K Number
K243499
Manufacturer
Date Cleared
2025-06-04

(204 days)

Product Code
Regulation Number
866.1640
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NG-Test® CTX-M MULTI is an in vitro rapid and visual immunochromatographic assay for the qualitative detection of CTX-M enzymes (groups 1, 2, 8, 9, and 25) from pure colonies of Enterobacterales suspected of ESBL production when grown on the following media:

  • 5% sheep blood agar or MacConkey agar (16-24 hours)
  • HardyCHROM™ ESBL agar (18-24 hours)

The NG-Test® CTX-M MULTI is intended as an aid for infection control in the detection of CTX-M enzymes-producing organisms (Enterobacterales) in healthcare settings. NG-Test® CTX-M MULTI is not intended to guide or monitor treatment. A positive or negative NG-Test® CTX-M MULTI test result does not rule out the presence of other mechanisms of antibiotic resistance. NG-Test® CTX-M MULTI should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing.

Device Description

NG-Test® CTX-M MULTI is an in vitro rapid and visual immunochromatographic assay for the qualitative detection of CTX-M enzymes (groups 1, 2, 8, 9, and 25) from pure colonies of Enterobacterales suspected of ESBL production after culturing on agar and processed in an extraction buffer. The device consists of a sample port, sample and conjugate pad, and nitrocellulose test strip, which are contained within a plastic cassette, in addition to reagents for liquid extraction. The result can be read 15 minutes after adding the sample to the sample well. A positive result on the NG-Test® CTX-M MULTI occurs when two red lines appear, one on the control (C) region and one on the test (T) region. A negative result occurs when only the control line is observed and indicates the sample does not contain any target CTX-M enzymes, or the CTX-M enzymes are present at a non-detectable level. If the control line does not appear, the test result is invalid.

Monoclonal antibodies that recognize the five major CTX-M groups are immobilized on a nitrocellulose membrane. Free monoclonal antibodies are present in the conjugate pad and labeled with colloidal gold. Upon addition of the processed sample to the sample pad, the capillary action of the nitrocellulose draws the sample through the mobile antibodies in the conjugate pad and the immobile antibodies on the test strip. The immobilized control antibodies capture any mobile antibodies that do not bind to the test line.

AI/ML Overview

This document describes the performance of the NG-Test® CTX-M MULTI device, an in vitro rapid immunochromatographic assay for detecting CTX-M enzymes.

Here is a breakdown of the requested information:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the device are implied by the reported sensitivity and specificity values. While explicit "acceptance criteria" are not listed as target percentages, the study results demonstrate high performance that led to clearance.

Table of Performance Data (Clinical Study - Blood and MacConkey Agar):

MetricAcceptance Criteria (Implied)Reported Performance
SensitivityHigh (e.g., >95%)100.0% (95% CI: 97.5% - 100.0%)
SpecificityHigh (e.g., >95%)99.4% (95% CI: 96.5% - 99.9%)

Table of Performance Data (Seeded Study - HC ESBL Agar):

MetricAcceptance Criteria (Implied)Reported Performance
SensitivityHigh (e.g., >95%)100.0% (95% CI: 97.5% - 100.0%)
SpecificityHigh (e.g., >90%)97.7% (95% CI: 87.9% - 99.6%)

Note: The lower bound of the specificity for HC ESBL agar (87.9%) is acknowledged as being below 90% due to a lower quantity of CTX-M negative isolates in this specific part of the study, and additional CTX-M negative isolates were evaluated in the cross-reactivity study to compensate.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Clinical Study (Blood and MacConkey Agar): 309 Enterobacterales isolates.
    • Seeded Study (HardyCHROM™ ESBL Agar): 193 clinical isolates.
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but the study was conducted at "three geographically diverse hospitals," implying the data is from a clinical setting.
    • Retrospective or Prospective: A mix was used:
      • "Prospectively-collected" bacterial isolates were used for the performance evaluation.
      • "Stock bacterial isolates" were also used. This suggests a combination of prospective collection and retrospective use of archived isolates.

