K Number
K243179
Date Cleared
2025-06-26

(269 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for steam sterilization. The recommended sterilization cycle is as follows:
• Pre-vacuum Steam: 4 minutes at 132°C (270°F); 30 minutes dry time.

The sterilization pouch and roll are made with medical grade paper and medical compound film. The sterilization pouch and roll maintain the sterility of the enclosed devices for up to 6 months post steam sterilization. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone a steam sterilization process. The steam sterilization indicator color will change from blue before sterilization to dark grey after sterilization.

Device Description

The subject Sterilization Pouch and Roll device has five types:
(1) Self-sealing sterilization pouches:
These pouches are made from a medical grade plastic film that is heat sealed on three sides. The fourth side has an adhesive strip that is used to seal the pouch. Release paper used in the pouch is a laminated sheet with composing structure of PE/paper/PE. It is a strip to cover the adhesive area and is released before seal the pouch. The medical grade paper conforms to recognized material standards and can be sterilized by steam. The Process Indicators Ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam.

(2) Sterilization pouches, Flat:
These pouches have the same components with the Self-sealing sterilization pouches, except the fourth side is left opened instead of an adhesive strip and will be heat-sealed when using.

(3) Sterilization pouches, Gusseted:
These pouches are the same with the Sterilization pouches, flat, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.

(4) Sterilization rolls, Flat:
These rolls are made from a medical grade paper and plastic film that are heat sealed on opposite two sides. It will be cut into the suitable length and the opened sides will be heat-sealed. The indicators printed on the medical grade paper are the same with the self-sealing sterilization pouches.

(5) Sterilization rolls, Gusseted:
These rolls are the same with the flat sterilization roll, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.

The Sterilization pouch and roll is composed of medical grade paper (60g/m2) and medical compound film (52μm), it is intended to be used to contain medical devices to be terminally sterilized by the Steam sterilization process. The recommended sterilization cycle parameters are:
Pre-vacuum Steam: 4 minutes at 132°C (270°F); 30 minutes dry time.

The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized for the Steam Sterilization Process. The heat-sealed pouch/roll are heat sealed prior to sterilization processing. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened.

The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 6 months post sterilization. The shelf life of the device is 3 months.

The Pouch/Roll is printed with a chemical indicator bar that changes from Blue to Dark grey when exposed to steam during the sterilization process. The steam chemical Indicator offers an addition way to verity processing in the sterilization cycle. The endpoint stability of the chemical indicator color change to dark grey is 3 months. The chemical Indicator should be used in addition to, not in place of, the biological indicator. The steam chemical indicator does not signify sterilization; it only indicates that the indicator has been exposed to steam.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary describe a Sterilization Pouch and Roll device, which is a Class II medical device. This is NOT an AI/ML device. Therefore, the questions related to AI/ML device performance metrics, such as effect size of human readers improving with AI, standalone algorithm performance, number of experts for ground truth, and training set information, are not applicable.

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Test ItemAcceptance Criteria (Endpoint)Reported Device Performance (Results)
Sterilant Penetration/Drying Time- Meets the requirement of SAL 10-6- The weight difference before sterilization and after drying shall not exceed 3%Pass
Biocompatibility - CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells. Pass
Biocompatibility - IrritationNon-irritatingUnder the conditions of the study, not an irritant. Pass
Biocompatibility - SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. Pass
Package AppearanceThe appearance of the sterilization pouch is clean, intact, and sealed on all four sides.Pass
Thickness52 ± 12% μmPass
Tensile Strength of PaperMD ≥ 4.4 KN/mCD ≥ 2.2 KN/mPass
Bursting TestBurst value > 3 Kpa or No BurstPass
Dye Penetration TestThe dye solution is no any leakage across the seal width of sterile barrier system. (No Infiltration)Pass
Seal Strength Test≥ 2.5 N/15mmPass
Vacuum Leakage TestNo water can penetrate the sterilization pouch.Pass
Microbial Barrier0 CFU No growthPass
Chemical Indicator FunctionalityColor of indicator changes from blue to dark grey after Steam sterilizationPass
Chemical Indicator Shelf Life & Endpoint StabilityShelf life: 3 monthsEndpoint stability: 3 monthsPass
Shelf Life (Product)Shelf life: 3 monthsPass
Maintenance of SterilityMaintains sterility of enclosed devices for up to 6 months post steam sterilization. (This is an indication from the "Indications for Use" and is generally supported by the package integrity and microbial barrier tests over time, often through accelerated aging studies).The document states "The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 6 months post sterilization." The "Maintenance of Sterility" is listed as a "Different" characteristic compared to the predicate (6 months vs. 3 years for EO and 6 months for Steam). The microbial barrier test is passed, supporting the general concept of sterility maintenance, but the specific validation for the 6-month claim is not detailed in the summary table.

