K Number
K243561
Date Cleared
2025-06-17

(211 days)

Product Code
Regulation Number
866.3987
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar.

All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions.

Positive results do not rule out bacterial infection or co-infection with other viruses.

Device Description

The Nano-Check™ Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for in vitro rapid, simultaneous qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly from anterior nasal swab specimens.

The assay kit consists of 25 test cassette devices, 25 reagent tubes, 25 ampules containing extraction buffer, 25 anterior nasal specimen collection swabs, one positive control swab, one negative control swab, one Instructions for Use, and one Quick Reference Instruction. An external positive control swab contains noninfectious influenza A, influenza B, and SARS-CoV-2 antigens dried onto the swab and an external negative control swab contains noninfectious blank universal viral transport media dried on the swab. The kit should be stored at 2°C - 30°C.

AI/ML Overview

Device Acceptance Criteria and Performance Study: Nano-Check Influenza+COVID-19 Dual Test

The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay for the qualitative detection and differentiation of influenza A, influenza B, and SARS-CoV-2 antigens in anterior nasal swab samples. The device's acceptance criteria and performance were established through extensive analytical and clinical studies.

1. Table of Acceptance Criteria and Reported Device Performance

The following table summarizes the key acceptance criteria and the performance achieved by the Nano-Check Influenza+COVID-19 Dual Test based on the provided 510(k) summary. Given that this is a qualitative assay, the primary performance metrics are Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) in clinical studies, and various measures of agreement/detection rates in analytical studies.

Performance Metric CategoryAcceptance Criteria (Implicit)Reported Device Performance
CLINICAL PERFORMANCE
SARS-CoV-2PPA ≥ 80% (typical for antigen tests), NPA ≥ 95%PPA: 87.6% (95% CI: 83.0% - 91.0%)NPA: 99.8% (95% CI: 99.5% - 99.9%)
Influenza APPA ≥ 80%, NPA ≥ 95%PPA: 86.9% (95% CI: 83.6% - 89.6%)NPA: 99.6% (95% CI: 99.1% - 99.8%)
Influenza BPPA ≥ 80%, NPA ≥ 95%PPA: 86.8% (95% CI: 79.4% - 91.9%)NPA: 99.7% (95% CI: 99.4% - 99.9%)
ANALYTICAL PERFORMANCE
Precision (Within-Lab)100% agreement for TN, HN, LP, MP levels across runs/operators100% agreement for all levels (SARS-CoV-2, Flu A, Flu B) per operator per run.
Precision (Between-Lot)Consistent results across lots, especially for moderate and high positivesFor C90 levels, agreement ranged from 83.3% to 100%. For 3X LOD levels, 100% agreement.
Reproducibility (Multi-site, Multi-operator)High agreement across sites and operators for all sample types (TN, HN, LP, MP)TN: 100%HN COVID: 100%HN Flu A: 100%HN Flu B: 99.4%LP COVID: 100%LP Flu A: 99.4%LP Flu B: 100%MP COVID: 100%MP Flu A: 100%MP Flu B: 100%
Cross-Reactivity/Microbial InterferenceNo cross-reactivity/interference at tested concentrationsNo cross-reactivity/interference observed with 50 pathogens (bacteria, fungi, viruses) and negative matrix.
Endogenous/Exogenous InterferenceNo interference with common substances at tested concentrationsNo interference observed with various nasal sprays, pain relievers, hand sanitizers, and other biological substances (except Hand sanitizer lotion, which caused false negative Influenza B when tested at 15% w/v).
Limit of Detection (LoD)Specific LoD values per virus strainSARS-CoV-2: 1.95×10² TCID₅₀/mL to 1.27×10⁴ TCID₅₀/mL (strain dependent)Influenza A: 2.8×10³ TCID₅₀/mL to 1.4×10⁵ CEID₅₀/mL (strain dependent)Influenza B: 1.04×10² TCID₅₀/mL to 2.25×10⁵ CEID₅₀/mL (strain dependent)WHO Standard SARS-CoV-2: 667 IU/mL
Analytical Reactivity (Inclusivity)100% detection for various strains at specified concentrations100% detection (3/3 replicates) for 14 SARS-CoV-2, 31 Flu A, and 16 Flu B strains at specified concentrations.
High Dose Hook EffectNo false negatives at high concentrationsNo high-dose hook effect observed for all tested viruses at concentrations up to 3.89×10⁴ TCID₅₀/mL (SARS-CoV-2), 2.8×10⁸ CEID₅₀/mL (Flu A), and 1.8×10⁷ TCID₅₀/mL (Flu B).
Competitive InterferenceNo interference between targets in co-infection scenariosNo competitive interference observed between SARS-CoV-2, Influenza A, and Influenza B at high/low titer combinations.
Specimen StabilityStable results for specified storage conditions/timesNasal swab samples stable for up to 48 hours at -20°C, 2-8°C, 23.5°C, and 30°C.
External Controls100% agreement with expected results for positive/negative controls100% agreement for all three lots of positive and negative external controls.

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Study Test Set Sample Size: A total of 1,969 subjects were enrolled in the clinical study.
  • Data Provenance: The data was collected from a multi-center, prospective clinical study in the U.S. between November 2022 and February 2025.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The device being reviewed is an in vitro diagnostic (IVD) test for antigen detection. For such devices, the "ground truth" in clinical performance studies is typically established by a highly sensitive and specific molecular assay (RT-PCR), rather than by human experts interpreting images or signals from the test device itself.

  • In this case, the ground truth for the clinical test set was established using an FDA-cleared RT-PCR method as the comparator.
  • The document does not specify the number of experts directly involved in establishing the RT-PCR ground truth or their qualifications beyond stating it was performed at a "reference laboratory as per the cleared instruction for use." This implies that qualified laboratory personnel, adhering to standardized RT-PCR protocols, established the ground truth.

4. Adjudication Method for the Test Set

Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving human interpretation (e.g., radiology reads) where discrepancies between readers need to be resolved. Since the Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunoassay interpreted visually by an operator, and the ground truth was established by an RT-PCR molecular assay, no explicit adjudication method for the test set is described or implied in the provided text. The comparison was directly between the device's visual results and the RT-PCR results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study (MRMC) is generally conducted for imaging AI devices to evaluate the impact of AI assistance on human reader performance. The Nano-Check Influenza+COVID-19 Dual Test is an in vitro diagnostic device for antigen detection, not an imaging AI device where human readers interpret complex images. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable to this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the performance presented for the Nano-Check Influenza+COVID-19 Dual Test in the clinical study is essentially standalone performance in the context of a rapid diagnostic test. While the test is visually interpreted by an operator, the performance metrics (PPA and NPA) are derived from the direct output of the device compared to the RT-PCR reference. There is no complex "algorithm" separate from the physical test strips' chemical reaction and visual readout. The operator simply reads the result displayed by the device. The "human-in-the-loop" here is the visual interpretation of a clear positive/negative line, not a complex decision-making process aided by AI.

