K Number
K243759
Date Cleared
2024-12-20

(14 days)

Product Code
Regulation Number
866.3365
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BIOFIRE Blood Culture Identification 2 (BCID2) Panel is a multiplexed nucleic acid test intended for use with BIOFIRE FILMARRAY 2.0 or BIOFIRE FILMARRAY TORCH Systems for the simultaneous qualitative detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants associated with antimicrobial resistance. The BIOFIRE BCID2 Panel test is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results. The following organism types and subtypes are identified using the BIOFIRE BCID2 Panel: Gram Positive Bacteria Enterococcus faecalis Staphylococcus spp. Streptococcus spp. Enterococcus faecium Staphylococcus aureus Streptococcus agalactiae (Group B) Listeria monocytogenes Staphylococcus epidermidis Streptococcus pneumoniae Staphylococcus lugdunensis Streptococcus pyogenes (Group A) Gram Negative Bacteria Acinetobacter calcoaceticus-baumannii complex Enterobacterales Bacteroides fragilis Enterobacter cloacae complex Haemophilus influenzae Escherichia coli Neisseria meningitidis (encapsulated) Klebsiella aerogenes Pseudomonas aeruginosa Klebsiella oxytoca Stenotrophomonas maltophilia Klebsiella pneumoniae group Proteus spp. Salmonella spp. Serratia marcescens Yeast Candida albicans Candida krusei Cryptococcus neoformans/gattii Candida auris Candida parapsilosis Candida glabrata Candida tropicalis The BIOFIRE BCID2 Panel contains assays for the detection of genetic determinants associated with resistance to methicillin (mecA/C and mecA/C in conjunction with MREJ, vancomycin (vanA and vanB), ß-lactams including penicillins, cephalosporins, monobactams, and carbapenems (blaCTX-M, blaIMP, blaKPC, blaNDM, blaOXA48-like, blaVIM) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In addition, the panel includes an assay for the mobilized genetic determinant mor-1, an emerging marker of public health importance. The antimicrobial resistance gene or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, b-lactams, and colistin exist. Antimicrobial Resistance Genes CTX-M KPC mecA/C NDM vanA/B IMP mcr-1 mecA/C and MREJ (MRSA) OXA-48-like VIM The BIOFIRE BCID2 Panel is indicated as an aid in the diagnosis of bloodstream infection and results should be used in conjunction with other clinical and laboratory findings. Positive results do not rule out co-infection with organisms not included in the BIOFIRE BCID2 Panel. The BIOFIRE BCID2 Panel is not intended to monitor treatment for bloodstream infection. Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the BIOFIRE BCID2 Panel, and for determination of species detected but not identified within complexes, groups, or genera by the BIOFIRE BCID2 Panel assays.

Device Description

The BIOFIRE Blood Culture Identification 2 (BCID2) Panel is designed to simultaneously identify 43 bacteria and yeast responsible for bloodstream infections, as well as select genetic determinants of antimicrobial resistance (see Table 1), in a timeframe (about an hour) that allows the test results to be used in determining appropriate patient treatment and management. The BIOFIRE BCID2 Panel is performed directly on positive blood culture samples. The BIOFIRE BCID2 Panel is compatible with BioFire's PCR-based in vitro diagnostic BIOFIRE FILMARRAY 2.0 and FILMARRAY TORCH systems for infectious disease testing. A specific software module (i.e., BIOFIRE BCID2 Panel pouch module) is used to perform BIOFIRE BCID2 Panel testing on these systems. A test is initiated by loading Hydration Solution into one port of the BIOFIRE pouch and positive blood culture specimen mixed with the provided Sample Buffer into the other port of the BIOFIRE BCID2 Panel pouch and placing it in a BIOFIRE System. The pouch contains all the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the BIOFIRE Software guides the user though the steps of placing the the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run. The BIOFIRE System contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis. Nucleic acid extraction occurs within the BIOFIRE pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the BIOFIRE system performs a nested multiplex PCR that is executed in two stages. During the first stage, the BIOFIRE System performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in single plex fashion in each well of the array. At the end of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data. The BIOFIRE Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information based on the provided FDA 510(k) summary for the BioFire Blood Culture Identification 2 (BCID2) Panel:

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily focuses on a software update to mitigate a risk of false negative results for C. tropicalis specifically. It states that the re-analysis with the modified software led to a "very minor adjustment to clinical sensitivity for S. epidermidis from 96.5% [93.0-98.2%] to 96.9%".