2. Sample Size Used for the Test Set and Data Provenance

The provided document details various performance tests such as sterilant penetration, biocompatibility, package integrity, and chemical indicator testing. However, it does not explicitly state the sample sizes used for each of these test sets. The tests are "bench testing" (non-clinical) and do not involve human subjects.

The data provenance is from non-clinical bench testing. The tests were performed to evaluate the performance and functionality of the device against a requirement specification. There is no mention of country of origin for the data or whether it was retrospective or prospective, as these terms are typically relevant for clinical studies or data collection in human populations.

3. Number of Experts and Qualifications for Ground Truth

Not applicable. This device is a physical sterilization pouch and roll, and its performance is evaluated through standardized laboratory and bench testing against established physical, chemical, and biological criteria, not by human expert interpretation of data or images. Ground truth is established by objective measurements and standardized test methods.

4. Adjudication Method for the Test Set

Not applicable. As described above, the evaluation relies on objective measurements and established standards rather than expert interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI/ML devices where human performance with and without AI assistance is being compared. This document is for a physical sterilization product.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

Not applicable. This device is a physical product, not an algorithm or software. Therefore, there is no "standalone" algorithm performance to evaluate.

7. The Type of Ground Truth Used

The ground truth for this device's testing is based on:

  • Standardized Test Methodologies and Acceptance Criteria: As outlined in the "Test Methodology" and "Acceptance Criteria or End Point" columns in Table 3 of the 510(k) summary. These include standards like ISO 17665-1, ISO 10993 series, ASTM F1886, ASTM F2251, ISO 1924-2, ASTM F1140, ASTM F1929, ASTM F88, ASTM D3078, DIN 58953-6, ISO 11140-1, and ASTM F1980.
  • Physical and Chemical Measurement: For properties like thickness, tensile strength, burst strength, seal strength, and the color change of the chemical indicator.
  • Microbiological Evaluation: For the microbial barrier test (e.g., 0 CFU No growth).
  • Biocompatibility Testing: Evaluating potential toxicity, irritation, and sensitization in laboratory settings (e.g., L-929 cells).
  • Sterilization Effectiveness: Meeting a specified Sterility Assurance Level (SAL 10-6).

8. The Sample Size for the Training Set

Not applicable. There is no AI/ML component described for this device, so there is no training set in the context of machine learning.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set.

FDA 510(k) Clearance Letter - Sterilization Pouch and Roll

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 26, 2025

Yichang Xinxin Paper Products Co., Ltd.
℅ Boyle Wang
Manager
Shanghai Truthful Information Technology Co., Ltd.
Room 1801, No. 161 East Lujiazui Rd., Pudong
Shanghai, 200120
China

Re: K243179
Trade/Device Name: Sterilization Pouch and Roll
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization wrap
Regulatory Class: Class II
Product Code: FRG, JOJ
Dated: May 20, 2025
Received: May 20, 2025

Dear Boyle Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243179 - Boyle Wang Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243179 - Boyle Wang Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Christopher K. Dugard -S

Christopher K. Dugard, M.S.
Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243179
Device Name: Sterilization Pouch and Roll

Indications for Use (Describe)

The Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for steam sterilization. The recommended sterilization cycle is as follows:
• Pre-vacuum Steam: 4 minutes at 132°C (270°F); 30 minutes dry time.

The sterilization pouch and roll are made with medical grade paper and medical compound film. The sterilization pouch and roll maintain the sterility of the enclosed devices for up to 6 months post steam sterilization. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone a steam sterilization process. The steam sterilization indicator color will change from blue before sterilization to dark grey after sterilization.

