K Number
K242996
Device Name
EndyMed PRO MAX
Date Cleared
2025-06-10

(257 days)

Product Code
Regulation Number
878.4400
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The PRO MAX system is a Radiofrequency device intended for use in dermatologic procedures with separate indications for each handpiece. The TC Handpieces (Shaper Max, Mini Shaper Max, iFine Max) are intended for: • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulation When used up to 65W, the Mini Shaper Max and Small Max handpieces are also intended for the non-invasive treatment of mild to moderate facial wrinkles and rhytides. The FSR Max Handpiece is intended for dermatological procedures requiring ablation and resurfacing of the skin. The Intensif MAX Handpiece is intended for use in dermatologic procedures for electrocoagulation and hemostasis. The EndyMed Contour MAX Handpiece is intended for the treatment of the following medical conditions; using the Contour MAX applicator for delivery of non-thermal RF combined with massage: • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulation • Temporary reduction in the appearance of cellulite
Device Description
ENDYMED PRO™ MAX is a computerized system that generates pulses of radio frequency (RF) energy that are emitted onto patient skin. The device consists of a console and seven handpieces that emit either ablative or non-ablative radiofrequency.
More Information

Not Found

No.

The document explicitly states "Not Found" for "Mentions AI, DNN, or ML" and describes a computerized system that generates radio frequency energy without any mention of AI capabilities.

Yes
The device is described as "intended for use in dermatologic procedures" and various handpieces are indicated for "relief of minor muscle aches and pain, relief of muscle spasm," "temporary improvement of local blood circulation," and "non-invasive treatment of mild to moderate facial wrinkles and rhytides," which are all therapeutic or medical purposes.

No

Explanation: The device description and intended use indicate that the PRO MAX system is used for various dermatological procedures and therapeutic applications (e.g., pain relief, wrinkle treatment, skin resurfacing), all of which are treatment-oriented rather than diagnostic.

No

The device is described as a "computerized system that generates pulses of radio frequency (RF) energy that are emitted onto patient skin" and consists of "a console and seven handpieces that emit either ablative or non-ablative radiofrequency." This clearly indicates the presence of physical hardware components that generate and deliver energy, making it a hardware device with integrated software, not a software-only medical device. The mention of IEC 60601 standards further confirms its nature as electrical medical equipment.

No. The device is a radiofrequency device used for dermatologic procedures; it does not perform tests on biological samples to provide diagnostic information.

N/A

Intended Use / Indications for Use

The PRO MAX system is a Radiofrequency device intended for use in dermatologic procedures with separate indications for each handpiece.

The TC Handpieces (Shaper Max, Mini Shaper Max, iFine Max) are intended for:
• Relief of minor muscle aches and pain, relief of muscle spasm
• Temporary improvement of local blood circulation

When used up to 65W, the Mini Shaper Max and Small Max handpieces are also intended for the non-invasive treatment of mild to moderate facial wrinkles and rhytides.

The FSR Max Handpiece is intended for dermatological procedures requiring ablation and resurfacing of the skin.

The Intensif MAX Handpiece is intended for use in dermatologic procedures for electrocoagulation and hemostasis.

The EndyMed Contour MAX Handpiece is intended for the treatment of the following medical conditions; using the Contour MAX applicator for delivery of non-thermal RF combined with massage:
• Relief of minor muscle aches and pain, relief of muscle spasm
• Temporary improvement of local blood circulation
• Temporary reduction in the appearance of cellulite

Product codes

OUH, GEI, PBX

Device Description

ENDYMED PRO™ MAX is a computerized system that generates pulses of radio frequency (RF) energy that are emitted onto patient skin. The device consists of a console and seven handpieces that emit either ablative or non-ablative radiofrequency.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Bench testing was performed to verify the accuracy of the output of RF energy.

Performance Testing - Tissue Heating Testing: The Shaper Max, iFine Max, Mini Shaper Max, and Contour Max handpieces were all subjected to tissue heating testing to verify their safety and effectiveness for the requested pain relief indication. The Small Max Handpiece was not included in the study because it has not been modified since its original clearance. The testing was performed on 3 subjects with 3 anatomical sites each. The test was performed at the device's high settings and demonstrated that each handpiece can maintain a therapeutic temperature of 39-45°C for at least 10 minutes of treatment.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

Imagine TC Skin Therapy System (K083461), Intensif Applicator (K130501), EndyMed Contour Handpiece (K161199), EndyMed FSR Applicator (K101510)

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.

