K Number
K243522
Date Cleared
2025-05-30

(197 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ViVi® Toga Premium is a component of the ViVi® System and is intended to be worn by surgical personnel to provide a barrier between the operating environment and the surgical personnel in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

Device Description

The ViVi® Toga Premium is a one-piece head and body cover that is worn by healthcare professionals. The ViVi® Toga Premium is a component of the ViVi® Surgical Helmet System and worn over the air-exchanging surgical helmet. The ViVi® Toga Premium is used with the ViVi® Helmet or the ViVi® Helmet HPL (High Power LED) with Light Emitting Diode (LED) light to provide a barrier between the operating environment and the surgical personnel to protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

ViVi® Toga Premium is designed and has been tested to meet the applicable AAMI PB70 standards for level 4 compliance. The AAMI standard does not cover apparel for the head, face, and eyes. Therefore, the hood area and the lens are exempt from classification under this standard.

AI/ML Overview

The provided FDA 510(k) clearance letter for the ViVI® Toga Premium is for a Class II surgical apparel device. This document primarily details the substantial equivalence of the ViVi® Toga Premium to a predicate device based on non-clinical performance data. It explicitly states that "No clinical evaluation is necessary for this device."

Therefore, this document does not describe a study involving an AI/Machine Learning (ML) algorithm with a test set, ground truth, expert adjudication, or MRMC studies. The device is a physical product (surgical toga/gown), and its "performance" is related to its physical properties and barrier capabilities, not diagnostic or predictive accuracy.

Given the information provided, it's not possible to fulfill the request for acceptance criteria and study details related to an AI/ML device. The "study" mentioned in the document refers to non-clinical laboratory testing of material properties.

However, I can extract the acceptance criteria and the "reported device performance" based on the provided document as it relates to this specific physical device.


Acceptance Criteria and Reported Device Performance for ViVi® Toga Premium (Surgical Apparel)

Based on the provided FDA 510(k) clearance letter for the ViVi® Toga Premium, the acceptance criteria and reported device performance relate to its physical properties and barrier capabilities, as it is a surgical apparel device and not an AI/ML device.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the standards and tests performed, with "PASS" indicating successful compliance.

Property or CharacteristicTest Method / StandardAcceptance Criteria (Implied)Reported Device Performance (ViVi® Toga Premium)
Flammability of clothing textiles16 CFR 1610Class 1 compliantClass 1 compliant PASS
Biocompatibility: CytotoxicityISO 10993-5CompliantCompliant PASS
Biocompatibility: SensitizationISO 10993-10CompliantCompliant PASS
Biocompatibility: IrritationISO 10993-23CompliantCompliant PASS
Tear ResistanceASTM D5587CompliantCompliant PASS
Tensile StrengthASTM D5034CompliantCompliant PASS
Seam StrengthASTM F88/F88MCompliantCompliant PASS
Laser ResistanceISO 11810:2015CompliantCompliant PASS
Linting (Lint and other particles generation in the dry state)ISO 9073-10:2005CompliantCompliant PASS
Water Vapor Transmission rateASTM E96/E96M-24ACompliantCompliant PASS
Barrier Performance (Liquid)ANSI/AAMI PB70:2012; ASTM F1671/F1671MLevel 4 compliantCompliant Level 4 PASS
Sterilization Assurance Level (SAL)Not explicitly detailed by test method but stated as requirement10⁻⁶10⁻⁶

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the exact sample sizes (number of garments or material samples) used for each individual non-clinical test. The testing is reported as "tests were conducted and completed," indicating standard laboratory evaluations.
  • Data Provenance: Not applicable in the context of clinical data. The tests are material and performance tests conducted in a laboratory setting. No geographical origin of "data" in the sense of patient data is mentioned, as this is a physical product. The manufacturer is THI Total Healthcare Innovation GmbH, based in Austria. The predicate device manufacturer is Zimmer Surgical, Inc.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Not applicable. This device is surgical apparel, and its "performance" is assessed through standardized physical and biological material testing, not through expert human interpretation of data like images or clinical outcomes. There is no "ground truth" in the diagnostic AI sense.