Table of Maximum Load

ItemSize (W)Size (L)Content / Max Load (lbs)
PZF-001, PZF-002, PZF-003, PZF-004, PZF-005, PZF-006, PZF-007, PZF-008, PRF-002, PRF-003Minimum: 55mmMinimum: 70mmMetal: 0.02
Maximum: 90mmMaximum: 280mmPlastic: 0.01
PZF-009, PZF-010, PZF-011, PZF-012, PZF-013, PZF-014, PZF-015, PZF-016, PZF-017, PZF-018, PZF-019, PRF-001, PRF-004, PRF-005, PRF-006, PRF-007, PRF-008, PRF-009Minimum: 57mmMinimum: 183mmMetal: 0.27
Maximum: 300mmMaximum: 610mmPlastic: 0.23
PZF-020, PZF-021, PZF-022, PZF-023, PZF-024, PRF-010Minimum: 300mmMinimum: 483mmMetal: 2.92
Maximum: 460mmMaximum: 635mmPlastic: 1.13
PRF-011400mm580mmMetal: 3.38
PRF-012420mm600mmPlastic: 1.13
PRFJ-001, PRFJ-002, PRFJ-003, PRFJ-004, PRFJ-005, PRFJ-006Minimum: 55mm30MMetal: 0.02
Maximum: 180mmPlastic: 0.01
PRFJ-007, PRFJ-008, PRFJ-009, PRFJ-010Minimum: 200mm30MMetal: 1.13
Maximum: 350mmPlastic: 0.86
PRFJ-011400mm30MMetal: 6.24
Plastic: 2.99

Page 5

ItemSize (W)Size (H)Size (L)Content / Max Load (lbs)
LRF-001100mm40mm300mmMetal: 0.21
LRF-002150mm50mm400mmPlastic: 0.05
LRF-003150mm50mm400mmMetal: 1.18
LRF-004200mm50mm480mmPlastic: 0.87
LRF-005300mm70mm500mmMetal: 3.78
Plastic: 1.82
LRFJ-001, LRFJ-002Minimum: 75mmMinimum: 35mm100MMetal: 0.12
LRFJ-003Maximum: 150mmMaximum: 50mm100MPlastic: 0.02
LRFJ-004, LRFJ-005, LRFJ-006, LRFJ-007Minimum: 200mmMinimum: 50mm100MMetal: 1.68
LRFJ-008Maximum: 350mmMaximum: 80mm100MPlastic: 0.87
400mm80mm100MMetal: 3.67
Plastic: 3.00

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 6

510(k) Summary

K243179

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's Information

Name: Yichang Xinxin Paper Products Co., Ltd.
Address: No. 308, Jianghan Avenue, Majiadian Street, Zhijiang City, Yichang City, Hubei Province, 443200, China.
Tel: +86 18086241888
Contact: Peng Xiao

Designated Submission Correspondent

Contact: Mr. Boyle Wang
Name: Shanghai Truthful Information Technology Co., Ltd.
Address: Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120, China.
Tel: +86-21-50313932
Email: Info@truthful.com.cn
Date Submitted: June 23, 2025

2.0 Device Information

Trade/Device name: Sterilization Pouch and Roll
Common name: Sterilization Pouch and Roll
Classification name: 1) Wrap, Sterilization
2) Indicator, Physical/Chemical Sterilization Process
Classification Product Code: 1) FRG
Subsequent Product Code: 2) JOJ
Regulation number: 1) 21 CFR880.6850
2) 21 CFR880.2800
Classification: Class II
Panel: General Hospital

3.0 Predicate Device Information

Predicate Device:
Manufacturer: Sigma Medical Supplies Corp.
Trade/Device Name: SIGMA Sterilization Pouch and Roll

Page 7

510(k) number: K180661

4.0 Device Description

The subject Sterilization Pouch and Roll device has five types:

(1) Self-sealing sterilization pouches:
These pouches are made from a medical grade plastic film that is heat sealed on three sides. The fourth side has an adhesive strip that is used to seal the pouch. Release paper used in the pouch is a laminated sheet with composing structure of PE/paper/PE. It is a strip to cover the adhesive area and is released before seal the pouch. The medical grade paper conforms to recognized material standards and can be sterilized by steam. The Process Indicators Ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam.

(2) Sterilization pouches, Flat:
These pouches have the same components with the Self-sealing sterilization pouches, except the fourth side is left opened instead of an adhesive strip and will be heat-sealed when using.

(3) Sterilization pouches, Gusseted:
These pouches are the same with the Sterilization pouches, flat, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.