7. The Type of Ground Truth Used

The type of ground truth used for the clinical performance study was an FDA-cleared molecular assay (RT-PCR method). This is a highly sensitive and specific laboratory-based test considered the gold standard for detecting viral nucleic acids, making it appropriate for establishing true positive and true negative cases of infection.

8. The Sample Size for the Training Set

The provided document describes the performance data for the test set (clinical study and analytical validation). It does not specify a separate training set sample size. This is expected because the Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunoassay, not a machine learning or AI model that requires a distinct training phase with a labeled dataset. The development and optimization of such assays rely on biochemical and immunological principles, followed by rigorous analytical and clinical validation.

9. How the Ground Truth for the Training Set Was Established

As noted above, there isn't a "training set" in the machine learning sense for this type of IVD device. The development of the assay (e.g., selecting antibodies, optimizing reagents) would involve internal R&D studies, using characterized viral samples and clinical specimens, but these are part of the development process rather than a formal "training set" with ground truth establishment for an AI algorithm. The performance data presented is from the validation against established reference methods.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 17, 2025

Nano-Ditech Corporation
James B Chang
President
259 Prospect Plains Rd., Bldg K
Cranbury, New Jersey 08512

Re: K243561
Trade/Device Name: Nano-Check Influenza+COVID-19 Dual Test
Regulation Number: 21 CFR 866.3987
Regulation Name: Multi-Analyte Respiratory Virus Antigen Detection Test
Regulatory Class: Class II
Product Code: SCA
Dated: November 15, 2024
Received: November 18, 2024

Dear James B Chang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K243561 - James B Chang Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243561 - James B Chang Page 3

Sincerely,

JOSEPH BRIGGS -S

Joseph Briggs, Ph.D.
Deputy Division Director
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K243561

Device Name: Nano-Check Influenza+COVID-19 Dual Test

Indications for Use (Describe)

The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar.

All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions.

Positive results do not rule out bacterial infection or co-infection with other viruses.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Nano-Check™ Influenza+COVID-19 Dual Test

I. Submitter

Nano-Ditech Corporation
259 Prospect Plains Rd., Bldg. K Cranbury,
NJ 08512 USA

Contact Person: James Chang (jchang@nanoditech.com)
Phone:1-609-409-3300
Date Prepared: March 19, 2025

II. Device

A. Trade, Proprietary, and Established Name: Nano-Check™ Influenza+COVID-19 Dual Test
B. Product Code: SCA - Multi-analyte respiratory virus antigen detection test
C. Classification: Class II
D. Regulation Number: 21 CFR 866.3987
E. Classification Name: Multi-analyte respiratory virus antigen detection test
F. Panel: MI-Microbiology
G. Purpose for Submission: New device 510(k) clearance for the Nano-Check™ Influenza+COVID19 Dual Test
H. Measurand: Nucleoprotein of influenza A and influenza B, nucleocapsid of SARS-CoV-2
I. Type of Test: Qualitative lateral flow immunoassay

III. Predicate Device

A. Trade Name: Healgen Rapid Check® COVID-19/Flu A&B Antigen Test
B. FDA 510k Number: DEN240029 (Cleared October 07, 2024)
C. Product Code: SCA

IV. Indications for Use

A. Intended Use(s):
See Indications for Use below

Page 6

B. Indication(s) for Use:

The Nano-Check™ Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar.

All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions.

Positive results do not rule out bacterial infection or co-infection with other viruses.

C. Special Conditions for Use Statement(s):

Rx - For Prescription Use Only
For In Vitro Diagnostic Use

D. Special Instrument Requirements:

N/A

V. Device/System Characteristics:

A. Device Description:

The Nano-Check™ Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for in vitro rapid, simultaneous qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly from anterior nasal swab specimens.

The assay kit consists of 25 test cassette devices, 25 reagent tubes, 25 ampules containing extraction buffer, 25 anterior nasal specimen collection swabs, one positive control swab, one negative control swab, one Instructions for Use, and one Quick Reference Instruction. An external positive control swab contains noninfectious influenza A, influenza B, and SARS-CoV-2 antigens dried onto the swab and an external negative control swab contains noninfectious blank universal viral transport media dried on the swab. The kit should be stored at 2°C - 30°C.

B. Principles of Operations:

The Nano-Check™ Influenza+COVID-19 Dual Test is designed to detect the extracted nucleoprotein antigen specific to influenza A, and influenza B and the extracted nucleocapsid antigen specific to SARS-CoV-2 in ANS specimens directly collected from patients exhibiting signs or symptoms of a respiratory infection.

The test strip enclosed in a cassette housing is comprised of the following components: sample pad,

Page 7

reagent pad, biotin pad, reaction membrane, and absorbent pad. The reagent pad contains colloidal gold conjugated with monoclonal antibodies (mAb) specific for SARS-CoV-2, influenza A, and influenza B target proteins. The biotin pad contains biotin conjugated with mAb specific for SARS-CoV-2. The reaction membrane contains the secondary antibodies for the proteins of Flu A and Flu B, and streptavidin for the biotinylated SARS-CoV-2 antibody. The whole strip is fixed inside a plastic cassette.

When the specimens are extracted and added to the sample well of the test device, Flu A, Flu B nucleoproteins and/or SARS-CoV-2 nucleocapsid antigen is present in the specimen, a complex form between the anti-Flu A/Flu B/ SARS-CoV-2 conjugate and the viral antigen will be captured by the streptavidin or specific anti-Flu A/Flu B mAb coated on the test line region (C/A/B line). The absence of the test line (C/A/B line) suggests a negative result. To serve as a procedural control, a red line will always appear in the control line region (CON) indicating that proper volume of sample has been added and membrane wicking has occurred. Any result without this control line is invalid.

VI. Comparison of Technological Characteristics with the Predicate Device

The table below compares the Nano-Check™ Influenza+COVID-19 Dual Test to the predicate devices of Healgen Rapid Check® COVID-19/Flu A&B Antigen Test.

Device & Predicate Device(s):Predicate: DEN240029Candidate
Device Trade NameHealgen Rapid Check® COVID-19/Flu A&B Antigen TestNano-Check™ Influenza+COVID-19 Dual Test

Similarities

| Indications for use/ Intended Use | The Healgen Rapid Check COVID-19/Flu A&B Antigen Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when | The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar. All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions. |

Page 8

determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider. Positive results do not rule out co-infection with other respiratory pathogens, and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.Positive results do not rule out bacterial infection or co-infection with other viruses.
Regulation number21 CFR 866.3987Same
DiseaseMulti-respiratory infections such as influenza and SARS-CoV-2Same
Intended Use PopulationIndividuals with symptoms of multi-respiratory tract infections such as influenza and SARS-CoV-2Same
UsageSingle use testSame
Assay Principle (Technology)ImmunochromatographicSame
AnalyteInfluenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigenSame
Test ResultQualitativeSame
Reading Time15 minutesSame
Result InterpretationVisually readSame
Specimen TypeAnterior nasal swab specimensSame

Differences

| Patient Use | Over the counter use/self-testing | For prescription use only and can be used in CLIA-waived settings |

VII. Performance Data

A. Analytical Performance

1. Precision/Reproducibility:

a) Within Laboratory Precision

The precision study 1 was undertaken by two operators with each operator conducting 24 test runs over a 12-day period, resulting in a total of 96 replicates per level. The study employed blind sample panels comprising true negative (TN), high negative (HN, 0.1× LoD), low positive (LP, 1× LoD), and moderate positive (MP, 3× LoD) samples for each analyte (SARS-CoV-2, influenza A, and

Page 9

influenza B). The precision testing was conducted according to the Package Insert and the qualitative results are summarized in Table 1 below.

