K Number
K243509

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-05-12

(181 days)

Product Code
Regulation Number
888.3660
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Archer^R PSI System is indicated as an orthopedic instrument to assist the physician in the intraoperative positioning of total shoulder replacement components and in guiding the drill and the cut of the bone.

Archer^R PSI System must only be used conjointly with Archer™ CSR Total Shoulder (K152825, K173812, K181287, K182500, K191811), Catalyst EA Convertible Stemmed Shoulder (K222317) and Archer™ R1 Reverse (K202611, K211991, K213349, K223655, K232583) components in the context of primary total shoulder replacement and following a delto-pectoral approach only. Archer^R PSI System is manufactured from a pre-operative planning validated by the surgeon in the 'Archer™ 3D Targeting' platform (K213779). Archer^R PSI System is indicated for patient population fulfilling the Archer™ CSR Total Shoulder, Catalyst EA Convertible Stemmed Shoulder and Archer™ R1 Reverse indications and for which CT images are available with identifiable placement anatomical landmarks and compliant with imaging protocol provided by Archer 3D Targeting.

The device is intended for single use only.

The device is intended for adult patients.

The device has to be used by a physician trained in the performance of surgery.

Device Description

The "Archer PSI System" device is a patient-matched additively manufactured single use surgical instrument (PSI). Archer PSI System is an instrument set containing a glenoid guide and its bone model and/or a humeral guide and its bone model. This patient-specific medical device is designed to fit the patient's anatomy to transfer a patient-specific pre-operative plan to the operating room. It is intended for surgical interventions in orthopaedic procedures for total shoulder arthroplasty.

The Archer PSI system instruments are designed from a draft treatment plan available via the Archer™ 3D Targeting' platform. Based on computed tomography (CT) of the shoulder anatomy, 3D CAD models of the bones and positioning and sizing of the glenoid and humeral components are submitted for evaluation to the surgeon. Upon the surgeon's approval, the guides and bone models are designed based on the validated planning and are manufactured using additive manufacturing.

AI/ML Overview

The provided FDA 510(k) summary for the "Archer PSI System" does not contain the detailed acceptance criteria or the specific study that directly proves the device meets those criteria in a quantitative manner as typically expected for medical device performance studies involving sensitivity, specificity, accuracy, etc. However, it does outline the types of non-clinical and cadaveric testing performed to demonstrate substantial equivalence to a predicate device.

Here's an attempt to structure the information based on the request, extracting what is available and noting what is not:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not overtly state quantitative "acceptance criteria" (e.g., "accuracy must be > 95%") nor does it provide "reported device performance" in terms of explicit metrics like sensitivity, specificity, or error rates. Instead, the "performance" is described in terms of demonstrating "substantial equivalence" through various engineering and cadaveric tests.

A more accurate representation, based on the provided text, would be:

Acceptance Criteria CategoryDescription (from document)Reported Device Performance (from document)
Mechanical IntegrityDemonstrate mechanical integrity post-processing.Testing was conducted.
Debris GenerationAssess debris generation.Testing was conducted.
Intra-Designer VariabilityAssess variability within a single designer's output.Testing was conducted.
Inter-Designer VariabilityAssess variability between different designers' outputs.Testing was conducted.
BiocompatibilityEnsure material biocompatibility.Assessment conducted.
Cleaning & SterilizationValidate cleaning and sterilization processes.Validations conducted.
Manufacturing CleaningValidate manufacturing cleaning processes.Validation conducted.
Packaging & Shelf-lifeValidate packaging integrity and shelf-life.Validation conducted.
Functional EquivalenceDemonstrate functional equivalence to manual techniques for positioning and guiding drill/cut.Cadaveric testing executed to demonstrate substantial equivalence between two techniques (manual and PSI, for both anatomic and for reverse techniques).
Pre-operative PlanningManufactured from a pre-operative planning validated by the surgeon in the 'Archer™ 3D Targeting' platform (K213779).The device design is based on surgeon-validated plans within the Archer™ 3D Targeting platform. This implies an acceptance of the planning accuracy by the surgeon.

