K Number
K243767

Validate with FDA (Live)

Date Cleared
2025-08-28

(265 days)

Product Code
Regulation Number
868.5895
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SV600, SV800 Ventilator is intended to be used in intensive care situations for long-term or during transport within a professional healthcare facility. The SV600, SV800 Ventilator is intended to provide ventilation assistance and breathing support for adult, pediatric and neonate patients with a minimum body weight of 0.5 kg. The SV600, SV800 Ventilator should be operated by properly-trained and authorized medical personnel. This equipment is not suitable for use in an MRI environment.

Device Description

The SV600 and SV800 Ventilators are pneumatically-driven and electronically-controlled ventilators. The Ventilators consists of a main unit (including pneumatic circuit, electronic system, mechanical structure, display, CO2 module, SpO2 module), trolley and support arm. The device also includes a neonatal flow sensor and neonatal flow sensor cable, which are used to measure the patient inspiration/expiration flow in neonatal ventilation modes.

AI/ML Overview

The provided FDA 510(k) Clearance Letter for the SV600, SV800 Ventilator describes modifications to an existing device, primarily the addition of neonatal ventilation capabilities and updated monitoring modules. However, the document does not contain the level of detail typically found in a clinical study report for evaluating acceptance criteria and device performance in the way you've requested for studies involving AI algorithms, image analysis, or expert consensus with specific metrics like sensitivity, specificity, or AUC.

This document focuses on providing evidence of substantial equivalence to predicate devices through technical comparisons and various forms of bench testing, software verification, and compliance with consensus standards. It does not present a performance study with acceptance criteria specific to an AI device's output (e.g., accuracy against ground truth, reader performance improvements).

Therefore, I cannot fill out your requested table and answer many of your specific questions as the information is not present in the provided text.

Here's what can be extracted based on the document's content:


Acceptance Criteria and Device Performance (Based on available information)

The document doesn't define "acceptance criteria" in terms of specific performance metrics (like sensitivity, specificity, accuracy) that would be typically found for an AI or diagnostic device. Instead, "acceptance criteria" are implied by compliance with various technical specifications and international standards. Device performance is generally reported as "meets specifications" or "is equivalent to the predicate."

Acceptance Criteria (Implied by Standards/Specifications)Reported Device Performance (Summary)
Biocompatibility: Conformance to ISO 10993 and ISO 18562 series.Testing performed, device meets standards.
Software Verification & Validation: Conformance to FDA Guidance for Software.Verification and validation conducted; product works as designed and meets design/performance checks.
Electromagnetic Compatibility (EMC) & Electrical Safety: Conformance to IEC 60601-1 and related collateral standards.Testing performed, device meets standards.
Functional & System Level Performance (Bench Testing): Meets accuracy specifications and demonstrates equivalence to predicate for new/modified features (e.g., neonatal ventilation parameters, CO2/SpO2 modules).Bench testing shows device meets accuracy specifications and is substantially equivalent to the predicate.
Usability: Conformance to IEC 60601-1-6.Testing performed, ensuring usability.
Alarm Systems: Conformance to IEC 60601-1-8.Testing performed, ensuring alarm system effectiveness.
Critical Care Ventilator Specifics: Conformance to ISO 80601-2-12.Testing performed, device meets standards.
Respiratory Gas Monitors: Conformance to ISO 80601-2-55.Testing performed, device meets standards.
Pulse Oximeter Equipment: Conformance to ISO 80601-2-61.Testing performed, device meets standards.
Ventilator Endurance: Conformance to ASTM F1100-90.Testing performed, ensuring endurance.
Key Technical Parameters (e.g., TV range, O2% range, SpO2 accuracy): Achieves specified ranges and accuracies (as detailed in the comparison tables).The comparison tables indicate specific ranges and accuracies for various parameters, demonstrating the device's adherence to these specifications (e.g., SpO2 accuracy of ±2% for adult/pediatric, ±3% for neonate in certain ranges).