FDA 510(k) Clearance Letter - EndyMed PRO MAX

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 10, 2025

EndyMed Medical Ltd.
Ohad Fisher
Regulatory Director
12 Leshem Street
Caesarea, 3088900
Israel

Re: K242996
Trade/Device Name: EndyMed PRO MAX
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: OUH, GEI, PBX
Dated: May 6, 2025
Received: May 7, 2025

Dear Ohad Fisher:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K242996 - Ohad Fisher Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K242996 - Ohad Fisher Page 3

Sincerely,

James H. Jang -S [Digitally signed by James H. Jang -S Date: 2025.06.10 12:52:56 -04'00']

James Jang, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K242996
Device Name: EndyMed PRO MAX

Indications for Use (Describe)

The PRO MAX system is a Radiofrequency device intended for use in dermatologic procedures with separate indications for each handpiece.

The TC Handpieces (Shaper Max, Mini Shaper Max, iFine Max) are intended for:
• Relief of minor muscle aches and pain, relief of muscle spasm
• Temporary improvement of local blood circulation

When used up to 65W, the Mini Shaper Max and Small Max handpieces are also intended for the non-invasive treatment of mild to moderate facial wrinkles and rhytides.

The FSR Max Handpiece is intended for dermatological procedures requiring ablation and resurfacing of the skin.

The Intensif MAX Handpiece is intended for use in dermatologic procedures for electrocoagulation and hemostasis.

The EndyMed Contour MAX Handpiece is intended for the treatment of the following medical conditions; using the Contour MAX applicator for delivery of non-thermal RF combined with massage:
• Relief of minor muscle aches and pain, relief of muscle spasm
• Temporary improvement of local blood circulation
• Temporary reduction in the appearance of cellulite

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(K) Summary

EndyMed PRO MAX

K242996

Applicant: EndyMed Medical Ltd.
Address: 12 Leshem Street
Caesarea, Israel
3088900
Contact Person: Ohad Fisher, Regulatory Director
Contact Information: ohadf@endymed.com
Preparation Date: June 08, 2025
Device Trade Name: EndyMed PRO MAX
K242996
Classification Name: Electrosurgical Cutting and Coagulation Device and Accessories
Regulation Number: 878.4400
Product Code: GEI, OUH, PBX
Regulatory Class: II
Legally Marketed Predicate/Reference Devices:

  • Imagine TC Skin Therapy System (K083461)
  • Intensif Applicator (K130501)
  • EndyMed Contour Handpiece (K161199)
  • EndyMed FSR Applicator (K101510)

Device Description:

ENDYMED PRO™ MAX is a computerized system that generates pulses of radio frequency (RF) energy that are emitted onto patient skin. The device consists of a console and seven handpieces that emit either ablative or non-ablative radiofrequency.

Indications for use:

The PRO MAX system is a Radiofrequency device intended for use in dermatologic procedures with separate indications for each handpiece.

The TC Handpieces (Shaper Max, Mini Shaper Max, iFine Max) are intended for:
• Relief of minor muscle aches and pain, relief of muscle spasm
• Temporary improvement of local blood circulation

When used up to 65W, the Mini Shaper Max and Small Max handpieces are also intended for the non-invasive treatment of mild to moderate facial wrinkles and rhytides.

The FSR Max Handpiece is intended for dermatological procedures requiring ablation and resurfacing of the skin.

The Intensif MAX Handpiece is intended for use in dermatologic procedures for electrocoagulation and hemostasis.

Page 6

510(K) Summary

EndyMed PRO MAX

K242996

The EndyMed Contour MAX Handpiece is intended for the treatment of the following medical conditions; using the Contour MAX applicator for delivery of non-thermal RF combined with massage:
• Relief of minor muscle aches and pain, relief of muscle spasm
• Temporary improvement of local blood circulation
• Temporary reduction in the appearance of cellulite