The summary does not provide a comprehensive table of acceptance criteria and the overall device performance for all targets. Instead, it highlights the impact of the software update on C. tropicalis detection and the resulting minor change in S. epidermidis sensitivity.

Acceptance Criteria (Implied / Pre-established for Original Device):

While not explicitly stated as "acceptance criteria" in this specific document, the overall context of a 510(k) clearance implies that the device must demonstrate comparable performance to its predicate. The key focus of this Special 510(k) is to show that the software update does not adversely affect, and in fact, improves, the device's performance, particularly concerning the false negative C. tropicalis issue. The acceptance criterion for this update seems to be that the corrected software resolves the C. tropicalis issue and does not significantly degrade performance on other targets.

Target Organism / MetricPre-update Reported Performance (where available)Post-update Reported Performance (where available)Implied Acceptance Criteria (for the update)
C. tropicalis detectionIntermittent false negatives with control M416Resolved (no false negatives with control M416)Elimination of intermittent false negatives for C. tropicalis with external control M416.
S. epidermidis sensitivity96.5% [93.0-98.2%]96.9%Maintain or slightly improve sensitivity, without degradation.
Overall clinical performanceCleared in K193519 (predicate)No overall change in study conclusions or claimsPerformance remains "as well as the predicate device."

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: The document refers to "reanalysis of the performance data (clinical and non-clinical studies) with the modified pouch module software." It explicitly mentions that the C. tropicalis issue was identified during testing of Maine Molecular Quality Controls, Inc. (MMQC) FilmArray BCID2 Control Panel M416. While the specific number of tests performed on this control or other test sets with the modified software is not given in this summary, the "reanalysis of the performance data" implies that the original clinical and non-clinical study data from the predicate device (K193519) was re-evaluated using the new software.
  • Data Provenance: The issue was identified through testing with an external control material (C. tropicalis synthetic control material). The "reanalysis of the performance data (clinical and non-clinical studies)" implies that data from the original (K193519) approval was used, which would have primarily involved prospective clinical samples from various geographic locations. However, this specific document does not detail the provenance of the original clinical data. The new data generated to address the C. tropicalis issue was from synthetic control material.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

  • This information is not provided in the given FDA summary. The document focuses on a software update to an already cleared device, addressing a technical issue related to melting temperature analysis. The ground truth for the original clinical studies would have been established through a combination of clinical methods (e.g., standard microbiological culture and identification, susceptibility testing), but those details are not part of this specific Special 510(k) summary.

4. Adjudication Method for the Test Set:

  • This information is not provided in the given FDA summary. For the original clinical studies related to the predicate device, an adjudication method (e.g., expert panel review of discrepancies) would likely have been employed, but this specific document does not detail it.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

  • This is not applicable. The device (BIOFIRE Blood Culture Identification 2 Panel) is a fully automated multiplex nucleic acid test. It does not involve human readers interpreting results in the same way an imaging AI device would. Therefore, an MRMC study with human readers and AI assistance is not relevant to this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes. The BIOFIRE BCID2 Panel operates as a standalone diagnostic system. The "pouch module software" automatically interprets the melt curve analysis and combines data with internal controls to generate a test result. The description explicitly states, "The BIOFIRE Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel." There is no "human-in-the-loop" performance component for result generation, although clinical interpretation of results in conjunction with Gram stain and other findings is required.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

  • For the specific issue addressed in this Special 510(k) regarding C. tropicalis, the ground truth was known composition of synthetic control material.
  • For the broader re-analysis of "clinical and non-clinical studies" (referring to the original predicate data), the ground truth for microbial identification in blood cultures typically involves standard microbiological culture (subculturing) and definitive identification methods (e.g., biochemical tests, mass spectrometry, sequencing), often supplemented by clinical context. The document emphasizes that "Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the BIOFIRE BCID2 Panel, and for determination of species detected but not identified within complexes, groups, or genera by the BIOFIRE BCID2 Panel assays," highlighting the role of traditional microbiology as a gold standard.

