K Number
K243138

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-02-04

(127 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Geon (S2) Nasal Aspirator is a hand-held, non-invasive, electronic medical device that provides negative pressure to remove nasal secretions and mucus from children (age 2-12 years old) at home environment.

Device Description

Geon Nasal Aspirator, model S2, is a hand-held, non-invasive, electronic medical device that provides negative pressure to remove nasal secretions and mucus from children (age 2-12 years old). Geon (S2) Nasal Aspirator is intended for over-the-counter (OTC) use at home environment. The main principle of operation is to utilize a pump to generate negative pressure in the suction system, which allows nasal secretions to flow into the device container.

AI/ML Overview

The Geon (S2) Nasal Aspirator is a hand-held, non-invasive, electronic medical device that provides negative pressure to remove nasal secretions and mucus from children (age 2-12 years old) at home environment.

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance (Geon S2)Predicate Device (K220126) Claimed PerformancePredicate Device (K220126) Head-to-head Test Performance
Suction Vacuum54kPa ~ 66kPa (tolerance: 3 kPa)52-64 Kpa57~60 Kpa
Noise55dB ~ 60dB (tolerance: 5 dB)55dB~60dB57~60dB

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify a separate "test set" in the context of clinical or comparative studies with a patient population. Instead, the performance evaluations were conducted through verification testing of the device's technical specifications.

  • Sample Size: The document doesn't explicitly state the number of units tested for suction vacuum and noise. However, it indicates that "For Model S2, no individual differences exceed the spective tolerance of suction vacuum and noise after the cleaning and disinfection procedure," implying multiple units or multiple measurements were taken to confirm consistency.
  • Data Provenance: The document does not provide details on the country of origin of the data or whether it was retrospective or prospective, as the testing described appears to be bench-top engineering verification rather than a clinical study.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications:

Not applicable. The ground truth for this device's performance is based on engineering specifications and direct physical measurements (suction vacuum and noise levels) rather than expert interpretation of clinical findings.

4. Adjudication Method for the Test Set:

Not applicable. As a bench-top engineering verification, there was no subjective judgment from experts requiring an adjudication method. The measurements against the specified tolerances determined success or failure.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not conducted. This type of study is typically used for diagnostic devices where human readers interpret medical images or data. The Geon (S2) Nasal Aspirator is a therapeutic/extraction device, and its performance is evaluated based on physical parameters like suction and noise, not diagnostic accuracy requiring human interpretation.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

Yes, the testing described is a standalone performance evaluation of the device. The device's performance (suction vacuum and noise) was measured directly on the device itself without human intervention in operating it beyond standard use, or human interpretation of outputs.

7. Type of Ground Truth Used:

The ground truth used was engineering specifications and direct physical measurements. The device's suction vacuum was measured in kPa and its noise level in dB, and these measurements were compared against pre-defined tolerance ranges established by the manufacturer.

8. Sample Size for the Training Set:

Not applicable. This device is a hardware product, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 4, 2025

Geon Corporation Henry Chen General Manager No. 12, Gung Ye Road Hsi Hu Chang Hwa Hsien, 514 Taiwan

Re: K243138

Trade/Device Name: Geon (S2) Nasal Aspirator Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: January 8, 2025 Received: January 8, 2025

Dear Henry Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Joyce C. Lin -S

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243138

Device Name

Geon (S2) Nasal Aspirator

Indications for Use (Describe)

Geon (S2) Nasal Aspirator is a hand-held, non-invasive, electronic medical device that provides negative pressure to remove nasal secretions and mucus from children (age 2-12 years old) at home environment.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) #:K243138510(k) SummaryPrepared on: 2025-01-08
-------------------------------------------------------------
Contact Details21 CFR 807.92(a)(1)
Applicant NameGeon Corporation
Applicant AddressNo. 12, Gung Ye Road, Hsi Hu, Chang Hwa Hsien 514 Taiwan
Applicant Contact Telephone(886)48852476
Applicant ContactMr. Henry Chen
Applicant Contact Emailinfo@geon.com.tw
Device Name21 CFR 807.92(a)(2)
Device Trade NameGeon (S2) Nasal Aspirator
Common NamePowered suction pump
Classification NamePump, Portable, Aspiration (Manual Or Powered)
Regulation Number878.4780
Product Code(s)BTA
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
-----------------------------------------------------------
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K220126GeonicBTA
Device Description Summary21 CFR 807.92(a)(4)
---------------------------------------------------

Geon Nasal Aspirator, model S2, is a hand-held, non-invasive, electronic medical device that provides negative pressure to remove nasal secretions and mucus from children (age 2-12 years old). Geon (S2) Nasal Aspirator is intended for over-the-counter (OTC) use at home environment.

Intended Use/Indications for Use21 CFR 807.92(a)(5)
---------------------------------------------------------

Geon (S2) Nasal Aspirator is a hand-held, non-invasive, electronic medical device that provides negative pressure to remove nasal secretions and mucus from children (age 2-12 years old) at home environment.

Indications for Use Comparison21 CFR 807.92(a)(5)
-------------------------------------------------------

The indication of use the subject devices (Model S2) is same as the predicate device K220126.

Technological Comparison

The Geon Nasal Aspirator (Model Name: S2) is substantially equivalent the predicates Geonic Nasal Aspirator (K220126) as a CFR 880.4780 Powered Suction Pump. The subject device and predicates (K220126) have the same Indication for use, Motor type (3V DC), and Type BF applied part, Contact parts and materials. The main principle of operation is to utilize a pump to generate negative pressure in the suction system, which allows nasal secretions to flow into the device container. All of these are a portable, DC powered devices intended to provide the suction fonction to aspirate children's nasal secretion from the same population of children (age 2-12 years old), The subject device has similar characteristics as the predicates (K220126), including Vacuum pressure, Storage container size, Noise level,

21 CFR 807.92(a)(6)

{5}------------------------------------------------

Suction Tip, Power source, Materials, Operating condition. The main difference between the subject device and the predicates (K220126) is the shape of the suction tip. It does not impact the safety and effectiveness.

The difference in design does not affect performance, or raise different questions of safety and effect and indirect tissue contacting materials of the subject device are same as the predicates (K220126), thus the subject device also meets the requirement of biocompatibility. The subject device complies with general requirements for basic safety and essential performance, and electromagnetic compatibility according to IEC 60601-1, IEC60601-1-2, and IEC 60601-1-11.

By definition, the subject device, (Geon (S2) Nasal Aspirator is substantially equivalent to the predicate device, K220126. Bench testing contained in the submission demonstrates the device does not raise different questions regarding its safety and effectiveness as compared to the predicate devices.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The evaluation of suction vacuum and noise were completed during verification testing of Geon (S2) Nasal Aspirator. The test reports showed that the device passed all the testing. For Model S2, no individual differences exceed the spective tolerance of suction vacuum and noise after the cleaning and disinfection procedure. As part of demonstrating the acceptable performance of the Geon (52) Nasal Aspirator in showing substantial equivalence to the predicate device:

  1. Suction vacuum

54kPa ~ 66kPa (tolerance: 3 kPa)

K220126: 52-64 Kpa (claimed), 57~60 Kpa (Head-to-head test)

    1. Noise
      55dB~ 60dB (tolerance: 5 dB)

K220126: 55dB60dB (claimed), 5760dB (Head-to-head test)

The device meets all the requirements for the specifications defined by the manufacturer.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.