K Number
K243080
Date Cleared
2024-10-29

(29 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

Device Description

The Riptide™ Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries, using continuous aspiration. The Riptide™ Aspiration System is composed of the following components and illustrated in Figure 1:

    1. React™ 68 Catheter (REACT-68), React™ 71 Catheter (REACT-71, REACT-71-115, REACT-71-125)
    1. Riptide™ Large Bore Aspiration Tubing (MAT-110-110)
    1. Riptide™ Aspiration Pump (MAP-1000)
    1. Riptide™ Collection Canister with Intermediate Tubing (MAC-1200)
AI/ML Overview

The provided text is a 510(k) Summary for the Riptide™ Aspiration System, which details the device's characteristics and its substantial equivalence to a predicate device. However, this document does not describe a study involving acceptance criteria for algorithmic performance, expert ground truth adjudication, or human reader effectiveness studies. Instead, it focuses on non-clinical bench testing to demonstrate compliance with safety and performance standards for a physical medical device (catheter system).

Therefore, most of the requested information regarding AI/algorithm performance and clinical study details cannot be extracted from this document because it does not pertain to such a device or study.

Here's what can be extracted based on the nature of the document:

1. A table of acceptance criteria and the reported device performance

The document lists various bench tests with pass/fail results. While these are performance metrics related to safety and functionality, they are for a physical medical device, not an AI algorithm.

TestTest Method SummaryReported Device Performance (Results)Acceptance Criteria (Implicitly "Pass")
Riptide™ Aspiration Pump
Electrical Safety (IEC 60601-1)Electrical safety testing conducted per IEC 60601-1 Edition 3.2 2020-08.PassCompliance with IEC 60601-1
Electromagnetic Compatibility (IEC 60601-1-2)Electromagnetic compatibility (EMC) testing conducted per IEC 60601-1-2 Edition 4.1 2020-09.PassCompliance with IEC 60601-1-2
Riptide™ Aspiration Pump and Riptide™ Collection Canister with Intermediate Tubing
Medical Suction Equipment (ISO 10079-1/10079-4)Medical suction equipment testing conducted per ISO 10079-1 Fourth Edition 2022-03 and ISO 10079-4 First Edition 2021-08.PassCompliance with ISO 10079-1 and 10079-4
React™ 71 Catheter
Design Validation (Compatibility with minimum length 136 cm guide catheter)The subject device was evaluated per FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices, Guidance for Industry and Food and Drug Administration Staff" (February 2016) and per FDA recognized consensus standards ISO 14971 (Third Edition 2019-12) and IEC 62366-1 (Edition 1.1 2020-06 CONSOLIDATED VERSION).PassCompatibility with guide catheter
Design Validation (Ability to navigate to the distal M1 segment of the MCA over a microcatheter)The subject device was evaluated per FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices, Guidance for Industry and Food and Drug Administration Staff" (February 2016) and per FDA recognized consensus standards ISO 14971 (Third Edition 2019-12) and IEC 62366-1 (Edition 1.1 2020-06 CONSOLIDATED VERSION).PassNavigability to distal M1 segment

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

This information is not applicable as the document describes bench testing of a physical device, not a performance study on a software algorithm with a test set of data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

Not applicable. Ground truth as typically defined for AI/software evaluations is not mentioned. The "design validation" tests refer to compliance with guidance and standards, not expert clinical interpretation of data.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable. No adjudication method for a test set of data is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not a study assessing AI assistance or human reader performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm, but a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable. The document refers to "non-clinical bench testing" and compliance with international standards (IEC, ISO) and FDA guidance for the safety and performance of the physical device components.

8. The sample size for the training set

Not applicable. There is no mention of a training set as this is not a machine learning device.

9. How the ground truth for the training set was established

Not applicable. There is no mention of a training set or ground truth establishment relevant to an algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 29, 2024

Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Tony Rizzardi Principal Regulatory Affairs Specialist 9775 Toledo Way Irvine, California 92618

Re: K243080

Trade/Device Name: Riptide Aspiration System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: September 27, 2024 Received: September 30, 2024

Dear Tony Rizzardi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243080

Device Name Riptide™ Aspiration System

Indications for Use (Describe)

The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary - K243080

510(k) OwnerMicro Therapeutics, Inc. d/b/a ev3 Neurovascular9775 Toledo WayIrvine, CA 92618, USAEstablishment Registration: 2029214
Contact PersonTony RizzardiPrincipal Regulatory Affairs SpecialistTelephone: (763) 505-3017Email: tony.rizzardi@medtronic.com
Date Summary Prepared29 October 2024
Trade Name of DeviceRiptide™ Aspiration System
Common Name of DeviceCatheter, Thrombus Retriever
Review PanelNeurology
Medical SpecialtyCardiovascular
Product CodeNRY
Regulation Number21 CFR 870.1250
Regulation NamePercutaneous Catheter
Device ClassificationClass II
Predicate DeviceRiptide™ Aspiration System (K201689)