8. The Sample Size for the Training Set:

  • This information is not provided in the given FDA summary, as the nature of this submission is a software update to an existing device, not a de novo algorithm development. The "pouch module software" utilizes pre-defined melting temperature (Tm) ranges and analysis parameters. The "reanalysis to include data generated from synthetic control materials" effectively served to refine or "retrain" specific parameters (Tm ranges) within the existing analytical framework, but the original training set size for the entire panel's algorithm is not detailed here.

9. How the Ground Truth for the Training Set Was Established:

  • This information is not provided in the given FDA summary. For the original development of the BIOFIRE BCID2 Panel, the ground truth for establishing the algorithms' detection parameters and Tm ranges would have been based on extensive analytical studies using well-characterized microbial isolates (strains with confirmed identities) at various concentrations, likely identified by definitive culture-based methods and sequencing. The update specifically states that "Data from the synthetic control material(s) had not been included in the establishment and validation of the assay Tm ranges used for analysis by the pouch module software," indicating that the "training" for the C. tropicalis Tm range was incomplete previously and was corrected by incorporating this synthetic control material data.

{0}------------------------------------------------

December 20, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a caduceus, a traditional symbol of medicine. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The word "FDA" is in a larger font size than the rest of the text.

BioFire Diagnostics, LLC Karli Plenert Sr. Director, Regulatory Affairs 515 Colorow Drive Salt Lake City, Utah 84108

Re: K243759

Trade/Device Name: BIOFIRE Blood Culture Identification 2 (BCID2) Panel (RFIT-ASY-0147 (30 pack) RFIT-ASY-0148 (6 pack)) Regulation Number: 21 CFR 866.3365 Regulation Name: Multiplex Nucleic Acid Assay For Identification Of Microorganisms And Resistance Markers From Positive Blood Cultures Regulatory Class: Class II Product Code: PEN, PAM, PEO Dated: December 5, 2024 Received: December 6, 2024

Dear Karli Plenert:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Bryan M. Bryan M. Grabias -S Grabias --> 10:36:58 -05'00'

Bryan Grabias, Ph. D. Acting Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243759

Device Name

BIOFIRE Blood Culture Identification 2 (BCID2) Panel (RFIT-ASY-0147 (30 pack) RFIT-ASY-0148 (6 pack))

Indications for Use (Describe) The BIOFIRE Blood Culture Identification 2 (BCID2) Panel is a multiplexed nucleic acid test intended for use with BIOFIRE FILMARRAY 2.0 or BIOFIRE FILMARRAY TORCH Systems for the simultaneous qualitative detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants associated with antimicrobial resistance. The BIOFIRE BCID2 Panel test is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results. The following organism types and subtypes are identified using the BIOFIRE BCID2 Panel: Gram Positive Bacteria Enterococcus faecalis Staphylococcus spp. Streptococcus spp. Enterococcus faecium Staphylococcus aureus Streptococcus agalactiae (Group B) Listeria monocytogenes Staphylococcus epidermidis Streptococcus pneumoniae Staphylococcus lugdunensis Streptococcus pyogenes (Group A) Gram Negative Bacteria Acinetobacter calcoaceticus-baumannii complex Enterobacterales Bacteroides fragilis Enterobacter cloacae complex Haemophilus influenzae Escherichia coli Neisseria meningitidis (encapsulated) Klebsiella aerogenes Pseudomonas aeruginosa Klebsiella oxytoca Stenotrophomonas maltophilia Klebsiella pneumoniae group Proteus spp. K243759 Lead Memo BioFire Diagnostics,... BIOFIRE Blood Cultur... Page 5 of 24 Salmonella spp. Serratia marcescens Yeast Candida albicans Candida krusei Cryptococcus neoformans/gattii Candida auris Candida parapsilosis Candida glabrata Candida tropicalis The BIOFIRE BCID2 Panel contains assays for the detection of genetic determinants associated with resistance to methicillin (mecA/C and mecA/C in conjunction with MREJ, vancomycin (vanA and vanB), ß-lactams including penicillins, cephalosporins, monobactams, and carbapenems (blaCTX-M, blaIMP, blaKPC, blaNDM, blaOXA48-like, blaVIM) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In addition, the panel includes an assay for the mobilized genetic determinant mor-1, an emerging marker of public health importance. The antimicrobial resistance gene or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, b-lactams, and colistin exist. Antimicrobial Resistance Genes CTX-M KPC mecA/C NDM vanA/B

IMP mcr-1 mecA/C and MREJ (MRSA) OXA-48-like VIM

The BIOFIRE BCID2 Panel is indicated as an aid in the diagnosis of bloodstream infection and results should be used in conjunction with other clinical and laboratory findings. Positive results do not rule out co-infection with organisms not included in the BIOFIRE BCID2 Panel. The BIOFIRE BCID2 Panel is not intended to monitor treatment for bloodstream infection.

Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the BIOFIRE BCID2 Panel, and for

{4}------------------------------------------------

determination of species detected but not identified within complexes, groups, or genera by the BIOFIRE BCID2 Panel assays.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a yellow-green gradient bottom half. The word "BIOMÉRIEUX" is written in white, sans-serif font in the center of the blue portion of the circle.

BIOFIRE® FILMARRAY® BIOFIRE Blood Culture Identification 2 (BCID2) Panel Special 510(k) Summary BioFire Diagnostics, LLC

Introduction:

The content of this Special 510(k) submission is limited to obtaining FDA clearance for the BIOFIRE Blood Culture Identification 2 (BCID2) Panel (K193519) with a software update to mitigate a risk of false negative results.

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitted by:

BioFire Diagnostics. LLC (bioMérieux) 515 Colorow Drive Salt Lake City, UT 84108

Contact: Karli Plenert, MBA Telephone: +1 385-414-4985 Email: Karli.Plenert@biomerieux.com

Date Submitted: December 5, 2024

Device Name and Classification:

Trade Name: BIOFIRE® FILMARRAY®Blood Culture Identification 2 (BCID2) Panel

Regulation Number: 21 CFR 866.3365

Classification Name: Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.

Predicate Devices:

K193519 - BIOFIRE Blood Culture Identification 2 (BCID2) Panel

Background:

In October of 2023, bioMérieux initiated a Field Safety Corrective Action (FDA Recall Event # 93360) due to reports of unexpected/false negative (Not Detected) results for C. tropicalis on the BIOFIRE BCID2 Panel when testing Maine Molecular Quality Controls, Inc. (MMQC) FilmArray BCID2 Control Panel M416, an external control material. In some portion of test runs using the MMQCI material, the C. tropicalis synthetic control material is amplified but reported as 'Not Detected' because the melting temperature (Tm) of the amplicon is above the Tm limit assigned to the assay.

{6}------------------------------------------------

Investigation determined two root causes for the intermittent false negative control results: 1) a shift to higher amplicon Tm associated with instrument calibration (FSCA 5788-1) and 2) minor incompatibility between the Tm range assigned to the Ctropicalis assay and the Tm of the amplicon generated by the synthetic control material. Data from the synthetic control material(s) had not been included in the establishment and validation of the assay Tm ranges used for analysis by the pouch module software.

Mitigation of the risk of false negative C. tropicalis results from the control material includes correction of instrument calibration and revision of the BCID2 Panel pouch module software to the Tm range of the Ctropicalis (and other) assays that incorporates the appropriate tolerance to the Tm of synthetic control material sequences.

Summary of Changes

Software Update:

bioMérieux has initiated a software (pouch module) update to adjust the Tm range for a subset of assays following re-analysis to include data generated from synthetic control materials. The revised Tm range(s) miligate the risk of false negative results associated with amplicon Tm.

Reanalysis of the performance data (clinical and non-clinical studies) with the modified pouch module software led to a change in sensitivity for S. epidermidis from 96.5% [93.0-98.2%] to 96.9%], as documented in the updated BIOFIRE BCID2 instructions for use. Re-analysis did not result in an overall change of the study conclusions or performance claims for the panel.

The pouch module revision also includes the addition of the following note to the results report:

Note: All BIOFIRE BCID2 Panel results are intended to be interpreted in conjunction with Gram stain results. In some cases, the Gram stain result and the BIOFIRE BCID2 Panel result may be discrepant. In these cases, the BIOFIRE BCID2 Panel results should be confirmed, e.g., by culture or other laboratory, epidemiological, or clinical findings. Blood culture media may contain non-viable organisms and/or nucleic acids that may lead to false positive BIOFIRE BCID2 Panel results.

Typically, these false positives present with more than one positive result from the BIOFIRE BCID2 Panel.