Page 7

510(K) Summary

EndyMed PRO MAX

K242996

Substantial Equivalence—Indications for use and Technical Comparison

Predicate and Reference device table by Applicator

HandpieceStatusPredicate NamePrimary PredicateSecondary Predicate
Shaper MaxModifiedImagine TC Skin Therapy SystemK083461K161199- Contour Handpiece
Small MaxUnchangedImagine TC Skin Therapy SystemK083461K161199- Contour Handpiece
iFine MaxNewImagine TC Skin Therapy SystemK083461K161199- Contour Handpiece
Mini Shaper MaxNewImagine TC Skin Therapy SystemK083461K161199 - Contour Handpiece
Contour MaxModifiedEndymed Contour HandpieceK161199N/A
Intensif MaxUnchangedIntensif ApplicatorK130501N/A
FSR MaxUnchangedFSR ApplicatorK101510N/A

Indications for Use and Technical Specifications Comparison

EndyMed ProMax System Comparison

SpecificationSubject DevicePredicate: Imagine TC Skin Therapy SystemPredicate: Contour HandpieceComparison
Mode of operationBipolar radiofrequencyBipolar radiofrequencyBipolar radiofrequencysame
Input Power110-120VAC, 50/60Hz, 2.7A or 220-240VAC, 50/60Hz, 1.3A110-120VAC, 50/60Hz, 2.7A or 220-240VAC, 50/60Hz, 1.3A110-120VAC, 50/60Hz, 2.7A or 220-240VAC, 50/60Hz, 1.3ASame

Page 8

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject DevicePredicate: Imagine TC Skin Therapy SystemPredicate: Contour HandpieceComparison
Max output power100W65W85WThe new generation of the device has a higher max power output than previous generations, but treatment temperature remains the same.
Console Cooling SystemFans Cooling systemWater cooling systemWater cooling systemdifferent
Thermal Camera (optional)SupportedUnsupportedunsupporteddifferent

Intensif Max Handpiece Predicate Comparison

SpecificationSubject Device (Intensif Max)Predicate Device (Intesif handpiece, K130501)Comparison
Indications for useThe Intensif Max Handpiece is intended for use in dermatologic procedures for electrocoagulation and hemostasis.The Intensif Applicator is intended for use in Dermatologic and General Surgical procedures for electrocoagulation and hemostasis.The subject device removes the term "general surgical" to better align with the intended use of the device.
Mode of ActionRF MicroneedleRF microneedleSame
Frequency1MHz1MHzSame
ModePulsedPulsedSame
Number of needles2525Same
Needle typeHyper: Non-insulatedNon-insulatedSame
Needle Diameter300 microns300 micronssame
Max treatment depth3.5mm3.5mmSame
Max power output25W25WSame
Total RF EnergyUp to 62 mJ/pinUp to 62 mJ/pin
Pulse Duration20-170 msec20-170 msec

Page 9

510(K) Summary

EndyMed PRO MAX

K242996

Intensif Max Handpiece Predicate Comparison

SpecificationSubject Device (Intensif Max)Predicate Device (Intesif handpiece, K130501)Comparison
Indications for useThe Intensif Max Handpiece is intended for use in dermatologic procedures for electrocoagulation and hemostasis.The Intensif Applicator is intended for use in Dermatologic and General Surgical procedures for electrocoagulation and hemostasis.The subject device removes the term "general surgical" to better align with the intended use of the device.
Mode of ActionRF MicroneedleRF microneedleSame
Frequency1MHz1MHzSame
ModePulsedPulsedSame
Number of needles2525Same
Needle typeHyper: Non-insulatedNon-insulatedSame
Needle Diameter300 microns300 micronssame
Max treatment depth3.5mm3.5mmSame
Max power output25W25WSame
Total RF EnergyUp to 62 mJ/pinUp to 62 mJ/pin
Pulse Duration20-170 msec20-170 msec
Duty Cycle50%50%Same
Patient Contacting MaterialsBiocompatibleBiocompatibleSame

iFine Max Handpiece Predicate Comparison

SpecificationSubject DevicePrimary Predicate (Imagine TC)Secondary Predicate (Contour Handpiece)Comparison
Indications for useThe PRO MAX system is a Radiofrequency device intended for use in dermatologic procedures with separate indications for each handpiece. The iFine Max Handpiece is intended for • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulationThe Imagine TC Skin Treatment System is a noninvasive device intended for use in Dermatologic and General Surgical procedures for the noninvasive treatment of mild to moderate facial wrinkles and rhytides.The EndyMed Contour Max Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage: • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulation • Temporary reduction in the appearance of cellulite.The subject device excludes the term 'general surgical' to better align with the intended use of the device. The Contour Handpiece predicate is cleared for use with the TC Imagine system. The new and modified handpieces share the pain relief indication for use that the Contour Handpiece is cleared for. They do not, however, include the vacuum function, and therefore do not include the cellulite reduction indication.

