K Number
K243419
Date Cleared
2025-05-08

(185 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Altus Spine Navigation System reusable instruments are indicated to be used during the preparation and placement of Altus Spine Valencia Pedicle Screw System and Altus Spine Monaco Pedicle Screw System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Altus Spine Navigation System reusable instruments are specifically designed for use with the Medtronic StealthStation® System S8, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

• For a comprehensive summary of indications, the IFU for the specific system of use must be directly referenced: Altus Spine Valencia Pedicle Screw System / Altus Spine Monaco Pedicle Screw System.

The Altus Spine Navigation System is intended to pair with either the Altus Spine Valencia Pedicle Screw System or Altus Spine Monaco Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The Altus Spine Navigation system may be used for noncervical pedicle fixation via posterior percutaneous approach with MIS instrumentation.

Device Description

The Altus Spine Navigation System reusable instruments are surgical instruments for use with the Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures for preparation and placement of pedicle screw system implants. The Altus Spine Navigation System includes the following instruments dedicated to screw placement: Screwdrivers and Taps. The Altus Spine Navigation System is to be used with the following Altus Spine Systems:

▪ Altus Spine Valencia Pedicle Screw System
▪ Altus Spine Monaco Pedicle Screw System

All instruments are made of stainless steel per ASTM F899. All instruments are reusable instruments provided non sterile. The Altus Spine Navigation System instruments are not compatible with implants from other manufacturers. The Altus Spine Navigation System are designed for use only with Medtronic StealthStation® System S8 (V2.1.0) and the Medtronic NavLock® Tracker System.

AI/ML Overview

Based on the provided FDA 510(k) clearance letter for the Altus Spine Navigation System, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
Recognition and Location Accuracy with Medtronic StealthStation® System S8 (via ASTM F2554)"The Altus Spine Navigation System was tested side-by-side with the primary predicate device and passed recognition and location testing with the Medtronic StealthStation® System S8 navigation system via ASTM F2554."
Dimensional Accuracy (with primary predicate device)"The subject device and predicate device were compared for dimensional accuracy and the critical length. The results demonstrate that the subject device is substantially equivalent to the predicate device in use and performance."
Critical Length (with primary predicate device)"The subject device and predicate device were compared for dimensional accuracy and the critical length. The results demonstrate that the subject device is substantially equivalent to the predicate device in use and performance."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the test set (number of instruments tested or number of trials). It only mentions that the Altus Spine Navigation System was "tested side-by-side with the primary predicate device."

Regarding data provenance:

  • Country of Origin: Not specified, but implied to be related to Altus Partners, LLC (Pennsylvania, USA) and Medtronic (USA).
  • Retrospective or Prospective: The testing described ("tested side-by-side") suggests a prospective performance evaluation conducted specifically for this 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

The document does not provide information on the number or qualifications of experts used to establish ground truth for the test set. The nature of the testing (recognition, location accuracy, dimensional accuracy, and critical length using ASTM F2554) suggests an objective, measurement-based ground truth rather than a subjective expert assessment of images or clinical outcomes.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set. Given the objective, measurement-based nature of the performance tests (ASTM F2554, dimensional comparisons), an adjudication process like 2+1 or 3+1 (typically used for subjective assessments by multiple readers) would not be applicable or necessary.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs. without AI assistance.

No, an MRMC comparative effectiveness study was not done.

This device (Altus Spine Navigation System) is a set of reusable surgical instruments designed to interface with an existing navigation system (Medtronic StealthStation® System S8). It is not an AI-powered diagnostic software or a device that directly assists "human readers" in interpreting medical images. Therefore, a study comparing human reader performance with and without AI assistance is not relevant to this device's function or clearance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done.

Yes, the performance tests described are essentially "standalone" device performance tests.

The tests evaluated the instruments themselves in their ability to be recognized and tracked accurately by the Medtronic StealthStation® System S8, and their dimensional accuracy. This is a characteristic of the device itself, independent of a specific surgeon's performance or interpretation during a procedure. It's about the technical performance of the physical instruments.

7. The Type of Ground Truth Used

The ground truth used for these performance tests was objective, measurement-based data derived from:

  • Standardized Test Methods: Adherence to ASTM F2554 for recognition and location accuracy.
  • Physical Measurements: Dimensional accuracy and critical length comparisons.

This is not a ground truth derived from expert consensus, pathology, or outcomes data, as those are typically used for diagnostic or therapeutic devices where subjective interpretation or clinical effect is being assessed.

8. The Sample Size for the Training Set

The concept of a "training set" is not applicable to this device. The Altus Spine Navigation System is a set of physical surgical instruments, not an AI/machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

As the concept of a training set is not applicable, this question is not relevant to the Altus Spine Navigation System.

