K Number
K243097

Validate with FDA (Live)

Date Cleared
2024-12-26

(87 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RM STAR EX with RMS Needle is intended for use in dermatologic procedures for electrocoagulation and hemostasis.

Device Description

The RM STAR EX with RMS Needle includes a system main device and a handpiece equitable with a bipolar electrode. The RF signal is generated from the main device which is then delivered to the handpiece and then to bipolar electrode. The RF signal is delivered to the target tissue using penetrating needle electrodes in the consumable tip. The bipolar electrode is placed in light contact with the epidermis while the handpiece is being held at right angles to the target tissue. As the RF signal passes through the skin, it generates an electrothermal reaction which is capable of coagulating the tissue. Using the consumable tip, RM STAR EX with RMS Needle creates heat within the target skin tissue via needle electrodes from the bipolar electrode.

AI/ML Overview

The provided document describes a 510(k) premarket notification for the "RM STAR EX with RMS Needle" device. This device is an electrosurgical cutting and coagulation device and accessories, intended for use in dermatologic procedures for electrocoagulation and hemostasis. The submission aims to demonstrate substantial equivalence to the predicate device "SYLFIRM X" (K213612).

However, the document does not contain information about acceptance criteria and a study that proves the device meets those criteria in the context of:

  • A table of acceptance criteria and reported device performance.
  • Sample size used for a test set (clinical data related to accuracy/performance against a ground truth).
  • Data provenance for a test set.
  • Number of experts and their qualifications for establishing ground truth.
  • Adjudication method for a test set.
  • Multi-reader multi-case (MRMC) comparative effectiveness study.
  • Standalone performance (algorithm only without human-in-the-loop performance).
  • Type of ground truth used (expert consensus, pathology, outcomes data).
  • Sample size for the training set (as this is not an AI/ML device in the context of diagnostic interpretation).
  • How ground truth for the training set was established.

The document primarily focuses on demonstrating the substantial equivalence of the RM STAR EX with RMS Needle to its predicate device through non-clinical performance data. This includes:

  • Biocompatibility Testing: Conducted according to ISO 10993-1, covering cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity.
  • Sterilization Validation Test: Performed using the biological indicator (BI) overkill method following ISO 11135 guidelines to confirm a sterility assurance level of 10^-6 for EO sterilization.
  • Shipping, Packaging, and Shelf Life: Included pyrogen tests (USP <161>, <85>), package integrity testing (ASTM D4169-22), and sterile barrier packaging testing (ASTM F88/F88M-15, F1929-15). Shelf life validated using ASTM F1980-21.
  • Electrical Safety and EMC Test: Evaluated according to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC TR 60601-4-2, EN 55011, and IEC 61000-3-2.
  • Software Validation Test: Performed for a device with a moderate level of software concern.
  • Pre-clinical Test (Animal Study): This is the closest to a performance study, assessing the safety and effectiveness on tissue.

Pre-clinical Test (Animal Study) Details:

  • Study Design: Radiofrequency (RF) microneedling was applied to liver, kidney, muscle, and skin of rabbits. Animals were euthanized at 3 hours, 5 days, and 22 days post-treatment.
  • Tissue Reactions:
    • Internal organ models: Minimal hemorrhage and fibrosis, no significant gross lesions.
    • Skin model: Mild dermal necrosis initially, good recovery with minimal fibrosis by Day 22.
  • Histological Findings: Localized and well-controlled tissue damage, signs of healing and regeneration, no major adverse reactions.
  • Conclusion: Demonstrated controlled tissue interaction, no adverse effects, and effective healing in both skin and internal organ models, confirming safety and effectiveness.

Differences with Predicate Device and Justification (related to performance):

The document highlights differences in Conduction Time, Depth, and Number of Electrodes compared to the predicate. For each, the differences are argued not to raise new questions of safety and effectiveness, supported by the animal test (Doc. No. NT24-00035, NT24-00036).

  • Conduction Time: Subject device (100-200 ms and continuous) shorter than predicate (120-300 ms). Argued not to raise different questions of safety and effectiveness because shorter conduction time.
  • Depth: Subject device (0-2.0 mm) less than predicate (0.3-4.0 mm). Argued not to raise different questions of safety and effectiveness because depth is not greater.
  • Number of Electrodes: Subject device (13 pins) less than predicate (25 pins). Argued not to raise different questions of safety and effectiveness because fewer electrodes.