(4) Sterilization rolls, Flat:
These rolls are made from a medical grade paper and plastic film that are heat sealed on opposite two sides. It will be cut into the suitable length and the opened sides will be heat-sealed. The indicators printed on the medical grade paper are the same with the self-sealing sterilization pouches.

(5) Sterilization rolls, Gusseted:
These rolls are the same with the flat sterilization roll, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.

The Sterilization pouch and roll is composed of medical grade paper (60g/m2) and medical compound film (52μm), it is intended to be used to contain medical devices to be terminally sterilized by the Steam sterilization process. The recommended sterilization cycle parameters are:
Pre-vacuum Steam: 4 minutes at 132°C (270°F); 30 minutes dry time.

The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized for the Steam Sterilization Process. The heat-sealed pouch/roll are heat sealed prior to sterilization processing. After completion of the sterilization process, the

Page 8

Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened.

The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 6 months post sterilization. The shelf life of the device is 3 months.

The Pouch/Roll is printed with a chemical indicator bar that changes from Blue to Dark grey when exposed to steam during the sterilization process. The steam chemical Indicator offers an addition way to verity processing in the sterilization cycle. The endpoint stability of the chemical indicator color change to dark grey is 3 months. The chemical Indicator should be used in addition to, not in place of, the biological indicator. The steam chemical indicator does not signify sterilization; it only indicates that the indicator has been exposed to steam.

The following table (Table 1~5) lists the model numbers of the sterilization pouch and roll by model, dimensions, and content/max. load (lbs.):

Table 1-The Model of Self-sealing sterilization pouches

ItemSpecification Size (W x L)Primary Package DimensionContent / Max Load (lbs)
No.Inchmmmm
PZF-0012.24"X2.76"57mmX70mm62X55X75
PZF-0022.17"X3.94"55mmX100mm60X55X105
PZF-0032.26"X5.12"57.5mmX130mm62X55X135
PZF-0042.26"X5.31"57.5mmX135mm62X55X140
PZF-0052.76"X10.24"70mmX260mm75X50X265
PZF-0063.54"X5.31"90mmX135mm95X50X140
PZF-0073.54"X6.5"90mmX165mm95X50X170
PZF-0083.54"X10.24"90mmX260mm95X50X265
PZF-0083.54"X12.36"90mmX314mm95X50X314
PZF-0105.31"X10.04"135mmX255mm140X50X260
PZF-0115.31"X7.2"135mmX183mm140X50X188
PZF-0125.31"X11.02"135mmX280mm140X50X285

Page 9

Plastic: 0.23
PZF-0135.31"X11.61"135mmX295mm140X50X300Metal: 0.27, Plastic: 0.23
PZF-0145.31"X13.19"135mmX335mm140X50X340Metal: 0.27, Plastic: 0.23
PZF-0155.91"X24.02"150mmX610mm155X50X615Metal: 0.27, Plastic: 0.23
PZF-0165.91"X14.17"150mmX360mm155X50X365Metal: 0.27, Plastic: 0.23
PZF-0177.87"X17.13"200mmX435mm205X50X440Metal: 0.27, Plastic: 0.23
PZF-0189.84"X15.55"250mmX395mm255X50X400Metal: 0.27, Plastic: 0.23
PZF-01911.81"X14.96"300mmX380mm305X50X385Metal: 0.27, Plastic: 0.23
PZF-02011.81"X19.49"300mmX495mm305X50X500Metal: 2.92, Plastic: 1.13
PZF-02114.02"X19.02"356mmX483mm360X50X488Metal: 2.92, Plastic: 1.13
PZF-02214.96"X25.0"380mmX635mm385X50X640Metal: 2.92, Plastic: 1.13
PZF-02315.35"X19.69"390mmX500mm395X50X505Metal: 2.92, Plastic: 1.13
PZF-02418.11"X24.02"460mmX610mm465X50X610Metal: 2.92, Plastic: 1.13

Table 2-The Model of Sterilization pouches, Flat

ItemSpecification Size (W x L)Primary Package DimensionContent / Max Load (lbs)
No.Inchmmmm
PRF-0012.24"X22.83"57mmX580mm62X55X585
PRF-0021.95"X7.87"75mmX200mm80X55X205
PRF-0031.95"X11.02"75mmX280mm80X55X285
PRF-0043.94"X16.54"100mmX420mm105X50X425
PRF-0055.91"X7.87"150mmX200mm155X50X205
PRF-0065.94"X11.81"150mmX300mm155X50X305
PRF-0075.94"X21.65"150mmX550mm155X50X555