Study Details (Based on available information in the 510(k) Summary)

  1. Sample size used for the test set and the data provenance:

    • The document does not specify sample sizes for test sets in the context of clinical performance evaluation (e.g., number of patients, number of readings). The "testing" mentioned refers to engineering, software, and standards compliance evaluations.
    • Data provenance is not provided, as this is not a clinical study report.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The document describes engineering and standards compliance testing, not a clinical study involving expert-established ground truth for performance evaluation of an AI component or diagnostic output.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a clinical study with an adjudication process for ground truth.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This document describes a ventilator, not an AI diagnostic or assistance tool in that context. While it includes "Intellicycle" and "Lung Recruitment (SI)" features, these are not presented as AI assistance augmenting human reader performance in a diagnostic capacity.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The document implies standalone performance testing for various technical aspects of the ventilator and its integrated modules (CO2, SpO2 sensors) against their respective specifications and standards. For example, SpO2 sensor accuracy is stated as "Measurement accuracy: 70 to 100%: ±2% (adult/pediatric mode)" for the Mindray SpO2 module. However, this is for sensor performance, not a complete "algorithm only" evaluation in the context of AI diagnostic output as typically measured.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the technical and performance testing, the "ground truth" would be established by reference standards, calibrated equipment, and design specifications. For example, a flow sensor's accuracy would be tested against a known, precise flow rate. For biocompatibility, the ground truth is defined by the toxicological profiles dictated by the ISO standards. For software, the ground truth is the functional requirements and design specifications.
    • There is no mention of expert consensus, pathology, or outcomes data being used to establish ground truth for clinical performance evaluation of an AI component.
  7. The sample size for the training set:

    • Not applicable. This document does not describe the development or validation of an AI algorithm with a training set. The "Intellicycle" feature is mentioned, but no details regarding its development, training data, or validation as an AI algorithm are provided.
  8. How the ground truth for the training set was established:

    • Not applicable, as no AI training set is described.

Summary Limitations:

The provided text is a 510(k) summary, which aims to demonstrate substantial equivalence to legally marketed predicate devices. It details technical changes, compliance with regulatory standards (e.g., biocompatibility, electrical safety, EMC, software V&V), and functional testing results. It does not outline a clinical performance study with the types of metrics and methodologies commonly associated with evaluating AI-driven diagnostic devices or those requiring expert consensus for ground truth. Therefore, many of your specific questions are not addressed by the provided document.

FDA 510(k) Clearance Letter - SV600, SV800 Ventilator

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 28, 2025

Shenzhen Mindray Bio-medical Electronics Co., LTD.
Li Lei
Manager Regulatory Affairs
Mindray Building, Keji 12th Road South,
Hi-tech Industrial Park, Nanshan
Shenzhen, Guangdong 518057
China

Re: K243767
Trade/Device Name: SV600, SV800 Ventilator
Regulation Number: 21 CFR 868.5895
Regulation Name: Continuous Ventilator
Regulatory Class: Class II
Product Code: CBK
Dated: August 1, 2025
Received: August 1, 2025

Dear Li Lei:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243767 - Li Lei Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243767 - Li Lei Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ethan L. Nyberg -S

Ethan Nyberg, Ph.D.
Assistant Director, Respiratory Devices Team
DHT1C: Division of Sleep Disordered
Breathing, Respiratory and
Anesthesia Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243767

Device Name: SV600, SV800 Ventilator

Indications for Use (Describe)

The SV600, SV800 Ventilator is intended to be used in intensive care situations for long-term or during transport within a professional healthcare facility. The SV600, SV800 Ventilator is intended to provide ventilation assistance and breathing support for adult, pediatric and neonate patients with a minimum body weight of 0.5 kg. The SV600, SV800 Ventilator should be operated by properly-trained and authorized medical personnel. This equipment is not suitable for use in an MRI environment.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number"

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 1 of 19

510(K) SUMMARY

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Mindray SV600, SV800 Ventilator is provided below.

1. SUBMITTER

Applicant: SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Mindray Building, Keji 12th Road South High-tech Industrial Park,
Nanshan Shenzhen 518057, P.R. China
Tel: +86 755 81888998
Fax: +86 755 26582680

Contact: Contact Person: Li Lei
Title: Manager Regulatory Affairs
Phone: +86 755 81885953
Fax: +86 755 26582680
E-mail: lilei.js@mindray.com

Date Prepared: December 3, 2024

2. DEVICE

Trade/Device Name: SV600, SV800 Ventilator
Common Name: Continuous Ventilator
Regulation Number: 21 CFR 868.5895
Device Class: Class II
Review Panel: Anesthesiology

3. PREDICATE DEVICES

Primary predicate:

  • K220107-Mindray SV600, SV800 Ventilator, ShenZhen Mindray Bio-Medical Electronics CO., LTD.

Secondary predicate:

  • K213799 - BeneVision N Series Patient Monitors (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1), ShenZhen Mindray Bio-Medical Electronics Co., LTD.
  • K083050 - Evita XL Ventilator, Drager Medical AG & Co. KG
  • K180098 - Servo-U Ventilator, Maquet Critical Care AB

Page 6

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 2 of 19

4. DEVICE DESCRIPTION

The SV600 and SV800 Ventilators are pneumatically-driven and electronically-controlled ventilators. The Ventilators consists of a main unit (including pneumatic circuit, electronic system, mechanical structure, display, CO2 module, SpO2 module), trolley and support arm. The device also includes a neonatal flow sensor and neonatal flow sensor cable, which are used to measure the patient inspiration/expiration flow in neonatal ventilation modes.