Summary of missing information requested:

The request asks for details typically associated with clinical studies or AI/ML performance evaluations (e.g., test set sample size, expert consensus ground truth, MRMC studies). The provided text describes a submission for an electrosurgical device based on engineering and pre-clinical animal studies, not a diagnostic device relying on human interpretation or AI algorithms requiring clinical validation of diagnostic accuracy metrics. Therefore, the specific criteria for "acceptance criteria and study that proves the device meets the acceptance criteria" in the context of diagnostic performance (as implied by the specific questions) are not applicable or present in this 510(k) summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 26, 2024

Bomtech Electronics Co., Ltd. % Juntaek Seo President Corazon 20, Jukjeon-ro, Giheung-gu, Gyeonggi-do Yongin, 16897 Korea. South

RE: K243097

Trade/Device Name: RM STAR EX with RMS Needle Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 30, 2024 Received: September 30, 2024

Dear Seo Juntaek:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Francisco Delgado -S 11:42:22 2024.12.26 '00,05-

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243097

Device Name RM STAR EX with RMS Needle

Indications for Use (Describe)

The RM STAR EX with RMS Needle is intended for use in dermatologic procedures for electrocoagulation and hemostasis.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

I.Submitter
510(k) Submitter:BOMTECH ELECTRONICS CO., LTD.
322, Gonghang-daero, Gangseo-gu, Seoul, Republic oKoreaPhone: +82 2 523 8295
Primary Contact:Juntaek Seoinfo@corazon.kr
Alternate Contact:Junseok LeeRA managerbomtech@bomtech.net
Date Prepared:December 23, 2024
II.Device Name
Device Trade Name:RM STAR EX with RMS Needle
Common Name:Electrosurgical cutting and coagulation device andaccessories
Regulation Number:21 CFR 878.4400
Product Code:GEI
Regulatory Class:Class II
Classification Panel:General and Plastic Surgery
III.Legally Marketed Predicate Device
  • Predicate Trade Name: SYLFIRM X

IV. Device Description

Predicate #:

The RM STAR EX with RMS Needle includes a system main device and a handpiece equitable with a bipolar electrode. The RF signal is generated from the main device which is then delivered to the handpiece and then to bipolar electrode. The RF signal is delivered to the target tissue using penetrating needle electrodes in the consumable tip. The bipolar electrode is placed in light contact with the epidermis while the handpiece

K213612

{5}------------------------------------------------

is being held at right angles to the target tissue. As the RF signal passes through the skin, it generates an electrothermal reaction which is capable of coagulating the tissue. Using the consumable tip, RM STAR EX with RMS Needle creates heat within the target skin tissue via needle electrodes from the bipolar electrode.

Model NumbersModel Name
Main Body + HandpieceRMS Needle
1RMS-13P31G
2RM STAR EXRMS-13P32G
3RMS-13P33G

The model names and numbers of the products applied for are as follows:

V. Indications for Use

The RM STAR EX with RMS Needle is intended for use in dermatologic procedures for electrocoagulation and hemostasis.

VI. Comparison of technological characteristics with the predicate device

The RM STAR EX with RMS Needle is substantially equivalent to the following predicate device:

Primary Predicate: For Indication for Use and Technological Characteristics

ItemRM STAR EXSYLFIRM XRemark
510(k) NumberK243097K213612-
ManufacturerBomtech Electronics Co., Ltd.VIOL Co., Ltd.-
Regulation Number No.21 CFR 878.440021 CFR 878.4400Same
Product CodeGEIGEISame
Indications for UseThe RM STAR EX with RMSNeedle is intended for use indermatologic procedures forelectrocoagulation and hemostasis.The SYLFIRM X is intendedfor use in dermatologicprocedures forelectrocoagulation andhemostasis.Same
Prescription or OTCPrescriptionPrescriptionSame
ESUGeneralRM STAR EX is a bipolarelectrosurgical device for tissuecoagulation using high-frequencyRF current.SYLFIRM X is a bipolarelectrosurgical device for tissuecoagulation using high-frequency RF current.Same
Major functions• Operation type: Coagulation• Operation type: CoagulationSame

{6}------------------------------------------------

RM STAR EX with RMS Needle

Traditional 510(k)

ItemRM STAR EXSYLFIRM XRemark
Active accessoryPerformancespecificationsOutput power: Max 16 W (50 Ω) Conduction time Mode Conduction time Level 1 100 msec Level 2 120 msec Level 3 160 msec Level 4 200 msec Level 5 Continuous Output frequency: 2 MHzOutput power: Max 16 W (50 Ω) Conduction time Mode Conduction time CW1 120 msec CW2 160 msec CW3 200 msec CW4 300 msec PW1 120 msec PW2 160 msec PW3 200 msec PW4 240 msec Output frequency: 2 MHzSimilar (Note 1)
PhysicalspecificationsInput voltage: AC 100 - 240 V, 50/60 HzInput voltage: AC 100 - 240 V, 50/60 HzSame
Monopolar or bipolarBipolarBipolarSame
PhysicalspecificationsDepth: 0 - 2.0 mm (unit: 0.5 mm)Depth: 0.3 - 4.0 mm (unit: 0.1 mm)Similar (Note 2)
DisposableYesYesSame
Consumable tipProvidedProvidedSame
Number of electrodesRMS-13P31G13-pinRMS-13P32G13-pinRMS-13P33G13-pinJ25BM 25-pin (5x5)J25BS 25-pin (5x5)J18BS 18-pin (3x6)Similar (Note 3)
BiocompatibilityThis is satisfied with biocompatibility according to ISO 10993.This is satisfied with biocompatibility according to ISO 10993.Same
Neutral electrodesNot applicableNot applicableSame

Note 1: Conduction time

The conduction time of the subject device is 100 msec to 200 msec and continuous. The conduction time of the predicate device is 120 msec to 300 msec.

