K Number
K242871
Manufacturer
Date Cleared
2024-11-21

(59 days)

Product Code
Regulation Number
888.3510
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ATTUNE Revision Hinge Knee is indicated in cases for cement use in patients who have reached skeletal maturity and for whom the surgeon has decided to resect both cruciate ligaments due to the following conditions:

  1. Severe instability, gross deformity and/or bone loss
  2. Failure of a previous knee reconstruction procedure.
  3. Trauma or tumor resection
  4. Absent or markedly insufficient collateral ligaments
Device Description

The ATTUNE™ Revision Hinge Knee, comprised of a femoral, housing assembly, and tibial insert component, is compatible with devices of the ATTUNE Knee System and the ATTUNE Revision Knee System and is designed to replace the natural articulating surface of the knee joint in total knee arthroplasty. The femoral component is a metal/ultra-high molecular weight polyethylene (UHMWPE) implant intended for cemented use. The housing assembly component is composed of both metal and antioxidant (AOX) UHMWPE subcomponents. The tibial insert component is composed of AOX UHMWPE.

AI/ML Overview

This document is a 510(k) Premarket Notification from the FDA for the "ATTUNE™ Revision Hinge Knee" device. It focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study demonstrating the device meets those criteria in the context of an AI/software device.

Therefore, many of the requested sections about acceptance criteria, device performance, sample sizes for test/training sets, ground truth establishment, expert involvement, and MRMC studies are not applicable to this particular document as it is for a mechanical implant and not an AI/software product. The information provided relates to the non-clinical testing of the physical implant's mechanical and material properties.

However, I can extract the relevant information regarding the tests performed for the determination of substantial equivalence (analogous to proving the device meets certain performance standards for a mechanical device).

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative acceptance criteria with reported performance values in a direct side-by-side comparison. Instead, it states that "The following tests were performed in support of the to demonstrate substantial equivalence of safety and efficacy with the predicate devices." This implies that the device performed equivalently to the predicate devices in these tests, satisfying the FDA's requirement for substantial equivalence.

Test TypeAcceptance Criterion (Implicit)Reported Device Performance (Implicit)
Shear TestingSubstantially equivalent to predicate devices.Performed in support of substantial equivalence and safety/efficacy.
Fatigue TestingSubstantially equivalent to predicate devices.Performed in support of substantial equivalence and safety/efficacy.
Wear PerformanceSubstantially equivalent to predicate devices.Performed in support of substantial equivalence and safety/efficacy.
Stability TestingSubstantially equivalent to predicate devices.Performed in support of substantial equivalence and safety/efficacy.
Contact Pressure TestingSubstantially equivalent to predicate devices.Performed in support of substantial equivalence and safety/efficacy.
BiocompatibilitySubstantially equivalent to predicate devices.Performed in support of substantial equivalence and safety/efficacy.
MRI Safety Evaluation TestingSubstantially equivalent to predicate devices (if applicable).Performed in support of substantial equivalence and safety/efficacy.

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not provide details on specific sample sizes for these tests (e.g., how many specimens were shear-tested or fatigue-tested). It also does not mention data provenance as these are mechanical/materials tests, not human data studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a submission for a mechanical medical device, not an AI/Software device requiring expert-established ground truth from medical images or clinical data. The "ground truth" for these tests would be the established engineering and material science standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for a mechanical device submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/software device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/software device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the mechanical tests listed (Shear, Fatigue, Wear, Stability, Contact Pressure, Biocompatibility, MRI Safety Evaluation), the "ground truth" would be established engineering standards, material science specifications, and regulatory requirements relevant to orthopedic implants.

8. The sample size for the training set

Not applicable. This is not a machine learning/AI device, so there is no training set mentioned.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

November 21, 2024

DePuy Ireland UC c/o Paige Myers Regulatory Affairs Specialist II DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, Indiana 46582

Re: K242871

Trade/Device Name: ATTUNE™ Revision Hinge Knee Regulation Number: 21 CFR 888.3510 Regulation Name: Knee Joint Femorotibial Metal/Polymer Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KRO Dated: September 20, 2024 Received: September 23, 2024

