(257 days)
The PRO Pen Microneedling System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years and older.
The PRO Pen microneedling system is a microneedling device and disposable 14-pin cartridges intended for use in adults. The device can be powered with a rechargeable battery or a provided power adapter. The device uses disposable single-use, sterile 14-pin (14-needle) cartridges made of polymer and stainless steel.
The provided FDA 510(k) Clearance Letter for the PRO Pen Microneedling System (K243800) does not describe specific acceptance criteria and the detailed study that proves the device meets those criteria in the way typically expected for an AI/CAD device. The document is a clearance letter for a physical medical device, a microneedling system, not an AI software.
Therefore, many of the requested items (like ground truth establishment, expert qualifications, adjudication methods, MRMC studies, standalone performance of an algorithm, and specifics about training/test sets for an AI) are not applicable to this type of device clearance document.
However, I can extract the information relevant to the performance tests conducted and the conclusions made regarding the device's adherence to regulatory requirements.
Here's a breakdown based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document describes performance tests in accordance with special controls for microneedling devices. While explicit numerical acceptance criteria are not provided in this summary, the conclusion of the testing is that the device "successfully passed all performance tests" and "complies with FDA requirements."
| Acceptance Criteria Category (based on special controls in 21 CFR 878.4430) | Reported Device Performance |
|---|---|
| Needle Penetration Depth Accuracy | Successfully passed. (Details not specified, but implies meeting the established accuracy requirements for microneedling devices). The device can be adjusted in 0.2mm increments up to a maximum penetration depth of 1.5mm. |
| Puncture rate characterization | Successfully passed. (Details not specified, but implies meeting the established rate charactersitics). |
| Vacuum and fluid ingress prevention evaluation | Successfully passed. Demonstrated reliable performance across all critical tests, including vacuum prevention and fluid ingress. |
| Sterility of patient contacting components | Successfully passed. |
| Aging distribution and shelf-life testing | Successfully passed. 14-pin cartridges have a shelf life of 3 years. |
| Electromagnetic compatibility and electrical safety | Successfully passed. Confirmed safe operation in healthcare environments. |
| In vitro cytotoxicity | Successfully passed. |
| Material characterization through extractables and leachables | Successfully passed. (Toxicological risk assessment of findings also passed). |
| Human patch testing for irritation | Successfully passed. No irritation reactions observed. Classified as non-irritating. |
| Human repeat insult patch testing (HRIPT) | Successfully passed. No sensitization reactions observed. Classified as non-sensitizing. |
| Cleaning and disinfection validation | Successfully passed. Procedures were validated, confirming the device can be safely cleaned and disinfected to reduce contamination risks. |
| Cross contamination protection | Complies with FDA requirements (Implied by Vacuum and fluid ingress prevention evaluation and Cleaning/disinfection validation). |
| Packaging durability | Complies with FDA requirements (Implied by aging/shelf-life testing and overall safety). Implied by successful vacuum and fluid ingress prevention which often relates to packaging integrity. |
| Needle retention | Demonstrated reliable performance. |
2. Sample size used for the test set and the data provenance
The document does not specify general "test set" sample sizes or data provenance (e.g., country of origin, retrospective/prospective) because this clearance is for a physical device, not an AI model. The "tests" mentioned are engineering and biocompatibility evaluations. For human patch testing (HRIPT), the sample size would typically be small (e.g., 50-200 subjects) for such tests, but no specific number is provided in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not an AI/CAD system requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for the device's performance is established by objective engineering standards, biological assays (cytotoxicity, extractables/leachables), and clinical observation (human patch testing for irritation/sensitization).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As this is not an AI/CAD system evaluating diagnostic images, there would be no adjudication method involving multiple human readers.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI system.
7. The type of ground truth used
The "ground truth" for the PRO Pen Microneedling System's performance is based on:
- Engineering Standards: For needle penetration depth, puncture rate, electrical safety, EMC.
- Biological/Chemical Testing Standards: For sterility, vacuum/fluid ingress, cytotoxicity, extractables/leachables, material characterization.
- Clinical Observation/Standard Dermatological Assessments: For human patch testing (irritation and sensitization reactions).
- Microbiological Standards: For cleaning and disinfection validation.
- Stability/Accelerated Aging Protocols: For shelf-life testing.
8. The sample size for the training set
Not applicable. This is not an AI system.
9. How the ground truth for the training set was established
Not applicable. This is not an AI system.
