K Number
K243063
Date Cleared
2025-02-10

(136 days)

Product Code
Regulation Number
882.5890
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is indicated for aesthetic use, including facial and neck stimulation or body skin stimulation.

Device Description

The subject device, the MyoLight Microcurrent Handpieces are intended to be used together with a main control unit that is equipped with a color touch screen that significantly facilitates the use of the device. There are two MyoLight Microcurrent Handpieces that connect to the main unit. The small handpiece (diameter: 3.9 cm, Length: 14 cm) is generally used on the face and the large handpiece (diameter: 6.2 cm, Length: 15 cm) is generally used on the body. The handpieces are different in dimensions but not function. The handpieces contain the electrodes that deliver microcurrent for aesthetic use. The subject device is intended to be used together with a main control unit that is equipped with a color touch screen that significantly facilitates the use of the handpieces. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about therapy type, remaining therapy time and main therapy parameters on the screen.

AI/ML Overview

The provided context does not offer information about the acceptance criteria or a study proving that the device meets specific acceptance criteria in the typical format of a clinical or performance study with metrics like sensitivity, specificity, or AUC.

The document is a 510(k) summary for a medical device (MyoLight Microcurrent Handpiece) and focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed performance study with acceptance criteria.

However, based on the non-clinical testing section, we can infer some "acceptance criteria" through compliance with recognized standards.

Here's the breakdown of the information that can be extracted, and where the requested information is not available:

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) for substantial equivalence, the "acceptance criteria" are implicitly met by demonstrating compliance with recognized standards and similarity to predicate devices. The "reported device performance" is essentially the device's conformance to these standards and its specifications aligning with the predicate devices.

Acceptance Criterion (Inferred from Standards & Equivalence)Reported Device Performance (MyoLight Microcurrent Handpiece)
Electrical Safety and Essential Performance (IEC 60601-1:2005/A1:2012)Compliant with IEC 60601-1 leakage current requirements and isolation requirements.
Electromagnetic Compatibility (EMC) (IEC 60601-1-2:2014/EN 60601-1-2:2015)Compliant with IEC 60601-1-2 (implied by listing the standard as tested).
Particular Requirements for Nerve and Muscle Stimulators (IEC 60601-2-10 Edition 2.1 2016-04)Compliant with IEC 60601-2-10 (implied by listing the standard as tested).
Biocompatibility (ISO 10993 and FDA Guidance)Biocompatibility tests performed (exact results not detailed, but implied to be acceptable).
Software Validation & VerificationSoftware V&V tests performed (exact results not detailed, but implied to be acceptable).
Bench Testing for PerformanceBench testing performed to verify performance (exact details not provided, but used to support substantial equivalence).
Specifications within Acceptable Range of Predicate Devices (e.g., Max output voltage, current, power density, net charge, phase charge, pulse duration, frequency)Max output voltage (30V @ 500 Ω, 33V @ 1,000 Ω, 35V @ 10,000 Ω), Max output current (60 mA @ 500 Ω, 33 mA @ 1000 Ω, 3.5 mA @ 10,000 Ω), Max current density (82 µA/mm² @ 500 Ω), Max average power density (0.012W/cm² @ 500 Ω), Net Charge (0 µC), Max phase charge (23 µC), Duration of primary (128, 256, 384 µs), Pulse Duration (128, 256, 384 µs), Frequency (1 ~ 63 Hz). These are compared to predicate devices and deemed "substantially equivalent."
Safety - Automatic Shut OffYES
Safety - Patient Contact MaterialStainless Steel 304, PMMA (biocompatibility validated).

2. Sample size used for the test set and the data provenance
The document does not describe a clinical test set in terms of patient samples. The testing described is non-clinical (bench testing, software V&V, and compliance with electrical safety and EMC standards). Therefore, there is no information on sample size or data provenance in the context of a clinical test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as no clinical test set using expert-established ground truth is described.

4. Adjudication method for the test set
Not applicable, as no clinical test set with adjudication is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a microcurrent handpiece, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable, as the device is not an algorithm for diagnostic or prognostic purposes. It is a physical device delivering electrical stimulation, with software controls for operation.

7. The type of ground truth used
For non-clinical testing:

  • Electrical Safety, EMC, and Nerve/Muscle Stimulator Standards: Compliance is based on objective measurements against defined limits in the respective IEC standards.
  • Biocompatibility: Likely based on laboratory assays and tests according to ISO 10993 standards.
  • Software Validation & Verification: Based on predefined software requirements specification and test protocols.
  • Bench Testing: Based on objective measurements of physical and electrical parameters against design specifications.