Device Description

The Riptide™ Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries, using continuous aspiration. The Riptide™ Aspiration System is composed of the following components and illustrated in Figure 1:

    1. React™ 68 Catheter (REACT-68), React™ 71 Catheter (REACT-71, REACT-71-115, REACT-71-125)
    1. Riptide™ Large Bore Aspiration Tubing (MAT-110-110)
    1. Riptide™ Aspiration Pump (MAP-1000)
    1. Riptide™ Collection Canister with Intermediate Tubing (MAC-1200)

Image /page/4/Figure/9 description: The image shows a medical device with four labeled parts. Part 1 is a thin, coiled tube. Part 2 is a connector between two tubes. Part 3 is the main unit of the device, which includes a knob and a pressure gauge. Part 4 is a container with measurement markings.

Figure 1. Riptide™ Aspiration System

{5}------------------------------------------------

Indications for Use

The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

AttributePredicate Device:Subject Device:
Riptide™ Aspiration System (K201689)Riptide™ Aspiration System (K243080)
General
Indication forUse StatementThe Riptide™ Aspiration System isintended for use in the revascularizationof patients with acute ischemic strokesecondary to intracranial large vesselocclusive disease (within the internalcarotid, middle cerebral M1 and M2segments, basilar, and vertebralarteries) within 8 hours of symptomonset. Patients who are ineligible forintravenous tissue plasminogenactivator (IV t-PA) or who fail IV t-PAtherapy are candidates for treatment.The Riptide™ Aspiration System isintended for use in the revascularizationof patients with acute ischemic strokesecondary to intracranial large vesselocclusive disease (within the internalcarotid, middle cerebral M1 and M2segments, basilar, and vertebralarteries) within 8 hours of symptomonset. Patients who are ineligible forthrombolytic drug therapy or who failedthrombolytic drug therapy arecandidates for treatment.
DevicesComprising theSystem• React™ 68 Catheter (REACT-68)• React™ 71 Catheter (REACT-71)• Riptide™ Large Bore AspirationTubing (MAT-110-110)• Riptide™ Aspiration Pump(MAP- 1000)• Riptide™ Collection Canister withIntermediate Tubing (MAC-1200)• React™ 68 Catheter (REACT-68)• React™ 71 Catheter (REACT-71,REACT-71-115, REACT-71-125)• Riptide™ Large Bore AspirationTubing (MAT-110-110)• Riptide™ Aspiration Pump(MAP- 1000)• Riptide™ Collection Canister withIntermediate Tubing (MAC-1200)
React™ 68 Catheter
Materials ofConstructionCommonly used medical grade plasticsand metals with hydrophilic coatingSame as K201689
Inner Diameter0.068"Same as K201689
Outer Diameter0.083"Same as K201689
Usable Length132 cmSame as K201689
AccessoryPeelable split-y introducer sheathSame as K201689
PackagingPolyethylene hoop/backing card, nylon-Tyvek™ pouch, SBS cartonSame as K201689
SterilizationEthylene oxide, SAL 10-6Same as K201689
Shelf Life2 yearsSame as K201689
React™ 71 Catheter
AttributePredicate Device:Riptide™ Aspiration System (K201689)Subject Device:Riptide™ Aspiration System (K243080)
Materials ofConstructionCommonly used medical grade plasticsand metals with hydrophilic coatingSame as K201689
Inner Diameter0.071"Same as K201689
Outer Diameter0.0855" (Max)Same as K201689
Usable Length132 cm115, 125, and 132 cm
AccessoryPeelable split-y introducer sheathSame as K201689
PackagingPolyethylene hoop/backing card, nylon-Tyvek™ pouch, SBS cartonSame as K201689
SterilizationEthylene oxide, SAL 10-6Same as K201689
Shelf Life2 yearsSame as K201689
Riptide™ Large Bore Aspiration Tubing
Materials ofConstructionCommonly used medical gradeplasticsSame as K201689
Inner Diameter0.110"Same as K201689
Outer Diameter0.188"Same as K201689
Usable Length112.0"Same as K201689
PackagingNylon-Tyvek™ pouch, SBS cartonSame as K201689
SterilizationEthylene oxide, SAL 10-6Same as K201689
Shelf Life3 yearsSame as K201689
Riptide™ Aspiration Pump
Vacuum Range0-29 inHg (0-98.2 kPa)Same as K201689
Flow RateMAP-1000 (60 Hz): 0-0.8 SCFM (0-23LPM)Same as K201689
VoltageMAP-1000: 110-115 Vac (US)Same as K201689
Frequency60 HzSame as K201689
Duty CycleNon-continuous97% (58.2 minutes on, 1.8 minutes off)Same as K201689
Applied PartClassificationType BFType CF
IEC 60601-1andIEC 60601-1-2ComplianceYes:IEC 60601-1 (Edition 3.1) andIEC 60601-1-2 (Edition 4.0)Yes:IEC 60601-1 (Edition 3.2 2020-08)IEC 60601-1-2 (Edition 4.1 2020-09)
ISO 10079-1andISO 10079-4ComplianceNoYes:ISO 10079-1 (Fourth edition 2022-03)ISO 10079-4 (First edition 2021-08)
PhysicalDimensionsLength: 16.1"Depth: 13.2"Height: 12.3"Same as K201689
WeightMAP-1000: Approximately 24 lbs (11 kg)Same as K201689
Noise60.5 dBa<70 dBa
AttributePredicate Device:Riptide™ Aspiration System (K201689)Subject Device:Riptide™ Aspiration System (K243080)
InternalComponentsCommonly used electrical components,fittings, and tubing.Same as K201689
SterilizationNon-sterileSame as K201689
Operating Life500 hoursSame as K201689
Riptide™ Collection Canister with Intermediate Tubing
Materials ofConstructionCommonly used medical grade plasticsSame as K201689
Volume1200 mLSame as K201689
Shelf Life3 yearsSame as K201689