Dear Paige Myers:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Peter G. Allen Digitally signed by Peter G.
-S Allen -S
Date: 2024.11.21 12:31:58
-05'00'

for: Lixin Liu

Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K242871

Device Name

ATTUNE™ Revision Hinge Knee

Indications for Use (Describe)

The ATTUNE Revision Hinge Knee is indicated in cases for cement use in patients who have reached skeletal maturity and for whom the surgeon has decided to resect both cruciate ligaments due to the following conditions:

    1. Severe instability, gross deformity and/or bone loss
    1. Failure of a previous knee reconstruction procedure.
    1. Trauma or tumor resection
    1. Absent or markedly insufficient collateral ligaments

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY

(As required by 21 CFR 807.92 and 21 CFR 807.93)

Submitter Information
NameDePuy Ireland UC
AddressLoughbeg, RingaskiddyCo. Cork Munster, IRELAND
Establishment RegistrationNumber3015516266
Name of contact personPaige Myers
E-Mail Addresspmyers1@its.jnj.com and DePuySynthesJointsRegulatoryAffairs@its.jnj.com
Alternate Contact PersonMichael Liao
Alternate Contact E-Mailmliao7@its.jnj.com
Alternate Contact Phone+1 925 8901497
Date prepared7 October 2024
Name of device
Trade or proprietary nameATTUNE™ Revision Hinge Knee
Common or usual nameTotal Hinge Knee System
Classification name21 CFR 888.3510 - Knee joint femorotibial metal/polymer constrainedcemented prosthesis
ClassII
Classification panel87 Orthopedics
RegulationClass II - 21 CFR 888.3510
Product Code(s)KRO: Prosthesis, Knee, Femorotibial, Constrained, Cemented,Metal/Polymer
Legally marketed device(s) towhich equivalence is claimedPrimary Predicate Device:K896048 – S-ROM/Noiles Mark 3Secondary Predicate Device:K091453 - DePuy LPS Universal Hinge Insert AssemblyReference Predicate Device:K160700 – ATTUNE™ Revision Knee SystemK983014 – Summit Acetabular System
Reason for 510(k)submissionPursuant to Section 510(k) of the Federal Food, Drug and Cosmetic Act, andin accordance with the Medical Device Amendments of 1976 and Subpart Eof Part 807, Title 21 of the Code of Federal Regulations and as per the FDAGuidance Format for Traditional and Abbreviated 510(k)s: Guidance forIndustry and Food and Drug Administration Staff (2019), DePuy Ireland UChas compiled a Traditional 510(k) Premarket Notification to obtain marketclearance for the ATTUNE Revision Hinge Knee
Device descriptionThe ATTUNE™ Revision Hinge Knee, comprised of a femoral, housing assembly, and tibial insert component, is compatible with devices of the ATTUNE Knee System and the ATTUNE Revision Knee System and is designed to replace the natural articulating surface of the knee joint in total knee arthroplasty. The femoral component is a metal/ultra-high molecular weight polyethylene (UHMWPE) implant intended for cemented use. The housing assembly component is composed of both metal and antioxidant (AOX) UHMWPE subcomponents. The tibial insert component is composed of AOX UHMWPE.
Intended use of the deviceThe ATTUNE Revision Knee Hinge is intended to provide a functional articulating knee joint in skeletally mature patients with a severely painful knee and impaired knee function and/or after a failed previous implant.
Indications for useThe ATTUNE Revision Hinge Knee is indicated in cases for cement use in patients who have reached skeletal maturity and for whom the surgeon has decided to resect both cruciate ligaments due to the following conditions:1. Severe instability, gross deformity and/or bone loss2. Failure of a previous knee reconstruction procedure.3. Trauma or tumor resection4. Absent or markedly insufficient collateral ligaments

{5}------------------------------------------------

{6}------------------------------------------------

SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

The ATTUNE Revision Hinge Knee is substantially equivalent to its primary predicate S-ROM/Noiles Mark 3 (K89048) and secondary predicate LPS Universal Hinge Insert (K091453). The primary and secondary predicates must be used together to form a rotating hinge knee construct formed by the components of the ATTUNE Revision Hinge Knee. In addition, the subject ATTUNE Revision Hinge Knee and a reference predicate, the ATTUNE Revision Knee System (K (60700) are very sinilar in devign features. The reference predicate device ATTUNE Revision Knee System is not a rotating hinge knee construct but, is the compatible knee system for the subject device.

