K Number
K243060
Date Cleared
2025-01-30

(125 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TeleRPM Gen2 Blood Glucose Monitoring System is comprised of the TeleRPM Gen2 Blood Glucose Meter and the TeleRPM Blood Glucose Test Strips. TeleRPM Gen2 Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Device Description

TeleRPM Gen2 Blood Glucose Monitoring System consists of TeleRPM Gen2 Blood Glucose Meter and the TeleRPM Blood Glucose Test Strips.

TeleRPM Control Solution, TeleRPM Lancing Device, TeleRPM Lancets are required for use but not included in meter box or test strips box and should be purchased separately. The TeleRPM Control Solution is for use with the above meter and test strip as a quality control check to verify that the meter and test strip are working together properly, and that the test is performing correctly. TeleRPM Lancing Device and TeleRPM Lancets are used for puncturing fingertip and then user can perform qlucose test with blood sample.

TeleRPM Gen2 Blood Glucose Monitoring System is designed to quantitatively measure the glucose concentration in fresh capillary whole blood from the fingertip. The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose oxidase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration value is reported in plasma equivalents and is displayed on meter screen.

AI/ML Overview

The provided text primarily focuses on the substantial equivalence determination for the TeleRPM Gen2 Blood Glucose Monitoring System to a predicate device. While it mentions the general types of studies conducted (robustness, precision, linearity, user evaluation, interference, stability, flex studies, software, cybersecurity controls, and a clinical usability study), it does not provide detailed acceptance criteria or numerical performance data as requested for several of your points.

Based on the information available:

1. A table of acceptance criteria and the reported device performance:

The document broadly states that the device "met the FDA SMBG OTC Guidance and industry standards" and that "these devices performed as intended and met associated guidance documents and industry standards." Specific numerical acceptance criteria and reported device performance for each study (precision, linearity, interference, etc.) are not detailed in the provided summary. For the user evaluation (clinical study), it states "the clinical performance met the FDA SMBG OTC Guidance."

Acceptance Criteria (General)Reported Device Performance (General)
Met FDA SMBG OTC Guidance and Industry StandardsPerformed as intended, met FDA SMBG OTC Guidance and industry standards.
User evaluation criteria metInexperienced lay persons able to obtain blood glucose readings, understand labeling, use system, interpret results and error messages. No adverse effects or complications.

2. Sample size used for the test set and the data provenance:

  • User Evaluation (Clinical Study): The document mentions "All participants" were able to understand the labeling, use the system, and interpret results. However, the exact sample size for the clinical usability evaluation is not specified.
  • Provenance: Not explicitly stated, but the company is located in Zhongshan, Guangdong, China. The testing location isn't specified, but it's reasonable to infer the studies were conducted by or on behalf of the manufacturer, likely in China or a region where they operate.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not provided in the summary. For a blood glucose monitoring system, the "ground truth" for glucose levels would typically be established by a laboratory reference method, not by experts adjudicating results.
  • For the usability evaluation, the "ground truth" is about successful interaction with the device, which is assessed through user performance and observation, not expert consensus on a measurement.

4. Adjudication method for the test set:

  • This information is not provided. For analytical performance, laboratory reference methods are used, not typically expert adjudication. For usability, the success of user interaction is observed and recorded.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This is not applicable to this device. A "Blood Glucose Monitoring System" measures blood glucose; it is not an AI-assisted diagnostic imaging device that involves "human readers." Therefore, an MRMC study comparing human readers with and without AI assistance was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The device performs a direct measurement of blood glucose. Its core function is a "standalone" algorithmic interpretation of the electrochemical reaction to display a glucose reading. This is its fundamental operation. There isn't a separate "human-in-the-loop" component in the direct glucose measurement process that would necessitate a distinction here.

