K Number
K242981
Device Name
Atellica IM Thyroglobulin (Tg)
Date Cleared
2025-06-20

(267 days)

Product Code
Regulation Number
866.6010
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Atellica IM Thyroglobulin (Tg) assay is for in vitro diagnostic use in the quantitative measurement of thyroglobulin in human serum and plasma (EDTA and lithium heparin) using the Atellica IM Analyzer. Thyroglobulin measurements are used as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.
Device Description
The Atellica IM Thyroglobulin (Tg) assay includes: - **Tg ReadyPack primary reagent pack:** - **Lite Reagent:** mouse monoclonal anti-human Tg antibody labeled with acridinium ester (~1.13 μg/mL); bovine serum albumin (BSA); mouse IgG; buffer; stabilizers; preservatives (7.5 mL/reagent pack). - **Solid Phase:** streptavidin-coated paramagnetic microparticles preformed with biotinylated mouse monoclonal antihuman Tg antibody (~267 μg/mL); BSA; mouse IgG; buffer; stabilizers; preservatives (15.0 mL/reagent pack). - **Ancillary Well Reagent:** BSA; bovine gamma globulin; buffer; preservatives (6.0 mL/reagent pack). - **Tg CAL:** After reconstitution, human thyroglobulin; BSA; buffer; stabilizers; preservatives (2.0 mL/vial). The following devices are sold separately: - **Atellica IM Tg MCM:** - **MCM 1:** After reconstitution, bovine serum albumin (BSA); buffer; stabilizers; preservatives (1.0 mL/vial). - **MCM 2–5:** After reconstitution, various levels of human thyroglobulin; BSA; buffer; stabilizers; preservatives (1.0 mL/vial).
More Information

Not Found

No
The device description and performance studies do not mention any AI, DNN, or ML components or algorithms. The device is a diagnostic assay that measures thyroglobulin levels using chemical reagents.

No.
The device is an in vitro diagnostic (IVD) assay used for quantitative measurement of thyroglobulin, which aids in monitoring thyroid cancer, not for treating or preventing disease.

Yes

The "Intended Use" section explicitly states that the Atellica IM Thyroglobulin (Tg) assay is an "in vitro diagnostic use" for the quantitative measurement of thyroglobulin, which is used as an "aid in monitoring differentiated thyroid cancer patients." This clearly indicates its diagnostic purpose.

No

The device is a laboratory assay (an in vitro diagnostic test) that measures a biomarker (thyroglobulin) in a patient's serum or plasma using specific chemical reagents and an analyzer. This involves physical components (reagents, microparticles), not just software.

Yes.

The device is explicitly stated as being for "in vitro diagnostic use" and measures a biomarker (thyroglobulin) in human bodily fluids (serum and plasma) to aid in monitoring a medical condition (differentiated thyroid cancer).

N/A

Intended Use / Indications for Use

The Atellica IM Thyroglobulin (Tg) assay is for in vitro diagnostic use in the quantitative measurement of thyroglobulin in human serum and plasma (EDTA and lithium heparin) using the Atellica® IM Analyzer.

Thyroglobulin measurements are used as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.

Product codes (comma separated list FDA assigned to the subject device)

MSW

Device Description

The Atellica® IM Thyroglobulin (Tg) includes the Tg ReadyPack® primary reagent pack (Lite Reagent and Solid Phase) and Ancillary Well Reagent. The Lite Reagent contains Mouse monoclonal anti-human Tg antibody labeled with acridinium ester, bovine serum albumin (BSA), mouse IgG, buffer, stabilizers, and preservatives. The Solid Phase contains Streptavidin-coated paramagnetic microparticles preformed with biotinylated mouse monoclonal antihuman Tg antibody, BSA, mouse IgG, buffer, stabilizers, and preservatives. The Ancillary Well Reagent contains BSA, bovine gamma globulin, buffer, and preservatives. The Atellica IM Tg MCM (MCM 1 and MCM 2–5) are sold separately. MCM 1 contains bovine serum albumin (BSA), buffer, stabilizers, and preservatives after reconstitution. MCM 2–5 contain various levels of human thyroglobulin, BSA, buffer, stabilizers, and preservatives after reconstitution.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Apparently healthy adults (≥ 22 years)
Post-thyroidectomy adults (≥ 22 years)

