(272 days)
OPXION Optical Skin Viewer is a non-invasive imaging system intended to be used for real-time visualization of the external tissues of the human body. The two-dimensional, cross-sectional, three-dimensional, and en-face images of tissue microstructures can be obtained.
OPXION Optical Skin Viewer is composed of two parts consisting of a handheld probe and a mainframe, connected by an optical fiber cable. The device comes with three accessories: a USB 3.0 cable, a power adapter, and a power cord. The Optical Skin Viewer needs to be connected to a laptop or a personal computer. The device uses Optical Coherence Tomography (OCT) technology with a Superluminescent diode, 840 nm, 6 mW light source.
Based on the provided FDA 510(k) clearance letter for the OPXION Optical Skin Viewer, an optical device that visualizes external tissue and is not an AI/ML powered device, the document does not contain the specific information requested about acceptance criteria and a study that proves the device meets the acceptance criteria in the context of AI/ML performance.
The 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (VivoSight Topical OCT System) primarily based on intended use, technology (Optical Coherence Tomography), and general performance (image quality accepted by a qualified medical professional for visualization).
Therefore, I cannot provide a table of acceptance criteria, sample sizes for test sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, or details about training sets, as these specific details are not present in the provided document, nor are they typically required for a Class II medical imaging device like this one unless it incorporates AI/ML for diagnostic or interpretive functions.
However, I can extract the general acceptance criteria and the type of study conducted for this device based on the provided text:
Acceptance Criteria and Study:
The document describes the device's performance in terms of its ability to produce images for visualization, rather than offering specific quantitative metrics for diagnostic accuracy, sensitivity, or specificity that would be typical for an AI/ML driven device.
Here's an interpretation based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (General) | Reported Device Performance (General) |
|---|---|
| Image quality confirmed and accepted by a qualified medical professional. | The OPXION Optical Skin Viewer demonstrated consistent performance in producing images of a quality that is substantially equivalent to that produced by the cited predicate device. The device successfully displayed anatomical features of skin. |
| No adverse events or safety concerns were reported. The scanning process was well-tolerated by all subjects. | |
| Safe and effective clinical imaging device capable of generating two-dimensional, cross-sectional, three-dimensional, and en-face images of external tissue microstructure. |
2. Sample Size Used for the Test Set and Data Providence
- Test Set Sample Size: The study included three subjects with healthy skin and five subjects with diseased skin conditions.
- Data Provenance: Not explicitly stated, but implies a prospective study given the "Study Design" description of scanning "each target area in three sessions." The country of origin of the data is not specified in the 510(k) summary.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: The document states that "Image quality was confirmed and accepted by a qualified medical professional." This implies at least one, but the exact number beyond "a" is not specified.
- Qualifications of Experts: Described as "a qualified medical professional." No specific specialty (e.g., dermatologist, radiologist) or years of experience are provided.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly stated as an adjudication method in the context of multiple readers reaching consensus. The acceptance criterion notes "Image quality was confirmed and accepted by a qualified medical professional," which suggests a single reviewer or possibly an internal review process where consensus was reached without a formal adjudication method described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
- MRMC Study: No, an MRMC comparative effectiveness study was not conducted or described in the provided document. The study focuses on the device's ability to produce images and its substantial equivalence to a predicate, not on how human readers perform with or without the device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Standalone Performance: This device is an imaging system for visualization, not an AI/ML algorithm that provides diagnostic outputs. Therefore, the concept of "standalone performance" of an algorithm is not applicable or described. Its "performance" is its ability to acquire and display images.
7. The Type of Ground Truth Used
- Type of Ground Truth: The ground truth for this device's performance evaluation was the visual assessment and acceptance of image quality by a qualified medical professional, based on the successful display of "anatomical features of skin" for both healthy and diseased conditions. This is a form of expert consensus/acceptance on display quality.
8. The Sample Size for the Training Set
- Training Set Sample Size: Not applicable. This is an optical imaging device, not an AI/ML algorithm that undergoes a training phase with a specific dataset.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as there is no mention of an AI/ML training set.