7. The type of ground truth used:

  • For analytical performance (precision, linearity, interference), the ground truth for blood glucose concentration would be established using a laboratory reference method (e.g., YSI analyzer). This is standard for blood glucose meter validation.
  • For the usability evaluation, the "ground truth" assesses whether users can successfully operate the device and interpret results, which is based on direct observation and participant feedback.

8. The sample size for the training set:

  • This device is a physical blood glucose meter and test strips relying on electrochemical principles, not a machine learning or AI model that requires a "training set" in the computational sense. Therefore, the concept of a training set sample size is not applicable here.

9. How the ground truth for the training set was established:

  • As above, the concept of a training set is not applicable.

{0}------------------------------------------------

January 30, 2025

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Guangdong Transtek Medical Electronics Co., Ltd. Mark Qian Quality Director Zone A, No. 105, Dongli Road, Torch Development District Zhongshan, Guangdong 528437 China

Re: K243060

Trade/Device Name: TeleRPM Gen2 Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: December 27, 2024 Received: December 27, 2024

Dear Mark Qian:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joshua Balsam -S

Joshua M. Balsam, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243060

Device Name

TeleRPM Gen2 Blood Glucose Monitoring System

Indications for Use (Describe)

TeleRPM Gen2 Blood Glucose Monitoring System is comprised of the TeleRPM Gen2 Blood Glucose Meter and the TeleRPM Blood Glucose Test Strips. TeleRPM Gen2 Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with 21 CFR 807.92.

The assigned 510(k) number is K243060

1. Submitter's Information:

1.1 Submitter's Name, Address and Telephone Number

Name: Guangdong Transtek Medical Electronics Co., Ltd.

Address: Zone A, No.105, Dongli Road, Torch Development District,

Zhongshan, 528437, Guangdong, China.

Telephone: + 86 0760 8828 2982

1.2 Contact Person:

Name: Mark Qian

Position: Quality Director

Email: mark.qian@vivachekbio.com

1.3 Date Updated for Summary

Jan 29, 2025

2. Devices Name

2.1 Proprietary Name:

TeleRPM Gen2 Blood Glucose Monitoring System

2.2 Common Name:

Glucose Test System

2.3 Classification Name:

Classification Name: Glucose Test System Regulation Number: 21 CFR 862.1345

{5}------------------------------------------------

Regulatory Class: Class II Product Code: NBW

2.4 Predicate (Legally Marketed) Device

Predicate Name: VivaChek™ Link Plus Blood Glucose Monitoring System. 510(k) Number: K233058. Applicant: VivaChek Biotech (Hangzhou) Co., Ltd. No reference device is used in this submission.

3. Device Description

TeleRPM Gen2 Blood Glucose Monitoring System consists of TeleRPM Gen2 Blood Glucose Meter and the TeleRPM Blood Glucose Test Strips.

TeleRPM Control Solution, TeleRPM Lancing Device, TeleRPM Lancets are required for use but not included in meter box or test strips box and should be purchased separately. The TeleRPM Control Solution is for use with the above meter and test strip as a quality control check to verify that the meter and test strip are working together properly, and that the test is performing correctly. TeleRPM Lancing Device and TeleRPM Lancets are used for puncturing fingertip and then user can perform qlucose test with blood sample.

TeleRPM Gen2 Blood Glucose Monitoring System is designed to quantitatively measure the glucose concentration in fresh capillary whole blood from the fingertip. The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose oxidase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration value is reported in plasma equivalents and is displayed on meter screen.

{6}------------------------------------------------

4. Intended Use:

TeleRPM Gen2 Blood Glucose Monitoring System is comprised of the TeleRPM Gen2 Blood Glucose Meter and the TeleRPM Blood Glucose Test Strips. TeleRPM Gen2 Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