5. INTENDED USE/INDICATIONS FOR USE

The SV600, SV800 Ventilator is intended to be used in intensive care situations for long-term or during transport within a professional healthcare facility. The SV600, SV800 Ventilator is intended to provide ventilation assistance and breathing support for adult, pediatric and neonate patients with a minimum body weight of 0.5 kg. The SV600, SV800 Ventilator should be operated by properly-trained and authorized medical personnel. This equipment is not suitable for use in an MRI environment.

6. SUBSTANTIAL EQUIVALENCE

Comparison of Indications

Comparing with the primary predicate SV600, SV800 Ventilator(K220107) and Secondary predicate Evita XL Ventilator (K083050) and Servo-U Ventilator (K180098), the indications for use for the subject device SV600, SV800 Ventilators are equivalent.

The BeneVision N Series Patient Monitor modules (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1), ShenZhen Mindray Bio-Medical Electronics Co., LTD cleared under K213799 are incorporated into thoes ventilators with no changes to the modules. This predicate supports the CO2 module and accessories, the SpO2 module and accessories, WiFi Module. A more detailed comparison of the features is included in the sections below.

The indications for the BeneVision N Series Patient Monitors modules are the same since the modules are incorporated with no changes in indications into the ventilators.

As a conclusion, the indications for use of the subject device SV600, SV800 Ventilators is the same as the primary predicate SV600, SV800 Ventilator as cleared in K220107, secondary predicate Evita XL Ventilator as cleared in K083050, Servo-U Ventilator as cleared in K180098 and the BeneVision N Series Patient Monitors as cleared in K213799.

Comparison of Technological Characteristics

The table below compares the key technological features of the subject devices to the primary predicate device (SV600, SV800 Ventilator (K220107)) and secondary predicate devices. The

Page 7

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 3 of 19

main differences between the subject device and the primary predicate SV600, SV800 Ventilators are:

  1. A neonatal ventilation option has been added including the ventilation mode nCPAP which is specific only to the neonate population.
  2. Modifications of certain parameters to support neonates
  3. Addition of a neonatal flow sensor.
  4. Addition of oxygen therapy.
  5. Addition of a backup air supply (blower) (optional).
  6. Addition of WiFi communication.
  7. Addition of neonate CO2 measurement.
  8. Addition of neonate Mindray SpO2 measurement.
  9. Addition of Masimo SpO2 and Nellcor SpO2 measurement.
  10. Modification of the main control board.
  11. Modification of the display.
  12. Modification of the structure of the expiratory valve assembly.
Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
Ventilation mode
V-A/CYesV-A/CSame
P-A/CYesP-A/CSame
PRVCYesPRVCSame
V-SIMVYesV-SIMVSame
P-SIMVYesP-SIMVSame
PRVC-SIMVYesPRVC-SIMVSame
CPAPYesCPAPSame
PSVYesPSVSame
DuoLevelYesDuoLevelSame
APRVYesAPRVSame
AMVYesAMVSame
VSYesVSSame

Page 8

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 4 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
nCPAPYes-Different, same as secondary predicate device Evita XL(K083050)
PSV-S/TYesPSV-S/TSame
Apnea VentilationYesYesSame
Specifications - Ventilator setting parameter
TV range
Adu: 100 to 4000 mlAdu: 100 to 4000 mlSame
Ped: 20 to 300 mlPed: 20 to 300 mlSame
Neo: 2 to 100 ml-Different, similar to secondary predicate device ServoU(K180098)
O2% range
Adu/Ped: 21 to 100 vol.%21 to 100 vol.%Same
Neo: 21 to 100 vol.%-Different, same as secondary predicate device Evita XL(K083050)
f range
Adu/Ped: 1 to 100 /min1 to 100 /minSame
Neo: 1 to 150 /min-Different, similar as secondary predicate device Evita XL(K083050)
fsimv range
Adu/Ped: 1 to 60 /min1 to 60 /minSame
Neo: 1 to 60 /min-Different, similar to secondary predicate device Evita XL(K083050)
Tinsp range
Adu/Ped: 0.1 to 10 s0.1 to 10 sSame