FDA 510(k) Clearance Letter - PRO Pen Microneedling System
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.08.00
Silver Spring, MD 20993
www.fda.gov
August 25, 2025
Dermalogica, LLC
Nelson Torres
Director, Medical Tools and Products
1535 Beachey Pl
Carson, California 90746
Re: K243800
Trade/Device Name: PRO Pen Microneedling System (6883)
Regulation Number: 21 CFR 878.4430
Regulation Name: Microneedling Device For Aesthetic Use
Regulatory Class: Class II
Product Code: QAI
Dated: May 15, 2025
Received: May 16, 2025
Dear Nelson Torres:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
2K243800 - Nelson Torres Page
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
3K243800 - Nelson Torres Page
Sincerely,
Jodie Giordano -S
Jodie Giordano, Ph.D.
Acting Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Indications for Use
Submission Number (if known)
K243800
Device Name
PRO Pen Microneedling System (6883)
Indications for Use (Describe)
The PRO Pen Microneedling System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years and older.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
1 of 2
Applicant Contact Details
Name: Dermalogica, LLC
Address: 1535 Beachey Place, Carson, CA 90746, USA
Contact Telephone: (210) 912-7017
Contact Name: Nelson Torres
Contact email: nelson.torres@dermalogica.com
Prepared: August 25, 2025
Device Details
Trade Name: PRO Pen Microneedling System (6883)
Common Name: Microneedling device for aesthetic use
Classification Name: Powered Microneedle Device
Regulation Number: 878.4430
Product Code(s): QAI
Legally Marketed Predicate Device
Predicate #: K230420
Name: Dr. Pen Microneedling System
Product Code(s): QAI
Device Description Summary
The PRO Pen microneedling system is a microneedling device and disposable 14-pin cartridges intended for use in adults. The device can be powered with a rechargeable battery or a provided power adapter. The device uses disposable single-use, sterile 14-pin (14-needle) cartridges made of polymer and stainless steel.
Intended Use/Indications for Use
The PRO Pen Microneedling System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years and older.
Indications for Use Comparison
Indications for use of the subject device are identical as for the predicate device.
Technological Comparison
The PRO Pen Microneedling System is identical to the predicate device in the following characteristics:
- Indications for use
- intended use location
- primary power source
- battery charger
- control mechanism
- puncture rate
- number of needles and needle gage
- needle material
- needle geometry
Page 6
2 of 2
- needle arrangement
- needle spacing
- sterile state and sterilization method
- cross contamination protection
The PRO Pen Microneedling System differs from the predicate device in the following characteristics:
-
Alternate Power Source: The PRO Pen can run with an AC adapter directly from mains power in addition to the rechargeable battery, as opposed to the predicate device, which can only be operated with a rechargeable battery.
-
Depth Settings and Max Needle Depth: The PRO Pen can be adjusted in 0.2mm increments up to a maximum penetration depth of 1.5mm, which is shorter than the predicate device's maximum penetration depth of 2.0mm
-
Shelf Life: The 14-pin cartridges of the PRO Pen have a shelf life of 3 years, which is greater than the predicate device's shelf life of 2 years.
Non-clinical and/or Clinical Tests Summary & Conclusions
The following performance tests were conducted with the PRO Pen Microneedling System in accordance with special controls for microneedling devices in 21 CFR 878.4430:
- Needle Penetration Depth Accuracy,
- Puncture rate characterization
- Vacuum and fluid ingress prevention evaluation,
- Sterility of patient contacting components,
- Aging distribution and shelf-life testing,
- Electromagnetic compatibility and electrical safety,
- In vitro cytotoxicity
- Material characterization through extractables and leachables, with a corresponding toxicological risk assessment of the findings
- Human patch testing for irritation
- Human repeat insult patch testing (HRIPT)
- Cleaning and disinfection validation
The PRO Pen Microneedling system successfully passed all performance tests listed above.
The PRO Pen Microneedling System complies with FDA requirements under special controls detailed in 21 CFR 878.4430, including needle penetration depth accuracy, puncture rate, cross contamination/fluid ingress prevention, packaging durability, cleaning and disinfection validation, electromagnetic compatibility, and shelf-life testing. Human patch and repeat insult patch testing of the patient-contacting materials yielded no irritation or sensitization reactions and concluded the materials can be classified as non-irritating and non-sensitizing. The system demonstrated reliable performance across all critical tests, including vacuum prevention, fluid ingress, packaging integrity, and needle retention. Cleaning and disinfection procedures were validated, confirming the device can be safely cleaned and disinfected to reduce contamination risks.
The device also passed EMC and electrical safety testing, confirming safe operation in healthcare environments.
Collectively, the results of all clinical and non-clinical performance testing conducted demonstrate that the PRO Pen Microneedling System is as safe, effective, and performs as well as the predicate device.
§ 878.4430 Microneedling device for aesthetic use.
(a)
Identification. A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.