Table 1. Precision Study 1 -Summary Result

SARS-CoV-2

RunTrue NegativeHigh NegativeLow PositiveModerate Positive
12121212
Run 124 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 24
Run 224 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 24
Total48 / 4848 / 4848 / 4848 / 4848 / 4848 / 4848 / 4848 / 48
% Agreement100%100%100%100%100%100%100%100%
95% CI92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%

Influenza A

RunTrue NegativeHigh NegativeLow PositiveModerate Positive
12121212
Run 124 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 24
Run 224 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 24
Total48 / 4848 / 4848 / 4848 / 4848 / 4848 / 4848 / 4848 / 48
% Agreement100%100%100%100%100%100%100%100%
95% CI92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%

Influenza B

RunTrue NegativeHigh NegativeLow PositiveModerate Positive
12121212
Run 124 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 24
Run 224 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 2424 / 24
Total48 / 4848 / 4848 / 4848 / 4848 / 4848 / 4848 / 4848 / 48
% Agreement100%100%100%100%100%100%100%100%
95% CI92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%92.6 - 100%

The study results demonstrated 100% agreement for all samples. These findings satisfied the acceptance criteria of the precision study

Study 2 was specifically conducted to assess between-lot variability. The study used negative samples (without virus analytes), low positive samples at C90, and moderate positive (3× LoD) samples for each analyte (SARS-CoV-2, influenza A, and influenza B). Samples were blinded and tested randomized. This supplemental precision testing was conducted by two operators, with each operator conducting 12 test runs per lot over 12 days with three lots, resulting in a total of 72 replicates per level. The lot and operator stratified results from this testing are included in Table 2 below.

Page 10

Table 2. Precision Study 2 -Summary Result

Between lot

AnalyteTest lineLot 1Lot 2Lot 3
PositiveAgreePositiveAgreePositiveAgree
NegativeCOVID-190/24100%0/24100%0/24100%
Flu A0/24100%0/24100%0/24100%
Flu B0/24100%0/24100%0/24100%
C90 COVID-19COVID-1920/2483.3%23/2495.8%21/2487.5%
Flu A0/24100%0/24100%0/24100%
Flu B0/24100%0/24100%0/24100%
C90, Flu ACOVID-190/24100%0/24100%0/24100%
Flu A20/2483.3%21/2487.5%23/2495.8%
Flu B0/24100%0/24100%0/24100%
C90, Flu BCOVID-190/24100%0/24100%0/24100%
Flu A0/24100%0/24100%0/24100%
Flu B22/2491.7%21/2487.5%24/24100%
3X LOD COVID-19COVID-1924/24100%24/24100%24/24100%
Flu A0/24100%0/24100%0/24100%
Flu B0/24100%0/24100%0/24100%
3X LOD Flu ACOVID-190/24100%0/24100%0/24100%
Flu A24/24100%24/24100%24/24100%
Flu B0/24100%0/24100%0/24100%
3X LOD Flu BCOVID-190/24100%0/24100%0/24100%
Flu A0/24100%0/24100%0/24100%
Flu B24/24100%24/24100%24/24100%

Between operator

AnalyteTest lineOperator 1Operator 2
PositiveAgreePositiveAgree
NegativeCOVID-190/36100%0/36100%
Flu A0/36100%0/36100%
Flu B0/36100%0/36100%
C90 COVID-19COVID-1931/3686.1%33/3691.7%
Flu A0/36100%0/36100%
Flu B0/36100%0/36100%
C90, Flu ACOVID-190/36100%0/36100%
Flu A32/3688.9%32/3688.9%
Flu B0/36100%0/36100%
C90, Flu BCOVID-190/36100%0/36100%
Flu A0/36100%0/36100%
Flu B33/3688.9%34/3694.4%
3X LOD COVID-19COVID-1936/36100%36/36100%
Flu A0/36100%0/36100%
Flu B0/36100%0/36100%

Page 11

3X LOD Flu ACOVID-190/36100%0/36100%
Flu A36/36100%36/36100%
Flu B0/36100%0/36100%
3X LOD Flu BCOVID-190/36100%0/36100%
Flu A0/36100%0/36100%
Flu B36/36100%36/36100%

Precision estimates for samples negative samples, low positive samples (C90), and moderate positive (3× LoD), are expected to be low due to the random errors of the testing procedure across different days and runs, paired with an operator's ability to read the line intensity for samples with very low analyte concentration.

Taken together, the results of both precision assessments demonstrate a test precision and a lot-to-lot precision that are consistent with the expectations for the analyte concentrations in the samples, the test's technology, and the test's LoD. The between-lot variability does not impact low-concentrated samples equal to or above 3 X LoD of the test.

b) Reproducibility

The reproducibility study was designed to evaluate site-to-site, operator-to-operator, and level-to-level variability and demonstrate that the Nano-Check™ Influenza +COVID-19 Dual Test can be performed consistently and correctly. This study was conducted at 4 distinct sites, each with two (2) or three (3) different operators, testing three (3) reagent lots, using a coded panel of contrived samples consisting of a true negative (TN), a high negative sample (HN, 0.1x LoD), a low positive (LP, 1.0 x LoD) and a medium positive (MP, 5.0 x LoD) sample for each analyte (SARS-CoV-2, influenza A, and influenza B). A total of 8 untrained operators at 3 CLIA waived sites and 3 operators in the internal site each tested using 150 coded samples (TN: 15, HN COVID: 15, HN-Flu A: 15, HN-Flu B: 15, LP-COVID: 15, LP-Flu A: 15, LP-Flu B: 15, MP-COVID: 15, MP-Flu A: 15, MP-Flu B: 15, respectively) over five nonconsecutive days. The summary of results is presented in Table 3 below.

