K Number
K242998

Validate with FDA (Live)

Date Cleared
2024-11-21

(56 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

GMReis Ankle Plating System are intended to be used for internal bone fixation for bone fractures, fusions, osteotomies and non-unions in the ankle.

Device Description

GMReis Ankle Plating System consists of plates and screws in a variety of designs and sizes that are commonly used in trauma and reconstructive surgery. The bone plates and screws are made from titanium alloy. The plates range in thickness from 1.8 to 3.6 mm, and the screws range in diameter from 2.7 to 3.5 mm. They are available on different sizes and shapes, according to the implantation site and the extension of the fracture. GMReis Ankle Plating System are for single use. The devices are provided non-sterile and must being properly cleaned and sterilized before use, according to the recommendations provided in the Instructions for Use.

AI/ML Overview

This document, K242998, describes a 510(k) premarket notification for the "GMReis Ankle Plating System." This is a medical device, specifically bone plates and screws, and the submission aims to demonstrate its substantial equivalence to previously marketed predicate devices.

The request asks for information about the acceptance criteria and the study that proves the device meets them, specifically in the context of an AI/software-driven medical device. However, the provided document does NOT contain information about an AI/software component, nor does it detail acceptance criteria related to a diagnostic or analytical performance study typical of AI/software devices.

Instead, this document focuses on establishing substantial equivalence for a physical orthopedic implant through:

  • Comparison of Technological Characteristics: Showing similarity in design, intended use, materials, manufacturing processes, and sterilization methods to predicate devices.
  • Performance Data (Mechanical Testing): Demonstrating the physical performance of the plates and screws according to well-established ASTM standards for metallic bone fixation devices.

Therefore, I cannot provide the requested information regarding AI/software specific acceptance criteria, test set sample sizes, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone AI performance, or training set details, as this information is not present in the provided document.

The "acceptance criteria" for this specific device (GMReis Ankle Plating System) are implicitly met by demonstrating that its mechanical performance is comparable to or better than the predicate devices, and that its materials and manufacturing processes are substantially equivalent.

Here's a breakdown of what can be extracted from the document, tailored to the spirit of your request but acknowledging the different nature of the device:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Implicit for a Bone Plate System)Reported Device Performance (as stated in Section VII)
Mechanical Performance:
- Plate performanceDemonstrated through mechanical testing according to ASTM F382.
- Screw performanceDemonstrated through mechanical testing according to ASTM F543.
Material/Biocompatibility:"substantially equivalent to the predicates with regard to materials and biocompatibility."
Intended Use:"same basic design and intended use" as predicate.
Manufacturing Process:"identical materials and identical manufacturing processes" as a previously cleared reference device (K182718).
Sterilization Method:"sterilized by the same methods" as predicate.

Note: The document states "Based on submitted testing data, the subject device is equivalent to the predicate devices." The specific numerical or statistical results of these mechanical tests that demonstrate equivalency are not provided in this summary, but the assertion is that the acceptance criteria for mechanical performance (i.e., meeting or exceeding predicate performance as per ASTM standards) were met.

2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • No information provided regarding sample sizes for human clinical data or diagnostic test sets, as this is a mechanical device primarily evaluated through engineering tests, not a diagnostic AI.
  • Data Provenance: The mechanical testing data would typically be generated in a lab setting, presumably by the manufacturer (GM Dos Reis Industria e Comercio Ltda) from Brazil, as they are the submitter. The document does not specify the country of origin of the test data directly, but it would be expected to come from their internal or contracted testing facilities.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not applicable. This pertains to establishing ground truth for diagnostic accuracy, which is not relevant for this physical device. Mechanical testing results are based on objective physical measurements against standards, not expert interpretation.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable. No expert adjudication is described or required for mechanical testing of this nature.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a bone plate system, not an AI diagnostic tool. No MRMC study was conducted.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable. This refers to AI algorithm performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable directly. For mechanical testing, the "ground truth" is defined by the objective performance requirements outlined in the ASTM standards (e.g., specific load-bearing capacities, fatigue life). Compliance with these standards serves as the "truth" for the device's physical performance.