Page 10

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject DevicePrimary Predicate (Imagine TC)Secondary Predicate (Contour Handpiece)Comparison
Mode of ActionNon-invasive RadiofrequencyNon-invasive RadiofrequencyNon-invasive Radiofrequency with vacuumSubject device does not include vacuum function
PolarityBipolarBipolarBipolarSame
ModeContinuousContinuousContinuousSame
Max output powerUp to 6WUp to 65WUp to 85WThe power output of the iFine handpiece is lower than the predicate device, but Tissue Heating Testing with FLIR camera demonstrates the handpiece can meet performance requirements for the pain relief indication.
Vacuum SpecificationsN/AN/AVacuum Pulse: Frequency: 2.5 Hz Vacuum Pressure: >250 mbar, 1-5 levelThe subject device does not include the vacuum function and therefore does not include the cellulite reduction indication.
Electrode Size69.6 mm2304.8 mm2 (Small HP)1381.2 mm2Different. Tissue Heating Testing with FLIR camera demonstrates the handpiece can meet performance requirements for the pain relief indication.
Power Density0.086 W/mm20.21 W/ mm20.06 W/ mm2Different. Tissue Heating Testing with FLIR camera demonstrates the handpiece can meet performance requirements for the pain relief indication.
Patient Contacting materialsBiocompatibleBiocompatibleBiocompatibleAll handpieces have been tested and determined to be biocompatible.

Page 11

510(K) Summary

EndyMed PRO MAX

K242996

Contour Max Handpiece Predicate Comparison

SpecificationSubject Device (Contour Max)Predicate Device (Contour Handpiece K161199)Comparison
Indications for UseThe EndyMed Contour Max Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage: • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulation • Temporary reduction in the appearance of cellulite.The EndyMed Contour Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage: • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulation • Temporary reduction in the appearance of cellulite.Same
Mode of ActionNon-invasive Radiofrequency with VacuumNon-invasive Radiofrequency with VacuumSame
PolarityBipolarBipolarSame
Frequency1Mhz1MHzSame
ModeContinuousContinuousSame
Max output powerUp to 100WUp to 85WThe max output power of the Contour MAX device has increased by 15% from the predicate Contour handpiece from 85W to 100W. Tissue Heating Testing with FLIR camera demonstrates the handpiece can meet performance requirements for the pain relief indication. See details on testing below.
Pulse Duration30 sec30 secSame

Page 12

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject Device (Contour Max)Predicate Device (Contour Handpiece K161199)Comparison
Electrode Size1381.2 mm21381.2 mm2Same
Power Density0.07 W/mm20.06 W/mm2Same
Vacuum Pulse frequency2.5Hz2.5HzSame
Vacuum Pressure>250 mbar, 1-5 level>250 mbar, 1-5 levelSame
Patient Contacting MaterialsBiocompatibleBiocompatibleAll handpieces have been tested and determined to be biocompatible.

Shaper Max Handpiece Predicate Comparison

SpecificationSubject DevicePrimary PredicateSecondary PredicateComparison
Indications for useThe Shaper MAX Handpiece is Indicated for • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulationThe Imagine TC Skin Treatment System is a noninvasive device intended for use in Dermatologic and General Surgical procedures for the noninvasive treatment of mild to moderate facial wrinkles and rhytides.The EndyMed Contour Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage: • Relief of minor muscle aches and pain, relief of muscle spasmThe subject device excludes the term 'general surgical' to better align with the intended use of the device. The Contour Handpiece predicate is cleared for use with the TC Imagine system. The new and modified handpieces share the pain relief indication for use that the Contour Handpiece is cleared for. They do not, however, include the vacuum function,