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for Cadaveric Testing: The document states that "Cadaveric testing was executed" but does not specify the sample size (number of cadavers or procedures) used for this testing.
  • Data Provenance: The cadaveric testing is implied to be prospective in nature, as it was "executed to demonstrate the substantial equivalence." There is no information provided regarding the country of origin of the cadaveric data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: The document states that "a pre-operative planning validated by the surgeon" is part of the process. For the cadaveric testing, it does not explicitly state the number of experts (e.g., surgeons) involved in establishing the "ground truth" or assessing the "substantial equivalence."
  • Qualifications of Experts: The document mentions that the device is to be used by a "physician trained in the performance of surgery." For the validation of the pre-operative plan, the expert is identified as "the surgeon." While this indicates a medical professional, specific qualifications (e.g., years of experience, subspecialty) are not provided.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the cadaveric testing or any other performance evaluation. The "validation by the surgeon" for the pre-operative plan suggests a form of single-expert consensus at the planning stage, but not for the overall performance assessment.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No.
  • Effect Size: As no MRMC study was performed, no effect size of human readers improving with AI vs. without AI assistance is reported. The Archer PSI System is a patient-specific instrument, not an AI diagnostic or assistive tool in the MRMC sense. The comparison was between manual surgical techniques and PSI-assisted techniques in cadavers.

6. Standalone (Algorithm Only) Performance Study

  • Was a standalone study done? Not explicitly in terms of an "algorithm only" performance study. The device itself is a physical, patient-specific instrument derived from a digital plan. The document describes "Cadaveric testing" which evaluates the combined PSI system (planning software output + physical guide) in a simulated surgical environment, not just the planning algorithm in isolation from its physical manifestation or use.
  • The "Intra- and Inter-Designer Variability testing" and "Mechanical Integrity" tests are standalone evaluations of aspects of the device's design and manufacturing, but not of the surgical guidance algorithm's performance on its own.

7. Type of Ground Truth Used

  • For Pre-operative Planning: The ground truth for the design of the PSI System is based on a "pre-operative planning validated by the surgeon" using CT images and anatomical landmarks. This can be considered a form of expert consensus/validation on the desired surgical outcome/instrument design.
  • For Cadaveric Testing: The "ground truth" for the cadaveric study would be the actual anatomical targets and the achieved drill/cut placements, compared to the planned placements and traditional manual techniques. While experts (surgeons) would perform and assess these, the ultimate "truth" is the physical reality within the cadaver. The document implies comparison to "manual techniques" as a reference.

8. Sample Size for the Training Set

The document does not provide any information regarding a "training set sample size." The Archer PSI System is described as being "designed from a draft treatment plan" and "manufactured from a pre-operative planning validated by the surgeon." This suggests a patient-specific design process rather than a machine learning model trained on a large dataset. The underlying "Archer™ 3D Targeting' platform (K213779)" (a separate cleared device) would be the system performing the planning, and its own 510(k) might contain training data details if it uses AI/ML. However, for the Archer PSI System itself, no training set information is present.

9. How Ground Truth for the Training Set Was Established

As no training set is mentioned for the Archer PSI System in this document, no information is provided on how its ground truth would have been established.

U.S. Food & Drug Administration

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

May 12, 2025

3D-Side
Florence Allé
Design & Development Director
Daalstraat 4
Maaseik, 3680
Belgium

Re: K243509
Trade/Device Name: Archer PSI System
Regulation Number: 21 CFR 888.3660
Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Code: QHE, KWS, PHX
Dated: April 11, 2025
Received: April 11, 2025

Dear Florence Allé :

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

FDA U.S. FOOD & DRUG ADMINISTRATION

May 12, 2025

3D-Side
Florence Allé
Design & Development Director
Daalstraat 4
Maaseik, 3680
Belgium

Re: K243509
Trade/Device Name: Archer PSI System
Regulation Number: 21 CFR 888.3660
Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Code: QHE, KWS, PHX
Dated: April 11, 2025
Received: April 11, 2025

Dear Florence Allé :

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243509 - Florence Allé Page 3

Sincerely,

Farzana Sharmin -S Digitally signed by Farzana Sharmin -S Date: 2025.05.12 16:26:38 -04'00'

Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Submission Number (if known)
K243509
Device Name
Archer PSI System

Indications for Use (Describe)

Archer^R PSI System is indicated as an orthopedic instrument to assist the physician in the intraoperative positioning of total shoulder replacement components and in guiding the drill and the cut of the bone.