4. Adjudication Method for the Test Set

  • Not applicable. As described above, there is no expert adjudication process for this type of device. Performance is determined by meeting pre-defined regulatory standards and test method specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • No. An MRMC study is a type of clinical study used to evaluate the performance of diagnostic devices (often imaging-based AI) by comparing the accuracy of multiple human readers with and without AI assistance across multiple cases. This does not apply to a surgical apparel device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was Done

  • Not applicable. This is not an AI algorithm.

7. The Type of Ground Truth Used

  • Not applicable in the context of AI/ML. The "ground truth" for this device's performance is derived from established international and national standards (e.g., ISO, ASTM, ANSI/AAMI, CFR), which define objective physical and chemical properties and test methodologies. Meeting these standards serves as the "truth" for device functionality.

8. The Sample Size for the Training Set

  • Not applicable. This is not an AI/ML device requiring a training set.

9. How the Ground Truth for the Training Set was Established

  • Not applicable. This is not an AI/ML device and does not have a training set or associated ground truth.

FDA 510(k) Clearance Letter - ViVi® Toga Premium

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 30, 2025

THI Total Healthcare Innovation GmbH
℅ Albert Rego
Consultant
Albert Rego PhD, Inc (d.b.a Rego Associates)
25401 Cabot Road, # 122
Laguna Hills, California 92653

Re: K243522
Trade/Device Name: ViVi® Toga Premium
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical apparel
Regulatory Class: Class II
Product Code: FYA
Dated: April 30, 2025
Received: April 30, 2025

Dear Albert Rego:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243522 - Albert Rego Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243522 - Albert Rego Page 3

Sincerely,

ALLAN GUAN -S

For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K243522

Device Name

  • ViVi® Toga Premium;
  • ViVi® Toga Premium XL-XXL (82220);
  • ViVi® Toga Premium M-L (82221);
  • ViVi® Toga Premium XXS-S (82222);
  • ViVi® Toga Premium XL-XXL (including customer logo) (82220V);
  • ViVi® Toga Premium M-L (including customer logo) (82221V);
  • ViVi® Toga Premium XXS-S (including customer logo) (82222V)

Indications for Use (Describe)
The ViVi® Toga Premium is a component of the ViVi® System and is intended to be worn by surgical personnel to provide a barrier between the operating environment and the surgical personnel in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

THI Total Healthcare Innovation GmbH
Gewerbestrasse 4
A-9181 Feistritz im Rosental
Tel: +43 4228 30100-0
Fax: +43 4228 30100-20
E-Mail: office@thigmbh.at

Page 1 of 5

K243522 510(k) SUMMARY

SUBMITTER:
THI Total Healthcare Innovation GmbH
Gewerbestrasse 4
9181 Feistritz
Austria
Tel.: +43 4228 30100

SUBMITTER CONTACT:
Albert Rego, Ph.D. Inc.
25401 Cabot Road
Suite 122
Laguna Hills, CA 92653
Phone (949) 632-8126

DATE:
May 30th, 2025

TRADE NAME:
ViVi® Toga Premium

PRODUCT CODE / DEVICE:
FYA – Gown, Surgical

REGULATION NUMBER AND DESCRIPTION:
21 CFR 878.4040 – Surgical Apparel

PREDICATE DEVICE:
TotalShield™II Zippered Surgical Toga and Hood (K192194 cleared on October 15th,2019)

Reference Device:
ViVi® Surgical Helmet System (K222214 cleared on October 17th, 2024)

DEVICE DESCRIPTION:
The ViVi® Toga Premium is a one-piece head and body cover that is worn by healthcare professionals. The ViVi® Toga Premium is a component of the ViVi® Surgical Helmet System and worn over the air-exchanging surgical helmet. The ViVi® Toga Premium is used with the ViVi® Helmet or the ViVi® Helmet HPL (High Power LED) with Light Emitting Diode (LED) light to provide a barrier between the operating environment and the surgical personnel to protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

ViVi® Toga Premium is designed and has been tested to meet the applicable AAMI PB70 standards for level 4 compliance. The AAMI standard does not cover apparel for the head, face, and eyes. Therefore, the hood area and the lens are exempt from classification under this standard.