3. Number of Experts and Qualifications for Ground Truth

  • The document does not specify the number of experts used to establish ground truth or their qualifications. The ground truth relies on laboratory methods (PCR and AST).

4. Adjudication Method for the Test Set

  • The document does not describe an adjudication method for the test set results. The comparison is between the device's reading and the reference method (PCR and AST).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • A MRMC comparative effectiveness study involving human readers with vs. without AI assistance was not conducted. This device is an immunochromatographic assay, not an AI-based diagnostic with human-in-the-loop performance.
  • However, a reproducibility study involved multiple readers: "The testing was done with one operator and two readers, blinded to each other's results, per site." This was to demonstrate reproducibility of the device's results, not to compare human performance with and without the device.

6. Standalone (Algorithm Only) Performance

  • This device is a standalone diagnostic test (an immunochromatographic assay). Its performance is evaluated directly without a human-in-the-loop component for result interpretation beyond visual reading. The reported sensitivity and specificity refer to the device's performance as a standalone test.

7. Type of Ground Truth Used

  • Reference standard:
    • PCR (Polymerase Chain Reaction): Used to detect blaCTX-M genes, considered the definitive molecular method. Isolates with a positive PCR result were also sequenced to determine the CTX-M variant.
    • Antimicrobial Susceptibility Testing (AST): Performed according to CLSI M100, 34th edition breakpoints. A positive comparator method result was defined as any Enterobacterales that produced a positive PCR result and was not susceptible to at least one of the antimicrobial agents.
    • Identification (ID): Performed using FDA-cleared ID systems.

8. Sample Size for the Training Set

  • The document does not specify a separate "training set" size. As this is an immunochromatographic assay, it doesn't involve machine learning model training in the typical sense. The analytical reactivity study and analytical specificity study essentially serve as extensive performance characterization using known strains, which could be conceptually similar to a "development" or "characterization" set for more traditional devices.
    • Analytical Reactivity: 57 strains of Enterobacterales characterized to harbor CTX-M enzymes.
    • Analytical Specificity: 55 organisms with other resistance mechanisms.

9. How the Ground Truth for the Training Set Was Established

  • For the analytical reactivity and analytical specificity studies (which characterize the device's performance with known strains, akin to a training/development set in ML contexts):
    • Strains were "characterized to harbor CTX-M enzymes" or "exhibit antimicrobial resistance mechanisms other than CTX-M." This characterization would typically involve established molecular methods (e.g., PCR, sequencing) and phenotypic susceptibility testing performed by a reference lab or research institution. The document states "Each organism was incubated..." and tested, implying these were well-characterized isolates.

FDA 510(k) Clearance Letter - NG-Test CTX-M MULTI

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.07.05
Silver Spring, MD 20993
www.fda.gov

June 4, 2025

NG Biotech
℅ Anna Klavins
Director of Technical Services and R&D
Hardy Diagnostics
1430 West McCoy Lane
Santa Maria, California 93455

Re: K243499
Trade/Device Name: NG-Test CTX-M MULTI
Regulation Number: 21 CFR 866.1640
Regulation Name: Antimicrobial Susceptibility Test Powder
Regulatory Class: Class II
Product Code: PTJ
Dated: May 8, 2025
Received: May 9, 2025

Dear Anna Klavins:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

2K243499 - Anna Klavins Page

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

3K243499 - Anna Klavins Page

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D. (ABMM)
Chief,
General Bacteriology and Antimicrobial Susceptibility Branch
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K243499

Device Name
NG-Test® CTX-M MULTI

Indications for Use (Describe)

NG-Test® CTX-M MULTI is an in vitro rapid and visual immunochromatographic assay for the qualitative detection of CTX-M enzymes (groups 1, 2, 8, 9, and 25) from pure colonies of Enterobacterales suspected of ESBL production when grown on the following media:

  • 5% sheep blood agar or MacConkey agar (16-24 hours)
  • HardyCHROM™ ESBL agar (18-24 hours)