Archer^R PSI System must only be used conjointly with Archer™ CSR Total Shoulder (K152825, K173812, K181287, K182500, K191811), Catalyst EA Convertible Stemmed Shoulder (K222317) and Archer™ R1 Reverse (K202611, K211991, K213349, K223655, K232583) components in the context of primary total shoulder replacement and following a delto-pectoral approach only. Archer^R PSI System is manufactured from a pre-operative planning validated by the surgeon in the 'Archer™ 3D Targeting' platform (K213779). Archer^R PSI System is indicated for patient population fulfilling the Archer™ CSR Total Shoulder, Catalyst EA Convertible Stemmed Shoulder and Archer™ R1 Reverse indications and for which CT images are available with identifiable placement anatomical landmarks and compliant with imaging protocol provided by Archer 3D Targeting.

The device is intended for single use only.

The device is intended for adult patients.

The device has to be used by a physician trained in the performance of surgery.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Page 1 of 4

K243509

510(K) SUMMARY (21CFR807.92)

SUBMITTER

Company Name: 3D-Side
Establishment registration number: 3013561205
Address: Rue André Dumont 5
1435 Mont-Saint-Guibert, BE

Submitter contact person: Florence Allé, Laurent Paul
Submitter contact e-mail: florence.alle@3dside.eu; laurent.paul@3dside.eu

Correspondent contact persons: Mieke Janssen, Florence Allé, Laurent Paul
Correspondent contact e-mail address: mieke@pra-consulting.be; florence.alle@3dside.eu; laurent.paul@3dside.eu
Correspondent contact telephone: 0032 496 56 41 31

Summary date: May 8, 2025

DEVICE NAME

Device trade name: Archer PSI System
Common name: Shoulder Arthroplasty Implantation System
Classification name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulation number: 21 CFR 888.3660
Classification product code: QHE
Subsequent product code: KWS, PHX

LEGALLY MARKETED PREDICATE DEVICE(S)

The predicate device to which substantial equivalence is claimed:

ItemDescription
Device Classification NameShoulder Prosthesis, Reverse Configuration
Device Trade NameMyShoulder™ Placement Guides
510(k) numberK190738
Original ApplicantMedacta International SA
Regulation Number21 CFR 888.3660, 21 CFR 888.3690, and 21 CFR 888.3670
Decision dateNovember 13, 2019
Classification product codePHX
Classification product code (secondary)KWS, HSD, MBF
510(k) Review PanelOrthopedic

Reference devices:

Trade name510(k) numberRegulation numberClassification product code
Catalyst CSR Shoulder SystemK152825888.3650KWT
Catalyst CSR 3 Peg GlenoidsK173812888.3650KWT
Catalyst CSR Shoulder SystemK181287888.3650KWT
Catalyst CSR Press-Fit Humeral ComponentsK182500888.3650KWT
Catalyst OrthoScience CSR Shoulder SystemK191811888.3650KWT
Catalyst EA Convertible Stemmed ShoulderK222317888.3660KWS
Catalyst OrthoScience R1 Reverse Shoulder SystemK202611888.3660PHX
Catalyst OrthoScience R1 Reverse Shoulder SystemK211991888.3660PHX
Catalyst OrthoScience R1 Reverse Shoulder SystemK213349888.3660PHX
Catalyst OrthoScience R1 Reverse Shoulder SystemK223655888.3660PHX
Catalyst OrthoScience R1 Reverse Shoulder SystemK232583888.3660PHX

DEVICE DESCRIPTION SUMMARY

The "Archer PSI System" device is a patient-matched additively manufactured single use surgical instrument (PSI). Archer PSI System is an instrument set containing a glenoid guide and its bone model and/or a humeral guide and its bone model. This patient-specific medical device is designed to fit the patient's anatomy to transfer a patient-specific pre-operative plan to the operating room. It is intended for surgical interventions in orthopaedic procedures for total shoulder arthroplasty.

The Archer PSI system instruments are designed from a draft treatment plan available via the Archer™ 3D Targeting' platform. Based on computed tomography (CT) of the shoulder anatomy, 3D CAD models of the bones and positioning and sizing of the glenoid and humeral components are submitted for evaluation to the surgeon. Upon the surgeon's approval, the guides and bone models are designed based on the validated planning and are manufactured using additive manufacturing.