Table 3. Reproducibility Study- Summary Result

SampleNo of Positive Result/No of Total Tested (% Positive Rate)Agreement
Site 1 (2Operators)Site 2 (3Operators)Site 3 (3Operators)Site 4 (3Operators)Total95% CI
TN0/30 (0%)0/45 (0%)0/45 (0%)0/45 (0%)165/165 (100%)97.7-100.0
HN COVID0/30 (0%)0/45 (0%)0/45 (0%)0/45 (0%)165/165 (100%)97.7-100.0
HN Flu A0/30 (0%)0/45 (0%)0/45 (0%)0/45 (0%)165/165 (100%)97.7-100.0
HN Flu B0/30 (0%)0/45 (0%)1/45 (2.2%)0/45 (0%)164/165 (99.4%)96.7-99.9

Page 12

LP COVID30/30 (100%)45/45 (100%)45/45 (100%)45/45 (100%)165/165 (100%)97.7-100.0
LP Flu A30/30 (100%)45/45 (100%)44/45 (97.8%)45/45 (100%)164/165 (99.4%)96.7-99.9
LP Flu B30/30 (100%)45/45 (100%)45/45 (100%)45/45 (100%)165/165 (100%)97.7-100.0
MP COVID30/30 (100%)45/45 (100%)45/45 (100%)45/45 (100%)165/165 (100%)97.7-100.0
MP Flu A30/30 (100%)45/45 (100%)45/45 (100%)45/45 (100%)165/165 (100%)97.7-100.0
MP Flu B30/30 (100%)45/45 (100%)45/45 (100%)45/45 (100%)165/165 (100%)97.7-100.0

All samples tested in the reproducibility study generated no significant difference between sites.

2. Linearity: This section is not applicable as this test device is a qualitative assay.

3. Cross-reactivity (Analytical Specificity):

a) Cross-reactivity and Microbial Interference Study

Cross-Reactivity and Interference study of the Nano-Check™ Influenza+COVID-19 Dual Test was conducted with 50 potential cross-reactive or interfering pathogens of bacteria (19), fungus (1), and viruses (28) and negative matrix (2). Each microorganism was tested in three (3) replicates in the absence or presence of 2x LoD concentrations of SARS-CoV-2, influenza A, and influenza B antigens. The concentrations of potentially interfering microorganisms tested and the results from the cross- reactivity study are presented in Table 4 below.

Table 4. Results of Cross-Reactivity/Microbial Interference

Microorganism (Strain)Concentration TestedPositive Sample (# Positive/ # Tested)Negative Sample (# Positive/ # Tested)Cross-Reactivity/ Microbial Interference
Bordetella pertussis1.0×10⁶ cfu/mL3/30/3No
Candida albicans1.0×10⁶ cfu/mL3/30/3No
Chlamydophila pneumoniae1.0×10⁶ IFU/mL3/30/3No
Corynebacterium diphtheriae1.0×10⁶ cfu/mL3/30/3No
Escherichia coli1.0×10⁶ IFU/mL3/30/3No
Haemophilus influenzae, B1.0×10⁶ cfu/mL3/30/3No
Lactobacillus acidophilus1.0×10⁶ cfu/mL3/30/3No
Legionella Pneumophila subsp. Pneumophila1.0×10⁶ cfu/mL3/30/3No
Moraxella catarrhalis1.0×10⁶ cfu/mL3/30/3No
Mycobacterium tuberculosis1.0×10⁶ cfu/mL3/30/3No
Mycoplasma pneumoniae1.0×10⁶ cfu/mL3/30/3No
Neisseria meningitidis1.0×10⁶ cfu/mL3/30/3No

Page 13

Neisseria mucosa1.0×10⁶ cfu/mL3/30/3No
Neisseria subflava1.0×10⁶ cfu/mL3/30/3No
Pseudomonas aeruginosa1.0×10⁶ cfu/mL3/30/3No
Staphylococcus aureus1.0×10⁶ cfu/mL3/30/3No
Staphylococcus epidermidis1.0×10⁶ cfu/mL3/30/3No
Streptococcus pneumoniae1.0×10⁶ cfu/mL3/30/3No
Streptococcus pyogenes1.0×10⁶ cfu/mL3/30/3No
Streptococcus salivarius salivarius1.0×10⁶ cfu/mL3/30/3No
Epstein-Barr Virus1.0×10⁵ cp/mL3/30/3No
Enterovirus 711.0×10⁵ TCID₅₀/mL3/30/3No
Enterovirus D 681.0×10⁵ TCID₅₀/mL3/30/3No
Human Herpesvirus8.0×10⁴ TCID₅₀/mL3/30/3No
Human Adenovirus 11.0×10⁵ TCID₅₀/mL3/30/3No
Human Adenovirus 21.0×10⁵ TCID₅₀/mL3/30/3No
Human Adenovirus 7, Gomen1.0×10⁵ TCID₅₀/mL3/30/3No
Human Coronavirus, 229E1.0×10⁵ TCID₅₀/mL3/30/3No
Human Coronavirus, NL637.0×10⁴ TCID₅₀/mL3/30/3No
Human Coronavirus, OC434.5×10⁴ TCID₅₀/mL3/30/3No
Human Metapneumovirus 3, B1, Peru2-20021.95×10⁴ TCID₅₀/mL3/30/3No
Human Metapneumovirus, TN/83-12111.0×10⁵ TCID₅₀/mL3/30/3No
Human Parainfluenza Virus 11.0×10⁵ TCID₅₀/mL3/30/3No
Human Parainfluenza Virus 21.0×10⁵ TCID₅₀/mL3/30/3No
Human Parainfluenza Virus 31.0×10⁵ TCID₅₀/mL3/30/3No
Human Parainfluenza Virus 4B1.0×10⁵ TCID₅₀/mL3/30/3No
Human RSV, A Long1.0×10⁵ TCID₅₀/mL3/30/3No
Human RSV, A 93201.0×10⁵ TCID₅₀/mL3/30/3No
Human RSV, A21.0×10⁵ TCID₅₀/mL3/30/3No
Human RSV, B 185371.0×10⁵ TCID₅₀/mL3/30/3No
Human RSV, B WV/14617/851.0×10⁵ TCID₅₀/mL3/30/3No
Human RSV, B11.0×10⁵ TCID₅₀/mL3/30/3No
Human Rhinovirus 1A, 20601.0×10⁵ PFU/mL3/30/3No
Measles Virus, Edmonston1.7×10⁴ TCID₅₀/mL3/30/3No
MERS-CoV, EMC/20121.0×10⁵ TCID₅₀/mL3/30/3No
Mumps Virus, MuV/Iowa.US/20061.0×10⁵ TCID₅₀/mL3/30/3No
Rhinovirus 20, 15-CV191.0×10⁵ TCID₅₀/mL3/30/3No
SARS-CoV1.0×10⁵ PFU/mL3/30/3No
Pooled Human Nasal WashN/A3/30/3No
Pooled Human Nasal FluidN/A3/30/3No

Coronavirus HKU1 was not tested for cross-reactivity due to a lack of availability. 20 clinical samples containing Coronavirus HKU1 were tested and all resulted as negative, however, the viral load/concentration of each sample is unknown.

Page 14

The study results showed that none of the evaluated microorganisms demonstrated cross-reactivity or interference with the Nano-Check™ Influenza+COVID-19 Dual Test at the tested concentrations.

b) Endogenous/Exogenous Interference

The Endogenous/Exogenous interference study of the Nano-Check™ Influenza+COVID-19 Dual Test was conducted using medically relevant endogenous and exogenous interferents. The SARS-CoV-2, influenza A, or influenza B were diluted in negative clinical matrix to 2x LoD separately and the potentially interfering substance was spiked to the concentration presented in Table 5 below. Three (3) replicates were tested for each sample prepared.