Page 9

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 5 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
Neo: 0.1 to 10 s-Different, same as secondary predicate device Evita XL(K083050)
I:E range
Adu/Ped: 4:1 to 1:104:1 to 1:10Same
Neo: 4:1 to 1:10-Different, same as secondary predicate device ServoU(K180098)
Tslope range
Adu/Ped: 0 to 2 s0 to 2 sSame
Neo: 0 to 2 s-Different, same as secondary predicate device Evita XL(K083050)
△Pinsp range
Adu: 1 to 100 cmH2O Note: when backup air supply works, can set to up to 80cmH2OAdu: 1 to 100 cmH2OSame
Ped: 1 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2OPed: 1 to 95 cmH2O
Neo: 1 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2O-Different, similar to secondary predicate device Evita XL(K083050)
PEEP range
Adu/Ped: 0 to 50 cmH2O0 to 50 cmH2OSame
Neo: 0 to 50 cmH2O-Different, same as secondary predicate device Evita XL(K083050)
△Psupp range
Adu: 0 to 100 cmH2OAdu: 0 to 100 cmH2OSame

Page 10

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 6 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
Neo: 0.1 to 10 s-Different, same as secondary predicate device Evita XL(K083050)
Adu/Ped: 4:1 to 1:104:1 to 1:10Same
I:E range
Neo: 4:1 to 1:10-Different, same as secondary predicate device Servo-U(K180098)
Adu/Ped: 0 to 2 s0 to 2 sSame
Tslope range
Neo: 0 to 2 s-Different, same as secondary predicate device Evita XL(K083050)
Adu: 1 to 100 cmH2O Note: when backup air supply works, can set to up to 80cmH2OAdu: 1 to 100 cmH2OSame
△Pinsp range
Ped: 1 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2OPed: 1 to 95 cmH2O
Neo: 1 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2O-Different, similar to secondary predicate device Evita XL(K083050)
Adu/Ped: 0 to 50 cmH2O0 to 50 cmH2OSame
PEEP range
Neo: 0 to 50 cmH2O-Different, same as secondary predicate device Evita XL(K083050)
Adu: 0 to 100 cmH2OAdu: 0 to 100 cmH2OSame
△Psupp range

Page 11

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 7 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
Note: when backup air supply works, can set to up to 80cmH2O
Ped: 0 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2OPed: 0 to 95 cmH2OSame
Neo: 0 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2O-Different, same as secondary predicate device Evita XL(K083050)
Phigh range
Adu/Ped: 0 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2O0 to 95 cmH2OSame
Neo: 0 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2O-Different, same as secondary predicate device Evita XL(K083050)
Plow range
Adu/Ped: 0 to 50 cmH2O0 to 50 cmH2OSame
Neo: 0 to 50 cmH2O-Different, same as secondary predicate device Evita XL(K083050)
Thigh range
Adu/Ped: 0.1 to 30 s0.1 to 30 sSame
Neo: 0.1 to 30 s-Different, same as secondary predicate device Evita XL(K083050)
Tlow range
Adu/Ped: 0.2 to 30 s0.1 to 30 sSame
Neo: 0.2 to 30 s-Different, similar to secondary predicate device Evita XL(K083050)

Page 12

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 8 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
Note: when backup air supply works, can set to up to 80cmH2O
Ped: 0 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2OPed: 0 to 95 cmH2OSame
Neo: 0 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2O-Different, same as secondary predicate device Evita XL(K083050)
Adu/Ped: 0 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2O0 to 95 cmH2OSame
Phigh range
Neo: 0 to 95 cmH2O Note: when backup air supply works, can set to up to 80cmH2O-Different, same as secondary predicate device Evita XL(K083050)
Adu/Ped: 0 to 50 cmH2O0 to 50 cmH2OSame
Plow range
Neo: 0 to 50 cmH2O-Different, same as secondary predicate device Evita XL(K083050)
Adu/Ped: 0.1 to 30 s0.1 to 30 sSame
Thigh range
Neo: 0.1 to 30 s-Different, same as secondary predicate device Evita XL(K083050)
Adu/Ped: 0.2 to 30 s0.1 to 30 sSame
Tlow range
Neo: 0.2 to 30 s-Different, similar to secondary predicate device Evita XL(K083050)