8. The sample size for the training set
Not applicable, as this is not an AI/machine learning device. The "training set" concept doesn't apply.

9. How the ground truth for the training set was established
Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 10, 2025

Raja Trading Company, Inc. % Yolando Smith Smith Assoicates 1469 Harwell Ave Crofton, Maryland 21114

Re: K243063

Trade/Device Name: Myolight Microcurrent Handpiece Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: January 15, 2025 Received: January 15, 2025

Dear Yolando Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Heather L. Dean -S

Heather Dean, PhD

Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243063

Device Name

Myolight Microcurrent Handpiece

Indications for Use (Describe)

The device is indicated for aesthetic use, including facial and neck stimulation or body skin stimulation.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of 21CFR Section 807.92.

510(k) Number: K243063

1. Date of Preparation

2/7/2025

2.

Name:RAJA Trading Company Inc.
Address:2801 Juniper Street, Suite 2, Fairfax, VA 22031
Contact Person:Robert J. Adipietro
Title:Vice President
Telephone:561-868-4600
Fax:561-258-0207
Email:rja@rajamedical.com

3. Identification of the Proposed Device

Trade Name:MyoLight Microcurrent Handpiece
Common Name:Microcurrent Stimulator
Classification Name:Stimulator, Transcutaneous Electrical, Aesthetic Purpose
Classification:Class II
Product Code:NFO
Regulation Number:21 CFR 882.5890
Review Panel:Neurology
AttachmentRegulation NameRegulation No.ProductcodeClass
MyoLight MicrocurrentHandpiecesStimulator, TranscutaneousElectrical, Aesthetic Purposes882.5890NFO2

Identification of Predicate Devices 4.

PrimaryPredicate DeviceManufacturerTrade NameProduct Code510KNumber
Aesthetic TENSDeviceBTL Industries, Inc.BTL-785BNF HandpieceNFOK232172
Reference PredicateDevice #1ManufacturerTrade Name510KNumber
Aesthetic TENSDeviceSalton, Inc.Rejuvenique Facial ToningSystemNFOK011935
Reference PredicateDevice #2ManufacturerTrade Name510KNumber
Aesthetic TENSDeviceAvazzia, Inc.EZZI-LIFTNFOK191951

5. Product Description

The subject device, the MyoLight Microcurrent Handpieces are intended to be used together with a main control unit that is equipped with a color touch screen that significantly facilitates the use of the device. There are two MyoLight Microcurrent Handpieces that connect to the main unit. The small handpiece (diameter: 3.9 cm, Length: 14 cm) is generally used on the face and the large handpiece (diameter: 6.2 cm, Length: 15 cm) is generally used on the body. The handpieces are different in dimensions but not function. The handpieces contain the electrodes that deliver microcurrent for aesthetic use.

{5}------------------------------------------------

The subject device is intended to be used together with a main control unit that is equipped with a color touch screen that significantly facilitates the use of the handpieces. The on-screen information guides the user step-bystep through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about therapy type, remaining therapy time and main therapy parameters on the screen.

The product is only sold to professionals licensed to administer the treatments including aestheticians, med spas, plastic surgeons and dermatologists and other licensed professionals. It is not intended to be sold to the consumer or for home use.

6. Indications for Use

The device is indicated for aesthetic use, including facial and neck or body skin stimulation.