Table 5. Results of Endogenous/Exogenous Interference Study

SubstancesInterferent Conc.Interference
Nasal Spray 115% v/vNo
Nasal Spray 215% v/vNo
Nasal Spray 315% v/vNo
Nasal Spray 415% v/vNo
Budesonide Nasal Spray15% v/vNo
Nasonex 24 hr Allergy15% v/vNo
Nasacort Allergy 24HR15% v/vNo
Sore Throat (Oral Pain Reliever spay)15% v/vNo
ZICAM® Oral mist15% v/vNo
Sore Throat Lozenges15% w/vNo
Zinc Cold Therapy15% w/vNo
Homeopathic Allergy Nasal Spray15% v/vNo
NasoGEL (Gel Spray)15% v/vNo
Nasalcrom® Nasal Allergy spray15% v/vNo
Histaminum 30C15% w/vNo
Skin relief hand cream1% w/vNo
Hand Soap Fresh Breeze Scent1% w/vNo
Antibacterial liquid Hand Soap15% w/vNo
Hand Sanitizer Gel15% w/vNo
Hand sanitizer lotion*10% w/vNo
Disinfectant Spray1% v/vNo
Acetylsalicylic acid3.00×10¹ μg/mLNo
Dexamethasone1.20×10¹ μg/mLNo
Mometasone furoate4.50×10⁻⁴ μg/mLNo
Mupirocin1.50×10⁰ μg/mLNo
Oseltamivir phosphate3.99×10⁻¹ μg/mLNo
Tobramycin3.30×10¹ μg/mLNo
Beclomethasone dipropionate5.04 μg/mLNo
Flunisolide870 μg/mLNo

Page 15

Molnupiravir3.29 mg/mLNo
Remdesivir240 μg/mLNo
Zanamivir30 mg/mLNo
Human Neutrophils5×10⁶ cells/mLNo
Mucin (Bovine submaxillary Glands, Type I-S)5 mg/mLNo
Whole Blood2.50%No
Biotin3500 ng/mLNo

*Hand sanitizer lotion tested at 15% w/v concentration resulted in false negative Influenza B results.

The study results showed no interference of these substances with the Nano-Check ™ Influenza+ COVID-19 Dual Test at the concentrations tested.

4. Assay Reportable Range: This section is not applicable as this test device is a qualitative assay.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

a) Internal Quality Control

The Nano-Check™ Influenza+COVID-19 Dual Test contains built in internal assay control. The appearance of the control line on the test ensures that sufficient flow of the sample occurred during the assay.

b) External Quality Controls

The Nano-Check™ Influenza+COVID-19 Dual Test contains one positive external control swab and one negative external control swab that allows for monitoring of the performance of the assay.

The positive control swab contains recombinant SARS-CoV-2 nucleocapsid, recombinant influenza A nucleoprotein, and recombinant influenza B nucleoprotein, and the negative control swab contains blank Universal Viral Transport media.

c) External Controls Lot-to-Lot Precision

The lot-to-lot precision for the external positive and negative control swabs was evaluated using three lots of external controls. The results from the study are summarized in Table 6 below.

Table 6. External Controls Lot-to-Lot Precision Study Results

External controlLot No.No. of Positive Result / No. of Total TestAgreement with Expected Results
COVID-19Flu AFlu B
Positive External Control0804-DN-3K24-P10 / 1010 / 1010 / 10100%
0804-DN-3K27-P10 / 1010 / 1010 / 10100%
0804-DN-3K30-P10 / 1010 / 1010 / 10100%

Page 16

Negative External Control0804-DN-3K24-N0 / 100 / 100 / 10100%
0804-DN-3K27-N0 / 100 / 100 / 10100%
0804-DN-3K30-N0 / 100 / 100 / 10100%

All lots of external controls generated 100% agreement with the expected results.

d) Specimen Stability

Two test samples were prepared for the specimen stability study: a negative sample (consisting of pooled human nasal fluid with no analyte) and a low positive sample (containing a diluted SARS-CoV-2, influenza A, and influenza B sample at 2 x LoD in the negative pooled human nasal fluid). Swabs were spiked with 50 μL of each sample and subjected to four different temperature conditions: room temperature (23.5°C), high room temperature (30 °C), 2°C to 8°C and frozen (-20°C). The samples were stored under each condition for various time intervals including 0 hours, 1 hour, 2 hours, 4 hours, 8 hours, 24 hours, and 48 hours. Subsequently, the exposed sample swabs were tested using 5 replicates for each exposure time, as shown in Table 7 below.

Table 7. Specimen Stability Results

Specimen Storage TemperatureTested SampleSpecimen Storage Time (No. of Positive/No. of Total Test)
0 hour1 hour2 hours4 hours8 hours24 hours48 hours
23.5°CSARS-CoV-25/55/55/55/55/55/55/5
Influenza A5/55/55/55/55/55/55/5
Influenza B5/55/55/55/55/55/55/5
Negative0/50/50/50/50/50/50/5
30°CSARS-CoV-25/55/55/55/55/55/55/5
Influenza A5/55/55/55/55/55/55/5
Influenza B5/55/55/55/55/55/55/5
Negative0/50/50/50/50/50/50/5
2°C to 8°CSARS-CoV-25/55/55/55/55/55/55/5
Influenza A5/55/55/55/55/55/55/5
Influenza B5/55/55/55/55/55/55/5
Negative0/50/50/50/50/50/50/5
-20°CSARS-CoV-25/55/55/55/55/55/55/5
Influenza A5/55/55/55/55/55/55/5
Influenza B5/55/55/55/55/55/55/5
Negative0/50/50/50/50/50/50/5

The results showed that the nasal swab samples were stable for up 48 hours under all temperature conditions, ranging from -20°C to 30°C. Therefore, freshly collected specimen swabs are recommended to be processed no later than one hour after specimen collection when kept at room temperature (15°C - 30°C) or within 24 hours when stored at 2°C to 8°C.

Page 17

6. Detection Limit:

The Limit of Detection (LoD) was determined with two SARS-CoV-2 Omicron variant strains (USA/MD-HP20874/2021, USA/COR-22-0631 13/2022), two influenza A strains (influenza A H1N1: A/California/04 /2009, influenza A H3N2: A/Victoria/361/2011) and two influenza B strains (influenza B/Hong Kong/330/2001, influenza B/Phuket/3073/13). Contrived samples were prepared by spiking the strains into a pooled negative nasal fluid matrix. A preliminary LoD was determined by spiking 50 µL of serially diluted sample onto swab heads and tested using the Nano-Check™ Influenza+COVID-19 Dual Test. The preliminary LoD initially determined by testing a two-fold serial dilution series of 3 replicates was confirmed by testing in 20 replicates. The results from these studies are summarized below in Table 8.