Page 6

THI Total Healthcare Innovation GmbH
Gewerbestrasse 4
A-9181 Feistritz im Rosental
Tel: +43 4228 30100-0
Fax: +43 4228 30100-20
E-Mail: office@thigmbh.at

Page 2 of 5

STATEMENT OF INTENDED USE:

The ViVi® Toga Premium is a component of the ViVi® System and is intended to be worn by surgical personnel to provide a barrier between the operating environment and the surgical personnel in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

COMPARISON TO PREDICATE:

The comparison of features and operation principles between ViVi® Toga Premium and predicate device TotalShield™ II Zippered Surgical Toga and Hood (K192194) and reference device ViVi® Hood as part of the ViVi® Surgical Helmet System (Surgical Gown) (K222214) is listed as follows:

CategorySubject Device ViVi® Toga Premium (Surgical Gown)Predicate Device TotalShield Surgical Helmet System (Surgical Gown)Reference Device ViVi® Surgical Helmet System (Surgical Hood)Comparison
Common NameViVi® Toga PremiumTotalShield™ II Zippered Surgical Toga and HoodViVi® Surgical Helmet System
ManufacturerTHI Total Healthcare Innovation GmbhZimmer Surgical, Inc.THI Total Healthcare Innovation Gmbh
510(k) NumberK243522K192194K222214
Registration Number21 CFR 878.404021 CFR 878.404021 CFR 878.4040Same
Device ClassClass IIClass IIClass IISame
Product CodeFYAFYAFXYSame as predicate device
Classification PanelGeneral HospitalGeneral HospitalGeneral HospitalSame
Indication For Use¹The ViVi® Toga Premium is a component of the ViVi® System and is intended to be worn by surgical personnel to provide a barrier between the operating environment and the surgical personnel in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.The TotalShield II Zippered Surgical Toga and/or TotalShield II Surgical Hood is for use with the TotalShield II Surgical Helmet and/or TotalShield Surgical Helmet with LED lighting as the TotalShield Surgical Helmet System that is intended to be worn by surgical personnel to provide a barrier between the operating environment and the surgical personnel in order to protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.The ViVi® Hood is a component of the ViVi® System and is intended to be worn by surgical personnel to provide a barrier between the operating environment and the surgical personnel in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.Same
Target user groupSurgical personnelOperating Room PersonnelSurgical personnelSimilar
Condition of use Surgical Hood or TogaSterile / Disposable / Single UseSterile / Disposable / Single Use (Gowns)Sterile / Disposable / Single Use (Gowns)Same

Page 7

THI Total Healthcare Innovation GmbH
Gewerbestrasse 4
A-9181 Feistritz im Rosental
Tel: +43 4228 30100-0
Fax: +43 4228 30100-20
E-Mail: office@thigmbh.at

Page 3 of 5

CategorySubject Device ViVi® Toga Premium (Surgical Gown)Predicate Device TotalShield Surgical Helmet System (Surgical Gown)Reference Device ViVi® Surgical Helmet System (Surgical Hood)Similarities/Differences
Materials Technology Product Features
Sterilization of GownYesYesYesSame
Sterilization method Surgical Hood or TogaEthylene oxideEthylene oxideEthylene oxideSame
Sterilization Assurance Level (SAL)10⁻⁶10⁻⁶10⁻⁶Same
Barrier PerformanceANSI/AAMI PB70:2012 Level 4 compliant ASTM F2407-06 (2013)AAMI/ANSI PB70:2012 Level 4 compliant ASTM F2407-06 (Reapproved 2013) Standard Specification for Surgical Gowns Intended for Use in Healthcare FacilitiesASTM F2407-06 (2013)Same
Method of Hood/Toga attachment to Surgical HelmetMechanical slot and hook and loop (Velcros®)Hook-and-loopMechanical slot and hook and loop (Velcros®)Similar
MaterialSMS nonwoven with PE filmBase material SMS Nonwoven with PE Film reinforcement in critical zonesSMS nonwoven with PE filmSimilar
BiocompatibilityISO 10993-5 Cytotoxicity ISO 10993-10 Sensitization ISO 10993-23 IrritationISO 10993-5 Cytotoxicity ISO 10993-10 Sensitization ISO 10993-23 IrritationISO 10993-5 Cytotoxicity ISO 10993-10 Sensitization ISO 10993-23 IrritationSame
SizesToga: 3 sizes (XXL-XL; L-M; S-XXS)Large, X-Large, 2X-Large, 3X-LargeHood: single sizeSimilar