In summary, the provided FDA 510(k) letter describes a traditional medical imaging device focused on visualization, not an AI/ML-powered device. Therefore, the detailed criteria and study designs typically associated with AI/ML device validation are absent from this document.
FDA 510(k) Clearance Letter - OPXION Optical Skin Viewer
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
June 23, 2025
OPXION Technology Inc.
℅ Ming-Yie Jan
Principal Consultant
RUSCERT Technology Co., LTD.
8F., No. 187, Lequn 2nd Rd., Zhongshan Dist.
Taipei City, 10462
Taiwan
Re: K242924
Trade/Device Name: OPXION Optical Skin Viewer (OPXSV1-01F)
Regulation Number: 21 CFR 892.1560
Regulation Name: Ultrasonic Pulsed Echo Imaging System
Regulatory Class: Class II
Product Code: NQQ
Dated: May 20, 2025
Received: May 22, 2025
Dear Ming-Yie Jan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K242924 - Ming-Yie Jan
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K242924 - Ming-Yie Jan
Page 3
Sincerely,
Jessica Carr -S
Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K242924
Device Name
OPXION Optical Skin Viewer (OPXSV1-01F)
Indications for Use (Describe)
OPXION Optical Skin Viewer is a non-invasive imaging system intended to be used for real-time visualization of the external tissues of the human body. The two-dimensional, cross-sectional, three-dimensional, and en-face images of tissue microstructures can be obtained.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
OPXION Tech. Incorporation
510(k) SUMMARY
Page 1 of 8 pages
510(k) SUMMARY
This 510 (K) summary is being submitted in accordance with requirements of Title 21 CFR Section 807.92.
A. 510(k) NUMBER K242924
B. DATE PREPARED June 16th, 2025
C. SUBMITTER OPXION Technology Inc.
22F., No. 29-2, Sec. 2, Zhongzheng E. Rd., Tamsui Dist., New Taipei City 251401, Taiwan
Phone number: +886-2-8809-1666
D. CONTACT PERSON
-
Primary Contact
Title: AVP
Name: Shao-Hua Huang
Tel: +886-2-8809-1666 # 2101
E-mail: becket.huang@opxiontech.com -
Official Correspondent
Title: Principal Consultant
Name: Ming-Yie Jan
Tel: +886287858989
E-mail: FDA.sub@ruscert.net
E. DEVICE
Proprietary Name: OPXION Optical Skin Viewer
Product Code: NQQ
Regulation Number: 892.1560
Regulation Name: Ultrasonic pulsed echo imaging system
Device Class: ClassII
Review Panel: General & Plastic Surgery
F. INDICATION(S) FOR USE
OPXION Optical Skin Viewer is a non-invasive imaging system intended to be used for real-time visualization of the external tissues of the human body. The two-dimensional, cross-sectional, three-dimensional, and en-face images of tissue microstructures can be obtained.
Page 6
OPXION Tech. Incorporation
510(k) SUMMARY
Page 2 of 8 pages
G. PRIMARY PREDICATE DEVICE
Proprietary Name: VivoSight Topical OCT System
Product Code: NQQ
Regulation Number: 892.1560
Regulation Name: Ultrasonic pulsed echo imaging system
510(k) Number: K093520
510(k) Submitter: Michelson Diagnostics Ltd
Device Class: ClassII
Review Panel: General & Plastic Surgery
H. DEVICE DECRIPTION
OPXION Optical Skin Viewer is composed of two parts consisting of a handheld probe and a mainframe, connected by an optical fiber cable. The device comes with three accessories: a USB 3.0 cable, a power adapter, and a power cord. The Optical Skin Viewer needs to be connected to a laptop or a personal computer.
