(270 days)
The Affirm 800 is used in viewing intra-operative blood flow in the cerebral vascular area including, but not limited to, assessing cerebral vessel branch occlusion, as well as intraoperative blood flow and vessel patency in bypass surgical procedures in neurosurgery.
The Affirm 800 module is designed to work in conjunction with the Class I Digital Surgical Microscope (DSM) RE3. The Affirm 800 module, a Class II device, comprises hardware and software components that enable the Digital Surgical Microscope to emit excitation light in specific wavelengths to activate the fluorescence properties of Indocyanine Green (ICG). The fluorescence signal emitted by the patient represents the distribution of the infrared dye in the patient's blood vessels during surgery. The emitted light is then captured by the optics of the digital microscope, passed through filters to remove unwanted wavelengths of light, and finally detected by the image sensors. This detected signal is then projected on a 3D monitor, which is part of the microscope system, enabling the surgeon to view the magnified image.
The integrated Affirm 800 fluorescence module is used to visualize infrared fluorescent areas in the surgical field and includes features to record and play back a video clip of the area of focus where fluorescent light is emitted. The module has been designed for excitation in the wavelength range from 740 nm to 800 nm and for fluorescence visualization in the wavelength range from 820 nm to 900 nm. The fluorescence feature generates an image in the infrared spectrum, which means it cannot be seen by the naked eye.
Here's a breakdown of the acceptance criteria and the study details for the Affirm 800 device, based on the provided FDA 510(k) Clearance Letter:
Note: The provided document is an FDA 510(k) Clearance Letter and a 510(k) Summary. These documents summarize the manufacturer's performance testing and justification for substantial equivalence. They don't typically include granular details about every aspect of the study design (like specific expert qualifications, adjudication methods, or sample sizes for specific training sets) that might be found in a full study report or a more comprehensive submission. Therefore, some information requested may not be explicitly available in this type of document and will be noted as "Not explicitly stated."
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Metric (as implied by test methods) | Acceptance Criteria (implied) | Reported Device Performance |
|---|---|---|---|
| Image Quality (NIR Fluorescence) | Sensitivity to ICG Molar Concentration: | Equivalent to predicate | Pass (for: Weber contrast, SNR, Fluorescence Pixel Intensity, LOD, LOQ, Signal linearity) |
| - Weber contrast | (implied by "confirm equivalent performance") | ||
| - Signal to Noise Ratio (SNR) | |||
| - Fluorescence Pixel Intensity | |||
| - Limit of detection (LOD) | |||
| - Limit of quantification (LOQ) | |||
| - Signal linearity (up to 3.2 micromolar) | |||
| Spatial & Temporal Noise | Spatial noise | Within acceptable limits | Pass |
| Temporal noise | |||
| Stereoscopic Display | Left and Right Eye Comparison | Consistent performance | Pass |
| Stereoscopic Crosstalk: | Within acceptable limits | Pass (for: Stereoscopic Extinction Ratio, Stereoscopic Crosstalk) | |
| - Stereoscopic Extinction Ratio | |||
| - Stereoscopic Crosstalk | |||
| ICG Depth Penetration | Sensitivity to NIR fluorescence depth penetration: | Equivalent to predicate | Pass (for: Weber Contrast, SNR) |
| - Weber Contrast | (implied by "confirm equivalent performance") | ||
| - Signal-to-Noise Ratio | |||
| Clinical Image Quality | Surgeons' image quality evaluation (scale 1-5) | Satisfactory clinical image quality | Pass (image quality evaluated in 5 porcine models) |
| Software Functionality | Software Verification & Validation (V&V) | Compliance with FDA guidance | Pass |
Study Details
-
Sample Size used for the test set and the data provenance:
- Sample Size (Clinical/Animal Study): 5 porcine models.
- Data Provenance: Not explicitly stated, but likely conducted by the manufacturer or a contracted research organization in a controlled setting, given it's an animal study. It is a prospective study by nature of being a "conducted" animal study.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not explicitly stated. The document mentions "Surgeons evaluated image quality." The exact number of surgeons is not provided.
- Qualifications of Experts: Only "Surgeons" are mentioned. No specific experience levels (e.g., "radiologist with 10 years of experience") are provided.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Adjudication Method: Not explicitly stated. The document only says "Surgeons evaluated image quality." It doesn't specify if this was an individual assessment, a consensus, or involved an adjudication process for discrepancies.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No, a multi-reader multi-case comparative effectiveness study was not explicitly mentioned or described. The performance testing focused on device characteristics and an animal study for image quality evaluation, not on human reader performance with or without AI assistance. The Affirm 800 as described is an imaging module, not an AI-assisted diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Not applicable in the context of an "algorithm only" device. The Affirm 800 is an imaging system designed for intra-operative visualization with a human surgeon-in-the-loop. Its performance is evaluated on its ability to acquire and display clear fluorescent images, not on an algorithm making a diagnostic decision by itself. However, the technical aspects of image quality (e.g., sensitivity, SNR, depth penetration) could be considered standalone measurements of the imaging capability.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Ground Truth: For the animal study, the "ground truth" for image quality was based on the subjective evaluation by surgeons on a defined scale (1 to 5). For the technical image quality tests, the "acceptance criteria" likely served as the ground truth, comparing measured values (e.g., Weber contrast, SNR) against pre-defined thresholds or performance of the predicate device. There is no mention of pathology or outcomes data as ground truth for this device's performance.