Page 8

THI Total Healthcare Innovation GmbH
Gewerbestrasse 4
A-9181 Feistritz im Rosental
Tel: +43 4228 30100-0
Fax: +43 4228 30100-20
E-Mail: office@thigmbh.at

Page 4 of 5

NON-CLINICAL PERFORMANCE DESCRIPTION:

During the design and development of the ViVi® Toga Premium, the following tests were conducted and completed.

Level 4 liquid barrier performance according to:

  • ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities
  • ASTM Method F1671/F1671M Standard Test Method for Resistance of Materials Used in Protective Clothing to Penetration by Blood-Borne Pathogens Using Phi-X174 Bacteriophage Penetration as a Test System

Biocompatibility according to:

  • ISO 10993-1 Biological evaluation of medical devices Evaluation and testing within a risk management process
  • ISO 10993-5 Biological Evaluation of Medical Devices - Part 5(2009): Tests for In Vitro Cytotoxicity
  • ISO 10993-10 Biological Evaluation of Medical Devices – Part 10(2021) Test for Skin Sensitization
  • ISO 10993-11 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity
  • ISO 10993-23 Biological Evaluation of Medical Devices – Part 23(2021): Tests for Irritation)

Laser Resistance of surgical drapes according to:

  • ISO 11810 Lasers and laser-related equipment — Test method and classification for the laser resistance of surgical drapes and/or patient protective covers — Primary ignition, penetration, flame spread and secondary ignition

Tear Resistance according to:

  • ASTM D5587 Standard Test Method for Tearing Strength of Fabrics by Trapezoid Procedure

Tensile test according to:

  • ASTM D5034 Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test)

Seam Strength according to:

  • ASTM F88/F88M Standard Test Method for Seal Strength of Flexible Barrier Materials

Lint and other particle generation according to:

  • ISO 9073-10:2005 Textiles — Test methods for nonwovens — Part 10: Lint and other particles generation in the dry state

Water Vapor Transmission according to:

  • ASTM E96/E96M-24A Standard Test Methods for Water Vapor Transmission of Materials

Page 9

THI Total Healthcare Innovation GmbH
Gewerbestrasse 4
A-9181 Feistritz im Rosental
Tel: +43 4228 30100-0
Fax: +43 4228 30100-20
E-Mail: office@thigmbh.at

Page 5 of 5

PERFORMANCE COMPARISON TO PREDICATE

Property or CharacteristicsTest MethodPredicate Device TotalShield Surgical Helmet System (Surgical Gown) K192194Subject Device ViVi® Toga Premium (Surgical Gown) K243522
Flammability of clothing textiles16 CFR 1610Class 1 compliant PASSClass 1 compliant PASS
Biological evaluationISO 10993-5 Cytotoxicity ISO 10993-10 Sensitization ISO 10993-23 IrritationCompliant PASSCompliant PASS
Tear ResistanceASTM D5587Compliant PASSCompliant PASS
Tensile StrengthASTM D5034Compliant PASSCompliant PASS
Seam StrengthASTM F88/F88MCompliant PASSCompliant PASS
Laser ResistanceISO 11810:2015Compliant PASSCompliant PASS
LintingISO 9073-10:2005Compliant PASSCompliant PASS
Water Vapor Transmission rateASTM E96Compliant PASSCompliant PASS
Barrier PerformanceANSI/AAMI PB70:2012Compliant Level 4 PASSCompliant Level 4 PASS

CLINICAL PERFORMANCE:

No clinical evaluation is necessary for this device.

CONCLUSION:

The conclusions drawn from the nonclinical tests demonstrate that the ViVi® Toga Premium is as safe, as effective, and performs as well as or better than the predicate device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.