Page 6

Page 2 of 4

K243509

Reference devices:

Trade name510(k) numberRegulation numberClassification product code
Catalyst CSR Shoulder SystemK152825888.3650KWT
Catalyst CSR 3 Peg GlenoidsK173812888.3650KWT
Catalyst CSR Shoulder SystemK181287888.3650KWT
Catalyst CSR Press-Fit Humeral ComponentsK182500888.3650KWT
Catalyst OrthoScience CSR Shoulder SystemK191811888.3650KWT
Catalyst EA Convertible Stemmed ShoulderK222317888.3660KWS
Catalyst OrthoScience R1 Reverse Shoulder SystemK202611888.3660PHX
Catalyst OrthoScience R1 Reverse Shoulder SystemK211991888.3660PHX
Catalyst OrthoScience R1 Reverse Shoulder SystemK213349888.3660PHX
Catalyst OrthoScience R1 Reverse Shoulder SystemK223655888.3660PHX
Catalyst OrthoScience R1 Reverse Shoulder SystemK232583888.3660PHX

DEVICE DESCRIPTION SUMMARY

The "Archer PSI System" device is a patient-matched additively manufactured single use surgical instrument (PSI). Archer PSI System is an instrument set containing a glenoid guide and its bone model and/or a humeral guide and its bone model. This patient-specific medical device is designed to fit the patient's anatomy to transfer a patient-specific pre-operative plan to the operating room. It is intended for surgical interventions in orthopaedic procedures for total shoulder arthroplasty.

The Archer PSI system instruments are designed from a draft treatment plan available via the Archer™ 3D Targeting' platform. Based on computed tomography (CT) of the shoulder anatomy, 3D CAD models of the bones and positioning and sizing of the glenoid and humeral components are submitted for evaluation to the surgeon. Upon the surgeon's approval, the guides and bone models are designed based on the validated planning and are manufactured using additive manufacturing.

Page 2 of 4

Page 7

Page 3 of 4

K243509

INDICATIONS FOR USE

Archer^R PSI System is indicated as an orthopedic instrument to assist the physician in the intra-operative positioning of total shoulder replacement components and in guiding the drill and the cut of the bone.

Archer^R PSI System must only be used conjointly with Archer™ CSR Total Shoulder (K152825, K173812, K181287, K182500, K191811), Catalyst EA Convertible Stemmed Shoulder (K222317) and Archer™ R1 Reverse (K202611, K211991, K213349, K223655, K232583) components in the context of primary total shoulder replacement and following a delto-pectoral approach only.

Archer^R PSI System is manufactured from a pre-operative planning validated by the surgeon in the 'Archer™ 3D Targeting' platform (K213779). Archer^R PSI System is indicated for patient population fulfilling the Archer™ CSR Total Shoulder, Catalyst EA Convertible Stemmed Shoulder and Archer™ R1 Reverse indications and for which CT images are available with identifiable placement anatomical landmarks and compliant with imaging protocol provided by Archer 3D Targeting.

The device is intended for single use only.
The device is intended for adult patients.
The device has to be used by a physician trained in the performance of surgery.

COMPARISON WITH THE PREDICATE DEVICE

The subject and predicate devices share the same intended use. Both are orthopedic instruments designed to assist with instrumentation positioning during total shoulder arthroplasty procedures (anatomic and reverse configurations) performed via the deltopectoral surgical approach.

They are manufactured from polyamide using additive manufacturing techniques, based on pre-operative planning validated by the surgeon through a pre-operative planning software. These instruments are compatible with specific total shoulder replacement implant components and their associated accessories.

The subject and predicate devices are comparable in terms of design characteristics, including contact surfaces and key functional features. However, the cut slot characteristics of the anatomic humeral component differ between the two systems due to differences in the geometry of the compatible humeral components.

Page 3 of 4

Page 8

Page 4 of 4

K243509

SUMMARY OF NON-CLINICAL AND CLINICAL PERFORMANCE DATA

The following assessments have been conducted to demonstrate that the subject device is substantially equivalent to the reference devices and performs in accordance with its intended use:

  • Biocompatibility assessment
  • Cleaning and sterilization validations
  • Manufacturing cleaning validation
  • Packaging and Shelf-life validation
  • Intra- and Inter-Designer Variability testing
  • Mechanical Integrity (Post-Processing) testing
  • Debris Generation testing

Cadaveric testing was executed to demonstrate the substantial equivalence between the two techniques (manual and PSI, for both anatomic and for reverse techniques) to validate the use of the subject device for the addressed surgical techniques.

CONCLUSION

The nonclinical and cadaver testing demonstrates that the subject device, Archer PSI System, is substantially equivalent to the predicate device cleared under K190738.

Page 4 of 4

§ 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”