The NG-Test® CTX-M MULTI is intended as an aid for infection control in the detection of CTX-M enzymes-producing organisms (Enterobacterales) in healthcare settings. NG-Test® CTX-M MULTI is not intended to guide or monitor treatment. A positive or negative NG- Test® CTX-M MULTI test result does not rule out the presence of other mechanisms of antibiotic resistance. NG-Test® CTX-M MULTI should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary

November 11th, 2024

The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:

I. General Information

Applicant: Milovan Stankov-Puges
CEO
NG Biotech
Z.A. Courbouton, secteur 1, Atelier relais le Tremplin
Guipry 35480
France
Phone: +33 (0) 2 23 30 17 83
Fax: +33 (0) 9 71 70 53 10
E-mail: msp@ngbiotech.com

Correspondent: Anna Klavins
Director of Technical Services and R&D
Hardy Diagnostics
1430 W. McCoy Lane
Santa Maria, CA 93455
Phone: 805-346-2766 x 5752
E-mail: KlavinsA@hardydiagnostics.com

II. Device Information

Device Trade Name: NG-Test® CTX-M MULTI
Common Name: NG-Test® CTX-M MULTI
Classification Name: Antimicrobial Susceptibility Test Powder
Regulation Number: 21 CFR 866.1640
Regulatory Class: II
Product Code: PTJ

III. Predicate Device

NG-Test CARBA 5 (K191889)

IV. Device Description

NG-Test® CTX-M MULTI is an in vitro rapid and visual immunochromatographic assay for the qualitative detection of CTX-M enzymes (groups 1, 2, 8, 9, and 25) from pure colonies of Enterobacterales suspected of ESBL production after culturing on agar and processed in an extraction buffer. The device consists of a sample port, sample and conjugate pad, and nitrocellulose test strip, which are contained within a plastic cassette, in addition to reagents for liquid extraction. The result can be read 15 minutes after adding the sample to the sample well. A positive result on the NG-Test® CTX-M MULTI occurs when two red lines appear, one on the

Page 1

Page 6

control (C) region and one on the test (T) region. A negative result occurs when only the control line is observed and indicates the sample does not contain any target CTX-M enzymes, or the CTX-M enzymes are present at a non-detectable level. If the control line does not appear, the test result is invalid.

Monoclonal antibodies that recognize the five major CTX-M groups are immobilized on a nitrocellulose membrane. Free monoclonal antibodies are present in the conjugate pad and labeled with colloidal gold. Upon addition of the processed sample to the sample pad, the capillary action of the nitrocellulose draws the sample through the mobile antibodies in the conjugate pad and the immobile antibodies on the test strip. The immobilized control antibodies capture any mobile antibodies that do not bind to the test line.

V. Intended Use/Indications of Use

NG-Test® CTX-M MULTI is an in vitro rapid and visual immunochromatographic assay for the qualitative detection of CTX-M enzymes (groups 1, 2, 8, 9, and 25) from pure colonies of Enterobacterales suspected of ESBL production when grown on the following media:

  • 5% sheep blood agar or MacConkey agar (16-24 hours)
  • HardyCHROM™ ESBL agar (18-24 hours)

The NG-Test® CTX-M MULTI is intended as an aid for infection control in the detection of CTX-M enzymes-producing organisms (Enterobacterales) in healthcare settings. NG-Test® CTX-M MULTI is not intended to guide or monitor treatment. A positive or negative NG-Test® CTX-M MULTI test result does not rule out the presence of other mechanisms of antibiotic resistance. NG-Test® CTX-M MULTI should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing.

VI. Substantial Equivalence Comparison

The following table demonstrates the substantial equivalence comparison of NG-Test® CTX-M MULTI.