Table 8. Limit of Detection (LoD)

VirusStrainLoDPositive Rate (%) (# Positive/# Tested
SARS-CoV-2USA/MD-HP20874/2021, Heat Inactivated1.95×10² TCID₅₀/mL (9.75×10⁰ TCID₅₀/swab)95%
USA/COR-22-063113/2022, Heat Inactivated1.27×10⁴ TCID₅₀/mL (6.35×10² TCID₅₀/swab)100%
Influenza AH1N1: A/California/04/20092.8×10³ TCID₅₀/mL (1.4×10² TCID₅₀/swab)100%
H3N2: A/Victoria/361/20111.4×10⁵ CEID₅₀/mL (7.0×10³ CEID₅₀/swab)95%
Influenza BVictoria: B/Hong Kong/330/20012.25×10⁵ CEID₅₀/mL (1.13×10⁴ CEID₅₀/swab)95%
Yamagata: B/Phuket/3073/131.04×10² TCID₅₀/mL (5.2×10⁰ TCID₅₀/swab)100%

Furthermore, the LoD was established using the 1st WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in real clinical matrix of nasal fluid. Initially, a preliminary LoD test was performed by spiking 50 µL of each diluted sample onto the sample collection swab head in three replicates. Following this, a confirmatory LoD test with 17 additional replicates was conducted at the preliminary LoD concentration, a total of 20 replications. To determine LoD concentration, additional dilutions were investigated bracketing the confirmed LoD concentration, with each dilution level undergoing 20 replications. It was determined that the LoD of the 1st WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in the nasal fluid matrix was confirmed to be 667 IU/mL (33.5 IU/swab).

7. Analytical Reactivity (Inclusivity):

A total of 14 strains of SARS-CoV-2, 31 strains of influenza A, and 16 strains of influenza B were tested in the inclusivity study. The virus stocks were diluted at the predetermined concentration in the pooled negative nasal fluid matrix and 50 μL were pipetted onto the swabs. The swabs were tested in 3 replicates in this study. The lowest concentration of each strain that resulted in 100% detection (3/3) is presented in Table 9 below.

Page 18

Table 9. Inclusivity

VirusVirus StrainsConcentrationUnits
SARS-CoV-2 (B.1.1.7, Alpha)USA/CA/CDC/5574/2020¹⁾6.77E+06GE /mL
USA/CA/CDC/5574/2020²⁾2.39E+04TCID₅₀/mL
England/204820464/20207.19E+03TCID₅₀/mL
SARS-CoV-2 (B.1.351, Beta)USA/MD-HP01542/2021³⁾7.20E+04GC/mL
USA/MD-HP01542/2021⁴⁾3.80E+06GC/mL
South Africa/KRISP-K0053 25/20201.90E+04TCID₅₀/mL
SARS-CoV-2 (B.1.617.2, Delta)USA/MD-HP05285/2021⁵⁾7.20E+07GC/mL
USA/MD-HP05285/2021⁶⁾5.00E+06GC/mL
USA/PHC658/20215.21E+02TCID₅₀/mL
SARS-CoV-2 (P.1, Gamma)Japan/TY7-503/20211.58E+04TCID₅₀/mL
USA/NY-Wadsworth-21033 899-01/ 20217.85E+03TCID₅₀/mL
SARS-CoV-2 (B.1.617.1, Kappa)USA/CA-Stanford-15_S02/ 20218.48E+04TCID₅₀/mL
SARS-CoV-2 (B.1.1.529, Omicron)USA/GA-EHC-2811C/20212.11E+07GC/mL
SARS-CoV-2 (JN.1, Omicron)USA/New York/PV96109/20 233.14E+03TCID₅₀/mL
Flu A H1N1A/Puerto Rico/8/348.00E+06CEID₅₀/mL
A/Brisbane/59/20074.45E+06CEID₅₀/mL
A/Denver/1/574.00E+05CEID₅₀/mL
A/San Diego/1/2009 pdm092.80E+04TCID₅₀/mL
A/Tijuana/4/092.45E+01TCID₅₀/mL
A/Solomon Islands/3/20064.45E+05CEID₅₀/mL
A/NWS/331.23E+05CEID₅₀/mL
A/FM/1/474.25E+05CEID₅₀/mL
A/New Jersey/8/761.70E+04CEID₅₀/mL
A/New Caledonia/20/19994.00E+05CEID₅₀/mL
A/Hawaii/66/20191.28HA
A/Hawaii/66/2019 X-345A1.28HA
A/Guangdong-Maonan/1536/ 20191.28HA
A/Guangdong-Maonan/1536/ 2019 CNIC-19090.64HA
A/Victoria/4897/2022(pdm09)1.58E+07EID₅₀/mL
A/Victoria/2570/2019 (pdm09)9.98E+05EID₅₀/mL
Flu A H1N2A/Swine/Ohio/09SW1477/20092.30E+04TCID₅₀/mL
Flu A H3N2A/Hong Kong/8/19681.40E+05CEID₅₀/mL
A/Aichi/2/19684.00E+05CEID₅₀/mL
A/Wisconsin/67/20057.00E+05CEID₅₀/mL
A/Hong Kong/4801/20149.60E+05CEID₅₀/mL
A/Netherlands/22/20038.00E+02TCID₅₀/mL
A/Netherlands/823/19921.44E+01TCID₅₀/mL
A/Brisbane/10/20071.38E+05CEID₅₀/mL

Page 19

A/Wisconsin/15/20095.00E+03CEID₅₀/mL
A/Sydney/5/974.45E+04CEID₅₀/mL
A/Port Chalmers/1/732.00E+05CEID₅₀/mL
A/Victoria/3/754.00E+05CEID₅₀/mL
A/Perth/16/2009 x A/Puerto Rico/8/19 34, NIB642.80E+06CEID₅₀/mL
A/Singapore/INFIMH-16-0019 /162.51E+03TCID₅₀/mL
Flu A H5N1A/mallard/Wisconsin/2576/20095.25E+05
Flu B (Victoria Lineage)B/Brisbane/60/20089.00E+04
B/Malaysia/2506/20041.12E+06
B/New York/1056/20033.20E+03
B/Washington/02/20191.28
Flu B (Yamagata Lineage)B/Florida/78/20152.80E+04
B/Texas/06/20114.45E+08
B/New York/1061/20048.00E+02
B/Christchurch/33/20048.00E+02
B/Sydney/507/20061.60E+05
B/Wisconsin/1/20101.80E+06
B/Florida/4/20063.50E+05
Flu B (Non-Victoria/ Yamagata)B/Colorado/6/171.78E+02
B/Taiwan/2/19624.45E+03
B/Lee/19409.00E+04
B/GL/1739/545.00E+04
B/Great Lakes/1739/19543.20E+04

