FL135 autoclave is a steam sterilizer composed of a pressure vessel that utilizes electric resistive heaters to produce the sterilizing agent, steam, from a chamber boiler. The autoclave contains an electronic control system that automatically initiates programmed Steam Flush Pressure Pulse (SFPP) phases to purge atmospheric air from the chamber followed by a temperature-controlled, timed steam exposure. A final drying phase, assisted by chamber internal and external heaters, completes the sterilization cycle. Exhaust air purged during the dry phase passes through a heat exchanger to condense water vapor and reclaim as much condensate as possible for reuse.

9

Image /page/9/Picture/0 description: The image shows the logo for FDI. The logo is orange and consists of the letters F, D, and I. The letters are connected and have a modern, geometric design. The logo is simple and eye-catching.

6. Intended Use

The Fort Defiance Industries FRONT-LINE Field Sterilizer (FL135) is an autoclave designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

| Model | Cycle | Maximum
Load | Exposure
Temperature1
(°F/°C) | Exposure
Time
(minutes) | Dry Time
(minutes) |
|-------|-------------------------------------|------------------|-------------------------------------|-------------------------------|-----------------------|
| FL135 | Immediate Use
(IUSS) | 25 lbs. | 270/132 | 4 | 0 |
| | | | 275/135 | 3 | |
| FL135 | Textiles | 3 Textile Packs2 | 250/121 | 30 | 15 |
| | | | 270/132 | 4 | |
| | | | 275/135 | 3 | |
| FL135 | Wrapped
Instruments /
Pouches | 25 lbs. | 250/121 | 30 | 30 |
| | | | 270/132 | 4 | |
| | | | 275/135 | 3 | |
| FL135 | Handpieces | 12 Handpieces | 270/132 | 4 | 30 |
| | | | 275/135 | 3 | |

1 The ANSI/AAMI ST55:2016 minimum temperature is listed for all cycles.

2 AAMI Standard Textile Pack (9 towels - ST 55 section 5.7.1.1).

10

Image /page/10/Picture/0 description: The image shows the logo for FDI. The logo is orange and features the letters "FDI" in a bold, sans-serif font. The letters are connected to each other, and there is a white line running through the middle of each letter. The logo is simple and modern, and the orange color is eye-catching.

7. Comparison of Technology

The previously cleared device and the FRONT-LINE Field Sterilizer model FL135 are traditional steam sterilizers of similar chamber size using AAMI approved Steam Flush Pressure Pulse (SFPP) dynamic air removal cycles. The previously cleared device and the new FL135 sterilizer use electric heaters to generate steam and touchscreen controllers for cycle control. Based on the information provided below, there are negligible differences noted between the new FL135 and the predicate with regards to chamber size, cycle parameters, and configuration of steam generation. The differences do not affect the safety or effectiveness of the new device compared to the predicate. Table 1 provides a summary of performance and safety characteristics of the proposed new device and the predicate.