I. COMPARISON OF CHARACTERISTICS WITH THE PREDICATE DEVICES
OPXION Optical Skin Viewer and predicate device K093520 are:
- Same intended use/ indication(s) for use
- Same intended population
- Same mode of action
- Same method of use
- Same duration of use
- Different technical characteristics
- Different physical characteristics design
The OPXION Optical Skin Viewer and the predicate device, VivoSight Topical OCT System (K093520), share the same intended use and have similar technological characteristics. They both use Optical Coherence Tomography (OCT) technology to create high-resolution real-time cross-sectional and enface images of tissue microstructure. Two devices utilize near-infrared light to produce OCT images with comparable axial and lateral resolutions. The differences in technical characteristics are based on the principle of design for each device. They are designed to meet the intended use of devices to deliver the performance. Those differences do not raise any new or different concerns about safety and effectiveness.
The OPXION Optical Skin Viewer and the predicate device, VivoSight Topical OCT System (K093520), share the same intended use and have similar technological characteristics. They both use Optical Coherence Tomography (OCT) technology to create high-resolution real-time cross-sectional and enface images of tissue microstructure. Two devices utilize near-infrared light to produce OCT images with comparable axial and lateral resolutions. The differences in technical characteristics are based on the principle of design for each device. They are designed to meet the intended use of devices to deliver the performance. Those differences do not raise any new or different concerns about safety and effectiveness.
Page 7
OPXION Tech. Incorporation
510(k) SUMMARY
Page 3 of 8 pages
Therefore, OPXION Technology Inc. concludes that OPXION Optical Skin Viewer is substantially equivalent to predicate devices regarding indications for use, design, and testing results, without raising any safety and efficacy risks or concerns.
Page 8
OPXION Tech. Incorporation
Page 4 of 9 pages
B. Substantial Equivalence Comparison Table-Basic Information
| Element of Comparison | Proposed Device OPXION Optical Skin Viewer | Predicate Device VivoSight Topical OCT System K093520 | Comparison |
|---|---|---|---|
| Product Code | NQQ | NQQ | Same |
| Intended Use/Indications for Use | OPXION Optical Skin Viewer is a non-invasive imaging system intended to be used for real-time visualization of the external tissues of the human body. The two-dimensional, cross-sectional, three-dimensional, and en-face images of tissue microstructures can be obtained. | VivoSight is a Multi-Beam Optical Coherence Tomography (OCT) system indicated for use in the two-dimensional, cross-sectional, real-time imaging of external tissues of the human body. | Same |
| Technology | Optical Coherence Tomography | Optical Coherence Tomography | Same |
| Near-Infrared Wavelength | Yes | Yes | Same |
| Light Source Center Wavelength | 840 nm | 1305 nm | Different |
| Lateral Resolution (Horizontal Optical Resolution) | 10 μm | < 7.5 μm | Different |
| Axial Resolution (Vertical Optical Resolution) | 7.5 μm | < 10 μm | Different |
| Maximum scan Range | 5mm x 5mm | 5mm x 5mm | Similar |
| Axial Scanning Range | Up to 2mm | Up to 2mm | Same |
| Scan time | 25 fps | >6fps | Different |
Page 9
OPXION Tech. Incorporation
Page 5 of 9 pages
| Element of Comparison | Proposed Device OPXION Optical Skin Viewer | Predicate Device VivoSight Topical OCT System K093520 | Comparison |
|---|---|---|---|
| Optical Radiation Safety | Class 1 Laser | Class 1 Laser | Same |
| Energy output | Superluminescent diode, 840 nm, 6 mW | Santec HSL-2000-11 wide sweep laser, ≥25 mW | Different |
| Power Supply | 100-240V, 50-60 Hz | 110-132 V, 50-60 Hz | Different |
| Weight and dimensions | 285mm × 135mm × 193mm 2.1 kg | 650 x 700 x 1650 mm 105 kg | Different |
Page 10
OPXION Tech. Incorporation
Page 6 of 9 pages
Discussion on Technical Characteristics Difference:
Difference in central wavelength and bandwidth: Both devices use near-infrared light sources, but there is a slight difference in the spectral range. The VivoSight Topical OCT System has a central wavelength of 1305 nm with a bandwidth of 150 nm, while the OPXION Optical Skin Viewer has a central wavelength of 840 nm with a bandwidth of ~45 nm. The specifications of devices are in the near-infrared light range without raising extra concerns on the safety of use. The spectral range of OPXION Optical Skin Viewer is also commonly used for clinical ophthalmology devices with illustrating the safety of 840-nm spectral range approved.