SpecificationsSubject DevicePrimaryPredicate DeviceReferencePredicateDevice # 1ReferencePredicateDevice # 2
510(k) NumberK243063K232172K011935K191951
ManufacturerRAJA TradingCompanyBTL Industries, IncSalton, Inc.Avazzia, Inc.
Device NameMyoLightMicrocurrentHandpieceBTL-785BNFHandpieceRejuveniqueFacial ToningSystemEZZI-LIFT
Classification &Subsequent ProductCodesNFONFONFONFO
Regulation Number882.5890882.5890882.5890882.5890
Classification Name:TranscutaneousElectrical NeverStimulator for PainReliefTranscutaneousElectrical NeverStimulator for PainReliefTranscutaneousElectrical NeverStimulator forPain ReliefTranscutaneousElectrical NeverStimulator forPain Relief
Common NameEsthetic TENS DeviceEsthetic TENSDeviceEsthetic TENSDeviceEsthetic TENSDevice
Indications for useThe device is indicatedfor aesthetic useincluding facial andneck or body skinstimulation.The device isindicated foraesthetic useincluding facial andneck or body skinstimulation.Indicates forCosmetic UseThe AvazziaOTC TENS foraesthetics,model BEST-AVI™ : EZZI-LIFTTM Deviceis indicated forover-the-counteraesthetic useincluding facialand neckstimulation orbody skinstimulation.
SpecificationsSubject DevicePrimaryPredicate DeviceReferencePredicateDevice # 1ReferencePredicateDevice # 2
Anatomic SitesFace, Neck, and BodyFace, Neck, andBodyFaceFace, Neck, andBody
Principal of OperationHandpieces withelectrodes that deliverelectrical stimulationattached to a mainbody that contains thepower supplies, themicroprocessor controlunit, and the touch-screen user interface.Handpieces withelectrodes thatdeliver electricalstimulation attachedto a main body thatcontains the powersupplies, themicroprocessorcontrol unit, and thetouch-screen userinterface.Face Mask withelectrodes that fitover users facethat deliverselectricalstimulationconnected to acontrol unit.Handheldportable devicewith electrodes,batteries anduser controls.
User InterfaceLCD Touch Screen onmain Control Unit withPower Source andMicroprocessorcontrolledLCD Touch Screenon main ControlUnit with PowerSource andMicroprocessorcontrolledControl Unitwith PowerSource andMicroprocessorcontrolled.LEDs andswitches
Compliance withVoluntary StandardsIEC 60601-1IEC 60601-1-2IEC 60601-2-10ISO 14971IEC 62366IEC 60601-1IEC 60601-1-2IEC 60601-2-10ISO 14971IEC 62366Not AvailableIEC 60601-1IEC 60601-1-2IEC 60601-2-10ISO 14971IEC 62366
Compliance with 21CFR 898YESYESYESYES
Power SourcesMains Power 110 -120, 220 - 240 VAC,50- 60 HzMains Power100-240V, 50-60Hz9V batteryTwo 1.5 V AAbatteries
Patient leakagecurrent: NormalConditionCompliant with IEC60601-1 leakagecurrent requirementsCompliant with IEC60601-1 leakagecurrentrequirementsNot AvailableCompliant withIEC 60601-1leakage currentrequirements
Method of ChannelIsolationCompliant with IEC60601-1 isolationrequirementsCompliant with IEC60601-1 isolationrequirementsElectrode Groupselected byrelaysCompliant withIEC 60601-1isolationrequirements
Synchronous orAlternating OutputBothBothAlternatingNot Available
Software / Firmware/ Micro-ProcessorControl?YesYesYesYes
Number of OutputModes1114
SpecificationsSubject DevicePrimaryPredicate DeviceReferencePredicateDevice # 1ReferencePredicateDevice # 2
Regulated Current orRegulated VoltageRegulated VoltageRegulated VoltageRegulatedVoltageRegulatedVoltage
Housing Materialsand ConstructionsMolded PC-ABSplastic material withscrew assemblyconstruction.PCBs and leadsinside plastic, orfoam case housing.ABS Plastic,latch assemblyPCBs insideplastic casehousing
Energy TypeElectrical StimulationElectricalStimulationElectricalStimulationElectricalStimulation
Voltage Current LevelYes - UncalibratedKnobYesYes -UncalibratedKnobNot Available
HandpieceDimensionsHandpiece #1: 62mm x147 mmHandpiece #2: 39mm x139.5 mm38.98" x 53.94" x34.65"4.5" X 3" X1.25"2.6" X 4.7" X1.35"
Handpiece WeightHandpiece #1: 1100 gHandpiece #2: 720 g2500 g80 grams5.4 oz.
RecommendedTreatment Time/Timer range20 MinutesRecommendedTreatment Time20 Minutes Timerrange20/16 minutesTimer RangeNot Available
Output Modes1114
Automatic Shut OffYESUnknownYESYES
Patient ContactYESYESYESYES
Contact MaterialStainless Steel 304PMMAConductive Ink(Silver)PCB Mask withGold PlatedBrass ElectrodesStainless Steel316
Waveform ShapePulsed Biphasicrectangular modulatedby trapezoidalPulsed Biphasicrectangularmodulated bytrapezoidalPulsed BiphasicrectangularPositive squarewave followedby a dampedsinusoidalwaveform ofvariabledurationdepending ondamping andbody loading
Max output voltage[V] (±20%) at500 Ω2,000 Ω10,000 Ω30V @ 500 Ω33V @ 1,000 ΩBTL-785-2, -8, -9Max. 48,5V @ 10 -373 Ω48,5V @ 500 Ω48,5V @ 2,000 Ω0 @ 10,000 Ω18V @ 500 Ω24V @ 1,000 Ω42V @ 500 Ω
SpecificationsSubject DevicePrimaryPredicate DeviceReferencePredicateDevice # 1ReferencePredicateDevice # 2
35V @ 10,000 ΩBTL-785-1, -7Max. 48,5V @ 10 -757 Ω48,5V @ 500 Ω48,5V @ 2,000 Ω0 @ 10,000 Ω28V @ 10,000 Ω122V @ 2,000 Ω348V @ 10,000 Ω
Max output current(+/-20%) a500 Ω2,000 Ω10,000 Ω60 mA @ 500 Ω33 mA @ 1000 Ω3.5 mA @ 10,000 ΩBTL-785-2, -8, -9Max. 130 mA @ 10- 373 Ohm97 mA @ 500 Ω24 mA @ 2000 Ω0 mA @ 10000 ΩBTL-785-1, -7Max. 64 mA @ 10 -757 Ω64 mA @ 500 Ω24 mA @ 2000 Ω0 mA @ 10000 Ω37.6 mA @ 500 Ω12.4 mA @ 1000 Ω2.8 mA @10,000 ΩMax. 0.5 mA0.363 mA @500 Ω0.117 mA @2000 Ω0.038 mA @10000 Ω
Max current density at500 Ω [μA/mm²]82 µA/mm²BTL-785-2 - 243μA/mm²BTL-785-8 - 335μA/mm²BTL-785-9 - 277μA/mm²BTL-785-1 - 168μA/mm²BTL-785-7 - 337μA/mm²464 μA/ mm²Built-in, YBrush: 8μA/mm²Pencil: 190μA/mm²
Max average powerdensity at 500 Ω[W/cm²]0.012W/ cm²BTL-785-2 - 0.047W/cm²BTL-785-8 - 0.065W/cm²BTL-785-9 - 0.054W/cm²BTL-785-1 - 0.033W/cm²BTL-785-7 - 0.065W/cm²0.0023 W/ cm²Built-in, YBrush: 0.0005W/cm²Pencil: 0.0035W/cm²
Net Charge per pulse0 µC (Pulse is biphasicsymmetrical)0 µC (Pulse isbiphasicsymmetrical)0 µC4 µC
Max phase charge perpulse at 500 Ω [μC]23 µCBTL-785-2, -8, -915 µC11.3 µC10 µC
SpecificationsSubject DevicePrimaryPredicate DeviceReferencePredicateDevice # 1ReferencePredicateDevice # 2
BTL-785-1, -710 μC
Duration of primary[μs](depolarizing phase)128 μs, 256 μs, 384 μs80 μsNot Available500 μs
Pulse Duration [μs]128 μs, 256 μs, 384 μs160 μs300 μs1100 μs
Frequency [Hz]1 ~ 63 Hz250 Hz8 Hz15-121 Hz