Page 13

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject DevicePrimary PredicateSecondary PredicateComparison
• Temporary improvement of local blood circulation Temporary reduction in the appearance of cellulite.and therefore do not include the cellulite reduction indication.
Mode of ActionNon-invasive RadiofrequencyNon-invasive RadiofrequencyNon-invasive Radiofrequency with vacuumSubject does not include vacuum function, and therefore does not include cellulite reduction indication
PolarityBipolarBipolarBipolarsame
Frequency1MHz1MHz1MHzSame
ModeContinuousContinuousContinuoussame
Max Output PowerUp to 100WUp to 65WUp to 85WCompared to the predicate Large HP, the electrode contact surface area ratio is 2.3 times bigger and the max output power is 1.5 times bigger. Because of this, the power density of the handpiece is the same as the predicate. Tissue Heating Testing with FLIR camera supports the addition of the pain relief indication that is also cleared for the TC Imagine System (Contour Handpiece). The specifications of the Shaper Max Handpiece are similar to the Contour handpiece. The output power is slightly lower, and the vacuum feature is not included in the Shaper Max handpiece.
Pulse Duration30 sec30 sec30 secSame.
Electrode Surface Area1971.1 mm2855.6 mm21381.2 mm2Different. Differences do not present additional risks to safety or efficacy.

Page 14

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject DevicePrimary PredicateSecondary PredicateComparison
Pulse Duration30 sec30 sec30 secSame. Different. Differences do not present additional risks to safety or efficacy. Performance testing demonstrates that the handpieces can perform equivalently.
Electrode Surface Area1971.1 mm2855.6 mm21381.2 mm2Tissue Heating Testing with FLIR camera supports the addition of the pain relief indication that is also cleared for the TC Imagine System (Contour Handpiece). The specifications of the Shaper Max Handpiece are similar to the Contour handpiece. The output power is slightly lower, and the vacuum feature is not included in the Shaper Max handpiece. Same. Different. Differences do not present additional risks to safety or efficacy.
Vacuum SpecificationsN/AN/AVacuum Pulse: Frequency: 2.5 Hz Vacuum Pressure: >250 mbar, 1-5 levelThe Vacuum function is not included with the shaper Max handpiece. The vacuum function is used for cellulite treatment and is not related to pain relief indication.
Power Density0.05 W/mm20.08 W/mm20.06 W/mm2Similar. Differences do not present additional risks to safety or efficacy.
Patient Contacting MaterialsBiocompatibleBiocompatibleBiocompatibleAll handpieces have been tested and determined to be biocompatible.

Mini-Shaper Max Handpiece Predicate Comparison

SpecificationSubject DevicePrimary Predicate – Imagine TC SMALL handpieceSecondary PredicateComparison
Indications for useThe Mini-Shaper MAX Handpiece is Indicated for • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulationThe Imagine TC Skin Treatment System is a noninvasive device intended for use in Dermatologic and General Surgical procedures for the noninvasive treatment of mild to moderate facial wrinkles and rhytides.The EndyMed Contour Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage:The subject device excludes the term 'general surgical' to better align with the intended use of the device. The new and modified handpieces include the wrinkles indication for use, but limit it to 65W power, the same that the predicate device is cleared for. The Contour Handpiece predicate is cleared for use with the TC Imagine

Page 15

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject DevicePrimary Predicate – Imagine TC SMALL handpieceSecondary PredicateComparison
Relief of minor muscle aches and pain, relief of muscle spasm Temporary improvement of local blood circulation Temporary reduction in the appearance of cellulite.system. The new and modified handpieces share the pain relief indication for use that the Contour Handpiece is cleared for. They do not, however, include the vacuum function, and therefore do not include the cellulite reduction indication.
When used up to 65W, the Mini-Shaper Max handpiece is intended for the non-invasive treatment of mild to moderate facial wrinkles and rhytides.Performance testing demonstrates that the handpieces can perform equivalently.
Vacuum SpecificationsN/AN/AVacuum Pulse: Frequency: 2.5 Hz Vacuum Pressure: >250 mbar, 1-5 levelThe Vacuum function is not included with the Mini Shaper Max handpiece. The vacuum function is used for cellulite treatment and is not related to pain relief indication.
Power Density0.05 W/mm20.08 W/mm20.06 W/mm2Similar. Differences do not present additional risks to safety or efficacy.
Patient Contacting MaterialsBiocompatibleBiocompatibleBiocompatibleAll handpieces have been tested and determined to be biocompatible.

