K Number
K243209

Validate with FDA (Live)

Device Name
nordicMEDiVA
Manufacturer
Date Cleared
2024-10-22

(21 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The nordicMEDIVA software is an advanced visualization and processing platform with a specific focus on providing algorithms designed to analyze functional and dynamic MRI data of the brain. The software runs on a server in a networked environment and is accessed by users via a standard web browser. It can communicate with other imaging platforms that support DICOM, and process medical image data acquired through DICOM-compliant imaging devices and modalities.

nordicMEDIVA is indicated for image analysis and visualization of functional and dynamic MRI data of the brain, presenting derived properties and parameters from the input image data in a clinically useful context.

Device Description

nordicMEDIVA is a software as a medical device (SaMD) for processing of MR images of the brain. Users will configure analysis pipelines, which are executed automatically when image data is received or manually by a user. The user can choose to send the results to other DICOM nodes for review or use nordicView for their review and export the results to PACS, neuro navigation systems, or other DICOM-compliant modalities. nordicMEDIVA is a server-client solution and can be installed on a local server at the customer's location or in a cloud-based setup. The software is containerized with Docker technology and operates on a GPU-enabled Linux host. This allows customers to manage the server environment themselves or use it as a Software as a Service (SaaS) hosted by NordiclmagingLab AS in the cloud. Customers can install the server on physical hardware or in their own cloud infrastructure. The device comprises a database, DICOM functionality, various APIs, a visualization engine, and medical image analysis modules. The device is not intended for long-term persistent storage of medical diagnostic data. The device incorporates rule-based algorithms for the calculation of metrics from dynamic MRI data. The device does not incorporate Al algorithms based on neural networks. The device connects to other imaging modalities, such as MR scanners, PACS, and surgical navigation systems.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the nordicMEDiVA device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Summative Usability Test: All scenarios met acceptance criteria.All scenarios from the summative usability test met the acceptance criteria completely, no new risks were found and existing risk control measures were proven to be effective.
Diffusion and Tractography: Results were as effective as the predicate device (nordicBrainEx K163324).The results from Diffusion and Tractography were evaluated in comparison with equivalent results from the predicate device, nordicBrainEx (K163324). The results were reviewed by internal and external clinical experts and proven to be as effective as the predicate device.
BOLD fMRI: Results were as effective as the predicate device (nordicBrainEx K163324).The results from BOLD fMRI were evaluated in comparison with equivalent results from the predicate device, nordicBrainEx (K163324). The results were reviewed by internal and external clinical experts and proven to be as effective as the predicate device.
DSC (Dynamic Susceptibility Contrast): Lin's Concordance Correlation Coefficient (CCC) for enhancing voxels ≥ 0.8 compared to the reference device (nordicMEDiVA K241608).The results from the DSC was confirmed to be the same as the reference device nordicMEDiVA (K241608) where a Lin's Concordance Correlation Coefficient of enhancing voxels was calculated with acceptance criteria being greater than or equal to 0.8 was applied. (The text states the results were "confirmed to be the same," implying the CCC met or exceeded the 0.8 acceptance criterion, though the specific achieved value is not provided.)

2. Sample Size and Data Provenance

  • Sample size for test set: Not explicitly stated in the provided text.
  • Data Provenance: Not explicitly stated. The text mentions "internal and external clinical experts" reviewing results, but doesn't specify the country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth

  • Number of experts: Not explicitly stated. The text refers to "internal and external clinical experts."
  • Qualifications of experts: Not explicitly stated, beyond being "clinical experts." Specific experience levels (e.g., "radiologist with 10 years of experience") are not provided.

4. Adjudication Method for the Test Set

The text does not specify an explicit adjudication method (e.g., 2+1, 3+1). It states "reviewed by internal and external clinical experts." This suggests a consensus-based review, but the exact process of reaching that consensus or resolving disagreements is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No, an MRMC comparative effectiveness study was not described. The studies focused on comparing the device's output to that of predicate devices and showing equivalence, rather than directly measuring human reader performance with and without AI assistance.
  • Effect size of human reader improvement: Not applicable, as an MRMC comparative effectiveness study was not performed.