DevicePredicate
NG-Test® CTX-M MULTING-Test® CARBA 5
510(k) DetailsN/AK191889
Product Code PTJProduct Code PTJ
21 CFR 866.164021 CFR 866.1640
"Antimicrobial Susceptibility Test Powder""Antimicrobial Susceptibility Test Powder"
Class IIClass II
MicrobiologyMicrobiology
Intended UseNG-Test® CTX-M MULTI is an in vitro rapid and visual immunochromatographic assay for the qualitative detection of CTX-M enzymes (groups 1, 2, 8, 9, and 25)NG-Test® CARBA 5 is an in vitro rapid and visual multiplex immunochromatographic assay for the qualitative detection and differentiation of five common carbapenemases (KPC,

Page 2

Page 7

from pure colonies of Enterobacterales suspected of ESBL production when grown on the following media:OXA-48-like, VIM, IMP, and NDM) from carbapenem non-susceptible pure bacterial colonies when grown on the following media:
• 5% sheep blood agar or MacConkey agar (16-24 hours).• 5% sheep blood agar or MacConkey agar (16-24 hours) for testing Enterobacterales (formerly Enterobacteriaceae) and Pseudomonas aeruginosa.
• HardyCHROM™ ESBL agar (18-24 hours).• HardyCHROM™ CRE agar (18-24 hours) for testing E. coli and KES (Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Enterobacter cloacae complex and Serratia marcescens).
The NG-Test® CTX-M MULTI is intended as an aid for infection control in the detection of CTX-M enzymes-producing organisms (Enterobacterales) in healthcare settings. NG-Test® CTX-M MULTI is not intended to guide or monitor treatment. A positive or negative NG-Test® CTX-M MULTI test result does not rule out the presence of other mechanisms of antibiotic resistance. NG-Test® CTX-M MULTI should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing.The NG-Test® CARBA 5 is intended as an aid for infection control in the detection of carbapenemase-producing Enterobacterales and Pseudomonas aeruginosa in healthcare settings. NG-Test® CARBA 5 is not intended to guide or monitor treatment for carbapenem non-susceptible bacterial infections. A positive or negative NG-Test® CARBA 5 test result does not rule out the presence of other mechanisms of antibiotic resistance. NG-Test® CARBA 5 should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing.

General Device Characteristic Similarities

InterpretationVisualSame
ControlsBuilt-in procedural control in every test stripSame
Intended Culture Media5% sheep blood agar, MacConkey agarSame
Sample TypeBacterial coloniesSame
Inoculum PreparationBy touching 3 colonies with a 1 µL loopSame

General Device Characteristic Differences

Intended Culture MediaHardyCHROM™ ESBL agarHardyCHROM™ CRE agar
AnalyteCTX-M enzymesCarbapenemase enzymes (KPC, OXA-48-like, VIM, IMP, NDM)

Page 3

Page 8

VII. Performance Data

The performance of the NG-Test® CTX-M MULTI was evaluated at three geographically diverse hospitals with prospectively-collected and stock bacterial isolates of Enterobacterales. The NG-Test® CTX-M MULTI performance was compared to reference method PCR and antimicrobial susceptibility testing (AST) according to the breakpoints described in the CLSI M100, 34th edition. The identification (ID) and AST of organisms was performed using FDA-cleared ID and AST systems. NG-Test® CTX-M MULTI quality control was performed daily in parallel with each day of testing. A total of 309 Enterobacterales were tested using phenotypic AST (disk diffusion and MIC) and reference method PCR. Isolates with a positive PCR result were also sequenced to determine the CTX-M variant. A positive reaction in the comparator method was defined as any Enterobacterales that produced a positive PCR result and was not susceptible to at least one of the antimicrobial agents. 308 isolates had concordant results between the NG-Test® CTX-M MULTI and reference method. One isolate of Citrobacter amalonaticus yielded a false-positive result.

The sensitivity for Enterobacterales from all sites combined for colonies tested on NG-Test® CTX-M MULTI from blood agar and MacConkey agar was 100.0% (95% confidence interval [CI]: 97.5% - 100.0%). The specificity was 99.4% (95% CI: 96.5% - 99.9%).

The sensitivity and specificity from the clinical study are summarized in Table 1.

Table 1. Comparison of NG-Test® CTX-M MULTI (Blood and MacConkey Agar) to Comparator Method – Study Device vs. Reference Method

Total IsolatesTPFP¹FNTNSensitivityLow 95%High 95%SpecificityLow 95%High 95%
30915110157100.097.5100.099.496.599.9

¹One isolate of Citrobacter amalonaticus was positive on NG-Test® CTX-M MULTI, not susceptible to at least one antimicrobial tested, and negative for blaCTX-M by PCR.