Mini-Shaper Max Handpiece Predicate Comparison

SpecificationSubject DevicePrimary Predicate – Imagine TC SMALL handpieceSecondary PredicateComparison
Indications for useThe Mini-Shaper MAX Handpiece is Indicated for • Relief of minor muscle aches and pain, relief of muscle spasm • Temporary improvement of local blood circulationThe Imagine TC Skin Treatment System is a noninvasive device intended for use in Dermatologic and General Surgical procedures for the noninvasive treatment of mild to moderate facial wrinkles and rhytides.The EndyMed Contour Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage:The subject device excludes the term 'general surgical' to better align with the intended use of the device. The new and modified handpieces include the wrinkles indication for use, but limit it to 65W power, the same that the predicate device is cleared for. The Contour Handpiece predicate is cleared for use with the TC Imagine

Page 16

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject DevicePrimary Predicate – Imagine TC SMALL handpieceSecondary PredicateComparison
When used up to 65W, the Mini-Shaper Max handpiece is intended for the non-invasive treatment of mild to moderate facial wrinkles and rhytides.Relief of minor muscle aches and pain, relief of muscle spasm Temporary improvement of local blood circulation Temporary reduction in the appearance of cellulite.system. The new and modified handpieces share the pain relief indication for use that the Contour Handpiece is cleared for. They do not, however, include the vacuum function, and therefore do not include the cellulite reduction indication.
Mode of ActionNon-invasive RadiofrequencyNon-invasive RadiofrequencyNon-invasive Radiofrequency with vacuumSubject does not include vacuum function, and therefore does not include cellulite reduction indication
PolarityBipolarBipolarBipolarSame
Frequency1MHz1MHz1MHzSame
ModeContinuousContinuousContinuousSame
Max Output PowerUp to 70WUp to 65WUp to 85WCompared to the predicate Small HP, in the Mini Shaper MAX HP, the electrode contact surface area ratio is 2.1 times bigger and the max output power is 1.17 times bigger. The power density for the Mini Shaper is lower than the predicate device, making the new handpiece safer than the previously cleared Small Handpiece. The wrinkles indication is limited to 65W to match the predicate. Tissue Heating Testing with FLIR camera supports the addition of the pain relief indication that is also cleared for the TC Imagine System (Contour Handpiece). The

Page 17

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject DevicePrimary Predicate – Imagine TC SMALL handpieceSecondary PredicateComparison
specifications of the Mini Shaper Max Handpiece are similar to the Contour handpiece. The output power is slightly lower, and the vacuum feature is not included in the Mini Shaper Max handpiece. Tissue Heating Testing with FLIR camera supports the addition of the pain relief indication that is also cleared for the TC Imagine System (Contour Handpiece). The
Pulse Duration30 sec30 sec30 secSame
Electrode Surface Area641.7 mm2304.8 mm2 (small Handpiece)1381.2 mm2Different. Power densities of the electrodes are similar. Differences do not present additional risks to safety or efficacy.
Vacuum SpecificationsN/AN/AVacuum Pulse: Frequency: 2.5 Hz Vacuum Pressure: >250 mbar, 1-5 levelVacuum function is not included with the Mini Shaper Max handpiece. The vacuum function is used for cellulite treatment and is not related to pain relief indication.
Power Density0.11 W/mm20.21 W/mm20.06 W/mm2Different. Differences do not present additional risks to safety or efficacy. Performance testing demonstrates that the handpieces can perform equivalently.
Patient Contacting MaterialsBiocompatibleBiocompatibleBiocompatibleAll handpieces have been tested and determined to be biocompatible.

Page 18

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject DevicePrimary Predicate – Imagine TC SMALL handpieceSecondary PredicateComparison
specifications of the Mini Shaper Max Handpiece are similar to the Contour handpiece. The output power is slightly lower, and the vacuum feature is not included in the Mini Shaper Max handpiece.
Pulse Duration30 sec30 sec30 secSame
Electrode Surface Area641.7 mm2304.8 mm2 (small Handpiece)1381.2 mm2Different. Power densities of the electrodes are similar. Differences do not present additional risks to safety or efficacy.
Vacuum SpecificationsN/AN/AVacuum Pulse: Frequency: 2.5 Hz Vacuum Pressure: >250 mbar, 1-5 levelVacuum function is not included with the Mini Shaper Max handpiece. The vacuum function is used for cellulite treatment and is not related to pain relief indication.
Power Density0.11 W/mm20.21 W/mm20.06 W/mm2Different. Differences do not present additional risks to safety or efficacy. Performance testing demonstrates that the handpieces can perform equivalently.
Patient Contacting MaterialsBiocompatibleBiocompatibleBiocompatibleAll handpieces have been tested and determined to be biocompatible.