Instructions for Use Update:

The instructions for use (IFU) for the BIOFIRE BCID2 Panel has been updated to reflect the very minor adjustment to clinical sensitivity for S. epidermidis due to the pouch module update and reanalysis. Additional updates are included for organism descriptions as well as analytical reactivity/specificity tables and descriptions that are driven by post-market monitoring and sequence surveillance programs. Changes include taxonomic updates (new species identification and/or re-classification) as well as adjustments to sequencebased reactivity predictions based on monitoring of available sequence data.

Intended Use:

The BIOFIRE® Blood Culture Identification 2 (BCID2) Panel is a multiplexed nucleic acid test intended for use with BIOFIRE® FILMARRAY® 2.0 or BIOFIRE® FILMARRAY® TORCH Systems for the simultaneous qualitative detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants associated with antimicrobial resistance. The BIOFIRE BCID2 Panel test is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results. The following organism types and subtypes are identified using the BIOFIRE BCID2 Panel:

Gram Positive Bacteria
Enterococcus faecalisStaphylococcus spp.Streptococcus spp.
Enterococcus faeciumStaphylococcus aureusStreptococcus agalactiae (Group B)
Listeria monocytogenesStaphylococcus epidermidisStreptococcus pneumoniae
BIOFIRE® FILMARRAY® Blood Culture Identification 2 PanelSpecial 510(k) SubmissionDecember 5, 2024BioFire Diagnostics, LLC

{7}------------------------------------------------

Staphylococcus lugdunensisStreptococcus pyogenes (Group A)
Gram Negative Bacteria
Acinetobacter calcoaceticus-baumannii complexEnterobacterales
Bacteroides fragilisEnterobacter cloacae complex
Haemophilus influenzaeEscherichia coli
Neisseria meningitidis (encapsulated)Klebsiella aerogenes
Pseudomonas aeruginosaKlebsiella oxytoca
Stenotrophomonas maltophiliaKlebsiella pneumoniae group
Proteus spp.
Salmonella spp.
Serratia marcescens
Yeast
Candida albicansCandida kruseiCryptococcus neoformans/gattii
Candida aurisCandida parapsilosis
Candida glabrataCandida tropicalis

The BIOFIRE BCID2 Panel contains assays for the detection of genetic determinants associated with resistance to methicillin (mecA/C and mecA/C in conjunction with MREJ), vancomycin (vanA and vanB), (S-lactams including penicillins, cephalosporins, monobactams, and carbapenems (ผลСТХ-М, ыЗМРС, №№М, №ОХААЗ-like, ble VIM) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In addition, the panel includes an assay for the detection of the mobilized genetic determinant mcr-1, an emerging marker of public health importance. The antimicrobial resistance gene or marker detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of resistance to methicililin, vancomycin, ß-lactams, and colistin exist.

Antimicrobial Resistance Genes
CTX-MKPCmecA/CNDMvanA/B
IMPmcr-1mecA/C and MREJ (MRSA)OXA-48-likeVIM

The BIOFIRE BCID2 Panel is indicated as an aid in the diagnosis of bloodstream infection and results should be used in conjunction with other clinical and laboratory findings. Positive results do not rule out co-infection with organisms not included in the BIOFIRE BCID2 Panel. The BIOFIRE BCID2 Panel is not intended to monitor treatment for bloodstream infection.

Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the BIOFIRE BCID2 Panel, and for determination of species detected but not identified within complexes, groups, or genera by the BIOFIRE BCID2 Panel assays.

Device Description:

The BIOFIRE Blood Culture Identification 2 (BCID2) Panel is designed to simultaneously identify 43 bacteria and yeast responsible for bloodstream infections, as well as select genetic determinants of antimicrobial resistance (see Table 1), in a timeframe (about an hour) that allows the test results to be used in determining appropriate patient treatment and management. The BIOFIRE BCID2 Panel is performed directly on positive blood culture samples.

{8}------------------------------------------------

The BIOFIRE BCID2 Panel is compatible with BioFire's PCR-based in vitro diagnostic BIOFIRE FILMARRAY 2.0 and FILMARRAY TORCH systems for infectious disease testing. A specific software module (i.e., BIOFIRE BCID2 Panel pouch module) is used to perform BIOFIRE BCID2 Panel testing on these systems.