The CTX-M variants detected during the NG-Test® CTX-M MULTI clinical study are outlined in Table 2.

Table 2. Overall CTX-M Variants Detected by NG-Test® CTX-M MULTI in Clinical Study (Blood Agar and MacConkey Agar)

CTX-M GroupOverall Variants Detected in US Clinical Trial
1CTX-M-1, -3, -15, -55, -178, -232, -V263A¹
2CTX-M-2
8CTX-M-8
9CTX-M-14, -24, -27, -65

¹The CTX-M variant in this isolate features a V263A substitution (valine to alanine at position 263).

The clinical trial isolates used to evaluate the NG-Test® CTX-M MULTI device from blood agar and MacConkey agar at the clinical sites were also used to evaluate the performance of NG-Test® CTX-M MULTI from HardyCHROM™ ESBL (HC ESBL) agar. NG-Test® CTX-M MULTI quality control was performed daily in parallel with each day of testing.

Page 4

Page 9

A total of 193 clinical isolates that met HC ESBL agar claims were tested with the NG-Test® CTX-M MULTI device using colonies recovered from culture of seeded stool samples in accordance with HC ESBL agar claims. Among the tested Enterobacterales, 192 showed concordant results between the NG-Test® CTX-M MULTI and the reference method. One false-positive result was identified, corresponding to the same discrepant isolate from the testing with blood agar and MacConkey agar.

The sensitivity for Enterobacterales tested on the NG-Test® CTX-M MULTI from HC ESBL agar, processed with both types of stool specimens, was 100.0% (95% CI: 97.5% - 100.0%). The specificity was 97.7% (95% CI: 87.9% - 99.6%).

The sensitivity and specificity from the seeded study are summarized in Table 3.

Table 3. Seeded Study Comparison of NG-Test® CTX-M MULTI (HC ESBL Agar – Raw and C&S Stool) to Comparator Method – Study Device vs. Reference Method

Total IsolatesTPFP¹FNTNSensitivityLow 95%High 95%SpecificityLow 95%High 95%
1931501042100.097.5100.097.787.9²99.6

¹One isolate of Citrobacter amalonaticus was positive on NG-Test® CTX-M MULTI, not susceptible to at least one antimicrobial tested, and negative for blaCTX-M by PCR.

²Lower bound is below 90% due to lower quantity of CTX-M negative clinical isolates that meet HC ESBL agar claims. Additional CTX-M negative isolates were evaluated in the cross-reactivity study.

VIII. Analytical Reactivity

NG-Test® CTX-M MULTI was evaluated using 57 strains of Enterobacterales characterized to harbor CTX-M enzymes. Each organism was incubated for 16 to 24 hours on blood agar and MacConkey agar, and for 18 to 24 hours on HC ESBL agar at 35°C. Each isolate was tested in triplicate from each type of media, and results were read 15 minutes after inoculating the extraction buffer mixed with bacteria into the sample port. The percent agreement across all target organisms was 100% (57/57), 100% (57/57), and 100% (56/56) from blood, MacConkey, and HC ESBL agar, respectively.

Variants detected during the analytical reactivity study are outlined in Tables 4 and 5.

Table 4. Analytical Reactivity Summary of Tested Variants

Number of strains tested on blood and MacConkey agarNumber of strains tested on HC ESBL agarCTX-M GroupNumber of individual variants tested on blood and MacConkey agarNumber of individual variants tested on HC ESBL agarCTX-M Variants Tested
5756¹
13130CTX-M-1, -3, -12, -15, -22, -30, -32, -55, -64, -79, -116
255CTX-M-2, -74, -75, -124
822CTX-M-40
91818CTX-M-9, -9-like, -14, -14b, -24, -27, -38
2566CTX-M-25, -100, -152, -160
Total6261

¹ One P. mirabilis strain with a characterized CTX-M-3 variant was susceptible to an antimicrobial agent in the HC ESBL formula, thus was not evaluated with NG-Test® CTX-M MULTI from HC ESBL.