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

A prospective, multi-center study was conducted using the Atellica IM Analyzer to assess the clinical performance of the Atellica IM Tg assay. A total of 291 serum samples were collected from 189 subjects diagnosed with differentiated thyroid cancer, 6 or more weeks following thyroidectomy or radioiodine ablation.
Sensitivity and specificity were calculated by comparing the Atellica IM Tg assay result to structural disease (SD). The assay result was considered positive if the Tg concentration was ≥ 0.2 ng/mL and negative if the Tg concentration was

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.

FDA 510(k) Clearance Letter - Atellica IM Thyroglobulin (Tg)

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 20, 2025

Siemens Healthcare Diagnostics, Inc.
Mey Vasquez
Regulatory Affairs Professional
511 Benedict Avenue
Tarrytown, New York 10591

Re: K242981
Trade/Device Name: Atellica IM Thyroglobulin (Tg)
Regulation Number: 21 CFR 866.6010
Regulation Name: Tumor-Associated Antigen Immunological Test System
Regulatory Class: Class II
Product Code: MSW
Dated: May 21, 2025
Received: May 22, 2025

Dear Mey Vasquez:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K242981 - Mey Vasquez Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K242981 - Mey Vasquez Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ying Mao -S

Ying Mao, Ph.D.
Branch Chief
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K242981

Device Name: Atellica IM Thyroglobulin (Tg)

Indications for Use (Describe):

The Atellica IM Thyroglobulin (Tg) assay is for in vitro diagnostic use in the quantitative measurement of thyroglobulin in human serum and plasma (EDTA and lithium heparin) using the Atellica IM Analyzer.

Thyroglobulin measurements are used as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Siemens Healthcare Diagnostics 510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 1 of 12

510(k) Summary of Safety and Effectiveness

Introduction: According to the requirements of SMDA 1990 and 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.

The assigned 510(k) Number: K242981

1. Date Prepared

June 12, 2025

2. Applicant Information

Siemens Healthcare Diagnostics Inc.
511 Benedict Avenue,
Tarrytown, NY 10591 USA

Contact: Mey Vasquez
Regulatory Affairs Professional
E-mail: mey.vasquez@siemens-healthineers.com

3. Regulatory Information

Assay
Trade NameAtellica IM Thyroglobulin (Tg)
DeviceThyroglobulin Test System
DefinitionThyroglobulin test system is an in vitro diagnostic device intended to measure Thyroglobulin in human serum and plasma. The test is intended to be used as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.
FDA ClassificationClass II
Review PanelClinical Chemistry
Product CodeMSW

Page 6

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 2 of 12

Regulation Number21 CFR 866.6010

4. PREDICATE DEVICE

Name of Device: Access Thyroglobulin
510(k): K241423

5. DEVICE DESCRIPTION

The following devices are included in the Atellica® IM Thyroglobulin (Tg):

MaterialDescription
Tg ReadyPack® primary reagent pack
Lite Reagent7.5 mL/reagent pack
Mouse monoclonal anti-human Tg antibody labeled with acridinium ester (~1.13 μg/mL); bovine serum albumin (BSA); mouse IgG; buffer; stabilizers; preservatives
Solid Phase15.0 mL/reagent pack
Streptavidin-coated paramagnetic microparticles preformed with biotinylated mouse monoclonal antihuman Tg antibody (~267 μg/mL); BSA; mouse IgG; buffer; stabilizers; preservatives
Ancillary Well Reagent6.0 mL/reagent pack
BSA; bovine gamma globulin; buffer; preservatives
Tg CAL2.0 mL/vial
After reconstitution, human thyroglobulin; BSA; buffer; stabilizers; preservatives