Page 13

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 9 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
Tpause(%) rangeAdu/Ped: OFF, 5% to 60%OFF, 5% to 60%Same
Flow range
Adu: 6 to 180 L/minAdu: 6 to 180 L/minSame
Ped: 6 to 30 L/minPed: 6 to 30 L/min
F-Trig range
Adu/ Ped: 0.5 to 15 L/min0.5 to 15 L/minSame
Neo: 0.1 to 5 L/min-Different, similar to secondary predicate device ServoU(K180098)
P-Trig range
Adu/ Ped: -20 to -1 cmH2O-20 to -1 cmH2OSame
Neo: -20 to -1 cmH2O-Different, same as secondary predicate device ServoU(K180098)
Exp% range
Adu/ Ped:1 to 70%1 to 70%Same
Neo: 1 to 70%-Different, same as secondary predicate device ServoU(K180098)
MV% range25 to 350%25 to 350%Same
Specifications - Ventilator monitoring parameter
TV range0 to 6000 ml0 to 6000 mlSame
TVspn range0 to 6000 ml0 to 6000 mlSame
MV range0 to 100 L/min0 to 100 L/minSame
MVspn range0 to 100 L/min0 to 100 L/minSame
MVleak range0 to 100 L/min0 to 100 L/minSame

Page 14

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 10 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
FiO215 to 100 vol.%15 to 100 vol.%Same
Airway pressure (Ppeak, Pplat, Pmean) range-20 to 120 cmH2O-20 to 120 cmH2OSame
PEEP range0 to 120 cmH2O0 to 120 cmH2OSame
ftotal/fspn/fmand range0 to 200 /min0 to 200 /minSame
Tinsp range0 to 60 s0 to 60 sSame
I:E range99:1 to 1:9999:1 to 1:99Same
Leak% range0 to 100%0 to 100%Same
RCexp range0 to 10 s0 to 10 sSame
RSBI range0 to 9999 1/(L•min)0 to 9999 1/(L•min)Same
WOBtot/ WOBvent/ WOBpat range0 to 20 J/L0 to 20 J/LSame
Ri range0 to 600 cmH2O/( L/s)0 to 600 cmH2O/( L/s)Same
Re range0 to 600 cmH2O/( L/s)0 to 600 cmH2O/( L/s)Same
Cstat range0 to 300 ml/cmH2O0 to 300 ml/cmH2OSame
Cdyn range0 to 300 ml/cmH2O0 to 300 ml/cmH2OSame
TVe/IBW range0 to 50 mL/kg0 to 50 mL/kgSame
NIF range-45 to 0 cmH2O-45 to 0 cmH2OSame
P0.1 range-20 to 0 cmH2O-20 to 0 cmH2OSame
PEEPi range0 to 80 cmH2O0 to 80 cmH2OSame
PIF range0 to 300 L/min0 to 300 L/minSame
PEF range0 to 180 L/min0 to 180 L/minSame
Specifications - O2 Therapy
Flow range
Adu: 10 to 60 L/min-Different, similar to secondary predicate device ServoU(K180098)
Ped: 2 to 25L/min-Different, similar to secondary predicate device ServoU(K180098)
Neo: 2 to 20 L/min-Different, similar to secondary predicate device

Page 15

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 11 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
device Servo-U (K180098)
Specifications - Auxiliary Pressure
PesI/PesE/ Paux2I/Paux2E range-40 to 120 cmH2O-40 to 120 cmH2OSame
PtpI/PtpE range-99 to 99 cmH2O-99 to 99 cmH2OSame
ΔPes range-99 to 99 cmH2O-99 to 99 cmH2OSame
Specifications - ATRC
Tube I.D.2.5 to 12 mm2.5 to 12 mmSame
Compensate1 to 100%1 to 100%Same
Specifications - Sigh
TypePressure sighPressure sighSame
△int.PEEP rangeOFF,1 to 40 cmH2OOFF,1 to 40 cmH2OSame
Specifications – Apnea Ventilation
Apnea VentilationYesYesSame
Specifications - Intellicycle
IntellicycleYesYesSame
Specifications - Lung Recruitment (SI)
Pressure Hold20 to 60 cmH2O20 to 60 cmH2OSame
Hold Time10 to 40 s10 to 40 s
Specifications – Static PV Loop
Pstart0 to 50 cmH2O0 to 50 cmH2OSame
Flow4 to 15 L/min4 to 15 L/minSame
Pmax1 to 80 cmH2O1 to 80 cmH2OSame
Vlimit100 to 2000 ml100 to 2000 mlSame
Specifications - Weaning Tools (SBT)
TypePSV ventilationPSV ventilationSame
SBT CriteriaApnea TimeApnea TimeSame

