(20 days)
Not Found
No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and the description focuses on mechanical and sensor performance, reprocessing, and simulated surgical use cycles.
No.
The device is a surgical system that assists in controlling instruments for various surgical procedures, rather than directly treating a disease or condition itself.
No
The device is described as assisting in the control of surgical instruments for manipulation of tissue and performing surgical procedures; it is not described as diagnosing conditions.
No
The device description explicitly states it is an "Endoscopic Instrument Control System" and describes physical instruments with articulating wrists and force feedback capabilities. The performance studies also detail testing related to cleaning, reliability (life), dielectric strength, and jaw force to failure, all of which are associated with physical hardware components, not software alone.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly describes a surgical system designed to assist in the control of endoscopic instruments for manipulating tissue during various surgical procedures. This is a direct intervention on the patient's body.
- Device Description: The description focuses on the mechanical and functional aspects of the surgical instruments and the system that controls them.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue samples) to provide information for diagnosis, monitoring, or screening. The device is used on the patient, not on a sample from the patient.
IVD devices are used to perform tests on samples taken from the body to provide information about a person's health. This device is a surgical tool used to perform procedures directly on the patient.
No
The document explicitly states "Predetermined Change Control Plan (PCCP) - All Relevant Information: Not Found." This indicates that the device is not authorized as a PCCP authorized device.
Intended Use / Indications for Use
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model IS5000) shall assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic surgical procedures and general thoracoscopic surgical procedures. The system is indicated for adult use.
It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Product codes
NAY
Device Description
The da Vinci Force Feedback Instruments are 8 mm, multi-use, endoscopic instruments intended for use with the da Vinci Surgical System, Model IS5000. These instruments have been designed with an articulating wrist that gives surgeons natural dexterity and range of motion, and have the ability to render forces measured at the instrument tip to the hand controls at the system Console.
The basis for this submission is a modification to the reprocessing instructions, in addition to an increase in the number of surgical uses (lives) and reprocessing cycles for the da Vinci Force Feedback Instrument family as listed in Table 1, which Intuitive intends to introduce into interstate commerce in accordance with 21 CFR Section 807.81.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult
Intended User / Care Setting
trained physicians in an operating room environment
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Cleaning Validation: Cleaning validation was performed to validate the efficacy of the manual cleaning process in accordance with the following standards and guidance documents:
- ANSI/AAMI ST98:2022 - Cleaning validation of health care products - Requirements for development and validation of a cleaning process for medical devices
- AAMI TIR12:2020 - Technical Information Report - Designing, testing and labeling reusable medical devices for reprocessing in health care facilities: A guide for device manufacturers
- ASTM F3208-20 - Standard Guide for Selecting Test Soils for Validation of Cleaning Methods for Reusable Medical Devices
- FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015 (amended on June 9, 2017)
Reliability (Life) Verification: Instruments were subjected to sufficient reprocessing and simulated surgical use cycles to simulate their design life. Testing included benchtop performance measurements, visual inspection, and evaluation of functional performance to verify instrument mechanical and sensor performance at each cycle.
Dielectric Strength Testing: Dielectric strength testing was performed in accordance with IEC 60601-2-2:2017 (Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories). This testing was also conducted to fulfill FDA Guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, issued on March 9, 2020. Testing was conducted after subjecting the instruments to sufficient reprocessing and simulated surgical use cycles to simulate their design life.
Jaw Force to Failure Verification: Jaw force to failure verification was performed after subjecting the instruments to sufficient reprocessing and simulated surgical use cycles to simulate their design life. This testing was conducted to fulfill FDA Guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, issued on March 9, 2020.
All testing passed and supports the subject device changes.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 16, 2024
Intuitive Surgical, Inc. Jennifer Karafin Senior Regulatory Affairs Specialist 1266 Kifer Road Sunnyvale, California 94086
Re: K243641
Trade/Device Name: da Vinci Surgical System (IS5000) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY Dated: November 25, 2024 Received: November 26, 2024
Dear Jennifer Karafin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
1
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
2
Sincerely,
Mark Trumbore -S
Digitally signed by Mark Trumbore -S Date: 2024.12.16 11:30:24 -05'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
3
Indications for Use
Submission Number (if known)
Device Name
da Vinci Surgical System (IS5000)
Indications for Use (Describe)
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model IS5000) shall assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures and general thoracoscopic surgical procedures. The system is indicated for adult use.
It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Contraindication:
Use of the force feedback needle driver is contraindicated in hysterectomy and myomectomy due to the risk of vaginal bleeding requiring hospital readmission and/or the need for additional procedures. The use of non-force feedback needle drivers is recommended for suturing in these procedures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
510(k) Summary
| 510(k) Owner: | Intuitive Surgical, Inc.
1266 Kifer Road
Sunnyvale, CA 94086 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Jennifer Karafin
Senior Regulatory Affairs Specialist
Intuitive Surgical, Inc.
1266 Kifer Road,
Sunnyvale, CA 94086
Phone: +1-315-447-6522
Email: jennifer.karafin@intusurg.com |
| Date Summary Prepared: | November 25, 2024 |
| Trade Name: | Da Vinci Force Feedback Instruments |
| Common Name: | Endoscopic instruments and accessories |
| Classification: | Class II, 21 CFR 876.1500, Endoscope and Accessories |
| Product Code: | NAY (System, Surgical, Computer Controlled Instrument) |
| Review Panel: | General & Plastic Surgery |
| Predicate Devices: | Da Vinci Force Feedback Instruments (K232610, K241284) |
| Device Description: | |
The da Vinci Force Feedback Instruments are 8 mm, multi-use, endoscopic instruments intended for use with the da Vinci Surgical System, Model IS5000. These instruments have been designed with an articulating wrist that gives surgeons natural dexterity and range of motion, and have the ability to render forces measured at the instrument tip to the hand controls at the system Console.