The following devices are sold separately:

MaterialDescription
Atellica IM Tg MCM:

Page 7

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 3 of 12

MCM 1:1.0 mL/vial
After reconstitution, bovine serum albumin (BSA); buffer; stabilizers; preservatives
Atellica IM Tg MCM:
MCM 2–5:1.0 mL/vial
After reconstitution, various levels of human thyroglobulin; BSA; buffer; stabilizers; preservatives

6. INTENDED USE/ INDICATIONS FOR USE

The Atellica® IM Thyroglobulin (Tg) assay is for in vitro diagnostic use in the quantitative measurement of thyroglobulin in human serum and plasma (EDTA and lithium heparin) using the Atellica® IM Analyzer.

Thyroglobulin measurements are used as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.

7. Special Conditions for Use Statement

For Prescription Use

8. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Candidate DevicePredicate
ItemAtellica IM Tg assayBeckman Coulter Access Thyroglobulin (K241423)
Intended UseThe Atellica® IM Thyroglobulin (Tg) assay is for in vitro diagnostic use in the quantitative measurement of thyroglobulin in human serum and plasma (EDTA and lithium heparin) using the Atellica® IM Analyzer.

Thyroglobulin measurements are used as an aid in monitoring | Access Thyroglobulin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems. This device is |

Page 8

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 4 of 12

Candidate DevicePredicate
ItemAtellica IM Tg assayBeckman Coulter Access Thyroglobulin (K241423)
differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.intended to aid in monitoring for the presence of persistent or recurrent/metastatic disease in patients who have differentiated thyroid cancer (DTC) and have had thyroid surgery (with or without ablative therapy), and who lack serum thyroglobulin antibodies.
Indications for UseSame as Intended Use (Candidate)Same (for Predicate)

Similarities

LoB0.039 ng/mL (0.059 pmol/L)≤ 0.03 ng/mL
LoD0.044 ng/mL (0.067 pmol/L)≤ 0.05 ng/mL
LoQ0.050 ng/mL (0.076 pmol/L)≤ 0.1 ng/mL
MeasurementQuantitativeSame
TechnologyChemiluminescentSame
Operating PrincipleFully automated Sandwich immunoassaySame
Sample typeSerum, EDTA plasma, lithium heparin plasmaHuman Serum and Plasma
StandardizationThe assay standardization is traceable to the Community Bureau of Reference (BCR) Certified Reference Material (CRM) 457.Community Bureau of Reference (BCR) Certified Reference Material (CRM) 457

Page 9

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 5 of 12

Candidate DevicePredicate
ItemAtellica IM Tg assayBeckman Coulter Access Thyroglobulin (K241423)
differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.intended to aid in monitoring for the presence of persistent or recurrent/metastatic disease in patients who have differentiated thyroid cancer (DTC) and have had thyroid surgery (with or without ablative therapy), and who lack serum thyroglobulin antibodies.
Indications for UseSame as Intended Use (Candidate)Same (for Predicate)

Similarities

LoB0.039 ng/mL (0.059 pmol/L)≤ 0.03 ng/mL
LoD0.044 ng/mL (0.067 pmol/L)≤ 0.05 ng/mL
LoQ0.050 ng/mL (0.076 pmol/L)≤ 0.1 ng/mL
MeasurementQuantitativeSame
TechnologyChemiluminescentSame
Operating PrincipleFully automated Sandwich immunoassaySame
Sample typeSerum, EDTA plasma, lithium heparin plasmaHuman Serum and Plasma
StandardizationThe assay standardization is traceable to the Community Bureau of Reference (BCR) Certified Reference Material (CRM) 457.Community Bureau of Reference (BCR) Certified Reference Material (CRM) 457

Page 10

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 6 of 12

Candidate DevicePredicate
ItemAtellica IM Tg assayBeckman Coulter Access Thyroglobulin (K241423)
Assigned values for calibrators are traceable to this standardization.
Clinical Cut-Off0.2 ng/mLNot applicable
Intended Use Population(s)Thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.Patients who have differentiated thyroid cancer (DTC) and have had thyroid surgery (with or without ablative therapy), and who lack serum thyroglobulin antibodies.