5. Comparison of Technological Characteristics with Predicate Device:

CharacteristicsPredicate: VivaChek ™ Link PlusBlood Glucose Monitoring System(K233058)Candidate: TeleRPM Gen2Blood Glucose MonitoringSystem
Similarities
Intended UseIt is intended to quantitativelymeasure the glucose concentrationin fresh capillary whole bloodsamples drawn from the fingertips. Itis intended for use by persons withdiabetes at home as an aid tomonitor the effectiveness ofdiabetes control.Same
Operation PrincipleElectrochemical biosensorSame
Detection MethodAmperometricSame
Strip ChemicalCompositionGlucose oxidaseSame
Measurement ResultPlasma GlucoseSame
SampleFresh capillary whole bloodSame
Unit of Measuremg/dLSame
MeasurementRange20-600 mg/dLSame
Sample Volume0.8µLSame
  • 5.1 Comparison of TeleRPM Gen2 Blood Glucose Monitoring System with the predicate:

{7}------------------------------------------------

Applicant: Guangdong Transtek Medical Electronics Co., Ltd.

Address: Zone A, No. 105, Dongli Road, Torch Development District, Zhongshan, Guangdong 528437, China Contact: Mark Qian Email: mark.qian@vivachekbio.com

Test Time5 secondsSame
Operating RelativeHumidity10-90% (non-condensing)Same
OperatingTemperature41-113°FSame
Hematocrit Range20-70%Same
Automatic Shutoff2 minutes after last actionSame
Power SourceRechargeable 3.7 Volt DC nominal,lithium polymer battery (5V inputcharge voltage)Same
Data TransmissionThis meter features automatictransfer of glucose results to theCloud using 4G wireless mobilecommunicationSame
Differences
Battery Capacity800 mAh1,300 mAh
Memory500 records2,000 records
Dimensions100.9mm x 61.6mm x 23.7mm105.2mm x 60.2mm x17.5mm
WeightApproximately 85gApproximately 123g

6. Laboratory (Nonclinical) and Clinical Studies

6.1 Studies/Tests with Performance Data

Study and testing activities included cleaning and disinfection, robustness, precision, linearity, user evaluation, interference evaluation, test strips stability, flex studies. The studies with performance data were provided in the submission to support of the substantial equivalence determination.

TeleRPM Gen2 Blood Glucose Monitoring System contains 4G module, the device complies with software and cybersecurity control guidance requirements based on the test reports and study reports.

6.2 Brief Discussion of Laboratory (Nonclinical) Studies:

Laboratory (nonclinical) studies were performed on the TeleRPM Gen2 Blood Glucose Monitoring System in accordance with the FDA SMBG OTC Guidance, CLSI guidelines and industry standards. Test data and results from laboratory studies showed that these devices performed as intended and met associated guidance documents and industry standards.

{8}------------------------------------------------

6.3 Brief Discussion of Clinical Study including Usability Evaluation:

The TeleRPM Gen2 Blood Glucose Monitoring System has the same intended use, work principle and technological characteristics as VivaChek™ Link Plus Blood Glucose Monitoring System, clinical study (user evaluation) and usability evaluation were conducted with intended users using the TeleRPM Gen2 Blood Glucose Monitoring System. Studies result indicated that inexperienced lay persons were able to obtain blood glucose readings when using the TeleRPM Gen2 Blood Glucose Monitoring System, and the clinical study data showed that the clinical performance met the FDA SMBG OTC Guidance. All participants had the ability of understanding of the labeling, could use of the system and could interpret the testing results and error messages based on the usability evaluation reports. In addition, there was no adverse effect or complication occurred or identified during the clinical study including usability evaluation. Therefore, these studies support the intended use as described in the proposed labeling.

7. Conclusion:

The nonclinical studies and clinical study results demonstrate that the TeleRPM Gen2 Blood Glucose Monitoring System performed as intended and met FDA SMBG OTC Guidance and industry standards. TeleRPM Gen2 Blood Glucose Monitoring System is as safe, as effective and perform as well as the predicate device.

Therefore, the candidate device TeleRPM Gen2 Blood Glucose Monitoring System is substantially equivalent to the predicate device VivaChek Link Plus Blood Glucose Monitoring System (K233058).

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.