SMALL Max Handpiece Predicate Comparison

SpecificationSubject DevicePrimary Predicate – Imagine TC SMALL handpieceComparison
Indications for useThe Small MAX Handpiece is intended for the noninvasive treatment of mild to moderate facial wrinkles and rhytides.The Imagine TC Skin Treatment System is a noninvasive device intended for use in Dermatologic and General Surgical procedures for the noninvasive treatment of mild to moderate facial wrinkles and rhytides.The subject device excludes the term 'general surgical' to better align with the intended use of the device. The new and modified handpieces include the wrinkles indication for use, but limit it to 65W power, the same that the predicate device is cleared for. The Small Max handpiece is identical to the cleared Small handpiece and has the same indication.
Mode of ActionNon-invasive RadiofrequencyNon-invasive RadiofrequencySame
PolarityBipolarBipolarSame
Frequency1MHz1MHzSame
ModeContinuousContinuoussame
Max Output PowerUp to 60WUp to 65WSame as Imagine TC Small handpiece
Pulse Duration30 sec30 secSame
Electrode Surface Area304.8 mm^2304.8 mm^2 (small Handpiece)Same as Imagine TC Small Handpiece
Vacuum SpecificationsN/AN/A
Power Density0.21 W/mm20.21 W/mm2Same as Imagine TC Small Handpiece, higher power density than Contour Max due to smaller electrode size.

Page 19

510(K) Summary

EndyMed PRO MAX

K242996

SMALL Max Handpiece Predicate Comparison

SpecificationSubject DevicePrimary Predicate – Imagine TC SMALL handpieceComparison
Indications for useThe Small MAX Handpiece is intended for the noninvasive treatment of mild to moderate facial wrinkles and rhytides.The Imagine TC Skin Treatment System is a noninvasive device intended for use in Dermatologic and General Surgical procedures for the noninvasive treatment of mild to moderate facial wrinkles and rhytides.The subject device excludes the term 'general surgical' to better align with the intended use of the device. The new and modified handpieces include the wrinkles indication for use, but limit it to 65W power, the same that the predicate device is cleared for. The Small Max handpiece is identical to the cleared Small handpiece and has the same indication.
Mode of ActionNon-invasive RadiofrequencyNon-invasive RadiofrequencySame
PolarityBipolarBipolarSame
Frequency1MHz1MHzSame
ModeContinuousContinuoussame
Max Output PowerUp to 60WUp to 65WSame as Imagine TC Small handpiece
Pulse Duration30 sec30 secSame
Electrode Surface Area304.8 mm^2304.8 mm^2 (small Handpiece)Same as Imagine TC Small Handpiece
Vacuum SpecificationsN/AN/A
Power Density0.21 W/mm20.21 W/mm2Same as Imagine TC Small Handpiece, higher power density than Contour Max due to smaller electrode size.
Patient Contacting MaterialsElectrode: Copper plated with hard Chrome or ST-316 Handpiece: PC Makrolon 2458 or Tritan Copolyester MXF121Electrode: ST-316 Handpiece: Steralloy #2463 ST-316electrodes are the same as the predicate device. Biocompatibility testing (Sensitization, Irritation, Cytotoxicity) was conducted on handpieces with both stainless steel and copper coated electrodes.

FSR MAX Handpiece Predicate Comparison

SpecificationSubject (FSR Max handpiece)Predicate Device (FSR Handpiece K101510)Comparison
Indications for UseThe FSR MAX Handpiece is intended for dermatological procedures requiring ablation and resurfacing of the skin.The FSR MAX Applicator is intended for dermatological procedures requiring ablation and resurfacing of the skin.Same
Mode of ActionAblative RadiofrequencyAblative RadiofrequencySame
PolarityBipolarBipolarSame
Frequency1MHz1MHzSame
ModepulsedpulsedSame
Max output power1-6W1-6WSame
Pulse duration10-60 msec10-60 msecSame
Electrode Size0.07 mm20.07 mm2Same
Power density84.8 W/mm284.8 W/mm2Same
Patient Contacting MaterialsBiocompatibleBiocompatibleSame

Page 20

510(K) Summary

EndyMed PRO MAX

K242996

SpecificationSubject (FSR Max handpiece)Predicate Device (FSR Handpiece K101510)Comparison
Patient Contacting MaterialsElectrode: Copper plated with hard Chrome or ST-316 Handpiece: PC Makrolon 2458 or Tritan Copolyester MXF121Electrode: ST-316 Handpiece: Steralloy #2463ST-316 electrodes are the same as the predicate device. Biocompatibility testing (Sensitization, Irritation, Cytotoxicity) was conducted on handpieces with both stainless steel and copper coated electrodes.