Differences

Calibration2 levels6 levels
Assay Range0.050–150 ng/mL (0.076–227 pmol/L)0.1 - 500 ng/mL
Hook EffectNo hook effect up to 80,000 ng/mL (121,200 pmol/L).No hook effect up to at least 40,000 ng/mL.
Sample Volume100 µL40 µL
Detection AntibodyMouse monoclonal anti-human Tg antibody labeled with acridinium ester.Mouse monoclonal anti-thyroglobulin-alkaline phosphatase (bovine) conjugate in a TRIS buffer with protein (bovine, murine).
Capture AntibodyBiotinylated mouse monoclonal anti-human Tg antibody that is bound to streptavidin-coated paramagnetic latex particles.Mouse monoclonal anti-thyroglobulin antibodies coupled to biotin in a HEPES buffer with protein (bovine and mouse).

Page 11

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 7 of 12

9. PERFORMANCE CHARACTERISTICS DATA

9.1. Detection Capability

The limit of blank (LoB), limit of detection (LoD), and the limit of quantitation (LoQ) were determined as described in CLSI protocol EP17-A2.

The Atellica IM Tg assay has a LoB of 0.039 ng/mL (0.059 pmol/L), a LoD of 0.044 ng/mL (0.067 pmol/L), and a LoQ of 0.050 ng/mL (0.076 pmol/L).

9.2. Precision

Precision was determined in accordance with CLSI Document EP05‑A3.

Samples were assayed in replicates of 2 with 2 runs per day using a 20‑day protocol. The following results are representative of the performance of the assay:

SampleN^aMean ng/mL (pmol/L)RepeatabilityWithin-Laboratory Precision
SD^b ng/mL (pmol/L)CV^c (%)SD ng/mL (pmol/L)CV (%)
Serum A800.078 (0.118)0.005 (0.008)6.40.007 (0.011)9.0
Serum B800.153 (0.232)0.003 (0.005)2.00.005 (0.008)3.3
Serum C801.79 (2.71)0.032 (0.048)1.80.042 (0.064)2.3
Serum D806.12 (9.27)0.082 (0.124)1.30.145 (0.220)2.4
Serum E8025.1 (38.0)0.575 (0.871)2.30.793 (1.20)3.2
Serum F8078.9 (120)0.976 (1.48)1.22.12 (3.21)2.7
Serum G80136 (206)2.59 (3.92)1.93.95 (5.98)2.9
Control 1803.83 (5.80)0.097 (0.147)2.50.151 (0.229)3.9
Control 28042.7 (64.7)1.24 (1.88)2.91.71 (2.59)4.0
Control 380121 (183)2.91 (4.41)2.44.38 (6.64)3.6

^a Number of measurements.
^b Standard deviation.
^c Coefficient of variation.

Page 12

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 8 of 12

9.3. Reproducibility

Reproducibility was determined using the Atellica IM Analyzer in accordance with CLSI document EP05‑A3. Testing was performed using 3 sites and 1 reagent lot. Samples were assayed in replicates of 3 with 2 runs per day using a 5‑day protocol (Number of measurements per sample = 90). The following results are representative of the performance of the assay:

SampleMean ng/mL (pmol/L)RepeatabilityBetween-RunBetween-DayBetween-SiteReproducibility
SD^a ng/mL (pmol/L)CV^b (%)SD ng/mL (pmol/L)CV (%)SD ng/mL (pmol/L)CV (%)SD ng/mL (pmol/L)CV (%)SD ng/mL (pmol/L)CV (%)
Serum A0.155 (0.235)0.005 (0.008)3.20.000 (0.000)0.00.000 (0.000)0.00.003 (0.005)1.90.006 (0.009)3.9
Serum B1.83 (2.77)0.031 (0.047)1.70.015 (0.023)0.80.008 (0.012)0.40.101 (0.153)5.50.107 (0.162)5.8
Serum C25.7 (38.9)0.321 (0.486)1.20.100 (0.152)0.40.098 (0.148)0.40.359 (0.544)1.40.501 (0.759)1.9
Serum D74.5 (113)1.12 (1.70)1.50.000 (0.000)0.00.243 (0.368)0.31.91 (2.89)2.62.23 (3.38)3.0
Serum E119 (180)1.76 (2.67)1.50.716 (1.08)0.60.845 (1.28)0.74.53 (6.86)3.84.99 (7.56)4.2
Control 14.35 (6.59)0.070 (0.106)1.60.049 (0.074)1.10.053 (0.080)1.20.065 (0.098)1.50.120 (0.182)2.8
Control 243.6 (66.1)0.812 (1.23)1.90.499 (0.756)1.10.000 (0.000)0.01.29 (1.95)3.01.60 (2.42)3.7
Control 3128 (194)2.53 (3.83)2.01.69 (2.56)1.30.000 (0.000)0.06.38 (9.67)5.07.07 (10.7)5.5

9.4. Linearity

Linearity testing was performed using the Atellica IM Analyzer in accordance with CLSI Document EP06‑ed2. The assay is linear for the measuring interval of 0.050–150 ng/mL (0.076–227 pmol/L).

9.5. Specimen Equivalence

Specimen equivalency was tested in accordance with the governing standard CLSI Document EP09c‑ed3. Specimen matrix studies were conducted to compare serum to the following specimen types and collection devices: dipotassium EDTA plasma, lithium heparin

Page 13

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 9 of 12

plasma, serum gel barrier tube, lithium heparin plasma gel-barrier tube. Each matrix comparison study included a minimum of 84 paired samples across the assay measuring interval. Results of these studies confirm that the performance of the assay is equivalent across the supported specimen types and collection devices.

9.6. Interferences

Hemolysis, Icterus, Lipemia (HIL)

Interference testing was performed using the Atellica IM Analyzer in accordance with CLSI document EP07‑ed3. Interference as defined by bias greater than 10% was not observed for the following substances when tested at analyte concentrations of 0.144–0.224 ng/mL (0.218–0.339 pmol/L) and 21.9–26.9 ng/mL (33.2–40.8 pmol/L).

SubstanceSubstance Test Concentration
Hemoglobin1000 mg/dL
Bilirubin, conjugated60 mg/dL
Bilirubin, unconjugated60 mg/dL
Lipemia (Intralipid)1900 mg/dL
Lipemia (triglycerides)1500 mg/dL

Other Substances

Interference testing was performed using the Atellica IM Analyzer in accordance with CLSI document EP07‑ed3.19 Interference as defined by bias greater than 10% was not observed for the following substances when tested at analyte concentrations of 0.125–0.235 ng/mL (0.189–0.356 pmol/L) and 21.6–27.8 ng/mL (32.7–42.1 pmol/L).

SubstanceSubstance Test ConcentrationSubstanceSubstance Test Concentration
Acetaminophen20 mg/dLItraconazole3 mg/dL
Acetylsalicylic Acid (Aspirin)65 mg/dLK3 EDTA5.4 mg/mL
AFP881.24 ng/mLLenvatinib Mesylate2.62 mg/dL
Amiodarone8.92 µmol/LLevodopa0.75 mg/dL
Ampicillin33.0 mg/dLMethyldopa2.25 mg/dL
Ascorbic Acid2590 mg/dLMetronidazole12.3 mg/dL
Biotin3510 ng/mLOctreotide Acetate (Sandostatin)5.2 ng/mL