FDA 510(k) Clearance Letter - Altus Spine Navigation System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 8, 2025

Altus Partners LLC.
Rand Baker
QA/RA Manager
1340 Enterprise Drive
West Chester, Pennsylvania 19380

Re: K243419
Trade/Device Name: Altus Spine Navigation System
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic instrument
Regulatory Class: Class II
Product Code: OLO
Dated: April 4, 2025
Received: April 4, 2025

Dear Rand Baker:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243419 - Rand Baker Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243419 - Rand Baker Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tejen D. Soni -S

For
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K243419

Device Name
Altus Spine Navigation System

Indications for Use (Describe)

The Altus Spine Navigation System reusable instruments are indicated to be used during the preparation and placement of Altus Spine Valencia Pedicle Screw System and Altus Spine Monaco Pedicle Screw System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Altus Spine Navigation System reusable instruments are specifically designed for use with the Medtronic StealthStation® System S8, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

• For a comprehensive summary of indications, the IFU for the specific system of use must be directly referenced: Altus Spine Valencia Pedicle Screw System / Altus Spine Monaco Pedicle Screw System.

The Altus Spine Navigation System is intended to pair with either the Altus Spine Valencia Pedicle Screw System or Altus Spine Monaco Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The Altus Spine Navigation system may be used for noncervical pedicle fixation via posterior percutaneous approach with MIS instrumentation.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

SUBMITTER: Altus Partners, LLC
1340 Enterprise Drive
West Chester, PA 19380
Phone: 610-356-6148
Fax: 610-300-3049

CONTACT PERSON: Rand Baker
QA/RA Manager
rbaker@altus-spine.com

DATE PREPARED: May 8, 2025

TRADE OR PROPRIETARY NAME: Altus Spine Navigation System

PRIMARY PREDICATE DEVICES:
Medtronic Sofamor Danek, USA Inc., Navigated CD Horizon® Solera® Screwdrivers and Taps (K170679)

ADDITIONAL PREDICATE DEVICES:
Altus Spine Monaco Pedicle Screw System (K230766)
Altus Spine Valencia Pedicle Screw System (K181281)

CLASSIFICATION & REGULATION #: Class II per 21 CFR §888.4560

PRODUCT CODES: OLO

DEVICE DESCRIPTION:

The Altus Spine Navigation System reusable instruments are surgical instruments for use with the Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures for preparation and placement of pedicle screw system implants. The Altus Spine Navigation System includes the following instruments dedicated to screw placement: Screwdrivers and Taps. The Altus Spine Navigation System is to be used with the following Altus Spine Systems:

▪ Altus Spine Valencia Pedicle Screw System
▪ Altus Spine Monaco Pedicle Screw System

All instruments are made of stainless steel per ASTM F899. All instruments are reusable instruments provided non sterile. The Altus Spine Navigation System instruments are not compatible with implants from other manufacturers. The Altus Spine Navigation System are designed for use only with Medtronic StealthStation® System S8 (V2.1.0) and the Medtronic NavLock® Tracker System.

INDICATIONS FOR USE:

The Altus Spine Navigation System reusable instruments are indicated to be used during the preparation and placement of Altus Spine Valencia Pedicle Screw System and Altus Spine Monaco Pedicle Screw System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Altus Spine Navigation System reusable instruments are specifically designed for use with the Medtronic StealthStation® System S8, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

Page 1 of 2 K243419

Page 6

• For a comprehensive summary of indications, the IFU for the specific system of use must be directly referenced: Altus Spine Valencia Pedicle Screw System / Altus Spine Monaco Pedicle Screw System.

The Altus Spine Navigation System is intended to pair with either the Altus Spine Valencia Pedicle Screw System or Altus Spine Monaco Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The Altus Spine Navigation system may be used for noncervical pedicle fixation via posterior percutaneous approach with MIS instrumentation.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:

The Altus Spine Navigation System has the same or similar technological characteristics as the predicate instruments including design (instrument functional length and NavLock® connection feature), intended use, material composition, function, range of sizes, and sterility. The subject system shares the same principle of operation as the predicates. The subject instruments differ from the primary predicate in tip geometries and dimensions in order to accommodate the spinal screws; however, the differences are minor and do not raise any new concerns of safety or effectiveness with respect to the predicate.

PERFORMANCE DATA:

The Altus Spine Navigation System was tested side-by-side with the primary predicate device and passed recognition and location testing with the Medtronic StealthStation® System S8 navigation system via ASTM F2554. The subject device and predicate device were compared for dimensional accuracy and the critical length. The results demonstrate that the subject device is substantially equivalent to the predicate device in use and performance.

SUBSTANTIAL EQUIVALENCE CONCLUSION:

The overall technology characteristics and mechanical performance data lead to the conclusion that the Altus Spine Navigation System is substantially equivalent to the predicate device.

Page 2 of 2 K243419

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).