Difference in light source nature: The OPXION Optical Skin Viewer uses a semiconductor superluminescent diode and the VivoSight Topical OCT System uses a swept source laser. This difference does not affect the safety of use but may impact factors such as cost, system size, and power consumption.
Difference in light source power: The lower the power of the light source, the less impact it has on the skin. OPXION Optical Skin Viewer have a light source output power of approximately 6 mW, while the VivoSight Topical OCT System uses a high-power sweep source laser with a power of 25 mW. In comparison, OPXION Optical Skin Viewer have a relatively safer light power level with an effective performance.
Difference in system size: OPXION Optical Skin Viewer has smaller reduced size and less weight with a portable feature. The difference in system size does not affect the safety of use and efficacy.
Page 11
OPXION Tech. Incorporation
Page 7 of 9 pages
J. SUMMARY OF PERFORMANCE DATA
The following performance data were provided to demonstrate the safety and efficacy of OPXION Optical Skin Viewer:
| Standards | Test Results |
|---|---|
| IEC 62304 | PASS |
| IEC 60601-1: 2005/A1:2012/A2:2020 | PASS |
| EN 60601-1:2006+A2:2021 | PASS |
| ANSI AAMI ES60601-1:2005/A2:2021 IEC | PASS |
| 60601-2-22:2007 +A1:2012(3.1) | PASS |
| IEC EN 60825-1 | PASS |
| ASTM D4169-22 | PASS |
| ISO 10993-5:2009 | PASS |
| ISO 10993-10:2010 | PASS |
| IEC 62366:2007+AMD1:2014 | PASS |
| IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 | PASS |
Page 12
OPXION Tech. Incorporation
Page 8 of 9 pages
K. Performance Testing Summary
The study was conducted to evaluate the performance of the OPXION Optical Skin Viewer. The primary objective was to assess the device's ability to capture high-resolution images of skin across a diverse patient population, including variations in age, gender, and anatomical skin location.
Study Design
The evaluation involved scanning each target area in three sessions of 20 seconds each, for a total scanning duration of 60 seconds per site. The study included three subjects with healthy skin and five subjects with diseased skin conditions.
Skin Conditions Evaluated
- Healthy skin: Scanned across different anatomical locations and demographic profiles (age and gender).
- Diseased skin: Conditions included rosacea, acne scars, and psoriasis affecting various anatomical sites.
Acceptance Criteria
Image quality was confirmed and accepted by a qualified medical professional.
Data Collection Parameters
The study recorded the following parameters for each subject:
- Demographics: Gender and age
- Anatomical site of scanning
- Imaging results: 2D images, full-depth en-face images, and en-face images at different depths
Page 13
OPXION Tech. Incorporation
Page 9 of 9 pages
Results
The OPXION Optical Skin Viewer demonstrated consistent performance in producing images of a quality that is substantially equivalent to that produced by the cited predicate device. The device successfully displayed anatomical features of skin.
No adverse events or safety concerns were reported. The scanning process was well-tolerated by all subjects.
Summary
The OPXION Optical Skin Viewer is a safe and effective clinical imaging device capable of generating two-dimensional, cross-sectional, three-dimensional, and en-face images of external tissue microstructure. Its non-invasive nature, and ability to image skin support its potential utility.
L. SUBSTANTIAL EQUIVALENCE CONCLUSION
Based upon the information presented, OPXION Technology Inc. concludes that the OPXION Optical Skin Viewer is substantially equivalent to the predicate device regarding indications for use, design, and testing results, without raising any new safety and efficacy risks or concerns.
N/A