The basis for this submission is a modification to the reprocessing instructions, in addition to an increase in the number of surgical uses (lives) and reprocessing cycles for the da Vinci Force Feedback Instrument family as listed in Table 1, which Intuitive intends to introduce into interstate commerce in accordance with 21 CFR Section 807.81.
| Device | Subject Device | | | Predicate Device | | | Predicate
Submission
Number |
|---------------------------------------------------------|----------------|-------------------|-------------------------------------|------------------|-------------------|-------------------------------------|-----------------------------------|
| | Model # | Number
of Uses | Number of
Reprocessing
Cycles | Model # | Number
of Uses | Number of
Reprocessing
Cycles | |
| Da Vinci Force Feedback
Large Needle Driver | 476006 | 6 | 9 | 475006 | 3 | 5 | K232610 |
| Da Vinci Force Feedback
Cadiere Forceps | 476049 | 6 | 9 | 475049 | 3 | 5 | K232610 |
| Da Vinci Force Feedback
Maryland Bipolar Forceps | 476172 | 6 | 9 | 475172 | 3 | 5 | K232610 |
| Da Vinci Force Feedback
Fenestrated Bipolar Forceps | 476205 | 6 | 9 | 475205 | 3 | 5 | K232610 |
| Da Vinci Force Feedback
ProGrasp Forceps | 476093 | 6 | 9 | 475093 | 4 | 6 | K241284 |
| Da Vinci Force Feedback
Mega SutureCut Needle Driver | 476309 | 6 | 9 | 475309 | 4 | 6 | K241284 |
Table 1: Subject and Predicate Device Number of Uses and Reprocessing Cycles
Image /page/4/Picture/9 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "I". The font is sans-serif and the letters are evenly spaced. The word is centered in the image.
©2024 Intuitive Surgical, Inc. All rights reserved. Product names are trademarks or registered trademarks of their respective holders.
5
Indications for Use:
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model IS5000) shall assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic surgical procedures and general thoracoscopic surgical procedures. The system is indicated for adult use.
lt is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Contraindication:
Use of the force feedback needle driver is contraindicated in hysterectomy and myomectomy due to the risk of vaginal bleeding requiring hospital readmission and/or the need for additional procedures. The use of non-force feedback needle drivers is recommended for suturing in these procedures.
Technological Characteristics:
The subject da Vinci Force Feedback Instruments are technologically similar to their predicate devices. The overall technological characteristics such as principles of operation are unchanged. The changes that are the basis for this submission include modifications to the performance specifications and design to increase the number of uses and reprocessing cycles.
Performance Data:
The following performance testing was conducted to demonstrate substantial equivalence to the predicate devices. All testing passed and supports the subject device changes:
Cleaning Validation: Cleaning validation was performed to validate the efficacy of the manual cleaning process in accordance with the following standards and guidance documents:
- . ANSI/AAMI ST98:2022 - Cleaning validation of health care products - Requirements for development and validation of a cleaning process for medical devices
- . AAMI TIR12:2020 - Technical Information Report - Designing, testing and labeling reusable medical devices for reprocessing in health care facilities: A guide for device manufacturers
- ASTM F3208-20 - Standard Guide for Selecting Test Soils for Validation of Cleaning Methods for Reusable Medical Devices
- . FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015 (amended on June 9, 2017)
Reliability (Life) Verification: Instruments were subjected to sufficient reprocessing and simulated surgical use cycles to simulate their design life. Testing included benchtop performance measurements, visual inspection, and evaluation of functional performance to verify instrument mechanical and sensor performance at each cycle.
Dielectric Strength Testing: Dielectric strength testing was performed in accordance with IEC 60601-2-2:2017 (Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories). This testing was also conducted to fulfill FDA Guidance Premarket Notification (510(k)) Submissions for Electrosurgical
Image /page/5/Picture/19 description: The image shows the word "INTUITIVE" in blue font. There is a small blue circle above the letter "U". The font is sans-serif and the letters are evenly spaced. The image is simple and clean, with a white background.
©2024 Intuitive Surgical, Inc. All rights reserved. Product names are trademarks or registered trademarks of their respective holders.
6
Devices for General Surgery, issued on March 9, 2020. Testing was conducted after subjecting the instruments to sufficient reprocessing and simulated surgical use cycles to simulate their design life.
Jaw Force to Failure Verification: Jaw force to failure verification was performed after subjecting the instruments to sufficient reprocessing and simulated surgical use cycles to simulate their design life. This testing was conducted to fulfill FDA Guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, issued on March 9, 2020.
Summary:
Based on the intended use, indications for use, technological characteristics, and performance data, the subject da Vinci Force Feedback Instruments are substantially equivalent to and are as safe, as effective, and perform as well as their predicate devices.
Image /page/6/Picture/7 description: The image shows the word "INTUITIVE" in blue font. There is a small blue circle above the letter "U". The font is sans-serif and the letters are evenly spaced.
©2024 Intuitive Surgical, Inc. All rights reserved. Product names are trademarks or registered trademarks of their respective holders.