6. Standalone Performance Study (Algorithm Only)

Yes, a form of standalone performance was implicitly done. The tests for Diffusion and Tractography, BOLD fMRI, and DSC were evaluations of the nordicMEDiVA software's output (algorithm only) compared to established predicate devices. This demonstrates algorithm-only performance in generating results equivalent to cleared devices.

7. Type of Ground Truth Used

The ground truth for the test set appears to be based on:

  • Predicate Device Equivalence: For Diffusion and Tractography, and BOLD fMRI, the ground truth was essentially the "equivalent results from the predicate device, nordicBrainEx (K163324)," as reviewed and confirmed by clinical experts.
  • Reference Device Equivalence/Metrics: For DSC, the ground truth was based on demonstrating "the same" results as the reference device nordicMEDiVA (K241608), quantified by a Lin's Concordance Correlation Coefficient of enhancing voxels.
  • Expert Confirmation: The final acceptance relied on confirmation by "internal and external clinical experts."

8. Sample Size for Training Set

The text does not provide any information regarding the sample size used for a training set. This is because the device does not incorporate AI algorithms based on neural networks. It explicitly states: "The device incorporates rule-based algorithms for the calculation of metrics from dynamic MRI data. The device does not incorporate AI algorithms based on neural networks." Therefore, there would be no "training set" in the context of deep learning models.

9. How Ground Truth for Training Set was Established

Given that the device uses "rule-based algorithms" and "does not incorporate AI algorithms based on neural networks," there would be no training set requiring ground truth establishment in the traditional machine learning sense. The performance is based on the correctness and validation of its rule-based calculations against established clinical principles and comparison to predicate device outputs.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

October 22, 2024

NordicNeurolab AS % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K243209

Trade/Device Name: nordicMEDiVA Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: October 1, 2024 Received: October 1, 2024

Dear Prithul Bom:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

Device Name

nordicMEDiVA

Indications for Use (Describe)

Intended use

The nordicMEDIVA software is an advanced visualization and processing platform with a specific focus on providing algorithms designed to analyze functional and dynamic MRI data of the brain. The software runs on a server in a networked environment and is accessed by users via a standard web browser. It can communicate with other imaging platforms that support DICOM, and process medical image data acquired through DICOM-compliant imaging devices and modalities.

Intended users

nordicMEDiVA must be operated according to its intended purpose and only by qualified medical personnel with the necessary knowledge in accordance with country-specific regulations, such as trained radiological technicians or physicians. Users must have received training in applicable areas, such as in MR technology and neuroanatomy.

Indications for use

nordicMEDIVA is indicated for image analysis and visualization of functional and dynamic MRI data
of the brain, presenting derived properties and parameters from the input image data in a clinically
useful context.

Contraindications

The software is not be used as an archiving device for patients' image data.

The software is not to be used as a sole basis for clinical decisions, but further evidence has to be taken into account.

The software is not to be used as a sole basis in emergency care applications.

The software is not to be used for applications other than the analysis and visualization of MR data, or data derived thereof, of the human brain.

Patient target population

nordicMEDiVA has no limitation concerning the patient population.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