¹⁾ Source: Bei Resources (Cat. #: NR-55245, Lot#:70043111), ²⁾ Source: ZeptoMetrix (Cat. #: 0810612CFHI, Lot#:328055), ³⁾ Source: Bei Resources (Cat. #: NR-553651, Lot#:70045299), ⁴⁾ Source: Bei Resources (Cat. #: NR-55350, Lot#:70045608), ⁵⁾ Source: Bei Resources (Cat. #: NR-56128, Lot#:70048021), ⁶⁾ Source: ATCC (Cat. #: VR-3342HK, Lot#:70048932

8. High Dose Hook Effect:

High-dose hook effect study was conducted to determine if a hook effect would be observed at high concentrations of the analyte (i.e., a false negative at high concentrations of SARS-CoV-2, influenza A, and influenza B) in a testing sample. The series of three or five concentrations with each two (2) strains of SARS-CoV-2, influenza A, and influenza B were prepared in pooled negative nasal fluid matrix and tested in three replicates on Influenza+COVID-19 Dual Test. The concentration of each virus in the prepared samples ranged from the maximum virus concentration possible (undiluted virus stock) to 2x LoD. All spiked samples were 100% positive, as expected, at all tested concentrations. The Nano-Check™ Influenza+COVID-19 Dual Test showed no high-dose hook effect when subjected to 3.89×10⁴ TCID₅₀/mL for SARS-CoV-2, USA/MD-HP20874/2021 (B.1.1.529, Omicron Variant), 2.53×10⁶ TCID₅₀/mL for SARS-CoV-2, USA/COR-22-063113/2022 (BA.5, Omicron Variant), 2.8×10⁶ TCID₅₀/mL for influenza A (A/California/04/2009, H1N1), 2.8×10⁸ CEID₅₀/mL for influenza A (A/Victoria/361/2011, H3N2), 1.8×10⁷ TCID₅₀/mL for

Page 20

influenza B (B/Hong Kong/330/2001, Victoria) and 4.17×10⁵ TCID₅₀/mL for influenza B (B/Phuket/3073/13, Yamagata).

9. Competitive Interference Study

For co-infection, SARS-CoV-2 at levels near LoD was tested in the presence of high levels of influenza A or influenza B and near LoD influenza A and influenza B in the presence of high levels of SARS-CoV-2. Additionally, the performance of Nano-Check™ Influenza+COVID-19 Dual Test was evaluated in the presence of high levels of influenza A and influenza B. Contrived high and low titer influenza A (H1N1 and H3N2) and B positive samples were. No competitive interference was observed between SARS-CoV-2 and influenza A and B as listed in Table 10 below.

Table 10. Competitive Inhibition

High Titer TargetLow Titer TargetLow Titer Target Percent Positivity
Virus NameConcentrationVirus NameConcentration
Flu A (H1N1)2.8 × 10⁵SARS-CoV-23.81 × 10⁴100%
Flu A (H1N1)2.8 × 10⁵Flu B (Victoria)6.75 × 10⁵100%
Flu A (H1N1)2.8 × 10⁵Flu B (Yamagata)3.12 × 10²100%
Flu A (H3N2)2.8 × 10⁶SARS-CoV-23.81 × 10⁴100%
Flu A (H3N2)2.8 × 10⁶Flu B (Victoria)6.75 × 10⁵100%
Flu A (H3N2)2.8 × 10⁶Flu B (Yamagata)3.12 × 10²100%
Flu B (Victoria)1.8 × 10⁶SARS-CoV-23.81 × 10⁴100%
Flu B (Victoria)1.8 × 10⁶Flu A (H1N1)8.4 × 10³100%
Flu B (Victoria)1.8 × 10⁶Flu A (H3N2)4.2 × 10⁵100%
Flu B (Yamagata)4.17 × 10⁵SARS-CoV-23.81 × 10⁴100%
Flu B (Yamagata)4.17 × 10⁵Flu A (H1N1)8.4 × 10³100%
Flu B (Yamagata)4.17 × 10⁵Flu A (H3N2)4.2 × 10⁵100%
SARS-CoV-22.53 × 10⁵Flu A (H1N1)8.4 × 10³100%
SARS-CoV-22.53 × 10⁵Flu A (H3N2)4.2 × 10⁵100%
SARS-CoV-22.53 × 10⁵Flu B (Victoria)6.75 × 10⁵100%
SARS-CoV-22.53 × 10⁵Flu B (Yamagata)3.12 × 10²100%
Flu A (H1N1)2.8 × 10⁵Flu B (Yamagata)3.12 × 10²100%
SARS-CoV-23.81 × 10⁴100%
Flu A (H1N1)2.8 × 10⁵Flu B (Victoria)6.75 × 10⁵100%
SARS-CoV-23.81 × 10⁴100%
Flu A (H3N2)2.8 × 10⁶Flu B (Yamagata)3.12 × 10²100%
SARS-CoV-23.81 × 10⁴100%
Flu A (H3N2)2.8 × 10⁶Flu B (Victoria)6.75 × 10⁵100%
SARS-CoV-23.81 × 10⁴100%
Flu B (Victoria)1.8 × 10⁶Flu A (H1N1)8.4 × 10³100%
SARS-CoV-23.81 × 10⁴100%
Flu B (Victoria)1.8 × 10⁶Flu A (H3N2)4.2 × 10⁵100%

Page 21

SARS-CoV-23.81 × 10⁴100%
Flu B (Yamagata)4.17 × 10⁵Flu A (H1N1)8.4 × 10³100%
SARS-CoV-23.81 × 10⁴100%
Flu B (Yamagata)4.17 × 10⁵Flu A (H3N2)4.2 × 10⁵100%
SARS-CoV-23.81 × 10⁴100%
SARS-CoV-22.53 × 10⁵Flu A (H1N1)8.4 × 10³100%
Flu B (Victoria)6.75 × 10⁵100%
SARS-CoV-22.53 × 10⁵Flu A (H1N1)8.4 × 10³100%
Flu B (Yamagata)3.12 × 10²100%
SARS-CoV-22.53 × 10⁵Flu A (H3N2)4.2 × 10⁵100%
Flu B (Victoria)6.75 × 10⁵100%
SARS-CoV-22.53 × 10⁵Flu A (H3N2)4.2 × 10⁵100%
Flu B (Yamagata)3.12 × 10²100%

9. Assay Cut-off:

Not Applicable

10. Carry-Over:

Carry-over contamination is not applicable to this test device as each sample uses an independent, new, single-use test cassette that is discarded after each run.

B. Comparison Studies:

1. Method Comparison with Predicate Device:

Not applicable. The performance of the Nano-Check™ Influenza+COVID-19 Dual Test was evaluated in a clinical study against an FDA-cleared molecular assay.

2. Matrix Comparison:

The Nano-Check™ Influenza+COVID-19 Dual Test is only intended for the qualitative detection of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen in direct anterior nasal swab specimens. As no other specimen or sample type is claimed herein.