-
The sample size for the training set:
- Training Set Sample Size: Not explicitly stated. The document describes verification and validation testing, but it does not detail a separate "training set" for a machine learning model, as the device is specified as an angiographic X-ray system module, not an AI algorithm for diagnosis.
-
How the ground truth for the training set was established:
- Ground Truth for Training Set: Not explicitly stated or applicable, as a discrete "training set" for an AI model (with associated ground truth) is not described in the context of this device's submission summary. The device's validation focuses on engineering specifications and direct performance measures rather than training an AI.
FDA 510(k) Clearance Letter - Affirm 800
Page 1
June 27, 2025
Digital Surgery Systems, Inc. (d.b.a True Digital Surgery)
Lu Ju
Regulatory Affairs Specialist
125 Cremona Dr Suite 110
Goleta, California 93117
Re: K243077
Trade/Device Name: Affirm 800
Regulation Number: 21 CFR 892.1600
Regulation Name: Angiographic X-Ray System
Regulatory Class: Class II
Product Code: IZI
Dated: May 27, 2025
Received: September 30, 2024
Dear Lu Ju:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Page 2
K243077 - Lu Ju Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K243077 - Lu Ju Page 3
Sincerely,
Adam D. Pierce -S [Digitally signed by Adam D. Pierce -S Date: 2025.06.27 12:14:55 -04'00']
Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K243077
Device Name
Affirm 800
Indications for Use (Describe)
The Affirm 800 is used in viewing intra-operative blood flow in the cerebral vascular area including, but not limited to, assessing cerebral vessel branch occlusion, as well as intraoperative blood flow and vessel patency in bypass surgical procedures in neurosurgery.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
K243077 Page 1 of 6
510(k) SUMMARY
1. SUBMITTER
OWNER/SUBMITTER: True Digital Surgery
125 Cremona Drive, Suite #110
Goleta, CA 93117
PRIMARY CONTACT: Lu Ju
805-978-5400 ext. 181 (phone)
lju@truedigitalsurgery.com
SECONDARY CONTACT: Deanna Najman
dnajman@truedigitalsurgery.com
DATE: June 25th, 2025
2. DEVICE
DEVICE TRADE NAME: Affirm 800
DEVICE CLASS: Class II
CLASSIFICATION NAME: System, X-Ray, Angiographic
REGULATION NUMBER: 892.1600
PRODUCT CODE: IZI
3. PREDICATE DEVICE
PREDICATE DEVICE: DIR 800 (K202391) from Aesculap
DEVICE CLASS: Class II
CLASSIFICATION NAME: System, X-Ray, Angiographic
REGULATION NUMBER: 892.1600
PRODUCT CODE: IZI
Page 6
K243077 Page 2 of 6
4. DEVICE DESCRIPTION
The Affirm 800 module is designed to work in conjunction with the Class I Digital Surgical Microscope (DSM) RE3. The Affirm 800 module, a Class II device, comprises hardware and software components that enable the Digital Surgical Microscope to emit excitation light in specific wavelengths to activate the fluorescence properties of Indocyanine Green (ICG). The fluorescence signal emitted by the patient represents the distribution of the infrared dye in the patient's blood vessels during surgery. The emitted light is then captured by the optics of the digital microscope, passed through filters to remove unwanted wavelengths of light, and finally detected by the image sensors. This detected signal is then projected on a 3D monitor, which is part of the microscope system, enabling the surgeon to view the magnified image.
The integrated Affirm 800 fluorescence module is used to visualize infrared fluorescent areas in the surgical field and includes features to record and play back a video clip of the area of focus where fluorescent light is emitted. The module has been designed for excitation in the wavelength range from 740 nm to 800 nm and for fluorescence visualization in the wavelength range from 820 nm to 900 nm. The fluorescence feature generates an image in the infrared spectrum, which means it cannot be seen by the naked eye.
5. INDICATIONS FOR USE
The Affirm 800 is used in viewing intra-operative blood flow in the cerebral vascular area including, but not limited to, assessing cerebral vessel branch occlusion, as well as intraoperative blood flow and vessel patency in bypass surgical procedures in neurosurgery.
6. SUMMARY/COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The Affirm 800 and its predicate device have the same intended use, and same design principles, identical materials, and similar technological characteristics. The technology used in the subject and predicate devices to illuminate the area of interest, optically focus the emitted light, filter unwanted wavelengths of light, and capture the light on the image sensor are identical. The software used to display the image on screen, as well as record and play back video is likewise similar.
The Affirm 800 and predicate device each utilize a surgical stereo microscope imaging system and have identical functions for viewing, recording, and replaying fluorescent images. Identical cameras are utilized as sensors and the infrared options are equivalent. Each device utilizes the same type of fluorescent agent, Indocyanine Green (ICG), to visualize blood flow. Using near infrared excitation, real-time visualization of images takes place in identical transmission ranges.
