K Number
K243193
Date Cleared
2024-11-26

(57 days)

Product Code
Regulation Number
870.5175
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CrossWise™ RF Transseptal Cannula and accessories are used to create an atrial septal defect in the heart. Secondary indications include infusing solutions including heparinized saline and mixtures of 50% contrast media and 50% saline.

Device Description

The CrossWise Multi-Use RF Adapter Cable (Model CW-1001) is used in Transseptal Puncture Procedures (TSP) to connect the CrossWise™ RF Transseptal Cannula (K241414) to a 3/32" electrosurgical (ES) pencil. The cable delivers radiofrequency (RF) energy to the CrossWise™ RF Transseptal Cannula to facilitate septal puncture. The CrossWise RF Transseptal Cannula is connected to a ValleyLabs Force2 ES Generator (K051644) via the CrossWise Multi-Use RF Adapter Cable (Model CW-1001) or CrossWise RF Adapter Cable (Model CW-1002 / K241414) and a commercially available 3/32" ES pencil. The CrossWise Multi-Use RF Adapter Cable has unique connectors on each end which cannot be connected to an improper device and/or accessory. The CrossWise Multi-Use RF Adapter Cable (Model CW-1001) and CrossWise RF Adapter Cable (Model CW-1002) are both compatible with all CrossWise RF Transseptal Access System models. The dimensions for the CrossWise Multi-Use RF Adapter Cable can be found on the package labels. Circa Scientific sterilized the CrossWise Multi-Use RF Adapter Cable via ethylene oxide (EO) for initial use and is intended for multiple patients use, following prescribed reprocessing instructions. The prescribed reprocessing includes cleaning and steam/autoclaye sterilization procedures that have been validated for up to10 uses when users follow the cleaning and sterilization methods described in the Cleaning and Sterilization Instructions sections of the CrossWise Multi-Use RF Adapter Cable IFU. The CrossWise Multi-Use RF Adapter Cable does not come into direct or indirect contact with the patient according to the definitions in Section 3.0 of ISO 10993-1, and Attachment G of the September 2023 FDA Guidance document.

AI/ML Overview

The provided document is a 510(k) Pre-Market Notification for a medical device called the "CrossWise™ Multi-Use RF Adapter Cable." This document primarily focuses on demonstrating the substantial equivalence of the new device to a previously cleared predicate device, especially regarding its reusability and materials.

The request asks for information typically found in studies validating an AI/ML medical device, specifically regarding acceptance criteria and performance data for an algorithm. However, this 510(k) submission does not describe an AI/ML device or algorithm. Instead, it describes a physical medical device (an adapter cable) and the bench testing performed to ensure its safety and effectiveness, particularly its reprocessing capabilities.

Therefore, many of the requested points, such as "effect size of how much human readers improve with AI vs without AI assistance," "standalone (i.e. algorithm only without human-in-the-loop performance)," "number of experts used to establish the ground truth," and "sample size for the training set," are not applicable to this type of device and submission.

However, I can extract information related to the acceptance criteria and the study that proves the device meets those criteria, reinterpreting the request for a physical device rather than an AI/ML algorithm.


Device Description:
The CrossWise™ Multi-Use RF Adapter Cable (Model CW-1001) is used in Transseptal Puncture Procedures (TSP) to connect the CrossWise™ RF Transseptal Cannula to a 3/32" electrosurgical (ES) pencil. The cable delivers radiofrequency (RF) energy to the cannula to facilitate septal puncture. The primary difference from the predicate device is its ability to be reprocessed (cleaned and re-sterilized) for up to 10 uses, and minor material modification to the connector shroud to withstand autoclave temperatures.


1. Table of Acceptance Criteria and Reported Device Performance

For this multi-use physical device, the "acceptance criteria" revolve around its ability to maintain its functionality and safety after repeated reprocessing cycles, showing substantial equivalence to the predicate device.