Page 5

Page 10

Table 5. Analytical Reactivity Summary

Genus speciesNumber of strains tested on Blood and MacConkey agarNumber of strains tested on HardyCHROM™ ESBL agarCTX-M Variants Tested
Citrobacter species11CTX-M-15
Enterobacter cloacae99CTX-M-9, -9-like, -15, -22, -30, -100
Escherichia coli1919CTX-M-1, -2, -3, -14, -15, -24, -27, -40, -55, -79, -100, -116
Klebsiella oxytoca55CTX-M-14, -15, -22, -30, -75, -100
Klebsiella pneumoniae1414CTX-M-3, -12, -14, -14b, -15, -22, -25, -38, -40, -64, -74, -124
Klebsiella variicola11CTX-M-152
Kluyvera ascorbata11CTX-M-124
Proteus mirabilis6CTX-M-3, -15, -32, -79, -160
Serratia marcescens11CTX-M-15

¹ One P. mirabilis strain with a characterized CTX-M-3 variant was susceptible to an antimicrobial agent in the HC ESBL formula, thus was not evaluated with NG-Test® CTX-M MULTI from HC ESBL.

IX. Analytical Specificity

NG-Test® CTX-M MULTI was evaluated with 55 organisms that exhibit antimicrobial resistance mechanisms other than CTX-M and were not susceptible to third-generation cephalosporins (cefotaxime, ceftazidime, ceftriaxone, cefpodoxime) and other agents (aztreonam and cefoxitin) from blood and MacConkey agar. 50 organisms were evaluated from HC ESBL. Some of the antimicrobial resistance mechanisms included were TEM, SHV, VEB, ACT, CMY, DHA, FOX, MIR, ACC, VIM, KPC, IMP, NDM, and OXA. Each organism was incubated for 16 to 24 hours on blood agar and MacConkey agar, or for 18 to 24 hours on HC ESBL agar at 35°C. Each organism was tested from each type of media, and results were read 15 minutes after inoculating the extraction buffer mixed with bacteria into the sample port. No cross reactivity was observed, thus the percent agreement for all non-target organisms evaluated was 100% (55/55), 100% (55/55), and 100% (50/50) from blood, MacConkey, and HC ESBL agar, respectively.

X. Incubation Study

In order to confirm that the NG-Test® CTX-M MULTI consistently delivered accurate results over a range of incubation times, five strains were tested from blood agar and MacConkey agar every two hours between 16 to 24 hours of aerobic incubation at 35°C. Five strains were also tested from HC ESBL agar every two hours between 18 to 24 hours of aerobic incubation at 35°C. All organisms produced the expected result on NG-Test® CTX-M MULTI at every time point. Test results were read 15 minutes after inoculating the buffer mixed with bacteria into the sample port.

XI. Refrigeration Storage Study

In order to determine if a refrigerated organism culture (plate agar media with colonies) can be used with NG-Test® CTX-M MULTI, five strains were cultured on blood agar, MacConkey agar, and HC ESBL agar, and evaluated over time from refrigerated storage. Each strain was incubated on blood and MacConkey agar at 35°C, with initial tests performed after 16 to 24 hours of incubation (Day 0). Initial testing for cultures on HC ESBL agar was conducted after 18 to 24 hours of incubation. All organisms

Page 6

Page 11

produced the expected result on the NG-Test® CTX-M MULTI for each day of refrigeration for up to 3 days.

XII. Reproducibility

Prior to initiating the clinical study, a panel of 20 blinded isolates provided by Hardy Diagnostics was tested at three distinct study sites on five days to demonstrate reproducibility and document proficiency in the performance of NG-Test® CTX-M MULTI. A >95% agreement with known test results was required to proceed with the clinical study. The testing was done with one operator and two readers, blinded to each other's results, per site. NG-Test® CTX-M MULTI successfully detected all target CTX-M positive isolates (100%) on every day of the reproducibility study.

XIII. Conclusions

The analytical and clinical data presented in this submission demonstrate that NG-Test® CTX-M MULTI is substantially equivalent to the predicate device.

Page 7

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).