Page 10

Plastic: 0.23
PRF-0087.87"X16.54"200mmX420mm205X50X425Metal: 0.27, Plastic: 0.23
PRF-0099.84"X14.17"250mmX360mm255X50X360Metal: 0.27, Plastic: 0.23
PRF-01011.81"X20.08"300mmX510mm305X50X515Metal: 2.92, Plastic: 1.13
PRF-01115.75"X22.83"400mmX580mm405X50X585Metal: 3.38, Plastic: 1.13
PRF-01216.54"X23.62"420mmX600mm425X50X605Metal: 3.38, Plastic: 1.13

Table 3-The Model of Sterilization pouches, Gusseted

ItemSpecification Size (W x L)Primary Package DimensionContent / Max Load (lbs)
No.Inchmmmm
LRF-0013.94"X1.57"X11.81"100mmX40mmX300mm105X90X305
LRF-0025.91"X1.97"X15.75"150mmX50mmX400mm155X100X405
LRF-0037.87"X1.97"X15.75"200mmX50mmX400mm205X100X405
LRF-0049.84"X2.36"X18.9"250mmX60mmX480mm255X110X485
LRF-00511.81"X2.76"X19.69"300mmX70mmX500mm305X120X505

Table 4-The Model of Sterilization rolls, Flat

ItemSpecification Size (W x L)Primary Package DimensionContent / Max Load (lbs)
No.Inchmm & Mmm
PRFJ-0012.17"X118.11"55mmX30M110X11.5X11.5
PRFJ-0022.95"X118.11"75mmX30M150X11.5X11.5
PRFJ-0033.94"X118.11"100mmX30M200X11.5X11.5
PRFJ-0045.0"X118.11"127mmX30M250X11.5X11.5
PRFJ-0055.9"X118.11"150mmX30M200X11.5X11.5
PRFJ-0067.09"X118.11"180mmX30M250X11.5X11.5

Page 11

PRFJ-0077.87"X118.11"200mmX30M200X11.5X11.5Metal: 1.13, Plastic: 0.86
PRFJ-0089.84"X118.11"250mmX30M250X11.5X11.5Metal: 1.13, Plastic: 0.86
PRFJ-00911.81"X118.11"300mmX30M300X11.5X11.5Metal: 1.13, Plastic: 0.86
PRFJ-01013.78"X118.11"350mmX30M350X11.5X11.5Metal: 1.13, Plastic: 0.86
PRFJ-01115.75"X118.11"400mmX30M400X11.5X11.5Metal: 6.24, Plastic: 2.99

Table 5-The Model of Sterilization rolls, Gusseted

ItemSpecification Size (W x L)Primary Package DimensionContent / Max Load (lbs)
No.Inchmm & Mmm
LRFJ-0012.95"X1.38"X393.7"75mmX35mmX100M383X383X77.5
LRFJ-0023.94"X1.57"X393.7"100mmX40mmX100M383X383X102.5
LRFJ-0035.91"X1.97"X393.7"150mmX50mmX100M383X383X152.5
LRFJ-0047.87"X1.97"X393.7"200mmX50mmX100M383X383X202.5
LRFJ-0059.84"X2.36"X393.7"250mmX60mmX100M383X383X127.5
LRFJ-00611.81"X2.76"X393.7"300mmX70mmX100M383X383X152.5
LRFJ-00713.78"X3.15"X393.7"350mmX80mmX100M383X383X177.5
LRFJ-00815.75"X3.15"X393.7"400mmX80mmX100M383X383X202.5

Page 12

5.0 Indication for Use Statement

The Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for steam sterilization. The recommended sterilization cycle is as follows:
• Pre-vacuum Steam: 4 minutes at 132°C (270°F); 30 minutes dry time.

The sterilization pouch and roll are made with medical grade paper and medical compound film. The sterilization pouch and roll maintain the sterility of the enclosed devices for up to 6 months post steam sterilization. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone a steam sterilization process. The steam sterilization indicator color will change from blue before sterilization to dark grey after sterilization.