| Performance
and/or Safety
Characteristic | New Device
Fort Defiance Industries
FRONT-LINE Field Sterilizer
Model FL135 K243801 | Predicate Device
Fort Defiance Industries FRONT-
LINE Field Sterilizer
Model FL135 (K213457) |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for
Use | The Fort Defiance Industries
FRONT-LINE Field Sterilizers
are autoclaves designed for
sterilizing heat- and moisture-
stable medical, dental, and
surgical materials, including
wrapped and unwrapped, solid,
porous, and hollow items (e.g.,
dental handpieces, suction pipes)
in healthcare facilities. | The Fort Defiance Industries
FRONT-LINE Field Sterilizers
are autoclaves designed for
sterilizing heat- and moisture-
stable medical, dental, and
surgical materials, including
wrapped and unwrapped, solid,
porous, and hollow items (e.g.,
dental handpieces, suction pipes)
in healthcare facilities. |
| Operating
Principle | Steam Flush Pressure Pulse
(SFPP).
Steam is the sterilization agent
(traditional method).
ANSI/AAMI ST55 | Steam Flush Pressure Pulse
(SFPP).
Steam is the sterilization agent
(traditional method).
ANSI/AAMI ST55 |
| Performance
and Safety
Standards
Followed | ASME Sec. VIII Div 1
UL 61010-1
EN 61326-1
UL 61010-2-040 | ASME Sec. VIII Div 1
UL 61010-1
EN 61326-1
UL 61010-2-040 |
| Performance
and/or Safety
Characteristic | New Device
Fort Defiance Industries
FRONT-LINE Field Sterilizer
Model FL135 K243801 | Predicate Device
Fort Defiance Industries FRONT-
LINE Field Sterilizer
Model FL135 (K213457) |
| | Factory Pre-Programmed | Factory Pre-Programmed |
| | Steam Flush Pressure Pulse
(SFPP) Conditioning Phase | Steam Flush Pressure Pulse
(SFPP) Conditioning Phase |
| Sterilization
Cycles | 30 min exposure @ 250°F Dry
time dependent on load | 30 min exposure @ 250°F Dry
time dependent on load |
| | 4 min exposure @ 270°F
Dry time dependent on load | 4 min exposure @ 270°F
Dry time dependent on load |
| | 3 min exposure @ 275°F
Dry time dependent on load | 3 min exposure @ 275°F
Dry time dependent on load |
| Dry Times -
Wrapped | 275°F: 30 min
270°F: 30 min
250°F: 35 min | 60 min
(all temperatures) |
| Dry Times -
Textiles | 15 min
(all temperatures) | 60 min
(all temperatures) |
| Dry Times -
Handpieces | 30 min
(all temperatures) | 60 min
(all temperatures) |
| Chamber Size
& Volume | FL135
13.5" dia. X 24" lg.
2.00 cf. (56.35 L) | FL135
13.5" dia. X 24" lg.
2.00 cf. (56.35 L) |
| Chamber
Design and
Construction | Single Wall Chamber ASME
Section VIII, Div. I certified
Stainless steel (316L) chamber
and aluminum door (6061-T6) | Single Wall Chamber ASME
Section VIII, Div. I certified
Stainless steel (316L) chamber
and aluminum door (6061-T6) |
| Shelves / Trays | Stainless Steel wire tray | Aluminum tray |
| Steam Source | Internal chamber boiler via
partitioned section | Internal chamber boiler via
partitioned section |
| Steam
Generation
Mechanism | Electric resistive heaters
submerged in chamber boiler | Electric resistive heaters
submerged in chamber boiler |
| Drying Process | External air pulled through 0.3-
micron HEPA filter then heated
by internal cartridge heaters and
external chamber wall heater for
drying | External air pulled through 0.3-
micron HEPA filter then heated
by internal cartridge heaters for
drying |
| Performance
and/or Safety
Characteristic | New Device
Fort Defiance Industries
FRONT-LINE Field Sterilizer
Model FL135 K243801 | Predicate Device
Fort Defiance Industries FRONT-
LINE Field Sterilizer
Model FL135 (K213457) |
| Control
Technology | Electronic controller touch
screen (PLC) | Electronic controller touch screen
(PLC) |
| Process
Monitors | Chamber pressure transmitter.
Dual element chamber
temperature sensor. | Chamber pressure transmitter.
Dual element chamber
temperature sensor. |
| Door Failsafe | Automatic Door Interlock | Automatic Door Interlock |
| Primary
Overtemp and
Overpressure
Control | PLC Controller - Utilizing
RTDs and Pressure
Transmitter | PLC Controller - Utilizing RTDs
and Pressure
Transmitter |
| Secondary
Overtemp
Control | Thermocouple Temperature
Cut-off Switch (TCO) | Thermocouple Temperature
Cut-off Switch (TCO) |
| Secondary
Overpressure
Control | ASME Pressure Safety Valve | ASME Pressure Safety Valve |
| Electrical
Input | 120VAC ± 10%,
1Φ, 50/60 Hz, 15 A
230-240VAC ± 10%,
1Φ, 50/60 Hz, 15 A | 120VAC ± 10%,
1Φ, 50/60 Hz, 15 A
230-240VAC ± 10%,
1Φ, 50/60 Hz, 15 A |

Table 1: Comparison of Technology

11

Image /page/11/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. The logo is simple and modern.

12

Image /page/12/Picture/0 description: The image shows the logo for FDI in orange. The logo is a stylized version of the letters FDI. The letters are bold and sans-serif. The letter D is incomplete, with a diagonal line cutting through the upper right corner. The letter I also has a diagonal line cutting through the upper right corner.