Page 7
K243077 Page 3 of 6
| Subject Device - True Digital Surgery Affirm 800 | Predicate Device - Aesculap DIR 800 | Substantial Equivalence Discussion |
|---|---|---|
| Submission# K243077 | K202391 | |
| Product Code IZI | IZI | Identical |
| Regulation number 892.1600 | 892.1600 | Identical |
| Class No. II | II | Identical |
| Indications for Use The Affirm 800 is used in viewing intra-operative blood flow in the cerebral vascular area including, but not limited to, assessing cerebral vessel branch occlusion, as well as intraoperative blood flow and vessel patency in bypass surgical procedures in neurosurgery. | The DIR 800 is an accessory for the Aesculap Aeos and is used in viewing intra-operative blood flow in the cerebral vascular area including, but not limited to, assessing cerebral aneurysm and vessel branch occlusion, as well as patency in neurosurgery. It also aids in the visual assessment of intraoperative blood flow and vessel patency in bypass surgical procedures in neurosurgery. | Similar |
| Basic System Function View, record, and replay fluorescent images of blood vessels in the cerebral area, view blood flow and vessel patency in bypass surgical procedures during neurosurgery. | View, record, and replay fluorescent images of blood vessels in the cerebral area, and view blood flow and vessel patency in bypass surgical procedures during neurosurgery. | Identical |
| Fluorescent Agent Indocyanine Green (ICG) | Indocyanine Green (ICG) | Identical |
| Visualization of Real-time Images Yes | Yes | Identical |
| Display Video and images are presented on 3D monitor. Picture-in-picture available. | Video and images are presented on 3D monitor. Picture-in-picture available. | Same |
| Light Source 2x white light LEDs and 13x IR LEDs | 2x white light LEDs and 13x IR LEDs | Identical |
| White Light Application 400 nm – 700 nm | 400 nm – 700 nm | Identical |
Page 8
K243077 Page 4 of 6
| Subject Device - True Digital Surgery Affirm 800 | Predicate Device - Aesculap DIR 800 | Substantial Equivalence Discussion |
|---|---|---|
| ICG Excitation 740 nm – 800 nm bandpass filter | 740 nm – 800 nm bandpass filter | Identical |
| ICG Detection Filter with high transmission from 820 nm – 900 nm | Filter with high transmission from 820 nm – 900 nm | Identical |
| Test Card Pre-operative check test card | Pre-operative check test card | Same |
| Distance of Imaging Head to Patient 200 mm – 300 mm | 200 mm – 300 mm | Same |
| Camera adaptation Integrated into the microscope head | Integrated into the microscope head | Same |
| Zoom Motorized 8.9:1 | Motorized 8.9:1 | Same |
| Auto detection of fluorescence influx Yes | Yes | Same |
| Autofocus Yes | Yes | Same |
| Color map Yes | No | Different |
Page 9
K243077 Page 5 of 6
7. SUMMARY OF PERFORMANCE TESTING
Design verification testing was performed to ensure that the subject device meets functional requirements, including design specifications, and to demonstrate that its performance is substantially equivalent to the predicate.
| Test | Test Method | Results |
|---|---|---|
| Image Quality (3D display and 3D glasses) | Characterize and confirm equivalent performance of the subject device compared to the predicate for the following:1. Sensitivity to NIR fluorescence in terms of ICG Molar concentration:- Weber contrast- Signal to Noise Ratio- Fluorescence Pixel Intensity for each concentration- Limit of detection- Limit of quantification- Signal linearity between limit of quantification and at least 3.2 micromolar concentration.2. Spatial and Temporal Noise- Spatial noise as measured from the image pipeline- Temporal noise measured using 30 black frames3. Left and Right Eye Comparison4. Sensitivity to NIR fluorescence in terms of depth penetration- Weber Contrast- Signal-to-Noise Ratio5. Stereoscopic Crosstalk- Stereoscopic Extinction Ratio- Stereoscopic Crosstalk | Pass |
Page 10
K243077 Page 6 of 6
| Test | Test Method | Results |
|---|---|---|
| 6. An animal study was conducted using 5 porcine models with a clinically relevant ICG dose. Surgeons evaluated image quality on a scale of 1 to 5, where 1 represented poor image quality and 5 represented very good quality, in visualizing vessels of various sizes. The study assessed image quality using a 3D display (part of the Class I surgical microscope) and 3D glasses. |
Software Verification and Validation Testing
Software verification and validation (V&V) testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions." The software for this device was considered as an enhanced documentation level, since in fluorescing modes, a failure or flaw of the imaging subsystem software could indirectly present a hazardous situation with a risk of death or serious injury to a patient prior to the implementation of risk control measures.
8. CONCLUSION
The indications for use of the subject device, Affirm 800, are equivalent to the indications for use of the predicate device, DIR 800. The technological characteristics and risk profile of the subject device are similar to those of the predicate device. Based on the similarities of the indications for use, technological, and design verification test results, the subject device is determined to be substantially equivalent to Aesculap's DIR 800 predicate device.
§ 892.1600 Angiographic x-ray system.
(a)
Identification. An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.