Page 16

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 12 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
High and low MV alarm High and low f alarmHigh and low MV alarm High and low f alarmdevice Servo-U (K180098)
Specifications - Other Functions
Manual BreathYesYesSame
Expiration HoldYesYesSame
Inspiration HoldYesYesSame
NebulizationYesYesSame
O2↑ (Oxygen Enrichment)YesYesSame
SuctionYesYesSame
Intrinsic PEEPYesYesSame
Dynamic lung (PulmoSight)YesYesSame
Specifications - Basic Alarm
High/Low Minute Volume AlarmYesYesSame
High/Low Tidal Volume AlarmYesYesSame
High/Low f AlarmYesYesSame
High/Low FiO2 AlarmYesYesSame
High/Low Airway Pressure AlarmYesYesSame
Pressure Limiting AlarmYesYesSame
Continuous Airway Pressure AlarmYesYesSame
Apnea AlarmYesYesSame
Specifications - Gas supply
Pipeline supply280 to 650 kPa for O2, Air280 to 650 kPa for O2, AirSame
Connector typeDISS body complying with CGA V-5DISS body complying with CGA V-5Same
Specifications – Sidestream CO2 Module

Page 17

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 13 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
ModuleExternal 1-slot Sidestream CO2 Module 2.0External 1-slot Sidestream CO2 Module 2.0Same
Water trap
DRYLINE II Water trap, Adult/pediatricDRYLINE II Water trap, Adult/pediatricSame
DRYLINE II Water trap, Neonate-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
Sampling Line
DRYLINE Gas Sampling Line, Neonate 2.5 m-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
DRYLINE Gas Sampling Line, Adult/pediatric 2.5 mDRYLINE Gas Sampling Line, Adult/pediatric 2.5 mSame
Airway adapter
Dryline airway adapter, straightDryline airway adapter, straightSame
Airway AdapterAirway AdapterSame
Sample flowrate
Neonate water trap: 90 ml/min-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
Adult/pediatric water trap: 120 ml/minAdult/pediatric water trap: 120 ml/minSame
Measurement range0 mmHg to 150 mmHg0 mmHg to 150 mmHgSame
Measurement accuracy0 mmHg to 40 mmHg: ±2 mmHg 41 mmHg to 76 mmHg: ±5% of reading 77 mmHg to 99 mmHg: ±10% of reading 100 mmHg to 150 mmHg: ± (3mmHg + 8% of reading)0 mmHg to 40 mmHg: ±2 mmHg 41 mmHg to 76 mmHg: ±5% of reading 77 mmHg to 99 mmHg: ±10% of reading 100 mmHg to 150 mmHg: ± (3mmHg + 8% of reading)Same
Specifications – Mainstream CO2 Module

Page 18

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 14 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
ModuleExternal 1-slot Sidestream CO2 Module 2.0External 1-slot Sidestream CO2 Module 2.0Same
Water trap
DRYLINE II Water trap, Adult/pediatricDRYLINE II Water trap, Adult/pediatricSame
DRYLINE II Water trap, Neonate-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
Sampling Line
DRYLINE Gas Sampling Line, Neonate 2.5 m-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
DRYLINE Gas Sampling Line, Adult/pediatric 2.5 mDRYLINE Gas Sampling Line, Adult/pediatric 2.5 mSame
Airway adapter
Dryline airway adapter, straightDryline airway adapter, straightSame
Airway AdapterAirway AdapterSame
Sample flowrate
Neonate water trap: 90 ml/min-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
Adult/pediatric water trap: 120 ml/minAdult/pediatric water trap: 120 ml/minSame
Measurement range0 mmHg to 150 mmHg0 mmHg to 150 mmHgSame
Measurement accuracy0 mmHg to 40 mmHg: ±2 mmHg 41 mmHg to 76 mmHg: ±5% of reading 77 mmHg to 99 mmHg: ±10% of reading 100 mmHg to 150 mmHg: ± (3mmHg + 8% of reading)0 mmHg to 40 mmHg: ±2 mmHg 41 mmHg to 76 mmHg: ±5% of reading 77 mmHg to 99 mmHg: ±10% of reading 100 mmHg to 150 mmHg: ± (3mmHg + 8% of reading)Same
Specifications – Mainstream CO2 Module