Page 14

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 10 of 12

SubstanceSubstance Test ConcentrationSubstanceSubstance Test Concentration
Cabozantinib-S-Malate15.3 mg/dLPerchlorate200 µg/mL
Carbimazole2.4 µg/mLPhenylbutazone32.1 mg/dL
Cefoxitin92.7 mg/dLPrednisolone8.31 µmol/L
Cyclosporine0.075 mg/dLPropranolol7.71 µmol/L
Doxycycline4.5 mg/dLPropylthiouracil7.2 µg/mL
Fluocortolone270 ng/mLRifampicin7.5 mg/dL
Fluorescein6 µg/mLRheumatoid Factor (RF)550 IU/mL
FSH40 mIU/mLSilwet L7200.03 mg/mL
Hydrocortisone984 ng/mLTBG (thyroxine binding globulin)210 µg/mL
Ibuprofen50 mg/dLTheophylline6 mg/dL
Imatinib (TKI)13.4 µg/mLThiamazole (Methimazole)300 ng/mL
Immunoglobulin G (IgG)2 g/dLTotal Protein30 mg/mL
Immunoglobulin M (IgM)0.3 g/dLTotal Protein120 mg/mL
Iodide38 mg/dLVEGF2835 pg/mL

9.7. Cross-Reactivity

Cross-reactivity was determined using the Atellica IM Analyzer in accordance with CLSI document EP07‑ed3. Cross-reactivity of samples spiked with various substances does not exceed 1.0% at analyte concentrations of 0.185–0.261 ng/mL (0.280–0.395 pmol/L) and 23.9–27.5 ng/mL (36.2–41.6 pmol/L).

SubstanceSubstance Test Concentration
T3100 ng/mL
T410 µg/mL
TSH235 mIU/mL
Galectin-31 µg/mL
Diiodothyronine (T2)55 µg/mL

Page 15

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 11 of 12

9.8. Reagent Stability

The on-board stability of the Atellica IM Tg reagents was determined to be 28 days on AIM with a calibration interval of 50 days.

The Atellica IM Tg calibrators when reconstituted were determined to be stable at 2-8°C and ≤ -20°C for 45 days and 60 days (thaw only once), respectively.

9.9. Sample Stability

  • After centrifugation, specimens in the primary collection device are stable for up to 3 days at 2–8°C. Samples in the primary collection device include serum stored on the clot, plasma stored on packed red cells, and samples processed and stored in gel-barrier blood collection tubes.
  • Separated serum samples are stable for up to 4 days at room temperature, and for up to 7 days at 2–8°C.
  • Separated plasma samples are stable for up to 3 days at room temperature, and for up to 4 days at 2–8°C.
  • Separated samples are stable at ≤ -20°C for up to 12 months and at ≤ -70°C for up to 24 months. Avoid more than 4 freeze-thaw cycles. Do not store in a frost-free freezer. Thoroughly mix thawed samples and centrifuge them before using.

9.10. High Dose Hook Effect

High Tg concentrations can cause a paradoxical decrease in the RLUs (high-dose hook effect). In this assay, no hook effect was observed up to 80,000 ng/mL (121,200 pmol/L).

9.11. Expected Values

Reference intervals were established on the Atellica IM Analyzer in accordance with CLSI Document EP28‑A3c. A total 321 apparently healthy patient serum samples which consisted of 164 healthy adult females and 157 healthy adult males were obtained.

Reference intervals were determined by calculating the 2.5th and 97.5th percentiles of the distribution of values.

GroupN^aReference Interval
Apparently healthy adults (≥ 22 years)3212.44–74.9 ng/mL (3.70–113 pmol/L)
Apparently healthy adult females1642.58–78.3 ng/mL (3.91–119 pmol/L)
Apparently healthy adult males1572.37–70.1 ng/mL (3.59–106 pmol/L)

Page 16

Siemens Healthcare Diagnostics
510(k) Summary

Siemens Healthcare Diagnostics, Inc
510(k) Submission – Atellica IM Tg Assay (K242981)
Page 12 of 12

Serum samples from 136 post-thyroidectomy adults were tested. Pregnant subjects and samples with detectable levels of anti-Tg were excluded from the study. The reference interval was determined by calculating the 95th percentile of the distribution of values.

GroupN^aReference Interval
Post-thyroidectomy adults (≥ 22 years)136