K Number
K243624

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-01-23

(59 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Acu-Sinch Knotless Mini, when used for fixation of bone-to-bone or soft-tissue to bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Acu-Sinch Knotless Mini is indicated for Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of hematoma distraction arthroplasty by providing stablization at the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis.

Device Description

The Acu-Sinch Knotless Mini is intended to provide fixation during the healing process of a trapeziectomy. Acu-Sinch Knotless Mini consists of two buttons, a suture strand, and instruments to aid in insertion. The buttons are manufactured from titanium alloy conforming to ASTM FI 36 (Ti-6AL-4V ELI). The suture is manufactured from Ultra High Molecular Weight Polyethylene (UHMWPE). The devices are sterile and for single use.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called Acu-Sinch Knotless Mini. It describes the device's indications for use, technological characteristics, and a summary of non-clinical tests conducted to demonstrate substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with specific numerical values for metrics like sensitivity, specificity, or accuracy, as would be common for diagnostic AI/ML devices. Instead, the acceptance criteria are framed in terms of substantial equivalence to a predicate device (Arthrex CMC Mini Tightrope). The key performance criteria evaluated through non-clinical testing are static and dynamic fatigue, and MRI compatibility.

Acceptance Criteria CategorySpecific Criteria/GoalReported Device Performance
Mechanical PerformanceDevice performs statistically equivalent to the predicate device in static and dynamic fatigue testing.Acu-Sinch Knotless Mini performed as well as, or better than the predicate device. (This indicates it met or exceeded the performance of the predicate in these tests).
MRI CompatibilityDevice meets safety requirements for image artifact, magnetically induced displacement force, radio frequency induced heating, and magnetically induced torque in the Magnetic Resonance Environment, per relevant ASTM standards.Tests conducted in accordance with ASTM F2052, ASTM F2119, ASTM F2182, and ASTM F2213 were performed. The document states these tests were conducted, implying compliance with the standards, but does not state specific numerical results or explicitly say "passed" for each. However, the overall conclusion that the device is "safe and effective for the indication" implies these tests were successful.
BiocompatibilityEndotoxin levels meet requirement of <20 EU/device.Endotoxin testing (LAL and MMP) was conducted and met the requirement of <20 EU/device.
Material DegradationDevice performance maintained after specified accelerated and real-time aging.Device performance was verified following 5-year accelerated and 2-year real-time testing and passed.

2. Sample Size Used for the Test Set and Data Provenance

The document describes non-clinical bench testing. For such tests (static and dynamic fatigue, MRI compatibility, endotoxin, aging), the "sample size" typically refers to the number of physical devices tested. The document does not specify the sample size (number of devices) used for these tests.

The data provenance is from laboratory bench tests, not human subjects. Therefore, there is no country of origin for data or retrospective/prospective distinction in the typical sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This question is not applicable as the described study is a non-clinical bench test of a physical medical device (Acu-Sinch Knotless Mini) and not a study involving human subjects, interpretation of medical images, or AI/ML model performance that would require expert-established ground truth. The "ground truth" for the mechanical and material tests are the predefined ASTM standards and performance benchmarks.

4. Adjudication Method for the Test Set

This question is not applicable for the same reasons as point 3. There is no human interpretation or diagnostic decision-making involved that would necessitate an adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. The device is a physical medical implant (knotless mini-anchor system), not an AI/ML diagnostic or assistive technology. Therefore, an MRMC study is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The device is a physical medical implant, not an algorithm or software. Its performance is inherent to its physical properties and design, not an algorithmic output.

7. The Type of Ground Truth Used

The ground truth used for this device's evaluation is primarily engineering specifications and established test standards.

  • For mechanical performance (fatigue): The "ground truth" is the performance of the legally marketed predicate device (Arthrex CMC Mini Tightrope) and general engineering principles of sufficient strength and durability for the intended orthopedic application. The study aimed to demonstrate "statistically equivalent" or "better" performance relative to the predicate.
  • For MRI compatibility: The "ground truth" is compliance with specific ASTM standards (F2052, F2119, F2182, F2213) which define acceptable limits for magnetic inducement, heating, and artifact generation.
  • For biocompatibility: The "ground truth" is the specified endotoxin limit (<20 EU/device).
  • For material degradation: The "ground truth" is maintaining performance after accelerated and real-time aging, indicating material stability over time.

8. The Sample Size for the Training Set

This question is not applicable. The device is a physical medical implant, not an AI/ML model that requires a training set.

