(121 days)
The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable.
- Osteoarthritis
- Avascular Necrosis ●
- Rheumatoid Arthritis ●
- Post-traumatic Arthritis
- Correction of functional deformity
The Catalyst CSR humeral and glenoid implants are intended for cemented use.
The Catalyst CSR Press-Fit humeral implants are intended for uncemented or cemented use.
The Catalyst CSR Shoulder System is a bone preserving total shoulder prosthesis designed for use in patients where the humeral head, neck and glenoid vault are of sufficient bone stock and there is an intact or reconstructable rotator cuff. The design requires minimal bone resection, thus giving the patient an alternative to other total shoulder designs where more bone is removed.
This submission adds Press-Fit Humeral Components to the CSR Shoulder System. The press-fit humeral components have a non-spherical polished surface for articulation with the glenoid component or the glenoid cavity of the scapula. The humeral components incorporate 4 pegs which assist with alignment and provide rotational stability. The four plane geometry of the back side of the humeral component matches four cut surfaces on the humeral head to recreate the geometry and thickness of the removed bone. The CSR Press-Fit Humeral Components are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75. The undersurface of the head and the proximal portion of the alignment pegs are coated with a plasma sprayed coating of CP Ti conforming to ASTM F1580.
The CSR humeral components are compatible with previously cleared CSR 3 Peg glenoid components.
Here's an analysis of the provided text regarding the acceptance criteria and study information for the Catalyst CSR Press-Fit Humeral Components:
Summary of Device and Context:
The document is an FDA 510(k) clearance letter and a 510(k) Summary for the Catalyst CSR Press-Fit Humeral Components, which are an addition to the existing Catalyst CSR Shoulder System. This device is a shoulder prosthesis intended for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint, where the humeral head, neck, and glenoid vault have sufficient bone stock, and the rotator cuff is intact or reconstructable. The key new feature here is the "Press-Fit" capability, allowing for uncemented or cemented use, and the device incorporates a plasma sprayed coating of CP Ti. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.
Detailed Analysis of Acceptance Criteria and Study Information:
Based on the provided text, the focus is on non-clinical testing to demonstrate substantial equivalence, as clinical testing was deemed unnecessary.
- Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Fixation strength in shear | Met the pre-determined acceptance criterion through Static Shear Testing. |
| Endotoxin limit (Bacterial Endotoxin Testing) | Met the acceptable endotoxin limit of <10EU/component or <20EU/device construct. |
| Torque-out (engineering analysis) | Deemed applicable from predicate Catalyst CSR Humeral Components. |
-
Sample Size Used for the Test Set and Data Provenance:
- Static Shear Testing: The document states "Static shear testing was conducted," but it does not specify the sample size for this test.
- Bacterial Endotoxin Testing: The document states "Bacterial Endotoxin Testing was performed," but it does not specify the sample size for this test.
- Data Provenance: The document does not provide information on data provenance (e.g., country of origin, retrospective/prospective). Given these are non-clinical lab tests, it's implied they were conducted in a controlled lab environment by the manufacturer or a contracted lab.
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- This question is not applicable as the document describes non-clinical engineering and bench tests, not studies requiring expert interpretation for ground truth.
-
Adjudication Method for the Test Set:
- This question is not applicable for the same reasons as above. No human expert adjudication was involved in the non-clinical tests described.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, an MRMC comparative effectiveness study was not done.
- The document explicitly states: "Clinical testing was not necessary to demonstrate substantial equivalence of the Catalyst CSR Press-Fit Humeral Components to the predicate devices."
- Therefore, there is no information on human reader improvement with or without AI assistance.
-
Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
- This question is not applicable. The device is a physical shoulder implant, not an algorithm or AI system. The testing performed was related to its physical properties and biocompatibility.
-
Type of Ground Truth Used:
- For the non-clinical tests mentioned:
- Static Shear Testing: The "ground truth" would be the universally accepted principles of mechanical engineering and material science, measured against a pre-determined acceptance criterion (likely based on established standards for orthopedic implants).
- Bacterial Endotoxin Testing: The "ground truth" is the predefined acceptable endotoxin limit (<10EU/component or <20EU/device construct), which is based on regulatory standards for medical devices to ensure safety.
- Torque-out: "Ground truth" from an engineering analysis, likely based on established mechanical principles and data from the predicate device.
- For the non-clinical tests mentioned:
-
Sample Size for the Training Set:
- This question is not applicable. The device is a physical implant, not an AI model requiring a training set. The term "training set" is typically used in the context of machine learning or AI development.
-
How the Ground Truth for the Training Set Was Established:
- This question is not applicable for the same reasons as above.