Page 19

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 15 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
ModuleExternal 1-slot Mainstream CO2 ModuleExternal 1-slot Mainstream CO2 ModuleSame
SensorCAPNOSTAT CO2 sensorCAPNOSTAT CO2 sensorSame
Airway adapter
Airway adapter, Adult/pediatricAirway adapter, Adult/pediatricSame
Airway adapter, Neonate-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
Measurement range0 mmHg to 150 mmHg0 mmHg to 150 mmHgSame
Measurement accuracy0 mmHg to 40 mmHg: ±2 mmHg 41 mmHg to 70 mmHg: ±5% of reading 71 mmHg to 100 mmHg: ±8% of reading 101 mmHg to 150 mmHg: ±10% of reading0 mmHg to 40 mmHg: ±2 mmHg 41 mmHg to 70 mmHg: ±5% of reading 71 mmHg to 100 mmHg: ±8% of reading 101 mmHg to 150 mmHg: ±10% of readingSame
Specifications – Mindray SpO2 Module
ModuleExternal 1-slot Mindray SpO2 moduleExternal 1-slot Mindray SpO2 moduleSame
Cable7pin SpO2 Cable7pin SpO2 CableSame
SpO2 sensor
SpO2 sensor, finger-clip, adultSpO2 sensor, finger-clip, adultSame
SpO2 sensor, finger-clip, pediatricSpO2 sensor, finger-clip, pediatricSame
SpO2 sensor, foot, bandageSpO2 sensor, foot, bandageDifferent, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
SpO2 Measured parametersMeasurement range: 0 to 100% Accuracy:70 to 100%: ±2% (adult/pediatric mode) 0% to 69%: Not specifiedMeasurement range: 0 to 100% Accuracy:70 to 100%: ±2% (adult/pediatric mode)Same

Page 20

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 16 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
0% to 69%: Not specified
Measurement range: 0 to 100% Accuracy: 70 to 100%: ±3% (neonate mode) 0% to 69%: Not specified-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
PR Measured parametersMeasurement range: 20 to 254 1/min; Accuracy: ±3 1/min.Measurement range: 20 to 254 1/min; Accuracy: ±3 1/min.Same
PI Measured parametersRange: 0.05 to 20 %Range: 0.05 to 20 %Same
Specifications – Masimo SpO2 Module
ModuleExternal 1-slot Masimo SpO2 module-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
CableRD SET MD 14-05, PC 5 ft RD SET MD 14-12, Patient Cable 12 ft RD to LNC Adapter Cable LNCS to RD Adapter 8pin masimo Cable (RD SET) 8-pin Masimo SpO2 Extension Cable (MASIMOLNCS)-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
SpO2 sensorLNCS DC-I Reusable sensor LNCS DC-IP Reusable SpO2sensor LNCS Multisite Reusable sensor LNCS Neo/Pt disposable sensor LNCS Neo disposable sensor LNCS Inf disposable sensor LNCS Pdtx adhesive sensor LNCS-Adtx disposable sensor-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)

Page 21

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 17 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
0% to 69%: Not specified
Measurement range: 0 to 100% Accuracy: 70 to 100%: ±3% (neonate mode) 0% to 69%: Not specified-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
PR Measured parametersMeasurement range: 20 to 254 1/min; Accuracy: ±3 1/min.Measurement range: 20 to 254 1/min; Accuracy: ±3 1/min.Same
PI Measured parametersRange: 0.05 to 20 %Range: 0.05 to 20 %Same
Specifications – Masimo SpO2 Module
ModuleExternal 1-slot Masimo SpO2 module-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
CableRD SET MD 14-05, PC 5 ft RD SET MD 14-12, Patient Cable 12 ft RD to LNC Adapter Cable LNCS to RD Adapter 8pin masimo Cable (RD SET) 8-pin Masimo SpO2 Extension Cable (MASIMOLNCS)-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
SpO2 sensorLNCS DC-I Reusable sensor LNCS DC-IP Reusable SpO2sensor LNCS Multisite Reusable sensor LNCS Neo/Pt disposable sensor LNCS Neo disposable sensor LNCS Inf disposable sensor LNCS Pdtx adhesive sensor LNCS-Adtx disposable sensor-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)

Page 22

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 18 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
LNCS Adtx disposable sensor LNCS Pdtx disposable sensor LNCS DCI reusable sensor LNCS DCIP reusable sensor LNCS NeoPt disposable sensor LNCS Neo disposable sensor LNCS Inf disposable sensor RD SET DCI, Adult Reusable Sensor RD SET DCIP, Pediatric Reusable Sensor RD Set TC-I SpO2 Reusable Tip-Clip Ear Sensor, 3ft RD SET Adhesive Sensor RD SET PDT Adhesive Sensor RD Set Infant Adhesive Sensor RD Set Neo Adhesive Sensor RD Set NeoPt Adhesive Sensor RD Set NeoPt-500 Non-adhesive sensor
SpO2 Measured parametersRange: 1 to 100% Accuracy: 70% to 100%: Adult/Pediatric: ±2% (without motion) Adult/Pediatric: ±3% (with motion) Neonate: ±3% 1 to 69%: not declaration. Low Perfusion (Pulse amplitude: >0.02%, Light penetration: >5%): Accuracy: ±2%-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
PR Measured parametersRange: 25 to 240 1/min Accuracy:-Different, same as secondary predicate