9. How the Ground Truth for the Training Set was Established

This question is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 23, 2025

Acumed LLC Marina Bull Regulatory Affairs Specialist 5885 NE Cornelious Pass Hillsboro, Oregon 97124

Re: K243624

Trade/Device Name: Acu-Sinch Knotless Mini Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: November 22, 2024 Received: November 25, 2024

Dear Marina Bull:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

RYAN TROMBETTA -S

For: Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243624

Device Name

Acu-Sinch Knotless Mini

Indications for Use (Describe)

Acu-Sinch Knotless Mini, when used for fixation of bone-to-bone or soft-tissue to bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Acu-Sinch Knotless Mini is indicated for Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of hematoma distraction arthroplasty by providing stablization at the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis.

Type of Use (Select one or both, as applicable)

IX.Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE F NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

'en agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 1 of 1

{4}------------------------------------------------

510(k) #:K243624Prepared on: 2024-11-23
Contact Details21 CFR 807.92(a)(1)
Applicant NameAcumed LLC
Applicant Address5885 NE Cornelious Pass Hillsboro OR 97124 United States
Applicant Contact Telephone970-846-6282
Applicant ContactMs. Marina Bull
Applicant Contact Emailmarina.bull@acumed.net
Device Name21 CFR 807.92(a)(2)
Device Trade NameAcu-Sinch Knotless Mini
Common NameWasher, Bolt, Nut
Classification NameSingle/multiple component metallic bone fixation appliances and accessories
Regulation Number888.3030
Product Code(s)HTN
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate#Predicate Trade Name (Primary Predicate is listed first)Product Code
K140328Arthrex CMC Min TightropeHTN
Device Description Summary21 CFR 807.92(a)(4)
The Acu-Sinch Knotless Mini is intended to provide fixation during the healing process of a trapeziectomy. Acu-Sinch Knotless Mini consists of two buttons, a suture strand, and instruments to aid in insertion. The buttons are manufactured from titanium alloy conforming to ASTM FI 36 (Ti-6AL-4V ELI). The suture is manufactured from Ultra High Molecular Weight Polyethylene (UHMWPE). The devices are sterile and for single use.
Intended Use/Indications for Use21 CFR 807.92(a)(5)
Acu-Sinch Knotless Mini, when used for fixation of bone-to-bone or soft-tissue to bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Acu-Sinch Knotless Mini is indicated for Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of hematoma distraction arthroplasty by providing stablization at the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis.
Indications for Use Comparison21 CFR 807.92(a)(5)

{5}------------------------------------------------

The Acu-Sinch Knotless Minj, has been compared to the predicate device, The Arthrex OMC Mini Tightrope within the 51 0(k) submission. The basis of substantial equivalence for the subject device is their similarities in intended use, material technology, operating principles, anatomical site, for implantation, performance and design. The analysis of difference between the subject device and predicate device supports substantial equivalence as the differences do not constitute a new intended use/indications for use, and the information the submission demonstrated that the subject device is comparable to the predicate device and does not raise different questions of safety or effectiveness.

The Acu-Sinch Knotless Mini has the same intended use/indications for use as the Arthrex CMC Mini Tightrope. Both have two buttons and a suture when used for bone or soft-tisue to bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair.

Technological Comparison

The Subject Device has the same technological characteristics as the Predicate Device and Predicate Devices make use of a suture to provide fixation between the buttons, located on the thumb metacarpal and index metacarpal.

The button and suture system is used for distributing suture tension over areas of ligament or tendon repair or in the suspensory fration of the thumb metacarpal at the base of the second metacarpal when the trapezium is removed.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Static and Dynamic fatigue testing was conducted to demonstrate that the proposed Acu-Sinch Knotless Mini performs statistically equivalent to the Predicate Device; Arthrex CMC Mini Tightrope cleared under KI 40328. Acu-Sinch Knotless Mini performed as well as, or better than the predicate device. " 1 1 1 8

WN force, torque, and image artifact testing were conducted in accordance with ASTM F2052 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment, ASTM F2119 Standard Test Method for Evaluation of In Image Artifacts from Passive Implants, ASTM F2182 Standard Test Method for Measurement of Measurement of Radio Frequency Induced Heating Near Passive Implants During Maging and ASTM F2213 Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment.

Endotoxin testing (LAL and MMP) was conducted and met the requirement of <20 EU/device.

Device performance was verified following 5-year accelerated and 2-vear real time testing and passed.

Clinical testing was not required to support substantial equivalence (Not Applicable).

Based on the results of the nonclinical bench testing described above, it was concluded that the Subject and Predicate Devices are equivalent in performance specifically for the intended use, hence was proven to be safe and effective for the indication.

Page 2 of 2

21 CFR 807.92(a)(6)

N/A