{4}------------------------------------------------

510(k) #:510(k) SummaryPrepared on: 2024-09-10
Contact Details21 CFR 807.92(a)(1)
Applicant NameNordicNeuroLab AS
Applicant AddressMollendalsveien 1 Bergen 5009 Norway
Applicant Contact Telephone+47 55 70 70 95
Applicant ContactMrs. Chandana Gurung Bhandari
Applicant Contact Emailchandana@nordicneurolab.com
Device Name21 CFR 807.92(a)(2)
Device Trade NamenordicMEDIVA
Common NameMedical image management and processing system
Classification NameSystem, Image Processing, Radiological
Regulation Number892.2050
Product Code(s)LLZ
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K163324nordicBrainExLLZ
K241608nordicMEDIVALLZ
K203518QuicktomeLLZ
Device Description Summary21 CFR 807.92(a)(4)
nordicMEDIVA is a software as a medical device (SaMD) for processing of MR images of the brain. Users will configure analysis pipelines,which are executed automatically when image data is received or manually by a user. The user can choose to send the results to otherDICOM nodes for review or use nordicView for their review and export the results to PACS, neuro navigation systems, or other DICOM-compliant modalities.nordicMEDIVA is a server-client solution and can be installed on a local server at the customer's location or in a cloud-based setup. Thesoftware is containerized with Docker technology and operates on a GPU-enabled Linux host. This allows customers to manage theserver environment themselves or use it as a Software as a Service (SaaS) hosted by NordiclmagingLab AS in the cloud. Customers caninstall the server on physical hardware or in their own cloud infrastructure.The device comprises a database, DICOM functionality, various APIs, a visualization engine, and medical image analysis modules. Thedevice is not intended for long-term persistent storage of medical diagnostic data.The device incorporates rule-based algorithms for the calculation of metrics from dynamic MRI data. The device does not incorporate Alalgorithms based on neural networks.
The device connects to other imaging modalities, such as MR scanners, PACS, and surgical navigation systems.

{5}------------------------------------------------

The following modules provide the main functions of the device.

Viewer: A browser-based user interface, accessed from destop clients, that provides tools for general image visualization, export, and relevant analysis tools for BOLD-fMRI, diffusion MRI (including tractography), and DSC-perfusion.

Task-based fMRI: BOLD task-based fMRI anall magnetic susceptibility changes in the human brain in areas with altered blood flow resulting from neuronal activity. The image processing requires the definition of a so-called design matrix which is used to calculate voxel-wise statistics conveying information about the probability of the voxel being involved in the execution of the given task.

The design matrix is is defined such that the timilation that was presented to that was presented to the patient during the scan. The task or stimulation presented during scan time is often refered to as "the paradigm". The design matrix can be defined manually by the user, or a paradigm from nordicAktiva - another product from NordicNeuroLab - can be used.

nordic ktiva is a software, marketed by Nordic Neurolab, that may be used during scan time to present the patient or subject being scanned. The use of nordicAktiva is not required.

Diffusion and tractography: The analysis of diffusion-weighted MRI data can be used to model water diffusion properties in the brain's white matter. In addition to the calculation of diffusion tensor imaging (DTI) derived maps such as fractional anisotropy (FA) and mean diffusivity (MD), tractography can be performed to create 3D reconstructions of the matter pathways. Both DTI and spherica deconvolution (SD) models are available for fiber tracking. Unlike DTI, the SD model can resolve multiple fiber populations at each voxel so can track through crossing fiber regions to give more accurate representations of white matter anatomy.

DSC: Calculations of perfusion-related parameters that provide information about the blood vessel structure and characteristics. Such maps include blood volume, blood flow, time to peak, mean transit time, and leakage.

Platform: The platform includes a database, DICOM functionality, various APIs, processing pipelines for the medical image analysis modules. Serves as a backbone component for the other modules of nordicMEDiVA.

Dashboard: a browser-based user interface accessed from desktop clients for and configuration.Reference to previous and similar generations of the product

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

Intended use

The nordicMEDiVA software is an advanced visualization with a specific focus on providing algorithms designed to analyze functional and dynamic MRI data of the brain. The software runs on a server in a networked environment and is accessed by users via a standard web browser. It can communicate with other imaging platforms that support DICOM, and process medical image data acquired through DICOM-compliant imaging devices and modalities.

Intended users

nordicMEDiVA must be operated according to its intended purpose and only by qualified medical personnel with the necessary knowledge in accordance with country-specific regulations, such as trained radiological technicians or physicians Users must have received training in applicable areas, such as in MR technology and neuroanatomy.

Indications for use

nordicMEDVA is indicated for image analysis and visualization of functional and dynamic MRI data of the brain, presenting derived properties and parameters from the input image data in a clinically useful context.