C. Clinical Study:

The clinical performance of the Nano-Check™ Influenza+COVID-19 Dual Test was evaluated in a multi-center, prospective clinical study in the U.S. between November 2022 and February 2025. The study only enrolled subjects who presented with symptoms of respiratory infection. A total of one thousand nine hundred sixty-nine (1,969) subjects were consecutively enrolled and tested by fourteen (14) operators across six (6) different clinical CLIA-waived sites. Two anterior nasal

Page 22

swabs were collected from each study subject during the same visit. The first AN swab specimen for the comparator method was collected by the operators from both sides of the nose. The collected AN swab specimen was stored/transported and tested at a reference laboratory with an FDA cleared RT-PCR method as per the cleared instruction for use. The second AN swab specimen was collected from both sides of the nose using the provided swab and was tested immediately using the Nano-Check™ Influenza+COVID-19 Dual Test by an operator at the site.

There were 1,969 subjects evaluated, 21.1% were collected from patients who are 5 years or younger, 43.8% from patients aged 6-21 years, 26.3 % from patients aged 22 to 60 years, and 8.8% from patients aged 61 years or older. Among the subjects, 51.6% were from female patients, while 48.4% were from male patients. Results obtained with the Nano-Check™ Influenza+COVID-19 Dual Test were compared to the results obtained with the RT-PCR comparator test to determine clinical sensitivity and specificity.

For SARS-CoV-2 detection, 265 of these specimens tested positive, and 1,704 tested negative by the FDA-cleared RT-PCR method. The comparison of Nano-Check™ Influenza+COVID-19 Dual Test results with the comparator method results showed a PPA of 87.6% (232/265) with a 95% CI of 83.0% - 91.0%, NPA of 99.8% (1,701/1,704) with a 95% CI of 99.5% - 99.9%.

For influenza A detection, 480 of these specimens tested positive, and 1,489 tested negative by the FDA-cleared RT-PCR method. The comparison of Nano-Check™ Influenza+COVID-19 Dual Test results with the comparator method results showed a PPA of 86.9% (417/480) with a 95% CI of 83.6% - 89.6%, NPA of 99.6% (1,483/1,489) with a 95% CI of 99.1% - 99.8%.

For influenza B detection, 114 of these specimens tested positive, and 1,855 tested negative by the FDA-cleared RT-PCR method. The comparison of Nano-Check™ Influenza+COVID-19 Dual Test results with the comparator method results showed a PPA of 86.8% (99/114) with a 95% CI of 79.4% - 91.9%, NPA of 99.7% (1,850/1,855) with a 95% CI of 99.4% - 99.9%.

Table 11. Clinical Performance of Nano-Check™ Influenza+COVID-19 Dual Test Compared to an RT-PCR Comparator

SARS-CoV-2 Detection

Nano-Check™ Influenza+COVID-19 Dual TestRT-PCR ComparatorTotal
PositiveNegative
Positive2323235
Negative331,7011,734
Total2651,7041,969

Positive Percent Agreement (PPA) = (232/265) x 100% = 87.6% (95% CI: 83.0% - 91.0%)
Negative Percent Agreement (NPA) = (1,701/1,704) x 100% = 99.8% (95% CI: 99.5% - 99.9%)
Positivity in Study Cohort = 13.46% (265/1,969; 95% CI: 12.0 - 15.0%)

Page 23

Influenza A Detection

Nano-Check™ Influenza+COVID-19 Dual TestRT-PCR ComparatorTotal
PositiveNegative
Positive4176423
Negative631,4831,546
Total4801,4891,969

Positive Percent Agreement (PPA) = (417/480) x 100% = 86.9% (95% CI: 83.6% - 89.6%)
Negative Percent Agreement (NPA) = (1483/1,489) x 100% = 99.6% (95% CI: 99.1% - 99.8%)
Positivity in Study Cohort = 24.4% (480/1,969; 95% CI: 22.5 – 26.3%)

Influenza B Detection

Nano-Check™ Influenza+COVID-19 Dual TestRT-PCR ComparatorTotal
PositiveNegative
Positive995104
Negative151,8501,865
Total1141,8551,969

Positive Percent Agreement (PPA) = (99/114) x 100% = 86.8% (95% CI: 79.4% - 91.9%)
Negative Percent Agreement (NPA) = (1,850/1,855) x 100% = 99.7% (95% CI: 99.4% - 99.9%)
Positivity in Study Cohort = 5.8% (114/1,969; 95% CI: 4.8 – 6.9%)

1. Clinical Sensitivity:

Please refer to Section VI.C (Clinical Studies) above for the clinical validation. The PPAs for each analyte in the Nano-Check™ Influenza+COVID-19 Dual Test are presented below.

a) SARS-CoV-2: 87.6% (232/265, 95% CI: 83.0% - 91.0%)
b) Influenza A: 86.9% (417/480, 95% CI: 83.6% - 89.6%)
c) Influenza B: 86.8% (99/114, 95% CI: 79.4% - 91.9%)

2. Clinical Specificity:

Please refer to Section VI.C (Clinical Studies) above for the clinical validation. The NPAs for each analyte in the Nano-Check™ Influenza+COVID-19 Dual Test are presented below.

a) SARS-CoV-2: 99.8% (1,701/1,704, 95% CI: 99.5% - 99.9%)
b) Influenza A: 99.6% (1,483/1,489, 95% CI: 99.1% - 99.8%)
c) Influenza B: 99.7% (1,850/1,855, 95% CI: 99.4% - 99.9%)

Page 24

D. Clinical Cut-Off:

Not Applicable since there is no clinical cutoff related to the presence of SARS-CoV-2, influenza A, and influenza B in patient samples.

E. Expected Value/Reference Range:

The rate of positivity as determined by the Nano-Check™ Influenza+COVID-19 Dual Test during the 2022-2025 clinical study was 21.5% for influenza A (423/1,969), 5.3% for influenza B (104/1,969), and 11.9% for SARS-CoV-2 (235/1,969).

F. Other Supportive Information:

1. Flex Studies

To assess the robustness and risk for false results of the test when deviating from the IFU/QRI test steps, flex studies were conducted that assessed all major aspects of the test procedure (sample volume, reading time, other deviations from the procedure [delay in mixing, delay in addition of sample to the well, incubation time] and variability of environmental test conditions that the test may be subjected to when in use (lighting, disturbance during use, temperature and humidity stress conditions). The test was performed with contrived positive nasal swabs generated by diluting SARS-CoV-2 virus, influenza A, and influenza B into a negative nasal fluid matrix at 2xLoD. The results demonstrated that the test system is robust and that false results can be expected to be reasonably mitigated through labeling.

VIII. Conclusion

The information presented in this Premarket Notification is complete and supports a substantial equivalence decision. Based on the comparison of technological characteristics and intended use, and as a result of the non-clinical and clinical performance testing completed on Nano-Check™ Influenza+COVID-19 Dual Test, the proposed device does not raise new questions of safety and effectiveness and supports the conclusion that the proposed device is substantially equivalent to the predicate device. The results of non-clinical and clinical testing demonstrate that the device is as safe, as effective, and performs as well as the predicate device. Therefore, the Nano-Check™ Influenza+COVID-19 Dual Test is substantially equivalent to the predicate device.

N/A