These differences in conduction time do not raise different questions of safety and effectiveness because conduction time of the subject device is shorter than that of the predicate device.

The safety and effectiveness of the subject device was supported by animal test (Doc. No. NT24-00035, NT24-00036).

Note 2: Depth

The depth of the subject device is 0 to 2.0 mm.

The depth of the predicate device is 0.3 to 4.0 mm.

These differences in depth do not raise different questions of safety and effectiveness because depth of the subject device is not greater than that of the predicate device. The safety and effectiveness of the subject device was supported by animal test (Doc. No. NT24-00035, NT24-00036).

Note 3: Number of electrodes

The number of electrodes of the subject device is 13 pins. The number of electrodes of the predicate device is 25 pins.

{7}------------------------------------------------

These differences in number of electrodes do not raise different questions of safety and effectiveness because the number of electrodes of the subject device is fewer than that of the predicate device.

The safety and effectiveness of the subject device was supported by animal test (Doc. No. NT24-00035, NT24-00036).

Based on the above comparison as well as the information provided in the SE Comparison Table in Section VI, supported by the appropriate performance data, the proposed device has been shown to be as safe and effective as the legally marketed predicate device for the proposed Indications for Use.

VII. Performance data

1. Biocompatibility Testing

In accordance with ISO 10993-1, the subject device is classified as an externally communicating device in contact with tissue/bone/dentin for limited duration (< 24 hours). The following testing was conducted.

    1. Cytotoxicity
    1. Sensitization
    1. Irritation
    1. Acute Systemic Toxicity
    1. Material-Mediated Pyrogenicity

2. Sterilization Validation Test

To confirm a sterility assurance level of 10-6 for EO sterilization, the validation process employed the biological indicator (BI) overkill method, following the guidelines outlined in ISO 11135. ISO 10993-1. ISO 10993-7, ISO 11138-1. ISO 11138-2, ISO 11737-1, ISO 11737-2, ISO 11140-1, AAMI TIR 15, AAAMI TIR 16, and AAMI TIR 28.

3. Shipping, Packaging, and Shelf Life

  • Pvrogen test was conducted in accordance with USP <161> Medical Devices -. Bacterial Endotoxin and Pyrogen Tests and USP <85> Bacterial Endotoxins Test.
  • Package integrity testing, after environmental conditioning and simulated . transportation using ASTM D4169-22, was conducted on final packaged and sterile devices. All packaging is acceptable for protection of final finished product and sterility maintenance.
  • Sterile Barrier Packaging Testing was performed on the proposed device: . 0 Seal peel test - ASTM F88/F88M-15

{8}------------------------------------------------

  • o Dye penetration ASTM F1929-15
  • Shelf life of 3 years is validated using the FDA-recognized standard ASTM . F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier System for Medical Devices.

4. Electrical Safety and EMC Test

The subject device has been evaluated for electrical safety and electromagnetic compatibility in accordance with the relevant standards, including IEC 60601-1:2005/AMD2:2020, IEC 60601-1-2:2014/A1:2020, IEC 60601-2-2:2017/AMD1:2023, IEC TR 60601-4-2:2016, EN 55011:2016 + A1:2017 (CISPR 11:2015/A1:2016), and IEC 61000-3-2:2018/AMD1:2020.

5. Software Validation Test

The level of concern of the software of RM STAR EX with RMS Needle is moderate, and the software validation tests were performed.

6. Pre-clinical Test

The RM STAR EX with RMS Needle was evaluated for safety and effectiveness through comprehensive pre-clinical testing using both internal organs and skin models.

    1. Study Design
      Radiofrequency (RF) microneedling was applied to liver, kidney, and muscle as well as the skin of rabbits to assess the localized effects of RF energy. The animals were euthanized at different time points (3 hours, 5 days, and 22 days post-treatment) to observe tissue response and healing.
    1. Tissue Reactions
      In the internal organ models, minimal hemorrhage and fibrosis were observed, with no significant gross lesions at any time point. In the skin model, mild dermal necrosis was observed shortly after treatment, but the tissue exhibited good recovery, with minimal fibrosis noted by Day 22.
    1. Histological Findings
      Histopathological analysis showed that the tissue damage was localized and well-controlled, with signs of tissue healing and regeneration over time. No major adverse reactions were observed in either the skin or internal organs.
    1. Conclusion
      The RM STAR EX with RMS Needle demonstrated controlled tissue interaction with no adverse effects and effective healing in both skin and internal organ models, confirming the safety and effectiveness of the device.

{9}------------------------------------------------

VIII. Conclusion

Based on the test results provided in this submission, including biocompatibility, sterilization validation, shipping, shelf life, EMC and electrical safety, software validation, and animal testing, Bomtech Electronics Co., Ltd. concludes that the RM STAR EX with RMS Needle is substantially equivalent to the predicate device as described herein.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.