Contraindications

The software is not be used as an archiving device for patients' image data.

The software is not to be used as a sole basis for clinical decisions, but further evidence has to be taken into account.

The software is not to be used as a sole basis in emergency care applications.

The software is not to be used for applications other than the analysis and visualization of MR data, or data derived thereof, of the human brain.

Patient target population nordicMEDiVA has no limitation concerning the patient population.

{6}------------------------------------------------

Indications for Use Comparison

The primary medical purpose conveyed in the intended use of both the predicates and the subject device is the same.

Although the wording of the intended use statements is slightly different from the predicate devices, the primary intention is the same. Both the subject, the predicate devices combined enable dynamic susceptibility contrast-enhanced (DSC)-, BOLDand diffusion-weighted magnetic resonance imaging (MR) data analysis, presenting the derived properties and parameters in a clinicaly valuable context. We identified no difference in the intended use statements that impacts the safety and effectiveness of the subject device.

Technological Comparison

nordicMEDVA has the same technological characteristics as the predicate device nordicBrainEx (K163324). Both the subject device and the predicate device:

  • Allow for the import and export of MR DICOM images from other DICOM-compliant imaging devices such as MR scanner, PACS and surgical navigation systems.

  • Has a graphical user interface for the clinical evaluation of analysis results.

  • Allow for co-registration of dynamic and functional MRI data to structural MRI data.

  • Allow for region of interest analysis (ROI), distance measurements, and the export of these measurements to PACS.

  • Allow for the generation of secondary data for use in surgical navigation systems.

  • Contain analysis modules for DSC-, Diffusion- and BOLD fMRI image processing.

The subject device differs from the predicate device in its technological characteristics in that:

  • The subject device is a thin client application deployed in a server environment, and the graphical user interface is a web browser. The predicate device is a desktop (thick client) application.

  • The subject device contains user-defined automated image data analysis, while in the predicate device, a user has to initiate image analysis manually.

The DSC - and BOLD fMRI analysis of the subject device has technological characteristics identical to those of the reference device, nordicMEDiVA (K241608).

These differences do not raise additional questions about the safety or effectiveness of the subject device.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Non-clinical performance testing has been performed on the nordicMEDiVA software and demonstrates compliance with the following International and FDA-recognized consensus standards and FDA guidance document:

  • ISO 14971 Medical devices Application of risk management to medical devices
  • IEC 62304 Medical device software Software life cycle processes
  • IEC 62366-1 Medical devices Part 1: Application of usability engineering to medical devices
  • IEC 82304-1 Health software Part 1: General requirements for product safety

nordicMEDVA was tested in accordance with verification processes. Verification and Validation tests have been performed to address intended use, the technological characteristics claims, requirement specifications and the risk management results.

Formative and summative usability studies have been performed with users from the intended audience.

The test results are summarized below:

  • All scenarios from the summative usability test met the acceptance criteria completely, no new risks were found and existing risk control measures were proven to be effective.

  • The results from Diffusion and Tractography were evaluated in comparison with equivalent results from the predicate device, nordicBrainEx (K163324). The results were reviewed by internal and external clinical experts and proven to be as effective as the predicate device

  • The results from BOLD fMRI were evaluated in comparison with equivalent results from the predicate device, nordicBrainEx (K163324). The results were reviewed by internal and external clinical experts and proven to be as effective as the predicate device

  • The results from the DSC was confirmed to be the same as the reference device nordicMEDVA (K241608) where a Lin's Concordance Correlation Coefficient of enhancing voxels was calculated with acceptance criteria being greater than or equal to 0.8 was applied.

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

{7}------------------------------------------------

Cybersecurity considerations related to nordicMEDiVA are included within this submission. nordicMEDiVA conforms to cybersecurity requirements by implementing a means to prevent unauthorized access, modification, misuse, denial of use of information stored, accessed or transferred from a medical device to an external recipient.

We believe that nordicMEDiVA is as safe and effective as the identified predicate devices and does not introduce new safety and effectiveness concerns.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).