8. The sample size for the training set

  • Not applicable. There is no "training set" as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

  • Not applicable. There is no "training set."

In summary, the provided FDA 510(k) clearance letter and summary for the GMReis Ankle Plating System clearly indicate that this is a physical implant device, and its safety and effectiveness (specifically substantial equivalence) are demonstrated through a comparison of physical characteristics and mechanical performance testing, not through AI/software validation studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 21, 2024

GM Dos Reis Industria e Comercio Ltda Guilherme Esteves Pontes Regulatory Affairs Analyst Avenida Pierre Simon de La Place, 600 Campinas, 13069320 Brazil

Re: K242998

Trade/Device Name: GMReis Ankle Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: September 26, 2024 Received: September 26, 2024

Dear Guilherme Esteves Pontes:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CHRISTOPHER FERREIRA -S

Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K242998

Device Name

GMReis Ankle Plating System

Indications for Use (Describe)

GMReis Ankle Plating System are intended to be used for internal bone fixation for bone fractures, fusions, osteotomies and non-unions in the ankle.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Section 5 - 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.

l. Submitter:

GM Dos Reis Industria e Comercio Ltda Avenida Pierre Simon de La Place 600 Campinas, São Paulo, Brazil 13069-320 Guilherme Esteves Pontes, Regulatory Affairs Analyst Phone: +55 (19) 3765-9900 Email: qualidade4@gmreis.com.br Date prepared: September 26, 2024

II. Device Name:

Trade Name:GMReis Ankle Plating System
Common Name:Plate, fixation, bone; Screw, fixation, bone
Classification Name:Single/Multiple Component Metallic Bone Fixation Appliances and Accessories
Device Class:II
Product Codes:HRS; HWC
Regulation Number:21 CFR 888.3030; 21 CFR 888.3040

III. Predicate Devices:

Legally marketed devices to which we are claiming "Substantial Equivalence" are the following: Fracture Plates (K123241) (Primary predicate device); Mini and Micro Fragments Reconstruction System - GMReis (K182718) (Reference device);

IV. Device Description:

GMReis Ankle Plating System consists of plates and screws in a variety of designs and sizes that are commonly used in trauma and reconstructive surgery. The bone plates and screws are made from titanium alloy. The plates range in thickness from 1.8 to 3.6 mm, and the screws range in diameter from 2.7 to 3.5 mm. They are available on different sizes and shapes, according to the implantation site and the extension of the fracture.

GM dos Reis Industria e Comercio Ltda. Pierre Simon de Laplace Ave., 600, Block 3F9677 Techno Park, Campinas, SP, Brazil, Zip Code 13069320 Phone: +551937659900, Email: gmreis@gmreis.com.br Website: www.gmreis.com.br

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo on the left and a table on the right. The logo is for GMREIS and includes a blue and red circular graphic with the text "GMREIS Qualidade para Vida" in blue. On the right side of the image, there is a table with the text "K242998 Page 2 of 2" at the top.

GMReis Ankle Plating System are for single use. The devices are provided non-sterile and must being properly cleaned and sterilized before use, according to the recommendations provided in the Instructions for Use.

V. Statement of Indications for Use of the Device:

GMReis Ankle Plating System are intended to be used for internal bone fractures, fusions, osteotomies and non-unions in the ankle.

VI. Comparison of Technological Characteristics with The Predicate Device:

The subject, the predicate and the reference devices have the same basic design and intended use. Both are packed using the same materials and are to be sterilized by the same methods. All of the subject device final finished components are manufactured in the same facilities using identical materials and identical manufacturing processes as used for the previously cleared reference device in K182718, and therefore are substantially equivalent to the predicates with reqard to materials and biocompatibility.

VII. Performance Data:

The performance of the subject devices are demonstrated through mechanical testing of plates and screws according to ASTM F382 and ASTM F543, respectively. Based on submitted testing data, the subject device is equivalent to the predicate devices.

VIII. Conclusions:

The GMReis Ankle Plating System has the basic design features and intended use as the predicate. Any differences between them are considered minor and do not raise questions concerning safety or effectiveness. Based on the information provided, GMReis has determined that the proposed device is substantially equivalent to the predicate device.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.