8. Summarv of Non-Clinical Testing

Fort Defiance Industries conducted validation studies in accordance with AAMI/ANSI ST55:2016 (FDA Recognition Number 14-518). Design outputs and testing demonstrate that the FRONT-LINE Model FL135 sterilizer meets all requirements of the standard.

Performance effectiveness of the sterilizer cycles and exposure time recommendations were demonstrated by the overkill method by passing 3 successive cycles at half the programmed exposure time to guarantee a sterility assurance level (SAL) of at least 10t6 probability of survival. Fort Defiance Industries validates its sterilization cycles using recommended practices, standards and guidelines developed by independent organizations such as the Association for the Advancement of Medical Instrumentation (AAMI). The Fort Defiance Industries FL135 Sterilizer has been validated to meet the requirements of ANSI/AAMI ST55:2016, Tabletop Steam Sterilizers (FDA Recognition Number 14-518). Table 2 summarizes the FL135 effectiveness testing and results performed by Fort Defiance Industries.

13

Image /page/13/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are connected to each other and have a modern, geometric design. The logo is simple and eye-catching.

Table 2: Performance Testing Summary

Test ProcedureAAMI ST55:2016 RequirementsAAMI ST55:2016 TestsMeasured Performance
Sterilizer Temperature Control & Pressure Measurement
TP-50704.4.3 -0/+6°F temp control5.4.3 - Temperature instrumented chamber tests to confirm steady state temperature control range and documentation of cold pointPASS
Software Verification and Validation4.4.7 Sterilizer fault conditions5.4.7 - Verified by inspection
TP-5110: Sterilizer Fault Conditions
TP-5120: General FunctionsPASS
TP-5130: Full Cycle Verification
Biological Performance - Wrapped Instrument
TP-51604.5 Biological performance of sterilizers5.5.4 Biological performance with wrapped instrumentsPASS
Moisture Retention - Textiles
TP-51904.7 Moisture Retention5.7.1 Textile test packsPASS
Moisture Retention - Wrapped Instrument
TP-52004.7 Moisture Retention5.7.2 Wrapped Instr. Test traysPASS
Moisture Retention - Pouches
TP-52104.7 Moisture Retention5.7.3 Paper-plastic peel pouchesPASS

14

Image /page/14/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are connected to each other and have a modern, geometric design. The logo is simple and eye-catching.

9. Safety

The Fort Defiance FRONT-LINE Field Sterilizer FL135 has been designed, constructed, and tested to meet the safety and performance requirements of various national safety codes and standards. The device complies with the following standards:

  • ANSI/AAMI ST55:2016, Tabletop Steam Sterilizers (FDA Recognition Number ● 14-518).
  • IEC 61010-1 Edition 3.1 2017-01, "Safety Requirements for Electrical Equipment ● for Measurement, Control and Laboratorv Use -- Part 1 General Requirements". (FDA Recognition Number 19-34)
  • IEC 61326-1:2020, "Electrical Equipment for Measurement, Control, and ● Laboratory Use – EMC Requirements – Part 1: General Requirements" (General Use)
  • IEC 61010-2-040:2020, "Safety Requirements for Electrical Equipment for . Measurement, Control and Laboratory Use – Part 2-040: Particular Requirements for Sterilizers and Washer-disinfectors used to Treat Medical Materials" (General Use)
  • ASME Boiler and Pressure Vessel Code, Section VIII, Division I, 2023 Edition, . Rules for Construction of Pressure Vessels (General Use)

Table 3 summarizes FRONT-LINE Field Sterilizer FL135 safety testing results performed by Fort Defiance Industries or a Third Party where indicated.

| FDI Test Procedure or
3rd Party | Standard | Tests | Measured
Performance |
|----------------------------------------|------------------------------------|-------------------------|-------------------------|
| 3rd Party Testing
Electrical Safety | IEC 61010-1 Edition 3.1
2017-01 | 3rd Party test protocol | PASS |
| 3rd Party Testing
EMC/EMI | IEC 61326-1:2020 | 3rd Party test protocol | PASS |

Table 3: Safety Testing Summary

15

Image /page/15/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all connected and have a modern, geometric design. The logo is simple and eye-catching.

10. Conclusion

The conclusion drawn from the nonclinical test results demonstrates that the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device, K213457, Class II (21 CFR 880.6880, Product code FLE).