Comparison of Technological Characteristics

{6}------------------------------------------------

{7}------------------------------------------------

Biocompatibility

There is no impact to the biocompatibility of the Riptide™ Aspiration System associated with the changes subject to K243080.

Performance Data – Bench

The following non-clinical bench tests were conducted for the Riptide™ Aspiration System:

TestTest Method SummaryResults
Riptide™ Aspiration Pump
Electrical Safety(IEC 60601-1)Electrical safety testing conducted per IEC 60601-1Edition 3.2 2020-08.Pass
ElectromagneticCompatibility(IEC 60601-1-2)Electromagnetic compatibility (EMC) testing conductedper IEC 60601-1-2 Edition 4.1 2020-09.Pass
Riptide™ Aspiration Pump and Riptide™ Collection Canister with Intermediate Tubing
Medical Suction Equipment(ISO 10079-1/10079-4)Medical suction equipment testing conducted per ISO10079-1 Fourth Edition 2022-03 and ISO 10079-4 FirstEdition 2021-08.Pass
React™ 71 Catheter
Design Validation(The length of the subjectdevice shall be compatiblewith minimum length 136 cmguide catheter)The subject device was evaluated per FDA guidance"Applying Human Factors and Usability Engineering toMedical Devices, Guidance for Industry and Food andDrug Administration Staff" (February 2016) and per FDArecognized consensus standards ISO 14971 (Third Edition2019-12) and IEC 62366-1 (Edition 1.1 2020-06CONSOLIDATED VERSION).Pass
Design Validation(The subject device should beable to navigate to the distalM1 segment of the MCA over amicrocatheter)The subject device was evaluated per FDA guidance"Applying Human Factors and Usability Engineering toMedical Devices, Guidance for Industry and Food andDrug Administration Staff" (February 2016) and per FDArecognized consensus standards ISO 14971 (Third Edition2019-12) and IEC 62366-1 (Edition 1.1 2020-06CONSOLIDATED VERSION).Pass

{8}------------------------------------------------

Performance Data – Animal

The determination of substantial equivalence is based upon non-clinical bench testing as there is no change to the intended use, fundamental scientific technology, or materials of construction.

Performance Data - Clinical

The determination of substantial equivalence is based upon non-clinical bench testing as there is no change to the intended use, fundamental scientific technology, or materials of construction.

Conclusion

There is no change to the intended use, fundamental scientific technology, principle of operation, and materials of construction of the subject Riptide™ Aspiration System in comparison to the legally marketed predicate Riptide™ Aspiration System (K201689). The design and labeling changes for Type CF applied part classification, meeting latest electrical safety and medical suction equipment standards (including IEC 60601-1, IEC 60601-1-2, ISO 10079-1, and ISO 10079-4), and addition of shorter length sizes of the React™ 71 Catheter (REACT-71-115 & REACT-71-125) to the system do not raise new questions of safety and effectiveness and have been tested through non-clinical bench testing using established scientific methods. The information provided in this submission supports a determination of substantial equivalence for the subject Riptide™ Aspiration System.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).