Acceptance Criteria CategorySpecific Test/Performance MetricAcceptance Criteria (Implicit from testing)Reported Device Performance (Implicit from "passed testing")
SterilizationInitial EO Sterilization SALSAL of $10^{-6}$Validated to meet SAL of $10^{-6}$
Subsequent Steam SterilizationValidated for up to 10 cyclesValidated for up to 10 additional steam sterilization cycles.
Physical IntegrityLabel Visual InspectionLabels must be present and legible.Devices passed visual inspection.
Device Visual InspectionNo damage, defects, or degradation.Devices passed visual inspection.
Dimensional VerificationDimensions within specified tolerances.Devices passed dimensional verification.
Functional PerformanceImpedanceMeets required electrical impedance.Devices passed impedance testing.
Bend RadiusMaintains structural integrity after bending.Devices passed bend radius testing.
ResistanceMeets required electrical resistance.Devices passed resistance testing.
RF Energy DeliveryCapable of delivering RF energy as intended.Devices passed RF energy delivery testing.
ContinuityMaintains electrical continuity.Devices passed continuity testing.
TensileWithstands specified tensile forces.Devices passed tensile testing.
DurabilityReprocessing DurabilityMaintains function after 10 reprocessing cycles and accelerated aging.Devices passed durability testing. (Implies all tests above passed after conditioning)
BiocompatibilityPatient ContactNo direct or indirect patient contact (inherent to device type).The device does not come into direct or indirect contact with the patient according to relevant standards.

2. Sample Size Used for the Test Set and the Data Provenance

  • Test Set Sample Size: The document does not explicitly state the numerical sample size for individual tests (e.g., "n=X cables tested"). It mentions "Devices" were tested and "testing was conducted on devices at 2 aging points: after manufacturing (T=0) and after 6 months of accelerated aging (T=6M AA)." It also states that devices were "subsequently conditioned to reflect the intended reusability of the devices after reprocessing."
  • Data Provenance: The data is from non-clinical bench testing performed by the manufacturer, Circa Scientific, Inc., to support the 510(k) submission. There is no indication of country of origin of the data beyond the manufacturer's location (Englewood, CO, USA). The testing is prospective as it's specifically designed to validate the new device's capabilities prior to market clearance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This concept is not applicable to this type of device validation. "Ground truth" established by experts (e.g., radiologists interpreting images) is relevant for AI/ML diagnostic tools. For this physical device, the "ground truth" is defined by established engineering and medical device standards for performance, sterilization, and material properties. The engineers and quality assurance personnel conducting and overseeing these tests are inherently acting as the "experts" whose work ensures compliance with these standards.

4. Adjudication Method for the Test Set

  • This concept is not applicable. Adjudication methods (e.g., 2+1, 3+1 consensus) are used for resolving disagreements in expert interpretations, typically in diagnostic or prognostic AI/ML studies. For bench testing of a physical device, results are typically objectively measured against pre-defined engineering specifications, not subject to subjective interpretation and adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This is not applicable. An MRMC study is designed to evaluate the impact of an AI diagnostic tool on human reader performance. The device here is a physical adapter cable, not an AI/ML algorithm.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This is not applicable. A standalone performance evaluation is performed for AI/ML algorithms. The device submitted is a physical product. Its "standalone" performance is assessed via the bench tests (e.g., electrical resistance, tensile strength) without human interaction being part of the performance metric itself, but this is a fundamentally different type of evaluation from an AI algorithm.

7. The Type of Ground Truth Used

  • The "ground truth" for this physical device is based on engineering specifications, material science properties, and established medical device performance standards (e.g., ISO 11135, ISO 17665, AAMI TIR12, ANSI/AAMI ST79, ANSI/AAMI ST98, IEC 60601-2-2). The physical and functional characteristics of the cable are objectively measured and compared against these pre-defined benchmarks to ensure safety and effectiveness for its intended use. There is no "pathology" or "outcomes data" in the sense of clinical study that establishes efficacy for this specific component; rather, component performance ensures the overall system works.

8. The Sample Size for the Training Set

  • This is not applicable. There is no "training set" as this is not an AI/ML device that learns from data.

9. How the Ground Truth for the Training Set Was Established

  • This is not applicable as there is no "training set."

§ 870.5175 Septostomy catheter.

(a)
Identification. A septostomy catheter is a special balloon catheter that is used to create or enlarge the atrial septal defect found in the heart of certain infants.(b)
Classification. Class II (performance standards).