Conclusion:
The FDA clearance for the Catalyst CSR Press-Fit Humeral Components relied on non-clinical testing demonstrating substantial equivalence to predicate devices. The key tests reported were static shear testing and bacterial endotoxin testing, both of which met pre-determined acceptance criteria. Clinical testing, which would involve human performance or outcomes, was explicitly stated as "not necessary." Therefore, many of the questions related to clinical study design, expert involvement, and AI performance are not relevant to this specific clearance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 11, 2019
Catalyst OrthoScience, Inc. Dale Davison Sr. VP of Manufacturing & Product Development 14710 Tamiami Trail North Naples, Florida 34110
Re: K182500
Trade/Device Name: Catalyst CSR Press-Fit Humeral Components Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder Joint Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KWT, HSD Dated: December 5, 2018 Received: December 7, 2018
Dear Dale Davison:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vesa
Digitally signed
by Vesa Vuniqi -
S
Vuniqi -S Date: 2019.01.11
14:05:05 -05'00'
Mark N. Melkerson For: Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K182500
Device Name
Catalyst OrthoScience CSR Shoulder System
Indications for Use (Describe)
The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable.
- Osteoarthritis
- Avascular Necrosis ●
- Rheumatoid Arthritis ●
- Post-traumatic Arthritis
- Correction of functional deformity
The Catalyst CSR humeral and glenoid implants are intended for cemented use.
The Catalyst CSR Press-Fit humeral implants are intended for uncemented or cemented use.
Type of Use (Select one or both, as applicable)
区Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
| K182500 | 510(k) Summary |
|---|---|
| Prepared: | December 4, 2018 |
| Submitter: | Catalyst OrthoScience, Inc.14710 Tamiami Trail NorthNaples, FL 34110 |
| Contact: | Dale DavisonSr. VP of Manufacturing & Product DevelopmentCatalyst OrthoScience, Inc.1-239-325-9976 ext 102ddavison@catalystortho.com |
| Proprietary Name: | Catalyst CSR Shoulder System |
| Common Name: | Shoulder Prosthesis |
| Classification Names: | 21 CFR 888.3650: Shoulder joint metal/polymer non-constrainedcemented prosthesis; Class II21 CFR 888.3690: Shoulder joint humeral (hemi-shoulder) metallicuncemented prosthesis; Class II |
| Product Codes: | KWT, HSD |
| Substantially Equivalent Devices | Catalyst OrthoScience CSR Shoulder System, K152825Catalyst OrthoScience CSR Shoulder System, K181287Exactech Equinoxe Resurfacing System, K131298Biomet Copeland Resurfacing Heads, K003044 |
Device Description:
The Catalyst CSR Shoulder System is a bone preserving total shoulder prosthesis designed for use in patients where the humeral head, neck and glenoid vault are of sufficient bone stock and there is an intact or reconstructable rotator cuff. The design requires minimal bone resection, thus giving the patient an alternative to other total shoulder designs where more bone is removed.
{4}------------------------------------------------
This submission adds Press-Fit Humeral Components to the CSR Shoulder System. The press-fit humeral components have a non-spherical polished surface for articulation with the glenoid component or the glenoid cavity of the scapula. The humeral components incorporate 4 pegs which assist with alignment and provide rotational stability. The four plane geometry of the back side of the humeral component matches four cut surfaces on the humeral head to recreate the geometry and thickness of the removed bone. The CSR Press-Fit Humeral Components are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75. The undersurface of the head and the proximal portion of the alignment pegs are coated with a plasma sprayed coating of CP Ti conforming to ASTM F1580.
The CSR humeral components are compatible with previously cleared CSR 3 Peg glenoid components.
Intended Use / Indications:
The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable.
- Osteoarthritis ●
- Avascular Necrosis
- Rheumatoid Arthritis ●
- . Post-traumatic Arthritis
- Correction of functional deformity
The Catalyst CSR humeral and glenoid implants are intended for cemented use.
The Catalyst CSR Press-Fit humeral implants are intended for uncemented or cemented use.
Summary of Technologies/Substantial Equivalence:
The Catalyst CSR Press-Fit Humeral Components are substantially equivalent to the predicate devices in regards to intended use and indications, materials, design and sizes. The press-fit indication and the addition of a plasma sprayed coating are similar to other predicate devices and do not rase new types of safety and effectiveness questions, nor are there new technological issues.
{5}------------------------------------------------
Non-Clinical Testing:
Static shear testing was conducted and demonstrated that the fixation strength of the Catalyst CSR Press-Fit Humeral Components, tested in shear, met the pre-determined acceptance criterion. An engineering analysis of torque-out of the predicate Catalyst CSR Humeral Components was determined to be applicable to the Catalyst CSR Press-Fit Humeral Components as well. Bacterial Endotoxin Testing was performed. The Catalyst CSR Press-Fit Humeral Components met the acceptable endotoxin limit of <10EU/component or <20EU/device construct.
Clinical Testing:
Clinical testing was not necessary to demonstrate substantial equivalence of the Catalyst CSR Press-Fit Humeral Components to the predicate devices.
N/A