7. Substantially Equivalent (SE) Comparison

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Similarities

The subject device is similar to the predicate product in:

  • same classification information,
  • same indications and intended use,
  • same principle of operation
  • same performance effectiveness and performance safety as the primary predicate device. ●
  • same waveform shape
  • . same maximum average power density
  • substantially equivalent maximum output voltage
  • substantially equivalent output current .

Differences

The differences do not affect the performance and effectiveness of the subject device. Differences in Maximum Voltage, Maximum Current, Frequency, Pulse Duration, Phase Charge, and Current Density do not affect the performance and safety of the subject device.

Therefore, the MyoLight Microcurrent Handpiece device is substantially equivalent to the predicate device.

8. Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The following tests were conducted:

  • IEC 60601-1:2005/A1:2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.

  • IEC 60601-1-2:2014/EN 60601-1-2:2015, Medical electrical equipment- Part 1-2: General requirements for basic A safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests.
  • IEC 60601-2-10 Edition 2.1 2016-04 Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.

  • Biocompatibility Tests per ISO 10993 and FDA Guidance.

  • Software Validation & Verification Test.

  • Bench Testing to verify the performance.

9. Clinical Testing

Not Applicable

10. Conclusion

Based on the comparison and analysis above, the proposed subject device is Substantially Equivalent (SE) to the predicate device.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).