A test is initiated by loading Hydration Solution into one port of the BIOFIRE pouch and positive blood culture specimen mixed with the provided Sample Buffer into the other port of the BIOFIRE BCID2 Panel pouch and placing it in a BIOFIRE System. The pouch contains all the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the BIOFIRE Software guides the user though the steps of placing the the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The BIOFIRE System contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the BIOFIRE pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the BIOFIRE system performs a nested multiplex PCR that is executed in two stages. During the first stage, the BIOFIRE System performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in single plex fashion in each well of the array. At the end of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

The BIOFIRE Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

Device Comparison:

Table 2 outlines the similarities and differences between the modified panels and the predicate panels.

ElementPredicate:BioFire Blood Culture Identification 2(BCID2) Panel (K193519)CurrentBioFire Blood Culture Identification 2(BCID2) Panel with software update
Intended UseThe BIOFIRE® Blood Culture Identification 2(BCID2) Panel is a multiplexed nucleic acid testintended for use with BIOFIRE® FILMARRAY®2.0 or BIOFIRE® FILMARRAY® TorchSystems for the simultaneous qualitativedetection and identification of multiple bacterialand yeast nucleic acids and select geneticdeterminants associated with antimicrobialresistance. The BIOFIRE BCID2 Panel test isperformed directly on blood culture samplesidentified as positive by a continuousmonitoring blood culture system. Results areintended to be interpreted in conjunction withGram stain results.Same
Specimen TypeBlood culture samples identified as positive bya continuous monitoring blood culture system.Same
Predicate:Current
ElementBioFire Blood Culture Identification 2(BCID2) Panel (K193519)BioFire Blood Culture Identification 2(BCID2) Panel with software update
OrganismsDetectedGram-positive Bacteria• Enterococcus faecalis• Enterococcus faecium• Listeria monocytogenes• Staphylococcus spp.• Staphylococcus aureus• Staphylococcus epidermidis• Staphylococcus lugdunensis• Streptococcus spp.• Streptococcus agalactiae (Group B)• Streptococcus pneumoniae• Streptococcus pyogenes (Group A)Gram-negative Bacteria• Acinetobacter calcoaceticus-baumanniicomplex• Enterobacterales• Bacteroides fragilis• Enterobacter cloacae complex• Haemophilus influenza• Escherichia coli• Neisseria meningitidis (encapsulated)• Klebsiella aerogenes• Pseudomonas aeruginosa• Klebsiella oxytoca• Stenotrophomonas maltophilia• Klebsiella pneumoniae group• Proteus spp.• Salmonella spp.• Serratia marcescensYeast• Candida albicans• Candida krusei• Cryptococcus neoformans/gattii• Candida auris• Candida parapsilosis• Candida glabrata• Candida tropicalisAntimicrobial Resistance Genes• СТХ-М• KPC• mecA/C• NDM• vanA/B• IMP• mcr-1• mecA/C and MREJ (MRSA)• OXA-48-like• VIMSame
AnalyteDNASame
TechnologicalPrinciplesHighly-multiplexed nested nucleic acidamplification test with melt analysisSame
InstrumentationBIOFIRE FILMARRAY 2.0 or BIOFIREFILMARRAY TORCHSame
Time to resultAbout 1 hourSame
Reagent StorageRoom temperatureSame
SamplePreparationMethodSample Processing is automated in theBIOFIRE System.Same
Shelf-Life12 months from Date of ManufactureSame
Test InterpretationAutomated test interpretation and reportgeneration. User cannot access raw data.Same
ElementPredicate:BioFire Blood Culture Identification 2(BCID2) Panel (K193519)CurrentBioFire Blood Culture Identification 2(BCID2) Panel with software update
ControlsTwo controls are included in each reagentpouch to control for sample processing andboth stages of PCR and melt analysis.Same
User ComplexityModerate/LowSame

Table 1. Comparison of the BioFire BCID2 Panel and the BioFire BCID2 Panel with Software Update

{9}------------------------------------------------

{10}------------------------------------------------

Conclusion:

The software modification and update to the labeling (instructions for use) do not affect the fundamental scientific technology, overall performance claims, or risk of the BIOFIRE BCID2 Panel. Therefore, the modified BIOFIRE BCID2 Panel performs as well as the predicate device.

§ 866.3365 Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.

(a)
Identification. A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.(b)
Classification. Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).