FSR MAX Handpiece Predicate Comparison

SpecificationSubject (FSR Max handpiece)Predicate Device (FSR Handpiece K101510)Comparison
Indications for UseThe FSR MAX Handpiece is intended for dermatological procedures requiring ablation and resurfacing of the skin.The FSR MAX Applicator is intended for dermatological procedures requiring ablation and resurfacing of the skin.Same
Mode of ActionAblative RadiofrequencyAblative RadiofrequencySame
PolarityBipolarBipolarSame
Frequency1MHz1MHzSame
ModepulsedpulsedSame
Max output power1-6W1-6WSame
Pulse duration10-60 msec10-60 msecSame
Electrode Size0.07 mm20.07 mm2Same
Power density84.8 W/mm284.8 W/mm2Same
Patient Contacting MaterialsBiocompatibleBiocompatibleSame

Page 21

510(K) Summary

EndyMed PRO MAX

K242996

Performance Testing

Verification and validation activities were successfully completed and establish that the EndyMed PRO MAX performs as intended. Testing included the following:

• IEC 60601-1:2005 + A1:2012; Medical Electrical Equipment- Part 1: General Requirements For Basic Safety And Essential Performance
• IEC 60601-1-2:2014; Medical Electrical Equipment- Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
• IEC 60601-2-2:2017; Medical Electrical Equipment –part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories
• IEC 62304:2006+A1:2015; Medical Device – Software Life Cycle Processes
• EN ISO 14971:2012; Medical Devices – Application Of Risk Management To Medical Devices
• ISO 11135 Second edition 2014-07-15: Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical device
• ISO 10993-1:2018: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
• ISO 11607-1:2019-02: Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including Amendment 1 (2023)]

Software verification and validation testing was conducted, and documentation provided in accordance with FDA's Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices.

Bench testing was performed to verify the accuracy of the output of RF energy.

Performance Testing - Tissue Heating Testing

The Shaper Max, iFine Max, Mini Shaper Max, and Contour Max handpieces were all subjected to tissue heating testing to verify their safety and effectiveness for the requested pain relief indication. The Small Max Handpiece was not included in the study because it has not been modified since its original clearance. The testing was performed on 3 subjects with 3 anatomical sites each. The test was performed at the device's high settings and demonstrated that each handpiece can maintain a therapeutic temperature of 39-45°C for at least 10 minutes of treatment.

Biocompatibility

A biocompatibility assessment has been conducted to determine the biocompatibility risks of the change in patient contacting materials, and appropriate testing has been conducted to demonstrate safety of these materials.

Page 22

510(K) Summary

EndyMed PRO MAX

K242996

Discussion

This submission is intended to combine multiple handpieces previously cleared for use into a single system, introduce minor modifications to the system handpieces, and include new handpieces. The EndyMed PRO MAX system is very similar to the previously cleared TC Imagine Skin Treatment System. The differences are in max output power and the addition of a thermal camera for temperature monitoring (optional). The original TC Imagine Skin Treatment System included 2 handpieces. Three additional handpieces have been added with individual 510(k)s over time (Contour Handpiece, Intensif Applicator and FSR Applicator). This submission adds two new handpieces (iFine Max and Mini-shaper Max).

The current generation of handpieces have all been verified for output power control. The increased wattage allows the device to more quickly reach the target temperature of 40-45°C. The new handpieces – iFine Max and Mini Shaper Max are smaller than the previously cleared handpieces, providing opportunities to treat smaller and harder to reach areas. The overall design and principle of operation are the same between devices. Many of the modified handpieces are closest in comparison to the Contour Handpiece (K161199), which was cleared for use with the Imagine TC Skin Therapy System. Therefore, the Contour Handpiece is included as a predicate device as well to establish a pain relief indication. Performance testing described below shows the TC handpieces and Contour Handpiece perform as intended to meet the pain relief indication standards.

The Intensif Max tip is unchanged and is substantially equivalent to the predicate device.