Table of Maximum Load:

ItemSize (W)Size (L)Content / Max Load (lbs)
PZF-001, PZF-002, PZF-003, PZF-004, PZF-005, PZF-006, PZF-007, PZF-008, PRF-002, PRF-003Minimum: 55mmMinimum: 70mmMetal: 0.02
Maximum: 90mmMaximum: 280mmPlastic: 0.01
PZF-009, PZF-010, PZF-011, PZF-012, PZF-013, PZF-014, PZF-015, PZF-016, PZF-017, PZF-018, PZF-019, PRF-001, PRF-004, PRF-005, PRF-006, PRF-007, PRF-008, PRF-009Minimum: 57mmMinimum: 183mmMetal: 0.27
Maximum: 300mmMaximum: 610mmPlastic: 0.23
PZF-020, PZF-021, PZF-022, PZF-023, PZF-024, PRF-010Minimum: 300mmMinimum: 483mmMetal: 2.92
Maximum: 460mmMaximum: 635mmPlastic: 1.13
PRF-011400mm580mmMetal: 3.38
PRF-012420mm600mmPlastic: 1.13
PRFJ-001, PRFJ-002, PRFJ-003, PRFJ-004, PRFJ-005, PRFJ-006Minimum: 55mm30MMetal: 0.02
Maximum: 180mmPlastic: 0.01
PRFJ-007, PRFJ-008, PRFJ-009, PRFJ-010Minimum: 200mm30MMetal: 1.13
Maximum: 350mmPlastic: 0.86
PRFJ-011400mm30MMetal: 6.24
Plastic: 2.99

Page 13

6.0 Technological Characteristic Comparison Table

Table 2- Comparison of Technology Characteristics

ItemSubject DevicePredicate DeviceComparison Analysis
Product NameSterilization Pouch and RollSIGMA Sterilization Pouch and Roll----
510(k) No.K243179K180661----
Product CodeFRG, JOJFRG, JOJSame
Regulation No.21 CFR 880.6850, 21 CFR 880.280021 CFR 880.6850, 21 CFR 880.2800Same
ClassIIIISame
Intended UseThe Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for steam sterilization. The recommended sterilization cycle is as follows: • Pre-vacuum Steam: 4 minutes at 132°C (270°F); 30 minutes dry time. The sterilization pouch and roll are made with medical grade paper and medical compound film. The sterilization pouch and roll maintain the sterility of the enclosed devices for up to 6 months post steam sterilization. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone a steam sterilization process. The steam sterilization indicator color will change from blue before sterilization to dark grey after sterilization.The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam autoclaves and via Ethylene Oxide (EO). The recommended steam sterilization cycle parameters are 30 minutes at 121°C. The recommended EO sterilization cycle is 4 hours at 55°C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintain the enclosed devices up until 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.Similar

Page 14

ItemSubject Device (K243179)Predicate Device (K180661)Comparison Analysis
Material CompositionMedical Grade Paper, CPP, PET, PU adhesive, Steam Process Indicator Print InkMedical Grade Paper, CPP, PET, PU adhesive, EO and Steam Process Indicator Print InkSame
Sterilization Cycles• Pre-vacuum steam: 4 minutes at 132°C (270°F); 30 minutes dry time.• Steam sterilization cycle parameters are 30 minutes at 121℃. • EO sterilization cycle is 4 hours at 55 ℃ with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L.Similar
TypesSelf-sealing sterilization pouches, Sterilization pouches, Flat Sterilization pouches, Gusseted, Sterilization rolls, Flat Sterilization rolls, GussetedSelf-sealing sterilization pouches Sterilization pouches, Flat Sterilization pouches, Gusseted Sterilization rolls, Flat Sterilization rolls, GussetedSame
Configuration/DimensionsVarious Sizes, Heat SealingVarious Sizes, Heat SealingSimilar
Sterilant Penetration EfficacyThe test meets the requirement of SAL 10-6The tests meet the requirement of SAL 10-6Same
Chemical Indicator (CI) Functionality and EndpointThe sterilant penetrated through the pouch configuration and affected the CI color change to the endpoint color.The sterilant penetrated through the pouch configuration and affected the CI color change to the endpoint color.Same
Device Design of CIThe color of Chemical Indicator changes from Blue to dark grey, when exposed to Steam.The color of Chemical Indicator changes from Blue to Greenish Black, when exposed to Steam.Similar