Page 23

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 19 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
±3 1/min (without motion) ±5 1/min (with motion)device BeneVision N Series Patient Monitors (K213799)
PI Measured parametersRange: 0.02 to 20 %-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
Specifications – Nellcor SpO2 Module
ModuleExternal 1-slot Nellcor SpO2 module-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
Cable8-pin, SpO2 extension cable, 2.5m-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
SpO2 sensorReusable SpO2 sensor, finger-clip, Adult Reusable Pediatric-Infant SpO2 sensor with wraps, Ped/Infant Reusable Adult/neonatal SpO2 sensor, adhesive wraps, Adult/Neo Nellcor MAX-A Disposable SpO2 sensor Nellcor MAX-P Disposable SpO2 sensor Nellcor MAX-I Disposable SpO2 sensor Nellcor MAX-N Disposable SpO2 sensor Nellcor reusable D-YS SpO2 sensor with wraps-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
SpO2 Measured parametersRange: 0 to 100% Accuracy:-Different, same as secondary predicate device BeneVision N

Page 24

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 20 of 19

Technical CharacteristicsSubject device SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Subject device)Primary predicate SV600, SV800 Ventilator Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (K220107)Comments
70% to 100%: Adult/Pediatric: ±2% Neonate: ±3% 0 to 69%: not declaration.Series Patient Monitors (K213799)
PR Measured parametersRange: 20 to 300 1/min Accuracy: 20 to 250 1/min, ±3 1/min; 251 to 300 1/min: No declaration-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
PI Measured parametersno declaration-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)
WiFi moduleSX-SDMAC-2832S+-Different, same as secondary predicate device BeneVision N Series Patient Monitors (K213799)

"-" means not applicable.

In conclusion, the differences in technological characteristics do not raise new questions of safety and effectiveness.

7. PERFORMANCE DATA

To establish the substantial equivalence of the SV600 and SV800 Ventilators, Mindray conducted functional and system level testing on the subject device. The testing provided an evaluation of the performance of the device relevant to each of the differences between the subject device and the predicate device. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate. Mindray has conducted testing to ensure the subject device meets relevant consensus standards. Mindray also conducted human factors testing to demonstrate that the device is safe and effective for the intended users, uses and use environments.

Biocompatibility Testing

Biocompatibility testing of the SV600, SV800 Ventilator was performed using the following standards:

Page 25

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 21 of 19

  • ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
  • ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
  • ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
  • ISO 10993-17 Second edition 2023-09 Biological evaluation of medical devices Part 17: Toxicological risk assessment of medical device constituents
  • ISO 10993-18 Second edition 2020-01 Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process
  • ISO 18562-1 First edition 2017-03 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
  • ISO 18562-2 First edition 2017-03 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter
  • ISO 18562-3 First edition 2017-03 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds

Software Verification and Validation Testing

Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of the SV600 and SV800 Ventilators was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product.

Electromagnetic Compatibility and Electrical Safety

EMC and Electrical Safety testing was performed using the following standards::

  • IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005 MOD) [Including Amendment 2 (2021)]
  • IEC 60601-1-2 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and

Page 26

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 22 of 19

essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

  • IEEE ANSI USEMCSC C63.27-2021American National Standard for Evaluation of Wireless Coexistence
  • IEC/TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
  • IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 60601-1-8 Edition 2.2 2020-07 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
  • ISO 80601-2-12 Second edition 2020-02 Medical electrical equipment - Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators
  • ISO 80601-2-55 Second edition 2018-02 Medical electrical equipment - Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors
  • ISO 80601-2-61 Second edition 2017-12 (Corrected version 2018-02) Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment

Bench Testing

To establish the substantial equivalence of the SV600 and SV800 Ventilators, Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification and is substantially equivalent to the predicate device.

In addition, Mindray has demonstrated device performance using the following standards:

  • IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 60601-1-8 Edition 2.2 2020-07 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

Page 27

Traditional 510(k): SV600, SV800 Ventilator

K243767
510(k) Summary
Page 23 of 19

  • ISO 80601-2-12 Second edition 2020-02 Medical electrical equipment - Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators
  • ISO 80601-2-55 Second edition 2018-02 Medical electrical equipment - Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors
  • ISO 80601-2-61 Second edition 2017-12 (Corrected version 2018-02) Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
  • ASTM F1100-90 (Reapproved 1997) Standard Specification for Ventilator Intended for Use in Critical Care (only Endurance Testing)

8. CONCLUSION

Based on the detailed comparison of specifications for each of the characteristics to the predicate devices, the performance testing and conformance with applicable standards, the SV600 and SV800 Ventilators can be found substantially equivalent to the predicate device.

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).