(30 days)
The Altus Spine Monaco Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The Altus Spine Monaco Pedicle Screw System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); turnor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used in a posterior percutaneous approach with MIS instrumentation, the Altus Spine Monaco Pedicle Screw System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); turnor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
The Altus Spine Monaco Pedicle Screw System consists of a system of implantable screws to be used with implantable rods for the purpose of aiding in spinal fusion. The Altus Spine Monaco Pedicle Screw System attaches to the vertebral body by means of screws to the non-cervical spinal and allows a surgeon to build a spinal implant construct with the intent to stabilize the spinal operative site during the fusion process of bone graft in the disc space. Implantable components are composed of titanium alloy meeting the requirements of ASTM F136 and cobalt chrome meeting the requirements of ASTM F1537. The device is supplied non-sterile and is intended for sterilization by hospital personnel.
The provided text is a 510(k) premarket notification for a medical device, the Altus Spine Monaco Pedicle Screw System. It describes the device's purpose, indications for use, and a summary of non-clinical tests conducted to demonstrate substantial equivalence to predicate devices.
However, the document does not contain information about:
- Acceptance criteria for an AI/ML-based device performance study.
- Any study involving a test set, expert readers, ground truth establishment, or human-in-the-loop performance with AI assistance.
- Sample sizes for training sets or provenance of data for AI models.
- Multi-reader multi-case (MRMC) comparative effectiveness studies.
- Standalone algorithm performance.
The provided text focuses on the mechanical and material properties of the pedicle screw system, demonstrating its substantial equivalence to previously cleared devices through bench testing (in accordance with ASTM F1717-21 for Dynamic Compression Bend, Static Compression Bend, and Static Torsion tests). This is typical for orthopedic implant submissions concerning physical integrity and function.
Therefore, I cannot extract the requested information regarding acceptance criteria and studies demonstrating a device meets those criteria for an AI/ML-based medical device from this document. The information requested (points 1-9) pertains to the evaluation of AI/ML software as a medical device (SaMD) or as part of a medical device, which is not what this document describes.
{0}------------------------------------------------
April 19. 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Altus Spine Rand Baker QA/RA Manager 1340 Enterprise Drive West Chester, Pennsylvania 19380
Re: K230766
Trade/Device Name: Altus Spine Monaco Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: March 20, 2023 Received: March 20, 2023
Dear Rand Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/5 description: The image shows the name "Eileen Cadel -S" in large, bold font on the left side. To the right of the name, there is text that reads "Digitally signed by Eileen Cadel -S Date: 2023.04.19 15:59:53 -04'00'". The word "for" is at the bottom right.
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K230766
Device Name Altus Spine Monaco Pedicle Screw System
Indications for Use (Describe)
The Altus Spine Monaco Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The Altus Spine Monaco Pedicle Screw System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); turnor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used in a posterior percutaneous approach with MIS instrumentation, the Altus Spine Monaco Pedicle Screw System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); turnor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
| SUBMITTER: | Altus Partners, LLC1340 Enterprise DriveWest Chester, PA 19380Phone: 610-356-6148Fax: 610-300-3049 |
|---|---|
| CONTACT PERSON: | Rand BakerQA/RA Managerrbaker@altus-spine.com |
| DATE PREPARED: | April 19, 2023 |
| COMMON NAME: | Pedicle Screw Spinal System |
| PROPRIETARY NAME: | Altus Spine Monaco Pedicle Screw System |
| PRIMARY PREDICATE DEVICES: | Altus Spine Pedicle Screw System (K211027) |
| ADDITIONAL PREDICATE DEVICES: | Altus Spine Pedicle Screw System (K181339)Vertebron PSS Pedicle Screw System (K051716) |
| CLASSIFICATION NAME: | 21 CFR §888.3070 Thoracolumbosacral Pedicle Screw System |
| PRODUCT CODES: | NKB |
| DEVICE CLASS: | Class II |
| MATERIAL: | The materials used are Titanium Alloy material that conforms to ASTMF136 and Cobalt Chrome that conforms to ASTM F1537. |
DEVICE DESCRIPTION:
The Altus Spine Monaco Pedicle Screw System consists of a system of implantable screws to be used with implantable rods for the purpose of aiding in spinal fusion. The Altus Spine Monaco Pedicle Screw System attaches to the vertebral body by means of screws to the non-cervical spinal and allows a surgeon to build a spinal implant construct with the intent to stabilize the spinal operative site during the fusion process of bone graft in the disc space. Implantable components are composed of titanium alloy meeting the requirements of ASTM F136 and cobalt chrome meeting the requirements of ASTM F1537. The device is supplied non-sterile and is intended for sterilization by hospital personnel.
INDICATIONS FOR USE:
The Altus Spine Monaco Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The Altus Spine Monaco Pedicle Screw System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
{4}------------------------------------------------
When used in a posterior percutaneous approach with MIS instrumentation, the Altus Spine Monaco Pedicle Screw System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with degeneration of the disc confirmed by history and radiographic studies): spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:
The Altus Spine Monaco Pedicle Screw System is substantially equivalent to the primary predicate in regards to implant materials and surgical technique.
SUMMARY OF NON-CLINICAL TESTS SUBMITTED:
Dynamic Compression Bend, Static Compression Bend, and Static Torsion tests were performed in accordance with ASTM F1717-21. Bench testing demonstrated that the Altus Spine Monaco Pedicle Screw System is substantially equivalent to the predicate device.
SUBSTANTIAL EQUIVALENCE CONCLUSION:
The Altus Spine Monaco Pedicle Screw System utilizes the same implant materials and surgical technique as the primary predicate device (K211027).
Altus Partners has determined that the modifications to the Altus Spine Monaco Pedicle Screw System do not alter the system function, strength and stability or materials. Therefore, the Altus Spine Monaco Pedicle Screw System is substantially equivalent to the predicate devices.
N/A