Page 15

ItemSubject Device (K243179)Predicate Device (K180661)Comparison Analysis
Thickness Variations (mm) ASTM F2251PassedPassedSame
Tensile strength ISO 1924-2PassedPassedSame
Burst Strength (kPa) ASTM F1140; ISO 11607-1> 3.0 Kpa PassedPassedSame
Bubble Leak Test ASTM D3078-02(2013)No Leakage PassedPassedSame
Seal Peel Test (N/15mm) ASTM F88/F88M; ISO 11607-1Min value= 2.08 PassedPassedSame
Dye penetration Test ASTM F1929; ISO 11607-1No Infiltration PassedPassedSame

Page 16

ItemSubject Device (K243179)Predicate Device (K180661)Comparison Analysis
Microbial Barrier Test DIN 58953-6PassedPassedSame
End point stability testing resultsThe color of chemical indicator is blue after 3 months before sterilization.The color of chemical indicator for EO sterilization indicator ink is Red, and the color of chemical indicator for steam sterilization indicator ink is Blue after 2-year shelf life before sterilization.Similar
The color of the chemical indicator is dark grey after 3 months post sterilization.The color of chemical indicator for EO sterilization indicator ink is yellow, and the color of chemical indicator for steam sterilization indicator ink is Blue after EO sterilized and 3 years shelf lifeSimilar
Maintenance of Sterility6 months3 years post EO gas sterilization 6 months post Steam sterilizationDifferent
Shelf Life3 months from date of manufacture3 years from date of manufacture for EO and Steam IndicatorsDifferent
BiocompatibilityConform with ISO 10993-1, ISO 10993-4, ISO 10993-5, ISO 10993-10, ISO 10993-23, ISO 10993-11.Conform with ISO 10993 standardsSame

The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device.

The comparison between the subject and predicate devices is based on the following:
• Same intended use
• Same indications for use
• Similar material types that meet ISO 10993 biocompatibility requirements
• Similar steam sterilization method
• Same fundamental technology/principal of operation/user interface
• The sterilization parameters of the subject device are different than those of the predicate device, but the steam sterilization validation results fully meet the requirements of ISO 17665-1.

Page 17

7.0 Summary of Non-Clinical Testing

Summary of non-clinical and performance testing bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The test results demonstrated the subject device is confirmed to be safe and effective for the intended use.

Table 3: Performance testing summary – Bench

Test ItemTest MethodologyAcceptance Criteria or End PointResults
Sterilant penetration/drying timeISO 17665-1: 2006; ISO 17665-2: 2009- Meets the requirement of SAL 10-6 - The weight difference before sterilization and after drying shall not exceed 3 %Pass
Biocompatibility testing
ISO 10993-5:2009Non-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells. Pass
ISO 10993-23:2021Non-irritatingUnder the conditions of the study, not an irritant. Pass
ISO 10993-10:2021Non-sensitizingUnder conditions of the study, not a sensitizer. Pass

Page 18

Test ItemTest MethodologyAcceptance Criteria or End PointResults
Package Integrity / Material Compatibility /Sterility Maintenance
Package appearanceASTM F 1886-16The appearance of the sterilization pouch is clean, intact and sealed on all four sides.Pass
ThicknessASTM F2251-1352±12%μmPass
Tensile strength of paperISO 1924-2:2008MD≥4.4 KN/m CD≥2.2 KN/mPass
Bursting testASTM F1140/F1140M-13Burst value > 3 Kpa or No BurstPass
Dye penetration testASTM F1929-2023The dye solution is no any leakage across the seal width of sterile barrier system. (No Infiltration)Pass
Seal strength testASTM F88-2023≧2.5N/15mmPass
Vacuum leakage testASTM D3078-02No water can penetrate the sterilization pouch.Pass
Microbial barrierDIN 58953-60 CFU No growthPass
Chemical Indicator TestingISO 11140-1:2014Color of indicator changes from blue to dark grey after Steam sterilizationPass

Page 19

Test ItemTest MethodologyAcceptance Criteria or End PointResults
Chemical Indicator TestingISO 11140-1:2014Shelf life: 3 months Endpoint stability: 3 monthsPass
Shelf lifeASTM F1980-21Shelf life: 3 monthsPass

8.0 Summary of Clinical Testing

No clinical study is included in this submission.

9.0 Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicated device, K180661.

Page 20

[This page appears to be blank]

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).