CharacteristicsSubject Device:ATTUNE Revision Hinge KneePrimary Predicate Device:S-ROM/Noiles Mark 3K896048Secondary Predicate Device:LPS Universal Hinge InsertK091453Reference Predicate Device:ATTUNE Revision Knee SystemK160700
Indications forUseThe ATTUNE Revision HingeKnee is indicated in cases forcement use in patients who havereached skeletal maturity and forwhom the surgeon has decided toresect both cruciate ligamentsdue to the following conditions:1. Severe instability, grossdeformity and/or bone loss2. Failure of a previous kneereconstruction procedure.3. Trauma or tumor resection4. Absent or markedlyinsufficient collateralligamentsThe Noiles (Mark 3) RotatingHinge Knee System is indicatedfor severe instability, grossdeformity and/or bone lossresulting from:1. failure of a previousprosthesis,2. trauma or tumorresection,3. absent or markedlyinsufficient collateralligamentsThe Noiles (Mark 3) RotatingHinge Knee System is intended forcement use.The DePuy LPS System isintended for use in replacement ofthe mid-shaft portion of the femur,proximal, distal and/or total femur,and proximal tibia, especially incases that require extensiveresection and replacement.Specific diagnostic indications foruse include:1. malignant tumors (e.g.,osteosarcomas,chondrosarcomas, giant celltumors, bone tumors)requiring extensive resectionand replacement:2. patient conditions ofnoninflammatorydegenerative joint disease(NIDJD), e.g. avascularnecrosis, osteoarthritis, andinflammatory joint disease(IJD), e.g., rheumatoidarthritis, requiring extensiveresection and replacement;3. revision cases for a failedprevious prosthesis requiringextensive resection andreplacement;4. severe trauma requiringextensive resection andreplacementCandidates for total kneereplacement include patients with• A severely painful and/orseverely disabled jointresulting from osteoarthritis,post-traumatic arthritis, orrheumatoid arthritis.• Moderate valgus, varus, orflexion deformities• Avascular necrosis of thefemoral condyle• A previous unsuccessfulknee replacement,osteotomy, or other kneeprocedureATTUNE Revision Knee Systemimplants are designed for use intotal knee arthroplasty forpatients with:• Absence or loss of bothcruciate ligaments• Moderate varus-valgus orflexion instability thatrequires a bearing surfacewith increased constraint inthe clinical judgment of thesurgeon.• Bone loss that requiressupplemental fixation in the
Material
Cobalt Chrome alloy, AOXUHMWPECobalt Chrome alloyUHMWPE, Cobalt Chrome alloyCobalt Chrome alloy, AOXUHMWPE
Fixation
CementedCementedN/ACemented, Cementless
Femur Sizes
3 through 8 (Lefts & Rights)XSmall, Small, Medium (Lefts & Rights)N/A1 through 10 (Lefts & Rights)
Tibia Insert Sizes
Sizes: 3 through 8Thicknesses (mm) 8 through 26(increasing by 2mm)N/ASizes: XSmall, Small, MediumThickness (mm):12,14,16,18,21,23,26,28,31Sizes: 1 through 10Thicknesses (mm): 6 through 26(increasing by 2mm)
HousingAssembly
Same as PredicatesAxel and Hinge Post AssemblyAxel and Hinge Post AssemblyN/A
SterilityAssurance Level
10-610-610-610-6
Sterile Method
Gamma RadiationGamma RadiationGamma RadiationGamma Radiation
SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
No clinical tests were conducted to demonstrate substantial equivalence.

{7}------------------------------------------------

{8}------------------------------------------------

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA

The following tests were performed in support of the to demonstrate substantial equivalence of safety and efficacy with the predicate devices:

  • Shear Testing •
  • Fatigue Testing ●
  • Wear Performance ●
  • Stability Testing .
  • Contact Pressure Testing .
  • Biocompatibility .
  • MRI Safety Evaluation Testing •

The ATTUNE Revision Hinge Knee is substant to is primary predicate S-ROMNoiles Mark 3 (K80048) and secondary predicate LPS Universal Hinge Insert (K091453).

§ 888.3510 Knee joint femorotibial metal/polymer constrained cemented prosthesis.

(a)
Identification. A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.